Certificate No: 1206290207 Job Dated: 29/05/2012
Job No: 120629090
Date of Calibration: 29/06/2012 Due Date: 29/06/2013 Page: 01 of 02
Customer Name & Address NAVYUGA ENGINEERING CO. LTD.
NOWGAM VILLAGE, BANIHAL TEHSIL,
RAMBAN DISTRICT, JAMMU-182146
Instrument Description
‘NAME ‘Temp. Indicator with Sensor (Oven)
MAKE Accrotecl Scientific Industries |
MODEL/TYPE : o
SLNO. : ~
1D.NO. : oy-o1 r
RANGEISIZE 5 0-250°¢
LEAST COUNT é
LOCATION
REF. CAL PROCEDURE! IS soP- T0903
‘CONDITION OF EQUIPMENT RECEIVED: OK
Environmental Condition i TEMPERATURE : (25 + 2)°C
HUMIDITY G0 +15)%
Standard Instruments Used
Name DMM with RTD Sensor
Make Fluke
Calibrated By: ERTL
Certificate No.: ERTL(NY'90(4)-2011-12/A2847
Valid up to: (09-04-2013
Notes
1. This Certificate refers only to the particular Item submitted for calibration & result reported inthe
certificate are valid atthe time of and under the stated conditions of measurement.
2. This Certificate shall not be reproduced, except in full without the permission of Managing Director
Speciro Analytical Labs Ltd. New Delhi, ~
th
cungugn? Mubarsaneess Meiers
= SubeetioTemee oa Conditions Overieat- 44, Okhia Indl. Area, Phase-ll,
New Dethi-110020 (INDIA)
Ph: 91-11-40522000, 41611000
Fax: 91-11-405031:
E-mail: care@spactro.in
URL: www.speciro.
150, 40503154
Job Dated: 29)06/2012 Jab No: 120629050
Date of Calibration: 29/06/2012 Due Date: 29/062013 Page: 02 of 02
Calibration Resulis,
* i ertaini
8.No. MUG Resins Corrected Std. Reading ((C) | PxPanded Uncertainity
CO) (C)
1 50 49.871 0.18
2, 100 99.957 0.25
3. 150 149,889 0.25
4 200 199.821 0.25
5. 230 299.674 0.75
Remarks:
1. *UUC: Unit Under Calibration
2. The Estimated Expanded Uncertainty has been caleulated at 95 % (approx.) Confidence level
with Coverage factor k =2 e
Nokes
'. This Certificate refers only tothe particular Item submitted for calibration & result reported inthe
Certificate are valid at the time of and under the siated conditions of measurement,
7. This Certificate shall uot be reproduced, except in full without the permission of Managing Director
Spectro Analytical Labs Lid. New Delhi
Andlyst °
cL-N04651.