Professional Documents
Culture Documents
1.Contact potential
When two kinds of homogeneous conductors A and B contact, due to the different density of
free electrons in a and B (set free electron density Na > NB), conductor a will diffuse free
electrons to conductor B through the contact, then a loses electrons, and B accumulates
electrons, so that potential difference is generated on both sides of the contact, and the
electrostatic field E is established, as shown in Figure 2-2, the left side of electrostatic
field E will prevent free electrons from continuing to diffuse.When the interaction of
diffusion force and electric field force is balanced.The diffusion of electrons stops
relatively, and the potential is finally generated between two sides of the contact. This
potential is called contact potential, which is represented by the symbol EAB (T), where t
is the temperature of the contact. The contact potential is related to the temperature of
the contact surface and the properties of the two conductors. The direction is shown in
Figure 2-2, from the electrode with small electronic density to the electrode with large
electronic density http://www.tksiliconerubbercable.com.
2.Thermoelectric potential
Because the free electron density of conductor will increase with the increase of
temperature, when the temperature of both ends of the same conductor is different (as shown
in Figure 2-3), the free electron density of the end with high temperature will be higher
than that of the end with low temperature, so there will be a similar spontaneous electron
diffusion process in contact potential between the two ends, and finally a potential
difference will be generated between the two ends of conductor to establish the potentialThe
potential is called thermoelectric potential, which is expressed by the symbol EA (T, to).
Its size is related to the temperature T, to and conductor properties at both ends of the
conductor, as shown in Figure 2-3, from the low temperature end to the high temperature
end.For the convenience of analysis, the thermoelectric potential is sometimes written in
the form of EA (T, to) = EA (T) - EA (to).
3.Thermal potential
To sum up, in the thermocouple circuit shown in Figure 2-1, when t > T0, Na > Nb, two
contact potentials (T) and EAB (to), two temperature difference potentials (T, to) and EB
(T, to) will be generated in the circuit.The direction of each potential is shown in the
figure.
At this time, the total potential of the loop, i.e. the thermal potential EAB (T, to), is
the algebraic sum of these contact potentials and the thermoelectric potential, i.e
EAB(t,to)= eAB(t) - eA(,to) - eAB(to) + eB(t,to)
= eAB(t)-[eA(t) - eA(to)] - eAB(to) + [eB(t) - eB(to)]
= [eAB(t) - eA(t) + eB(t)] - [eAB(tO) - eA(to) + eB(to)]
= fAB(t) – fAB(to)
Because the thermoelectric potential is smaller than the contact potential and T > to, the
contact potential EAB (T, to) accounts for the largest percentage in the total potential EAB
(T, to). Therefore, the direction of the total potential EAB (T, to) depends on the
direction of EAB (T).Because of the high electron density of a, a is the positive electrode
and B is the negative electrode. In the positive hot electrode, the direction of the
potential is from the hot end to the cold end.
The above formula shows that when the materials of the two thermoelectric poles are
selected, the thermoelectric potential is the difference between the two functions related
to the contact temperature respectively.If the cold end temperature to remains the same,
then FAB (to) = C (constant), then EAB (T, to) = FAB (T) - C, the thermoelectric potential
and the hot end temperature t have a one-to-one correspondence.Therefore, when the
thermoelectric potential EAB (T, to) is measured, the value of the measured temperature T
can be determined, which is the principle of thermocouple temperature measurement.
For the convenience of use, the corresponding relationship between the hot end temperature
and the thermoelectric potential of the standardized thermocouple can be found in the
function table.This kind of function table is made by experiment under the condition of cold
end temperature of 0 ℃, which is called thermocouple graduation table.The thermocouple
scale can be used to express the thermoelectric characteristics of thermocouples.See table
1-5 for the graduation table of several common thermocouples.It should be noted that when to
is not equal to 0 ℃, the index table cannot be used to directly check the value of EAB (T,
to) or EAB (T, to)
Attached table 1: graduation table of Pt Rh 10 Pt thermocouple (graduation s; reference end
temperature 0 ℃; MV)
Attached table 2: platinum rhodium 13 platinum thermocouple graduation table (graduation R;
reference end temperature 0 ℃; MV)
The thermoelectric Thermocouple is not matched with display More thermocouple or display
potential is larger instrument instrument to match
than the actual value
Compensation wire is not matched with Replace the compensation lead to
(the indicated value
thermocouple match
of the display
instrument is higher) There is DC interference signal entering Eliminate DC interference
Xue Fu Lian
The correct use of thermocouples can not only accurately get the temperature value, ensure the
product is qualified, but also save the material consumption of thermocouples, save money and ensure
the product quality.Incorrect installation, thermal conductivity and time lag are the main errors of
thermocouple in use.
