Chapter 1 Foreword ................................................................................ 2


Chapter 2 Function and operation..................................................................... 3
2.1 Simple operation..............................................................................................................3
2.2 FUNTION INTRODUCTION............................................................................................ 4
2.2-1 FUNTION MODULE DIAGRAM....................................................................................4
2.2-2 connection port of complete machine.................................................................... 7

Chapter 3 MAINTENANCE ............................................................................11


3.1 ROUTINE MAINTENANCE.......................................................................................... 11
3.1.1 Clean the Appearance................................................................................................. 11
3.1.2 Replace Fuse............................................................................................................... 12
3.2 DAILY MAINTENANCE............................................................................................... 12
3.3 WEEKLY MAINTENANCE.......................................................................................... 12
3.4 MONTHLY MAINTENANCE....................................................................................... 13
3.5 YEARLY MAINTENANCE...........................................................................................13
3.6 MAINTENANCE BEFORE TRANSPORT OR FOR THE INSTRUMENT THAT
WILL NOT BE USED FOR A LONG TIME............................................................................ 13

Chapter4 troubleshooting.............................................................................. 15
Chapter5 Liquid System diagram................................................................21

edition:A/0

1-
Chapter 1 Foreword

This manual includes the information about functionality daily operation,


and maintenance. To achieve the optimal performance, be sure to operate and
maintain the instrument according to this manual.

Functionality
Operation
maintenance.

In the chapter 2-there are diagrams related to express function and


operation,

In the chapter 3- Maintenance is mentioned


In the chapter 4- This includes Errors encountered during operation of the
machine and trouble shooting

2-
Chapter 2 Function and operation

2.1 Simple Operation

First: front panel

Display

Sample needle

Power light

Start Key

2-1 Diagram Front Panel

3-
Second:Operation Process

1) After installing the machine,put the power on and enter into the system

2) Test blank once,after that it will show the data for the background check .Repeat

blank test till the values of all the five parameters namely WBC, RBC, HgB,PLT and

MCV are within the acceptable range.

3) Once Blank is in acceptable range patient samples can be tested . Put the patient

sample under the sample probe and press the start key . Ensure that bubble is not

entering during the aspiration procedure . Move the sample tube along with the

sample probe so that sample enters correctly into the measuring chamber.

4) Before the test is completed no other operation can be executed . After completion

the needle come back to the home position and is ready for the next sample

aspiration.

5) Press the key “print”. The results are printed on the thermal printer paper

6) After finishing measure sample click the right key,the menu will pop-up

last bar of the menu is “shutdown” key,press this key,it will show “please

shutdown power “ a few minutes later, operator can turn off the power.

4-
2.2 ) Function Introduction

2.2-1. Function Module Diagram

1. Interface board

P104 bus

Motor control CPLD--->printer driver Printer driver


One-way
and CPLD total
WBC-RBC
drive unit control unit CPLD<--- W¡¢R

enlarge

Electromagnetic
Two-way
CPLD<---HGB HGB ·enlarge
Valve drive

Analog power supply

Simple Introduction
1) CPLD total control unit.
Control every function module how to communicate with P104.
2) Motor control and driver
Control motor working and amplify signal
3) Electromagnetic valve driver.
Amplify the signal of electromagnetic valve driver
4) Printer driver
Amplify the signal of printer driver.
5) WBC-RBC amplify
Amplified WBC-RBC signal that can be earlier to change it to digital signal.

5-
6) HGB amplify
1) Amplified HGB signal that can be earlier to change it to digital signal.
2) Supply power to the HGB lamp.
7) P104 bus.
The bus that Connect computer board with CPLD of interface board.
8) Analog power supply.
1) Supply power to each part,+48V ,+12V ,+8V , 8V.
2) Protect each part supply power.

2) Mechanism Fasten Board Module Introduction

1) Function Diagram

Sampling pump
TWO WAY

Sampling
needle Liquid
Two way

and road

bracket And flow


Sensor Vacuum
Two way
Measure board pump
cup

2 . Simple Introduction
1) Liquid tubing board and flow sensor board.
1) Liquid tubing board.

Each valve control the liquid tubing in measurement.

2) Flow sensor board.

