Certificate of Calibration
Report Number : 17080132
"27317180008
Reference Standards used in this calibration are traceable to the National Institute of Standards and
Technology (NIST) or other recognized National Metrology institute (NMI's). This calibration is compliant to
ANSI/NCSL 2540-1-1994. The calibration report applies only to the item described. Uncertainties of the
measurements are based upon a coverage factor of k=2 providing an approximate 95% confidence level. In
tolerance or pass conditions are based on the test results falling within specified limits with no reduction by
the uncertainty of the measurement.
Agilent Digital Multimeter 3458a_— | mvasoag291| 7/7/2017 7772018 S-B2NIN-20-1
Fluke Multifunction Calibrator| 57204. 2016203 5/3/2017 5822018 4D11775
Additel Quartz Pressure 1.694-AP4OKQ-PS 00516090001 | 7/13/2017 | 10/13/2017 17071401
‘The recommended calibration interval for this instrument is 12 months from the date of verification, Your particular
quality assurance requirements may supersede this recommendation.
Environmental Conditions Performed by: Bor e
‘Temperature: 23.5°C é
Humidity: 59% RH
Approved ByReport Number : 17080132
Test Results
Pressure
Reference [indicated Reading] Difference | Low limit [High limit :
Parameter Coneition
(psi) (psi) (psi) si) (psi)
Tinearty mm Evy 05 399 40.1 pass
70000 100087 #7 3960.0 To040.0 pass
20000 2010.3, 103 19960.0 | 200400 pass
30000 300124 14 29960.0 | 300400 pass
3999.9 400140 141 39959.9 | _40030.9 pass
Tiystersis [30000 3004.0, 40 29960.0 | _30040.0 ass
20000 2003.7 37 19960.0 | 200400, ass
70000 10009.6 96 9960.0 10040.0 pass
02 04 05; 398 402 pass
DC Voltage Measurement
Accuracy: +(0.01%RD + I.5mV)
Range [Reference DUT Difference | Low limit | High limi ae
o w o. wo wo w
-30,0000__| 30.0003 | 0.0003 | 30.0035 | 29.9955 pass
10,0000} _-10.0001 | 0.0001 | -10.0025 | 9.975 pass
5.0000 ~5.0000 0.0000 | 5.0020 | 4.9980, pass
30%030 [0.0000 0.0000 0.0000 | 0.0015 0.0015 pass
5.0000 5.0001 0.0001 4.9980) 5.0020 pass
10,0000 70,0002 0.0002 9.9975 10,0025 pass
30.0000, 30.0008, 0.0003 | 299955 | 30,0045 pass
DC Current Measurement
Accuracy: +(0.01%RD + 1.5pA)
Range | _ Reference DUT Difference [Low limit | High aaa
(mA) (mA) (mA) (mA) (mA) (mA)
“30.0000 30.0001 | 0.0001 | 30.0045 | 29.9955 pass
-10.0000 | _-10.0000 0.0000 | _-10.0025 | 9.9975 pass
5.0000 4.9999 0.0001 5.0000 | 4.9980 ass
30:30 [7 0.0000 0.0000 0.0000 | 0.0015 0.0015 pass
5.0000 5.0001 0.0001 4.9980 5.0020 pass
10,0000 10,0001 0.0001 9.9975 10.0025 pass
30.0000, 30.0001 0.0001 29.9955 | 30.0045 pass
24V DC Output
Accuracy: #0.5V
Range Dor Reference | Difference | Lowlimit | High mit cae
W) Ww) W) mw) Ww) W)
24 24 74.0220 0.0220 | 23.5000__| 24.5000, pass