You are on page 1of 9

Drug Testing

Research article and Analysis

Received: 8 November 2016 Revised: 9 February 2017 Accepted: 19 March 2017 Published online in Wiley Online Library: 12 May 2017

(www.drugtestinganalysis.com) DOI 10.1002/dta.2195

Development of a dried blood spot HPLC-PDA


method for the analysis of linezolid and
ciprofloxacin in hospital-acquired
pneumonia patients
Vincenzo Ferrone,a Maura Carlucci,b Roberto Cotellese,b Paolo Raimondi,b
Annadomenica Cichella,b Lorenzo Di Marco,a Salvatore Genovesea
and Giuseppe Carluccia*
This study developed a high performance liquid chromatography (HPLC) method involving dried blood spotting (DBS) as a
sampling method for the therapeutic drug monitoring of antimicrobial combination therapy with linezolid and ciprofloxacin.
DBS for standards, quality control samples, and patient samples was excised and then extracted using a mixture of
methanol/water/formic acid 80:20:0.1 (v/v/v), respectively. Linezolid (LZD) and ciprofloxacin (CPR) were measured by HPLC
using a Kinetex EVO C18 (100 × 4.6 mm I.D. 2.6 μm particle size). Mobile phase consisted of 10 mM ammonium acetate (A)
and a mixture of acetonitrile and methanol (B), both containing 0.1% triethylamine, in gradient elution. Detection was carried
out at 251 nm for linezolid (LZD) and 277 for ciprofloxacin (CPR). Ulifloxacin was used as an internal standard. The internal
standard, LZD, and CPR were eluted in 7.90, 7.18, and 8.89 min, respectively. Total run-time was 20 min. Calibration curves
were constructed in the range of 0.05–30 μg/mL for LZD and 0.02–10 μg/mL for CPR, respectively. The intra- and inter-day
precision (RSD values) did not exceed 9.43%, the intra- and inter-day accuracy (accuracy %) ranged between 96.2 and
106.2%. Haematocrit (Hct) effects were investigated to obtain a linear correlation between haematocrit values and volume
of blood. Copyright © 2017 John Wiley & Sons, Ltd.

Keywords: linezolid; ciprofloxacin; dried blood spot; method validation; HPLC-PDA

Introduction quinolone derivative obtained by addition of a fluorine atom in


position 6 and a piperazine substituent in position 7 with different
Pneumonia is a respiratory infection of the lung affecting primarily chemical structure and spectrum of activity.[9] CPR stabilizes
the microscopic air sacs known as alveoli.[1] It is defined as being enzyme/cleaved DNA complexes within the cell, terminating DNA
nosocomial or hospital-acquired pneumonia (HAP) if it becomes replication. Ciprofloxacin, alone or in combination with other
apparent more than 48 h after hospital admission.[2] HAP affects antimicrobial drugs, provides effective treatment for a variety of
0.5 to 1% of inpatients and is the most common healthcare- infections, in particular urinary, respiratory, and gastrointestinal
associated infection contributing to death.[3] Pneumonia in tract.[10] Fluoroquinolones generally show concentration-
ventilated patients is defined as ventilator-associated pneumonia dependent bacteria-killing activity that depends on the ratio of
(VAP) and is considered a subset of HAP. The mortality rate of maximum drug concentrations to minimum inhibitory
patients with VAP is 24% to 50% which increases to 76% if infection concentration.[11]
is caused by multidrug-resistant pathogens.[4] To prevent multidrug Different methods describing the determination of
resistance, an antimicrobial combination therapy may be used. fluoroquinolones or LZD concentrations in biofluids have been
Linezolid (LZD) or (N-[[(5S)-3-[3-Fluoro-4-(4-morpholinyl)phenyl]-2- reported.[12–20]
oxo-5-oxazolidinyl]methyl]acetamide) (Figure 1a) is an High performance liquid chromatography (HPLC) with UV,
antimicrobial agent from the oxazolidinone class. It is especially fluorescence, mass spectrometry (MS), or tandem mass
active against multidrug-resistant Gram-positive organisms such
as methicillin-resistant Staphylococcus aureus (MRSA) and certain
* Correspondence to: Giuseppe Carlucci, Dipartimento di Farmacia, Università
Gram-negative and anaerobic bacteria.[5,6] degli Studi ‘G. d’Annunzio’ Chieti-Pescara, Via dei Vestini, 66100 Chieti, Italy.
LZD inhibits bacterial ribosomes by binding to the 23S ribosome E-mail: g.carlucci@unich.it
and preventing the 30S–50S fusion.[7] The achievement of a
minimum plasma concentration (Cmin) >2 μg/mL has been a Department of Pharmacy, Università degli Studi, G.d’Annunzio’ Chieti-Pescara,
via dei Vestini, 66100 Chieti, Italy
proposed as theoretical threshold to ensure LZD efficacy.[8]
Ciprofloxacin (CPR) or N (1-cyclopropyl-6-fluoro-4-oxo-7-
1611

b Department of Oral Health Sciences and Biotechnology, Università degli Studi, G.


(piperazin-1-yl)-quinoline-3-carboxylic acid) (Figure 1b) is a d’Annunzio’ Chieti-Pescara, via dei Vestini, 66100 Chieti, Italy

Drug Test. Analysis 2017, 9, 1611–1619 Copyright © 2017 John Wiley & Sons, Ltd.
Drug Testing
and Analysis V. Ferrone et al.

by Suzhou Bichal Biological Technology Co. Ltd (Jiangsu, China).


