You are on page 1of 9

A Practical Approach to

Validation of HPLC Methods


Under Current Good
Manufacturing Practices
GHULAM A. SHABIR


Introduction Regulatory analytical procedures are of two types: com-
Analytical methods validation is an important regulatory pendial and noncompendial. The noncompendial analytical
requirement in pharmaceutical analysis. High-Performance procedures in the USP are those legally recognized as regu-
Liquid Chromatography (HPLC) is commonly used as an latory procedures under section 501(b) of the Federal Food,
analytical technique in developing and validating assay Drug and Cosmetic Act. When using USP analytical meth-
methods for drug products and drug substances. Method val- ods, the guidance recommends that information be provided
idation provides documented evidence, and a high degree of for the following characteristics: specificity of the method,
assurance, that an analytical method employed for a specific stability of the analytical sample solution, and intermediate
test, is suitable for its intended use. Over recent years, regu- precision. Compendial analytical methods may not be stabil-
latory authorities have become increasingly aware of the ne- ity indicating, and this concern must be addressed when de-
cessity of ensuring that the data submitted to them in appli- veloping a drug product specification, because formulation
cations for marketing authorizations have been acquired based interference may not be considered in the monograph
using validated analytical methodology. The International specifications. Additional analytical tests for impurities may
Conference on Harmonization (ICH) has introduced guide- be necessary to support the quality of the drug substance or
1,2
lines for analytical methods validation. The U.S. Food and drug product. Noncompendial analytical methods must be
Drug Administration (FDA) methods validation draft guid- fully validated. The most widely applied validation charac-
3-5
ance document, as well as United States Pharmacopoeia teristics are accuracy, precision (repeatability and intermedi-
(USP) both refer to ICH guidelines.
6
ate precision), specificity, detection limit,
These draft guidances define regulatory quantitation limit, linearity, range, and sta-
and alternative analytical procedures and Figure 1 bility of analytical solutions.
stability-indicating assays. The FDA has
____________________ The parameters that require validation
The chemical structure of
proposed adding section CFR 211.222 on and the approach adopted for each partic-
ethyl 4-hydroxybenzoate.
analytical methods validation to the cur- ular case are dependent on the type and
rent Good Manufacturing Practice applications of the method. Before under-
7
(cGMP) regulations. This would require taking the task of method validation, it is
pharmaceutical manufacturers to estab- necessary that the analytical system itself
lish and document the accuracy, sensitiv- is adequately designed, maintained, cali-
8
ity, specificity, reproducibility, and any brated, and validated. The first step in
other attribute (e.g., system suitability, method validation is to prepare a proto-
stability of solutions) necessary to vali- col, preferably written with the instruc-
date test methods. tions in a clear step-by-step format. This

Equipment and Instrumentation Qualification 29


Ghulam A. Shabir

8
approach has been reported previously. In
this paper, it is intended to review and Figure 2
______________________________________________
demonstrate practical approaches to ana- Graph measured peak area versus ethyl 4-hydroxyben-
lytical method validation in detail with zoate concentration demonstrating linearity.
reference to an HPLC assay of ethyl 4-hy-
droxybenzoate (Figure 1). Ethyl 4-hy-
droxybenzoate alone or in combination
with other esters of p-hydroxybenzoic
acid, or with other antimicrobial agents, is
used as a preservative in pharmaceutical
formulations.

