Professional Documents
Culture Documents
SEP-10S Plus
SYRINGE INFUSION PUMPS
SERVICE MANUAL
BS037016EN-P02
CONTENTS
1. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. TECHNICAL DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3. EXTERNAL VIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4. KEYBOARD DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5. TECHNICAL DESCRIPTION
5.1. PRINCIPLE OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5.2. ELECTRICAL SCHEMATIC DIAGRAM OF THE PUMP . . . . . . . . . . . . . . 9
5.3. ELECTRICAL SCHEMATIC DIAGRAM
OF ELECTRONIC BOARD B7043 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
6. SETUP MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7. CHECKING PARAMETERS
7.1. MEANS REQUIRED FOR PARAMETERS CHECKING . . . . . . . . . . . . . 19
7.2. CHECKING INFUSION VOLUME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
7.3. CHECKING INFUSION RATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.4. CHECKING BOLUS VOLUME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.5. CHECKING OCCLUSION PRESSURE . . . . . . . . . . . . . . . . . . . . . . . . . 21
9. TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
2
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
10.11. REPLACING SYRINGE SIZE SENSOR . . . . . . . . . . . . . . . . . . . . . . . 35
10.12. REPLACING BUZZER UNIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
10.13. REPLACING POWER SUPPLY INLET . . . . . . . . . . . . . . . . . . . . . . . . 36
10.14. REPLACING SUPPLY UNIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
ANNEX A
ELECTRICAL SCHEMATIC DIAGRAM
OF THE PUMPS SP-12S Pro and SEP-12S Plus . . . . . . . . . . . . . . . . . . . . . . . . . . 40
ANNEX B
ELECTRICAL SCHEMATIC DIAGRAM OF THE PUMP SEP-10S Plus . . . . . . . . . 42
ANNEX C
INTERCONNECTION DIAGRAM OF THE PUMPS . . . . . . . . . . . . . . . . . . . . . . . . 44
ANNEX D
ELECTRICAL SCHEMATIC DIAGRAM OF ELECTRONIC BOARD B7043 . . . 46
ANNEX E
ASSEMBLY DRAWING OF ELECTRONIC BOARD B7043 . . . . . . . . . . . . . . . . . . 52
ANNEX F
DRAWING OF CABLES FOR MULTIFUNCTIONAL CONNECTOR (MFC) . . . . . . 54
ANNEX G
DESCRIPTION OF ERRORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
ANNEX H
PREVENTIVE MAINTENANCE CHECKLIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
ANNEX I
COMPONENT LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
ANNEX J
ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
ANNEX K
COMMUNICATION PROTOCOL. TECHNICAL DESCRIPTION . . . . . . . . . . . . . . 59
3
Service manual
1. INTRODUCTION
NOTE:
The manufacturer’s authorized personnel may carry
out the technical servicing of the pump only!
4
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
2. TECHNICAL DATA
Power supply:
220–230 VAC ±10%, 50/60 Hz, or 115 VAC ±10%, 50/60 Hz,
or internal rechargeable battery
Fuses:
T80 mA/L250 V – for 220–230 VAC
T160 mA/L250 V – for 115 VAC
Power consumption:
10 VA (max)
Internal battery:
9.6 V 1300 mAh NiMH battery
Cordless work time:
8 h (minimum) at 5 ml/h infusion rate;
2 h (minimum) at 100 ml/h infusion rate.
Battery recharging time: 24 hours.
Classification: Class II, CF, splash proof.
RS232 (optional)
12 VDC supply connection (optional)
Nurse Call connection (optional)
Operating temperature range:
+5 to +40 °C.
Storage temperature range:
-20 to +40 °C.
Air pressure:
60 to 106 kPa.
Relative humidity:
90% max., no condensation (operation and storage).
Dimensions:
135 x 305 x 195 mm
Weight:
2.6 kg.
5
Service manual
3. EXTERNAL VIEW
Fig. 1.
