Main unit:
Cont ra ange l(CX235- 1) Test wire File holder
Function option
NO.
② ③ Normal Automatic1 Automatic2 Apex Slow down Start/stop
① Mode module
M1 ü - - - - -
② Forward & reverse sign: display for forward or reverse rotation
③ Apex locator module M2 ü ü - - - -
④ Rotation speed setting value
M3 ü - ü - - -
⑤ Torque setting value
⑥ Current torque M4 - - - ü - -
⑦ Program number
⑧ Battery power display M5 ü - - ü - -
M6 ü ü - ü - -
Ⅴ Assemble M7 ü ü - ü ü -
1. Please put the instrument on a solid place. M8 ü ü - ü - ü
2. Insert the foot switch connector into the hole on the main unit
3. Insert the motor wire connector into the hode on the main unit. M9 ü ü - ü ü ü
4. Connect th e 1:1 contra angle to the motor and clamp the file into
the contra angle. And then put the handle on the handle base. This unit has 9 modes and consists of 6 function options:
5. Insert the adaptor into the hole on the main unit NORMAL:Manually motor functions
AUTOMATIC1:automatic reverse1,When the torque goes to the setting value,
Notice: when the instrument is being used, please don't use the the file reverses automatically and stops automatically upon
adaptor for avoiding the electromagnetic interference. withdrawal
AUTOMATIC2:automatic reverse2,When the torque goes to the setting value,
6. If only use the apex locator function ( M4 ), please insert the test the file reverses automatically and starts rotating
wire ( dual holes) to the hole on the main unit,and connect the file hodler and automatically upon withdrawal
stainless electrode to test wire. APEX:Apex locator
5 6
SLOW DOWN:When the file is approaching to apex, the file automatically Press SET button,apex module flashing,the apex position flash last set position,
slows down for safer treatment indicating having been entering into the setting mode,adjust the apex
position with pressing “ADJUST +” and “ADJUST -” button, the range
START/STOP: The file starts rotating automatically when inserted into the
within 0.0-1.0. Press SET again to memorize the apex position,the apex module
canal and it stops automatically upon withdrawal
stops flashing,setting completed.
Notice: It will quit the setting mode after 2S without any press
ⅱ. Program settings operation ,and the numerical value has been set will be lost.
Before use, the user should set the speed and torque and save the programs.
There are 9 programs can be set by the user. Ⅶ Use Method
Setting method:
Notice: The program only can be set in M1、M2、M3 modes
(without apex locator function) ⅰ. Preparation before the use
Step1. Press SET button to enter into the setting mode, speed value flashing, then 1. Check each component connection
press “ADJUST +” button or “ADJUST - ”button to set the speed, 2. Press POWER button, make sure the information displayed in the LCD screen
speed range:125、250,375,500,625rpm,press SET button again, is completed
speed value stops flashing and it' s been saved; 1) LCD displays last working mode.
2) LCD displays last program.
Step2.when torque value flashing,press “ADJUST +”button or“ADJUST -”
3) LCD displays apex module in M4 mode
button to set the torque value,the torque range:from 6 to5 2 mN. m,
4) Indicating the power of battery quantity, the battery is running when
press SET button,torque value stops flashing and it's been saved;
it is charging
Step3. When program number flashing, press “ADJUST +” button or 3. Press MODE button to select the working mode
“ADJUST -” button to set program number, press SET button, 4. Press PROGRAM button to select the working program
program number stops flashing and it' s been saved , setting is
completed。 ⅱ. Operation
Notice: It will quit the setting mode after 2S without any press
operation, and the numerical value, which had been set,
Notice: when the instrument is being used, please don' t use the
will be lost.
adaptor for avoiding the electromagnetic interference.
ⅲ.Apex position setting M1:Manual motor function, file will rotate when pressing foot switch and will
Dentist can adjust the apex position if necessary. For example you can set the stop when the torque reaches the setting torque. Release the foot
apex position 0.5mm away to the apex hole and unit will memorize this switch, the file stops.
position automatic and endo file will automatic reverse or stop when file reaches M2:Motor with automatic reverse 1 function,press foot switch, file rotates
this position. forward,when the torque reaches the setting torque,file reverses
Set method: automatically,and will stop when the torque is disappeared. Release
Notice: Dentist can set the apex position only in M4 mode (apex the foot switch, the file stops.
M3:Motor with automatic reverse2 function,press foot switch, file rotates
locator mode).
forward, when the torque reaches the setting torque,file reverses
7 8
M5: Apex locator and motor function, press foot switch and the file
automatically, and will rotate forward when the torque is disappeared. rotates forward, the file stops working when it reaches the setting
Release the foot switch, the file stops. apex position.
Notice: when the instrument is in M4 - M9, please don't use M6: Apex locator & auto reverse1 motor function, press foot
the adaptor for it cannot get an exact value. switch the file rotates forward and reverses when it reaches the
M4:In this mode, only apex locator functions, methods of operation are as follows: setting apex position and stops without touching the root canal.
M7: Apex locator &autoreverse1 motor & slowdown function, press
NOTICE: it cannot get an exact value of the apex length when the foot switch the file rotates forward and automatically slows down
treatment is in an unstable condition. So, please insert when approaching to apex and reverses automatically when it
the file into the root-canal and take an X-ray photo to reaches the setting apex position and stops without touching the
make sure the veracity. root canal.
