You are on page 1of 10

International Journal of Laboratory Hematology

The Official journal of the International Society for Laboratory Hematology

ORIGINAL ARTICLE INTERNAT IONAL JOURNAL OF LABORATO RY HEMATO LOGY

Internal Quality Control Practices in Coagulation Laboratories:


recommendations based on a patterns-of-practice survey
A. MCFARLANE*, B. ASLAN*, A. RABY*, K. A. MOFFAT † , R. SELBY ‡ , R. PADMORE §

*Institute for Quality S U M M A RY


Management in Healthcare
(IQMH), Toronto, ON, Canada Introduction: Internal quality control (IQC) procedures are crucial

Hamilton Regional Laboratory for ensuring accurate patient test results. The IQMH Centre for Pro-
Medicine Program, Hamilton,
ON, Canada ficiency Testing conducted a web-based survey to gather informa-

Sunnybrook Health Sciences tion on the current IQC practices in coagulation testing.
Centre and University Health Methods: A questionnaire was distributed to 174 Ontario laborato-
Network, Toronto, ON, Canada ries licensed to perform prothrombin time (PT) and activated partial
§
Ottawa Hospital—General
Campus, Ottawa, ON, Canada thromboplastin time (APTT).
Results: All laboratories reported using two levels of commercial QC
Correspondence: (CQC); 12% incorporate pooled patient plasma into their IQC pro-
A. McFarlane, Institute for gram; >68% run CQC at the beginning of each shift; 56% follow-
Quality Management in Health-
care (IQMH), 393 University ing maintenance, with reagent changes, during a shift, or with
AvenueSuite 1500, Toronto every repeat sample; 6% only run CQC at the beginning of the day
M5G 1E6, Ontario, Canada. and 25% when the instruments have been idle for a defined period
Tel.: +1-416-323-9540;
of time. IQC run frequency was determined by manufacturer rec-
Fax: +1-416-323-9324;
E-mail: amcfarlane@iqmh.org ommendations (71%) but also influenced by the stability of test
(27%), clinical impact of an incorrect test result (25%), and sam-
ple’s batch number (10%). IQC was monitored using preset limits
doi:10.1111/ijlh.12397 based on standard deviation (66%), precision goals (46%), or
allowable performance limits (36%). 95% use multirules. Failure
Received 18 February 2015;
actions include repeating the IQC (90%) and reporting patient
accepted for publication 19 May
2015 results; if repeat passes, 42% perform repeat analysis of all patient
samples from last acceptable IQC.
Keywords Conclusion: Variability exists in coagulation IQC practices among
Coagulation, quality control, Ontario clinical laboratories. The recommendations presented here
hemostasis, laboratory practice would be useful in encouraging standardized IQC practices.

optimal quality control (QC) program, can minimize


INTRODUCTION
analytical errors and better ensure accurate patient
The laboratory plays a key role in providing test results and the quality of test performance [2]. Over
results that help guide the management of patient time, QA has evolved from just running QC on analy-
care [1]. Laboratories that design and implement good sis to included aspects that are outside of the analytic
quality assurance (QA) practices, which includes an process, such as pre-analytical, postanalytical, and

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738 729
730 A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES

