You are on page 1of 12

1118 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 53, no.

6, june 2006

A New Formalism for the Quantification of


Tissue Perfusion by the
Destruction-Replenishment Method in
Contrast Ultrasound Imaging
Marcel Arditi, Member, IEEE, Peter J. A. Frinking, Member, IEEE, Xiang Zhou,
and Nicolas G. Rognin, Member, IEEE

Abstract—A new formalism is presented for the time of measurement. These authors observed that moni-
destruction-replenishment perfusion quantification ap- toring the UCA echo levels in intermittent imaging mode,
proach at low mechanical index. On the basis of physi-
at moderately high mechanical index (MI) resulting in par-
cal considerations, best-fit methods should be applied us-
ing perfusion functions with S-shape characteristics. These tial destruction of microbubbles, allowed them to establish
functions are first described for the case of a geometry with a correspondence between the values of “video intensity”
a single flow velocity, then extended to the case of vascular versus pulsing interval and local flow parameters. With
beds with blood vessels having multiple flow velocity val- the advent of sensitive, contrast-specific imaging modes
ues and directions. The principles guiding the analysis are, such as B-mode pulse-inversion, power pulse inversion, or
on one hand, a linearization of video echo signals to over-
come the log-compression of the imaging instrument, and,
contrast pulse sequencing [2], other authors noted that a
on the other hand, the spatial distribution of the transmit- similar correspondence could be established by using a low
receive ultrasound beam in the elevation direction. An in MI in the replenishment phase, and real-time imaging (i.e.,
vitro model also is described; it was used to confirm ex- at rates of typically 8 to 25 frames per second), thus sim-
perimentally the validity of the approach using a commer- plifying considerably the method [3].
cial contrast agent. The approach was implemented in the
form of a computer program, taking as input a sequence Published literature has consistently reported using a
of contrast-specific images, as well as parameters related best-fit of the replenishment signal with an expression de-
to the ultrasound imaging equipment used. The generated scribing the dilution kinetics of an indicator in a single-
output is either flow-parameter values computed in regions- compartment volume, often referred to as the growing mo-
of-interest, or parametric flow-images (e.g., mean velocity,
noexponential function, or sometimes as the “single-phase
mean transit time, mean flow, flow variance, or skewness).
This approach thus establishes a base for extracting infor- exponential association equation” [1], [3]–[6]. These heuris-
mation about the morphology of vascular beds in vivo, and tic approaches were based on analyses of the video grey
could allow absolute quantification provided that appropri- levels of ultrasound imaging instruments, which system-
ate instrument calibration is implemented. atically undergo nonlinear compression (also called log-
compression), before they are displayed on a video mon-
itor. Fitting these video data with the monoexponential
I. Introduction function thus allowed producing flow-estimates related to
the actual local organ perfusion, which were then judged
new formalism is presented for the quantification of
A blood perfusion of organs, based on the analysis of
the replenishment kinetics following the destruction of a
satisfactory.
However, there is no physical basis to model the re-
plenishment signal as a monoexponential function: the de-
microbubble-based ultrasound contrast agent (UCA) in tection of the arrival of microbubbles within the imaging
the imaging plane. Perfusion parameters are derived from a plane following destruction is a process that differs radi-
best-fit of a parametric function of time with echo-derived cally from the process of dilution of an indicator within a
signals during the replenishment phase. This method is single compartment. Instead, it is governed by the imaging
mostly useful in assessing in vivo perfusion in the microvas- probe diffraction properties in the elevation direction, in
culature, in which velocities are too small to be measured relation with the steady UCA-concentration replenishment
by Doppler techniques. This approach was first described as a function of time and in the vicinity of the imaging
by Wei et al. in 1998 [1]. Locally destroying microbubbles plane following destruction. Recently, Lucidarme et al. [7],
essentially served the purpose of providing what may be noting a marked discrepancy between observed S-shape re-
considered as a “negative-bolus” of agent in an organ oth- plenishment echo-signals as a function of time and the mo-
erwise under an essentially constant perfusion during the noexponential model, proposed a complex, time-delayed,
multicompartment model to account for possible destruc-
Manuscript received July 21, 2005; accepted December 12, 2005. tion inducing progressively delayed UCA arrival at indi-
The authors are with Bracco Research SA, Ultrasound
Physics, Plan-les-Ouates, Geneva, Switzerland (e-mail: vidual locations within the image plane. Our suggestion is
marcel.arditi@brg.bracco.com). that, although this effect may be real and playing a signif-

0885–3010/$20.00 
c 2006 IEEE
arditi et al.: tissue perfusion by destruction-replenishment method 1119

