Name of the Specifications Quantity


Equipment
1 Elisa Reader  Filter based ELISA reader with halogen lamp as light 1
source and single channel optics / single detector
with silicon photo detectors for reliable results.
 Halogen Lamp should have minimum usage option,
i.e. it’s should get ‘on’ while reading and should be
‘off’ mode even the instrument is on standby mode.
 Should have a linear measurement range of 0 to 4
Abs.
 Should be compatible to wavelength range from 340
to 850nm
 Should have a photometric accuracy of ±2 % or
better.
 Should have a resolution of 0.001Abs.
 Should have variable speed plate shaking capability.
 Should have easy access 8 position filter wheel.
 Instrument should be supplied with 3 standard filters
of 405nm, 450nm and 620nm.
 Should have automatic filter selection option.
 Should have automatic calibration before each
reading.
 Should able to read 96 well format micro well plate
in 6 seconds.
 Capable of doing multi standard tests and controls.
 Should be capable of reading U, V and flat type
wells.
 The instrument should run in stand-alone mode OR
with computer & software controlled.
 Internal software should have all the basic application
related to ELISA like quality control, qualitative
classification, standard graph etc.
 System should able to run in stand-alone mode using
4.5 inch color display and keypad.
 Also system should able to run with computer &
software controlled mode.
 The instrument should have a memory of 99 inbuilt
protocols in stand-alone mode.
 The instrument should have 2 USB ports, one for the
easy data transfer and another to connect printer
directly to the unit.
 System should have USB port for external printer
connectivity optionfor the data printout.
 Software CD should have unlimited user license to
analyze the data in different user / computer systems.

Page 1 of 4
Accessories

 Compatible online UPS and Computer


Data Analysis Software :

 Analysis software should be supplied with the


instrument and has unlimited user system license.
 Software should have standard curve fit, qualitative
classifications, parallel line analysis etc.
 Software should have option for area selection. i. e
different protocols at different area of the same plate.
 Database based software to run backups of all data,
restore back up data (in case of hardware failure of
original computer).

Warranty-Two year
AMC: 3 year

Page 2 of 4
2. CO2 Incubator  Temperature range Ambient +5°C to 45°C .
 Temperature Uniformity ±0.3 °C at 37 °C.
 Interior Volume: Minimum 150 liters.
 CO2 control range: 0-20%. Control: Better than ±0.1%,
measures with T/C sensor
 Readability and Stability at least 0.1%.
 Stainless steel shelves and supports should be readily
removable without tools for easy cleaning, autoclaving or
adjustment.
 Heat sterilization cycle of at least 180⁰C for one hours to
safely kill all organisms within the chamber.
 Built in system audible alarm should be activated if the outer
door is opened during the Heat Sterilization cycle and the
temperature is 60°C or greater to ensure safety in the lab.
 The Microprocessor Control System should give clear cut
written message for the errors, should guide through the cycle
with start-up and cycle status messages during
decontamination phase.
 Post decontamination process, the system should return to
normal operating status without the need for manual cleaning.
 HEPA filtered chamber airflow.
- 100% filtration of chamber air every 60 seconds to
clean room particulate and air borne contamination to
best air quality standards thereby providing constant
protection against contaminations.
 Built-in system with adjustable timeframe to notify the user
when it’s time to replace the filter and should be easily
removable without tools.
 Uniform distribution of direct heat at every part of the inner
chamber.
 There should be microbiological filters on all gas inlets and
outlets and sample ports to reduce the potential contamination
sources.
 Programmable tracking alarms for temperature and CO2.
Alarms should be custom configurable.
 All probes and sensors located inside the chamber should be
easily located.
 Peltier/cooling device to maintain ambient 37 °C even at
higher temperature outside.
 System should be compatible to 230V/50-60 Hz.
 Warranty: 1 year
 Accessories
 One Compatible CO2 Cylinder and regulator
 ISO certified
 Compatible stablizer