AlarmScout 961/962 (FM)

AlarmScout 961/962 (FM)


The AlarmScout 961/962 series are used to detect high or
low level switch points in a broad range of viscous to light
products. Switch points can be used for safety functions such
as alarm or trip. Pulsed signal technology provides superior
performance in applications suffering from foam or turbulence.
Installation is easy and straightforward: no configuration or
commissioning needed.

The AlarmScout 961 has a tip sensitive set-point. The


AlarmScout 962 offers 2 set-points on the same transducer,
a tip sensitive set-point and a second set-point via a flow-
through upper gap. The unit is used for level alarm or to control
a pump in an auto fill/empty mode.

The AlarmScout 961/962 has no moving parts and is equipped


with advanced diagnostics that continuously check the trans-
ducer and electronics, even in submerged condition. Both the
AlarmScouts can be used in SIL-2 rated loops.
Technical specifications
Performance
Response time : 0.5 s typical
Repeatability : ± 2 mm (0.078”)
Ambient Temperature : -40 °C up to +70 °C (-40 °F up to +160 °F)
Electromagnetic Compatibility : Meets CE requirements (EN-61326: 1997 + A1 + A2) and NAMUR NE 21
Electronics
Input Voltage
- mA - version (Pos 5 = 5) : 2 wire loop powered , 11 - 35 V DC
- Relay - version (Pos 5 = 2 or 7) : 102 - 265 V AC 50/60 Hz or 18 - 32 V DC
Power consumption : < 3 Watt (relay version) – < 1 Watt (mA version)
Output
- mA - version (Pos 5 = 5) : 961: 8 mA (safe), 16 mA (alarm) ± 1 mA
962: 8 mA (safe), 12 mA (lower gap alarm), 16 mA (upper gap alarm) ± 1 mA
961/962: ≤ 3,6 or ≥ 22 mA error signal
- Relay - version (Pos 5 = 2 or 7) : 961: one 5 A DPDT relay
962: two 5 A SPDT relays
961/962: one 5 A SPDT malfunction relay
Time delay : 0,5 to 45 s adjustable (in addition to transducer response time)
Indication : LED’s for process alarm status, malfunction (error of transducer, electronics or electrical noise
interference) and wet/dry status of transducer (961 with relay only)
Selftest
- Automatic : Continuously verifies electronics, transducer and noise interference
- Manual : Via pushbutton for checking alarm output(s) and error output/function
Housing material : IP66, cast aluminium or cast stainless steel
Approvals: FM/CSA : Class I, Div. 1; Groups B, C & D
Class II, Div. 1; Groups E, F & G
Class III, NEMA Type 4X, T6
SIL (Safety Integrity Level) : Functional safety to SIL 2 in accordance to IEC 61508 – SFF > 90 %
Full FMEDA report and declaration sheets available at request
Electrical data : Ui = 28,4 V, li = 94 mA, Pi = 0,67 W (mA version)
Equivalent data : Ci = 10,4 nF (961) / Ci = 60 nF (962), Li = 400 µH (mA version)
Shock/Vibration : ANSI/ISA-S71.03 Class SA1 (shock), ANSI/ISA-S71.03 Class VC2 (vibration)
Net weight : Alu: 1 kg (2.2 lbs) – electronics only
SST: 2,5 kg (5.5 lbs) – electronics only
Transducer
Material : 316/316L SST (1.4401/1.4404)
Actuation length : From 2” up to 130”
Process temp. : -40 °C to +165 °C (-40 °F to +325 °F)
Max pressure : 103 bar @ +165 °C (1500 psi @ +325 °F)
Flanged models are downrated to the design pressure of the selected flange
Humidity : 0-99 %, non-condensing
PED : The PED directive 97/23/EC (pressure equipment directive).
Safety accessories per category IV module H1.
A COMPLETE CONFIGURATION CONSISTS OF 2 MODEL CODES
1. AlarmScout electronics
2. AlarmScout transducer