other issues that may impact patient results [1, 2]. tion laboratories, as in all clinical laboratories, have a
Internal quality control (IQC) remains the fundamen- significant role in ensuring accurate and precise
tal component of most laboratory QA programs and patient test results for the management of patient care
provides documented evidence that the laboratory’s and improvement of laboratory services [9].
results are accurate and acceptable for release [3]. The Institute for Quality Management in Health-
Although, external QC and external quality assess- care (IQMH) [formerly Quality Management Program
ment (EQA) or proficiency testing programs are help- —Laboratory Services (QMP-LS)] provides ISO 17043-
ful in assuring accuracy and, to some extent, accredited proficiency testing for medical laboratories
precision, the daily IQC demonstrates day-to-day [10]. In 2012, a web-based patterns-of-practice survey
method consistency between external challenges and was conducted to gather information on current IQC
the original method validation [4]. practices in Ontario laboratories for coagulation test-
Currently, there are many challenges facing labora- ing.
tories, such as increased workload and scope of test- The scope of this review is not to provide the basics
ing, usually with no additional resources or staffing. of QC, but to provide guidance and recommendations
In addition, implementation of automated platforms to clinical laboratories for the planning and imple-
for higher throughput and increasing use of auto-veri- mentation of a good IQC program in coagulation test-
fication requires close monitoring of systems to ensure ing. Standardization of IQC practice within and across
accurate results are reported [5, 6]. laboratories will help provide a means for detecting
The interpretation of IQC data has evolved from important errors and minimize false rejection of test
manual plotting of data and visual inspection of Le- results to ensure quality patient care.
vey–Jennings graphs to the use of multiple-rule IQC
computer software. The use of integrated software
METHODS
programs rapidly alert users to potential IQC changes
thereby ensuring better reliability of results. The A web-based questionnaire was created and distributed
advent of advanced software, within or interfaced to to 174 Ontario clinical laboratories that are licensed to
the laboratory information system (LIS), has greatly perform prothrombin time (PT), reported as interna-
accelerated the ability of laboratories to deploy more tional normalized ratio (INR) and activated partial
advanced IQC practices. Ongoing education is thromboplastin time (APTT). Results were collected
required to ensure appropriate understanding and use electronically using QViewTM and assessed using MS
of IQC when developing and implementing advanced Excelâ (Microsoft, Redmond, WA, USA). All 174 labo-
software programs [5, 6]. ratories responded. One hundred and forty-one (81%)
National and international guidelines focusing on laboratories were affiliated with either small or med-
good laboratory practice are available for the manage- ium hospitals with less than 500 beds; 12 (7%) partici-
ment of IQC. It is important for laboratories to con- pants were from large hospitals with greater than 500
sider the appropriate requirements and guidance beds; 17 (10%) were ‘community’ (private) laborato-
documents including the International Organization ries; and four (2%) participants were affiliated with
for Standardization (ISO), specifically ISO 15189, and ambulatory care centers. The average number of
the Clinical and Laboratory Standards Institute (CLSI) patient samples tested for PT/INR and APTT per day
in the development of an IQC program [7, 8]. A well- varied from less than 20 to greater than 600 (Figure 1).
developed IQC program will continuously monitor the
performance of an assay to ensure the reported results
R E S U LT S
are accurate, reliable, and reproducible. The design of
the IQC system in a routine coagulation laboratory
Types of quality control materials
consists of a number of critical decisions that will ulti-
mately determine its overall effectiveness. Coagulation Laboratories were asked what types of quality control
tests are essential in the diagnosis of coagulopathies materials they used. All laboratories used commercial
and to monitor the effectiveness of some anticoagu- quality control materials with 131 (75%) running
lant therapies. Quality assurance practices in coagula- commercial QC material from their analyzer’s manu-

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738
A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES 731

APTT PT/INR

Average number of tests reported daily


>600
401 – 600

201 – 400

101 – 200

51 – 100

21 – 50
Figure 1. Average number of
patient samples tested for PT ≤20
and APTT in Ontario 0% 10% 20% 30% 40% 50% 60% 70%
laboratories. Laboratories (%)

facturer and 56 (32%) using third party QC from these indicated that both College of American Pathol-
another manufacturer’s source. Additionally, 13 (7%) ogist (CAP) and LA requirements were used, one
participants reported that their laboratory used com- (0.06%) participant reported use of IQMH Consensus
mercial QC from their analyzer’s manufacturer and Practice Recommendations, two (1%) participants
another source. A minority (22; 12%) of laboratories used recommended or regulatory guidelines which
used an in-house QC of pooled patient plasma. Of were not specified, and one laboratory (0.06%) used
these, 18 (10%) used patient controls at times by itself the CLSI guideline H47-A2. Five (3%) participants
or four (2%) in addition to running a commercial QC reported that QC testing frequency was based on past
material. Two (1%) participants indicated that they experience, and one participant reported that the
used in-house material, one for checking precision length of workday dictated QC frequency. In six (3%)
and the other ran in-house normal plasma pool every organizations, the medical laboratory director or he-
2 h for monitoring the stability of the analytical sys- matopathologist determined frequency of QC runs.
tem between commercial QC runs. All laboratories Other factors considered included the following: the
ran at least one normal and one abnormal CQC daily. stability of the test (46; 27%), clinical impact of incor-
Almost half of the participants reported that their lab- rect test result (42; 25%), and the number of samples
oratory ran one normal and one level of abnormal potentially requiring repeat analysis if QC fails
commercial QC, PT/INR: 81 (47%) and APTT: 79 (17; 10%) (Figure 2).
(46%). The other participants reported that they ran The results were similar for how laboratories estab-
one normal and two levels of abnormal control, PT/ lish the frequency of commercial QC for the APTT test.
INR: 93 (53%) and APTT: 92 (54%). Of 169 laboratories who responded, the majority 119
(70%) based QC frequency on manufacturer recom-
mendations. Other factors considered included the fol-
Frequency of internal quality control runs
lowing: the stability of test (45; 27%), clinical impact
Laboratories were asked how they established the fre- of incorrect test result (42; 25%), and the number of
quency for running commercial controls for PT. Of samples potentially requiring repeat analysis if QC fails
the 172 laboratories who submitted answers for this (18; 11%). Seventy-two (41%) participants selected
question, 122 (71%) follow manufacturer recommen- ‘other’ as an option in the survey with 50 (30%) who
dations for establishing QC frequency of PT runs, 72 reported that they establish frequency of QC runs
(42%) chose alternate options, with 31 (18%) select- based on IQMH LA requirements, 2 (1%) of these
ing one other option, and 41 (24%) selecting multiple indicated that both College of American Patholo-
options. Fifty (29%) laboratories indicated that they gist (CAP) and LA requirements were used, one
use the IQMH Laboratory Accreditation (LA) Stan- (0.06%) participant reported use of IQMH Consensus
dards for determining QC run frequency, two (1%) of Practice Recommendations, two (1%) participants