icant role in the determination of the replenishment func- portionality can be obtained by signal linearization. When
tion in some cases, it is not the fundamental reason for dealing with log-compressed video signals, such lineariza-
explaining the S-shape replenishment characteristic: our tion can be obtained by reverting the compression law and
explanation is actually considerably simpler, as will be ex- subsequent squaring. When dealing with radio frequency
posed hereafter. Other authors [8], [9], noting the sequen- (rf) or demodulated echo signals, such linearization can be
tial arrival of UCA in B-mode images, caused by the exis- obtained by squaring these signals. In both cases, propor-
tence of various flow velocities and entry angles within the tionality between signal amplitude (i.e., echo power) and
image plane, developed a formalism for the replenishment, UCA concentration is established. (This proportionality is
also ignoring the effects of probe transmit (Tx) and receive verified at constant depths, and in the situation in which
(Rx) diffraction properties in the determination of the re- attenuation of the ultrasound beam due to the UCA is
plenishment function. In their work published in 2004, Pot- negligible).
devin et al. [10] correctly recognized the influence of the
acoustical beam point-spread function on the replenish- B. Linear and Phased-Array Geometries
ment function, and stressed the importance of performing
the perfusion analysis on acoustical intensity signals. They The acoustic transmit-receive sensitivity distribution in
also pointed at the possibility of associating the shape of elevation is discussed because it constitutes the basis of
the replenishment function with perfusion patterns typical the present approach. As is well-known from the classical
of healthy or diseased tissue. However, these authors did diffraction theory [12], the acoustic pressure distribution in
not develop a complete formalism for deriving physically the elevation direction y, in the field of a focusing aperture
meaningful perfusion parameters. The object of this pa- with rectangular geometry, excited in a continuous-wave
per is to present such a formalism, based on fundamental mode, is approximately determined by the function:
physical principles. p(y) ∼
= Γ · sinc(KT x y), (1)
Ignoring the ultrasound equipment properties and set-
tings, such as receiver gain, log-compression, Tx and Rx where the “sinc” function is defined as sinc(q) ≡
focusing, and so on, perfusion estimates remain user- (sin(πq)/πq) and KT x is the transmit beam diffraction
and instrument-dependent. With the new formalism in- parameter in [m−1 ]. All symbols used in this paper are
troduced here, it will be shown how user- and instrument- defined in Table I.
independent relative flow estimates may be achieved. With In the case of pulsed wave excitation, as used in echo-
appropriate probe calibration, this formalism also paves graphic imaging modes, the main lobe of the peak-pressure
the way for obtaining absolute perfusion parameters, pro- distribution is in close agreement with the continuous wave
vided that proper beam, destruction zone, and attenuation case at a frequency near the center (or mean) frequency of
calibrations are implemented. the excitation waveform.
Following a brief reminder of basic beam-diffraction The transmit-receive sensitivity distribution, expressed
considerations, in the case of linear- or phased-array aper- in terms of echo-signal power, can be determined by the
tures with rectangular geometries, the new formalism for combined effects of the transmit and receive distributions,
the replenishment signal will be presented for the simple which, in general, may be different. In transmit, the nor-
and purely theoretical case of single-flow velocities or tran- malized acoustic power distribution PT x (y) in the eleva-
sit times. It will then be extended to the more realistic case tion direction, in the field of an ultrasound transducer,
of flow with multiple transit times, known to prevail in is approximately determined by the square of the pressure
tissue perfusion. Experimental verification of the new for- distribution given above. For a rectangular aperture, it can
malism will be outlined, before exposing its applicability thus be expressed by a function of the form:
within perfusion quantification software.
PT x (y) ∼
= sinc2 (KT x y). (2)
In practice, and for the purpose of quantifying reperfu-
II. Beam Considerations sion parameters, this power distribution can be approxi-
mated, in the main lobe of a focused aperture with rect-
A. Preamble: Linearity angular geometry, by a Gaussian-shaped function G(YT x ),
defined as:
When addressing the analysis of echo signals arising
G(YT x ) = e−(1.94·YT x ) ,
2

from reflections by UCA microbubbles, it is useful to re- (3)


member that, as long as the backscattered ultrasound where YT x ≡ KT x y is a unitless quantity.
energy originates from an ensemble of randomly spaced The close correspondence of G(YT x ) with sinc2 (YT x ) is
and statistically uncorrelated scatterers, the instantaneous illustrated in Fig. 1(a) for YT x values ranging from −1 to
echo power is a quantity proportional to the local concen- +1.
tration of microbubbles [11]. In the context of perfusion In the receive mode, a similar sensitivity distribution is
quantification, it is thus important to preserve that pro- given by:
portionality when interpreting parametric values related
G(YRx ) = e−(1.94·YRx ) ,
2
to the UCA concentration as a function of time. This pro- (4)
1120 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 53, no. 6, june 2006

TABLE I
Symbols Used.

SI
Symbol Quantity unit
x space coordinate within image plane, m
(lateral)
direction orthogonal to beam
y space coordinate across image plane, m
(elevation)
direction orthogonal to beam
z space coordinate within image plane, m
(depth)
direction along the beam
p, q dummy, generic variable —
(A)
p(y) ultrasound pressure distribution Pa
E ultrasound echo power W
P, P E, G unitless power distribution functions —
D special extent of microbubble destruction, m
in elevation (y) direction
f ultrasound frequency Hz
c speed of sound m s−1
λ ultrasound wavelength m
τ flow transit time from edge to center of s
destroyed zone
m mean of the natural logarithm —
s standard deviation of the natural —
logarithm
τmean mean of the lognormal distribution of s
transit times
vmean mean of the lognormal distribution of m s−1
velocities
σ2 variance of the lognormal distribution —
γ skewness of the lognormal distribution —
PD probability density or relative concentra- m−1
tion of microbubbles
A steady-state amplitude —
a half-aperture width in y direction m (B)
v local flow velocity (vx , vy , vz components) m s−1
KT x , KRx parameters on transmit and receive, re- m−1 Fig. 1. Gaussian beam approximation and its normalized integral.
spectively 2a/λz (A) Transmit power-sensitivity profile in elevation, sinc2 (YT x ) and
K transmit-receive parameter determined m−1 Gaussian G(YT x ). (B) Gaussian transmit-receive power sensitiv-
by K 2 = KT2 x + KRx2
ity profile GP E (Y ) in elevation and cumulative normal distribu-
Y unitless variables Y ≡ K y — tion E(Y ).
θ angle between flow direction and normal —
to image plane
β “velocity” term of the monoexponential s−1
function where YRx is defined, in analogy with the transmit case, by
t time s YRx ≡ KRx y and computed considering the receive focus-
Γ arbitrary proportionality constant — ing aperture. In harmonic modes often used for the specific
GL grey level value — detection of UCA echoes, the frequency to be considered
V, Vmax echo amplitude and maximum echo —
amplitude
in KRx is, accordingly, the second-harmonic frequency.
LC dynamic range of log compression dB In the pulse-echo case, the power sensitivity of the ul-
trasound transducer to off-axis targets in the elevation di-
rection may, in a first approximation, be determined by
the product of the transmit and receive distributions. The
transmit-receive power sensitivity-pattern thus can be ap-
proximated by a Gaussian function GP E (Y ) as:

GP E (Y ) = e−(1.94·Y ) ,
2
(5)
arditi et al.: tissue perfusion by destruction-replenishment method 1121

with Y ≡ Ky and K being a transmit-receive parameter to a clinician for assessing local tissue pathologies. The fol-
as defined in Table I. lowing two sections address the cases of single and multiple
flow values, successively.
C. Other Beam Geometries or Imaging Methods