Experimental
✔ Chemicals and reagents.
All chemicals and reagents were of the
highest purity. HPLC-grade acetonitrile was
obtained from Merck (Darmstadt, Ger-
many). Water was purified with a Millipore
Milli-Q system (Watford, UK). Ethyl 4-hy-
droxybenzoate (Batch #1005425) was sup-
plied by Lancaster Synthesis (Morecambe, England). ✔ Preparation of mobile phase.
The mobile phase was prepared by adding 650 ml of
✔ HPLC instrumentation. HPLC-grade acetonitrile in 1000 ml of water (65:35, v/v).
The HPLC system used for the validation studies con- The mobile phase was filtered under a vacuum through 0.45
sisted of a Waters Alliance 2690 Separations Module to a mm nylon filters and degassed before use. Also, the mobile
996 photodiode-array (PDA) detector. The control of the phase continuously was degassed with an on-line degasser.
HPLC system and data collection was by a Compaq com-
puter equipped with Waters® Millennium32 software (ver- ✔ Preparation of standard and sample solutions.
sion 3.20). The second HPLC system used for intermediate Ethyl 4-hydroxybenzoate (100 mg) was weighed accu-
precision studies consisted of Perkin Elmer: model series rately and added to a 100 ml volumetric flask before being
200 UV visible detector, series 200 LC pump, series 200 au- dissolved in acetonitrile. A 2.0 ml aliquot of stock solution
tosampler, and a series 200 peltier LC column oven were was diluted to 100 ml in the mobile phase, yielding a final
used to chromatograph the solutions. The data was acquired concentration of 20 µg/ml. Standard solutions for the eval-
via PE TotalChrom Workstation data acquisition software, uation of ethyl 4-hydroxybenzoate linearity were prepared
(Version 6.2.0) using PE Nelson series 600 LINK interfaces. over a concentration range of 5.0-40 µg/ml, to 25, 50, 75,
Both HPLC systems including software (Food and Drug Ad- 100, 150, and 200% in the mobile phase.
ministration (FDA), 21 Code of Federal Regulations (CFR)
Part 11) were validated prior to use for the test method vali- Results and Discussion
dation.
All chromatographic experiments were performed in the ✔ Validation of the chromatographic method:
isocratic mode. A C18 symmetry analytical column from Linearity and range
Waters (located in Milford, MA, United States) 3.9 x 150 The linearity of the method should be tested in order to
mm, 5 mm particle size was used. The mobile phase con- demonstrate a proportional relationship of response versus
sisted of a mixture of acetonitrile, water solution (65:35, analyte concentration over the working range. The linearity
v/v). The flow rate was set to 1.0 ml/min, and the oven tem- range for evaluation depends on the purpose of the analytical
perature to 25˚C. The injection volume was 20 µl, and the test method. The ICH guidelines specified a minimum of
detection wavelength was set at 254nm. five concentration levels, along with certain minimum spec-

30 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir

Figure 3 n = 3) and the following regression equation


______________________________________________
was found by plotting the peak area (y) ver-
Results of assessment of the linearity of the HPLC method
sus the ethyl 4-hydroxybenzoate concentra-
for the assay of ethyl 4-hydroxybenzoate employing the an-
tion (x) expressed in µg/ml: y = 29935x +
alytical working standard dissolved in mobile phase
51338 (r2 = 1.000). The demonstration coef-
ficient (r2) obtained for the regression line
Concentra- Concentration EP peak area Peak area demonstrates the excellent relationship be-
tion (µg/ml) as percent of as mean of RSD (%) tween peak area and concentration of ethyl 4-
20 µg/ml 3 injections hydroxybenzoate (Figure 2). The data ob-
tained from linearity experiments are pre-
5 25 792862 0.13
sented in Figure 3. The range is derived from
10 50 1535889 0.16
linearity studies, and depends on the intended
15 75 2308902 0.21
20 application of the test method. It is estab-
100 3057149 0.13
30 150 4546415 0.10 lished by confirming that the assay procedure
40 200 6027790 0.25 provides an acceptable degree of linearity,
accuracy, and precision when applied to sam-
Correlation coefficient: r2 = 1.000; Equation for ples containing amounts of analyte within, or
regression line: y = 29935x + 51338 (n = 3) at the extremes of the specified range, of the
test method. The range is normally expressed
in the same units as the test results obtained
by the method. In this study, the data obtained
Figure 4
______________________________________________ during the linearity and accuracy studies was
Accuracy/recovery of ethyl 4-hydroxybenzoate from used to assess the range of the assay method.
samples with known concentration The precision data for this assessment was the
precision of the three replicate samples ana-
Sample Percent of Recovery (%) RSD (%) of lyzed at each level in the accuracy studies.
number nominal (n = 3) area response
The valid analytical range of the method is
factor
that range of concentrations, which pass the
1 50 99.67 0.16 linearity and accuracy criteria, and yields an
2 75 99.78 0.21 RSD of < 2%. The linearity data described
3 100 99.85 0.13 earlier demonstrates acceptable linearity for
4 150 99.87 0.10 ethyl 4-hydroxybenzoate over the range of 80
Mean recovery: 99.8%; RSD 0.09% to 120% of the target concentration. The RSD
values obtained for the recovery of ethyl 4-hy-
ified ranges. For assay, the minimum specified range is from droxybenzoate at 50, 75, 100, and 150% of target are 0.16,
80-120% of the target concentration. For an impurity test, 0.21, 0.13, and 0.10%, respectively. Each value was the result
the minimum range is from the reporting level of each im- of three individual sample preparations and analysis. These
purity to 120% of the specification. Acceptability of linear- data support a method range of 80 to 120% of the target con-
ity data is often judged by examining the correlation coeffi- centration.
cient and y-intercept of the linear regression line for the re-
sponse versus concentration plot. The regression coefficient Accuracy/recovery studies
(r ) is > 0.998 is generally considered as evidence of accept-
2
The accuracy of an analytical method is the closeness of
6
able fit of the data to the regression line. The y-intercept test results obtained by that method to the true value. Accu-
should be less than a few percent of the response obtained for racy is usually determined in one of four ways. First, accu-
the analyte at the target level. The Percent Relative Standard racy can be assessed by analyzing a sample of known con-
Deviation (RSD), intercept, and slope should be calculated. centration (reference materials), and comparing the mea-
In the present study, linearity was studied in the concentra- sured value to the true value. The second approach is to com-
tion range 5.0-40 µg/ml (25-200% of nominal concentration, pare test results from the new method with results from an