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738
732 A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES

APTT PT/INR
IQMH laboratory accreditation, CAP, CLSI or
other guidelines
Stability of parameter/test to allow for
retesting
Clinical impact of incorrect test result if
error is undetected
Number of samples potentially requiring
repeat analysis

Manufacturer's recommendation Figure 2. Sources used in


0% 10% 20% 30% 40% 50% 60% 70% 80%
determination of frequency of
Percent of participant laboratories
IQC runs.

used recommended or regulatory guidelines which commercial controls and at the end of each analytical
were not specified, and one laboratory (0.06%) used run. Twenty-two (13%) participants noted that they
the CLSI guideline H47-A2. Five (3%) participants determined the stability time frame for in-house QC
reported that QC testing frequency was based on past materials, and six (3%) participants did not. For the
experience, and one participant reported that the APTT, 147 (86%) did not use in-house patient QC
length of workday dictated QC frequency. In six (3%) materials. However 15 (9%) laboratories specified that
organizations, the medical laboratory director or he- they perform in-house QC with each defined analyti-
matopathologist determined frequency of QC runs cal run, seven (4%) when an analyzer has not been
(Figure 2). used for a defined time, three (2%) at the beginning
Most laboratories responded that the commercial of each day, two (1%) at the beginning of each shift,
QC for the PT and APTT was run at the beginning of and three (2%) had other responses which included
each shift: 119 (69%) and 116 (68%), respectively, every one or 2 h between commercial controls, and at
and when the analyzer had not been in use for a least three times during a shift. Nineteen (11%) par-
defined time period: 44 (25%) and 43 (25%), respec- ticipants noted that they determined the stability time
tively. A small number of participants only ran com- frame for in-house QC materials, and six (3%) partici-
mercial QC for the PT and APTT at the beginning of pants did not.
the day: 12 (7%) and 11 (6%), respectively. In addi- Participants were asked how many patient samples
tion, some laboratories defined other times that com- they would run before retesting the commercial QC
mercial QC would be run for the PT 102 (59%) and for both the PT and APTT. The most common practice
APTT 95 (56%). These times included the following: was to run the commercial QC following an analysis
postmaintenance, after a reagent change, in the mid- batch of <20 samples: PT (75;43%), APTT (97;56%),
dle of each shift, with every repeat sample, following followed by 21–40 samples: PT (33;19%), APTT
a service on the instrument, every 100 patient sam- (31;18%); 41–60 samples PT (21;12%), APTT
ples, or at time prespecified intervals of either 1, 2, 3, (12;7%); 61–80 samples: PT (8;5%), APTT (6;3%);
or 4 h. and >80 samples: PT (14;8%), APTT (4;2%) (Fig-
The majority (146; 84%) of participants reported ure 3). These results correlate well to the average
that they do not use in-house patient QC materials number of patient samples being run per day (Fig-
for PT analysis. However, 16 (9%) laboratories speci- ure 1).
fied that in-house QC was performed with each
defined analytical run, nine (5%) reported in-house
Monitoring internal quality control results
QC was performed when an analyzer had not been
used for a defined time, three (2%) ran at the begin- All participants reported that preset QC limits were
ning of each day, two (1%) at the beginning of each used for coagulation tests, but there was a variation in
shift, and six (3%) had other responses which stated the procedures used by Ontario laboratories to deter-
that in-house QC was run every one or 2 h between mine QC limits. The majority (114; 66%) of laborato-

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738
A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES 733

APTT PT/INR
Other - every 2, 4 or 8 h or after
reagent change
Only run commercial quality control(s)
once a day
>80 patient samples

