MR-96

MICROPLATE READER

Operation Manual
MR-96 Microplate Reader Operation Manual(2.0)

copyright
© 2005 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Service Manual, the issued Date is 2006-11 (Version: 2.0).

Intellectual Property Statement

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns
the intellectual property rights to this Mindray product and this manual. This manual may refer to information
protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the
rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights
of third parties.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the
information in this manual in any manner whatsoever without the written permission of Mindray is strictly
forbidden.

Release, amendment, reproduction, distribution, rent, adaption and translation of this manual in any
manner whatsoever without the written permission of Mindray is strictly forbidden.

, are the registered trademarks or trademarks owned by Mindray in China and other countries.

All other trademarks that appear in this manual are used only for editorial purposes without the intention of
improperly using them. They are the property of their respective owners.

Responsibility on the Manufacturer Party

Contents of this manual are subject to changes without prior notice.

All information contained in this manual is believed to be correct. Mindray shall not be liable for errors
contained herein nor for incidental or consequential damages in connection with the furnishing,
performance, or use of this manual.

Mindray is responsible for safety, reliability and performance of this product only in the condition that:

• all installation operations, expansions, changes, modifications and repairs of this product are conducted by
Mindray authorized personnel;

-i-
MR-96 Microplate Reader Operation Manual(2.0)

• the electrical installation of the relevant room complies with the applicable national and local requirements;
• the product is used in accordance with the instructions for use.

Upon request, Mindray may provide, with compensation, necessary circuit diagrams, calibration illustration list
and other information to help qualified technician to maintain and repair some parts, which Mindray may define
as user serviceable.

Note
This equipment is to be operated only by medical professionals trained and authorized by Mindray or
Mindray-authorized distributors.

Warning
It is important for the hospital or organization that employs this equipment to carry out a reasonable
service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.

Warranty

THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR


IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY
PARTICULAR PURPOSE.

Exemptions

Mindray's obligation or liability under this warranty does not include any transportation or other charges
or liability for direct, indirect or consequential damages or delay resulting from the improper use or
application of the product or the use of parts or accessories not approved by Mindray or repairs by
people other than Mindray authorized personnel.

This warranty shall not extend to:


any Mindray product which has been subjected to misuse, negligence or accident;
any Mindray product from which Mindray's original serial number tag or product identification
markings have been altered or removed;
any product of any other manufacturer.

- ii -
MR-96 Microplate Reader Operation Manual(2.0)

Return Policy

Return Procedure

In the event that it becomes necessary to return this product or part of this product to Mindray, the
following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a Customer Service
Authorization (Mindray) number. The Mindray number must appear on the outside of the shipping
container. Returned shipments will not be accepted if the Mindray number is not clearly visible. Please
provide the model number, serial number, and a brief description of the reason for return.

2. Freight policy: The customer is responsible for freight charges when this product is shipped to Mindray for
service (this includes customs charges).

3. Return address: Please send the part(s) or equipment to the address offered by Customer Service department

Company Contact

Manufacture: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,

Nanshan,Shenzhen518057,P.R.China,

Phone: +86 755 26582479 26582888

Fax: +86 755 26582934 26582500

Safety Precautions

1. Meaning of Signal Words


In this manual, the signal words BIOHAZARD, WARNING, CAUTION and NOTE are used
regarding safety and other important instructions. The signal words and their meanings are defined as follows.
Please understand their meanings clearly before reading this manual.

- iii -
MR-96 Microplate Reader Operation Manual(2.0)

Signal word Meaning

Read the statement following the symbol. The statement is alerting


BIOHAZARD
you to a potentially biohazardous condition.

WARNING Read the statement following the symbol. The statement is alerting

you to an operating hazard that can cause personal injury.

CAUTION Read the statement following the symbol. The statement is alerting you to a
possibility of system damage or unreliable results.
Read the statement following the symbol. The statement is alerting you to
NOTE
information that requires your attention.

Labels Used on the System

The labels attached to some MR-96 panels use symbols with the text to clarify the meaning of the text. The
table below explains the symbols on the labels.

Serial Number

Manufacturer

Date of Manufacture

In Vitro Diagnostic equipment

Biohazard Warning: risk of potentially biohazardous


infection

Warning: risk of personal injury or equipment damage

~ Alternating current (AC)

Graphics

All graphics, including screens and printout, are for illustration purposes only and must not be used for any
other purpose.