For example, the installation position and insertion depth of thermocouples cannot reflect the
real temperature of furnace, in other words, thermocouples should not be installed too close to doors
and heating places, and the insertion depth should be at least 8-10 times of the diameter of the
protective tube; the gap between the protective sleeve and the wall of thermocouples is not filled
with insulation material, which causes the hot overflow or cold air intrusion in the furnace, so the
thermocouple protective tube and the wall holeThe gap between them shall be blocked with insulating
materials such as fire-resistant mud or asbestos rope to avoid the influence of cold and hot air
convection on the accuracy of temperature measurement; the cold end of thermocouple shall be too
close to the furnace body to make the temperature exceed 100 ℃; the installation of thermocouple
shall avoid strong magnetic field and strong electric field as far as possible, so the thermocouple
and power cable shall not be installed in the same conduit to avoid interference and error;
thermocouple cannot be installedIn the area where the measured medium rarely flows, when measuring
the gas temperature in the tube with thermocouple, the thermocouple must be installed against the
flow direction and fully contacted with the gas.
For example, if the thermocouple is insulated, the dirt or salt slag between the protective tube
and the stay wire plate are too much, resulting in poor insulation between the thermal dipole and the
furnace wall, which is more serious under high temperature. This will not only cause the loss of the
thermal potential, but also introduce interference, resulting in errors sometimes up to Baidu.
Because of the thermal inertia of the thermocouple, the indicating value of the instrument lags
behind the change of the measured temperature, especially in the rapid measurement.Therefore,
thermocouples with thin thermoelectric pole and small diameter of protective tube should be used as
far as possible.When the temperature measurement environment permits, the protective pipe can even be
taken away.Because of the measurement lag, the amplitude of temperature fluctuation detected by
thermocouple is smaller than that of furnace temperature fluctuation.The larger the measurement lag
is, the smaller the amplitude of thermocouple fluctuation is, and the greater the difference with the
actual furnace temperature is.When a thermocouple with a large time constant is used to measure or
control the temperature, although the temperature displayed by the instrument fluctuates very little,
the actual furnace temperature may fluctuate a lot.In order to measure temperature accurately,
thermocouple with small time constant should be selected.The time constant is inversely proportional
to the heat transfer coefficient and directly proportional to the diameter of the hot end of the
thermocouple, the density of the material and the specific heat. To reduce the time constant, in
addition to increasing the heat transfer coefficient, the most effective way is to minimize the size
of the hot end.In use, the protective casing with thin wall and small inner diameter is usually made
of materials with good thermal conductivity.In the more precise temperature measurement, the bare
wire thermocouple without protective sleeve is used, but the thermocouple is easy to be damaged, so
it should be corrected and replaced in time.
In case of high temperature, if there is a layer of coal ash on the protective tube and dust is
attached to it, the thermal resistance will increase and block the heat conduction. At this time, the
temperature indication is lower than the true value of the measured temperature.Therefore, the
outside of the thermocouple protection tube should be kept clean to reduce the error.
All the comrades who are engaged in temperature measurement know the working principle of
thermocouples - two different conductors form a closed circuit. When the temperature at both ends of
the circuit is equal, the thermoelectric potential of the circuit is zero.When the temperature of the
working end and the overstock end of the thermocouple is not equal, the heat generated in the rebel
loop is potential.The thermoelectric potential is related to the temperature difference between the
two ends.The free end temperature of the thermocouple is not a constant value.All manufacturers use
thermocouple special wire compensation wire.By extending the thermocouple to a constant temperature,
the thermoelectric potential of the thermocouple is compensated.
In the process of daily inspection, many electricians do not know the function of compensation wire,
only know that compensation wire is the special wire connecting thermocouple and instrument.It is
wrong to connect the compensation wire with thermocouple and instrument at will, which will cause
great error.Take a machining factory in our city as an example. At that time, the process of
workpiece required 920 ℃.The operator also set the value according to the process requirements, and
the instrument indication is normal.After the liquid fire, the workpiece appears striation, which
makes a furnace of workpiece scrapped.In the analysis of the accident, it is found that the
compensation wire is connected reversely. After correction, the temperature in the furnace is 973
℃.At that time, the free end temperature of the mass transformer couple was 45 ℃, and the indoor
temperature was 18 ℃.The temperature error caused by reverse connection is as follows:
=1.817mV =0.718mV
thermocouple compensation wire can cause considerable temperature error after being connected
reversely.Therefore, do not connect the compensation wire in reverse.
