Updated Service Organization Procedures
Service Training Schedule 2011
New Version TosohCatalogue v6.0
Integrated Service & Application Platform
Newsletter December 2010
We are pleased to announce the release of the new ISAP_DVD_Deptember2010: Integrated Service and
Application Platform in DVD format for our Partners.
The enclosed DVD contains all service and application related documents for the Tosoh analysers AIA-360,
AIA-600II, AIA-1800, AIA-2000, G7, and G8. It also contains the PIANO and TosohCatalogue software.
This New ISAP_DVD needs to be copied and handed over to ALL Service Engineers and
Application Specialists working with Tosoh instruments! Discard the previous version September
2010 and only use this version!
In this Newsletter you will find all REVIEWED instructions you need to follow when:
All above mentioned instructions and documents are also available in the
ISAP_DVD/SERVICEORGANIZATION.
In chapter M3 you will find a detailed overview of all new released documents since September
2010.
This document explains the instructions our Partners have to follow after receiving the ISAP.
1. Explanation
Every Distributor Service Manager will receive 3 or 4 times a year a new ISAP_DVD (Integrated
Service and Application Platform) containing all updated service and application related documents
together with a Newsletter explaining all Critical Work Procedures and an overview of all new
released documents since the last ISAP edition.
Only use the last ISAP_DVD and make sure to discard the previous version.
Make sure that the new ISAP_DVD is copied and supplied to all Tosoh qualified Service Engineers
and Product Specialists.
If important technical information is available in between 2 ISAP_DVD releases, the Service Manager
will receive the Technical Bulletins by E-mail. It is important to spread this information to all Tosoh
qualified Service Engineers.
In order to follow up and to make sure that you actually received the ISAP_DVD, we ask you to
complete the TosohBioscienceISAPResponse Form and return it at the earliest convenience. This
Form is enclosed in the package together with DVD and Newsletter. This form is also available on the
ISAP_DVD in the folder NEWSLETTERS/ISAP_DVD_ResponseForm.
2. Procedure
Step2: Make Copies of the ISAP_DVD and spread to all qualified Service Engineers, Product
managers and Product Specialists.
This document explains the definitions and instructions to follow when one of our Partners
needs support in solving a technical or application problem: Support is available by Phone,
E-mail or by an Intervention from Tosoh.
1. SUPPORTLINE:
1.1. Definitions:
The SupportLine is available from Monday till Friday from 9h until 17h (European time)
After working hours you can speak in a message, this will generate a WAV file that will be send in the
form of an E-mail to the responsible Tosoh Service and Application specialist.
The SupportLine is NOT available during the weekends on Saturday and Sunday and will not be
available on the legal holidays in 2011:
2. E-MAIL SUPPORT:
2.1. Definitions:
Step 2: Select the type of Problem and complete corresponding section 2A, 2B, 2C or 2D
o Select the kind of problem you want to report, only 1 selection possible
o Now complete the corresponding section!
DistributorName_SerialNumber_YYYYMMDD
Step 5: Reporting the solution by the Distributor in section 5 of the D14 document
o Fill in the date when the problem was solved.
o Fill in the actions that you have undertaken to solve the reported problem.
o Fill in the cause of the problem and select if the cause was Instrument or Application related (only
select one possible cause).
o Fill in further explanations about the cause and main cause of the problem.
3.1. Definitions:
As a result of the service efforts performed by the Distributor‟s service organization, the request for
assistance can be used in the following cases :
As a result of an intervention caused by an instrument malfunction and where a proper
troubleshooting is not possible, assistance can be requested by phone, e-Mail or remote
access as explained in chapter 1 and 2.
Upon the efforts of the service engineer on site, the request for support may be raised -
without the special demand of spare parts or special delivery of parts- to restore the
instrument.
In consequence to the escalation procedure, at the demand of the distributor an intervention can be
requested from Tosoh or Tosoh qualified staff members when the problem is not solved despite the
remote assistance via telephone, e-mail and/or remote access.
The distributor commits to cover the costs linked to the request for support. He therefore accepts to
support the costs related to the arrival of a technical- and/or application support specialist. In order to
create a cost estimate, it is important that the demand for support does also contain a clear
description of the problem and an overview of the history of the service call by completing the
D14_SRS document. Based on the correctness of the supplied information, Tosoh will provide an
estimated cost definition sheet which is not restricted or binding for both parties. If due to incomplete
information, the cost would be higher than the estimate, a mutual acceptance of this access will be
worked out by Tosoh.