M8: Apex locator & auto reverse1 motor& automatically start and
1. The plug of the testing wire should be completely plugged into stop function ( no need foot switch in the mode). The file starts to
hole on the main unit. rotate forward when it entering into the root canal and reverse
* When you need to pull out the test wire, please don' t pull the wire but when it reaches the setting apex position, and stops without
hold its pin and pull it out. touching the root canal.
M9: Apex locator & auto reverse1 motor & automatically start and
2. Clip the metal part of the file with the holder and hang the stainless
stop & slowdown ( no need foot switch in the mode). The file
electrode up at any side of the patient's mouth.
starts to rotate forward when it entering into the root canal and
3.Insert the file into the root- canal,and circumrotate the file toward the
automatically slows down when approaching to apex and reverses
apex of the root-canal. And then the distance between the tip of the file
automatically when it reaches the setting apex position and stops
and the apex of the root-canal will be displayed on the LCD screen. When
without touching the root canal.
the figure on the LCD screen shows 0.5mm, please fasten the file with
Reverse: Press the REVERSE button, file will reverse and you can
the rubber vernier caliper on the reference point of the tooth crest rather
change the forward and reverse by press the REVERSE button.
than keep up probing into the root-canal. Then measure the distance
Notice:When the torque reaches or exceeds the setting torque in
from the bottom of rubber vernier caliper to the tip of the file. M5 - M9, and no matter the root canal enlarged file
* Root- canal length is the file distance when the LCD displays 0.5. reaches the set data or not, the file would stop or
4. Deciding the working length of root canal reserves automatically.
When the figure on the LCD screen shows 0.5mm, then measure the
distance from the bottom of rubber vernier caliper to the tip of the
ⅲ. Power manage
file. Take note of this figure. So subtract 0.5-1.0mm from the above
data is the working length of root-canal. 1. Battery power
When the file drills through the root-canal, please make the final value in LCD displays the current battery power, the number of lattice shows more, and the
terms of the one measured in 0.5mm. power is fuller. If there is no electricity the lattice will flicker,which means the
The working length of root-canal varies from each other for the reason of battery power is about to run out , at this status the beeper will ring per 3S, and
different shapes of teeth and root-canal. unit will shut down after 60S.
5. When the file reaches the position to 2.0mm away from the apex of the
2. Battery recharging
root-canal, there will be a continuous alarm.
9 10
To fully recharge the battery, you should connect the unit to adaptor at least
for 4 hours or more (The new battery requires at least 8 hours or more) , Resistance in Increase the torque by
root-canal is too manual adjusting the
the sign will be running during the battery charging. File stops much forward/reverse mode
automatically
Notice: 1. when the instrument is connected to the adaptor, the beeper Bad condition in Change to non-apex
will beep for three times, showing the connection is correct. root canal (M5) locator mode (M1-M3)
2. when the instruments in M4-M9, please don't charge the The unit is in Restart and compare
unstable condition with x-ray film
battery.
3. Automatic power off Inaccurate apex Bad condition in Clean the liquid and
value root canal remnant
The instrument will shut down automatically when it ceases working for 3 minutes.
Electromagnetic Don't use the adaptor
In that case, all display and function will stop. Interference for charging
Notice: When power off, the LCD will show recharging sign when Not in Select
charging. auto-forward/reverse auto-forward/reverse
File can't mode mode
reverse
automatically Preset torque is too much Decrease the torque
Ⅷ Maintenance and sterilization File has not reached File will reverse only
the preset apex when the file reach the
1. If not use the unit for long time, please open the battery cover and remove (M6-M9) preset apex location
the battery, keep the instrument clean and store it somewhere dry and clean. The file hasn't been into The file will start when
2. Parts like contra angle and stainless electrode, which have direct contact with the root canal goes into the root canal
File can't start Root canal is too dry Infuse some normal saline
the patient, should be disinfected by autoclave each time after using.
(M8,M9) The pin or electrode is in Clean or change the
3. The crust of the instrument and the wire should be disinfected by the ethanol. bad contact pin/electrode
Oral mucosa is too dry Partial humidify oral mucosa
Don't use the adaptor for
Ⅸ Troubleshooting Auto start/stop
out of control
Electromagnetic
Interference
charging
Change to non-apex
If the equipment is not normal work, before call our after-sales center, please (M8,M9) locator mode (M1-M3)
check the table below. The file Increase the torque by
Resistance in
Phenomenon Reason analyses Problem shooting reverses root-canal is manual adjusting the
Can't power on Low battery Recharge the battery continually too much forward/reverse mode
Check the wire and Clean the liquid and remnant
Bad contact in foot Bad condition in
make sure the foots The file root canal (M6-M9) Change to non-apex locator
switch reverses mode (M1-M3)
switch works well
File doesn't run continually The preset apex location
Resistance in Increase the torque by Reset the apex location
root-canal is too manual adjusting the is too high (M6-M9)
much forward/reverse mode
11 12
The opening of root
The file canal is too small(M6-M9) Enlarge it
reverses
continually The file is too big (M6-M9) Change a smaller size
Electromagnetic Don't use the adaptor for
Interference charging
The using time Not fully charged Charge the battery at
of battery is least 4 hours
too short Battery is aged Replace the battery
Ⅻ Mark illustration
Class Ⅱ equipment
Type B equipment
CE marked product
13