Methanol, acetonitrile, triethylamine, and ammonium acetate were
obtained from Carlo Erba Reagents (Milan, Italy). Hydrochloric acid
was from VWR International (Milan, Italy). All chemicals were of
analytical-reagent grade or better.
All dilutions were prepared with HPLC-grade water obtained by
passage through an Elix 3 and Milli-Q Academic water purification
system (Millipore, Bedford, MA, USA). Bond Elut DMS card was
supplied by Agilent Technologies (Santa Clara, CA, USA). Blank
whole blood from healthy donors was kindly supplied by S.S.
Annunziata Hospital Transfusion Centre (Chieti, Italy).

Chromatographic conditions
Figure 1. Chemical structure of (a) linezolid, (b) ciprofloxacin, and (c)
ulifloxacin, the internal standard. The chromatographic apparatus for the determination of these
analytes was a Waters system consisting of a model 600 pump
spectrometry (MS/MS) detection are the techniques most used. and an UV–Vis photodiode array detector model 2996 (Waters,
Conventional solid-phase extraction (SPE) and liquid–liquid Milford, MA, USA). A Rheodyne model 7125i injector (Rheodyne,
extraction (LLE), although being efficient methods to extract LZD Cotati, CA, USA) equipped with 20 μL loop and a degasser system
and CPR from biological fluids, are impractical in routine analyses model DG-4400 (Phenomenex, Torrance, CA, USA) was used.
as they are laborious and solvent- and time-consuming. In addition, Chromatographic separation was achieved using a Kinetex EVO
some of them have deleterious effects on the environment due to C18 (100 × 4.6 mm I.D. 2.6 μm particle size) column protected by
the large quantities of organic solvents that need to be used and a disposable Security Guard Gemini precolumn (3.0 × 4.0 mm)
discarded. (Phenomenex, Torrance, CA, USA) maintained at 25 ± 1°C, using a
Due to the limitation of performing therapeutic drug monitoring thermostatically-controlled column heater. Mobile phase consisted
in hospitals which do not have a therapeutic drug monitoring in 10 mM ammonium acetate buffer adjusted to pH 3.5 with acetic
(TDM) laboratory, dried samples spot devices are becoming a valid acid (phase A) and a mixture of acetonitrile and methanol in a ratio
option to overcome these problems. In recent years, dried blood of 80/20 (v/v) (phase B), both phases were added with 0.1% (v/v) of
spots (DBS) have gained increasing attention due to the triethylamine. To perform the best separation of the analytes, a
development of new applications. Bang was the first to make linear gradient elution program was used; the gradient began from
practical and effective use of DBS for the determination of blood 0 to 4 min with 90% and 10% of phase A and B, respectively, from 5
sugar levels. However, this sampling technique did not achieve its to 10 min became 65% and 35% of phase A and B, respectively, and
breakthrough as a cheap screening method until the 1960s, when returned to the original conditions between 11 and 13 min
it was applied for the detection of phenylketonuria.[21] Less invasive followed by 7 min of re-equilibration of the column to the initial
sampling, simplified sample preparation, and easier storage are the condition. The flow rate was set at 1.0 mL/min. The solvents were
reasons for the recent increase of interest in DBS techniques. filtered before use through a 0.45 μm WTP membrane, while
Despite the numerous applications using DBS,[22–26] the exact ammonium acetate solution was filtered through a WCN 0.5 μm
quantification is still a frequently discussed issue.[27–31] membrane (Whatmann, Maidstone, UK). Detection was carried
A first concern is if and how venous and capillary blood out at 251 nm for LZD and 277 for CPR. Empower v.2 Software
concentrations of analyte correlate.[29] Another challenge in DBS- (Waters Spa, Milford, MA, USA) was used for setting-up the analysis
based sampling is the spot volume, because in clinical samples and for data management.
obtained from a finger-prick, the blood volume is not controlled
and will vary.[28,30] Other issues to be investigated are the site of
Preparation of calibration standards and quality controls
punching and the haematocrit (Hct) effect. The latter is the most
widely discussed DBS-related problem due to the accuracy of the Stock solutions of 2 mg/mL of LZD, CPR and I.S. were prepared by
Hct measurement and to the blood/plasma distribution ratio of appropriate dissolution of the reference powders into methanol.
an analyte. To date, the spot volume, the site of punching, and Then the stock solutions were diluted with Milli-Q water to obtain
the Hct effect must be evaluated when a DBS sampling technique the working solutions. Calibration curve was prepared by spotting
is performed. [28–31] on filter paper spiked control blood with LZD, CPR, and I.S. to obtain
At present, no methods for the quantification of both CPR and a concentration of 0.05, 0.15, 0.50, 2.00, 5.00, 15.00, and 30.00 μg/
LZD in DBS are reported. The aim of this study is to describe the first mL for linezolid and 0.02, 0.10, 0.25, 0.50, 1.00, 2.50, and 10.00 μg/
set-up, application, and validation of an HPLC method to quantify mL for ciprofloxacin. Quality control samples were prepared in a
the TDM of LZD and CPR involving dried blood spot as a similar way at the concentration of 0.10, 1.00, and 25.00 μg/mL
biosampling procedure. for LCD and 0.05, 0.75, and 7.50 μg/mL for CPR, the I.S. (Figure 1c)
concentration was 5 μg/mL. The I.S. was chosen because it is not
co-administrated with the investigated analyte and is not abused
Experimental design by the patients.
Chemicals and reagents
Sample preparation
LZD, (CAS 165800–03-3) and CPR (CAS 85721–33-1) were
purchased from Sigma-Aldrich (Milan, Italy), while ulifloxacin (CAS Venous blood was taken and collected in EDTA-coated 7.5
1612

no. 112984–60-8), used as the internal standard (I.S.), was supplied ampoules and was used for the DBS-specific validation except for