During the intervention the involved Distributor SE or AS has to be present.
3.3. Validation:
A validation of the escalation process is required, once the estimated costs are accepted and confirmed
by date and signature of the distributor. Also when the estimated costs are not accepted, a confirmation
will be required with a clear disagreement to the offered cost calculation. The intervention in this case
will not be executed by Tosoh and the incident will be terminated within the Tosoh call registration .
Step1: Sent a complete filled in D14_SRS (see chapter 2.2) to distributor.support@tosoh.com and
explain why a TSEN intervention is needed. Provide all history of the instrument: Interventions, errors,
Step2: The Tosoh Distributor Department will provide extra information by E-mail or Hotline in order to
verify if all possible solutions have been tried out by the Distributor and if an Intervention is justified.
Step3: If an intervention by TSEN is justified. TSEN will send the D15_ECDS after negotiation with an
indication of the estimated cost based on the information provided.
Step5: Only after receiving a signed off D15_ ECDS, TSEN will contact the Distributor to take further
steps in organizing the intervention.
This document explains the definitions and instructions to follow when a Distributor installs a
new Tosoh Instrument.
Every New first Installation of a Tosoh Instrument needs to be reported by the Distributor with the
Service Report provided ( ISAP/SERVICEORGANIZATION/GENERALDOC/D12_ServiceReport)
After receiving the Service Report D12 the Installation and date will be registered by Tosoh on your
personalized Installed base document D08. This D08 document will be forwarded to each distributor
on a quarterly base.
The installation report date will be used to calculate the warranty period and „end of warranty visit‟.
2. Procedure:
Step 2: Print out the document and let the customer sign.
This document explains the definitions and instructions to follow when a Distributor wants to
claim a warranty and how to obtain the end of warranty maintenance kit.
If the 1 installation occurs within a delay of 6 months after shipment date, the end of warranty date
st
will be fixed to 1 year (new instruments) or 6 months (refurbished instruments) after installation date.
If the 1 installation occurs later than 6 months after shipment date, the warranty date will be fixed 1
st
year + 6 months (new instruments) or 1 year (refurbished instruments) after shipment date.
This report (D12) should be sent to TOSOH within a delay of 2 months following this installation.
If no installation report is received by TOSOH, the end of warranty date will be fixed 1 year new
instruments) or 6 months (refurbished instruments) after shipment date.
2. Warranty of parts
If during the defined warranty period one or more parts of the instruments are malfunctioning, these
parts have to be ordered at Tosoh in the regular way using Document D11 (see
ISAP/SERVICEORGANIZATION/GENERALDOC/D11)
Reporting of the malfunctioning should be done using customer signed Document D12 (or equivalent)
within a period of 2 months after detection of the malfunctioning.
Based on the correct reporting, Tosoh will issue a credit note for the same amount as parts under
warranty have been ordered.
Tosoh preserves the right to request the return of the defective part(s) or to directly contact the end-
user for further information.
The End of Warranty Visit is the maintenance visit made at the end of the manufacturer warranty
period (as defined under 1.).
The End of Warranty visit compensation will concern the cost of the maintenance kit used during this
visit.
The End of Warranty visit compensation will concern only new instruments bought directly to TOSOH
(Warranty is not valid for instrument acquired over third parties. Warranty is not transferrable to third
parties or sub-distributors).
The End of Warranty Visit should be made before the end of warranty date, if this date is overdue, the
compensation will not be applied.
The End of Warranty visit compensations request will be made on a quarterly basis, by sending to
TOSOH a copy of the End of warranty visits service reports signed by the end-user.
This request should be sent to TOSOH before the 10 of following month (e.g. 10 of April, July,
th th
The end of warranty visit should be reported using customer signed Document D12 and should clearly
show:
o The character of the visit (Maintenance or End of Warranty).
o The serial number of the instrument.
o The maintenance date.
o The maintenance kit part number.
The end of warranty compensation will concern only the maintenance kit cost, i.e. the Maintenance kit
ordered by Distributor and used during the maintenance as a single part (single part number).