61 – 80 patient samples

41 – 60 patient samples

21 – 40 patient samples
Figure 3. Number of patient
samples run prior to retesting <20 patient samples
the commercial QC. 0% 10% 20% 30% 40% 50% 60%

ries used the standard deviation (SD) of the QC mate- 12x, five (15%) use 8x, 1 (0.5%) uses 7x, one (0.5%)
rial to set the QC limits. However, many participants 12x and 7T (seven consecutive data points for a single
indicated a combination of methods for determining level of control show an increasing or decreasing pat-
QC limits was used, including precision goals from tern), one (0.5%) uses 7T, and one (0.5%) uses 14s
manufacturer recommendations (79; 46%), published and 15s. Two (1%) participants that selected other did
precision goals (45; 26%), and allowable performance not specify the rules that they used.
limits (APL) (62; 36%). Of the 107 laboratories that Laboratories used various methods to monitor the
reported their precision goals, the majority (94; 88%) IQC results: 114 (66%) used a QC program incorpo-
of participants use the limit of ≤5% coefficient of vari- rated into their LIS, 78 (45%) used a QC software
ation (CV) for both the PT and APTT. The remaining from the QC provider, 69 (40%) used the QC software
laboratories (13; 12%) reported a wide variation of on the coagulation analyzer, 21 (12%) used manual
precision goals ranging from 6% to 25% for both PT documentation (pen and paper), and five (3%) used
and APTT (Figure 4). an in-house computer software program built in MS
Nine (5%) participants reported that their labora- Excelâ.
tory used only one QC rule to determine acceptability
for reporting patient results; of these, five (56%) use
Actions taken when internal quality control fails
12s and four (44%) use 13S. All other laboratories use
multirule procedures (Table 1). The majority (165; Laboratories were asked to choose from a list of trou-
95%) of laboratories use a combination of the 12s, 13S, bleshooting actions taken when QC fails. Participants’
22s, R4s, 41s, and 10x rules. Additionally, 34 (20%) lab- responses are summarized in Table 2. Almost all of
oratories reported also using the following QC rules: the responding laboratories (greater than or equal to
14 (8%) laboratories use two of three 2SD rule (two 90%) repeat the QC and, if it passes, patient results
of last three results outside of 2SD), nine (5%) use are reported; participants responded that they would

Precision goals -PT/ INR Precision goals -APTT


80 80.0
Number of laboratories
Number of laboratories

60 60.0

40 40.0

20 20.0

Figure 4. Precision goals for PT 0 0.0


3 3.5 4 4.5 5 7 8 10 15 20 25 1.4 2 3 3.5 4 5 6 8 10 15 20 25
and APTT.
Precision goals (%) Precision goals (%)

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738
734 A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES

Table 1. QC acceptability rules used by Ontario Table 3. Documentation of investigation of failed QC


Laboratories
Number of Percent
No. of Laboratories Action Laboratories (%)
(total number of
laboratories = 174) % of Laboratories Record level and lot 165 95
number of QC material
12S 153 88 Analyzer name 154 89
13S 156 90 Rule violated 153 88
22S 152 87 Lot number of reagents 124 71
R4S 129 74 Date and time testing 109 63
41S 93 53 is resumed
10x 64 37 Root cause analysis 102 59
Other 34 20 Authorization for 72 41
Total 174 testing to resume
Lot number of calibrators 39 22

Table 2. Troubleshooting actions when QC fails


Participants were asked what information would be
Number of Percent
documented in the investigation of failed QC, and
Action Laboratories (%)
these responses are summarized in Table 3. Almost all
Repeat the QC, and if it 168 97 (170; 98%) participants document actions taken for
passes, report results troubleshooting, but only 24 (14%) of participants
Open new QC 166 95
include all details listed in their investigation reports.
Look for trending 161 93
Discontinue testing until 156 90 The majority (165; 95%) of laboratories record the
controls are within limits level and lot number of QC material, analyzer name,
Repeat all patient samples 73 42 the rule violated, and the lot number of reagents, and
from last acceptable QC the date and time testing was resumed and the root
Other actions for failed 44 25
cause analysis was performed. A minority (72; 41%)
QC results:
Change reagent 17 10 of laboratories document authorization for testing to
Repeat consecutive samples 8 5 resume and lot number of calibrators.
until no discrepancy Of 168 laboratories that responded to a question
is observed regarding using risk assessment for improvements to
Contact technical 5 3 patient safety, 120 (71%) reported that their labora-
service
tory used risk management techniques. In the coagu-
Calibrate if no analyzer 4 2
issue has been identified lation laboratories, QC issues were used by 131 (77%)
Use options depending 4 2 of participants as a mechanism for identification of
on the failure potential risks for patient safety.
Repeat effected samples 2 1
Confirm calibration 2 1
Repeat random samples 1 0.5 DISCUSSION
Inform charge technologist 1 0.5
A high percentage of medical decisions are influenced
by laboratory results [1]. The results of coagulation
open a new vial of QC, would look for trending of tests are essential for the diagnosis of coagulation dis-
results, would discontinue testing until QC was within orders, assessment of a patient’s risk for bleeding, and
acceptable limits, and 73 (42%) of participants would to monitor effectiveness of some anticoagulant thera-
repeat all patient samples from last acceptable QC. pies [3, 9]. As such, the QA practices for QC in coagu-
There were 44 (25%) participants that reported they lation laboratories, as in all clinical laboratories, have
would take other actions for failed QC results, which a significant role in ensuring accurate and precise
are itemized in Table 2. patient test results [9]. A world leader in quality con-