wileyonlinelibrary.com/journal/dta Copyright © 2017 John Wiley & Sons, Ltd. Drug Test. Analysis 2017, 9, 1611–1619
Drug Testing
DBS-HPLC-PDA analysis of linezolid and ciprofloxacin and Analysis

the correlation of the capillary dried blood spot (CDBS) with venous Linearity
dried blood spot (VDBS) where punched discs of VDBS and CDBS
Linearity was evaluated using freshly prepared spiked matrix
from patients were analyzed for their CPR and LZD levels.
samples in the range of 0.05–30 μg/mL for LZD and 0.02–10 μg/
DBS samples were collected at the same time as routine TDM
mL for CPR. Each calibration curve consisted in a seven-calibrator
samples for plasma and did not require extra visits to the clinic or
concentration. Daily calibration curves were constructed using the
additional fingerpricks.
ratios of the observed peak areas of LZD and CPR to the I.S., to
Fingerprick blood was directly spotted onto the paper cards by
describe the relationships between concentration of the analytes
placing two-to-three drops of blood on the DBS card, omitting a
and the detector response, a linear regression analysis with
direct contact between the finger and the card. To prepare DBS
weighting factors consisted in 1/x2 values.
spiked samples, 25 μL of blood, previously spiked with the analytes
and the I.S., was transferred onto a Bond Elut DMS card by an
Eppendorf pipette, and was left to dry 3 h at room temperature
and then transferred into a sealed plastic bag containing desiccant Precision and accuracy
and stored at 20C° until analysis. Validation was performed using Accuracy and precision were evaluated in six replicates at lower
blood with Htc value of 45% except for the evaluation of the Htc limit of quantification (LLOQ), low-level quality control (LQC),
percentage effect on the analysis. Extraction from DBS was medium-level quality control (MQC), and high-level quality control
performed using a mixture of methanol/water/formic acid (HQC) on the same day and three analytical batches of QCs level
80:20:0.1 (v/v/v), respectively. on six consecutive days, respectively. Accuracy was determined
A 9-mm diameter disk was punched out from DBS and by the mean of the measured QC concentration to the theoretical
transferred into a 2.0 mL Eppendorf containing 300 μL of extraction QC value and reported in percentage. Precision was defined by
solution and was sonicated for 25 min at 35°C followed by the relative standard deviation percentage (RSD%).
centrifugation at 4000 rpm for 10 min to obtain a clear supernatant.
The supernatant was then transferred into a clean tube, evaporated
to dryness under a gentle stream of nitrogen and reconstituted to
Limit of detection and limit of quantification
25 μL with mobile phase, 20 μL was injected into the HPLC system.
The limit of detection (LOD) was defined as the concentration
which gave a signal-to-noise ratio of 3 while the limit of
Method validation quantification was determined by injecting six times a spiked
sample with decreasing concentration of the analyte. LOQ was
The proposed method was validated according to the guidance for regarded as the minimum amount of analytes that gave a precise
industry on the validation of bioanalytical methods.[32] measurement (accuracy within ±20% and RSD% <20%).

Figure 2. A typical chromatogram of LZD, CPR, and I.S. after injection of: (a) a blank DBS extract, (b) a sample of patient in standard therapy with intravenous
1613

linezolid (Zyvox 600 mg twice daily) and ciprofloxacin (ciproxin 400 mg twice daily) 4 h after administration of these drugs, (c) a blank DBS extract spiked with
5 μg/mL of internal standard and both LZD and CPR.

Drug Test. Analysis 2017, 9, 1611–1619 Copyright © 2017 John Wiley & Sons, Ltd. wileyonlinelibrary.com/journal/dta
Drug Testing
and Analysis V. Ferrone et al.

Selectivity form of pretreatment, which can retain several matrix compounds


in the paper.
The selectivity of the method was tested by comparing the
Choosing the most suitable extraction solvent is of primary
chromatograms of six different batches of blank plasma.
importance to achieve good extractability and selectivity of the
Figure 2 shows a representative chromatogram of a blank plasma
target analytes in DBS. First, to select the extraction solvent,
sample. There was no chromatographic interference from
methanol ethanol and acetonitrile were tested using 300 μL for
endogenous compounds at the retention times of linezolid and
each organic solvent. Methanol was chosen due to the best
ciprofloxacin. The chromatographic peaks were well resolved to
extraction yield compared to acetonitrile and ethanol. The
the baseline. During the selectivity assay, other co-administered
extraction solution was chosen by testing different solvent
drugs in spiked blood which were not found to interfere with
compositions, at various percentages of organic composition;
the assay.
methanol/water/formic acid 80:20:0.1 (v/v/v), respectively, gave
the best extraction recovery, as shown in Figure 3. Different
Recovery extraction times were tested, from 5 to 50 min every 5 min. A
plateau in the extraction yield was reached after 25 min
Recovery of LZD and CPR was evaluated by comparing the mean (Figure 4), therefore 25 min were chosen as optimum for the
peak areas of different post extracted QC samples with those extraction time. The last parameter tested was the extraction
obtained by spiking the post extracted blank matrix with the temperature, in the range of 25–50°C, every 5°C; the optimum
analytes at the QCs levels of concentration. was found to be 35°C.
Experiments regarding the homogeneity of drug distribution
Stability were carried out, and no significant differences were found, which
was due to the low blood volume spotted (25 μL).
Stability of LZD and CPR were investigated during sample
collection, after long-term storage and short-term storage, and
through several freeze and thaw cycles according to the guidance
for industry on the validation of bioanalytical methods.[32]