Parts replaced during end of warranty visit, other than contained in the maintenance kit, will not be
included in the compensation. The compensation of these parts may be reclaimed following the regular
warranty procedure (see 2.).
Any compensation on maintenance kits will be made on a quarterly basis, through a credit note, after
validation by TOSOH.
This document explains the instruction that need to be followed strictly when returning a Tosoh
instrument to Tosoh Headquarters.
1. Explanation:
All Tosoh instruments returned from a distributor to Tosoh HQ need to be decontaminated and
labelled with the decontamination label.
These are mandatory requirements from our transport companies and our supply chain:
(1) Distributor will be held responsible for all claims caused by the fact the instrument
was not decontaminated.
(2) When a non decontaminated, non labelled instrument will be received, a fee of 800
Euro for in-house decontamination will be charged to the responsible distributor.
2. How to proceed:
Step 1: Request and inform distributor.support@tosoh.com the reason why you want to return an
instrument
Step 2: Follow the decontamination procedures that are available on the ISAP_DVD:
AIA_SERVICE/AIA_GeneralService/AIA_Decontamination_Waste
GENERALSERVICE/Decontamination_v20100601
GENERALSERVICE/DecontaminationLabel
Make sure the Decontamination Label is located on the instrument as described in the procedure, printed
with name and signature of person who performed decontamination.
Step 3: Follow the instrument specific de-installation procedures available on the ISAP_DVD
Step 4: Include a Service Report to the instrument explaining reason of return and visual damages, send
the service report to distributor.support@tosoh.com
Step 5: Use the ORIGINAL boxes to pack your instrument and check if all accessories are available; in
case the boxes are not available please contact distributor.support@tosoh.com
1. Explanations:
Information about Spare Parts, Tools, Accessories, Consumables and Reagents can be found in the
TosohCatalogue (ISAP/TOSOHCATALOG) or in service manuals available on the ISAP.
All inquiries about pricing, availability, orders for Spare Parts, Reagents and Instruments
should be send to:
customerservice.diag.eu@tosoh.com
2. How to proceed:
Every Distributor need to be able to install, troubleshoot, repair and maintain Tosoh Bioscience
Instruments.
Only qualified Service Engineers who participated training can service Tosoh Instruments.
All qualified Service Engineers who received a Service Training from Tosoh will be logged in our
database and will receive a Certificate.
Service Trainings are organized at Tosoh HQ or can be organized at the distributor site on special
request.
All training participants to the training sessions should correspond to the following profile: Educational
level of bachelor in science with special attention for Electronics and Mechanics, basic tooling skills
like soldering, field experience and self management.
Step 2: Complete the registration form Service Training Academy 2010 Registration form.
Remarks:
Contact travel.eu@tosoh.com for all further requirements and questions regarding hotel booking, taxi
pickups and flights
Contact travel.eu@tosoh.com immediately in case you can not attend the training for any
reason
This document provides lists of general and specific tools that every distributor Tosoh Qualified
Service Engineer needs to have in their personal stock.
1. EXPLANATION:
Every Distributor needs to provide all Service Engineers with the General Tools to be able to perform
technical interventions and maintenances. Non-compliance can lead to disqualification of the work
performed.
For every Tool listed in the tables below an indicator explains whether the tool is: Mandatory, Strongly
advised or optional. All Tools can be ordered at Tosoh or similar items can be bought locally.
General Tools are Tools that are required to work on all Tosoh equipment and are split up in:
IT, Check&Repair, Working, Measure, Lubricants, Cleaner, Decontaminate, Mics.
Specific Tools are Tools necessary to perform specific adjustments or repairs and are listed per
instrument.
For more detailed information all Tools can be looked up in the TosohCatalogue on the ISAP_DVD.
Specific Instrument Toolkits will be provided as a „gift‟ with the active participation to the service training.
2. GENERAL TOOLS:
AIA-1800
AIA-21
AIA-600II
AIA-2000
G5
G7
This document explains and defines the Minimum Spare Parts every Distributor needs to
have in stock to ensure fast repairs and quality interventions. It also provides the different
spare part lists.
Every Distributor has the commitment to have a minimum of Spare Parts in his Country for the following
reasons:
o Saving time and costs in the process of ordering and transportation of spare parts
o Faster troubleshooting and repair at first intervention
o Avoiding problem escalation and customer satisfaction
Tosoh has defined the Minimum Country Stock based on experience and turnover of used parts.
