IMPORTANT AOC-20I/S FUNCTIONS

FUNCTION STAND ALONE W/AUTOSAMPLER W/DUAL TOWER CLASS VP CLASS 5K


F31 0 0 0 0 0
F40 0 1 1 0 0
F41 0 0 1 0 0
F50 N/A N/A 0-8 (1) N/A N/A
F51 N/A N/A 1 (2) N/A N/A
F52 N/A N/A 76 (2) N/A N/A
F53 N/A N/A 49 (2) N/A N/A
F62 0 0 1 0 0
F80 2 N/A N/A 2 2
F81 0 N/A N/A 0 0
F82 1 N/A N/A 1 1
F83 1 N/A N/A 0 1
F84 1 N/A N/A 0 0
F85 0 N/A N/A 0 0
F92 0 0 1 0 0
F93 (3) 0 1 1 0 0
F94 (4) 1 1 1 1 1

1. Function 50: “000” Odd numbered vials to main, even numbered vials to sub (slave).
“001” Even numbered vials to main, odd numbered vials to sub.
“002” All vials to main injector.
“003” All vials to sub-injector.
“004” Same vial first to main injector, then to sub-injector.
“005” Vial Nos. 1 to 75 to main injector and Nos. 76 to 150 to sub-injector.
“006” Vial Nos. 76 to 150 to main injector and Nos. 1 to 50 to sub-injector.
“007” 1.5 mL vials to main injector and 4 mL vials to sub-injector.
“008” 4 mL vials to main injector and 1.5 mL vials to sub-injector.

2. Functions 51, 52, and 53: These Functions are used only for dual injectors and only in Stand Alone.

3. Function 93: This Function is set to 0 (default), when the AOC-20S is being used.

4. Function 94: This Function sets the injection position on the GC-14 or GC-17.
When using the dual tower, do not set slave to F94 = 2. Leave as default.

5. Class 5K: Functions 83 and 84 are set to these parameters only when using Class5K Ver. 2.20 or later.
When using Class 5K Ver. 2.11 or earlier, set F 83 = 0, and F 84 = 1.