HAMILTON-GS VUP SN _ HAMILTON-G5 Ventilator Test Report IPP SN Customer Name: Date: (YYYY/MM/DD): Service Manual Version: 624093) General Maintenance Instructions 1, Record information in spaces provided, 2. Check (tick) check boxes if function is correct. Oxygen Cell replaced Fan Filter cleaned or replaced Gas Inlet Filters replaced Internal Battery Extended Battery LCD Display Backlight replaced Test Configurations (TC) oooooo Dooocao Locations Serve Board Senor Hoard E.. ie 5 Ta sens 20 al Pesta dns Cad Stam 624093/00 This printout is not authorized by HAMILTON MEDICAL AG. Unit ‘VUProcessor Checks 4: voprocessor | VUP Ver. checks VRC Ver MIP 1 Ver: ESM Mosel ESM HWeRev: FSM Voltage 5000m\: ‘VUP-Motherooard HW-Rev: Software Checksum: Must display VUP OK. ox IP 2 Ver ESM Sin CPU Teme: ESM Voage 3200m\: Operating Hours Unit Test/Funetion Specifications 2: Relay and | Emergency Alarm Emergency Alarm Sounds oko Snergency Checks | Relay ON/OFF Relay Switches ON and OFF ou HE Relay ON/OFF LE Relay Switches ON and OFF oko 2: vRC-voP DPT RAM Test Must cisplay OK oko Communication — | yup.vkc test: Must cisplay OK OK checks ‘VRC-VUP Test Mast cisplay OK oko DDPRAM R/V Test: Must cisplay OK oka settest Must cisoay 0 oko DPRAM Collisions: 1000 displayed in one or both fields OK Incremented Value: Incremental count from 0720 t0 235255 OK 4: A/D Converter || ADC 11 MaxiMin Values in Range (7250 to 7750) OKO check ADC 12 MaxdMin Values in Range (-7250 to -7750) OK Measure -10 VDC on the Sensor Board from TP2- Gnd TPH -9.93:0-10.07 oka Apc 14 Max/Min Values in Range (980 10-1010) OK. ADC 15 Max/Min Values in Range (980 to 1010) OK ADC 16 Max/Min Values in Range (10 to 10), OK. 5: D/A Converter |inputo ADC 9 1010 10 OK check ADC 10 9800 to 10200, oko Input4o ADCs 3800 to 4200 oKa ADC 10 +5800 to 6200 oka Input 100 ADCs £9800 to 10200 oka ADC 10 “10 t0 10 Ko 1524093100 This printout is not authorized by HAMILTON MEDICAL AG. Unit Test/Function Specifications 6: zero and —_|[zero calibrations adjust | Adjust to: Full-scale pai | 0.020.1 mbar oka Cadipeation Paw 1 mba oka Note aux 0.00.1 mbar oka Disconnect the thing eP Servo 23 mba oka from the Seo Va a Mer | 05m on DVM Measure T°6-Gnd TP OKA Note AP Flow Sensor (Mixer ON). | (TCH) 0.0 + 40.0 bar oka Reconnect the Seo Vahe ung andconrec:the TeutSeale Calibrations ‘Adjust | Adjust Equal to Pressure Gauge: pat (80 mbar pressure) | (Tc2) (76-84 mba) OK aw (80 mbar pressure) | (TC2) (-Ppat 205 mba) OKO Paxx(80 mbar pressure) | 2) (-Ppat =05 mba) OK AP How Sensor Siler (2.5 mbar probe). | {TC3)= to gauge and postive 200 uber OK oP low Sensor Bue (25 mbar probe) | (Tea) = to gauge but negative + 200 par OKO a? Servo 200 mbar pressure) (TCS) x 1 mbar oka 102 cot select Calibrate 0, Cell Burton 02 Celis calibrated oka Golibration € Automated - Cal status 0 oka checks Actwate Oxygen Knob, set to 50% - now Deselect Atiacs Av ane Oxygen] Os Cell Calibration Checks supples Select Tank Gas Button 0, Gain | 75 to.285 oKa 0, Cet: 129v to a25v ox owgen | 47% 1053% oKa Selectinfet airButton 0, Gay | 75 toss Ka o,cek —Josvto17v oka owygen: | 19% 10 23% ox Press inlet, 2utton 0, Gain’ | 75 to 285 oka O,cet —Jasvtossy oka onygen: oka 4: Mixer Tank Tightness Prank Drop from 300 t0 280 mbar > 8 seconds OK Calibration ¢ — iixer Valve Checks 80m Flow | 010 300 mus oka necks 100m 00 200 ms oka tow Mixer Calibration Fer Mixer, perform Calibrate Miner 3 mes before adjusters. Start shave 250 mar ana just down Press Calibrate Mixer