Rear view of the pumps SP-12S Pro, SEP-12S Plus and SEP-10S Plus
1 – syringe driver arm 6 – handle
2 – slot for inserting the push- 7 – cap
button of the syringe 8 – mounting pole clamp
plunger 9 – fuse holder
3 – rubber bellows 10 – mains inlet
4 – slot for inserting the finger 11 – MFC*
grips of the syringe barrel 12 – Audio volume control*
5 – syringe clamp * – optional
7
Service manual
4. KEYBOARD DESCRIPTION
8
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
5. TECHNICAL DESCRIPTION
9
Service manual
5.2.2. Power supply
u 220-230 VAC (or 115 VAC) voltages via the mains filter
MF1 and fuse F1 is fed to the primary winding of the
step-down transformer TR1. Voltage from the secondary
winding of the transformer TR1 reduced to 10-11 VAC is
rectified by the diode bridge D1 and capacitor C1 (11-
13,5 VDC) and via the fuse F1(A2) is fed to the remain-
ing part of the circuit. The resistor R1 sets the charging
current for the rechargeable batteries GB1, GB2 which
shall be in range of (70-130) mA. When the pump is
powered from the internal battery, power supply current
goes via the diode D2. The fuse F2(A2) protects batteries
by limiting charging and load currents. There is fed a sig-
nal to the electronic board B7043 via the diode D3 and
the resistor R2 informing the pump is powered from the
external mains of 50/60 Hz, 220-230 VAC (or 115 VAC).
The electronic board B7043 controls the buzzer Z1.
5.2.3. Keyboard
u The keyboard is composed of 21 keys (SEP-10S Plus
20 keys) in form of a matrix having 7 inputs and 3 out-
puts. Electronic board B7043 performs scanning of the
keyboard and reading of information from it. The key
ON/OFF controls the circuit of electronic power supply
switch located on the electronic board B7043.
5.2.4. Drive
u The stepper motor SM1 rotates the lead screw with a nut
on it converting rotating movement of the motor to lin-
ear movement of the tube. The tube supported by four
guiding bearings is connected to the syringe driver arm.
Marginal positions of the tube with syringe driver arm
are limited by:
w limit switch S1 at the right extreme position which is
actuated by the end of the tube;
10
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
w rotational optical sensor TAX1 at the left extreme po-
sition when the tube covers the raster of rotating cou-
pling of the optical sensor TAX1.
In case of occlusion when pressure in the syringe in-
creases to the prohibited level, it stops movement of the
syringe driver arm and at the same rotation of the step-
per motor SM1.
Rotation of the stepper motor is controlled by the optical
sensor TAX1 which reads a signal, reflected from the raster
of rotational coupling. This signal is fed to the micro-
controller U18 located on the electronic board B7043, and
this signal switches on the emergency signal. There is some
delay between beginning of occlusion on one hand and
stepper motor SM1 stopping and emergency signal actua-
tion on the other hand depending on infusion rate, length,
thickness and elasticity of extension tube.
Xmin. PREALARM!
SYRINGE EMPTY!
5.3.9. Switches
u Currents for both phases flow to the common conductors
of motor phase coils via accumulating reactors L4 and L5.
Direction of motor rotation is defined by appropriate or-
der of current switching in both phase coils. This switch-
ing is implemented by current switches T8, T9 and T10,
T13 for the 1st and the 2nd phases accordingly.
16
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
6. SETUP MENU
NOTES:
1. Entered character can be reset by means of the C or NO
keys.
2. Old drug name can be restored by means of the PROG key
until new name is confirmed.
ATTENTION !
It is recommended to minimize number of parameters, types
of syringes, drug names and other functions leaving only that
necessary for work. It will help to avoid errors in parameters
programming and thereby decrease patient’s risk.
u If the high occlusion level value after actions specified in section 7.5 are
performed is beyond specified limits, occlusion level shall be calibrated.
For this sake factual value of high occlusion level pressure obtained dur-
ing testing is entered into pump.
Typical value for CORRECT END is 60 kPa.
18
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
7. CHECKING PARAMETERS
NOTE:
Programming infusion parameters, filling the sy-
ringe and extension tube with drug solution, insert-
ing the syringe into the pump and starting infusion
are described in the OPERATING INSTRUC-
TIONS of the pump.
19
Service manual
7.3. CHECKING INFUSION RATE
u Take the 50 ml BD Plastipak syringe with extension tube
connected to it and fill it with distilled water up to the
point of 55 ml. Insert the syringe into the pump and con-
firm the 50 ml BD Plastipak syringe.
Program the following infusion parameters:
w infusion time — 12 min,
w volume limit — 10 ml,
w occlusion pressure level — HIGH.
Expel the air from the extension tube. Insert the free end
of the extension tube into the graduated test-tube and
start the infusion. Measure infusion time by the stop-
watch and record the volume in the test-tube at the end
of infusion. Calculate the infusion rate. It shall be
50 ml/h ± 1 ml/h (± 2%).
20
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
7.5. CHECKING OCCLUSION PRESSURE
u Take the 50 ml BD Plastipak syringe with an extension
tube connected to it and fill it with distilled water up to
the point of 30 ml. Insert the syringe into the pump and
confirm the 50 ml BD Plastipak syringe.
Program the following infusion parameters:
w infusion rate — 50 ml/h,
w volume limit — 30 ml,
w occlusion pressure level — LOW.