- iv -
MR-96 Microplate Reader Operation Manual(2.0)

Safety Symbols

This chart explains the symbols used in this manual.

When you see … Then …


Read the statement following the symbol. The statement is
WARNING: alerting you to an operating hazard that can cause personal
injury.

Read the statement following the symbol. The statement is


BIOHAZARD: alerting you to a potentially biohazardous condition.

Read the statement following the symbol. The statement is


CAUTION: alerting you to a possibility of system damage or unreliable
results.

Read the statement following the symbol. The statement is


NOTE: alerting you to information that requires your attention.

Read the statement following the symbol. The statement is


ELECTRIC alerting you to risk of electric shock.
SHOCK:

-I-
MR-96 Microplate Reader Operation Manual(2.0)

Table of Contents

1. INSTALLATION .............................................................................................................................................................2

1.1. UNPACKING ..............................................................................................................................................................2


1.2. INSTALLATION ENVIRONMENT REQUIREMENTS .......................................................................................................2
1.2.1 Transportation and Storage................................................................................................................................2
1.2.2. Operation Environment......................................................................................................................................2
1.3. CONNECTION ...........................................................................................................................................................2

2. FUNCTION......................................................................................................................................................................3

2.1. INTRODUCTION .............................................................................................................................................................3


2.2. REAR VIEW (FIGURE 1)................................................................................................................................................3
2.3. FRONT VIEW (FIGURE 2)..............................................................................................................................................4

3. OPERATION ...................................................................................................................................................................4

3.1. POWER-ON ..............................................................................................................................................................4


3.2. MAIN MENU..............................................................................................................................................................5
3.3. MEASURE .................................................................................................................................................................5
3.4. PROGRAM ................................................................................................................................................................7
3.4.1. Edit Test..........................................................................................................................................................7
3.4.2. New Test.......................................................................................................................................................12
3.4.3. Delete Test ...................................................................................................................................................13
3.5. QUALITY CONTROL .....................................................................................................................................................13
3.6. SETUP .........................................................................................................................................................................15
3.6.1. Date and Time...................................................................................................................................................15
3.6.2. Vibrancy .............................................................................................................................................................15
3.6.3. Light Source ......................................................................................................................................................16
3.6.4. Printer ...........................................................................................................................................................16
3.6.5. Filter ..............................................................................................................................................................16
3.7. SERVICE .................................................................................................................................................................17
3.7.1. Data Transfer...............................................................................................................................................17
3.7.2. Result Review .............................................................................................................................................17
3.7.3. Standard Review.........................................................................................................................................17
3.8. RAPID MEASURE....................................................................................................................................................18

4. MAINTENANCE ..........................................................................................................................................................19

4.1. NOTICE...................................................................................................................................................................19
4.2. MAINTENANCE .......................................................................................................................................................19

5. TECHNICAL SPECIFICATIONS...............................................................................................................................19

6. TROUBLESHOOTING................................................................................................................................................20

7. WARNING .....................................................................................................................................................................21

8. PACKAGING ................................................................................................................................................................21

-1-
MR-96 Microplate Reader Operation Manual(2.0)

1. Installation

1.1. Unpacking

When you receive your MR-96, carefully unpack the package and check the device. If you see any

damages, file a claim immediately with your local distributor or our local agency.

Put away the packaging material for future use.

Check the MR-96 and accessories against the packing list. If you find anything missing or

unconformable, please contact our local distributor or local agency.

1.2. Installation Environment Requirements

1.2.1 Transportation and Storage

Ambient temperature (-20 - 55)℃

Relative humidity ≤93%

Atmospheric pressure (50 - 106) kPa

1.2.2. Operation Environment

The MR-96 should be free of dust, vibration, corrosive and flammable gas, interference by strong

current and magnetic field as well as direct sunlight. Moreover, enough space should be kept

behind the device for good ventilation.

Ambient temperature: 10 - 40℃

Relative humidity: ≤85%

Atmospheric pressure: 86 - 106kPa

Power requirements: AC 220V, 50Hz(AC 110V,60Hz)

1.3. Connection

Check if the voltage conforms to requirements of the device. Make sure the power jack is

grounded correctly; otherwise you should connect a single ground wire to the Grounding port on

the device.

