Professional Documents
Culture Documents
Abstract— This paper highlights the use of Calibration Constant uncertainties involved for certain measurement points other
of Test & Measuring (T&M) equipment derived during than all test points. Therefore, modeling of T&M equipment
calibration process, in measurement inference. The equipment performance by mathematical equation and its curve reduces
user is much interested to find the true value of measurand the difficulties of user in real time measurement. The simplest
against the measured quantity. In this paper, the Best-fit-curve is and the most common type of best fit curve is a straight line.
proposed to model a set of test results obtained during As discussed in [4], data interpolation is achieved by the best
calibration by Least Square Error (LSE) based parabolic fit straight line to the data points obtained during calibration,
regression. Calibration Constant becomes more dominant rather which is represented by straight line equation as
than the detailed multi page test results to determine the actual
value of measurement with certain degree of accuracy. This 1
reduces interpolation error while calculating the true value other
than testing points and also the measurement time. The Test set- Where, ‘y’ represents the standard or actual value and ‘x’ is
up is automated and the measured parameters are traceable to the measured value. ‘ a ’ and ‘ b ’ are the calibration constants
national standards. determined by least square method. The method of least
squares assures that the best fit is one which the sum of squares
Keywords- Best-fit-curve, Calibration Constant, Interpolation, of the residual is minimum [6],[9]. The calibration line as fitted
Least square error, Parabolic function, Regression is then used to obtain the actual value ‘ y ’ for measured value
I. INTRODUCTION of ‘ x ’.
Calibration is the process by which integrity of the quality However, linear assumption for interpolating points may
of measurement remains intact. Before taking any final not appropriate for accurate measurement. In [15], the
recommendations or decisions, the due importance is to be cumulative distribution function in the transferred domain of a
given to calibrated measurement data & the associated sum of log normal random variables is approximated using
uncertainty. Due to the properties of material and so frequent least squares method via piecewise linear functions. This
use of T&M equipment, the performance may get deteriorated method increases the complexity in the computation with more
with respect to time. There are also other factors which number of basic functions.
predominates the measurement result in a measurement Probability density function (PDF) based progressive
process such as gain, offset and linearity errors. Several polynomial calibration [1] evaluates calibration co-efficient
methods have been adopted for linearization of equipment directly for smart sensor linearization is one of the possible
performance to get better measurement inference. Internal solution for linearization.
processing circuit of equipment is being modified and updated
The least square based parabolic regression method
to facilitate the same. Electrical noise may introduce certain
presented in this paper seems to be an attractive solution for
degree of non-linearity in the low level precision signal computation of calibration constants in measurement because it
measurements process. Noise level shall be minimized at the requires minimal time with less computational complexity for
source by proper designing source coding of software filters accurate measurement inference.
and moreover by the use of appropriate shielding & grounding
techniques. Equipment manufacturer provides the discrete test II. LEAST SQUARE BASED PARABOLIC
points for calibration of various performance parameters. REGRESSION & CALIBRATION CURVE
During recalibration, the calibration engineer brings out the Polynomial regression is used to fit data points by the
standard results of equipment under test & a certificate of the equation
test result to that effect is used to user for its application.
2
The practical application of the calibration data in real time
measurement plays a crucial role to assure the quality in Co-efficients ‘ a ’, ‘ b ’, ‘ c ’…. are determined to make the
measurement process. The engineer handling the T&M curve best fit for data points. When ‘ a ’ & ‘ b ’ are to be fit
equipment takes decision depending upon calibration results. [10], the equation describes a line, and polynomial regression
But it becomes cumbersome for the engineer to go through is identical to linear regression as explained by Eq. (1).
calibration certificate in detail and find correction and
1
978-1-4244-6653-5/10/$26.00 ©2010 IEEE 334
2010 5th International Conference on Industrial and Information Systems, ICIIS 2010, Jul 29 - Aug 01, 2010, India
The T&M equipment readings (i.e. measured values) Eq. (12) can be written as,
against a set of standard values are taken along ‘X’-axis and
standard values are along ‘Y’-axis. A set of data series are 13
represented by the parabolic equation as Where,
3
The calibration curve obtained by fitting above equation is
then used to obtain the actual value of ‘ y ’ for measured value
of ‘ x ’.
, 14
The method of least squares assures that the best fit is one,
which the sum of squares of the vertical residual is a minimum
[9]. Here ‘ a ’, ‘ b ’& ‘ c ’ are considered as calibration
constants for a particular set of measurement. Eq. (3) may be
written as
; , , 4
For a set of paired data , ; 1,2, … … … , , the
residual error 'is defined by , 14
; , , 5
2
335
2010 5th International Conference on Industrial and Information Systems, ICIIS 2010, Jul 29 - Aug 01, 2010, India
of data not considered for the curve fitting is being estimated IV. TEST RESULT AND DISCUSSION
by, The test result of UUT at different test points and
∑ ; , , associated estimated value (both linear and parabolic
1 18 regression using least square method) is represented in Table
∑
I. Using calibration data obtained as average value on UUT
The accuracy of calibration curve as fitted is closely and the corresponding value on standard, the best fit curve is
associated with the value of . To minimize the effect of plotted based on least square parabolic regression (Fig. 2).