In the case of transducer geometries different from rect- III. Single Flow Values
angular apertures—such as apodized rectangular, sparse
array, synthetic array, annular array, or axicon focusing— A flow direction perpendicular to the imaging plane is
the spatial distribution still can be approximated by a first assumed. As UCA microbubbles replenish the slice
Gaussian-shape function in the elevation main lobe. For volume following destruction, the linearized echo signal
any other geometry, actual values of KT x and KRx may be amplitude is determined by the growing proportion of mi-
determined, either from theoretical considerations or ex- crobubbles re-entering that volume, and weighted by the
perimentally, by measuring actual transmit-receive beam transmit-receive beam sensitivity pattern. For a uniform
sensitivity patterns, from the best-fit of Gaussian func- concentration of UCA microbubbles of unity value, having
tions. re-entered the slice until position y ≡ Y /K, the normal-
ized echo power signal E(Y ), resulting from the detection
with the beam sensitivity GP E (Y ) then can be expressed
D. Replenishment Kinetics
as the integral:

As mentioned previously, most prior investigators have
∼ 1.94 Y
based their perfusion quantification approach on the analy- E(Y ) = √ GP E (Y  )dY  . (7)
π −∞
sis of the video signal observed during UCA replenishment
following the destruction of microbubbles in a volume con- The normalized echo power signal E(Y ) can simply be
taining the tomographic image plane. Basing their anal- found with the aid of the error function erf (q), defined
yses on log-compressed, echo-signal data, these authors as [13]:
chose, as a heuristic model of replenishment kinetics, a  q
2
e−p dp.
2
monoexponential function describing the video grey level erf (q) = √ (8)
π 0
values as a function of time, GL(t), according to:
The error function verifies: erf (0) = 0, erf (−q) =
GL(t) = A(1 − e−βt ), (6) −erf (q) and lim erf (q) = 1.
q→∞

where the time origin is taken at the instant immediately The physical situation of microbubbles replenishing the
following the last UCA destruction pulses. The values A, β, destroyed slice is equivalent to assigning −∞ to the lower
and Aβ have commonly been interpreted as quantities pro- integration limit of (8). The normalized echo power signal
portional to “blood volume,” “blood velocity,” and “blood is thus determined by the cumulative normal distribution
flow” within the analyzed region. Considering the nonlin- function, renamed in short perf in the present context of
ear gain applied to the echo signals, it may be expected perfusion estimates. This perf function is defined as:
 q
that such analysis of the contrast-replenishment echo sig- 1
e−p dp,
2

nals may be influenced by experimental parameters such perf (q) ≡ √ (9)


π −∞
as log-compression, receive gain, tissue attenuation, UCA
concentration, and other instrument settings. which verifies the following properties: lim perf (q) = 0,
q→−∞
In contrast, the formalism exposed in this paper is perf (0) = 0.5, and lim perf (q) = 1. These properties
q→∞
founded on the well-established relationship between the
are equivalent to noting that half of the steady-state echo
observed echo-power as a function of time and the instan-
power is achieved when the bubbles reach the beam axis
taneous UCA concentration. In this way, the actual UCA
from either side, and steady state is achieved after re-
replenishment kinetics is estimated by a best-fit approach
plenishment of the whole destroyed section. Furthermore,
based on the spatial transmit-receive sensitivity distribu-
perf (q) can be expressed simply in terms of the error func-
tion in the pulse-echo mode along the elevation direction.
tion erf (q) as:
The replenishment signal in the echographic image is thus
determined by two factors. The first one is the re-entry of perf (q) = 0.5 · (1 + erf (q)). (10)
fresh UCA microbubbles in the region in which they were
destroyed. The second one is the transmit-receive sensitiv- With the above definitions, the normalized echo power
ity pattern of the ultrasound probe in the elevation direc- signal then may be expressed as:
tion. E(Y ) ∼
= perf (1.94 Y ). (11)
Despite the great complexity of organ perfusion (mi-
crovasculature structure, random directions of flow, super- Graphically, this is an S-shape function representing
position of different flow values, etc.), it turns out that the the cumulated echo power under the Gaussian transmit-
estimated perfusion parameters yield useful information in receive beam GP E (Y ) sensitivity profile, integrated from
the in vivo case, which may provide valuable information −∞ to Y , as illustrated in Fig. 1(b).
1122 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 53, no. 6, june 2006

It is now considered that the MI during the destruc-


tion frames is sufficiently high to destroy a majority of
the UCA bubbles in the imaging plane (this is generally
the case in destruction-replenishment UCA-quantification
studies since the highest MI possible is normally chosen).
As a consequence, a region of destroyed microbubbles of
thickness D extends symmetrically on either side of the
image plane in the elevation direction y and is delimited
by sharp edges.
Immediately following the application of the destruc-
tion frames, a flow of fresh microbubbles starts replenish-
ing the destroyed slice in the y direction at a flow velocity
v. Noting that y = vt, the normalized echo power signal
(i.e., the replenishment function) during the nondestruc-
tive monitoring phase as a function of time, E(t), is thus
also represented by a perf function, now expressed as:
  
D
E(t) = A · perf 1.94 · Kv t − , (12) Fig. 2. Replenishment function E(t), in percent, following destruction
2v frames, at four flow velocities. The time origin is at the end of the
destruction frames.
with a time origin at the instant immediately following the
last destruction frame. By introducing the definition of a
transit time, τ , as the time required to replenish half the such that the resulting distribution of transit times follows
bubble destruction slice: a lognormal probability density distribution. This proba-
D bility density function, P D(τ ), is commonly expressed as:
τ≡ , (13)
2v
[ln(τ ) − m]2

the replenishment function E(t) becomes: e 2s2
  P D(τ ) = √ , (15)
KD τ s 2π
E(t) = A · perf 1.94 · (t − τ ) . (14)

where, in this particular case, m and s2 are the mean and
In the general case, when the direction of replenishment variance of the distribution of the natural logarithm of
is at an arbitrary angle θ with respect to the y direction, v transit times, τ , respectively [16]. The normalization of
must be replaced by vθ = v cos θ in (12) and (13). However, the lognormal distribution function is such that:
when the perf function is evaluated in terms of τ (14), the  ∞
angle dependence is without object, following the defini- P D(τ )dτ = 1. (16)
tions of τ and D as given above. 0
Thus, the replenishment function for a single flow value
has an initial “heel”, corresponding to the beginning of The lognormal distribution of transit times is character-
the reperfusion process when the microbubbles begin to ized by its mean τmean , variance σ2 and skewness γ, de-
be detected by the edge of the elevation beam. A central, fined by:
fairly linear portion corresponds to fresh bubbles crossing s2
the mid-plane of the imaged volume. A final “shoulder” τmean = em+ 2 ,
portion, approaching the steady-state value of complete s2 +2m s2
reperfusion, corresponds to fresh bubbles finishing to re-
2
σ =e (e − 1), and (17)
  2

