Uin mV 0 100 150 200 250 300 350 400 450

f-masurat Hz 0 991.00 1485.00 1982.00 2478.00 2976.00 3474.00 3970.00 4470.00


f-teoretic Hz 0 1000 1500 2000 2500 3000 3500 4000 4500
ε-f % 0 -0.09 -0.15 -0.18 -0.22 -0.24 -0.26 -0.3 -0.3
Uout-
masurat mV 0 99.70 149.20 198.98 248.70 298.60 348.40 398.60 448.20
Uout-teoretic mV 0 100 150 200 250 300 350 400 450
ε-u % 0 -0.03 -0.08 -0.10 -0.13 -0.14 -0.16 -0.14 -0.18
500 550 600 650 700 750 800 850 900
4969.0 5469.5 5969.9 6470.8 6971.5 7473.60 7974.20 8476.80 8977.60
0 0 0 0 0
5500 6000 6500 7000 7500 8000 8500 9000
5000
-0.31 -0.305 -0.301 -0.292 -0.285 -0.264 -0.258 -0.232 -0.224
548.21 598.30 649.40 698.50 748.73 798.81 849.08 899.16
498.20
550 600 650 700 750 800 850 900
500
-0.18 -0.18 -0.17 -0.06 -0.15 -0.13 -0.12 -0.09 -0.08
10000
9000
8000
7000
F-masurat[hz]

6000
5000
4000
3000
2000
1000
0
0 100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900
U-in[V]

10000
9000
8000
7000
F-teoretic[hz]

6000
5000
4000
3000
2000
1000
0
0 100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900
U-in[V]

0
0 100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900
-0.05

-0.1

-0.15
ε-f

-0.2

-0.25

-0.3

-0.35
U-in[V]
1000
900
800
700
Uout-masurat[V]

600
500
400
300
200
100
0
0 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00
10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

f-teoretic[hz]

1000
900
800
700
600
Uout-teoretic[V]

500
400
300
200
100
0
0 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00
10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90

f-teoretic[hz]

0.00
00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00
10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90
-0.02
-0.04
-0.06
-0.08
ε-u

-0.10
-0.12
-0.14
-0.16
-0.18
-0.20
f-teoretic[hz]
Concluzi: Graficele valorilor teoretice si graficele valorilor masurata sunt asemanatoare (diferentele
fiind nesemnificative).

Senzor de temperatura cu fibre optice cu cristal fluorescent de rubin

 [C] 0 10 20 30 40 50 60 70
 [ms] 5.345 5.277 5.215 5.144 5.068 4.976 4.878 4.768

Senzor de temperatura cu fibre optice cu cristal fluorescent de LiCAF

 [C] 22 66 102 141 168 204 231 254


 [s] 140.92 140.91 140.81 138.49 127.57 78.14 33.31 14.28
Rubin

5.4
5.3
5.2
5.1
5
t [ms]

4.9
4.8
4.7
4.6
4.5
4.4
0 10 20 30 40 50 60 70
q [°C]

LiCAf
160

140

120

100
t [ms]

80

60

40

20

0
22 66 102 141 168 204 231 254
q [°C]

Concluzii:LiCAF fata de rubin are timpul de raspuns mai rapid la temperatiri mai mari.

Regimul dinamic al unui termostat asociat unei diode laser.

0.8

0.7

0.6

0.5
U(q) [°C]

0.4

0.3

0.2

0.1

0
0.1 5.5 10.916.321.727.132.537.943.348.754.159.564.970.375.781.186.591.997.3
-0.1
t[s]
Nume:Ilie Ionut-Costin

Grupa:131IA

Labarator 3