Connect the pressure gauge to the free end of the exten-
sion tube. Start the infusion. As soon as the audible alarm
is activated and there is displayed the message
OCCLUSION!!!
the pressure gauge shall read the pressure of (30 or 40)* kPa
± 15 kPa.
Press the key C. Reprogram the occlusion pressure to
MEDIUM. Start again the infusion. As soon as the audi-
ble alarm is activated and there is displayed the message
OCCLUSION!!!
NOTE:
If measured values of pump parameters are out of
ranges permitted, contact the Manufacturer.
* depends on the installed occlusion presure set (for SP-12S Pro only).
21
Service manual
8. CHECKING ALARM SIGNALS
22
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
All alarm signals are accompanied by the audible alarm.
Alarm signals listed in the Table are checked in the fol-
lowing ways:
OCCLUSION!!!
OCCLUSION or END
* depends on the installed occlusion presure set (for SP-12S Pro only).
23
Service manual
Start the infusion. Disconnect the mains cord. There shall
be heard the audible alarm and there shall be displayed
the message
NO MAINS!!!
check power cord!
LOW BATTERY!
LOW BATTERY!
* In case the pump was in use for more than 2 years, this time may be shorter.
24
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
and
VERY LOW BATTERY
5 min. PREALARM!
END of INFUSION!
KOR X.X ml/h
END of INFUSION!
KOR X.X ml/h
Press the PROG key and restart the infusion with the pa-
rameters set in step 6. Press the START/STOP key. 2
minutes later there shall be heard the audible alarm and
there shall be displayed the message
25
Service manual
STOP X.X ml
KOR X.X ml/h
27
Service manual
9. TROUBLESHOOTING
5 In the syringe insertion Faulty end switch S1 Check the end switch and
mode, the syringe driver (Fig. 4 pos. 11). replace, if required, as
arm stops with noise described in the sections
when it reaches its right 10.1 and 10.5.
hand side position.
6 The pump doesn’t Faulty pre-alarm switch S2 Check the pre-alarm switch
display the messages (Fig. 4 pos. 8). and replace, if required, as
"X min. PREALARM!" described in the sections
and 10.1 and 10.6.
"SYRINGE EMPTY!"
although the syringe is
filled with more than
50% of its nominal
volume.
29
Service manual
No SYMPTOM CAUSE CORRECTIVE ACTIONS
7 Although the syringe is Syringe sensor S3 is faulty Disassemble the pump as
inserted, there is described in the section 10.1.
displayed the message Replace the syringe sensor as
INCORRECT described in the section 10.7.
INSERTION!
8 When a syringe is in- Faulty syringe size sensor Calibrate syringe sensor (see
serted, a wrong syringe (Fig. 4 pos. 3). section 6) or replace it (if
size is displayed, or calibration was not success-
there is a message ful) as described in the
ILLEGAL SYRINGE! sections 10.1 and 10.11.
CHANGE SYRINGE
10 Audible alarm of the Faulty buzzer Z1 in the Check the buzzer and replace,
pump fails to operate. power supply. if required, as described in
the sections 10.1 and 10.12.
12 The pump fails to ope- Faulty electronic board Replace the electronic board
rate when switched on B7043. B7043 as described in the
with ON/OFF key. No sections 10.1 and 10.8
message or the message (see also Annex F).
"ERROR XXX" is dis-
played. Audible alarm is
activated.
14 When fixing the pump The plastic cap (Fig. 2 Remove the cross knob of the
to a stand, the stand pos. 4) of the cross knob is mounting clamp rotating it
surface is damaged cracked. counterclockwise until the
cap falls down. Rotate the
cross knob clockwise with the
cylinder of 20mm x 20mm
inserted and pressing the new
cap between the cross knob
30
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
No SYMPTOM CAUSE CORRECTIVE ACTIONS
screw and the cylinder until
the cap seats on the screw.
31
Service manual
10. PUMP REPAIR
32
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
10.4. REPLACING OPTICAL SENSOR TAX1
(FIG. 4)
u Disassemble the pump according to the section 10.1. Re-
move the screw (5) and then remove the optical sensor
(6). Note the order of conductor connection and unsolder
them. Re-solder the conductors to the new optical sensor
(6) in the same order. Pull insulation tubes onto the sol-
dering places. Spread an universal glue onto the optical
sensor (6). Fasten the optical sensor (6) by means of the
screw (5) so as to maintain the distance between the op-
tical sensor (6) and the coupling (4) equal to 4 ± 0.1 mm.
Reassemble the pump in reverse order to the section 10.1.