plenish the imaged volume in the elevation direction. This γ = es2 − 1 2 + es .
behavior is illustrated in Fig. 2, for an example with a du-
ration of destruction frames of 1 s, A = 100, D = 8 mm, The probability density function in (15) can also can
K = 0.276 mm−1 , θ = 0◦ and different values of v at 0.5, be expressed, as an alternative, in terms of flow velocity v,
1, 2, and 4 mm s−1 , respectively. instead of transit time τ . In the Appendix, a law of trans-
formation is derived for converting (15) into a lognormal
distribution expressed in flow velocity v.
IV. Multiple Flow Values
In the presence of multiple flow transit times, the
A. Lognormal Distribution of Transit Times replenishment function E(t) may, in this case, be ex-
pressed as:
As mentioned by, for example, Karshafian et al. [4], [14]  ∞  
KD
or Qian and Bassingthwaighte [15], the vascular arrange- E(t) = A P D(τ )perf 1.94 · (t − τ ) dτ.
ment of blood vessels in most regularly perfused organs is 0 2τ
(18)
arditi et al.: tissue perfusion by destruction-replenishment method 1123

Fig. 4. Replenishment functions corresponding to the four distribu-


Fig. 3. Lognormal distribution of transmit times τ with m = 0.835 tions τ1 to τ4 of Fig. 3, as well as the overall replenishment function
and s = 0.726, corresponding to τmean = 3 s and σ = 2.5, decom- determined by the sum of the four contributions.
posed into four distribution segments τ1 to τ4 , separated at τ = 2,
4, and 6 s, respectively.
resented by lognormal probability distributions of transit
times. In these cases, and as discussed by Karshafian et
In this context, E(t) is named the lognormalperf replen- al. [4], [14], the analyses of the replenishment functions by
ishment function, and it is expressed in terms of elemen-
best-fit approaches outline differences in skewness of the
tary perf functions, weighted by the probability of density
lognormal probability distribution, which then may bear
distribution P D(τ ). diagnostic significance in the assessment of their state of
In Fig. 3, an example of a lognormal probability dis- perfusion.
tribution of transit times with a mean at τmean = 3 s is
shown, considering four arbitrary transit time intervals,
between 0 and 2 s, 2 and 4 s, 4 and 6 s, and above 6 s. The V. Experimental Verification
four corresponding contributions of these subranges to the
overall replenishment function E(t) are illustrated in Fig. 4 A. Objective
as the numbered curves Eτ1 (t) to Eτ4 (t), as obtained by in-
tegration of (18) in the corresponding τ intervals. It may A controlled-flow system was assembled to be imaged
be observed that the individual replenishment functions in a contrast-specific ultrasound imaging mode. By ap-
of the four relatively narrow distributions have an appear- plying the destruction-replenishment technique, the im-
ance close to the S-shape perf function of a single-flow age database collected was analyzed in two ways: one us-
value. In contrast, the overall replenishment function dis- ing the monoexponential replenishment model function on
plays a distorted or skewed appearance, with a more rapid the video signals, and the other using the new approach
“heel” at the onset of replenishment and slower “shoulder” as described above. The objective was to assess how re-
near the end of the replenishment. It reflects the contri- liable each approach was while outlining their respective
butions of the various transit times, each with their own advantages and limitations. In particular, the same mea-
weights as determined by the relatively wide probability surements were repeated at two different receive-gains and
distribution function. log-compression settings in the imaging instrument, to as-
sess the instrument-dependence of the techniques.
B. Other Distributions of Transit Times
B. Materials and Method

If flow quantification is to be applied to tissues that A variation on the model described by Veltmann et al.
are believed not to obey lognormal distributions of tran- [17] was used to establish stable and reproducible flow
sit times, but are better described by another probabil- conditions to be imaged by a commercially available ul-
ity distribution (e.g., first passage, random walk, gamma trasound scanner. The general setup is shown in Fig. 5. A
variate, or arbitrary distributions), (15) then may be ex- bundle of 170 microfibers with 240-µm lumen diameter and
pressed by using this other distribution for P D(τ ). There 18-cm length (AN69 type, Hospal SA, Lyon, France) was
is, however, strong evidence that the perfusion of most modified by removing its clear polymethyl methacrylate
physiological tissues, including abnormal structures such (PMMA) tubing over half its circumference over a couple
as neoplasms or neovasculature in tumors, are well rep- of centimeters in length. In this way, the microfibers were
1124 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 53, no. 6, june 2006

values of log-compression and two relative receiver-gain


values (0 and −10 dB), in order to simulate the effects
of different user choices of instrument settings. The log-
compression function, expressed as a function of the input
echo amplitude variable V , was:
 
20 dB V LC
GL(V ) = 255 log10 10 20 dB ,
LC Vmax
(19)
Fig. 5. Double-flow circuit (HI and LO), modified from Veltmann
et al. [17], allowing a stable agent concentration in the microfiber- where Vmax was chosen as the maximum echo amplitude,
bundle at all flow values imposed by the downstream peristaltic and kept constant for the whole data set. LC was set at
pump. values of 40 and 70 dB (using the 0 dB gain only), and
any negative values of GL were truncated to zero.
Image sequences of 8 s of UCA replenishment were an-
exposed directly, in water immersion, to the ultrasound alyzed in two different ways. In one data path, the com-
field of a linear-array probe (Esaote LA532E, 2.6/5.2 MHz pressed GL images were analyzed using an in-house video-
Tx/Rx-frequencies), coupled to an EsaTune instrument densitometric program in a region-of-interest (ROI), hand-
(Esaote, Florence, Italy). The ultrasound scanner was op- drawn to include the whole fiber bundle region, to gener-
erated at a frame rate of 43 Hz, in harmonic B-mode, at ate what was called video replenishment functions. As an
a low acoustic power (MI 0.07), with the probe placed at alternative data path, the compressed GL images were lin-
45-mm distance from the center of the fiber bundle. With earized by applying, in sequence, the inverse of (19) and
an aperture width in elevation of 5.5 mm, these imaging squaring, to obtain replenishment functions proportional
conditions (i.e., using a transmit frequency of 2.6 MHz to the instantaneous echo-power from the UCA, E(t), with
and a receive frequency of 5.2 MHz for KT x and KRx , the aid of the function E(GL):
respectively) determined a value K = 0.480 mm−1 . The
bundle was characterized as having a linear relationship  2
E(GL) = Vmax 10( 255 −1) 20 dB ,
GL LC