button Mixer Check Press Check Mixer button, Joxygen knob in several pesitions. Tank Overpressure Valve ress Check Mixer button and Act Joxygen Knob Adjust Onygen to 50% 624093/00 This printout is not authorized by HAMILTON MEDICAL AG. Agjust dP Mine FullScale until Prank max displays 330 to 350 mbar Value = Oxygen knob 25% Prank Max Displays < 500 rebar ox. oxa oxa Unit Test/Function Specifications 9: Flow Sensor Cixeuit Checks Note Fo Sensor Atazera Vales eee every 5 seconde ON ana OFF dP Flow Sensor Autozero Valve Check Seal Siber Flow Sensor Check Seal Blue Flow Sensor Check Rinse Flow Check Check for equal Rinse Flows ‘Auxiliary Rinse Activation Check You must confirm the position of the Metal Plate is switched in the OFF postion and there are no bubbles. Only if the instrument is specifically ‘configured for Auxiliary Pressure Measurements should there be any bubbles. (rcs) OFF Range -40 to 40 par oKa (ON 5040 10 $200 par oka 17) OFF Range 40 t0 40 oar oka (ON -5040 to -§200 par oxo (res) Rate of bubble flow equal rom both tubes OK Note (CB) Two equal length tubing (Ser) res) No bubbles oa Bubbles (only if configured) oa I there are bubbles, check the Metal Plate position. If itis positioned for ne Auxiliary Rinse Flow, look for leaks in the system. 10: Patient overpressure Valve Checks Servo Valve Tightness Check Safety Valve Block Tightness Check {Bombar pressure) Patient Overpressure Valve Check {Continuous pumping Hand Pum) (C10) 20s Capilary Tube <4 mbar (TC11) 90 moar to.43 mar >10 seconds oxo oa (TC11) overpressure Valve opens and pressure ‘must drop at 104 t0 11 mbar maximum pressure oka Note (ert) Tbng = 110 em y 22mm = 345 11: ambient ‘Ambient Valve Check Valve checks ‘Ambient Vale is Open when Unit 13s | (TCY2) Inhale = able io inhale freely OK. selected Pat (e12) Inhale oa (e12) Exhale oka Select Close Ambient Valve (TCA2} Inhale» not possible oka (e12) Exhale = not possible oka This printout is not authorized by HAMILTON MEDICAL AG. De-select Close Ambient Valve 624093100 Inhale to confirm the Ambient Valve fs open OK Unit 12: Inspiratory Valve Checks Note For the flow Sensor Test/Function Inspiratory Valve. Select Adult Button Select 20ml/s Button ir Delivery Check Specifications {TC10) w/20ms Capillary) Pressure 23 mbar (Flow Range 1.15 - 1.125 Yin) Catton scat the Ajust 20nUs on Servo Board oka Spoor sien dekectonboaedonibe vss Sooml/s Button {IC10) wScOms Orfce Pressure 22 mbar Deals a eo the {Few Range 29-31 rin) een nsactane ‘Adjust 00m on Servo Board (corrected for altitude) oka lattuce [[ 0] 50] 100] +50] 200] 250] 00] 280] aoa] 50] soo] $80] a0] oso] eco Cenecton|| 43 [+4036 [133 |s20[s27]s23 [.20[s17[ +14 [r10[so7[+04 [sor] 0 fatawde [| 700] 200] 200] 1000] 1100] 1200) 1300] 1400] 1500) 1600) 700] 1800] 1900|2000] 2100 (eneaion|)-02 | 09] 15[21|27|-2[30| 4a] so[ss| ei |r| 72|7a[ ee Flow Sensor Calibration Note Attach Pavent Circuit Automated - Cal statu ox Leave he Patent Beating (Creat alae 0 perform he eprtory Valve Linear Chece Select Calibrate Flow Sensor Button Inspiratory Valve Linearity Check 20 mls Button 50 mls Button 100 ms Button 200 ml’ Button Flow pat Ppat_ Flow Ranges limin (0 40.mis) (0 5 mbarkt.15-1.125)0K (30-70 ms)(0- Smba) 24-356) OKO (80-120 mls) (0- Smbay (64-66) OK 160-240 mls} (0- § mbar) (10.8 - 13.2) OK 500 mils Button (425-575 ms) (0- Sma) @9-31) OKO 1000 mls Button (@50- 1150.) (0-10 mbar) (54-66) OKO. closed (0- Sms) (0~ S mbar) oka 13: Expizatory __|/Expiratory Valve Calibration Check