Detailed lists with pictures and alternative parts can be viewed and printed out from the
TosohCatalogue access program available on the ISAP.
Every Distributor need to check on a regular base the availability of the Spare Parts he has in stock.
The Minimum Country Stock will be subject for future external audits.
The Minimum Country Stock is used to perform fast interventions within the Distributor territory; the
parts are not to be used to carry out scheduled maintenances. Maintenances are carried out with parts
or kits specifically ordered to perform the planned maintenance.
The Minimum Country Stock is split up in two stocks: Service Engineer stock and Warehouse Stock.
o Warehouse stock: Spare parts that need to be in the warehouse are listed in column
„Stock Warehouse‟. These are statistically low rated defective parts.
In the worksheet „CrossTablePartsInstruments‟ a list is available of Spare Parts that are used in
different Tosoh Instruments.
The Spare Parts are alphabetically listed per Group and then per FunctionDescription.
All Minimum Spare Part Lists can also be viewed and printed out by using the „TosohCatalogue‟
available on the ISAP
This document explains the work procedures for distributors on how to report there installed
base, instrument status & Origin, software upgrades, installations, maintenances and updates.
Every Distributor Service Manager will receive on a quarterly base an E-mail with his personal Tosoh
Installed Base in Excel format:
In this report or document D08 the distributor need to report all changes within his Tosoh Installed Base:
Installation date, Software Upgrade, Updates, Instrument Status & Origin and Scheduled Maintenances.
Refer to chapter 7 to find the detailed instructions on how to complete the report.
The format of the list is protected and can not be changed only extra rows can be added to add new
instruments you received from Tosoh or bought from third parties.
The Excel sheet needs to be returned before the specified day as listed above.
2. Software Upgrades:
Why to keep your software up to date:
4. Maintenance:
Two columns are foreseen were the dates of ‘last maintenance’ and next „planned maintenance’ needs
to be reported. All maintenance procedures are available on the ISAP.
As Tosoh is estimating the forecast of future reagent deliveries we need to know exactly which instruments
are active. An instrument is active when it‟s consuming reagents. Active [y/n] can be selected in the list.
Three columns are foreseen were the distributor needs to report if the instrument is installed, still in the
Warehouse or Scrap. An instrument is „Scrap‟ when it no longer exists, sold to 3 party or when it is
confirmed that it will never be used again (demolished).
The distributor is committed to report all Tosoh instruments he bought from 3 parties and from TSEN in his
territory and needs to add them to the list.
Two columns are foreseen to report ‘Bought from 3 Party’ or ‘Sold to 3 Party’
6. Mandatory Updates:
Tosoh will inform the distributor when mandatory updates needs to be performed. A Technical Bulletin will
be send out with the details and actions to perform
Two columns are foreseen to report the „Update Date’ and „TB number’
Remarks:
The format of the list is protected and can not be changed only extra rows can be
added to add new instruments.
Return the excel list by E-mail before the specified day (see table above)
The goal is that ALL instruments need to be upgraded to the software version as listed in
table 2 above ‘Goal end 2010’
Distributors not complying with the above instructions will be contacted personally
Detailed procedures and explanations about software upgrades are also available in the service manuals,
training manuals or technical bulletins.
Remarks:
Upgrading an instrument to a new version does not automatically install the new test file
parameters or sometimes system parameters!
New test file parameters or sometimes system parameters need to be set manually
according to the released TB.
Some new system parameters can only be set after a CLR function (for G7 and G8) or
Program reset as indicated in the TB. In this case a complete data backup or parameter
print out needs to be performed in order not to loose the instrument settings and/or
calibrations.
Always double check the test file specifications in the instrument with the latest test file
overview in the ISAP before starting to calibrate this test:
AIA_Application/‟SelectInstrument‟/TestFiles
If an instrument needs to be reinstalled, refurbished or facing software problems it is
recommended to perform a program erase and reinstall the complete new version with all
new system parameters and latest test files! Readjustment and recalibrating will be
required!
Please check all technical bulletins for system parameter updates or modifications if you
are not sure if previous installed software was performed correctly! An overview is given
here below per instrument when upgrading the software from a specific version.
