The first chapter Function Description&Specification
1 component
The instrument is composed by two parts :the host (controller) and the sensor, it is connected by 5
core shielded cable between them
Display time
3 password protection
4 calibration method
The level calibration method: set 0 distance between the bottom of sensor and level.
Provides a set of 4-20mA output (0-20mA reservations), 16 bit high precision D/A, up to 750 ohms
load.
4mA and 20mA current point is set automatically by the user according to their demand.
6 relay outputs
(1) pulse output relay (K1): when total value reached pulsed flow accumulated value, closed 1
times (50ms)
2
(3) low alarm relay (K3)
(4) the backup relay (K4):design according to the customer request when the need
7 communication
8 historical records
Specification
1 flow range: 0 l/s to 100 m3/ s (determined by species and specifications for weir groove matched)
3 flow accuracy: 3%
6 level resolution: 1 mm
Pulse output relay (K1): total value reached pulsed flow accumulated value, closed 1 times (50ms)
Backup relay (K4): design according to the customer request when the need.
To avoid the corrosion of liquid gas, or the need for isolation protection.
2 sensor installation
Between the sensor and the controller comes standard with a 10 meters 5-core cable connection,
according to the corresponding color controller connected to the corresponding cable.
4
3 wiring
Liquid level
Symbol of receiving the echo Symbol of relay
signal
2 key
Changing display content: in the main screen interface, pressing the key to enter the password
interface in the user’s parameters interface, clicking the button to switch to the next parameter
interface.
Pressing the button to enter the set state if Need to enter your password or to modify the
parameters, at this time there will be a flashing cursor, then press the button, the flashing cursor
digit add '1' changing.
The cursor moves to the next digit,both and together could modify the parameters or
input the password.
5
Press this button to confirm after modifying the parameters (there will not be a cursor blinking)
after confirmation, press the button again to return to the main display interface
6
The fourth chapter Parameter Interface and Instructions
Max_Dist:
refers to distance from the zero water level to sensor. The definition of zero water level is below.
After the completion of the installation,please enter the correct empty distance here.
Blink: refers to the distance which could not be measured from the bottom of the sensor which is
decided by the characteristics of the sensor. And which has been set when out of factory. Usually be
30-50 cm.
Disp: can be set to the distance displaying the sensor to the surface and the reference display of
signal sampling.
00 user define Parshall flume, placement C and N value according to the actual.
7
01~25 standard Parshall flume (see Appendix one tank code & size.).
26 V shaped groove
There will be corresponding tips for the position of the model of the current selection of groove
according to this set code.
Code selection:
26 Triangular Weir
27 Rectanglar Weir X.XX m , X.XX says the settings of the weir width B.
Damp: damping coefficient calculation: input digital '0-9' instrument, “0- “ means without damping,
“9-” means the maximum damping coefficient.
Rate: the maximum change value allowed per level measurement, unit: mm. The value is smaller,
the anti interference is better, but the response speed is slower.
Level_T:starting effective level and the highest effective level of the measurement of instantaneous
flow, when the liquid level is less than the effective initial level value, instantaneous flow is as 0,
8
when the liquid level is greater than the highest effective level value, then the corresponding flow to
the highest effective level values is as the instantaneous flow. Unit:mm.
K1 setting:
T (alarm delay)
C (clearing the total cumulative flow, namely clearing the cumulative value on the main display )
T value is 0, when the flow accumulates to value A, the relay K1 is pull one time (50 ms).
T value is not 0, when the flow accumulates value A, the relay K1 will disconnect after pulls and
delays T seconds , and re accumulate the flow .
C value for the 'No' means not clearing the cumulative flow value; for 'Yes' means measuring
cumulative flow and clearing 0.
When the flow reaches 1.20m3, K1 pull, 10 seconds later the K1 reply, and clearing 0 for total
cumulative flow.
K2_High: control relay setting. Pull in point must be greater than or equal to separate point. Unit:
meters.
K3_Low: control of relay setting. Pull in point must be less than or equal to separate point. Unit is
meters.
Test: if you do not wish to produce cumulative flow in the installation test, it could be set to 01.
normal operation is 0.