Precise ration for the measured volume.

6-
2) Sample needle.

Aspirate outer sample.

3) Measurement cup

The place where measuring sample is kept

4) Vacuum pump

Let the related liquid from measured cup to vacuum room.

5) Sampling pump

Create force when aspirating sample

7-
Chapter 3- MAINTENANCE

As other precision instrument, only careful daily service and periodic

maintenance can keep the instrument in a good working status, and can we get the

reliable results and have few errors

This chapter introduces some preventive methods for service and maintenance.

If you want to have more information, please contact the customer service department of

MERIL DIAGNOSTICS PVT LTD.

According to the requirements for maintenance of the instrument during the procedures,

we divide the preventive service and maintenance into the following types: daily, weekly,

monthly, yearly and maintenance according to actual need.

Warning:

It is important for the hospital or organization that employs this instrument to

carry out a reasonable maintenance schedule. Neglecting of this may result in machine

breakdown.

3.1 ROUTINE MAINTENANCE

3.1.1 Clean the Appearance


Wipe the appearance of the instrument with neutral detergent or distilled water.

Caution:

 Avoid using corrosive acids, alkali, and volatile organic solvent such as: acetone,
ether, chloroforms to wipe the appearance of instrument. Only neutral detergent

should be used.

 Avoid wiping the inner of the instrument.

8-
3.1.2 Replace Fuse
The fuse is installed in the fuse-box on the side of power switch. Open the box to

replace the fuse

Appointed specification fuse: AC 250V 2A

Warning:

Only appointed specification fuse can be used.

3.1 Daily Maintenance

It can be divided into two types: run and shutdown.

 Run

The instrument has installed daily maintenance procedure. On running, it can execute

auto cleaning procedure according to the quantity of the sample to keep the

instrument in good working status.

The general setting principles are:

Working time > 8 hours, auto-cleaning time=8 hours;

4 hours<Working time<8 hours, auto-cleaning time=4 hours;

Working time< 4 hours, auto-cleaning time=2 hours;

The auto-cleaning time will decrease 1 hour year by year.


 Shutdown

When the instrument is shutdown, it will run daily shutdown auto cleaning procedure.

You only need to clean the workbench and wipe the appearance of instrument when

the power-off.

3.2 Weekly Maintenance


This instrument has installed weekly maintenance procedure.

Act as follows:

a) Prepare E-Z cleanser and probe cleanser.

9-
b) In the main operation window, press the right key of the mouse. In the menu area,

select "Service/Maintenance". Operate as the screen shows.

c) Turn off the power supply.

3.3 Monthly Maintenance

It is necessary to clean the dust of the instrument monthly.

Operate as follows:
a) Turn off the power supply, and pull out the power line.

b) Hold up the air filter cover with slotted screwdriver. Take off the cover and the

net.

c) Clean them with neutral detergent brush and put them in the shade.

d) Airing naturally .Brush away the dust clinging on the air filter grid at the rear

panel of instrument.

e) Put the air filter grid on the air filter net carefully. Planish them.

Notice:

Only can use the neutral detergent cleaning the air filter. Avoid airing it with heat.

3.4 Yearly Maintenance

It's necessary to perform preventive maintenance once every year. Because of

the high -requirements of yearly maintenance, the maintenance should be performed by

the authorized engineer of MERIL DIAGNOSTICS PVT LTD. Please contact the customer

service department MERIL DIAGNOSTICS PVT LTD of before yearly maintenance.

10 -
MAINTENANCE BEFORE TRANSPORT OR FOR THE INSTRUMENT

THAT WILL NOT BE USED FOR A LONG TIME

If the instrument is not going to be used in 2 weeks or longer, or needs to be


packed and transported, perform he following procedures:
 In the main operation window, press the right key of the mouse, choose
"Service/Park" in the menu.

 Operate as screen shows.


 Rotate the bottle covers of reagents and store them in proper condition . User
should perform efficient action to prevent the material from deteriorating.

 Plug the stopples which were pulled out at the first installation into the
corresponding tube connector.

 Pull out diluent connection tube, rinse connection tube, lyse connection tube and
waste connection tube and clean them with distilled water. Dry them in shady

place,then packing them into plastics.