Connect the printer to the Printer port on the device through a data cable.
- 2 -
MR-96 Microplate Reader Operation Manual(2.0)

Insert one end of the power cord into the Power Jack on the rear panel of the device and other end

to the power supply.

2. Function

2.1. Introduction

The MR-96 Microplate Reader is an 8-channel optical measuring system, which can be used to

measure the density of vertical light. It is a specialized analyzer to hold ELISAs and then take

readings and process data.

The MR-96 adopts the touch screen which is comfortable to operate.

With the advanced high speed microprocessor in the world and man-machine conversation

design on the software, the MR-96 makes the operation simpler and more convenient, result

calculation quicker and waiting time of operator much shorter.

2.2. Rear View (Figure 1)

Printer Port Serial RS232 Port Power Jack Power On/Off Grounding

Figure 1

-3-
MR-96 Microplate Reader Operation Manual(2.0)

2.3. Front View (Figure 2)

Touch Screen Sample Compartment

Figure 2

3. Operation

3.1. Power-on

When you place the Power On/Off to ON ( | ), the device will begin self-check and then enter the Main

Menu screen (Figure 3) if everything is normal. After powering on the device, you should wait 10m for it to

preheat, so that each component and the lamp within the device become steady.

- 4 -
MR-96 Microplate Reader Operation Manual(2.0)

Jan.18,2005
Main Menu
18:18:18

Measure Program

Quality Control Setup

Service Rapid Measure

Figure 3

3.2. Main Menu

The Main Menu displays the function keys and current date and time. You can enter the submenu

screen by touching corresponding function key.

3.3. Measure

Touch the Measure button, the Select Test screen as shown in Figure 4 is displayed. Touch

Pgdn to display the tests of other pages circularly (42 tests in total, 3 pages). Touch the Exit

button to return to the Main Menu.

Measure Select Test

01 HBsAg 02 HBsAb

03 HBcAg 04 HBcAb

05 06

07 08

09 10

11 12

Pgdn Exit

-5-
MR-96 Microplate Reader Operation Manual(2.0)

Figure 4 Select Test screen

After you select the tests, the Plate No. and Vibrancy screen (Figure 5) is displayed.

Plate No. 00

Vibrancy

Exit OK

Figure 5 Plate No. and Vibrancy screen

Touch Plate No. and enter the Plate No. in the Plate No. screen (Figure 6). As shown in figure 6, the

CLR button is to delete what you have entered. After entering, touch YES to confirm and return to

the Plate No. and Vibrancy screen. If you touch Exit , then the Plate No. is 00. Touch OK to begin

measuring and analyzing the ELISA plate. If you need to mix the sample by vibrating the plate before

measuring, touch Vibrancy and then the OK button.

Plate No. ____________

1 2

3 4

5 6

7 8

9 0

. -

YES CLR

Exit

Figure 6 Plate No. screen

- 6 -
MR-96 Microplate Reader Operation Manual(2.0)

3.4. Program

The Program screen includes Edit Test, New Test and Delete Test (Figure 7). Touch Exit to

return to the Main Menu.

Program

Edit Test

New Test

Delete Test

Exit

Figure 7 Program screen

3.4.1. Edit Test

Touch Edit Test to select the test that you want to edit (Figure 4), then enter the Edit Test

screen (Figure 8).

Note: After setting, touch OK to save your settings.

-7-
MR-96 Microplate Reader Operation Manual(2.0)

Edit Test HBsAg

Prim. Wav. 405nm

Sec. Wav. 630nm

Plate Layout

Calculation Mode

Exit OK

Figure 8 Edit Test screen

Touch the Prim. Wav. button, a menu will then appear to ask you to select the wavelength.

Please select as instructed by the package insert.

NOTE: Don’t close all the filters; otherwise the alarm message “Lamp L” will appear.

Touch the Sec. Wav. button, a menu will then appear to ask you to select the wavelength.

Please select as instructed by the package insert.

In the Plate Layout screen (Figure 9-1), the right column is for selecting the well position

you want to set, use the and buttons to select different

well columns. The box on top right corner displays the current strip being set. Touch

Exit to return to the screen as shown in figure 8.