To be sure if all former system parameters are set from previous versions by the SE you
need to check all system parameter updates since the introduction of the instrument
The „TosohCatalogue‟ will enable you to search for all technical bulletins related to different
subjects: Test file Updates, Software updates, System parameter updates,…
If software problems are suspected or if instruments causing unexplained problems
complete software reset might be very useful. Installing the latest software after a complete
reset will set all parameters at the default value and will add all latest system parameters.
In this case all calibrations will be lost.
The software upgrade is done by a Smart Media Card Max 32Mbyte. A special card
reader is needed to download the software from the ISAP to your PC: 940086
Download the new software from the ISAP to the SM that was delivered with the
instrument or by a SM card that you ordered: 021139. Do not format the SM card on
your PC nor on the instrument, just copy all software files from the folder
AIA360/SOFTWARE/ version… from the ISAP to the SM card.
Insert the SM card in the instrument slot at the right side with gold contacts facing to
the front.
Start the instrument in the normal mode and perform a DATA BACKUP first: MAINTE
MENU/ DATA BACKUP all data will be saved on the SM as AIA360.dat.
Now start the instrument with PAPER FEED to enter the software upgrade mode.
A screen appears with 4 programs that can be downloaded: 1. Main, 2. Slave, 3. Touch
Panel, 4. Installation. The main program always needs to be uploaded.
The other installed programs have a version number and are mentioned on the screen.
Check the actual version of these programs with the new versions mentioned in the TB
of the latest software release! Upload if necessary.
After the software upgrade re-start the instrument in the normal mode.
Remarks:
After setting new parameters, exit the screen to save new parameters,
then turn off and turn on the instrument again!
The addition of test file parameter information mentioned in the TB‟s
should be done manually and are not updated automatically when
upgrading the software.
Please note if one uses Test file initialize in Exam menu, all new test
specifications can be added when installing the new software but the
calibration information will be erased!
From soft 1.11: T042: parameter 57 was introduced: „Calib Void Period‟. (Please set on 60)
The software upgrade is done with a max 16Mbyte SM card. A card reader for your PC
can be ordered: 940086
This instrument is not delivered with the SM card! If you don‟t have one you will need to
order one and use it to upgrade all your installed base of AIA-600II: 019400
If the card contains old software, then just overwrite it with the new software from the
DVD, do not format the card!
If the card is not delivered by Tosoh then it might be necessary to format the card first if
it‟s a complete new one:
Start the instrument in Boot mode level 2: Start while pressing Assay Monitor &
Special menu.
Perform SM format.
Switch off instrument and card is now formatted and ready to load the software
from the DVD.
Start the instrument in the test mode and make a printout of the machine conditions to
be on the save side if something should happen.
Read TB of the latest released version for new features and remarks
Insert the SM in the instrument at the right side panel slot with contacts facing to the
back.
Start instrument in the Boot mode level 1: Start while pressing Assay Monitor to enter
the software upgrade mode.
Press 1 to update
Select Set Flash Card No [yes] and press enter.
Wait until directory is listed and loading is processed
After update, list disappears
Press 2 to exit to boot menu.
More information about software update can be found in the AIA-600II training guide.
AIA-600II new system parameters or modifications to check after installing latest software:
Remarks:
After setting new parameters, exit the screen to save new parameters
then turn off and turn on the instrument again!
The addition of test file parameter information mentioned in the TB‟s
should be done manually and are not updated automatically when
upgrading the new software.
Please note if one uses 2: Memory initialize at 8: Support utility, test
specifications can be added when installing the new software but the
calibration information, and all other adjustments will be erased! This
option can be useful during instrument refurbishing to set everything at
default value.
From soft 2.02: T014: Parameter 33: BF-wash dead volume (please set to 0)
34: Pickup H offset (normally same value as 3)
35: H offset cupstand (needs to be adjusted)
36: Cup Waste Height (needs to be adjusted)
From soft 3.12: T056: Parameter 37: FT3 Auto Check (must be -1)
39: BF- Wash Inhale offset (needs to be adjusted)
40: Sampler tube size (Must be -1)
41: Sampler tube Bottom pulse (Must be -1)
From soft 3.19: T072: Parameter 42: RS232C out data (no result):-1:blanc output;1:0 output
43: RS232C out data(<L,>H):-1:blanc;1:0 output;2:Range;3: < or >
From soft 3.22: T047: Parameter 47: Calibvoidperiod: (please set to 60!)
