WHITE PAPER
Despite the common need for testing, purpose- 2. V2 is closed and V4 is opened, evacuating
the boost chamber.
built explosive decompression systems remain
expensive and hard to find. Users are limited to 3. V2 is closed and V1 is opened to initiate the
purchasing a slow response altitude chamber, explosive decompression.
incorporating expensive high-power vacuum
pumps, contracting out testing, or even forgoing 4. V5 and V3 are opened to vent and reset the
system.
real world testing entirely and settling for
numerical simulations.
Ideal Vacuum Products 5910 Midway Blvd. Albuquerque, NM 87109 (505)872-0037 www.idealvac.com
Page No. 1
White Paper: Performance and Analysis of a Modular Vacuum
Chamber for Explosive Decompression Testing
𝑉𝑏
𝑃𝑡,𝑖 − 𝑃𝑠,𝑓 = 𝑉𝑡
(𝑃𝑠,𝑓 − 𝑃𝑏,𝑖 ). (6)
Ideal Vacuum Products 5910 Midway Blvd. Albuquerque, NM 87109 (505)872-0037 www.idealvac.com
Page No. 2
White Paper: Performance and Analysis of a Modular Vacuum
Chamber for Explosive Decompression Testing
The boost chamber and test chamber selected To measure the decompression time, the boost
were the Ideal Vacuum 12x12 and 6x6 modular chamber was instrumented with two piezoelectric
chambers, respectively. This configuration yields discs, along with the capacitance manometer.
a boost ratio of approximately 9.5. One disc was placed near the entrance of the
boost chamber to capture the effect of the air
Implementation of Figure 1 was as follows:
rushing in from the test chamber. The second disc
● M1: 1,000 Torr capacitance manometer was positioned near the back of the boost
● M2: Bourdon gauge chamber so an average response could be
● P1: SH-110 scroll pump determined. The response from the discs were
● V2-5: Manual butterfly valve measured by a Tektronix TDS 3014B 100 MHz
oscilloscope. Figure 3 shows the test setup.
For comparison, V1 was switched between a
pneumatic ISO-100 gate valve and an Ideal The pressure in both chambers was measured
Vacuum KF-50 CommandValve (an electronic using a 1,000 Torr capacitance manometer (note:
butterfly valve). These components are readily the use of a second pressure transducer in Figure
available and require no customization to 1 was for automation/ease-of-use).
accomplish the demonstrated testing.
Test Routine:
The change in test chamber pressure was 1. V1 and V2 are open and the vacuum
determined by using Equations 7 and 8 to relate pump evacuates both chambers to the
altitude to pressure. The initial altitude in the test initial boost chamber pressure; 0.5 Torr
was 8,000 ft (2,438.4 m), which corresponds to in this case.
565.8 Torr (75.4 kPa). The final altitude was
60,000 ft (18,288 m), which when related to 2. Close V1 and open V5 until the pressure
rises to the initial condition for the test
atmospheric pressure was 54.3 Torr (7.2 kPa).
chamber; 565.8 Torr.
Using Equation 6, the boost chamber starting
pressure was about 0.5 Torr (0.67 kPa). 3. Open V1 and collect the data.
Results
The first decompression test used the electronic
butterfly valve as V1 to separate the test and
boost chamber. Figure 4 shows the results of the
decompression. The time -20 ms to 0 ms is the
time before the V1 opening. The pressure is stable
and the piezoelectric discs are producing 0 volts.
V1 opens and the piezoelectric disc in the path of
the flow responds first, followed by the
Figure 3: Explosive decompression test setup. Left-to-Right:
12x12 modular vacuum cube, the CommandValve electronic
piezoelectric disc measuring the average
butterfly valve, and the 6x6 modular vacuum cube. response. The gauge follows after a 20 to 25 ms
lag due to delayed response and conversion
latency.
Ideal Vacuum Products 5910 Midway Blvd. Albuquerque, NM 87109 (505)872-0037 www.idealvac.com
Page No. 3
White Paper: Performance and Analysis of a Modular Vacuum
Chamber for Explosive Decompression Testing
Discussion
With both the pneumatic gate valve and
electronic butterfly valve meeting the goal of full
pressure equilibrium within 0.5 s, this test and
system configuration is appropriate for explosive
decompression testing.
Ideal Vacuum Products 5910 Midway Blvd. Albuquerque, NM 87109 (505)872-0037 www.idealvac.com
Page No. 4
White Paper: Performance and Analysis of a Modular Vacuum
Chamber for Explosive Decompression Testing
References
[1] Wikipedia Contributors, "Uncontrolled decompression," 24
September 2019. [Online]. Available:
https://en.wikipedia.org/w/index.php?title=Uncontrolled_deco
mpression&oldid=916341251
Ideal Vacuum Products 5910 Midway Blvd. Albuquerque, NM 87109 (505)872-0037 www.idealvac.com