- 8 -
MR-96 Microplate Reader Operation Manual(2.0)

Automatic A01 Blank 01

B01 Pos. Control 01

C01 Sample 01

D01 Sample 02

E01 Sample 03

F01 Sample 04

Exit G01 Sample 05

H01 Sample 07

Figure 9-1 Plate Layout screen

After selecting the well position, the screen as shown in figure 9-2 will be displayed. The left

column is for setting the filling type and filling No. of the selected well position (Sample: 1-96;

Standard: 1-10; Quality Control: 1-10; Neg./Pos. control: 1-12; Blank well: 1-12), use the

Pgdn button to view each No. of filling type. Touch the selected well position once more to

exit its filling, refer to A01 in figure 9-2.

Standard 01 A01 Blank 01

Quality Control 01 B01 Pos. Control 01

Neg. Control 01 C01 Sample 01

Pos. Control 01 D01 Sample 02

Blank 01 E01 Sample 03

Sample 01 F01 Sample 04

Pgdn G01 Sample 05

H01 Sample 07

-9-
MR-96 Microplate Reader Operation Manual(2.0)

Figure 9-2 Plate Layout Menu

You can set the sample position quickly by using the Automatic button as shown in

figure 9-1. Since this function sets No. only for the original sample wells or blank wells, you

should touch Delete All (Figure 10) to delete all settings of the well positions before

auto filling. Then enter the Blank, Standard and Quality Control, etc, touch Automatic

and select Hor. Strip or Ver. Strip and No. of wells that you want to repeat.

Hor. Strip

Ver. Strip

Delete All

Exit

Figure 10 Automatic screen

Touch the Calculation Mode button as shown in figure 8, then select a method in the right list

appeared.

1) Absorbance: Read the absorbance of the sample directly, no submenu is displayed.

2) Qual Calculation: You should set Qual Grouping, Cut-off, Neg. Range, Pos. Range, Validity

Confirm and so forth. Please make settings as instructed by the package insert.

a) Qual Grouping 1-4: You can divide the qualitative results into four groups (Negative-,

Positive+, Weak Positive-+ and Strong Positive ++) in accordance with the cut-off you

have set.

b) Cut-off 1-3: Enter values of a, b and c in the CutOff formula a*NC+b*PC+c.

Where,

NC - Absorbance of the negative control well


- 10 -
MR-96 Microplate Reader Operation Manual(2.0)

PC - Absorbance of the positive control well

c) Neg. Range: When the absorbance exceeds the required range, the device will calculate

automatically with the negative control limit you have set.

d) Pos. Range: When the absorbance exceeds the required range, the device will calculate

automatically with the positive control limit you have set.

e) Validity Confirm 1-3: Set the range of absorbance for blank wells, positive and negative

control wells. The device will not calculate results when the absorbance exceeds the

range.

3) Quant Calculation: Some parameters are added on the basis of qualitative settings, such as

STD Concentration, Unit, Reference, Calibration Rule and Factor. Please make settings as

instructed by the package insert.

a) STD Concentration: You can set the concentrations of up to 8 standards. When setting,

you should begin with 1 from low concentration to high concentration. If No. of standards

is less than 8, you must set the remaining concentrations to 0.

b) Unit: Enter the concentration unit of the selected test.

c) Reference H: Enter the upper limit for the selected test.

d) Reference L: Enter the lower limit for the selected test.

e) Calibration Rule (C - Concentration, A - Absorbance, k - Slope, b - Intercept)

Factor

Formula: A =kC +b.

k and b should be entered by the operator manually.

One Point Linear

A标
Formula: A =kC, K =
C标

This calibration rule requires only one standard to obtain the parameter k.

Two Point Linear

A2 − A1 A − A1
Formula: A =kC+b, K = , b = A1 − ( 2 )C1
C 2 − C1 C 2 − C1

This calibration rule requires two standards. C1 and C2 are concentrations of standard 1

and 2; A1 and A2 are their absorbance.


- 11 -
MR-96 Microplate Reader Operation Manual(2.0)

Multipoint Linear

Formula: A =kC+b.

You can get values of k and b through the method of least squares.

This calibration rule requires at least 3 standards.

Nonlinear

Draw a broken line passing through the standard concentration and its absorbance.

Measure the absorbance of the waiting samples and check relevant concentration on the

broken line.

This calibration rule requires at least 3 standards.

Multipoint Percent

The maximum absorbance of the standards is 100%.

This calibration rule requires at least 3 standards as the multi-point broken line.

Exponent

Formula: A = k ln C + b

This calibration rule requires at least 3 standards.

Logarithm

Formula: A = ke
bC

This calibration rule requires at least 3 standards.