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738
A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES 735

trol, Dr. James Westgard, stated ‘Do the right QC sample can be used as a risk mitigation tool and is a
right’ and when this is applied, there is an assurance good verification of precision. Results from a stable
of the quality of test results [11]. The incorrect use patient sample are compared against previous results
and monitoring of quality control strategies can be and should be within predetermined and validated
associated with serious patient risk when a wrong limits. If the difference exceeds the limits, an investi-
result is reported and can add additional costs associ- gation should be performed to determine the cause.
ated with performing unnecessary repeat or follow-up The use of patient samples may also be used for inter-
tests [1]. Performing the QC ‘right’ is a shared respon- and intralaboratory agreement between different ana-
sibility within the laboratory from the bench technol- lyzers and they are useful to eliminate matrix effects
ogist to laboratory supervisors, managers, and when reagents are changed [14–17]. However, labora-
directors whose responsibility lies in overall quality tories need to consider sample stability and the impact
assurance within a quality management system [1, 2]. on coagulation test results over time (i.e., loss of fac-
Coagulation testing is considered to be one of the tor activity that may negatively impact reproducibility
most complex in diagnostic laboratory testing and of the assays) [18].
needs to be well controlled for providing reliable It is important to determine the stability of control
results for screening, diagnosis, and monitoring ther- materials and effects of different reagents that may be
apy of hemostasis [12]. The summary of the results used in testing. Commercial control materials offer the
from the 2012 IQMH patterns-of-practice survey advantage of longer storage stability but do have the
shows significant variability of IQC practices across disadvantage of possible matrix effects and may be
Ontario laboratories and suggests that standardization prone to errors from inaccurate reconstitution of
could be beneficial for detecting errors, minimizing lyophilized samples [2, 15]. QC material may slightly
false rejection of test results, and optimizing produc- degenerate after opening, and a small amount of drift
tivity [9, 13]. may be seen before outdating of the QC material [17].
Choosing the ‘right’ QC is the first step for a good The frequency of running controls is also an impor-
IQC program [10]. In the results of our survey, all tant part of initiating a good QC program [8]. The
participating laboratories reported that they used com- majority of coagulation laboratories in Ontario indi-
mercial QC materials with approximately one-third cated that commercial QC was run at the beginning of
reported using third party QC and a minority of labo- each shift with 25% of participants reporting that they
ratories also using an in-house QC comprised of also ran their QC if the analyzer has not been in use
pooled patient plasma. Third party QC has the advan- for a certain length of time. A small number of partic-
tage of being an independent assessment for the ana- ipants only ran commercial QC at the beginning of
lytical performance over multiple reagents and the day. Additionally, participants reported that QC
calibrator lot number changes as it is not optimized may be run following maintenance, reagent change,
for any specific instrument or reagent system. Choos- middle of each shift, and with every repeat sample.
ing the appropriate QC material is essential to ensur- From the results of our 2012 survey, the frequency of
ing optimal error detection [8]. As there is no perfect running QC is largely based on manufacturer recom-
control material, several factors should be taken into mendations. However, other factors considered
consideration when choosing QC materials, such as included the following: stability of test, clinical impact
matrix, cost, stability, ease of use, and the laboratory’s of incorrect test results, the number of samples poten-
intended application [8, 10, 11]. Part of choosing the tially requiring repeat analysis, directive from the lab-
right QC is to include QC levels at or near to clinical oratory director, and past practices and/or the length
decision points as recommended in ISO 15189 [7]. of the workday. About one-third of the participants
In addition to commercial controls, patient controls also reported using laboratory accreditation require-
can be run with each batch, or periodically during ments and regulatory guidelines for determination of
analytical runs, to show reproducibility over time. IQC frequency. The most common guidance cited in
This offers an economical approach to rapidly identify- the laboratory responses was IQMH. Both IQMH and
ing any issue that may develop between running CLSI recommend running a minimum of one normal
commercial controls. Repeat testing of a stable patient and one abnormal control during every 8 h of contin-