DBS-specific validation
A spot volume range of at least 15–40 μL should be investigated
during the validation. The spotting volumes should be tested at a
minimum of two concentration levels (low and high) in triplicate.
In all cases, the spot volume for calibration standard should be
equal to the expected spot volume of the clinical samples. When
these tests show that the effect of spot volume falls within
acceptable ranges, there is no need to use accurately spotted
volumes for clinical application. [30]
Possible differences between punches taken from the centre or
from the periphery of the spot should be investigated during DBS
method development and validation. Spot homogeneity was
investigated by preparing DBS at two concentration levels (low
Figure 3. Evaluation of the extraction yield as a function of the percentage
and high) in triplicate and punching from the centre and the of methanol in the extraction solution.
periphery of the DBS. When deviated analysis results are observed
from the peripheral punches, a bigger puncher could be tested to
overcome this problem. Whole blood was centrifuged at 1500 g
for 15 min to separate the plasma and red blood cells. Increasing
volumes of plasma were added to whole blood providing
decreasing Hct levels within 30–55%. The non-centrifuged whole
blood with 45% Hct served as 100% control. These were then
analyzed for %Hct levels on an ABX Pentra XL 80. The blood was
then spiked with CPR and LZD, providing 3 QC concentration
(low, medium, and high) at each investigated Hct value, and
spotted onto collection cards. Punches (9 mm) taken from the
centre of each spot were subsequently analyzed.

Results and discussion


Optimization of extraction from DBS
The use of DBS is a valid alternative compared to traditional blood
sampling; the analysis of a few drops of blood can also be suitable
for pharmacokinetic studies, TDM, and analyses involving children.
1614

The simple blood adsorption onto the paper can be considered a Figure 4. Extraction yield % as a function of the extraction time.

wileyonlinelibrary.com/journal/dta Copyright © 2017 John Wiley & Sons, Ltd. Drug Test. Analysis 2017, 9, 1611–1619
Drug Testing
DBS-HPLC-PDA analysis of linezolid and ciprofloxacin and Analysis

Evaluation of haematocrit effect, spot volume, and punch very broad, which made their integration rather difficult. Good
localization separation of all analytes and internal standard was obtained using
a Kinetex EVO C18 column. This semi-porous silica column (packed
Haematocrit (Hct) may have an important role on the spot area and
with core-shell particles) provided a shorter diffusion path for
in the matrix effect, therefore these parameters should be
analytes, minimised peak broadening at flow rates and had a high
evaluated. As the Hct levels change, the viscosity of the blood also
permeability. Compared to the results obtained using other
changes. The variability of viscosity leads to differences in flux and
columns, the application of Kinetex EVO C18 afforded higher
diffusion properties of blood, which can directly impact the size of
separation efficiency. Hence, it was possible to analyse LZD and
the blood spots formed. The use of a non-cellulose paper as Bond
CPR more rapidly, with good resolution and peak shape. To achieve
Elut DMS card may decrease the bioanalytical bias introduced as
the efficient separation of the analyzed compounds and the IS,
a haematocrit variability, but it must be evaluated. According to a
different mobile phases (methanol, acetonitrile, water with additive
previous study,[34] different Hct values: 30% to 55%, respectively,
ammonium acetate), first in isocratic mode, then with a gradient
were investigated to obtain a linear correlation between
programme, were tested. Isocratic separation was found the be
haematocrit values and volume of blood. The haematocrit effect
not suitable for the analysis of linezolid and ciprofloxacin due to
is described by the following equation:
the overlapping of the drug peaks. It was determined that the
optimal mobile phase consisted of a mixture of methanol-
acetonitrile and 10 mM ammonium acetate buffer (pH 3.5) in a
V Std
C corrected ¼ C observed (1) gradient elution programme. Gradient elution is typically used to
V Std þ bðHct  45Þ
improve the peak separation and eliminate significant matrix
effects due to co-elution of the analyte and the endogenous plasma
components. Furthermore, to increase the peak shape, different
where Ccorrected is the concentration after the correction for Hct triethylamine (TEA) percentages in the range of 0.5 and 1% (v/v)
value, Cobserved is the found concentration before its correction, were added to the mobile phase. The optimum was achieved by
Vstd is the blood volume at the standardized condition (45% Hct) using 1% (v/v) of TEA. Under this condition, the peak shape of the
and b is the regression coefficient between blood volume and analytes was symmetric and the retention times of ciprofloxacin
haematocrit percentage. and linezolid were well separated. To detect the analytes with high
The equation obtained between Vest and Hct is: sensitivity, each drug was monitored at its maximum wavelength.
Vest = 20.93 + 0.1133 x (Hct-45) where b value is the slope By applying the chromatographic condition herein reported, the
(0.1133) and Vstd the intercept (20.93). total run time in this assay was 20 min and the retention times for
Using uncorrecting concentration between the high and low Hct, CRP, LZD were 7.1 min and 8.9 min, respectively. The retention time
the error was approximately ±27%, while correcting values for Hct for internal standard was 7.9 min. The retention times had
percentage using Eqn (1), the bias were lowered to not more than consistently excellent reproducibility of less than 0.16%. The
10% according to the Food and Drug Administration (FDA) resolution of the developed HPLC-PDA method was satisfactory,
guidelines which request ±15%.[32] as all analytes were well base-line separated. The peaks were
The accuracy and precision of quantitation of LZD and CPR symmetric with a symmetry factor As ≤1.34. All the evaluate
extracted from a 9-mm diameter punch taken from a DBS sample parameters were within the required range, as established in FDA
derived from 15, 25, and 40 μL aliquots of human blood were guidelines.[32]
investigated. For all examined blood samples, accuracy and
precision were less than the defined acceptance criteria of 15%.
Furthermore, compared to those from 25 μL, the difference Linearity
between the accuracy values for the extracts from 15 and 40 μL
were all less than 7.3%. This indicates that there is no notable Linearity was evaluated using freshly prepared spiked matrix
difference in distribution of analyte and blood across spots when samples in the range of 0.05–30 μg/mL for LZD and 0.02–
a 9-mm punch is taken from the centre of the blood spot. 10 μg/mL for CPR. The correlation between LZD and CPR
concentration (x) and the ratios of the observed peak areas of
linezolid and ciprofloxacin to the internal standard (y) led to a
Chromatographic condition
mean slope (±S.D.) of 0.093 (± 0.004) for LZD and 0.182
During the method development, four reversed-phase columns (± 0.008) for CPR, respectively, and a mean y intercept (±S.D.) of
Spherisorb ODS (150 × 4.6 mm I.D. 5 μm), Gemini C18 0.0019 (± 0.0001) for LZD and 0.00257 (±0.00010) for CPR. Each
(150 × 4.6 mm I.D. 5 μm), Supelcosil LC-18 (150 × 4.0 mm I.D. calibration curve consisted of a seven-calibrator concentration.
5 μm), and Kinetex EVO C18 (100 × 4.6 mm I.D. 2.6 μm) were tested To describe the relationships between concentration of the
with different mobile phase compositions. LZD and CPR were not analytes and the detector response, a linear regression analysis
separated with Spherisorb ODS and Supelcosil columns. Using with weighting factors consisting in 1/x2 values was used. The
Gemini C18, the peak shapes for all the analysed compounds were results are shown in Table 1.

