Professional Documents
Culture Documents
art ic l e i nf o a b s t r a c t
Article history: Glycated proteins, such as glycated hemoglobin (HbA1c) or glycated albumin (GA) in the blood, are es-
Received 16 June 2016 sential indicators of glycemic control for diabetes mellitus. Since GA, compared to HbA1c, is more sen-
Received in revised form sitive to short term changes in glycemic levels, GA is expected to be used as an alternative or together
1 August 2016
with HbA1c as a surrogate marker indicator for glycemic control. In this paper we report the develop-
Accepted 2 August 2016
ment of a sensing system for measuring GA by combining an enzyme analysis method, which is already
used in clinical practice, with electrochemical principles. We used fructosyl amino acid oxidase, hex-
Keywords: aammineruthenium(III) chloride as the electron mediator, and an inexpensive and economically at-
Diabetes mellitus tractive screen-printed carbon electrode. We used chronoamperometry to measure protease-digested GA
Glycated albumin
samples. The developed sensor strips were able to measure protease-digested samples containing GA in
Fructosyl amino acid oxidase
very small sample volumes (1.3 μL) within about 1 min. We also prepared enzyme sensor strips suitable
Hemoglobin A1c
Point-of-care testing for clinical use in which the enzyme and the mediator were deposited and dried on. This sensor system
Screen-printed carbon electrode showed a clear correlation between the GA concentration and the resulting current. The strips were
stable following 3 months of storage at 37 °C. We conclude that this disposable enzyme sensor strip
system for measuring GA is suitable for point-of-care test (POCT) applications.
& 2016 Elsevier B.V. All rights reserved.
http://dx.doi.org/10.1016/j.bios.2016.08.005
0956-5663/& 2016 Elsevier B.V. All rights reserved.
Please cite this article as: Hatada, M., et al., Biosensors and Bioelectronics (2016), http://dx.doi.org/10.1016/j.bios.2016.08.005i
2 M. Hatada et al. / Biosensors and Bioelectronics ∎ (∎∎∎∎) ∎∎∎–∎∎∎
Please cite this article as: Hatada, M., et al., Biosensors and Bioelectronics (2016), http://dx.doi.org/10.1016/j.bios.2016.08.005i
M. Hatada et al. / Biosensors and Bioelectronics ∎ (∎∎∎∎) ∎∎∎–∎∎∎ 3
of 15% or 30% along with 100 kU/mL of protease in 100 mM P.P.B. temperature for 1 min to facilitate the enzyme reaction coupled
(pH 8.0) was incubated for 1 h at 37 °C to digest GA and release ε- with reduction of the mediator. Afterwards, þ100 mV vs. Ag/AgCl
FK. The mixture was then ultrafiltrated at 4 °C at 14,000 g for was applied to the electrode to reoxidize the reduced mediator.
20 min to remove the protease. Subsequently, 1.3 μL of mixture Figs. 2a and b show the response curve and calibration curve,
containing the sample of protease-digested GA, 60 U/mL of FAOx, respectively, of the Z-FK measurements. Good reproducibility was
and 300 mM of Ru complex in 100 mM P.P.B. (pH 8.0) was ab- observed. The resulting currents decreased gradually to reach a
sorbed into the spacer layer (120 mm thick) of the SPCE. After in- steady state current (Cottrell current), and the current values
cubation for 1 min at room temperature, a potential of þ100 mV correlated positively with Z-FK concentration. The calibration
vs. Ag/AgCl was applied, and the resulting current was observed. curve shows the plot of currents obtained at 20 s after applying
potential versus the Z-FK concentrations. From the calibration
2.3. Preparation and electrochemical measurement of fructosyl ly- curve, a linear response was obtained over the entire Z-FK con-
sine or glycated albumin using enzyme sensor strips centration range (0–500 μM), with a linear regression coefficient
of R2 ¼0.999.
