Operating Instructions

HOT- AIR STERILIZERS


DRYING CABINETS

Models with
Mechanical Air Movement

Type 215 Type 225 Type 255

Sterilization - Disinfection
Sterilization means to free objects from breedable The only result of disinfection is that an object
microbes; in the process not only the vegetative, but cannot disinfect any more. It is then free from
the resistive forms of germs must be eliminated. vegetative forms of disease-causing agents, but not
Sterilization by hot air requires a minimum free from bacterial spores.
temperature of 180°C (German DIN 58947). During Medical instruments which come into contact with
sterilization, spores of type A, B, C, D (according to blood or which injures the integrity of human body
the German federal public health department) or surface, must therefore always be sterilized before
spores of resistivity level III (according to Prof. use (tetanus and danger of gangrene infection) –
Konrich) have to be killed. disinfection is not sufficient.

What can be sterilized in hot-air?


In hot-air, it is possible to sterilize objects made of immediately be disinfected and then thoroughly be
incombustible (inorganic) materials, such as metal, cleaned. Before placing them in the sterilizer, the
glass, porcelain, stoneware, enamelware, etc., e.g. items have to be dried. Do not place items on
instruments, syringes, cannulas, and so on (pay cellulose because this might lead to build-up of heat
attention to the manufacturer's instructions). After and thus to an obstruction of heat egalization in the
use, the instruments, syringes, etc. should apparatus.

page 1
1. Operation of the apparatus
Set up the temperature regulator (1) to the sterilizing for a possible adjustment of the regulating knob (see
temperature of 180°C. If the indicated value for "Hints for fault corrections" below).
temperature is different from the one set up, check

1.1. Switching on power _______________________________________________


The apparatus' with timer is switched on by adjusting to "I"), the sterilizer is turned on for permanent
the timer (5) to the time given in the operating time operation.
table. The permanent switch (6) of sterilizers of type The light of signal lamp (4) indicates that power is
215-255 must always be in position "0", if the timer is on.
in operation.
The signal lamp (2) optically shows the automatic
For longer processes (or if the timer is defective), working procedure of the temperature regulator. It
power can be switched on by using switch (5) for the keeps the adjusted temperature to a constant level
required time. The timer is out of operation. By by turning the heating on or off (not current in
turning the timer knob counter-clockwise (to the left general). Lamp (2) goes on and out respectively.

1.2. Switching off power ______________________________________________


The apparatus is automatically switched off at the Permanent operation of sterilizers of type 71 and
end of the timer (5). type 91 is ended (timer in position "I") by setting the
If a sterilization process was carried out by using the timer knob back to "0". Both signal lamps (2 and 4)
permanent switch (6), set it back to position "0". go out if power is turned off.

1.3. Mechanical Air Movement __________________________________________


Sterilizers with mechanical air movement have on by the turning on of power and switched off re-
motor-driven fans which are automatically switched spectively.

1.4. Drying procedure _________________________________________________


Set up temperature to 100°C by turning the regula- As far as sterilizers of type 215-255 are concerned,
tion knob. Shake out pipettes to be dried and lay the door can be engaged into a position of the door
them in not too wet and – if possible – a little sloping. hinge where the door will be kept open to enable the
Escape of damp air is possible as follows: drying process.

2. Description of the apparatus


2.1. Front view _______________________________________________________
Type 215 - 255

1
2
4
6

1 Thermostat 3 Thermometer 5 Timer - On-Off Switch


2 Signal lamp heating 4 Signal lamp main 6 Permanent switch

page 2
Operating Instructions

Table of Operating Times

Apparatus Maximum load Mechanical


Type (without basket/trays) air movement

215 / 225 / 255 70 Min. (3 trays)


3,0 kg
85 Min. (6 trays)

The given operating times are based on sterilizations Use containers from aluminium not stainless steel
of medical instruments with trays or open containers. container. Determine the additional time by a test
Sterilization processes with closed containers need run. The safety margin should be 10%.
longer operating times (see below).
2.2. German VDE-requirements _________________________________________
In accordance with the valid VDE-requirements, this It may be repaired only by the manufacturer or a
autoclave is not suitable for areas with danger of licensed service unit.
explosion.

3. Attention! Warning
• Before opening the door switch off the appliance, • During and after the sterilisation process some
either using the main switch (6) or by turning the parts of the housing may become very hot and
time switch (5) to “0”. The signal lamp (4) is no could cause burns. Danger!
longer on. • This appliance is not suitable for the sterilization
• Be careful when loading and unloading the hot of liquids and for the warming up of rinse
sterilizer. Trays and sterilised items are hot! solvents !
• The ventilator only comes to rest some 45 secs
after switching off the appliance.