Table 1. Mean linear regression parameters for linezolid and ciprofloxacin obtained using seven calibrators (n = 6)
2
Compound Slope (± S.D) Intercept (± S.D) r Range (μg/mL) Weighting LOQ (μg/mL) LOD (μg/mL)
2
Linezolid 0.093 ± (0.004) 0.0019 ± (0.0001) 0.9995 0.05–30 1/x 0.05 0.015
2
1615

Ciprofloxacin 0.182 ± (0.008) 0.0257 ± (0.0010) 0.9998 0.02–10 1/x 0.02 0.007

Drug Test. Analysis 2017, 9, 1611–1619 Copyright © 2017 John Wiley & Sons, Ltd. wileyonlinelibrary.com/journal/dta
Drug Testing
and Analysis V. Ferrone et al.

Selectivity Table 3. Recoveries of linezolid and ciprofloxacin in dried blood spot

The analytical method should be able to differentiate the analytes Nominal Mean Measured Recovery RSD%
of interest and I.S. from endogenous components in the matrix or concentration concentration (%)
(μg/mL) ± S.D.
other components in the sample. Normally, absence of interfering
components is accepted where the response is less than 20% of LINEZOLID
the lower limit of quantification for the analytes and 5% for the 0.10 0.096 ± 0.006 96.7 6.25
internal standard. Six independent blank dried blood spot 1.00 0.97 ± 0.08 97.3 8.24
samples were analyzed and no interferences were observed at 25.00 24.31 ± 0.34 97.4 1.39
the retention time of the selected analytes and of the internal CIPROFLOXACIN
standard. Resolution values ranged from 1.87 (± 0.04) for CPR/I. 0.05 0.046 ± 0.004 92.4 8.64
S. to 2.33 (±0.06) for LZD/I.S. at the higher point of the calibration 0.75 0.71 ± 0.05 94.67 7.04
curve. A representative chromatogram of a blank dried blood 7.50 7.19 ± 0.44 95.88 6.11
spot sample is shown in Figure 2a, while a blank spiked with
the internal standard and both LZD and CPR are shown in
Figure 2c.
except for LLOQ which should be within 20% of the nominal value.
Precision, expressed as relative standard deviation percent (RSD%),
Limit of detection (LOD) and limit of quantification (LOQ) should be demonstrated within a single run (intra-day precision)
The limit of detection (LOD) was the spiked blood concentration and in different runs (inter-day precision) using the same runs
that had a signal-to-noise ratio > 3, while the limit of quantification and data for the demonstration of accuracy. RSD% should not
(LOQ) was considered the lowest standard of the calibration curve, exceed 15% for the QC samples, except for the LLOQ which should
determined by injecting a spiked sample six times with decreasing not exceed 20%. Intra-day precision and accuracy were evaluated
concentrations of the analytes. The LOQs of the method were by analyzing six times in the same day blank samples spiked with
0.02 μg/mL and 0.05 μg/mL for CPR and LZD, respectively, with the studied analytes at 3 different concentration levels (LLOQ 0.05
good precision (RSD ≤ 9.43 for LZD and ≤7.42 for CPR) and accuracy QCL 0.10, QCM 1.00, and QCH 25.00 μg/mL) for LZD and (LLOQ
(within 92.6 and 106.3 for LZD and 95.4 and 104.0 for CPR). The 0.02, QCL 0.05, QCM 0.75, and QCH 7.5 μg/mL) for CPR, respectively.
LODs were 0.008 μg/mL for CPR and 0.015 μg/mL for LZD. The For inter-day precision, the assays described above were repeated
results are shown in Table 2. six times over six different days. Table 2 displays the overview of
the intra-day (RSD%), inter-day (RSD%) and mean relative errors
of the methods. The intra- and inter-day accuracies were within
Precision and accuracy
96% and 106.2%. The intra- and inter-day (RSD%) were not more
The data for intra- and inter-day precision and accuracy obtained than 9.43%. These results indicated that the validated assay was
from QC samples at four different levels of analytes (LLOQ, LQC, precise, accurate and reproducible.
MQC, and HQC) are shown in Table 2. The QC samples were
analyzed against the calibration curve, and the obtained
Extraction recovery
concentrations were compared with the nominal value. The
accuracy and precision are the most important criteria for judging Recovery yield of linezolid and ciprofloxacin from blank DBSs
the performance of an analytical method. Accuracy was reported spiked with the analytes at the QCs concentrations was evaluated;
as percentage of the nominal value and evaluated for the values recoveries for linezolid were within 96.7% and 97.3% while for
of the QC samples obtained within a single run (intra-day accuracy) ciprofloxacin were within 92.4% and 95.88%. Recovery of LZD and
and in different runs (inter-day accuracy). The mean concentration CPR was evaluated by comparing the mean peak areas of different
should be within 15% of the nominal value for the QC samples, QC samples post-extracted with those obtained by spiking the