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738
736 A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES

uous operation with the level of controls spanning the closeness of agreement between replicate measure-
clinically relevant reporting range. They both also rec- ments. It is expressed quantitatively in terms of
ommend running a control specimen immediately fol- imprecision either as the standard deviation (SD) or
lowing instrument maintenance, reagent change, or a the coefficient of variation (CV) of results in a set of
lot number change. Additionally CLSI suggests run- replicate measurements [8, 15, 16]. Most often QC
ning QC more frequently in high-volume testing cen- limits are based on the standard deviation of the QC
ters by running one of the two or three controls levels results but may consider other sources such as preci-
once every 4 h [19, 20]. sion goals provided by manufacturers or published lit-
Manufacturers may offer recommendations for the erature or may be determined from allowable
frequency of running QC, but it is the responsibility performance limits. Clinicians expect that the labora-
of the laboratory to evaluate and validate the effec- tory report results that are accurate and reproducible
tiveness of the control process used in the laboratory in order to allow the correct medical decision to be
[3, 11]. The length of the analytical run is at the dis- made [23]. It is the responsibility of the laboratory to
cretion of the laboratory and the medical director; determine the appropriate allowable performance lim-
however, the decision for length of the analytical run its for each test [24]. In order to do this, laboratories
should take into consideration test volumes and time need to understand how precision goals were derived,
between samples. Periodic reassessment of the analyti- for instance limits will be more stringent if validated
cal performance should be at a frequency that will on the same instruments under the same conditions,
monitor the system and increase the likelihood of or may be extremely wide when validated under dif-
detecting system errors, decreasing the potential ferent conditions, such as different laboratories with
release of incorrect results [3, 8, 11, 21, 22]. The con- different operators or when comparing different
cept of an analytical run is the time interval or the instrument/reagent combinations [1, 25].
number of measurements over which accuracy and The majority of participants used the limit of ≤5%
precision are expected to remain stable [15]. coefficient of variation (CV) for both the PT and APTT.
All Ontario laboratories reported the use of preset However, our survey demonstrated a wide variation
QC limits for the assessment of QC results. The major- of precision goals ranging from 6% to 25% for both
ity of survey participants use a combination of the 12s, PT and APTT.
13s, 22s, R4s, 41s, and 10X rules. Additionally, laborato- The majority of laboratories reported they would
ries responded also using the following QC rules: 2 of repeat the QC when values are outside of the target
3 2SD rule (2 of last 3 results outside of 2SD); 12x, 8x, limits. Most would open a new QC sample/vial, and if
12x, and 7T (seven consecutive data points for a single the repeat passes, patient results would be reported.
level of control show an increasing or decreasing pat- Other strategies for troubleshooting failed QC included
tern); and 7T, 14s, and 15s. Monitoring of QC should looking for trending, discontinue testing, and repeat-
be a planned process and reviewed periodically for ing all patient samples from last acceptable QC. The
detection of systematic and random errors. A system- development of a good QC plan should include identi-
atic error is a change in a test system that is always in fication of a problem in an assay and maintaining an
one direction and causes a shift in the mean. System- accurate documentation of actions taken [15].
atic errors are detected by the 22s, 2 of 32s, 31s, 41s, 6X, The concept of risk management analysis has
and 10X. A systematic error could occur after a QC become part of the medical laboratory’s day-to-day
sample is tested and could remain undetected for a operations for improved patient safety by reducing the
period of time before the next QC measurement. A risk of patient harm [23–27]. The laboratory must
random error is inherent in any process. It can be identify events that could cause the testing process to
either positive or negative and results in a change in be susceptible to errors [3, 23]. A comprehensive dis-
precision. Random errors are detected by the 13s, 15s, cussion on risk management is beyond the scope of
and R4s rules [8, 11, 15]. These errors can be detected this document; however, the QC process should
on careful review of quality control charts. include identifying the potential hazards and fre-
Precision goals are often used to determine the quency of occurrence, analyze the risk of each hazard,
reproducibility of results. Precision is defined by the implement controls, and monitor the process [23, 25].