33
Service manual
10.7. REPLACING SYRINGE SENSOR S3
(FIG. 1)
u Remove the screw (4) and the cover (1) (see Fig. 3). Dis-
connect the flexible cable (5) from the connector (4).
Remove the syringe sensor (3) with its connector (4).
Glue the new syringe sensor (3) with its connector (4) on
the same place at the same depth. Connect the flexible
cable (5) to the connector (4). Fasten the cover (1) with
the screw (4) (see Fig.3).
34
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
10.10 . R E P L A C I N G M I C R O C O N T R O L L E R
(ANNEX D)
u Disassemble the pump according to the section 10.1.
Note the orientation of the key on the body of micro-
controller U18 on the electronic board B7043. By means
of the special tool lift carefully the microcontroller U18
out from its socket on the electronic board B7043.
Insert the microcontroller into the socket on the elec-
tronic board B7043 so as to have the same orientation
of the key of its body as it was on the old micro-
controller. Reassemble the pump in reverse order to the
section 10.1.
10.11 . R E P L A C I N G S Y R I N G E S I Z E S E N S O R
(FIG. 4)
u Disassemble the pump according to the section 10.1.
Remove the screw (8) and remove the syringe clamp (9)
(see Fig. 3). Disconnect the syringe size sensor (3) con-
nector P8 from electronic board B7043 (9). Remove two
screws (1) and lift the syringe size sensor (3) with other
parts. Loosen two screws (2) and pull out the syringe
size sensor (3). Install, fasten and reconnect the new sy-
ringe size sensor (3) executing above-described actions
in reverse order. Reassemble the pump in reverse order
to the section 10.1. Perform syringe size calibration (see
Chapter 6)
10.12 . R E P L A C I N G B U Z Z E R U N I T
(FIG. 2)
u Disassemble the pump according to the section 10.1.
Detach buzer unit (17) (potentiometer and buzer) from
the housing. Glue the new buzer unit to the housing in
the same location. Glue potentiometer using ther-
moglue.
Reassemble the pump in reverse order to the section 10.1.
35
Service manual
10.13 . R E P L A C I N G P O W E R S U P P L Y I N L E T
(FIG. 2)
Disassemble the pump according to the section 10.1.
Unsolder wires from power supply inlet (1). Remove two
screws (2) and remove the power supply inlet (1). Install
the new power supply inlet (1) and re-solder wires to it
executing above-described actions in reverse order. Push
insulating tubes onto soldering places. Re-assemble the
pump in reverse order to section 10.1.
10.14 . R E P L A C I N G S U P P L Y U N I T
(FIG. 2)
u Disassemble the pump according to the section 10.1.
Disconnect connectors P13, P14 (11) and P11, P12 (6)
from supply unit (7). Remove four screws (10). Remove
the supply unit (7). Install the new supply unit (7) and
re-connect connectors executing above-described actions
in reverse order. Re-assemble the pump in reverse order
to section 10.1.
Fig.1
1. Keyboard
2. Electronic board B7043
3. Syringe sensor *
4. Connector *
5. Flexible cable *
6. Screws M4 * – optional
36
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
Fig.2
1. Power supply inlet 10. Screws ∅ 3x6,5
2. Screw M3x10 with nuts M3 11. Battery connector
3. Fuse unit with fuse T80mA 12. Battery fuse
4. Plastic cap 13. Screw ∅ 3x10
5. Cross knob 14. Cover
6. Mains connector 15. Battery
7. Supply unit 16. Buzzer connector
8. Fuse T1A 17. Buzzer
9. Supply unit connector 18. Filter unit
37
Service manual
Fig.3
1. Cover 7. Housing
2. Screw M3x8 8. Screw M2,5x12
3. Plate 172 9. Syringe clamp
4. Screw ∅ 3x10 10. Tube
5. Support 11. Screws M3x16
6. Rubber bellows 12. Plate 146
38
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
Fig.4
1. Screw M4x8
2. Screw M3x10
3. Syringe size sensor
4. Coupling
5. Screw M3x8
6. Optical sensor
7. MFC
8. Prealarm switch S2
9. Electronic board B7043
10. Screws ∅ 3x6,5
11. End switch S1
39
Service manual
ANNEX A
40
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
*
* – optional
41
Service manual
ANNEX B
42
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
* – optional
43
Service manual
ANNEX C
INTERCONNECTION
DIAGRAM
OF THE PUMPS
44
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
* – optional
45
Service manual
ANNEX D
46
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
47
Service manual
ANNEX D
48
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
49
Service manual
ANNEX D
50
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
51
Service manual
ANNEX E
52
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
* – optional
53
Service manual
ANNEX F
Connection data:
Ground (GND)
Normally Closed (NC)
Common (COM)
Normally Open (NO)
* – optional
NOTE:
Use standard D Type — 9 pin connectors.