Table 2. Intra-day and inter-day accuracy and precision for linezolid and ciprofloxacin at the LLOQ and QCs value

Theoretical Intra-day (n = 6) Inter-day (n = 6)


concentration
(μg/mL) Mean Measured Accuracy (%) Precision Mean Measured Accuracy (%) Precision (RSD%)
concentration (RSD%) concentration
(μg/mL) ± S.D. (μg/mL) ± S.D.

LINEZOLID
0.05a 0.046 ± 0.003 92.6 6.49 0.053 ± 0.004 106.2 9.43
0.10 0.096 ± 0.004 96.3 4.15 0.095 ± 0.005 95.2 5.25
1.00 1.03 ± 0.06 103.45 5.82 1.05 ± 0.06 105.3 5.71
25.00 24.87 ± 0.57 99.48 2.29 25.4 ± 1.15 101.6 4.61
CIPROFLOXACIN
0.02a 0.019 ± 0.002 95.3 7.42 0.021 ± 0.03 104.0 7.17
0.05 0.048 ± 0.003 96.3 6.25 0.053 ± 0.005 106.2 9.43
0.75 0.74 ± 0.02 98.66 2.70 0.78 ± 0.07 104.3 8.97
7.50 7.53 ± 0.19 100.4 2.52 7.51 ± 0.43 100.1 5.72
1616

a
LLOQ concentration.

wileyonlinelibrary.com/journal/dta Copyright © 2017 John Wiley & Sons, Ltd. Drug Test. Analysis 2017, 9, 1611–1619
Drug Testing
DBS-HPLC-PDA analysis of linezolid and ciprofloxacin and Analysis

Table 4. Stability of linezolid and ciprofloxacin in dried blood spot at Short-term stability in DBSs
different temperature
Spiked blank DBS were stored at room temperature and 20°C,
QCs Storage Mean Accuracy Precision +4°C and 35°C up to one month from the sampling date, then
Concentration temperature measured (%) (RSD%)
(μg/mL) (°C) concentrationa
subjected to the extraction treatment previously described.
(μg/mL) ± S.D. Stability studies showed no significant differences between room
temperature and refrigerated temperature. Nor were any
Linezolid significant differences shown when DBSs were stored at 37°C.
0.10 20 0.098 (±0.002) 98.00 2.04 These results, shown in Table 4, are very satisfactory due to the very
1.00 20 0.97 (±0.05) 97.00 5.15 low difference between DBSs stored at 20°C with the same
25.00 20 24.91 (±0.23) 99.64 0.93 sample stored at +35°C.
0.10 4 0.097 (±0.001) 97.00 1.03 The maximum loss of the two analytes being less than 7% is
1.00 4 0.99 (±0.03) 99.00 3.03 attributable to the water loss during the blood spot drying that
25.00 4 24.59 (±0.37) 98.36 1.50 minimizes most enzymatic processes responsible for the loss in
0.10 Room 0.097 (±0.003) 97.00 3.09 analyte compared to other matrices. Freeze–thaw stability of LZD
1.00 Room 0.98 (±0.06) 98.00 6.12 and CPR was determined by assaying the three QC samples in
25.00 Room 23.44 (±0.24) 93.76 1.01 triplicate over three freeze–thaw cycles.
0.10 35 0.095 (±0.003) 95.00 3.16 Three aliquots at each concentration were stored at 20° for 24 h
1.00 35 0.95 (± 0.04) 95.00 4.21 then thawed unassisted at room temperature. When completely
25.00 35 24.68 (±0.33) 98.70 1.33 thawed, the samples were refrozen under the same conditions for
Ciprofloxacin 24 h. No degradation was observed after three freeze–thaw cycles.
0.05 20 0.047 (±0.002) 94.00 4.26
0.75 -20 0.73 (±0.02) 97.33 2.74
7.50 -20 7.28 (±0.26) 97.07 3.57 Carry-over
0.05 4 0.046 (±0.003) 92.00 6.52
Carry-over should be addressed and minimised during method
0.75 4 0.71 (±0.06) 94.67 8.45
development. It was assessed by injecting blank samples after a
7.50 4 7.33 (±0.29) 97.73 3.96
calibration standard at the upper limit of quantification.32 The
0.05 Room 0.045 (±0.002) 90.00 4.44
carry-over of the blank sample following the high concentration
0.75 Room 0.73 (±0.07) 97.33 9.59
was <6% of the lower limit of quantification for LZD and CPR while
7.50 Room 7.24 (±0.15) 96.53 2.07
it was <2% for the I.S.
0.05 35 0.047 (±0.004) 94.00 8.51
0.75 35 0.69 (±0.02) 92.00 2.90
7.50 35 7.12 (±0.21) 94.93 2.95 Clinical application
a
Results are expressed as average of the analysis in triplicate. Before being applied to patients, a correlation between VDBS and
CDBS needs to be established. Hence, 9-mm punched discs of VDBS
and CDBS were analyzed for their CPR and LZD levels. The
post-extracted blank matrix with the analytes at the QCs levels of correlation coefficients (r2) of 0.964 for CPR and 0.951 for LINZ show
concentration. The. a very good comparability, indicating that finger-prick sampling is
results are reported in Table 3. The mean (±SD) extraction sufficient to measure both compound (Figure 5). In routine TDM,
recovery for IS was 93.36 ± 1.24% (n = 6), The use of an IS in the most of the drug levels are reported as plasma values. Therefore,
extraction procedure is crucial to compensate for variability in if DBS levels are reported as whole blood values, a correlation
extraction efficiency. In the present study, ulifloxacin was chosen between DBS and plasma is needed.
as the I.S. Ulifloxacin exhibits good absorbance at the wavelength To demonstrate the applicability of this assay and to
of the other analytes. It also displayed appropriate compare it against a validated method[33] with plasma matrix
chromatographic retention with its peak sufficiently separated from as the ‘gold standard’, matched plasma and DBS samples were
that of LZD and CPR (Figure 2c). analyzed.
1617