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738
A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES 737

Table 4. Recommendations for IQC in Coagulation

Recommendations

Selection of control material


The QC material needs to reflect the matrix and range of analytes commonly reported for patient samples and at or
near the clinical decision values.
A minimum of two levels, at least one normal and one abnormal must be run.
It is acceptable to use patient controls in combination with commercial controls.
Commercial controls that are lyophilized or freeze dried must be reconstituted as per manufacturer’s directions.
QC should demonstrate stability over the claimed shelf life and the interval after opening or reconstitution.
Define the length if the analytical run and frequency of controls
The length of the run should be defined for each specific analytical system and procedure.
The commercial QC material must be run at least once per day. It is recommended to run controls for routine
coagulation with each batch or at appropriate times throughout a 24-hour period.
Use of Quality Control Rules
Select at minimum the 12s, 13S, 22s, R4s, 41s and 10X QC rules to detect clinically important errors and do not allow
for false rejections.
The 12s rule is a warning rule; it is not recommended to be used as a rejection rule, as it causes the most false
rejections [1].
Actions when quality control fails
Failure of a QC rule requires that patient testing should be stopped. It is common practice to hold and not release
patients’ results until the QC results have been evaluated, thus reducing the risk of reporting erroneous results.
Depending on the size of the batch, the placement of quality control samples at the beginning and also at the end of
an analytical run may help detect shifts in performance.
In laboratories that use auto-validation/auto-verification, results may be released as soon as they are run. Good
practice includes running controls at the beginning of the analytical run to ensure the system is performing
correctly and to include additional controls throughout the analytical run to detect immediate errors and to
monitor changes in the analytical performance and environment.
In determining the intervals, keep in mind that if there is a failure of the QC run, and investigation demonstrates an
analyzer fault, all patient samples since the previous “good” quality control event should be rerun.
Risk management
Document an accurate history of actions that have taken place in the testing system.
Maintain records of all actions that have impact on performance of the system.
Monitor and perform periodic review of records for troubleshooting and to prevent occurrence of QC failures.
Risk analysis should include identifying where failures have occurred, the potential hazards and frequency of
occurrence, analyze the risk of each hazard and determine the QC required to minimize the risks and monitor the
process [3, 5, 8].

One approach is to map the QC process in order to detected, all testing on that analyzer for that assay
identify where QC failures may occur and brainstorm should be discontinued and effective corrective actions
potential causes and determine the amount of QC implemented before resuming testing. Protocols
needed to minimize patient risk [23, 25]. Documenta- should be implemented to assess the impact on previ-
tion records should include information on scheduled ously tested specimens since the last successful QC
maintenance, reagent lot number changes and expira- event [3, 15].
tion dates, calibration records, quality control results,
summary of statistics including monthly means and
CONCLUSIONS
standard deviations, as well as cumulative means,
standard deviations, and control target limits. In addi- Our data illustrate the wide variability in QC practices
tion, records should include QC problems such as for coagulation testing in Ontario. Laboratories should
trends and shifts, corrective actions taken, and trou- establish an IQC program that ensures patient safety
bleshooting reports [3, 13, 23]. When a QC failure is by monitoring, detecting, and minimizing errors that

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738
738 A. MCFARLANE ET AL. | INTERNAL QUALITY CONTROL PRACTICES IN ONTARIO COAGULATION LABORATORIES

may cause harm to a patient and documentation of results. This review uses survey results from Ontario
investigation of failed QC. A good QC program should laboratories to provide guidance and recommenda-
include a ‘minimum’ of two IQC samples, normal and tions (Table 4) for clinical laboratories in the planning
abnormal, to optimize the probability of error detec- and implementation of a good IQC program in coagu-
tion, but minimize the probability of false error detec- lation testing. Standardization of IQC practice within
tion. It is important to establish a review system to and across laboratories will help provide a means for
include benchmarks to monitor the QC program for detecting important errors and minimize false rejec-
its effectiveness over time for confidence that the ana- tion of test results to ensure quality patient care.
lytical system is producing accurate and good quality