1. Principle of thermocouple temperature measurement
measured medium; the other is called cold end, also known as reference end or
Contact potential is the electromotive force formed at the contact due to the
conductors contact, the free electron diffuses from the conductor with high density
to the conductor with low density, and one side of the free electron loses positive
charge at the contact, and the other side of the free electron has negative charge.
When the diffusion reaches dynamic balance, a stable contact potential is formed at
both sides of the contact surface.The value of contact potential depends on the
properties of two different conductors and the temperature of the contact point.The
contact potentials AB (T) and ab (0) of the two contacts can be expressed asEET
Where: K - Boltzmann constant; - electric quantity per unit charge; NAT, and
at0, BT0 - electronic density of the two materials when the temperature is and 0,
respectively.ENBTNNTTA, B
Thermoelectric potential is a kind of electromotive force produced by the
temperature at both ends of the same conductor is different, the electronic energy
at the high temperature end is greater than that at the low temperature end, so the
number of electrons from the high temperature end to the low temperature end is
more than that from the low temperature end to the high temperature end. As a
result, the high temperature end is positively charged due to the loss of
electrons, and the low temperature end is negatively charged due to the gain of
redundant electrons. Therefore, the contact potential is formed at both ends of the
conductor, and its size is determined by the following factorsThe formula is:
Where, NAT and are the electronic density of conductor and conductor respectively,
EAB(T, T0)=eAB(T)+eB(T,T0)-eAB(T0)-eA(T,T0)
thermocouple is much smaller than that of the contact potential, which can be
This relation is very useful in practical measurement, that is, as long as the
temperature of two materials and two contacts, but not to the size, shape and
the circuit. Will the connecting wires and instruments affect the
third conductor is the same, there is no effect on the total thermal potential
of the circuit.
(T, TC) and (T, t0) of the thermocouple at the contact temperature of C and 0,
that is, EAB (T, t0) = EAB (T, TC) + EAB (TC, t0)TABT, tEABABT, tTCTEABEAB
circuit, the thermoelectric potential with the reference terminal temperature not 0
3、 Structure of thermocouple
connection, no fixed device and other forms according to the connection form
when it is installed.
The main advantages of armored thermocouple are small thermal capacity, fast
industrial sectors.
3Membrane thermocoupleThin film thermocouple is a kind of special thermocouple
made of two kinds of thin film thermoelectric materials, which are evaporated
contact of thin film thermocouple can be made very small (can be thin to 0.01-
0.1 μ m), has the characteristics of small thermal capacity, fast reaction
speed and so on. The thermal response time reaches the microsecond level, which
is suitable for the surface temperature of small area and the dynamic
temperature
related to the temperature of the hot end and the cold end.Therefore, only when
thermocouple has a single value function relationship with the hot end
temperature of 0 ℃, but in the actual use process, the cold end temperature is
often not 0 ℃, so the cold end temperature must be treated to eliminate the
Then, when the hot end temperature is, the relationship between the
obtained according to the intermediate temperature law as follows: (T, 0) = EAB (T,
away from the site, so that the cold end temperature 0 is relatively
compensation wire is used, which is usually made of two kinds of cheap metal
compensation wire and the matched thermocouple are required to have the same
thermoelectric characteristics.T
2. cold end temperature correction method: The cold end of the thermocouple can
the compensation wire, but as long as the cold end temperature 0 is not equal
corrected.When the working end temperature is, the relationship between the
It can be seen that the corrected value of the measured potential value
value to be measured can be found out from the graduation table.In the
worldEABEABEABTEABTT
furnace temperature.It is known that the cold end temperature is 0 = 30 ℃, and the
measured thermal potential (0) is 33.29mv, and the temperature of the heating
furnace is calculated.TEABT,T
+ EAB (T0, 0) = 33.29 + 1.203 = 34.493mv, and from the graduation table of Ni
and precise measurement, the cold end is usually put into a 0 ℃ thermostat or
a container filled with ice water mixture to keep the cold end temperature at 0
method): The compensation bridge method uses the unbalanced voltage generated
the change value of the thermal potential caused by the temperature of the cold
which are wound by manganese copper wire with smaller temperature coefficient
of resistance, and the resistances Cu which are wound by copper wire with
compensation bridge and the cold end of the thermocouple are in the same
ambient temperature. When the thermoelectric potential (T, t0) changes due to
the change of the cold end temperature, the resistance value of Cu changes with
the change of the cold end temperature. If the resistance of the bridge arm and
the bridge current are properly selected, the unbalanced voltage AB generated
by the bridge can compensate the thermoelectric potential change due to the
compensation.UabRRRREABRUT