between the fluid volume flow ϕ in milliliters per min- (20)


utes and the mean flow velocity v in millimeters per sec-
ond by the relation ϕ = 0.46 v. While gravity feeding a with E(GL) truncated at (Vmax )2 for any values GL > 255.
HI-flow circuit with deionized water at a constant rate of The echo-power signals so computed then were averaged
270 mL/minute, SonoVueTM microbubbles (Bracco SpA, in the same ROI, effectively representing the mean square
Milan, Italy) were injected in this circuit at a constant (rms2 ) amplitude in that region, resulting in what was
rate of 0.5 mL/minute by using a VueJect infusion pump called linearized replenishment functions.
(Bracco Research SA, Geneva, Switzerland), determining The recorded data sets were analyzed using the curve
a dilution factor of 1:1620, equivalent to approximately 2.5 fitting toolbox (v.1.1.2) of Matlab. The monoexponential
105 bubbles per milliliter. A small fraction of that flow was function GL(t) of (6) and the perf function, E(t) of (14)
sucked into a LO-flow circuit from a small mixing chamber were used as fitting functions. The GL(t) function had A
(approximately 2 mL), using a MiniPuls 3 peristaltic pump and β as fitting parameters, and the E(t) function had A
(Gilson, Villiers-le-Bel, France) placed downstream from and τ . Note that the perf function model was used be-
the bundle. In this way, a UCA suspension with stable con- cause the simple geometry of the bundle of parallel fibers
centration was forced through the bundle, with nine differ- determined single flow velocity values in one direction. The
ent flow values between 0.46 and 13.8 mL/minute, corre- experimental conditions were such that there were no satu-
sponding to mean flow velocities between 1 and 30 mm s−1 . rated values in both the video and linearized data sets, and
With the imaging probe positioned to scan a cross sec- negligible offset levels. Each measurement was repeated
tion of the fiber bundle, at an angle θ of 60◦ , sequences of three times, and sets of parameters were computed and
real-time, echo rf-data were captured by a Femmina digital averaged for both functions. Estimates of the flow veloci-
interface (University of Florence, Italy [18]) coupled to the ties then were considered as:
EsaTune instrument by a fiberoptic link, at a sample rate
of 40 MS/s. The UCA destruction sequences consisted of D D
vest ≡ = , (21)
43 frames of destructive imaging (maximum acoustic out- 2τ cos 60◦ τ
put, for a duration of 1 s). Three recordings were made at
each of nine flow values. The rf-data sequences then were for the perf function, and as β for the monoexponential
demodulated using the Hilbert transform, scan-converted, function. The value of D was found to be D = 7.2±0.1 mm.
log-compressed and stored as 8-bit grey scale TIF im- It was measured in a separate experiment by optical ob-
ages with 256 possible grey levels (GL), using an in-house servation of the extent of destroyed bubbles embedded in
program written in Matlab (The Mathworks Inc., Natick, an agarose gel, exposed at the same distance as the fibers,
MA). Three sets of TIF images were generated, using two but perpendicular to the ultrasound imaging beam.
arditi et al.: tissue perfusion by destruction-replenishment method 1125

Fig. 6. Results of curve fits with the perf function on linearized data,
at flow velocities of 4 and 2 mm s−1 .

C. Results
Fig. 7. Flow-velocity estimates vest as a function of flow velocity
In Fig. 6, two linearized replenishment curves are between 1 and 30 mm s−1 using the perf function. The actual esti-
mated values of D (7.2 mm) and θ (60◦ ) were used in the fits. The
shown, for individual measurements performed at 4 and slope a1 of the linear regression on vest is 0.924.
2 mm s−1 mean flow velocities. Parametric perf functions
were fitted to the data, determining sets of correspond-
ing A and τ values. The S-shape characteristics of the
replenishment may be clearly recognized. The compara- and monoexponential models in terms of transit time or
tively large noise amplitude at the plateau level is caused flow-velocity estimates, in the case of a single flow direc-
by fluctuations in the instantaneous agent concentrations tion. In these conditions, the perf model is clearly able to
and speckle patterns in the ROI, magnified by the lin- provide more consistent and robust estimates of flow ve-
earization process. In Fig. 7, the values of vest are plotted locity, because it is based on a physical description of the
against the nine flow velocity values v. The linear regres- UCA replenishment process as measured by a given imag-
sion slope coefficient a1 was found to be 0.924, a result ing beam. Different gain values for the estimation of flow
in close agreement with the value of 1.0 expected with an velocity with the perf model were not considered, as this
absolute quantification method. is irrelevant: the perf analysis is intrinsically a linear pro-
In Fig. 8, replenishment functions at 8 mm s−1 are cess. Changing the gain value, or equivalently, the concen-
shown for GL amplitude data, obtained with 40 dB log- tration of UCA bubbles, results in identical flow velocity
compression, at two relative front-end gain settings of 0 dB estimates with the perf model, in contrast with the mo-
and −10 dB (applied before log-compression). The cor- noexponential model applied on video grey levels, which
responding best-fit parametric monoexponential functions produces estimates that are instrument-setting dependent.
are able to adjust well to the experimental data near the The monoexponential model also was applied on the lin-
plateau region, but the fits are poor in the early phases earized replenishment data set (results not shown). In this
of replenishment. Fig. 9 illustrates the system gain depen- case, the best-fit analysis was performed with and with-
dence as β is plotted against the nine values of v; although out a manual adjustment of the time origin to the instant
the linearity in each case may be considered as adequate, of detectability of the UCA following destruction. This op-
an overestimate of 50% on the slope a1 is made at 0 dB tional adjustment was considered in an attempt to improve
gain, compared to the slope value a2 obtained at −10 dB the quality of the fits, especially at the low-flow velocities.
gain. Similarly, Fig. 10 shows the β estimates using a fixed In both cases, the quality of the fits and the linearity of
gain of 0 dB, but at two different LC values of 40 and the β estimates with flow velocity v were still poorer than
70 dB. Here, also, estimates of the linear regression slopes when applying the monoexponential model on compressed
are significantly affected by the log-compression settings. video grey levels.
In reality, the shape of the UCA destruction zone might
D. Discussion be more complex than was considered here, e.g., with a
Gaussian or truncated Gaussian profile as described in
In the previous section, experimental results obtained [10]. In the present formalism, the value of D represents a
in a simple flow model were presented. The simplified ge- destruction zone delimited by sharp edges. In our experi-
ometry allowed comparative assessments of both the perf ence, this assumption proved to be justified, as supported
1126 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 53, no. 6, june 2006