54
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
ANNEX G
DESCRIPTION OF ERRORS
NOTE:
When eliminated fault repeats consult the manufacturer.
55
Service manual
ANNEX H
Service
Manual Parameter Value
section
7.2 Infusion volume accuracy %
7.3 Infusion rate accuracy %
7.4 Bolus volume accuracy %
7.5 Occlusion pressure accuracy: Low kPa
Med kPa
High kPa
Service
Manual Alarm signal Comment
section
8(1) OCCLUSION!!!
8(2) OCCLUSION or END
8(3) NO MAINS!!! Check power cord
8(4) LOW BATTERY
8(5) VERY LOW BATTERY
8(6) X min. PREALARM!
8(7) END of INFUSION
8(8) KOR X.X ml/h
8(9) SYRINGE EMPTY!
8(10) CLAMP OPENED!
8(11) ILLEGAL SYRINGE! Change SYRINGE!
8(13) ATTENTION! 2 min INACTIVE
8(14) STANDBY TIME ELAPSED
56
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
ANNEX I
COMPONENT LIST
No Description Code
1 Battery GB1, GB2 B6640004
2 Rubber bellows B8703002
3 Optical sensor TAX1 V5440101
4 End S1 and prealarm S2 switch V5500900
5 Syringe sensor S3 B6660016
6 Electronic board B7043 B3087043
7 Keyboard V6675033 (for SEP-10S Plus)
V6675034 (for SP-12S Pro)
V6675042 (for SEP-12S Plus)
8 Microcontroller U18 B6270010
9 Syringe size sensor B6666002
10 Buzzer unit Z1 B6710005 (with volume control)
B6710006 (without volume control)
11 Power supply inlet V5520180
12 Supply unit B6340011
13 Fuse T80 mA F1 V5501002 (for 230 VAC mains)
Fuse T160 mA F1 V5501000 (for 115 VAC mains)
14 Fuse T1A F1, F2(A2) V5501020
15 Plastic cap B8123009
16 Filter unit B6730006
17 Cross knob B6370007
18 Display MD1 V5500442
19 LED yellow D26, D27, D28 V5430100
20 LED green D29, D30 V5430200
21 Mains cable V5570010 (European style)
22 Syringe clamp B8127002
23 Plate 172 B8600172
24 Plate 146 B8600146
25 Support B8126028 (without tooth)
B8126046 (with tooth)
57
Service manual
ANNEX J
ACCESSORIES
58
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps
ANNEX K
COMMUNICATION PROTOCOL
TECHNICAL DESCRIPTION
u Communication protocol is used for downloading „Event
Log” from the syringe pump in response to commands
from a computer.
u The protocol uses a variation of the HDLC format for
data transfer. Following format is used to transfer com-
mands from computer to the syringe pump:
BYTES CODE DESCRIPTION
1 0xC0 Flag
1 0x02 —
1 0x13 Control
1 0x00 Status (reserved)
1 — Command code (see below)
1 0x00 —
2 — CRC
1 0xC1 Flag
u Following commands are used for downloading „Event
Log” from syringe pump to computer:
CODE DESCRIPTION
0x01 Send Model and SN
0x30 Send Event Log Size
0x31 Send (repeat) frame
0x32 Send next frame
0x33 Quit
u Following format is used for transferring data from sy-
ringe pump to computer:
BYTES CODE DESCRIPTION
1 0xC0 Flag
1 0x03 —
1 0x13 Control
1 0x00 Status (reserved)
1 — Code of command from PC pump is responding to
1 0x00 —
45 — Data (5 events)
2 — CRC
1 0xC1 Flag
59
Service manual
u Technical specification of the RS232 Interface is shown
in the 1 table
1 table
RS 232 Specification
Connector 9-pin D-Socket
TX Pin 3
RX Pin 2
Baud rate 9600 baud
Bit format 1 start, 8 data, no parity, 1 stop
Character format Binary/ASCII (see below)
Ground Pin 5
NOTE:
ASCII is used to transfer information regarding pump
model in response to command „Send Model and SN”
2 table
PC Pump
Pin 3 (Tx) Pin 2 (Rx)
Pin 2 (Rx) Pin 3 (Tx)
Pin 5 (Ground) Pin 5 (GND)
Pin 4 Nc Pin 4, Pin 7 (Jumper)
Pin 7 Nc
60
SP-12S Pro, SEP-12S Plus and SEP-10S Plus syringe infusion pumps