random measurement errors, multiple point calibration is
preferred although it takes more time. From Table I, it is found that zero estimated error is more
in linear regression in comparison to parabolic regression,
III. EXPERIMENTAL SET UP whereas error at 6V & 9V is more in parabolic with respect to
The test is undertaken with TRMS (True Root Mean linear estimation. “Fig. 3” shows the trend of curve fitting
Square) Multi meter “RishMulti 18S”. UUT (Unit Under Test) error.
performance is compared with its higher standard, calibrator
Fluke 5520A. The cable loss is taken care by using the standard TABLE I. CALIBRATION DATA AND ESTIMATED VALUE
measurement cable. The measurement parameters are traceable Average Value on Measured yest yest
to National Standard. To get better measurement performance Value Standard Error (Parabolic) (Linear)
of UUT, Test Uncertainty Ratio (TUR) is maintained at 4:1 at on UUT (V) (V) (V) (V) (V)
each calibration test points. To minimise the human error, the xi yi xi - yi
0 0 0 0.0730 -0.3844
test procedure is partially automated through calibration
5.25 5.0 0.25 5.0158 4.9736
software MET/CAL version 7.2[11]. Five sets of 6.15 6.0 0.15 5.8840 5.8921
measurements are taken at the respective test points and the 7.28 7.0 0.28 6.9828 7.0454
parameter uncertainty is estimated as per Guide to the 8.34 8.0 0.34 8.0223 8.1272
Expression of Uncertainty in Measurement (GUM). The GUM 9.19 9.0 0.19 8.8620 8.9947
approach considers a standard deviation and the confidence 10.21 10.0 0.21 9.8768 10.0357
interval in which the measurand is supposed to lie with a given 11.34 11.0 0.34 11.0102 11.1890
level of confidence [4]. The distribution of measurement 12.48 12.0 0.48 12.1634 12.3524
results is assumed to be Gaussian. 13.46 13.0 0.46 13.1626 13.3526
15.23 15.0 0.23 14.9856 15.1590
The selection of test points is being guided by the 17.24 17.0 0.24 17.0845 17.2104
manufacturer. Using manufacturer’s specification, the 19.12 19.0 0.12 19.0753 19.1291
calibration curve is fitted as explained above. Although there is 19.8 20.0 -0.20 19.8019 19.8231
a scope of selecting PASS/FAIL depending upon the allowable
TABLE II. INTERPOLATION RESULT
limit with logic conditions, it is always preferred to limit the
test result to 70% of the allowable limit to extract better data Average Value on yest(int) yest(int) yint - yintP yint - yintL
for best fit calibration curve. As calibration simulates the Value on Standard (Parabolic) (Linear)
performance of UUT by a mathematical model, the UUT (V) (V) (V) (V) (V) (V)
xint yint yintP yintL
environment factor is also considered by controlling the stable 2.496 2.25 2.3970 2.1629 -0.147 0.0871
environmental condition of calibration facility. The data flow 3.695 3.50 3.5301 3.3866 -0.0301 0.1134
in a calibration system is shown in “Fig. 1”. Serial and Parallel 5.494 5.35 5.2506 5.2226 0.0994 0.1274
port interface unit of Standard communicates with central 7.494 7.25 7.1920 7.2638 0.058 -0.0138
processor in both direction for executing the calibration 10.494 10.35 10.1607 10.3256 0.1893 0.0244
software and sends the stimulus for UUT and measured data 12.493 12.05 12.1766 12.3256 -0.1266 -0.2756
are stored in the database of central processor. Statistical 14.456 14.0 14.1855 14.3691 -0.1855 -0.3691
16.28 16.0 16.0782 16.2306 -0.0782 -0.2306
computation is performed by MATLAB software.
17.67 17.5 17.5375 17.6493 -0.0375 -0.1493
20
STANDARD Actual Reading
18
Estimated Reading
16
I/O Keypad Processor and Central
Data Base
for
Standard Value(V)
12
Calibration
Serial & Parallel Software
10
UUT 2
3
336
2010 5th International Conference on Industrial and Information Systems, ICIIS 2010, Jul 29 - Aug 01, 2010, India
1.2
On computation, the values for ‘ a ’, ‘ b ’& ‘ c ’were
computed as 0.0730, 0.9217 and 0.0038 respectively. From the
Curve Fitting Error
1 tabulated data, standard deviation of error was found 0.1753.
Considering the interpolation measurement, the standard
0.8
deviation of error for parabolic regression was found 0.1236.
Curve Fitting Error(V)
0.6 Similarly, same set of data points are used for computing
the best fit line using straight line equation Y = a + b X. The
values for ‘ a ’ & ‘ b ’ obtained are -0.3844 and 1.0206
0.4
0.05
same concept can be extended to get fixed calibration constant
0
for particular equipment considering the total range into
different continuous segments, which will be of future concern
-0.05 for analysis.
-0.1
ACKNOWLEDGMENT
-0.15 The authors are greatly thankful to Sri S.P. Dash, Scientist-
H(Outstanding Scientist), Director, ITR for allowing us to use
-0.2
2 4 6 8 10 12 14 16 18 the Calibration Facility of the establishment to achieve the
result and also thankful to Dr. S.N. Banerjee, Additional
Interpolation Value(V)
Figure 5: Interpolation Curve Fitting Error Trend Director & Sri B.K. Paltasingh, Scientist-F, GD (QA) for
providing constant guidance and moral support.
4
337
2010 5th International Conference on Industrial and Information Systems, ICIIS 2010, Jul 29 - Aug 01, 2010, India
5
338