Fig. 8. Results of curve fits with the monoexponential function at


Fig. 10. Estimates of β with the monoexponential function, as a
the flow velocity of 8 mm s−1 . Two different front-end gains of 0 dB
function of flow velocity between 1 and 30 mm s−1 , at two different
and −10 dB were used. The log-compression was set to LC = 40 dB.
log-compressions (LC) of 40 and 70 dB, with a front-end gain of 0 dB.
The estimated slopes a1 and a2 of the linear regression functions are
influenced by the choices of log compression settings.

compared to the monoexponential model. However, these


flow estimates remain relative, and not absolute, consider-
ing the fact that the A estimates remain dependent on the
actual UCA concentration, system gain and the round-trip
ultrasound attenuation between the probe and the depth
of analysis.
The improvements of the perf analysis on echo-power
data may thus be summarized as allowing possibly abso-
lute estimates of flow velocity or transit time, and rela-
tive estimates of blood-volume flow, independently of log-
compression settings of the imaging instrument. Further-
more, the lognormalperf analysis allows estimates of the
morphology of tissue vascularization in an ROI, possibly
opening new avenues of tissue characterization for tumor
staging, antiangiogenic drug efficacy assessment, or differ-
ential lesion diagnosis.
Fig. 9. Estimates of β with the monoexponential function, as a func-
tion of flow velocity between 1 and 30 mm s−1 , at different front-end
gains of 0 dB and −10 dB, at a log-compression with LC = 40 dB. VI. Application in Echo-Contrast Imaging
The slopes a1 and a2 of the linear regression functions are strongly
influenced by the choices of gain values.
A. Linearizing Video Data

by the good agreement between the perf function and the Although the experimental examples described above
experimental replenishment data. were collected using a dedicated rf-grabber, the new for-
Relative estimates of blood-volume flow have commonly malism may be implemented on video data, collected on a
been associated with the product Aβ of the monoexpo- personal computer using any video acquisition hardware.
nential model. Clearly, such products cannot physically In some cases, equipment manufacturers provide a tool
represent blood flow when applied to compressed video to generate echo data without log-compression (e.g., Q-
grey level data, because the plateau value A is not pro- lab by Philips Medical Systems, Bothell, WA). Otherwise,
portional to local UCA concentration (or regional blood linearizing the video data to generate echo power data
volume, RBV), and the estimate of β is subject to the de- may be implemented in one of several different ways. One
pendence on settings discussed above. Using the perf or way is to obtain the log-compression law from the equip-
lognormalperf analyses, the expression of A/τ or A/τmean ment manufacturer, and to apply the corresponding anti-
are improved estimates of the local blood-volume flow, log function. If that information is not available, another
arditi et al.: tissue perfusion by destruction-replenishment method 1127

option is to record images of a fixed tissue-mimicking ma- analyses carried out with the lognormalperf model func-
terial at increasing gain values, often displayed as decibels tion provided successful curve fitting and useful quantifi-
on the monitor of the echo instruments to estimate the cation information when applied to assess perfusion within
log-compression law by analyzing the resulting grey lev- user-drawn ROI in both preclinical and clinical settings.
els within individual pixels or small groups of pixels. Yet
another method consists of recording video sequences of
echo-contrast images produced at known and increasing VII. Conclusions
concentrations of UCA, in order to derive the grey level
A new formalism has been presented for the
map as a function of UCA concentration, thus allowing lin-
destruction-replenishment, perfusion-quantification ap-
earization of the data. In any case, the preferred approach
proach. Based on a physical analysis of the transmit-
is for the equipment manufacturer to provide direct access
receive imaging conditions during the UCA replenishment,
to the raw echo data (rf-signals, quadrature-demodulated
the model function to be used for a best-fit analysis
signals, linearized image data, or similar), or to incorpo-
on echo-power data should be with S-shape characteris-
rate the quantification analysis within the instrument.
tics. More precisely, this function is a linear combination
of elementary cumulative normal distribution functions
B. Practical Considerations
weighted by the probability density distribution of flow
transit times within the region being analyzed. The char-
In a clinical application of the UCA destruction-
acteristics of such an elementary function are determined
replenishment method, it can happen that the echo level
by the spatial transmit-receive sensitivity distribution in
immediately after the last destruction frame is not negli-
the elevation direction, and the extent of the zone in which
gible. This can be caused by two factors: a residual tissue
UCA microbubbles have been destroyed, also in the eleva-
echo signal, due to incomplete cancellation by the contrast-
tion direction. Using a simple in vitro flow model, it was
specific imaging mode, or a nonzero UCA echo signal, due
shown how absolute flow velocity estimates may be made
to incomplete destruction. The first situation can be han-
with knowledge of the extent of the bubble destruction
dled by subtracting baseline echo power data, recorded
zone.
before UCA administration, from the replenishment data.
Compared to previously described approaches, the new
The second situation can be handled by including D as
formalism is at least as simple to implement in a software
a fitting parameter. In this case, the estimate of abso-
program, and it has the advantage of providing perfusion
lute flow velocity remains possible, provided constraints
parameters that are independent of user settings or instru-
are considered on the acceptable values of D as a function
ment characteristics. As a counterpart, information about
of depth.
the destruction zone and beam profile is required, as well
as log-compression characteristics if raw-echo data are not
C. Parametric Imaging
available.
The formalism is applicable to the computation of
The new formalism, described above in the case of com-
perfusion-parameter values either in ROIs, or within in-
puting best-fit functions of replenishment echo data in
dividual pixels or groups of pixels for parametric imaging.
ROIs, also is applicable at the scale of individual pixels or
Furthermore, the approach presented allows the extraction
small groups of pixels, as required in the case of display-
of information about the morphological distribution of vas-
ing parametric images depicting the spatial distribution
cular beds in vivo, and thus could become a useful tool of
of flow parameters (e.g., local RBV, volume flow, veloc-
tissue characterization for tumor staging, antiangiogenic
ity, transit time, etc.). In this case, the signal-to-noise ra-
drug efficacy assessment, or differentiating pathologies.
tio of echo-power data naturally is degraded, due to large
fluctuations in signal amplitude during the replenishment
phase. However, alternative approaches are possible, and Appendix A
a detailed description of the use of the present formalism Correspondence Between Lognormal
for the purpose of constructing parametric flow images is Probability Density Distributions
in preparation and will be the object of a separate article. Expressed in τ or in v.