The mixture containing 60 U/mL of FAOx, 300 mM of Ru com- The protease-digested products of GA were then measured
plex, and 0.25% of sucrose in 100 mM P.P.B. (pH 8.0) was deposited using the same electrochemical method. Samples containing ratios
onto the SPCE and dried at 25 °C for more than 30 min. A spacer of either 15% or 30% GA to total albumin were used in the present
(120 mm thick) and a cover were attached afterwards. study. The lower value represents approximately the standard
For the measurement, a volume of 1.3 μL of each sample (Z-FK value of GA, whereas the higher value represents the GA propor-
or protease-digested GA) in 100 mM P.P.B. (pH 8.0) was absorbed tion observed during serious diabetes (Japan Diabetes Society,
into the spacer layer of multiple enzyme sensor strips and in- 2014; Tahara, 2009). The ε-FK concentrations of prepared GA were
cubated at room temperature for 1 min. Subsequently, a potential estimated as 220 μM and 432 μM for 15% GA and 30% GA re-
of þ100 mV vs. Ag/AgCl was applied, and the current was spectively. Fig. 3a shows the response curve obtained by measur-
observed. ing the mixture of the protease digestion product of the GA, FAOx,
and Ru complex. There was a clear difference between the re-
2.4. Evaluation of the storage stability of the enzyme sensor strips sulting currents obtained with samples of high and low GA con-
centrations. Protease-digested non-glycated albumin (non-GA)
Enzyme sensor strips were prepared as above. To evaluate the samples containing known Z-FK concentrations were also mea-
storage stability under different temperatures, the sensors were sured, and a calibration curve was plotted (Fig. 3b) to evaluate the
placed in a plastic container containing silica gel. They were stored impact of the presence of protease-digested materials on the
at 25 °C, 37 °C, or 50 °C in a shaded location. After 91 days, various sensor response. A slight decrease in the sensitivity (slope) and
Z-FK concentrations were measured using the enzyme sensor linear regression coefficient (R2 ¼ 0.993) was observed (Fig. 3b) in
strips. The stability was evaluated by comparing the 91-day the presence of protease-digested materials compared with those
readings with the results obtained from measurements with sen- in the absence of protease-digested materials (Fig. 2b). The ε-FK
sors immediately after preparation. concentrations contained in protease-digested GA were calculated
by applying the current value to the calibration curve formula. The
calculated ε-FK concentrations were 51 and 86 μM for 15% and
3. Results 30% GA, respectively. However, ε-FK concentrations for these
samples calculated from GA concentrations, that were spectro-
s
3.1. Electrochemical measurement of fructosyl lysine or glycated scopically determined using the reference method Lucica GA-L,
albumin were 55 and 108 μM, respectively (assuming that 2 molecules of ε-
FK were released per one GA molecule). Because the calibration
To evaluate the electrochemical measurement principle used in curve was based on Z-FK, whereas the protease digested GA
the present study, we first attempted to measure a synthetic sample contained ε-FK, this slight difference might have resulted
analog of ε-FK, namely Z-FK. After mixing FAOx, mediator, and from the difference of reactivity between Z-FK and ε-FK with
Z-FK, the mixture was applied on SPCE and incubated at room FAOx. Since the calibration curve would ultimately be based on GA
a b
1200 300
y=0.54 x + 14
1000 250
R2=0.999
Current (nA)
200
Current (nA)
800
600 150
400 100
200 50
0 0
0 20 40 60 0 200 400 600
Time (s) Z-FK concentration (µM)
Fig. 2. a) The response curve of the Z-FK measurement. A solution of 1.3 μL containing various Z-FK concentrations, 60 U/mL of FAOx, and 300 mM of Ru complex in
100 mM P.P.B. (pH 8.0) was measured. The concentration range of Z-FK was 0, 50, 100, 300, 500 μM (down to up). (n¼ 3), b) The calibration curve of the Z-FK measurement.
Currents obtained at 20 s after applying potential were plotted versus Z-FK concentration.
Please cite this article as: Hatada, M., et al., Biosensors and Bioelectronics (2016), http://dx.doi.org/10.1016/j.bios.2016.08.005i
4 M. Hatada et al. / Biosensors and Bioelectronics ∎ (∎∎∎∎) ∎∎∎–∎∎∎
a b
250
200 30% GA y = 0.45 x + 18
200 R2=0.993
15% GA
150
Current (nA)
Current (nA)
150
100 5% GA 100
100
50
50
50
0
30% GA
0 50 100 150
0 0
0 10 20 30 40 50 60 70 0 100 200 300 400 500
Time (s) Z-FK concentration (µM)
Fig. 3. a) The response curve obtained by measuring the mixture of the protease digestion product of GA, FAOx, and Ru complex. Dark lines were obtained from 30% GA and
light lines were obtained from 15% GA respectively. (n ¼ 3), b) Calibration curve of protease digested non-GA samples containing known Z-FK concentrations measurement
(n¼ 3). The current values obtained with protease digested 15% GA (circle) and 30% GA (square) were plotted on the calibration curve. The inset shows enlarged curve.