Copy the complete NEXUS folder to another location on the Hard Disk or other
medium.
Read the latest software upgrade TB about new features and remarks.
Directly from the DVD (if PC does not have DVD reader, copy first to memory stick or
CD) go to the latest software version for AIA-1800 and double click Setup.exe.
Follow instruction in T082 for installing the new software!
AIA1800 new system parameters or modifications after installing latest software 4.38:
Remarks:
The addition of test file parameter information mentioned in the TB‟s
should be done manually and are not updated automatically when
upgrading the new software.
The test file information is available on the ISAP in the folder:
APPLICATION/AIA-1800/TESTFILES and new tests can be added to
the instrument.
New test file information from an existing test as mentioned in the TB‟s
can be added but doing so will erase the already set information for
this test like „reference ranges‟, „reschedule ranges‟ and so on. Input
the changes manually or printout this test file, add the test file from the
ISAP and restore manually the parameters from the printout.
From soft 4.00: T048: Mandatory mechanical upgrade when upgrading from soft 4.00
From soft 4.11: T054: Nexusconst/Mode[p03]/ parameter 16:Sampling Check (set disable )
From soft 4.31: TB82: CheckControlFlagType =0 (no indication H and L); =1 (indication H and L)
Copy the complete MAXIA folder to another location on the Hard Disk or other medium.
Read the latest software upgrade TB about new features and remarks.
Directly from the DVD (if PC does not have DVD reader, copy first to memory stick or
CD) go to the latest software version for AIA-2000 and double click Setup.exe.
Follow instruction in T026 for installing the new software!
AIA2000 new system parameters or modifications after installing latest software 1.44:
Remarks:
The addition of test file parameter information mentioned in the TB‟s
should be done manually and are not updated automatically when
upgrading the new software.
The test file information is available on the ISAP in the folder:
APPLICATION/AIA-2000/TESTFILES and new tests can be added to
the instrument.
New test file information from an existing test as mentioned in the TB‟s
can be added but doing so will erase the already set information for
this test like „reference ranges‟, „reschedule ranges‟ and so on. Input
the changes manually or printout this test file, add the test file from the
ISAP and restore manually the parameters from the printout.
Validate that the following settings have been set in the MaxiaConst.ini file by going into the
MaxiaConst editor of the mainte program. These settings can also be found in the new
Installation Guideline.
Remarks:
The system.prm version in software version 3.01 is different as the system.prm version
in software version 2.00 and lower. Therefore all system parameters will be automatically
initialized when downloading the new software in an instrument installed with version
2.00 and lower! This means all specific customer, engineer settings will be lost!
Read T092 to be informed about the new features in this software!
When upgrading to version 3.01 from a version equal to or lower then 2.00 only
follow procedure below!
Do not change other parameters then those listed here below as some are factory set!
Remarks:
The system.prm version in this software version 3.01 is the same as the system.prm
version in software version 3.00. Therefore you can simply load this new software version
and there are no new parameter settings from version 3.00 to version 3.01 as explained
in T092.
Read T092 to be informed about the new changes.
If you are not sure if the software 3.00 was installed correctly last time or when
facing problems then perform the software upgrade to version 3.01 as explained in
chapter E!
Download the software version 3.01 files from the ISAP to a floppy disk that was
formatted by the G7.
Insert the soft 3.01 disk and start up the instrument with the main power switch and press
power on the key path. Wait until the instruments starts up.
To be 100% sure if previous versions were installed correctly it might be very
useful to verify all parameter settings in the table below!
Remarks:
The system.prm version in software version 5.02 is different as the system.prm version
in software version 4.00 and lower. Therefore all system parameters will be automatically
initialized when downloading the new software in an instrument installed with version
4.00 and lower! This means all specific customers, engineer settings and results will be
lost!
Read T082 to be informed about the new features in this software!
If results need to be saved follow the instructions in T072.
When upgrading to version 5.02 from a version equal to or lower then 4.00 only
follow procedure below!