REFERENCES Definitions; Approved Guidelines, 3rd edn, on coagulation testing. Int J Lab Hematol
C24–A3. Wayne, PA: NCCLS, 2006. 2015;37:551–9.
1. Carlson RO, Amirahmadi F, Hernandez JS. 9. Kitchen S, Olsen JD, Preston FE. Quality in 19. CLSI. One-stage Prothrombin Time (PT)
A primer on the cost of quality for Laboratory Hemostasis and Thrombosis. Test and Activated Partial Thromboplastin
improvement of laboratory and pathology West Sussex UK: Wiley-Blackwell, 2009. Time (APTT) Test; Approved guideline, 2nd
specimen processes. Am J Clin Pathol 10. International Organization for Standardiza- edn. CLSI document H47-A2. Wayne, PA:
2012;138:347–54. tion ISO 17043:2010 Conformity Assessment – Clinical and Laboratory Standards Institute;
2. Favaloro EJ, Funk DM, Lippi G. Pre-analyt- General Requirements for Proficiency Testing 2008.
ical variables in coagulation testing associ- Available from http://www.iso.org 20. IQMH –OLA requirements V.6
ated with diagnostic errors in hemostasis. 11. Westgard JO. Basic QC Practices 3rd edn. 21. Parvin CA, Robbins S. Evaluation of the
Lab Med 2012;43:1–10. Madison, WI: Westgard QC Inc.; 2010. performance of randomised versus fixed
3. Bonar R, Favaloro EJ, Adcock DM. Quality 12. Lippi G, Plebani M, Simundic AM. Quality time schedules for quality control proce-
in coagulation and haemostasis testing. in laboratory diagnostics: from theory to dures. Clin Chem 2007;53:575–80.
Biochem Med 2010;20:184–99. practice. Biochem Med 2010;20:126–30. 22. Hatjimihail AT. Estimation of the optimal
4. College of American Pathologists, Trouble- 13. Sonntag O. Quality in the analytical phase, Statistical Quality Control Sampling Time
shooting problems with the Prothrombin Special issue: quality in laboratory diagnos- intervals using a residual risk measure,
Time (Protime/INR), the activated Partial tics: from theory to practice. Biochem Med PLoS One, 2009;4:e5770
Thromboplastin Time (aPTT) and the 2010;20:147–53. 23. National Committee for Clinical Laboratory
Fibrinogen Assay. Available online last 14. National Committee for Clinical Laboratory Standards. Laboratory Quality Control
accessed November 21, 2013 at http:// Standards. Evaluation of Matrix Effects; Based on Risk Management; Approved
www.cap.org/apps/docs/laboratory_accredi- Approved Guidelines, 2nd edn. EP14-A2. Guideline, EP23-A. Wayne, PA: NCCLS,
tation/build/pdf/coag_troubleshooting.pdf Vol 25, No 4 Wayne, PA: NCCLS; 2005. 2011.
5. Koepke JA. Technologies for coagulation 15. WHO, Process control: quality control for 24. Erasmus RT, Zemlin AE. Clinical audit in
instruments. Lab Med 2000;31:211–5. quantitative tests, Laboratory Quality the laboratory. J Clin Pathol 2009;62:
6. Zhao Y, Yang L, Zheng G, Cai Y. Building Management System: Handbook. Geneva, 593–7.
and evaluating the autoverification of coag- Switzerland: WHO Press, World Health 25. Powers D. Laboratory Quality Control
ulation items in the laboratory information Organization, 2011;Version 1.1: 77–90. Requirements Should be Based on Risk
system, Clinical Laboratory, 2013 e-publica- 16. CLSI. Evaluation of Precision Performance Management Principles. Lab Med,
tions available online last accessed at of Quantitative Measurement Methods; 2005;36:633–8.
November 25 available at: https://www. Approved Guidelines, 3rd edn. CLSI docu- 26. Yundt-Pacheco J, Parvin CA. Validating the
clin-lab-publications.com/files/eaop/2014_01+ ment EP5-A3, Wayne, PA: Clinical Labora- performance of QC procedures. Clin Lab
02/130109-Zhao.pdf tory Standards Institute; 2014. Med 2013;33:75–88.
7. International Organization for Standardiza- 17. Jennings I, Kitchen DP, Woods TAL, 27. Tankovic AK, Silvestri J, Mails M, Najork
tion ISO 15189:2012(E). Medical Laborato- Kitchen S, Preston FE, Walker ID. Stability C. Total quality in laboratory diagnostics:
ries – requirements for quality and of coagulation proteins in lyophilized the role of commercial companies, Special
competence, Available from http://www.i- plasma. Int J Lab Hematol 2015;37:495– issue: quality in laboratory diagnostics:
so.org 502. from theory to practice. Biochem Med,
8. National Committee for Clinical Laboratory 18. Gosselin RC, Honeychurch K, Kang HJ, 2010;20:207–14.
Standards. Statistical Quality Control for Dwyre DM. Effects of thawing conditions
Quantitative Measurements: principles and

© 2015 John Wiley & Sons Ltd, Int. Jnl. Lab. Hem. 2015, 37, 729–738

You might also like