Figure 5. Concentrations of (a) CPR and (b) LZD determined from punched 9-mm DBS discs of CDBS versus VDBS.

Drug Test. Analysis 2017, 9, 1611–1619 Copyright © 2017 John Wiley & Sons, Ltd. wileyonlinelibrary.com/journal/dta
Drug Testing
and Analysis V. Ferrone et al.

Figure 6. Correlation between plasma concentrations, obtained using a validated method, of (a) CPR and (b) LZD, and DBS concentrations.

As shown in Figure 6, a good correlation between the [6] A. Marchese, G. C. Schito. The oxazolidinones as a new family of
antimicrobial agent. Clin. Microbiol. Infect. 2001, 7, 66.
DBS and the plasma concentration was found. By applying
[7] D. Livermore. Linezolid in vitro: mechanism and antibacterial spectrum.
[CPR calculated plasma] = ([CPR DBS] – 0.02869)/ 1.417 and J. Antimicrob. Chemoter. 2003, 51, 9.
[LZD calculated plasma] = ([LZD DBS] -0.3749) / 1.201 it is possible to [8] F. Pea, P. Viale, P. Cojutti, B. Del Pin, E. Zamparini, M. Furlanut.
correlate the plasma concentration of the investigated analytes Therapeutical drug monitoring may improve safety outcomes of
with their DBS concentration for the TDM. long-term treatment with linezolid in adult patients. J. Antimicr.
Therapy. 2012, 67, 2034.
The proposed method was successfully applied for the [9] P. C. Sharma, A. Jain, S. Jain. Fluoroquinolone antibacterials: a review on
quantification of linezolid and ciprofloxacin in human dried chemistry, microbiology and therapeutic prospects. Acta Pol. Pharm.
blood spot samples of patients in standard therapy with 2009, 66, 587.
intravenous LZD (Zyvox 600 mg twice daily) and CPR (ciproxin [10] A. Forrest, D. E. Nix, C. H. Ballow, M. C. Birmingham, J. J. Schentag.
Pharmacodynamics of Intravenous Ciprofloxacin in Seriously Ill
400 mg twice daily) afferent to S.S. Annunziata Hospital (Chieti, Italy)
Patients. Antimicrob. Agents Chemother. 1993, 37, 1073.
(Figure 2b). Written informed consent was obtained before blood [11] G. Hayem, P. X. Petit, M. Levacher, C. Gaudin, M. F. Kahn, J. J. Pocidalo.
sampling. Cytofluorometric analysis of chondrotoxicity of fluoroquinolone
antimicrobial agents. Antimicrob. Agents Chemoter. 1994, 38, 243.
[12] G. La Marca, F. Villanelli, S. Malvagia, D. Ombrone, S. Funghini,
M. De Gaudio, S. Fallani, M. I. Cassetta, A. Novelli, E. Chiappini,
Conclusions M. De Martino, L. Galli. Rapid and sensitive LC-MS/MS method for the
analysis of antibiotic linezolid on dried blood spot. J. Pharm. Biomed.
A simple, precise, accurate HPLC-DAD method for the Anal. 2012, 67-68, 86.
quantification of linezolid and ciprofloxacin in human dried blood [13] J. Zander, B. Maier, M. Zoller, D. Teupser, M. Vogeser. Quantification
spot sample was developed and fully validated. The proposed of linezolid in serum by LC-MS/MS using semiautomated sample
preparation and isotope dilution internal standardization. Clin.
method ensures satisfactory sensibility and specificity for the Chem. Lab. Med. 2014, 52, 381.
therapeutic drug monitoring of both LZD and CPR. The main goal [14] O. A. Phillips, M. F. Abdel-Hamid, N. A. Al-Hassawi. Determination of
of this work is the development of a method ensuring the linezolid in human plasma by LC–MS–MS. Analyst. 2001, 126, 609.
simultaneous determination of two analytes and the application [15] G. Carlucci. Analysis of fluoroquinolones in biological fluids by high
of a sample preparation with the DBS technique. Furthermore, performance liquid chromatography: Review. J. Chromatogr. A. 1998,
812, 343.
DBS disregards the need of blood samples, facilitating the shipping [16] V. Ferrone, M. Carlucci, P. Palumpo, G. Carlucci. Development and
and storage to prevent bacterial degradation. Additionally, high validation of an MEPS-UHPLC-PDA method for determination of
efficiency and reproducibility using a common available DAD ulifloxacin in human plasma and urine of patients with peripheral
detection indicate that the present method can be an important arterial disease. J. Pharm. Biomed. Anal. 2016, 128, 313.
[17] M. Simon, N. Thuo, N. Karisa, A. Mutori, G. O. Kokwaro, K. Maitland.
analytical tool during routine determination of ciprofloxacin and Determination of ciprofloxacin in human plasma using high
linezolid. performance liquid chromatography coupled with fluorescence
detection: Application to a population pharmacokinetics study in
children with severe malnutrition. J. Chromatogr. B 2011, 879, 146.
[18] J. Vella, F. Busuttil, N. Sammut Bartoloa, C. Sammut, A. Serracino-Inglott,
References L. M. Azzopardi, G. LaFerla. A simple HPLC-UV method for the
determination of ciprofloxacin in human plasma. J. Chrom. B. 2015,
[1] A. McLuckie. Respiratory Diseases and its Management. Springer Verlag, 989, 80.
2009. [19] V. Ferrone, M. Carlucci, P. Palumbo, G. Carlucci. Bioanalytical method
[2] American Thoracic Society. Infectious Diseases Society of America. development for quantification of ulifloxacin, fenbufen and felbinac
Guidelines for the management of adults with hospital-acquired, in rat plasma by solid-phase extraction (SPE) and HPLC with PDA
ventilator-associated, and healthcare-associated pneumonia. Am. J. detection. J. Pharm. Biomed. Anal. 2016, 123, 205.
Respir. Crit. Care Med. 2005, 171, 388. [20] W. Li, F. L. Tse. Dried blood spot sampling in combination with LC-MS/
[3] P. Gross, H. Neu, P. Aswapokee, C. Van Antwerpen, N. Aswapokee. MS for quantitative analysis of small molecules. Biomed. Chromatogr.
Deaths from nosocomial infections: Experience in a university 2010, 24, 49.
hospital and a community hospital. Am. J. Med. 1980, 68, 219. [21] R. Guthrie, A. Suzi. A simple phenylalanine method for detecting
[4] J. Chastre, J. Y. Fagon. Ventilator associated pneumonia. Am. J. Respir. phenylketonuria in large population of newborn infants. Pediatrics.
Crit. Care Med. 2002, 165, 867. 1963, 32, 338.
1618