Q_ unit: only setting the instantaneous flow unit; can be set to '0-4';
9
2-L/min (liters / minute);
3- m3/min (m 3/minute)
4- m3/S (m 3/ seconds);
Flow_sp: relative to the instantaneous flow, input value = actual value/display value. General set
up well out of factory,unnecessary modify it.
Time setting
Time_sev: time error correction, each additional 1, additional 1 second for daily time.
24mA_err:set to “Yes”, when the liquid level exceeds the maximum value of "effective level", the
output current is 24MA, set to “No”, the parameter is invalid .
To find the hours flow, daily flow, monthly flow, annual flow according to , in the query traffic
record condition, you could find the forward no displayed record according to .
Hour flow : hour cumulative flow per hour, can store 128 pcs.
In the main interface state, press the key TWO times to enter the querying interface, and then
press the key to find the circulating query hourly flow record.
Daily flow: daily flow display cumulative daily flow, can store 64 pcs.
After entering the querying hour flow interface, if press again,you could enter the querying
interface on daily flow, then press to find the circulating query daily flow records.
10
Monthly flow: monthly flow displays cumulative flow per month, can store 32 pcs.
After entering the the daily flow querying interface, press again to enter monthly interface,
then press the key to find the circulating query monthly flow records.
The annual flow: the annual flow display cumulative flow of every year, can store 4 pcs.
After Entering the query interface in the monthly flow, press again, to enter the query
interface to view the annual flow.
2 measuring sensors to 0 level distance, this value is set into the empty distance value. (this
parameter must be accurate, directly affecting the level of accuracy of measurement. )
3 To choose of the corresponding 'sink' code according to actual condition, if it is the Parshall flume
can also be set to '00' code and then set into the C and N values of Parshall flume. (set correctly
should display the correct level and flow value)
4 Set up the Current parameter according to the actual needs, 485 interface parameters, or relay
parameter. (unnecessary to mention it.)
5 To inspect parameters "debugging"if is' 1 ' or not, if is' 1 ', pls set it into' 0'. If the parameter is set
to '0', the cumulative flow will be increased, otherwise the cumulative flow will remain unchanged.
Select the type of measuring weir groove, need to consider the flow in the channel, the flow state of
the water in the channel, and whether form the free flow or not. If the flow rate is less than 40
liters/sec, to choose the angled triangle weir in general. If the flow rate is more than 40 liters/sec, to
choose the Parshall cell in general. If the flow rate is more than 40 liters/sec,and the water level in
channels is large at the same time, you could use the rectangular weir.
After the installation of the weir plate, there will form a "small waterfall" in the weir downstream.
Weir groove could be made from glass steel, and processing size should be accurate. The key of
Triangular weir and rectangular weir is the size, and more exactly. The side surface towards the
entering water should be smooth. Parshall groove size must be accurate,and the interior surface
should be smooth. The key of throat is the size, and more accurate.
11
1, Right angled triangle weir
Below is the processing map of the right angled triangle weir. Water level and flow correspondence
such as the below table. Note that the upstream channel width of the installation right angled
triangle weir is 600mm, the height of triangle vertex angle and the upper reaches canal bottom is
250mm. Water level--flow list has existed in the weir plate of the flow meter.
Table 1#: right angled triangle weir flow stage corresponding table level unit: m flow unit: L / S
level 0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09
flow 0 0.014 0.080 0.221 0.454 0.794 1.243 1.828 2.552 3.426
level 0.1 0.11 0.12 0.13 0.14 0.15 0.16 0.17 0.18 0.19
flow 4.459 5.618 6.984 8.531 10.267 12.200 14.234 16.563 19.107 21.873
level 0.2 0.21 0.22 0.23 0.24 0.25 0.26 0.27 0.28 0.29
12
The water level of triangular weir in the channels of the installation and triangular weir zero
Triangular weir is positioned on the channel as above. Weir plate is to the vertical,and is to be
installed in the channel axis. When machining triangular weir, may make the vertex into a round,
pay attention to the equal zero position in determining the water level. Node level triangular weir
zeros should on the crossover point in triangle weir side extend line. The instrument probe should
be installed in the upstream position 0.5m ~ 1m distance to the weir plate.