samples of known concentrations, this difference is not significant spacer layer of enzyme sensor strips and applying þ100 mV vs.
for practical use. Furthermore, GA might not be completely di- Ag/AgCl after incubation for 1 min at room temperature. Fig. 4a
gested and less ε-FK might have been released before the elec- shows the response curve of Z-FK measurements. The resulting
trochemical measurement with the current conditions for pro- current was gradually decreased until it reached a steady state
tease digestion. The protease digestion condition of the commer- current, as observed in the measurement of the mixture of Z-FK,
s
cial assay kit Lucica GA-L, which was used as the reference FAOx, and Ru complex (solution condition) (Fig. 2a), and correlated
spectroscopic measurement, has been optimized, so that maximal positively with Z-FK concentration. Once again, good reproduci-
ε-FK was released. This might lead to less ε-FK in the sample used bility was observed. Fig. 4b shows the calibration curve. The linear
for the electrochemical method compared to the sample used for response between Z-FK concentrations and current values was
the spectroscopic method. Therefore, further optimization of the obtained over the entire concentration range of Z-FK (0–500 μM).
condition for protease digestion of GA, suitable for electrochemical Slight decreases in the sensitivity (slope) and linear regression
measurement should be necessary to develop a POCT device em- coefficient (R2 ¼0.993) were observed compared with the solution
ploying this disposable enzyme sensor strip. condition (Fig. 2b). This might be because of the time and process
These results suggested that the current electrochemical required to dissolve of enzyme and mediator on the sensor strips
method is principally applicable for the measurement of GA. by the sample solution, which might affect the homogeneity of the
enzyme reaction, and consequently affect the kinetics of the
3.2. Electrochemical measurement of fructosyl lysine or glycated electrochemical reactions.
albumin using enzyme sensor strips Fig. 5a shows the response curve for the measurement of
protease-digested GA samples using the enzyme sensor strips.
We then investigated the performance of enzyme sensor strips, There was a clear difference between the resulting currents of high
which were prepared initially by depositing the enzyme and and low concentration GA samples. Once again, protease-digested
mediators on SPCEs, followed by a drying process. Various Z-FK non-GA samples containing known Z-FK concentrations were
concentrations and two different GA samples digested by protease measured, and the calibration curve was plotted (Fig. 5b) to
were measured by absorbing 1.3 μL of sample solution to the evaluate the impact of the presence of protease-digested materials
300
1000
250
y=0.49 x + 31
800 R2=0.990
Current (nA)
Current (nA)
200
600
150
400
100
200
50
0
0
0 20 40 60 0 200 400 600
Time (s) Z-FK concentration (µM)
Fig. 4. a) The response curve of Z-FK measurement with enzyme sensor strips which were prepared by depositing the enzyme and mediators on SPCE, and were dried. A
solution of 1.3 μL containing various Z-FK concentrations in 100 mM P.P.B. (pH 8.0) was measured. The concentration range of Z-FK was 0, 50, 100, 300, 500 μM (down to up).
(n¼ 3), b) The calibration curve of Z-FK measurement with enzyme sensor strips. The currents obtained at 20 s after applying potential are plotted versus Z-FK concentration.
Please cite this article as: Hatada, M., et al., Biosensors and Bioelectronics (2016), http://dx.doi.org/10.1016/j.bios.2016.08.005i
M. Hatada et al. / Biosensors and Bioelectronics ∎ (∎∎∎∎) ∎∎∎–∎∎∎ 5
a b
250
140 y = 0.41 x + 23
30% GA
120 200 R2=0.990
15% GA
100
Current (nA)
150
Current (nA)
80
5% GA 100
60 100
50
40
50
20 0
30% GA 0 50 100 150
0 0
0 20 40 60 0 100 200 300 400 500
Time (s) Z-FK concentration (µM)
Fig. 5. a) The response curve obtained by measuring the solution of the protease digestion product of GA with enzyme sensor strips. Dark lines were obtained from 30% GA
and light lines were obtained from 15% GA, respectively. (n ¼3), b) Calibration curve of protease digested non-GA samples containing known Z-FK concentrations mea-
surement with enzyme sensor strips (n¼ 3). The current values obtained with protease digested 15% GA (circle) and 30% GA (square) were plotted on the calibration curve.
