You are on page 1of 5

5.

4 Bourdon and Helical Pressure Sensors PI

B. G. LIPTÁK (1969, 1982, 1995) J. E. JAMISON, S. EDVI (2003)

PT

Flow Sheet Symbol

Applications: Sensing elements for the measurements of medium to high pressures using atmo-
spheric reference. For pressure gauges and switches using these elements, see the
corresponding sections in this chapter.

Design Pressure: Up to 100,000 PSIG (690 MPa)

Design Temperature: Up to 600°F (316°C) with stainless steel Bourdon tube; lower with others; ±1% zero
shift per 50°F (28°C) ambient variation

Materials of Construction: Phosphor bronze, alloy steel, 316 and 403 stainless steel, beryllium copper, Monel,
Ni-Span C

Inaccuracy: 1/4 to 5% of span

Range: Minimum span 15 PSI (100 kPa); maximum span 100,000 PSI (690 MPa)

Costs: $50 to $300 for indicators and switches, $500 to $1,200 for direct recorders, and
$750 to $1,800 for direct controllers and transmitters. Quartz helix units range for
$2,500 for an electronic transmitter to $18,000 for an IEEE and RS-232 program-
mable, stable-alone station.

Partial List of Suppliers: ABB Automation Technology (formerly Kent-Taylor) (www.abb.com)


Ametek Inc., U.S. Gauge Div. (www.ametekusg.com)
Anderson Instrument Co. Inc. (www.andersoninstrumentco.com)
Bailey, a Unit of ABB (www.abb.com)
Barton Instruments Systems (www.barton-instruments.com)
Circle Seal Controls Inc. (www.circle-seal.com)
Dresser Industries, Instrument Div. (www.dresserinstruments.com)
Dwyer Instruments Inc. (www.dwyer-inst.com)
Helicoid Instruments Div. of Bristol Babcock (www.bristolbabcock.com)
Fischer & Porter, a Unit of ABB (www.abb.com)
Fisher Controls Int., a Div. of Emerson Process Management
(www.emersonprocess.com)
The Foxboro Co. (www.foxboro.com)
Honeywell Inc. (www.honeywell.com)
Marsh Instrument Co. (www.marshbellofram.com)
Marshalltown Instruments Inc. (www.marshbellofram.com)
Mensor Corp. (quartz helix) (www.e-pressure.com)
Mid-West Instrument (www.midwestinstrument.com)
Moeller Instrument Co. (www.moellerinstrument.com)
Moore Products, now part of Siemens Inc. (www.sea.siemens.com)
Noshok Inc. (www.noshok.com)
OCI Instruments Inc. (www.ociinstruments.com)
Palmer Instruments Inc. (www.palmerinstruments.com)
Perma-Cal Corp. (www.perma-cal.com)

731
© 2003 by Béla Lipták
732 Pressure Measurement

Ruska Instrument Corp. (quartz helix) (www.ruska.com)


Rosemount Inc., a Div. of Emerson Process Management (www.emersonprocess.com)
Span Instruments Inc.
Techscience International
Texas Instruments (quartz helix)
H.O. Trerice Co. (www.hotrerice.com)
Validyne Engineering Corp. (www.validyne.com)
Wallace & Tiernan Inc. (www.usfwt.com)
Weiss Instruments Inc. (www.weissinstruments.com)
Weksler Instruments Corp. (www.dresserinstruments.com)
Wika Instrument Corp. (www.wika.com)
Yokogawa Corp. of America (www.yca.com)

In 1852, E. Bourdon patented a curved or twisted tube that 10 1000 100,000 PSI
if held and pressurized at its open end, produced movement 100 10,000
at its closed end (tip travel).
Flat Oval Pointed Oval

THE BOURDON TUBE

Bourdon tube-type pressure elements are used to detect Flat Arch = 0.2 in.
higher pressures (spans exceeding 30 PSI, or 200 kPa),
FIG. 5.4b
because their spring gradient is insufficient for detecting Cross-sections of Bourdon tubes for various pressure ranges. (Cour-
lower pressures or vacuums. Bourdon tubes are manufactured tesy of WIKA Instrument Corporation.)
in C, helical, and spiral forms (see Fig. 5.4a). The C-Bourdon
element is made by winding a tubular element circularly to
form a segment of a circle, the helical element is made by strain-hardened alloys (brass, bronze, 304 and 316 stainless
winding the tube several turns into a helix, and the spiral is steel, or Monel), from heat-treated alloys (403, 4130, or 8630
formed by winding two or three turns in a spiral around the stainless steel), and from precipitation-hardened alloys (beryl-
same axis. In each of these cases the circular cross section lium copper, Inconel X, Ni-Span C, and K Monel). Some of the
of the tube is flattened. When these elements are pressurized, more desirable features of the Bourdon tube elements include
their cross section tends to become more circular, which tends their fast speed of response (usually 0.1 s full scale) and good
to straighten their shape (see Figure 5.4b). Helical and spiral sensitivity (0.01% of maximum pressure when unrestrained).
elements are used when it is desired to reduce the stress in Their limitations include a nonlinearity of about 0.5% full
the tube and to increase the tip movement. scale, which can be compensated mechanically or electroni-
The design of Bourdon tubes is based on empirical equa- cally; a 0.2 to 0.5% hysteresis on cycling pressures, which
tions and practical observations. The tubes can be made from can be minimized by heat treatment; and a sensitivity to
temperature variations, which can be reduced by the use of
Bourdon Tubes bimetallic compensators or by the use of element materials
(Ni-Span C) having moduluses of elasticity that are insensi-
tive to temperature variations.
C-Shaped Helical Spiral
S S S
C-Bourdon Pressure Sensors