Power function

Formula: A = kC
b

This calibration rule requires at least 3 standards.

f) Factor Setup: Enter the calibration parameters k and b which are mainly used in the factor

calculation method.

3.4.2. New Test

Touch the New Test button to select the test that you want to add or modify (refer to figure 4

for Select Test screen), and then enter the New Test screen (Figure 11).

- 12 -
MR-96 Microplate Reader Operation Manual(2.0)

New Test _____________

1ABCabc 2DEFdef

3GHIghi 4JKLjkl

5MNOmno 6PQRpqr

7STUstu 8VWXvwx

9Yzyz +-*/

0.,() NEXT

YES CLR

Exit

Figure 11 New Test screen

In the New Test screen, if you touch a button repeatedly, the screen will display the next

number, letter or symbol in circular way. NEXT is used to move the cursor

rightward to enter the next character. If you enter a wrong character, delete it with the

CLR button and enter again. Touch YES to save what you have entered and

Exit to return to the Program screen (Figure 7) without saving your entering.

3.4.3. Delete Test

Touch the Delete Test button, select the test that you want to delete (Figure 4).

3.5. Quality Control

Touch the Quality Control button, select the test that you want to view (Figure 4), then the QC

plot of this test will be displayed (Figure 12.). Touch the Settings button to set the QC parameters

of this test (V - Mean, SD - Standard Difference). If you want to print the QC data and QC plot, touch

the Print button.

- 13 -
MR-96 Microplate Reader Operation Manual(2.0)

hAFP Quality Control V: 0.000 SD:0.000


+3SD
+2SD

-2SD
-3SD

Settings Delete Data

Print Exit

Figure 12 QC Plot screen

After touching the Delete Data button, you need to select Del. Last Month or Del. All Data, as

shown in figure 13. Touch Exit to return to the screen as shown in figure 12.

Del. Last Month

Del. All Data

Exit

Figure 13 Delete Data screen


- 14 -
MR-96 Microplate Reader Operation Manual(2.0)

3.6. Setup

In the Setup screen (Figure 14), you can set the date and time, vibrant speed, light source, printer and

filter.

Setup

Date and Time Vibrancy

Light Source Printer

Filter

Exit

Figure 14 Setup screen

3.6.1. Date and Time

Touch the Date and Time button to set Year, Month, Day, Hour and Minute one by one. Touch YES

to confirm your settings and proceed to the next option until you finish all settings.

3.6.2. Vibrancy

The Vibrancy screen (Figure 15) includes Vibrant Speed and Vibrant Time. There are three speeds

available as High, Normal and Low. The vibrant time should not be longer than 60s.

- 15 -
MR-96 Microplate Reader Operation Manual(2.0)

Vibrant Speed Normal

Vibrant Time 30s

Exit

Figure 15 Vibrancy screen

3.6.3. Light Source

This button is used to turn on/off the lamp temporarily for the sake of the lamp’s lifespan.

NOTE: The lamp should be turned on and preheats for 10m before testing; otherwise the

alarm message “Lamp L” will appear.

3.6.4. Printer

This button is used to power on/off the printer, so you can select whether to print the results immediately

after testing.

3.6.5. Filter

The MR-96 reads the air blank absorbance of each filter in the 4 channels (F5 is spare), which is used to

check the photometric system. The default absorbance of the 405 - 630 filters is less than 0.500. However,

it may increase gradually in the result of dust, humidification and moldiness when the device is used for a

long time. If the absorbance is greater than 1.000, you should clean the photometric system or replace the

accessories like the filters.

- 16 -
MR-96 Microplate Reader Operation Manual(2.0)

3.7. Service

There are three options in the Service menu:

Data Transfer

Result Review

Standard Review

3.7.1. Data Transfer

Touch the Data Transfer button to transport the test data to the computer which is connected to the

MR-96 through the serial RS232 port.

3.7.2. Result Review

Touch the Result Review button and enter the No. of Plate that you want to search, then touch

YES to display the results. Touch the button to display circularly the results of column 0-6

and 7-12.

3.7.3. Standard Review

After touching the Standard Review button and selecting the test that you want to search, the screen

will display the standard absorbance and relevant concentration, as well as the calibration curve, as shown

in figure 16. However, if the selected test doesn’t use the Quant Calculation Mode, the results will not be

displayed).