 Pull out the power wire,packing it in plastics after cleaning with neutral detergent.
 Put the instrument and parts packed in plastics into packing cases.

11 -
Chapter-4 TRAUBLESHOOTING

Malfunction
Reason Treatment program
phenomenon

A:
1)Checking the tube connector whether is
it leakage
2)Enter into submenu :
the situation which liquid flow
Mechanic:(password:6800)
into the flow sensor board.
checking each solenoid
A .when the liquid flow too
valve whether it is normal , if
WBC,RBC fast
not, change it.
bubble B .when liquid flow into flow
B)when taking count of measurement
sensor, both lamp is always
time ,check the lamp: when there is liquid
on or bottom lamp is always
the lamp will be on, no liquid the lamp will
on
be off. If the lamp is always on, press
several pieces of gummed papers until it
is normal. for other reason, change the
board.

A) The most possible is aperture clog :1)


Enhance flush
Detecting the situation
2) Cauterant (aperture damaged
which liquid flow into the flow
by caustic reagent )aperture
sensor board.
3) Clean the ruby and reassemble the
.
measurement cup module
A)The liquid flow along wall.
WBC and B) Use EZ cleaner to clean wall of flow
B)The liquid flow is normal.
RBC clog sensor board module.
C )The liquid flow slowly.
C) check whether the two lamp is on, if
D) wether is the second foot
the lamp is always off, change the the
Of interface board
board.
connector J50 is connected
D)Resulting from leakage in the vacuum
with ground
pump E connect the second foot to the
ground.
A) Use E-Z cleaner to clean wall of
Detecting the situation in
flow sensor board module
which liquid flow into the flow
B) Change the flow sensor board
WBC or RBC sensor board
module
clog A) The liquid flow along wall
C)Enter into special(password:
B) The liquid flow is normal
6868), the setting time is always
C) The liquid flow slowly.
9.5 S

12 -
A) Whether the HBG lamp is
on. A) If the lamp is off , check whether the 1
B) whether the HGB receiver feet of U57 is 5V, if not ,change the
HGB
line have problem. interface board
Malfunction
C)Enter into mechanic B) Reconnecting the receiver line
HGB bubble
(password:6800),check C)Adjust the resistance VR6 to normal
whether the HGB voltage is range 4.25—4.31V
normal range 4.25—4.31

A) Detect whether others


A) If it is stable ,follow B step. If it is
items is stable
Hemoglobin unstable, check the sampling volume
B)Enter into mechanic
unstable B) If the HGB blank voltage is larger than
(password:6800) check the
0.2V, change the HGB lamp
HGB blank voltage fluctuation

A) Ground line state.


B whether it is shared socket
with large efficient machine
C whether environment
temperature is suit for the
request of the machine A detect the voltage of the groundwire
D the constant: maintenance and zero curve whether it is less than
once every week 5V.if it is more than 5V ,reassemble
E whether the reagent is B it is separated from the large efficient
PLT value is expired date or sordid machine
too high in F check every module C environment temperature is suit for the
ground test especially the test cup request of the machine
module whether it is rusting, D if not, use the function “park” clean all
whether the fasten board is the tubing
separated from ground.(how E change the reagent
to check the separation: unfix F descaling ,reassemble
the signal wire, check
whether the fasten board is
separated from cabinet. The
right assemblage is
separation

13 -
A ground line state
B whether it is shared socket
with large efficient machine
C whether environment
temperature is suit for the
request of the machine
A detect the voltage of the groundwire
D the constant
and zero curve whether it is less than
maintenance :once every
5V.if it is more than 5V ,reassemble
week
B it is separated from the large efficient
E whether the reagent is
machine
expired date or sordid
PLT unstable C environment temperature is suit for the
F check every module
request of the machine
especially the test cup
D if not, use the function “park” clean all
module whether it is rusting,
the tubing
whether the fasten board is
E change the reagent
separated from ground.(how
F descaling ,reassemble
to check the separation: unfix
the signal wire, check
whether the fasten board is
separated from cabinet. The
right assemblage is
separation