Equipment and Instrumentation Qualification 31


Ghulam A. Shabir

concentration. The ICH2 recommends collecting data from a


Method validation minimum of nine determinations over a minimum of three
provides documented concentration levels covering the specified range (e.g., three
concentrations, three replicates each).
evidence, and a high In the present study, a number of different solutions
degree of assurance, that were prepared with known added amounts of ethyl 4-hy-
droxybenzoate and injected in triplicate. Percent recover-
an analytical method ies of response factor (area/concentration) were calcu-
lated. The results of accuracy studies are shown in Figure
employed for a specific 4, and it is evident that the method is accurate within the
test, is suitable for its desired recovery range.
intended use.
Specificity of the Assay and Degradation
of Active Constituent
existing alternate well-characterized procedure that is known In order to design a chromatographic system for the
to be accurate. The third approach is based on the recovery analysis of an active component of a pharmaceutical prod-
of known amounts of analyte. This is performed by spiking uct, it is essential to have a good knowledge of; (1) sus-
analyte in blank matrices. For assay methods, spiked sam- ceptibility of the drug to degradation and its degradation
ples are prepared in triplicate at three levels over a range of pathway; (2) assay interference by possible degradants or
50-150% of the target concentration. The percent recovery synthesis precursors; and (3) assay interference by chemi-
should then be calculated. The fourth approach is the tech- cals employed in sample preparation and excipients in the
nique of standard additions, which can also be used to deter- formulation.
mine recovery of spiked analyte. This approach is used if it Degradation products may be formed by acid/base hy-
is not possible to prepare a blank sample matrix without the drolysis, oxidation, Ultraviolet (UV) irradiation, heat, light,
presence of the analyte. Accuracy criteria for an assay etc., however, it is not within the scope of this paper to dis-
method (FDA) is that the mean recovery will be 100 ± 2% at cuss in detail the elucidation of degradation pathways.
each concentration over the range of 80-120% of the target In the present study, initially, a reference standard of

Figure 5
________________________________________________________________
HPLC chromatogram of ethyl 4-hydroxybenzoate.