Valve Calibration |! aczvate Reset Expiratory Valve Button | (TC13) Resets Gain to default (128) OK & Checks Activate Calibrate Expiratory Valve aan Button (C13) Automated - Ca Status 6) oxo Net Expiratory Valve Linearity Check Paw/Ppat the PEEPICPAP Knob 0 (a= 15 mbay ox PEEPICPAP Knob 5 @.0- 6.5 mbar) oKa PEEPICPAP Knob 10 (9.0-11.5 mbar) ox PEEPICPAP Knob 30, (28.5-32.0 mbar) oka 624093/00 This printout is not authorized by HAMILTON MEDICAL AG. Unit Test/Function Specifications 14: Nebulizer checks Note Nebulizer vabe cles every 5 seconds ON and ‘Nebulzer Pressure Checks Nebulizer Vale switches ON and Oi Nebulizer Valve Leak Check Opens and Closes 5 second intervals. OK. (TC14) When the Nebulizer Valve ON, Pressure Gauge ? bar) oxa No bubbles or maximum of 3 bubbles per 10 OF seconds oa 15: Internal || Internal Battery and PMS Checks Battery and PMS | Design cap Ah Should cisplay 8.000 oa Cyele count Record Cycle count ste. Record PMS Rew Temperature C -<60°C Record Temperature 16: Extended _| Extended Battery Checks Battery Design cap Ah Should cisplay 6.600 oKa yale count Record Cycle count she. Record Battery SN Temperature C <<60°C Record Temperature Unit 'Processor Checks 18: Processor | Ver: Checks IPP <-> VUP connecrea oxo FPGA Code: Displays Variant A, revision 6 oxo ESM Model oxa For ESM SIN <600: Displays EMOIAA or EMOTAS For ESM SIN >600: Displays EMOTA3 ESM Voltage 5000mV: IPP-Motherooard KW-Rev: Software Checksum: Must display BP OK oxo NOS 0: Displays Rev: 3.0 oxa ESM SIN CPU Temp: <70°C ESM Voltage 3300m\: Unit Test/Function Specifications 19: LED Display _|| Select color Buttons ON/OFF Displays corect color (no flaws) oa Checks Leo on/OFF Switches display backlight OFF/ON oka Message & Patient Alarm Dials Not Required This printout is not authorized by HAMILTON MEDICAL AG. 624093100 20: user [control Knob Checks Interface & Select Knob ‘Alarmtone Checks Audible Alarm Check High Alarm Medium Alaren Low Alarm Hard-Key Buttons Check Interaction Panel Hard Key Checks Press each button Interaction Panel Hard Key LED Checks Press each Interaction Panel Hard Key Interaction Panel Dimming Checks amp blue Lamp red amp yellow White LED Color LED LeD test 624093/00 This printout is not authorized by HAMILTON MEDICAL AG. 16 Clicks Right (slaw) from Center 16 Clicks Let (slowly from Center 5 continuous beeps 3 beeps then 2 beeps, repeats 25sec. 2 beeps, no repeat Confirms Hard-Keys function correctly Butions change color Each time pressed, changes color Blue lamp dims Rec! lamp dims Yellow lamp dims White LED dims Color LED dims Lea dims ox oka oka ox oka oxa oka oKa. oxa oka oxo ox oka. oKa TestFunction Visual Checks Grounds are secure oxo Contact strios are secure oka Internal Cable Checks || crounc (earth pin to: Potential Equalization Terminal < 19 oka croune (earth pinto: Communication Interface Connector < 19 oka [Groune earth pin to: Tank Gverpressure Valve Cover < 1082 oxa croune (earth) pint: Fan Sereen Cover < 100% oxo [Groune earth pin to: Interaction Panel < 1082 oka Grounc (earth pin to:Column < 1002 oxu Type B Current / Value Nc SRC Earth Leakage Current general mA, 0s 1 oKa Enclosure Leakage Current mA o4 os: oka Electrical Safety Tests || avers Leakage Curent/ mA o4 os. oka TE 60601-1 Patient Leakage Current /mA. - - oka Mains on applied part Protective Earth Resistance ohms 02 Ka Induding Power Cord Insulating Resistance Mohn >200 oka 624093100 This printout is not authorized by HAMILTON MEDICAL AG.

You might also like