The inset shows enlarged curve.
Please cite this article as: Hatada, M., et al., Biosensors and Bioelectronics (2016), http://dx.doi.org/10.1016/j.bios.2016.08.005i
6 M. Hatada et al. / Biosensors and Bioelectronics ∎ (∎∎∎∎) ∎∎∎–∎∎∎
an enzyme analysis method which is already used in clinical POCT devices used for monitoring lipid panels (Méndez-González
practice, with the combination of FAOx, Ru complex as the elec- et al., 2010). We also prepared enzyme sensor strips that are
tron mediator, and an inexpensive and economically attractive practical and can be easily used, which showed a clear correlation
SPCE, which will be utilized for POCT of GA. We developed this between measured current and GA concentrations. The storage
sensor strip for GA measurement by referring to the formulation of stability of these enzyme sensor strips was 3 months at 37 °C. This
the sensor strip utilized in commercial SMBG. In this type of storage capability was achieved with chemical components used
measurement, the enzyme, mediator and substrate react in a mL- to develop this electrochemical principle only on a laboratory
scale volume, and the potential is applied after starting the reac- scale. Further steps to stabilize the system for longer shelf life
tion. In these cases, the enzyme reaction reaches its equilibrium storage of the sensor strips could include adding chemical stabi-
relatively fast, usually before the potential is applied. The response lizers to the enzymes, optimizing the drying procedure, and de-
current then is an indicator for the concentration of the reduced signing an envelope or cartridge for storage.
mediator at equilibrium. Therefore, the mediator must be present One critical issue of the current status of the method is the so-
enough not to limit the enzyme reaction. Consequently, in SMBG, called pretreatment time required to prepare a sample of protease-
several tens to hundreds mM of mediator is often used. For ex- digested GA to be measured by the enzyme sensor strips. The
s
ample, in one of our group's contribution reported previously, current commercially available enzyme analysis kit, Lucica GA-L
60 mM Ru complex was used as mediator (Yamaoka and Sode, also uses a protease digestion procedure. With the Lucia system,
2007a) and in another, 60, 120 and 240 nmol/sensor (120, 240, the required time for digestion is 5 min and protease is not re-
480 mM) Ru complex were compared and 240 nmol/sensor was moved before adding FAOx contained reagent. Therefore, by op-
chosen as the optimal concentration (Yamaoka and Sode, 2007b). timizing the reaction conditions such as pH or temperature, the
s s
In the commercially available OneTouch Ultra Test Strips (for period for GA pretreatment can be shortened. Furthermore, the
SMBG), more than 22 μg (67 mM) ferricyanide is used as mediator step for protease removal also can be omitted before electro-
(LifeScan, Inc., 2008). We investigated the impact of the Ru com- chemical measurement. So the entire analysis procedure will be-
plex concentration on the Z-FK measurements (S1), and used come simpler and ideally, an on-chip reaction for protease diges-
300 mM as the minimal concentration of Ru complex not limiting tion will be developed.
the reaction. The FAOx and protease digestion procedures we The measurement units of GA are specified to be the proportion
employed in the present study are identical to those utilized in of the GA concentration relative to the total albumin concentra-
clinical practice. Therefore, the process presented here is a realistic tion, and therefore, the albumin concentration has to be de-
method to measure GA in blood samples. Furthermore, electro- termined to achieve clinically relevant values. Currently, a spec-
s
chemical methods are often used for POCT systems because of troscopic method is used in Lucica GA-L for measuring total al-
their suitability for miniaturized systems, good sensitivity, re- bumin. Combining this spectroscopic method with the electro-
producibility, and low cost, and, moreover, their compatibility with chemical GA measurement system in this study can be applicable
enzymatic methods. These advantages of the electrochemical for quick practical application. Ideally, an electrochemical mea-
systems are evidenced in the fact that electrochemical sensors are surement system of albumin should be combined. Electrochemical
dominating commercialized SMBG systems, although SMBG sys- methods for albumin measurement were also reported (Barisci
tems were first developed based on reflectometer technology et al., 1998; Chuang et al., 2006; Luque et al., 2007; Cieplak et al.,
(Newman and Turner, 2005; Yoo and Lee, 2010; Clarke and Foster, 2015; Yasukawa et al., 2015), which may lead to the development
2012). of a principle suitable for disposable sensing system. Therefore, the
In general, electrochemical measurements are subject to in- future combination of an electrochemical principle for total albu-
terferences, such as ascorbic acid and other electroactive sub- min concentration and our electrochemical principle for GA sen-
stances in blood. However, Ru-complex electron transfer allowed sing will realize an electrochemical POCT device to measure GA
us to conduct electrochemical measurements with a relatively low units specified as the proportion of the concentration relative to
operational oxidation potential (þ 100 mV vs Ag/AgCl), which the total albumin concentration.