32 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir

Figure 6
________________________________________________________________
HPLC chromatogram for placebo. The analyte peak was eluted at 1.58 minutes.

ethyl 4-hydroxybenzoate was chro-


Figure 7
matographed. Figure 5 clearly demonstrates ______________________________________________
that ethyl 4-hydroxybenzoate is well sepa- Demonstration of the repeatability of the HPLC assay for
rated from any potential interference. Assay ethyl 4-hydroxybenzoate as shown by the results of 10
interference was investigated by injecting replicate injections of one solution at 100 percent of the
placebo. No interfering peaks (Figure 6) test (20 mg/ml) concentration
were observed. Therefore, this method was
Injection RT (min) Peak height Peak area
specific for ethyl 4-hydroxybenzoate. number (µV) (µV s)
1 1.57 855847 3109735
Precision 2 1.58 858249 3100787
3 1.58 854532 3099540
Precision is the measure of the degree
4 1.58 856705 3103544
of repeatability of an analytical method
5 1.57 857058 3101464
under normal operation, and is normally
6 1.57 854755 3099731
expressed as the percent relative standard 7 1.57 855098 3102575
deviation for a statistically significant 8 1.58 854078 3103159
number of samples. Precision may be per- 9 1.58 856416 3104217
formed at three different levels: repeatabil- 10 1.58 849916 3091891
ity, intermediate precision, and repro- Mean 1.58 855265 3101665
ducibility. RSD (%) 0.18 0.25 0.14

Repeatability three repetitions each), or from a minimum of six determi-


nations at 100% of the test or target concentration. A preci-
Repeatability (intra-day assay precision) is the results of sion criterion for an assay method is that the instrument pre-
the method operating over a short time interval under the cision (RSD) will be ≤1%, and for the impurity assay, at the
same conditions (intra-assay precision). It should be deter- limit of quantitation, the instrument precision (repeatability)
mined from a minimum of nine determinations covering the will be ≤ 5%. Documentation in support of precision studies
specified range of the procedure (for example, three levels, should include the standard deviation, relative standard devi-

Equipment and Instrumentation Qualification 33


Ghulam A. Shabir

Figure 8
_____________________________________________________________
Demonstration of the intermediate precision of the HPLC assay for ethyl
4-hydroxybenzoate results in relative percent purity area

HPLC system 1 HPLC system 2

Sample S1 S2 S3 S1 S2 S3
(50%) (100%) (150%) (50%) (100%) (150%)
Operator 1, day 1 99.83 99.79 99.76 99.76 99.83 99.83
Operator 1, day 2 99.76 99.74 99.74 99.82 99.80 99.78
Operator 2, day 1 99.71 99.76 99.77 99.75 99.76 99.69
Operator 2, day 2 99.53 99.62 99.57 99.79 99.81 99.82
Mean 99.71 99.73 99.71 99.78 99.80 99.78
(HPLC systems)
Mean (Operators) 99.79 99.79 99.78 99.70 99.74 99.71
RSD (criteria ≤ 2%) S1+S1 S2+S2 S3+S3
HPLC systems 0.05 0.05 0.05
Operators 0.06 0.04 0.05

ation, coefficient of variation, and confidence interval. In this cent purity mean values at 50, 100, and 150% of the test con-
study, precision of the method was evaluated through the re- centration. The RSD values presented in Figure 8 were less
peatability of the method (intra-assay precision) by assaying than 1% for both systems and operators, and illustrated the
ten replicate injections of ethyl 4-hydroxybenzoate at the good precision of the analytical method.
same concentration (20 µg/ml), during the same day, under
the same experimental conditions. The RSD values of the re- Reproducibility
tention time, area, and height of ethyl 4-hydroxybenzoate
peak were found to be < 0.3%, as presented in Figure 7. Reproducibility1 is determined by testing homogeneous
samples in multiple laboratories, often as part of inter-labora-
Intermediate Precision tory crossover studies. An example of reproducibility criteria
for an assay method could be that the assay results obtained
Intermediate precision (inter-day variation) is the results in multiple laboratories will be statistically equivalent, or the
from within lab variations, due to random events, such as dif- mean results will be within 2% of the value obtained by the
ferent days, analysts, equipment, etc. In determining interme- primary testing lab. For an impurity method, results obtained
diate precision, experimental design should be employed, so in multiple laboratories will be statistically equivalent, or the
that the effects (if any) of the individual variables can be mon- mean results will be within 10% (relative) of the value ob-
itored. Precision criteria for an assay method is that the intra- tained by the primary testing lab for impurities. Repro-
assay precision will be ≤ 2%, and for impurity assay, at the ducibility is not normally expected if intermediate precision
limit of quantitation, the instrument precision will be ≤ 5%, is performed.
and the intra-assay precision will be ≤10%. In this study, in-
termediate precision (within-laboratory variation) was Limit of Detection and Quantitation
demonstrated by two operators, using two HPLC systems,
and evaluating the relative percent purity data across the two The Limit of Detection (LOD) and Limit of Quantitation
HPLC systems at three concentration levels (50%, 100%, (LOQ) tests for the procedure are performed on samples
150%) that cover the ethyl 4-hydroxybenzoate assay method containing very low concentrations of analyte. LOD is de-
range (5.0-40 µg/ml). The mean and RSD across the systems fined as the lowest amount of analyte that can be detected
and analysts were calculated from the individual relative per- above baseline noise; typically, three times the noise level.