2, Rectangular weir
Rectangular weir can be processed according the following drawing. Water level and flow
corresponding relationship as shown in table 2、3、4 and 5. Relationship between water level and
flow of rectangular weir mainly depends on the weir width "b". Also is relative with the upstream
channel width "B" and the weir height "P". Below are 4 commonly types of rectangular weir, and
could satisfy with most of the application places generally.
13
Table 2#: 0.25 m rectangular weir level flow corresponding table level unit: m flow unit: L / S
level 0.000 0.010 0.020 0.030 0.040 0.050 0.060 0.070 0.080 0.090
flow 0.0000 0.4376 1.2397 2.2812 3.5181 4.9250 6.4849 8.1855 10.018 11.973
level 0.100 0.110 0.120 0.130 0.140 0.150 0.160 0.170 0.180 0.190
flow 14.047 16.232 18.526 20.924 23.423 26.020 28.712 31.497 34.373 37.338
level 0.200 0.210 0.220 0.230 0.240 0.250 0.260 0.270 0.280 0.290
flow 40.390 43.527 46.749 50.054 53.441 56.907 - - - -
Table 3#: 0.5 m rectangular weir level flow corresponding table level unit:m flow unit: L / S
level 0.000 0.010 0.020 0.030 0.040 0.050 0.060 0.070 0.080 0.090
flow 0.000 0.877 2.485 4.575 7.058 9.883 13.018 16.437 20.122 24.058
level 0.100 0.110 0.120 0.130 0.140 0.150 0.160 0.170 0.180 0.190
flow 28.233 32.637 37.261 42.097 47.140 52.382 57.820 63.449 69.264 75.262
level 0.200 0.210 0.220 0.230 0.240 0.250 0.260 0.270 0.280 0.290
flow 81.440 87.794 94.322 101.021 107.888 114.922 122.121 129.482 137.004 144.686
level 0.300 0.310 0.320 0.330 0.340 0.350 0.360 0.370 0.380 0.390
flow 152.524 - - - - - - - - -
14
Table 4#: 0.75 m rectangular weir level flow corresponding table level unit: m flow unit: L / S
level 0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16 0.18
flow 0 3.748 10.668 19.718 30.544 42.945 56.793 71.997 88.487 106.21
level 0.2 0.22 0.24 0.26 0.28 0.3 0.32 0.34 0.36 0.38
flow 125.13 145.21 166.41 188.73 212.13 236.61 262.14 288.72 316.34 344.98
level 0.40 0.42 0.44 0.46 0.48 0.50 0.52 0.54 0.56 0.58
Table 5#: 1 m rectangular weir level flow corresponding table level unit: m flow unit: L / S
level 0.000 0.020 0.040 0.060 0.080 0.100 0.120 0.140 0.160 0.180
flow 0.0000 4.9780 14.141 26.092 40.345 56.626 74.756 94.605 116.08 139.09
level 0.200 0.220 0.240 0.260 0.280 0.300 0.320 0.340 0.360 0.380
flow 163.59 189.53 216.85 245.54 275.54 306.85 339.43 373.26 408.33 444.63
level 0.400 0.420 0.440 0.460 0.480 0.500 0.520 0.540 0.560 0.580
flow 482.13 520.83 560.71 601.76 643.98 687.36 - - - -
Rectangular weir is positioned on the channel as shown below. Weir plate to the vertical, to be
installed on the channel axis. The instrument probe should be installed in the position 0.5m ~ 1 m
away from the weir plate.
3, Parshall flume
The structure of the Parshall groove is the following figure. The Marked size of it is the throat width
"b".First of all, according to the maximum flow application needs, from "appendix 2#, tank water
level-flow formula" to find the appropriate throat width "b" for Parshall flume. Again from
“appendix 1#, groove structure size" to find out other sizes equal to the" b " corresponding throat
width. Such as "L", "N", "B1" and "L1" etc.
15
The relationship between the Parshall flume water level and flow is usually like formula: Q=C*han .
According to the throat width "b", from "appendix 2#, tank water level - flow formula" to find out
the corresponding formula. Point to point to enter into the water level value, calculate the
corresponding flow.
The instruments have been pre stored the tank level flow meter list. Table 6# and 7# are discharge
relation of throat width 0.025m、0.051m for water level.