4. Attention! Misoperation
The hot-air sterilisszation method is fully accepted about hot-air sterilizers still show a growth of testing
(German DIN 58947). Nevertheless, this method is germs. A more detailed examination demonstrates
still criticized because bacteriological examinations that this fact is mostly caused by misoperation.

5. Fault No. 1: Operating time too short!


The operator assumes the correct fact that a tem- the killing time begins which endures 30 minutes
perature of 180°C for a period of 30 minutes takes (with 180°C). This period contains an extra charge
effect on the items to be sterilized. But the fault lies for safety (German DIN 58947, part 5).
in the exclusive observation of the thermometer. As All these conclusions of times are included in the
soon as the thermometer shows a temperature of operating times given above (see table). The operat-
180°C, the operator wrongly begins to count the 30 ing times are based on sterilizations of medical
minute period after which the sterilizer can be instruments under use of trays or open containers.
switched off. This procedure only considers the pre- Sterilization processes with closed containers need
heat time but not the equalization time. It is wrong to longer operating times. If the items are sterilized in
assume that the killing time starts immediately after closed cassettes made of aluminium with thin walls,
the conclusion of the pre-heat time. Between killing the operating times have to be elongated for 15
and pre-heat time, the operator has to take the minutes. On principle, we recommend not to use
equalization time into account. This is the time closed containers. For example, we have measured
needed for the required temperature reaching all a temperature of only 160°C inside of plates after 3½
places of the apparatus and of the items to be steril- hours!
ized. Only after conclusion of the equalization time

page 3
6. Hints about fault corrections

6.1. Timer does not automatically switch off power ________________________


If the timer is defective, the sterilizer can still be ends from the flat plugs of the timer and unscrew the
operated in a half-automatic way with the help of the timer from the front scale plate. Connect and install
permanent switch (6). Before exchange of the timer, the new timer in the same way.
it is absolutely necessary to pull out the main plug. A new timer can only be ordered at specialist shops.
Loosen the screw of the timer knob and remove the For quick completion, it is necessary to know the
knob. Unscrew the front scale plate and remove it type of sterilizer and its work number.
with care. Then remove the stick hulls of the cable

6.2. Temperature set up with the regulation knob does not tally with
indication of the thermometer _______________________________________
Hot-air sterilizers from MELAG are delivered with in order to obtain a higher temperature, or
temperature calibrated at 180°C by adjusting the counter-clockwise for lower temperature. Provi-
regulator axis so that the thermometer really sional markings on the scale will help adjustment.
shows a value of 180°C. Then the thermostat A correct calibration is carried out as follows:
knob is sticked on the axis so that the white
Remove the grey cap from the thermostat knob.
indicator of the knob points to the 180°C- mark on
After loosening the clamp screw below, the knob
the scale. In this position, the thermostat knob is
can be removed from the axis. Close the door and
fixed to its axis by fastening the clamp screw.
switch on the sterilizer. Now adjust the tem-
If the thermostat knob is unintentionally be turned perature regulator by turning the axis (perhaps
with force beyond the resistance due to improper with small pliers) so that the thermometer shows
treatment, e.g. during cleaning or unpacking of the the desired temperature (e.g. 180°C) after con-
gadget, or if the screw is loosened in the taper, the clusion of pre-heat and equalization time. Put the
knob can slip on the regulator axis. Respectively, thermostat knob back on the axis so that the white
the real temperature inside the vessel (indicated indicator points to a value on the scale which
on the thermometer) is different from the one set agrees with the temperature shown on the
up on the regulator scale. thermometer. Finally screw the knob's screw tight
A quick correction can easily be achieved by without changing the position of the axis.
turning the thermostat knob either little clockwise

6.3. The appliance does not heat up _____________________________________


If your hot air steriliser does not work, check the If the signal lamp (4) does not light up to show that
internal fuses. To change fuses, first disconnect the power is on, check the cable, power socket,
the appliance from the mains power supply. Press and mains fuse.
down the tongue on the fuse container on the rear If the signal lamp (4) lights up but not signal lamp
side of the oppliance and pull the container out. (2), check the setting of the control knob (see 1.
Inspect the fuses and exchange if necessary. above).

7. Appendix

7.1. Spare parts ______________________________________________________


Article No. Description
32985 door seal (type 215-225)
92995 door seal (type 255)
32890 thermostat
32930 timer
33470 thermometer
16450 mains fuse 6,3A 2x

page 4 Rev. 2 – 99/0233 BA_GB_215_225_255.doc