34 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir

Figure 9
________________________________________________________________
HPLC chromatogram for limit of detection (2 ng/ml)

Figure 10
________________________________________________________________
HPLC chromatogram for limit of quantitation (5 ng/ml).

Figure 11
________________________________________________________________________
Stability of ethyl 4-hydroxybenzoate in solution (n = 6)

Time RT (min) Peak area Peak Height Percent Percent of initial


(hour) RSD (%) RSD (%) recovery
0 0.14 0.70 0.80 99.88
24 0.18 0.15 0.27 99.82
48 0.29 0.30 0.51 99.78 99.35

Equipment and Instrumentation Qualification 35


Ghulam A. Shabir

Figure 12
_______________________________________________________
Demonstration of the system suitability of the HPLC assay for ethyl
4-hydroxybenzoate

Results
System Suitability Acceptance
Parameter Criteria
HPLC system 1 HPLC system 2

Injection precision for RSD ≤ 1% 0.15 0.11


area (n = 10)

Injection precision for


retention time (min) RSD ≤ 1% 0.18 0.11

USP tailing (T) for ethyl T≤2 1.05 1.03


4-hydroxybenzoate peak

Theoretical plates (N) for


ethyl 4-hydroxybenzoate N = > 2000 5276 6628
peak

LOQ is defined as the lowest amount of analyte which can mal lighting conditions for 48 hours, and were shown to be
be reproducibly quantitated above the baseline noise, that stable with no significant change in ethyl 4-hydroxybenzoate
gives S/N = 10. concentrations over this period (Figure 11). This is indicated
In this study, LOD for a 20 µl injection of ethyl 4-hy- by <1% changes in area between T = 0 hours and T = 48
droxybenzoate standard (signal to noise = 3) was 2.0 ηg/ml hours. Based on these data that show quantitative recovery
(Figure 9), and the LOQ (signal to noise = 10) was 5 ηg/ml through 48 hours, ethyl 4-hydroxybenzoate solutions can be
(Figure 10) and RSD < 2% (n = 6). assayed within 48 hours of preparation.

Stability of Analytical Solutions System Suitability


Samples and standards should be tested over at least a 48 System suitability tests are an integral part of HPLC
hour period (depends on intended use), and quantitation of methods, and are used to verify that the accuracy and preci-
components should be determined by comparison to freshly sion of the system are adequate for the analysis to be per-
prepared standards. A stability criterion for assay methods is formed.
that sample and standard solutions and the mobile phase will Parameters, such as plate count, tailing factor, resolution,
be stable for 48 hours under defined storage conditions. Sta- and repeatability (RSD of retention time and area for six rep-
bility is considered to be acceptable when the change in the etitions) are determined and compared against the specifica-
standard or sample response is within 2% relative to freshly tions set for the method. The parameter to be measured and
prepared standards. In this study, the stability of ethyl 4-hy- their recommended limits4,9 obtained from the analysis of the
droxybenzoate solutions was investigated. Therefore, test so- system suitability sample are shown in Figure 8. In the pre-
lutions of ethyl 4-hydroxybenzoate were prepared using the sent study, the system suitability test was performed on both
conditions cited in Section 2.4. They were chromatographed HPLC systems to determine the accuracy and precision of
at the beginning, and after 24 and 48 hours. The stability of the system, by injecting ten injections of a solution contain-
ethyl 4-hydroxybenzoate and the mobile phase were calcu- ing 20 µg/ml of ethyl 4-hydroxybenzoate. RSD for peak area
lated by comparing area response and area percent of two and retention time < 1%, tailing factor (T) < 2 and theoreti-
standards at 20 µg/ml over time. Standard solutions stored in cal plate (N) were > 5000 for both HPLC systems, as can be
a capped volumetric flask on a laboratory bench under nor- seen in Figure 12.