Table 6#: throat 0.025 m Parshall flume water level&flow corresponding table level unit: m flow unit: L / S
level 0.000 0.010 0.020 0.030 0.040 0.050 0.060 0.070 0.080 0.090
flow 0.0000 0.0480 0.1405 0.2634 0.4114 0.5814 0.7712 0.9794 1.2046 1.4458
level 0.100 0.110 0.120 0.130 0.140 0.150 0.160 0.170 0.180 0.190
flow 1.7023 1.9733 2.2582 2.5565 2.8677 3.1914 3.5271 3.8747 4.2336 4.6037
level 0.200 0.210 0.220 0.230 0.240 0.250 0.260 0.270 0.280 0.290
flow 4.9846 5.3762 5.3762 5.3762 5.3762 5.3762 - - - -
16
Table 7#:throat 0.051 m Parshall flume water level flow corresponding table level unit: m flow unit: L / S
level 0.000 0.010 0.020 0.030 0.040 0.050 0.060 0.070 0.080 0.090
flow 0.0000 0.0959 0.2807 0.5263 0.8221 1.1617 1.5411 1.9571 2.4071 2.8892
level 0.100 0.110 0.120 0.130 0.140 0.150 0.160 0.170 0.180 0.190
flow 3.4018 3.9434 4.5127 5.1088 5.7307 6.3775 7.0484 7.7429 8.4602 9.1997
level 0.200 0.210 0.220 0.230 0.240 0.250 0.260 0.270 0.280 0.290
flow 9.9610 10.744 11.547 12.370 13.214 13.214 - - - -
17
18
Appendix 1#: sink code table and Parshall flume structure size
A 3 0.076 0.152 0.057 0.259 0.457 0.305 0.178 0.305 0.025 0.46
L 4 0.152 0.305 0.114 0.400 0.610 0.407 0.394 0.610 0.076 0.61
L 5 0.228 0.305 0.114 0.575 0.864 0.576 0.381 0.457 0.076 0.77
6 0.25 0.60 0.23 0.78 1.325 0.883 0.55 0.92 0.08 0.80
7 0.30 0.60 0.23 0.84 1.350 0.902 0.60 0.92 0.08 0.95
S 8 0.45 0.60 0.23 1.02 1.425 0.948 0.75 0.92 0.08 0.95
T 9 0.60 0.60 0.23 1.20 1.500 1.0 0.90 0.92 0.08 0.95
A 10 0.75 0.60 0.23 1.38 1.575 1.053 1.05 0.92 0.08 0.95
N 11 0.90 0.60 0.23 1.56 1.650 1.099 1.20 0.92 0.08 0.95
D 12 1.00 0.60 0.23 1.68 1.705 1.139 1.30 0.92 0.08 1.0
A 13 1.20 0.60 0.23 1.92 1.800 1.203 1.50 0.92 0.08 1.0
R 14 1.50 0.60 0.23 2.28 1.95 1.303 1.80 0.92 0.08 1.0
D 15 1.80 0.60 0.23 2.64 2.10 1.399 2.10 0.92 0.08 1.0
16 2.10 0.60 0.23 3.00 2.25 1.504 2.40 0.92 0.08 1.0
17 2.40 0.60 0.23 3.36 2.40 1.604 2.70 0.92 0.08 1.0
18 3.05 0.91 0.343 4.76 4.27 1.794 3.68 1.83 0.152 1.22
19 3.66 0.91 0.343 5.61 4.88 1.991 4.47 2.44 0.152 1.52
L
20 4.57 1.22 0.457 7.62 7.62 2.295 5.59 3.05 0.229 1.83
A
21 6.10 1.83 0.686 9.14 7.62 2.785 7.32 3.66 0.305 2.13
R
22 7.62 1.83 0.686 10.67 7.62 3.383 8.94 3.96 0.305 2.13
G
23 9.14 1.83 0.686 12.31 7.93 3.785 10.57 4.27 0.305 2.13
E
24 12.19 1.83 0.686 15.48 8.23 4.785 13.82 4.88 0.305 2.13
25 15.24 1.83 0.686 18.53 8.23 5.776 17.27 6.10 0.305 2.13
19
Appendix 2#: The formula for water level and flow of Parshall flume
Cate Serial Throat Flow formula Water level Flow range critical
gory width Q=Chan range h(m) Q(L/S) submergence
b(m) (L/S) Min Max Min Max degree %
20