1. Identification code for AlarmScout electronics


Pos 1, 2, 3, 4 Instrument designation
X 9 6 1 AlarmScout 961, electronics for single setpoint 9M1 transducers
X 9 6 2 AlarmScout 962, electronics for dual setpoint 9M2 transducers
Pos 5, 6, 7 Input power
5 0 A 24 V DC 2-wire loop powered electronics with current shift output
2 D A 18 - 32 V DC line powered electronics with 5 A gold flash relays output
7 D A 102 - 265 V AC line powered electronics with 5 A gold flash relays output
Pos 8 Accessories
0 Blind housing cover
1 Housing cover with glass window (only for aluminium housing: if Pos 11 = 0 or 1)
Pos 9 Mounting
0 Integral mount electronics
Pos 10 Approvals
1 FM explosion proof (only for loop powered electronics: if Pos 6 = 0)
3 FM explosion proof (only for line powered electronics: if Pos 6 = D)
Pos 11 Housing / cable entry
1 Cast aluminium housing with M20 x 1,5 cable entry (2 entries - one plugged)
0 Cast aluminium housing with 3/4” NPT cable entry (2 entries - one plugged)
3 Cast stainless steel with M20 x 1,5 cable entry (2 entries - one plugged)
2 Cast stainless steel with 3/4” NPT cable entry (2 entries - one plugged)

X 9 6 1 7 D A 0 0 3 1 Typical identification code


Typical identification code
X 9 6 A 0 Your identification code

2. Identification code for AlarmScout transducer


Pos 1, 2, 3, 4 Instrument designation
X 9 A 1 AlarmScout 961, transducer with single setpoint
X 9 A 2 AlarmScout 962, transducer with dual setpoints
Pos 5 Transducer materials
A 316/316L (1.4401/1.4404) stainless steel
Pos 6, 7 Process connection
Threaded
1 1 3/4” NPT
ANSI Flanges
2 3 1” 150 lbs ANSI RF
2 4 1” 300 lbs ANSI RF
4 3 2” 150 lbs ANSI RF
4 4 2” 300 lbs ANSI RF
Pos 8 Sensor type
A Standard sensor: min -40 °C / max + 165 °C (-40 °F / +325 °F)
Pos 9 , 10, 11 Actuation length (see dimensional drawing)
0 0 2 Minimum length = 2 inch (only if Pos 4 = 1)
0 0 6 Minimum actuation length “A” = 6 inch (see dim. drawing 962) (only if Pos 4 = 2)
1 3 0 Maximum length = 130 inch

X 9 A 1 A 1 1 A 0 6 0 Typical identification code


Typical identification code
X 9 A A A Your identification code

NOTE: consult factory if CSA approval is required.


Dimensional drawing
102 102
(4) (4)

113 (4.43) without window


113 (4.43) without window 123 (4.85) with window
123 (4.85) with window

cable entry (both sides)


(see position 11 of model code)

60 (2.35) Typical
46 (1.8) Typical

Actuation
length Actuation length
for BSP Actuation length
connection for NPT connection

6 (0.25)

6 (0.25)
Ø 22,9 (0.90)

Ø 22,9 (0.90)
102 102
961 - Threaded connection (4) 961 - Flanged connection (4)

113 (4.43) without window


113 (4.43) without window
123 (4.85) with window
123 (4.85) with window

60 (2.35) Typical
46 (1.8) Typical
Actuation length
“B” for BSP Actuation length Actuation
connection “B” for NPT length “B”

Actuation length Actuation length “A”


“A” for BSP Actuation length “A”
connection for NPT connection

6 (0.25)
6 (0.25) Ø 22,9 (0.90)
Ø 22,9 (0.90)
962 - Threaded connection 962 - Flanged connection
NOTE: all dimensions in mm (inch)
min actuation length between “A” and “B” = 8 cm
min length for “B” = 7 cm
allow 200 mm (8”) clearance for cover removal
anchor sensor for lengths > 80 inch
For More Information
To learn more about Honeywell Enraf’s
solutions, contact your Honeywell Enraf
account manager or visit
www.honeywellenraf.com

Americas Asia Pacific


Honeywell Enraf Americas, Inc. Honeywell Pte Ltd.
2000 Northfield Ct. 17 Changi Business Park Central 1
Roswell, GA 30076 Singapore 486073
USA Phone: +65 6355 2828
Phone: +1 770 475 1900 Email: enraf-sg@honeywell.com
Email: enraf-us@honeywell.com

Europe, Middle East and Africa


Honeywell Enraf
Delftechpark 39
2628 XJ Delft
The Netherlands
EN-12-06-ENG
Phone: +31 (0)15 2701 100 APRIL 2012
Email: enraf-nl@honeywell.com © 2012 Honeywell International Inc.

You might also like