Figure 5.4c illustrates a C-Bourdon tube used in direct indi-


P cating gauges, which usually have an arc of 270 degrees. The
P
P process pressure is connected to the tube support end of the
tube, while the tip end is sealed. Because of the difference
between inside and outside radii, the Bourdon tube presents
different areas to pressure, which causes the tube to tend to
straighten when pressure is applied.
The resulting tip motion is nonlinear because less motion
results from each increment of additional pressure. This non-
linear motion has to be converted to linear rotational pointer
response. This is done mechanically by means of a geared
FIG. 5.4a sector and pinion movement. The tip motion is transferred to
The main types of flexible measuring elements. (Courtesy of WIKA the tail of the movement sector by the connector link. The
Instrument Corporation.) angle between the connecting link and the sector tail is called

© 2003 by Béla Lipták


5.4 Bourdon and Helical Pressure Sensors 733

A See Fig. 5.6e


Force Bar Fulcrum
Section A-A A Stop & Seal
Pivot
Pinion
Flexure

Movement a Connecting
Link Atm. Process
Sector
Pressure
Traveling
Socket Angle C -Bourdon

Process
Pressure FIG. 5.4e
Force balance C-Bourdon pressure sensor.
FIG. 5.4c
Pressure gauge with C-Bourdon pressure element. of these devices is a function of the Bourdon tube diameter,
design quality, and calibration procedures. It can vary from
the traveling angle. This angle changes with tip movement in ±0.1 to ±5% inaccuracy, with the majority of these units
a nonlinear fashion, compensating for the nonlinearity of the failing in the area of ±1% (Table 5.11f).
tip movement. It is designed to minimize backlash and pro- The quality of C-Bourdon measurement can be influ-
vide smooth roll-on and roll-off characteristics in the geared enced by several factors. The large overhang in the element
sector. It is also designed to provide pinion movement by makes it susceptible to shock or vibration. If a Bourdon
using fine pitch gears or by eliminating the gears altogether element has been calibrated on air and in the process it is
and using a cam sector that positions on a roller surface. filled with liquid, weight of the fluid in the overhang will
This design eliminates the gears that eventually wear out introduce an error. In some applications, the tubes will need
and reduces the play that occurs when teeth are worn. On to be flushed, for which purpose tube tip bleed valves are
gear and pinion designs, the operation has been improved by available. On liquid service, air is likely to be trapped in the
the use of nylon and Teflon materials. tube end, which acts as a cushion and results in sluggish
Table 5.4d lists Bourdon tube materials, noting some of performance. Bourdon designs are available with an internal
the important characteristics by the letters P (poor), F (fair), capillary tube that allows entrapped air to escape from the tip.
G (good) and also indicating the maximum pressure range Figure 5.4e illustrates the design that uses a C-Bourdon
that the Bourdon tube can detect. element in a force balance transmitter. (See Figure 5.6e for
As stated in Table 5.4d, Ni-Span C seamless tubing, a more detailed view of a force balance transmitter mecha-
drawn to the profile in Figure 5.4c, is the best choice for nism.) The pressure applied to the Bourdon tube tends to
Bourdon tube applications in all cases except those at very straighten it and move its tip to the left. The flexure transmits
high pressure or those that are extremely corrosive. Direct the resulting force to the lower end of the force bar.
indicators or motion balance transmitters using C-Bourdon Due to the force balance nature of the unit, the force bar
elements are available with spans from 0–15 PSI to 0–20,000 is constantly balanced; the sensing Bourdon does not move
PSI (0–100 kPa to 0–140 MPa) and can be used for positive, as long as the pressure sensed is within the range of the
negative, or compound pressure ranges, but the indication on instrument. If the range is exceeded, the lower end of the force
the vacuum side will not be accurate or sensitive. The accuracy bar moves to the left, where it is supported by a limit stop.