14 -
A ground line state
B whether it is shared socket
with large efficient machine
C whether environment
temperature is suit for the
request of the machine A detect the voltage of the groundwire
D the constant and zero curve whether it is less than
maintenance :once every 5V.if it is more than 5V ,reassemble
week B it is separated from the large efficient
E whether the reagent is machine
expired date or sordid C environment temperature is suit for the
All the items
F check every module request of the machine
unstable
especially the test cup D if not, use the function “park” clean all
module whether it is rusting, the tubing
whether the fasten board is E change the reagent
separated from ground.(how F descaling ,reassemble
to check the separation: unfix
the signal wire, check
whether the fasten board is
separated

from cabinet. The right


assemblage is separation

A check whether the voltage in the first


foot of connector J36 is 1.2V.if not,
change the sensor. Check the voltage of
A the vacuum pump move three foot in connector , when the sheet
in the same direction. metal enter into sensor, the voltage is
Vacuum
B the vacuum pump skid 4.0-5.0V; when the sheet metal is out of
pump skid
C when the vacuum pump sensor, the voltage is 0.0-1.0V . if
and noise
stop moving, the piston not ,change it
bounce B as the rack and the gear suffer wear
and tear ,change them.
C the waste tubing flow
disfluency ,dredge it

15 -
A if the value of HGB is
zero, the lyse is not enough
in the tube.
B enter into mechanic A change the piston gasket ring of
(password: 6800), check sampling room, and prime lyse.
WBC no whether the function B reconnect the cup board, if no use,
value or the “sampling” “break” is normal. change the interface board
value is”*” C If the value of HGB is not C if the screen show the pulse diagram,
zero, check the pulse of adjust the resistance VR8 clockwise or
WBC,RBC,PLT(enter into anticlockwise,
special, the password is
6666, press F5 to watch the
pulse of WBC )
A if the value of HGB is
zero, the lyse is not enough
in the tube. A reconnect the cup board, if no use,
B If the value of HGB is not change the interface board
RBC no value zero, check the pulse of B if the screen show the pulse diagram,
WBC,RBC,PLT(enter into adjust the resistance VR8 clockwise or
special, the password is anticlockwise,
6666, press F6 to watch the
pulse of RBC )

A enter into mechanic


(password:6800),check
whether the function
“sampling” “break” is normal
PLT no value B If the value of HGB is not A reconnect the cup board, if no use,
zero, check the pulse of change the interface board
WBC,RBC,PLT(enter into B if the screen show the pulse diagram,
special, the password is the same solution as “PLT TOO HIGH”
6666, press F7 to watch the
pulse of PLT )

A check the voltage of 1


feet of J33 whether it is 1.2V, A if not, the sensor have broken, change
B wheter the voltage of 3 it
distort no
feet is greater than 3.8 when B if the voltage far from the abnormal
lyse
the tube have liquid or less range ,change the sensor\Opposite,
than 2.7V when it have not adjust the resistance R153 or R144
liquid

16 -
A check the voltage of 1
feet of J33 whether it is 1.2V A if not, the sensor have broken, change
B wheter the voltage of 3 it
distort no
feet is greater than 3.8 when B if the voltage far from the abnormal
diluent
the tube have liquid or less range ,change the sensor\Opposite,
than 2.7V when it have not adjust the resistance R151or R142
liquid
A check the voltage of 1
feet of J33 whether it is 1.2V, A if not, the sensor have broken, change
B wheter the voltage of 3 it
distort no
feet is greater than 3.8 when B if the voltage far from the abnormal
rinse
the tube have liquid or less range ,change the sensor\Opposite,
than 2.7V when it have not adjust the resistance R152 or R143
liquid

A use the quality control to adjust the


not be good A inappropriate magnification
diagram of WBC
at B inappropriate sampling
B check the sampling
classification C inappropriate store of blood
C detect the store time

A the connector of screen have poor


A no display but other normal
Only blank in contact
B no display and others is
the screen B the cup board have poor contact or
abnormal
was broken
A alarm when it is not use theA there are some problem in the
Full waste waste sensor interface board
alarm B no alarm when it is not use B the waste sensor broken or the
the waste sensor waste sensor have bubble

REC NO Printer paper running out or not A check if there is printer paper
PAPER Properly installed B Check if the printer is well installed

17 -