[5] G. M. Eliopoulos, R. C. Moellering. in Antibiotics in Laboratory Medicine, [22] D. Blessborn, S. Romsing, Y. Bergqvist, N. Lindegardh. Assay for
3rd Edn. The Williams & Wilkins Co., Baltimore, 1991. screening for six antimalarial drugs and one metabolite using dried

wileyonlinelibrary.com/journal/dta Copyright © 2017 John Wiley & Sons, Ltd. Drug Test. Analysis 2017, 9, 1611–1619
Drug Testing
DBS-HPLC-PDA analysis of linezolid and ciprofloxacin and Analysis

blood spot sampling, sequential extraction and ion-trap detection. [29] A. Sharma, S. Jaiswal, M. Shukla, J. Lal. Dried blood spots: Concepts,
Bioanalysis. 2010, 2, 1839. present status, and future perspectives in bioanalysis. Drug Test. Anal.
[23] S. Aburuz, J. Millership, J. Mc Elnay. Dried blood spot liquid 2014, 6, 399.
chromatography assay for therapeutic drug monitoring of metformin. [30] N. G. L. Jager, H. Rosing. Procedures and practices for the validation of
J. Chromatogr. B. 2006, 832, 202. bioanalytical methods using dried blood spots: A review. Bioanalysis.
[24] E. Kralj, J. Tronteli, T. Paji, A. Kristl. Simultaneous measurement of 2014, 6, 2481.
imatinib, nilotinib and dasatinib in dried blood spot by ultra high [31] P. M. De Kesel, N. Sadones, S. Capiau, E. Lambert, C. P. Stove. Hemato-
performance liquid chromatography tandem mass spectrometry. critical issues in quantitative analysis of dried blood spot: Challenges
J. Chromatogr. B. 2012, 903, 150. and solutions. Bioanalysis. 2013, 5, 2023.
[25] G. Lawson, E. Cocks, S. Tanna. Bisoprolol, ramipril and simvastatin [32] FDA Food and Drug Administration of the United States. Guidance for
determination in dried blood spot using LC-HRMS for assessing Industry - Bioanalytical Method Validation. US Department of Health
medication adherence. J. Pharm. Biomed. Anal. 2013, 83, 1. and Human Services, Center for Drug Evaluation and Research and
[26] M. Hatami, E. Karimnia, K. Farhadi. Determination of salmeterol in dried Center for Veterinary Medicine, 2003.
blood spot using ionic liquid based dispersive liquid-liquid [33] V. Ferrone, M. Carlucci, R. Cotellere, P. Raimondi, A. Cichella, L. Di Marco,
microextration coupled with HPLC. J. Pharm. Biomed. Anal. 2013, 85, 283. G. Carlucci. Development and validation of a fast micro-extraction by
[27] T. Panchal, N. Sponeer, M. Barfield. Ensuring the collection of high packed sorbent UHPLC-PDA method for the simultaneous
quality dried blood spot samples across multisite clinical studies. determination of linezolid and ciprofloxacin from patients with
Bioanalysis. 2017, 9, 209. hospital-acquired pneumonia. Talanta. 2017, 164, 64.
[28] M. O’Mara, B. Hudson-Curtis, K. Olson, Y. Yueh, J. Dunn, N. Spooner. The [34] D. H. Vu, R. A. Koster, J. W. Alffenaar, J. R. Brouwers, D. R. Uges.
effect of hematocrit and punch location on assay bias during Determination of moxifloxacin in dried blood spots using LC-MS/
quantitative bioanalysis of dried blood spot samples. Bioanalysis. MS and the impact of hematocrit and blood volume.
2011, 3, 2335. J. Chromatogr. B 2011, 879, 1063.

1619

Drug Test. Analysis 2017, 9, 1611–1619 Copyright © 2017 John Wiley & Sons, Ltd. wileyonlinelibrary.com/journal/dta

You might also like