D. Clinical Areas of Interest A law of transformation is derived for converting a log-


normal probability density distribution expressed in terms
As mentioned by many investigators, the destruction- of a transit-time variable τ into a lognormal distribution
replenishment, perfusion-quantification technique finds expressed in flow velocity v. The reverse transformation
applications in many areas of medical contrast ultrasound. is found by reciprocity in similar terms. In the following,
The formalism described above has been implemented the angle θ formed between the flow direction and the el-
within an in-house developed software program, which has evation direction is considered to be zero. The case of an
been tested in a wide variety of pathological conditions arbitrary angle θ may be treated, as indicated in the dis-
(myocardial perfusion deficits, renal artery stenosis, liver cussion about (13) and (14), by replacing v hereafter by
disease, prostate and breast cancer, etc.). In most cases, vθ = v cos θ.
1128 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 53, no. 6, june 2006

The m and s parameters of the lognormal distribution Acknowledgment


in transit times of (15), P D(τ ), are renamed mτ and sτ ,
to be distinguished from the corresponding parameters mv The authors wish to thank E. Gaud, A. Broillet, F. Yan,
and sv of the following lognormal distribution expressed in T. Messager, and M. Schneider of Bracco Research, as well
velocity P D(v): as F. Bertora and M. Cerofolini of Esaote Biomedica, for
their precious support, inputs, and collaboration.
[ln v−mv ]2
− 2s2v
e
P D(v) = √ . (A1)
v sv 2π References
Considering that: [1] K. Wei, A. R. Jayaweera, S. Firoozan, A. Linka, D. M. Skyba,
and S. Kaul, “Quantification of myocardial blood flow with
 
D D −D ultrasound-induced destruction of microbubbles administered as
v= , ln v = ln − ln τ and dv = dτ, a constant venous infusion,” Circulation, vol. 97, pp. 473–483,
2τ 2 2τ 2 1998.
(A2)
[2] P. Rafter, P. Phillips, and M. A. Vannan, “Imaging technologies
and techniques,” Cardiol. Clin., vol. 22, pp. 181–197, 2004.
the probability distribution in v, for positive dτ incre- [3] T. Schlosser, C. Pohl, C. Veltmann, S. Lohmaier, J. Goenechea,
ments, yields: A. Ehlgen, J. Koster, D. Bimmel, S. Kuntz-Hehner, H. Becher,
and K. Tiemann, “Feasibility of the flash-replenishment concept

2 in renal tissue: Which parameters affect the assessment of the
− ln τ + ln D 2 − mv contrast replenishment?,” Ultrasound Med. Biol., vol. 27, pp.
− 937–944, 2001.
e 2s2v [4] R. Kharchakdjian, P. N. Burns, and M. Henkelman, “Fractal
P D(v)dv = √ dτ modeling of microbubble destruction-reperfusion in unresolved
τ sv 2π vessels,” in Proc. IEEE Ultrason. Symp., 2001, pp. 1669–1673.

 2
ln τ − −mv + ln D 2
[5] S.-J. Rim, H. Leong-Poi, J. R. Lindner, D. Couture, D. Elle-
− gala, H. Masson, M. Durieux, N. F. Kasse, and S. Kaul, “Quan-
2s2v (A3)
e tification of cerebral perfusion with real-time contrast-enhanced
= √ dτ ultrasound,” Circulation, vol. 104, pp. 2582–2587, 2001.
τ sv 2π [6] T. H. Murthy, P. Li, E. Locvicchio, C. Baisch, I. Dairywala,
= P D(τ )dτ, W. F. Armstrong, and M. Vannan, “Real-time myocardial blood
flow imaging in normal human beings with the use of myocardial
contrast echocardiography,” J. Amer. Soc. Echocardiogr., vol.
provided that: 14, pp. 698–705, 2001.
[7] O. Lucidarme, S. Franchi-Abella, J. M. Correas, S. L. Bridal,
[ln(τ )−mτ ]2 E. Kurtisovski, and G. Berger, “Blood flow quantification with
− 2s2τ contrast-enhanced US: Entrance in the section phenomenon—
e
P D(τ ) = √ , (A4) phantom and rabbit study,” Radiology, vol. 228, no. 2, pp. 473–
τ sτ 2π 479, 2003.
[8] M. Krix, F. Kiessling, N. Farhan, K. Schmidt, J. Hoffend, and
with: S. Delorme, “A multivessel model describing replenishment ki-
netics of ultrasound contrast agent for quantification of tissue
  perfusion,” Ultrasound Med. Biol., vol. 29, no. 10, pp. 1421–
D 1430, 2003.
mτ = −mv + ln and sτ = sv ≡ s.
2 [9] M. Krix, C. Plathow, F. Kiessling, F. Herth, A. Karcher, M. Es-
(A5) sig, H. Schmitteckert, H. U. Kauczor, and S. Delorme, “Quan-
tification of perfusion of liver tissue and metastases using a mul-
The correspondences between the mean of the lognor- tivessel model for replenishment kinetics of ultrasound contrast
agents,” Ultrasound Med. Biol., vol. 30, no. 10, pp. 1355–1363,
mal distributions are found with the definitions: 2004.
[10] T. C. Potdevin, J. B. Fowlkes, A. P. Moskalik, and P. L. Car-
s2 s2 son, “Analysis of refill curve shape in ultrasound contrast agent
vmean = emv + 2 and τmean = emτ + 2 , studies,” Med. Phys., vol. 31, no. 3, pp. 623–632, 2004.
(A6) [11] N. De Jong and L. Hoff, “Ultrasound scattering of Albunex mi-
crospheres,” Ultrasonics, vol. 31, no. 3, pp. 175–181, 1993.
[12] L. E. Kinsler, A. R. Frey, A. B. Coppens, and J. V. Sanders,
which yield: Fundamentals of Acoustics. New York: Wiley, 1982.
    [13] W. Gautschi, “Error function and fresnel integrals,” in Handbook
D s2 s2 of Mathematical Functions. M. Abramowitz and I. A. Stegun,
−mv +ln 2 + 2 D − mv + 2 +s2
τmean = e = e . Eds. New York: Dover, 1972, pp. 297–329.
2 [14] R. Karshafian, P. N. Burns, and M. R. Henkelman, “Transit time
(A7) kinetics in ordered and disordered vascular trees,” Phys. Med.
Biol., vol. 48, no. 19, pp. 3225–3237, 2003.
and: [15] H. Qian and A. Bassingthwaighte, “A class of flow bifurcation
models with lognormal distribution and fractal dispersion,” J.
D 2
Theoretical Biol., vol. 205, pp. 261–268, 2000.
τmean = es , (A8) [16] D. Zwillinger and S. Kokoska, “Continuous Probability Distri-
2vmean butions,” in CRC Standard Probability and Statistics Tables and
Formulae. Boca Raton, FL: Chapman and Hall/CRC, 2000, pp.
which, together with the relations in (A5), provides the 143–144.
[17] C. Veltmann, S. Lohmaier, T. Schlosser, S. Shai, A. Ehlgen, C.
rules for converting the probability density distributions Pohl, H. Becher, and K. Tiemann, “On the design of a capillary
from τ to v and conversely. flow phantom for the evaluation of ultrasound contrast agents
arditi et al.: tissue perfusion by destruction-replenishment method 1129