should minimize the impact of such electroactive ingredients in a Our research group is also engaged in the development of
sample solution. Other electroactive interferences might be pro- various biomolecules dedicated to sensing glycated proteins. In
tein-derived interfering substances including oxidizable amino particular, we have reported the construction of engineered FAOx/
acids, such as cysteine. Because about 60–70% of plasma protein is FPOx (Kim et al., 2010, 2012; Ferri et al., 2004, 2005, 2013). In
albumin, the majority of amino acids would be derived from al- contrast to the wild-type enzymes, the engineered enzymes
bumin. Albumin consists of 585 amino acid residues and has 34 strongly preferred artificial molecules to oxygen as electron ac-
cysteine residues. Therefore, any interferences due to a release of ceptor. By employing these types of engineered FAOx enzymes,
cysteine or other oxidizable amino acids during the protease di- then an increase in sensitivity will also likely be achieved.
gestion should occur when measuring protease digested albumin To elevate the research presented in this article into a com-
or GA. The present results indicated that the background current, mercial product, then further study will be required to use human
namely the y-intercept of the calibration curve of Z-FK measure- serum or plasma as the sample measured by these enzyme sensor
ments, were nearly the same whether they contained (Fig. 2b) or strips under development. Furthermore, a monitor will need to be
did not contain (Fig. 3b) protease-digested albumin. Therefore, it created with an algorithm to convert current to a GA concentra-
appeared that this measurement system was not affected by in- tion. These developments will result in the development of an
terfering substances that could be contained in protease-digested electrochemical POCT sensor for GA.
albumin.
The development of a cost-effective SPCE will facilitate mass
production of GA sensor strips, which can be fabricated using the 5. Conclusion
same type of procedure that is used for BG monitor systems. The
sensor strips described in this article could measure a protease- In our study, we developed a sensing system for measurement
digested GA sample with a very small sample volume (1.3 μL) of GA by combining an enzyme analysis method, which is already
within about 1 min after sample preparation by protease diges- used in clinical practice, with electrochemical principles. FAOx, Ru
tion. These volume and measurement time conditions are ap- complex as the electron mediator, and an inexpensive and eco-
proximately the same as those required by currently marketed nomically attractive SPCE were used in this study. The prepared
Please cite this article as: Hatada, M., et al., Biosensors and Bioelectronics (2016), http://dx.doi.org/10.1016/j.bios.2016.08.005i
M. Hatada et al. / Biosensors and Bioelectronics ∎ (∎∎∎∎) ∎∎∎–∎∎∎ 7
enzyme sensor strips were able to measure a protease-digested GA Fang, L., Li, W., Zhou, Y., Liu, C.-C., 2009. Sens. Actuators B 137, 235–238.
sample with a very small sample volume (1.3 μL) within about Goldstein, D.E., Little, R.R., Lorenz, R.A., Malone, J.I., Nathan, D., Peterson, C.M.,
Sacks, D.B., 2004. Diabetes Care 27, 1761–1773.
1 min after sample preparation by protease digestion. The system Guthrow, C.E., Morris, M.A., Day, J.F., Thorpe, S.R., Baynes, J.W., 1979. Biochemistry
had a storage stability of 3 months at 37 °C. These results indicate 76, 4258–4261.
that the developed sensor has a legitimate potential for a POCT Hare, M.J., Shaw, J.E., 2016. Curr. Diabetes Rev. 12, 51–57.