TABLE 5.4d
Pressure Ranges and Other Characteristics of the Different Bourdon Tube Materials (1 PSI =
6.9 kPa = 0.069 bars)
Tube Material Corrosion Spring Rate Temp. Coefficient Hysteresis Maximum Pressure
Phosphorus bronze P F P F 800 PSIG
Beryllium copper P G P G 5000 PSIG
316 Stainless steel G P P P 10,000 PSIG
403 Stainless steel G P P P 20,000 PSIG
Ni-Span C G G G G 12,000 PSIG
K-Monel G P P P 20,000 PSIG

P = poor; F = fair; G = good.

© 2003 by Béla Lipták


734 Pressure Measurement

Feedback
Process Nozzle
Bellows
Pressure Flapper

Restriction

Air Vent
Output
Supply
FIG. 5.4f Process Diaphragm
Pressure
Spiral Bourdon element.

The Bourdon tube is made of Ni-Span C, and all other wetted FIG. 5.4g
components are stainless steel. Available spans vary from Motion balance pressure transmitter with spiral element.
1000 to 12,000 PSI (6.9 to 83 MPa) with 150% overrange
protection. The inaccuracy of this transmitter is ±0.5% in the
lower and ±1% in the higher ranges. Maximum ambient
Process
temperature is limited to 250°F (120°C).
Pressure

Spiral Bourdon Pressure Sensors Moving


Tip
The free end motion of the C-Bourdon tube is insufficient to
operate some of the motion balance devices, such as the FIG. 5.4h
transmitters. The spiral element shown in Figure 5.4f is Helical Bourdon pressure sensor.
essentially a series of C-Bourdon tubes joined end to end.
When pressure is applied, this flat spiral tends to uncoil and position. For each value of process pressure there is a corre-
produces a greater movement of the free end, requiring no sponding definite flapper position and output pressure, hence,
mechanical amplification. This increases the sensitivity and the name motion balance. The inaccuracy of these motion
accuracy of the instrument because no lost motion or friction balance pressure transmitters is generally placed at ±0.5%.
is introduced through the links and levers. As shown in Figure 5.4g, they will read the process pressure
Standard spiral materials include bronze, steel, stainless even if the air supply is lost.
steel, beryllium copper, Monel, and Ni-Span C. Spans as low
as 10 PSI (69 kPa) are available for positive, negative, or
Helical Bourdon Pressure Sensors
compound ranges. The various manufacturers use different
break points between the spiral and the helical elements. Figure 5.4h shows the construction of a helical Bourdon ele-
Some start using helical elements at above 200 PSI (1.4 MPa) ment. This sensor produces an even greater motion of the
spans; others stay with the spiral element up to 4,000 PSI free end than the spiral element, eliminating the need for
(27 MPa) and higher. Unflattened, heavy-wall, spiral tube mechanical amplification.
springs that detect pressures up to 100,000 PSIG (690 MPa) Other advantages of this design include the high over-
are also available in direct indicators. range protection available. For example, a 0 to 1,000 PSIG
Figure 5.4g is a simplified illustration of an old, pneu- (0 to 6.9 MPa) element may be safely exposed to 10,000
matic motion balance transmitter operated by a spiral ele- PSIG (69 MPa) pressure, and it is suitable for pressure mea-
ment. While the flapper is substituted with an electronic surement on continuously fluctuating services. The range of
motion sensor in the majority of today’s pressure transmitters, the helical coil is affected by the diameter, wall thickness,
the nature of the sensor’s operation has not changed. The number of coils used, and construction materials. High-pressure
sketch would also apply if the process pressure was detected elements might have as many as 20 coils, while low-span
by a C-Bourdon or a helical element. sensors can have only 2 or 3 coils.
In this unit, the air supply passes through a restriction Available materials include bronze, beryllium copper, Ni-
before being applied to the top of the relay diaphragm and Span C, and stainless steel. The various suppliers have dif-
the nozzle. An increase in process pressure tends to straighten ferent practices concerning the pressure levels at which they
out the spiral, which causes the flapper to move closer to the change from spiral to helical elements. For this reason, helical
nozzle. This increases the nozzle back pressure sensed by the elements are available with spans from 0–30 PSI to 0–80,000
relay diaphragm, which will move down, opening up the air PSIG (0–200 kPa to 0–550 MPa), depending on the supplier.
supply to the output. The increased output pressure is felt by For pressures exceeding 5,000 PSIG (35 MPa), heavy-duty
the feedback bellows and restores the flapper to its throttling stainless steel is the standard element material.