36 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir

Conclusion
Article Acronym Listing
It is clear from the various guidelines issued by regula-
tory authorities that analytical methodology should be thor- CFR: Code of Federal Regulations
oughly validated under Current Good Manufacturing Prac- cGMP: current Good Manufacturing Practice
tice (cGMP). HPLC assay of active ingredients in pharma- FDA: Food and Drug Administration
ceutical products, and subsequent method validation, can be HPLC: High Performance Liquid Chromatography
complex and time-consuming. However, a well-defined pro- ICH: International Conference on Harmonization
tocol and documented validation plan simplifies and short- LOD: Limit of Detection
ens the process, while also providing regulatory agencies LOQ: Limit of Quantitation
with evidence that the analytical system and method is suit- PDA: Photodiode Array
able for its intended use. This paper is intended to provide RSD: Relative Standard Deviation
guidance on how to perform method validation for HPLC USP: United States Pharmacopoeia
that generates both useful and meaningful data that meets all UV: Ultra Violet
FDA, USP, and ICH validation requirements for pharmaceu-
tical analysis. ❏

Acknowledgements References
I thank Abbott Laboratories and MediSense for permis- 1. Text on Validation of Analytical Procedures. ICH, Q2A, FDA,
sion to publish this article. I would also thank to Dr Nigel Federal Register, Vol. 60, (March), 1995. p. 11260.
Forrow for his comments on the text. 2. Validation of Analytical Procedures: Methodology. ICH, Q2b.
FDA, Federal Register, Vol. 62, (May), 1997. p. 27463.
3. Analytical Procedures and Methods Validation: Chemistry,
About the Author Manufacturing and Controls Documentation, FDA, Federal
Register (Notices) 65 (169), August, 2000, p. 52776.
4. Validation of Chromatographic Methods, Reviewer Guid-
Ghulam Shabir, BSc (Hons), MSc, FIQA, CIM, is a
ance, Centre for Drug Evaluation and Research, FDA, 1994.
Principal Scientist in Analytical R&D at Abbott Lab-
5. Guideline for Submitting Samples and Analytical Data for
oratories, MediSense Products UK, where he is re-
Methods Validation. FDA, February 1987.
sponsible for analytical methods development and 6. Validation of Compendial Methods. USP 25-NF 20, (1225),
validation and instrument qualification. He has over United States Pharmacopeal Convention, Rockville, MD,
17 years of experience in the pharmaceutical indus- 2002, p. 2256.
tries, and has received the Technical Excellence 7. Current Good Manufacturing Practice for Finished Pharma-
and the Spot Award from Abbott Laboratories. He ceuticals, Proposed Rules, FDA, HHS, 21 CFR Part 211,
has also won the Education Award from the Insti- Federal Register, Vol. 61, 87 (May), 1996. p. 20106.
tute of Manufacturing UK. Mr. Shabir has written for 8. Shabir, G.A., “Validation of HPLC Methods for Pharmaceuti-
over 26 industry publications, and his work has cal Analysis: Understanding the Differences and Similarities
been presented at the world’s premier international between Validation Requirements of the U.S. FDA, USP and
conferences. He is a Fellow of the Institute of Qual- ICH.” Journal of Chromatography A. 987 (2003). pp. 57-66.
ity Assurance, Companion of the Institute of Manu- 9. Chromatography. USP 23 (621), United States Pharmaco-
facturing, and Committee member of the Society of peal Convention, Rockville, MD, 1994, p. 1776.
Chemical Industry Separation Science & Technol-
ogy UK.

Mr. Shabir can be reached by phone at +44(0) 1993


863099 or by email at ghulam.shabir@abbott.com

Originally published in the May, 2004 issue of the Journal of Validation Technology

Equipment and Instrumentation Qualification 37

You might also like