Hirokawa, K., Nakamura, K., Kajiyama, N., 2004. FEMS Microbiol. Lett. 235, 157–162.
application. Further studies, such as optimization of protease di-
Hashiba, H.,J., 1976. Agric. Food Chem. 24, 70–73.
gestion procedure will advance the progression of this sensor to- Japan Diabetes Society, Treatment Guide for Diabetes 2014–2015. 〈http://www.fa.
ward clinical use. kyorin.co.jp/jds/uploads/Treatment_Guide_for_Diabetes_2014-2015.pdf〉.
Jain, U., Chauhan, N., 2016. Biosens. Bioelectron., doi:10.1016/ j.bios.2016.02.033 [e-
pub ahead on line]
Kameya, M., Sakaguchi-Mikami, A., Ferri, S., Tsugawa, W., Sode, K., 2015. J. Diabetes
Acknowledgments Sci. Technol. 9, 183–191.
Kameya, M., Tsugawa, W., Yamada-Tajima, M., Hatada, M., Suzuki, K., Sakaguchi-
Mikami, A., Ferri, S., Klonoff, D.C., Sode, K., 2016. Biotechnol. J. 2016, 1–8.
We acknowledge i-SENS (Seoul, Republic of Korea) for kindly
Kim, S., Nibe, E., Ferri, S., Tsugawa, W., Sode, K., 2010. Biotechnol. Lett. 32,
supplying screen printed carbon electrodes. We also acknowledge 1123–1129.
Asahi-Kasei Pharma (Tokyo, Japan) for kindly supplying samples Kim, S., Nibe, E., Tsugawa, W., Kojima, K., Sode, K., 2012. Biotechnol. Lett. 34,
including FAOx, Z-FK, GA, recombinant albumin, and protease, and 491–497.
Koga, M., 2014. Clin. Chim. Acta 43, 96–104.
furthermore, for helping with the determination of GA con- Kohzuma, T., Koga, M., 2010. Mol. Diagn. Ther. 14, 49–51.
s
centration by Lucica GA-L. Kouzuma, T., Usami, T., Yamakoshi, M., Takahashi, M., Imamura, S., 2002. Clin. Chim.
Acta 324, 61–71.
Liu, G., Iyengar, S.G., Gooding, J.J., 2012. Electroanalysis 24, 1509–1516.
LifeScan, Inc., 2008. OneTouchs Ultras Test Strips, 〈https://www.onetouch.com/
Appendix A. Supplementary material sites/d7admin-onetouch.jnj.com.professional/files/docfile/1391164268C_UBlue
TestStrip_Instructions_for_use.pdf〉.
Supplementary data associated with this article can be found in Luque, G.L., Ferreyra, N.F., Rivas, G.A., 2007. Talanta 71, 1282–1287.
Méndez-González, J., Bonet-Marqués, R., Ordóñez-Llanos, J., 2010. Point Care: J.
the online version at http://dx.doi.org/10.1016/j.bios.2016.08.005. Near-Patient Test. Technol. 9, 102–107.
Nanjo, Y., Hayashi, R., Yao, T., 2007. Anal. Chim. Acta 583, 45–54.
Newman, J.D., Turner, A.P.F., 2005. Biosens. Bioelectron. 20, 2435–2453.
Ogawa, K., Stoöllner, D., Scheller, F., Warsinke, A., Ishimura, F., Tsugawa, W., Ferri, S.,
References
Sode, K., 2002. Anal. Bioanal. Chem. 373, 211–214.
Qu, L., Xia, S., Bian, C., Sun, J., Han, J., 2009. Biosens. Bioelectron. 24, 3419–3424.
Andersen, J.T., Dalhus, B., Viuff, D., Ravn, B.T., Gunnarsen, K.S., Plumridge, A., Rajkumar, R., Katterle, M., Warsinke, A., Möhwald, H., Scheller, F.W., 2008. Biosens.
Bunting, K., Antunes, F., Williamson, R., Athwal, S., Allan, E., Evans, L., Bjørås, M., Bioelectron. 23, 1195–1199.
Kjærulff, S., Sleep, D., Sandlie, I., Cameron, J., 2014. J. Biol. Chem. 289, Sakaguchi, A., Tsugawa, W., Ferri, S., Sode, K., 2003. Electrochemistry 71, 442–445.
13492–13502. Sakurabayashi, I., Watano, T., Yonehara, S., Ishimaru, K., Hirai, K., Komori, T., Tagi, M.,
Abidin, D., Liu, L., Dou, C., Datta, A., Yuan, C., 2013. Anal. Methods 5, 2461–2469. 2003. Clin. Chem. 49, 269–274.
