Models with
Mechanical Air Movement
Sterilization - Disinfection
Sterilization means to free objects from breedable The only result of disinfection is that an object
microbes; in the process not only the vegetative, but cannot disinfect any more. It is then free from
the resistive forms of germs must be eliminated. vegetative forms of disease-causing agents, but not
Sterilization by hot air requires a minimum free from bacterial spores.
temperature of 180°C (German DIN 58947). During Medical instruments which come into contact with
sterilization, spores of type A, B, C, D (according to blood or which injures the integrity of human body
the German federal public health department) or surface, must therefore always be sterilized before
spores of resistivity level III (according to Prof. use (tetanus and danger of gangrene infection) –
Konrich) have to be killed. disinfection is not sufficient.
page 1
1. Operation of the apparatus
Set up the temperature regulator (1) to the sterilizing for a possible adjustment of the regulating knob (see
temperature of 180°C. If the indicated value for "Hints for fault corrections" below).
temperature is different from the one set up, check
1
2
4
6
page 2
Operating Instructions
The given operating times are based on sterilizations Use containers from aluminium not stainless steel
of medical instruments with trays or open containers. container. Determine the additional time by a test
Sterilization processes with closed containers need run. The safety margin should be 10%.
longer operating times (see below).
2.2. German VDE-requirements _________________________________________
In accordance with the valid VDE-requirements, this It may be repaired only by the manufacturer or a
autoclave is not suitable for areas with danger of licensed service unit.
explosion.
3. Attention! Warning
• Before opening the door switch off the appliance, • During and after the sterilisation process some
either using the main switch (6) or by turning the parts of the housing may become very hot and
time switch (5) to “0”. The signal lamp (4) is no could cause burns. Danger!
longer on. • This appliance is not suitable for the sterilization
• Be careful when loading and unloading the hot of liquids and for the warming up of rinse
sterilizer. Trays and sterilised items are hot! solvents !
• The ventilator only comes to rest some 45 secs
after switching off the appliance.
4. Attention! Misoperation
The hot-air sterilisszation method is fully accepted about hot-air sterilizers still show a growth of testing
(German DIN 58947). Nevertheless, this method is germs. A more detailed examination demonstrates
still criticized because bacteriological examinations that this fact is mostly caused by misoperation.
page 3
6. Hints about fault corrections
6.2. Temperature set up with the regulation knob does not tally with
indication of the thermometer _______________________________________
Hot-air sterilizers from MELAG are delivered with in order to obtain a higher temperature, or
temperature calibrated at 180°C by adjusting the counter-clockwise for lower temperature. Provi-
regulator axis so that the thermometer really sional markings on the scale will help adjustment.
shows a value of 180°C. Then the thermostat A correct calibration is carried out as follows:
knob is sticked on the axis so that the white
Remove the grey cap from the thermostat knob.
indicator of the knob points to the 180°C- mark on
After loosening the clamp screw below, the knob
the scale. In this position, the thermostat knob is
can be removed from the axis. Close the door and
fixed to its axis by fastening the clamp screw.
switch on the sterilizer. Now adjust the tem-
If the thermostat knob is unintentionally be turned perature regulator by turning the axis (perhaps
with force beyond the resistance due to improper with small pliers) so that the thermometer shows
treatment, e.g. during cleaning or unpacking of the the desired temperature (e.g. 180°C) after con-
gadget, or if the screw is loosened in the taper, the clusion of pre-heat and equalization time. Put the
knob can slip on the regulator axis. Respectively, thermostat knob back on the axis so that the white
the real temperature inside the vessel (indicated indicator points to a value on the scale which
on the thermometer) is different from the one set agrees with the temperature shown on the
up on the regulator scale. thermometer. Finally screw the knob's screw tight
A quick correction can easily be achieved by without changing the position of the axis.
turning the thermostat knob either little clockwise
7. Appendix