at very low flow velocities,” Ultrasound Med. Biol., vol. 28, no. Currently, he is working for Bracco Research SA, Geneva, Switzer-
5, pp. 625–634, 2002. land, where he is involved in microbubble physics and contrast ul-
[18] M. Scabia, E. Biagi, and L. Masotti, “Hardware and software trasound imaging. He is working on advancing the understanding
platform for real-time processing and visualization of echo- of the interaction of ultrasound beams and microbubbles, promot-
graphic radiofrequency signals,” IEEE Trans. Ultrason., Fer- ing novel detection schemes and exploring new applications of ul-
roelect., Freq. Contr., vol. 49, no. 10, pp. 1444–1452, 2002. trasound contrast agents like molecular imaging, sonoporation, and
ultrasound-directed drug delivery.

Marcel Arditi (M’82) was born in Alexan- Xiang Zhou was born in Wuhan, P.R. China,
dria, Egypt, in 1951. He graduated in 1975 in 1972. He obtained his M.D. degree in the
with a M.Sc. degree in physics at the Univer- Tongji Medical College, Huazhong Science
sity of Geneva, Switzerland. He then obtained and Technology University, Wuhan, Hubei,
his Ph.D. degree in 1982 at the Department P.R. China, in 2000. His M.D. thesis fo-
of Medical Biophysics, University of Toronto, cused on perfusion quantification and func-
Toronto, ON, Canada, specializing in several tional imaging studies on kidney tumors, us-
aspects of medical ultrasound imaging (annu- ing ultrasound contrast agents. In 2002, he
lar array transducers, beam forming, and sig- worked as a postdoctoral fellow at Bracco
nal processing). Research SA, Geneva, Switzerland, where he
From 1982 to 1985, he was with SRI In- contributed to studies on perfusion quantifi-
ternational in Menlo Park, CA, where he con- cation.
tributed to research projects for various industrial and governmental In 2003 and 2004, he was a clinical fellow at the Friedrich-
organizations, still in the ultrasound imaging field. In 1985, he joined Alexander-Universität Erlangen-Nürnberg, Germany, where he stud-
Battelle-Europe in Geneva, Switzerland, and managed contract re- ied the hepatic transit time of ultrasound contrast agents for the
search projects for corporations in the optical and automotive indus- evaluation of liver tumor radiofrequency ablation.
tries, in fields related to image and signal processing. Since 1989, he Dr. Zhou is now a clinician at the Union Hospital, Tongji Medical
has been associated with Bracco Research SA, also in Geneva, where College, Huazhong University of Science and Technology (HUST),
he contributed to the acoustical specifications and characterization Wuhan, Hubei, P.R. China.
of novel microbubble-based contrast agents for medical ultrasound.
His current interests focus on the quest for new contrast-specific
detection methods and for making medical ultrasound a reliable
quantitative modality.
Nicolas G. Rognin (M’05) was born in Vo-
iron, France, in 1974. He received his M.Sc.
degree in signal, image, and speech process-
ing from the National Institute of Applied
Peter J. A. Frinking (M’02) was born in Sciences, INSA-Lyon, France, in 1998, with a
Bussum, The Netherlands, in 1966. After sec- thesis on brain perfusion using magnetic res-
ondary school, he received his M.Sc. degree onance imaging.
in applied physics from the Delft Univer- He completed his Ph.D. degree in 2002 at
sity of Technology, Delft, The Netherlands, in Creatis (Centre de Recherche et d’Applica-
1995. He received his Ph.D. degree in medical tions en Traitement de l’Image et du Signal),
biophysics at the Thoraxcentre of the Eras- Lyon, France, a unit of the National Centre
mus University Rotterdam, Rotterdam, The for Scientific Research (CNRS), focusing on
Netherlands, in 2000. During his stay in Rot- simulation and quantification tools in ultrasound contrast imaging.
terdam, he worked on the acoustic character- From 2000 to 2002, he had a teaching position in computer sciences at
ization of new ultrasound contrast agents, de- Claude Bernard University, Lyon, France. Since 2003, Dr. Rognin has
veloping novel detection methods for contrast been with Bracco Research SA, Geneva, Switzerland. His research in-
ultrasound imaging, and he did preliminary work on ultrasound me- terests include ultrasound contrast imaging, parametric imaging, and
diated drug delivery. image registration.

You might also like