© 2003 by Béla Lipták


5.4 Bourdon and Helical Pressure Sensors 735

Process Its resolution and sensitivity are high while its hysteresis
Pressure is negligible. It is presently used in laboratories, but the
principle should make it suitable for industrial installations
in which precision pressure detection is desired. The cost is
Fused Quartz
high and the unit is sensitive to, and thus should be protected
Helical Element from, vibration. One of its drawbacks is the slow response
Fused Quartz speed of 2 min for full-scale travel.
Mirror Photo Cells
Lamp
Bibliography

American National Standard for Gauges: Pressure Indicating Dial Type:


Gear
Elastic Element, ASME B40.1, New York, 1991.
Digital Babichev, G.G., Kozlovskiy, S.I., Romanov, V.A., and Sharan, N.N., “Pressure
Counter Transducers with Frequency Output on the Base of Strain-Sensitive
Helical
Unijunction Transistors,” Paper 2.31, 1st IEEE International Confer-
Element &
ence on Sensors (IEEE Sensors 2002), Orlando, FL, June 2002.
Mirror
Bailey, S.J., “Pressure Sensors and Transmitters Affected by Technological
Change,” Control Engineering, January 1984.
von Beckerath, A., Eberlein, A., Julien, H., Kerstein, P., and Kreutzer, J.,
Balancing WIKA Handbook on Pressure and Temperature Measurement, U.S. ed.,
Lamp Lawrenceville, GA: Wika Instrument Corp., 1998.
Motor
“Bourdon Pressure Gauges,” Measurements and Control, December 1991.
Photo Cells
Brombacher, W.G., “40 Years of Pressure Measurement,” Instruments and
Control Systems, September 1967.
FIG. 5.4i
Buckon, L., “Considerations in Selecting a Pressure Calibration Device,”
Fused quartz helical pressure sensor.
Paper #91–0449, Instrumentation, Systems, and Automation Society
Conference, Toronto, 1991.
Helical coils can also be used as the element in differential Buffenmyer, W.L., “Selecting Bourdon-Tube Gauges,” Instruments and
pressure sensors if one of the pressures is acting on the outside Control Systems, February 1961.
surface and the other on the inside of the coil. Elliott, T.C., “Temperature, Pressure, Level, Flow: Key Measurements in
Power and Process,” Power, September 1975.
Motion balance instruments with helical elements include
Faust, E., “Pressure Calibration Standards: How Accurate Should They Be?”
direct indicators, recorders, controllers, switches, and transmit- Hydrocarbon Processing, January 1994.
ters with either pneumatic or electronic outputs. Their inaccu- “Gas Purged D/P Transmitters,” InTech, November 1992.
racy is normally in the area of ±1% of span. Giacobbe, J.B., “Selecting and Working Bourdon-Tube Materials,” Instru-
ment Manufacturing, July 1952.
Fused Quartz Helix Sensors As shown in Figure 5.4i, the Giitein, K., “A Dimensional Analysis Approach to Bourdon Tube Design,”
process pressure causes the free end of the helix to rotate Instrumentation Practice, September 1952.
together with the mirror attached to it. The quartz helix is Gillum, D., “Industrial Pressure, Level, and Density Measurement,”
Research Triangle Park, NC: ISA Press, 1995.
mounted above a large precision gear that can rotate concentri-
Hall, J., “Monitoring Pressure with Newer Technologies,” Instruments and
cally around the pressure sensor. A lamp and a pair of photo Control Systems, April 1979.
cells are attached to the gear. The light from the lamp is reflected Hughes, T.A., “Pressure Measurement,” EMC series, downloadable PDF,
from the mirror and distributed equally on the two photo cells. Instrumentation, Systems, and Automation Society, Research Triangle
When the process pressure changes, the free end of the helix Park, NC, 2002.
rotates the mirror, unbalancing the photo cells. This signal “Improved Online Calibration,” InTech, December 07, 2001.
Instrumentation, Systems, and Automation Society, “Industrial Measure-
causes the balancing motor to turn the gear until the light falling
ment Series: Pressure” (video VHS, PAL & NTSC), Research Triangle
on the photo cells is equally distributed once more. Conse- Park, NC, 2002.
quently, the gear follows the movement of the free end of the Instrumentation, Systems, and Automation Society, “Pressure: Indicators
helix without mechanical contact. The rotation of the gear is and Transmitters,” CD-ROM, Research Triangle Park, NC, 2002.
registered by a digital counter that is calibrated in pressure units. Johnson, D., “Pressure Sensing: It’s Everywhere,” Control Engineering,
This instrument detects vacuum, absolute, gauge, and April 2001.
−3 Marrano, S.J., “How to Choose and Apply Pressure Transmitters,” Control,
differential pressures between 1 µm (10 mmHg) and 500
March 2000.
PSIG (3.5 MPa) in several ranges with a minimum span of “Pressure Systems,” Chemical Engineering, January 29, 1968.
5 PSI (35 kPa). The device is repeatable to two parts in “Step-Shaped Diaphragm Seal Reduces Temperature Error,” Control, March
100,000 or inaccurate to ±0.01%. 2000.

© 2003 by Béla Lipták

You might also like