Apiwat, C., Luksirikul, P., Kankla, P., Pongprayoon, P., Treerattrakoon, K., Pai- Sakaguchi, A., Ferri, S., Tsugawa, W., Sode, K., 2007. Biosens. Bioelectron. 22,
boonsukwong, K., Fucharoen, S., Dharakul, T., Japrung, D., 2016. Biosens. Bioe- 1933–1938.
lectron. 82, 140–145. Sakaguchi-Mikami, A., Ferri, S., Katayama, S., Tsugawa, W., Sode, K., 2013. J. Appl.
Bunn, H.F., Gabbay, K.H., Gallop, P.M., 1978. Science 200, 21–27. Microbiol. 114, 1449–1456.
Barisci, J.N., Hughes, D., Minett, A., Wallace, G.G., 1998. Anal. Chim. Acta 371, 39–48. Sode, K., Ohta, S., Yanai, Y., Yamazaki, T., 2003. Biosens. Bioelectron. 18, 1485–1490.
Chawla, S., Pundir, C.S., 2011. Biosens. Bioelectron. 26, 3438–3443. Stratton, I.M., Adler, A.I., Neil, H.A.W., Matthews, D.R., Manley, S.E., Cull, C.A.,
Chawla, S., Pundir, C.S., 2012. Anal. Biochem. 430, 156–162. Hadden, D., Turner, R.C., Holman, R.R., 2000. BMJ 321, 405–412.
Chopra, A., Tuteja, S., Sachdeva, N., Bhasin, K.K., Bhalla, V., Suri, C.R., 2013. Biosens. The Diabetes Control and Complications Trial Research Group, 1993. N. Engl. J. Med.
Bioelectron. 44, 132–135. 329, 977–986.
Chuang, M.C., Liu, C.C., Yang, M.C., 2006. Sens. Actuators B 114, 357–363. Tahara, Y., 2009. Diabetes Res. Clin. Pract. 84, 224–229.
Clarke, S.F., Foster, J.R., 2012. Br. J. Biomed. Sci. 69 (2), 83–93. Tsugawa, W., Ishimura, F., Ogawa, K., Sode, K., 2000. Electrochemistry 68, 869–871.
Cieplak, M., Szwabinska, K., Sosnowska, M., K.C., C.B, Borowicz, P., Noworyta, K., Tsugawa, W., Ogawa, K., Ishimura, F., Sode, K., 2001. Electrochemistry 69, 973–975.
D'Souza, F., Kutner, W., 2015. Biosens. Bioelectron. 74, 960–966. Ueda, Y., Matsumoto, H., 2015. J. Diabetes Sci. Technol. 9, 177–182.
Franco, R.S., 2012. Transfus. Med Hemother. 39, 302–307. Yamazaki, T., Ohta, S., Yanai, Y., Sode, K., 2003. Anal. Lett. 36, 75–89.
Ferri, S., Miura, S., Sakaguchi, A., Ishimura, F., Tsugawa, W., Sode, K., 2004. Mar. Yamaoka, H., Sode, K., 2007a. J. Diabetes Sci. Technol. 1, 28–35.
Biotechnol. 6, 625–632. Yamaoka, H., Sode, K., 2007b. Open Biotechnol. J. 1, 26–30.
Ferri, S., Sakaguchi, A., Goto, H., Tsugawa, W., Sode, K., 2005. Biotechnol. Lett. 27, Yoo, E.H., Lee, S.Y., 2010. Sensors 10, 4558–4576.
27–32. Yonehara, S., Inamura, N., Fukuda, M., Sugiyama, K., 2015. J. Diabetes Sci. Technol. 9,
Ferri, S., Kim, S., Tsugawa, W., Sode, K., 2009. J. Diabetes Sci. Technol. 3, 585–592. 200–205.
Ferri, S., Miyamoto, Y., Sakaguchi-Mikami, A., Tsugawa, W., Sode, K., 2013. Mol. Yasukawa, T., Kiba, Y., Mizutani, F., 2015. Anal. Sci. 31, 583–589.
Biotechnol. 54, 939–943.
Please cite this article as: Hatada, M., et al., Biosensors and Bioelectronics (2016), http://dx.doi.org/10.1016/j.bios.2016.08.005i