- 17 -
MR-96 Microplate Reader Operation Manual(2.0)

Standard Review hAFP

Concentration Absorbance ABS

1 1.0 1.000

2 2.0 2.000

3 3.0 3.000

4 4.0 4.000

5 5.0 5.000

6 6.0 6.000

7 7.0 7.000

8 8.0 8.000

k 1.000

b 0.000

Multipoint Linear STD

Exit

Figure 16 Standard Review screen

3.8. Rapid Measure

This function is used to read the absorbance of the waiting samples quickly. After touching the

Rapid Measure button, select the primary and secondary wavelength and enter the No. of samples.

Touch OK to start testing (Figure 17).

Prim. Wav. 450nm

Sec. Wav.

Vibrancy

Exit OK

Figure 17 Rapid Measure screen

- 18 -
MR-96 Microplate Reader Operation Manual(2.0)

4. Maintenance

4.1. Notice

To ensure reliable performance and excellent operation of the MR-96, the operator must operate and

maintain strictly as instructed by this Operation Manual.

4.2. Maintenance

The MR-96 Microplate Reader is a precision optical measuring system. Therefore, it’s necessary to

perform the daily maintenance, which includes:

1) Keeping the operating environment dry and clean. Covering the sample compartment after using

to prevent from dust.

2) Keeping the mechanical and electrical parts of the device away from water and reagents.

3) Switching off the power supply while cleaning the surface and screen of the device. Be sure to use

soft cloth with noncorrosive detergent.

5. Technical Specifications

1) Wavelength: Accuracy (±2nm), Half width (≤12nm), Peak transmittance (≥35%).

2) Wavelength range: 405 nm, 450 nm, 492 nm, 630nm. 4 free positions for optional filters

3) Reading range: 0 - 3.5A, Measurement range: 0 - 2.5A

4) Absorbance resolution: 0.001A (Internally: 0.0001A)

5) Absorbance accuracy: ±0.03A (0 - 2.5A)

6) Repeatability (CV, %) ≤1%

7) Linearity: Error requirements

Wavelength Range (nm) Absorbance (A) Linearity Error %

405 - 630 0.0 - 2.5 ≤ ±2.5%

8) Stability: ≤ 0.005A/10min

9) Reading Speed: Single wavelength <20s/plate (96-well plate); dual wavelength <35s/plate

(96-well plate).

- 19 -
MR-96 Microplate Reader Operation Manual(2.0)

10) Vibrancy: Vibrant Speed (High, Normal, Low); Vibrant Time (0 - 60s).

11) Power: 100VA

6. Troubleshooting

Symptoms Causes Actions to Take

The power cord is not connected Re-connect the power cord

Can’t start up correctly properly

The fuse is broken Replace the fuse

The LCD display or inverter goes Contact our local distributor or


No display, no backlight
wrong local agency

The optical coupler is with dust Remove the dust and replace the

The sample compartment works or goes wrong optical coupler

abnormally The sample compartment is Reposition the ELISA plate

blocked

Sound is abnormal during Foreign matter exists Remove the foreign matter

operation Mechanical movement is Contact our local distributor or

abnormal local agency

The printer is not connected Check the connection between


Can’t print
properly the printer and the device

The lamp is aging or goes wrong Replace the lamp

The filter is aging or goes wrong Replace the filter


Readings are abnormal
The electric circuit goes wrong Contact our local distributor or

local agency

- 20 -
MR-96 Microplate Reader Operation Manual(2.0)

7. Warning

The MR-96 is a mini in vitro diagnosis instrument. It may bring risks to the patients indirectly, and

accompanying risks to the operator and environment. To avoid these risks, be sure to read the

following instructions carefully.

This device adopts AC 220V which may bring electrical harm to human body. Don’t open

the rear cover before switching off the power supply. If you need to open the rear cover to maintain

the device, you’d better contact our service personnel or professionals.

For the patient samples may contain germina or virus which will cause disease, be

sure to dispose of the waste in accordance with relevant regulations of your hospital.

NOTE: Don’t operate the device unless you have read this Operation Manual carefully.

8. Packaging

The package of this product includes the following information:

Product name, model, weight, package volume, manufacturing date, quantity.

Company name, address, post code.

Information indicated with symbols: FRAGILE, THIS WAY UP, KEEP AWAY FROM RAIN, DO

NOT ROLL.

- 21 -