The software upgrade is done with a Smart Media Card of Max128Mbyte. A card reader
can be ordered: 940086
Download the software version 5.02 files from the ISAP to the Smart Media that was
delivered with the instrument. Do not format the card on your PC or the instrument,
just copy, paste and replace the software files in the folder version 5.02 with the ones on
the SM card.
Enter password MAINTE and print out all parameters in all the different modes the
customer is working in. If the customer is using Variant and Thallasemia you need to print
out 2 lists!
Go into the “MAINTE” folder, double tap the top left corner and go into the ENG MENU.
Note the counter settings for the Life and Test.
Enter the password CLR. Return to the Standby screen and turn off the instrument with
the power key, then switch off the main unit by means of the main circuit breaker.
Insert the SM card and start up the instrument with the main power switch and press
power on the key path. Wait until the instruments starts up. Now the latest system.prm
V5.02 is installed on the instrument and all parameters are initialized!
Enter the password CMODE0110 to activate Variant and B-thall mode.
Some customer and engineering parameter settings that were initialized needs to be set
back in all the different modes. Verify the following parameters for all modes the
customer is working in (variant and/or B-thall):
In the utility screen select RS232 and correct the parameters from the printout.
In the utility screen select BCR and correct the parameters from the printout
In the utility screen select FLAG and correct all the flag settings from the printout!
In the utility screen select PASSWORD.
Only in case the live counter was reset: Enter the PASSWORD: “RESETL xxxxxx”
whereby xxxxxx = the old value for the Life count
Only in case the live counter was reset: Enter the PASSWORD: “RESETT yyyyyy”
whereby yyyyyy = the old value for the Test count
After setting all these parameters and to make sure that all changes take effect, return to
the Standby screen and turn off the instrument with the power key on the front, then
switch off the main unit by means of the main circuit breaker.
Remarks:
The system.prm version in this software version 5.02 is the same as the system.prm
version in software version 5.00 and higher. Therefore you can simply load this new
software version but you need to set some extra parameters as explained in T082 and
listed in the table below see number 5.
Read T082 to be informed about the new changes.
If you are not sure if the software 5.02 was installed correctly last time or when
facing problems then perform the software upgrade to version 5.02 as explained in
chapter G!
The software upgrade is done with a Smart Media Card of Max128Mbyte. A card reader
can be ordered: 940086
Download the software version 5.02 files from the ISAP to the Smart Media that was
delivered with the instrument. Do not format the card on your PC or the instrument,
just copy, paste and replace the software files in the folder version 5.02 with the ones on
the SM card.
Insert the SM card and start up the instrument with the main power switch and press
power on the key path. Wait until the instruments starts up.
Set the parameters manually as explained in T082 or use table below (see Nr:5)
To be 100% sure if previous versions were installed correctly it might be very
useful to verify all parameter settings in the table below!
This platform contains all service and application related information in a new structured way. The
application information is now available in separate folders for AIA and HPLC.
ISAP_DISTR_DECEMBER2010:
GENERALSERVICEDOC
TOOLS
AIA_APPLICATION
AIA21_APPLICATION
AIA360_APPLICATION
AIA600II_APPLICATION
AIA1800_APPLICATION
AIA2000_APPLICATION
AIA_COMPETITION
AIA_STUDIES
AIA_APPLICATIONBULLETINS
AIA_GENERALAPPLICATION
AIA_SERVICE
AIA21
AIA360
AIA600II
AIA1800
AIA2000
AIA-GENERALSERVICEDOC
HPLC_APPLICATIE
G7_APPLICATIE
G8_APPLICATIE
HPLC_APPLICATIONBULLETINS
HPLC_COMPETITION
HPLC_GENERALAPPLICATION
HPLC_STUDIES
HPLC_SERVICE
G5
G7
G8
HPLC-GENERALSERVICEDOC
NEWSLETTERSUPDATES
The Tosoh Catalogue is an access application database with advanced search functions and report
functions giving a total overview of all spare parts, consumables, accessories, reagents, technical
bulletins,… available with pictures. The search result can be printed out or exported in another format like
excel.
Each search function will give a list with parts and also will give alternative parts that can be ordered if the
original is not available.
In addition to the spare part database a technical document database has been introduced so from now
on you can view and look up technical bulletins by different categories.
Both spare parts and technical bulletin databases are linked so that specific information or technical
bulletins can be shown from a spare part.
2. What is new?
Home Page: