You are on page 1of 7

CUSTOMER SERVICE DEPARTMENT FAST FAX

To: FAROArm User From: Customer Service Department


FORM: 176-XXX
1 Pages total including this page. Contact us immediately if pages are missing or not legible.

Instructions for Calibrating the USB FARO Arm with


Guidance On

The guidance feature has been designed to show users how to minimize calibration error by maximizing
the probe's range of motion during the calibration process.

With the ball probe securely seated in the calibration cone, Guidance prompts users to take a fixed
number of points while keeping the unit's elbow to one side and sweeping the handle over a
predetermined path.

With Guidance ON, the motion of the handle is restricted to a specific path in 3D space thus obtaining a
successful calibration with Guidance might prove to be more challenging to some users; especially when
using probe extensions or if there are environmental constraints that prevent them from positioning the
handle as far down as the routine requires.

Users are encouraged to calibrate with and without guidance in order to determine which method gives
them the most consistent results.

The following procedure describes calibrating a USB arm with Guidance. The procedure is the same in
all versions of the CAM 2 software and all of the FARO external software’s that use the FARO calibration
plug-in.

To turn the Guidance on, the user will need to Edit the current probe and check the checkbox labeled
Guidance as shown below, and then accept the change by clicking on the Modify Button. Note: The
Modify button functions the same as the OK button on most pop up windows.

Revised: June 29, 2011 LM-06REF176-128.doc


© 2011 FARO Page 1 of 7
The user will then need to choose the Hole Calibration from the list of choices on the left side of the
Probe Calibration screen. The screen below will be displayed. (Actual pictures of a user calibrating with
Guidance have been placed to assist in determining the correct orientation)

Revised: June 29, 2011 LM-06REF176-128.doc


© 2011 FARO Page 2 of 7
When the arm and probe are oriented in the correct position on the calibration cone the following screen
will be displayed and the user can take points as long as the arm and probe continues to be in the correct
orientation.

Revised: June 29, 2011 LM-06REF176-128.doc


© 2011 FARO Page 3 of 7
The three sections of the screen below are to assist the user in orienting the arm in the correct position.
(I have added the numbers) Section 1 is the view as if the user were looking from the top of the
calibration cone. Section 2 is the view from the side, near the calibration cone. Section 3 is the view
from the side of the cone and will only display green when the probe is oriented at the correct angle on
the calibration cone. The user will need to reference all three of these to take all of the points in each of
three orientations.

As the user takes the points, the green band will disappear as shown below.

Revised: June 29, 2011 LM-06REF176-128.doc


© 2011 FARO Page 4 of 7
If the Arm is not oriented correctly for the measurement, the following screen will appear indicating
which direction the Arm is located relative to the correct area.

The pictures below show all of the different orientations that were necessary for this user to calibrate
with Guidance on.

Revised: June 29, 2011 LM-06REF176-128.doc


© 2011 FARO Page 5 of 7
Revised: June 29, 2011 LM-06REF176-128.doc
© 2011 FARO Page 6 of 7
If you have further questions, please contact the customer service department by telephone at 1-800-736-2771 or by e-mail at
support@faro.com.

FAX QUESTIONS TO: (407) 333-8056

Revised: June 29, 2011 LM-06REF176-128.doc


© 2011 FARO Page 7 of 7

You might also like