You are on page 1of 4

1, PM2.

5, PM10 output at the same time


2, unique dual-band data acquisition and automatic calibration technology, high consistency
3, the use of advanced laser anti-attenuation technology to ensure long-term stability of equipment
4, the device is 10-30V DC wide voltage Power supply, standard Modbus-RTU protocol

DC power supply (default) :10-30VDC


Transmitter circuit Operating temperature: -20C-60C, 0% RH-80% RH
Communication interface:

 485 communication (modbus) protocol baud rate: 2400.4800 (default) .9600


 Data bit length: 8 bits
 Parity mode: none
 Stop bit length: 1 bit
 Default modbus communication address: 1
 Support function code: 03
Parameter settings: Configured with the supplied configuration software via the 485 interface
Resolution: 0.1ug / cubic meter
Measuring range: PM2.5 / PM10 0-1000ug / cubic meter
Response speed: <= 90S
Warm-up time: <= 2min
Wiring instructions
Cable color Description
Power Brown Power+ (10-30vdc )
black Power -
yellow 485-A
Communication
Blue 485-B
Communication protocol
Basic communication parameters
Code 8-bit binary
Data bit 8 bit
Parity bit NO
Stop bit 1 bit
Error check CRC
Baud rate 2400bit/s, 4800bit/s, 9600 bit/s default is 4800bit/S
4.2 Data frame format definition
Adopt Modbus-RTU communication protocol, the format is as follows:
Initial structure ≥ 4 bytes of time
Address code = 1 byte
Noise Transmitter Instruction Manual (Type 485) V2.0
Function code = 1 byte
Data area = N bytes
Error check = 16-bit CRC code
End structure ≥ 4 bytes of time
Address code: is the address of the transmitter, which is unique in the communication network (factory default
0x01).
Function code: The instruction function of the command sent by the host. This transmitter only uses function
code 0x03 (read register data).
Data area: The data area is the specific communication data. Note that the 16bits data high byte is in front!
CRC code: Two-byte check code.
Host inquiry frame structure:
Register start Check code
Address code Function code Register length check code low
address high
1 bytes 1 bytes 2 bytes 2 bytes 1 bytes 1 bytes

Slave transponder frame structure


Function No 1 Data NO2 Data No N Data
Address code effective byte Check code
code area area area
1 bytes 1 bytes 1 bytes 2 bytes 2 bytes 2 bytes 2 bytes

Register address

Register address PLC or configuration


content operation
address
PM2.5
0000 H 40001 Upload data is the real Read only
value
PM10
0000 H 40002 Upload data is the real Read only
value

4.4 Communication protocol examples and explanations


Example: Read the PM value of device address 0x01

Inquiry frame:
Check code
Address code Function code Start address Data length check code low
high
0x01 0x03 0x00 0x00 0x00 0x02 0xC4 0x0B

Response frame: (eg read 85ug/m3 for PM2.5 and 156ug/m3 for PM10)

Function check Check code


Address code effective byte humidity temperature
code code low high
0x01 0x03 0x04 0x00 0x55 0x00 0x9c 0xEA 0x4A

PM2.5 calculation:
Air quality (PM2.5/P10) transmitter instruction manual (type 485) V2.0
PM2.5:0055 H (hexadecimal) = 85=> PM2.5 = 85ug/m3
PM10 calculation:
PM10:009CH (hexadecimal) = 156 => PM10 = 156ug/m3