Professional Documents
Culture Documents
Bradford1999 2
Bradford1999 2
Characterization of noise sources for two generations of computed radiography systems using powder and
crystalline photostimulable phosphors
Med. Phys. 34, 3345 (2007); 10.1118/1.2750973
Key words: computed radiography ~CR!, digital radiography, modulation transfer function ~MTF!,
noise power spectrum ~NPS!, detective quantum efficiency ~DQE!
27 Med. Phys. 26 „1…, January 1999 0094-2405/99/26„1…/27/11/$15.00 © 1999 Am. Assoc. Phys. Med. 27
28 Bradford, Peppler, and Dobbins III: Characteristic of a radiography system 28
TABLE I. Kodak screen type and sizes studied. slowscan directions for each screen type and size studied.
~Other publications have referred to these two directions by
Screen Type Screen size
various terms. The ‘‘fastscan’’ direction has been also
GP-25 35 cm343 cm termed the ‘‘laser scan’’ or ‘‘scan’’ direction. The ‘‘slows-
18 cm324 cm can’’ direction has been also described as the ‘‘plate scan’’
HR 18 cm324 cm
or ‘‘subscan’’ direction.! This was accomplished by attach-
ing a 10 mm slit camera ~Radiation Measurements, model:
07-624! to the front surface of the cassette with the side
studied. ~The study file contains the 12-bit logarithmic data, marked ‘‘focus side’’ facing the focal spot. The slit’s long
look-up tables, display images and masks.! axis was slightly angled ~;5°–8°! with respect to the central
The Kodak Model 400 reader is able to scan an area axes of the CR plate to provide adequate sampling for the
smaller than the actual screen size. This option allowed us to composite LSF. Shielding ~2 mm of lead! was placed around
study two different scan conditions with the use of one the slit camera to block any forward-directed scatter. The
screen. For example, to measure the system’s performance cassette/slit assembly was oriented vertically on the tabletop
for the 18 cm324 cm plate, a specific portion of the perpendicular to the edge of the table, and therefore, the
35 cm343 cm plate was exposed. When this larger plate was x-ray beam, using a square corner and positioned at an SDD
then scanned, the 18 cm324 cm option was selected. By us- of ;170 cm. The vertical orientation eliminated backscatter
ing one plate for two different sizes, effects from slight sen- since there was nothing behind the cassette. This simple cas-
sitivity variations in the phosphor from one screen to another sette alignment method was quite adequate and was con-
were removed.8 Unfortunately, this process could be done firmed by the consistency of the maxima of the line spread
for just the GP-25 screen type since only the smaller functions obtained. One at a time the HR (18 cm324 cm)
18 cm324 cm size was available for the HR screen. and GP-25 (35 cm343 cm) screens were exposed. After a
Before every measurement, each screen was flash erased ten min delay the screens were read and then erased. The
to remove any accumulated signals due to extraneous back- GP-25 screen was exposed a second time at this same dis-
ground radiation and/or the naturally occurring isotopes tance for the 18 cm324 cm GP-25 measurement. Determina-
present in the phosphor. After each screen exposure, there tion of the mAs factors were based on the goal to produce a
was a ten min delay before reading to minimize the effects of suitable LSF down to the 0.1% level.
latent image decay.9 The screens were cleaned prior to use. Due to the slit camera’s construction ~see Fig. 1!, the
The x-ray source used to expose each plate was a Phillips relief edges on each jaw ~4°! allowed a tolerance in the ac-
x-ray tube ~Model SRO 33/100! with a Phillips Maximus curacy of alignment. However, due to transmission through
C850 ‘‘Classic’’ ~three phase, twelve pulse! generator. For the beveled edges, the width of this slit was effectively larger
all plate exposures, the x-ray unit was set to 70 kVp and than the nominal 10 mm. To determine the effect of the bev-
confirmed within 2 kV by an RMI kVp meter ~Model 240A!. eled edges on the resulting MTF curves, repeat MTF mea-
Filtration of 0.5 mm Cu was applied during all measure- surements ~in both the fastscan and slowscan directions!
ments. The aluminum equivalent half-value layer for this fil- were conducted using a straight edge slit with the 100 mm/
tered spectrum was 7 mm. Exposure levels ~‘‘in air’’! were pixel HR Kodak plate. Comparisons between the MTF
measured with an ADCL ~Accredited Dosimetry Calibration curves made by the two slit types produced a correction fac-
Lab! calibrated exposure meter ~ionization chamber: Kei- tor that was applied to the less accurate MTF curves obtained
thley, Model 96035; electrometer: Keithley, Model 602!. with the beveled slit. The correction factor is described be-
The x-ray tube was aimed horizontally with the central axis low.
of the beam parallel to the edge of the x-ray table and the The greater availability of this commercial slit and its
cassette was mounted vertically on the table perpendicular to more forgiving alignment requirements ~as opposed to the
the beam using a square corner and the edge of the table. straight edge slit! will allow others to more easily reproduce
this MTF measurement method. Due to these advantages, we
chose to use the beveled slit for all of our MTF measure-
B. MTF measurements
The method of Fujita et al.6 as adapted by Dobbins et al.4
was applied to measure the presampling MTF for each plate
type and size. The line spread function ~LSF! was obtained
by acquiring images of a narrow slit in both the fastscan and
TABLE II. Image, pixel, and study file sizes for the two CR screen sizes
studied.
FIG. 6. Two-dimensional NPS for the Kodak GP-25 ~100 mm/pixel! plate at
~a! 0.03, ~b! 0.3, ~c! 3, and ~d! 30 mR. ~Scaling for each image has been
independently adjusted.!
relatively flat except for a few peaks that are most prominent
at the lowest exposure levels ~i.e., the 0.03 mR images!.
Because the noise peaks are most prominent at the low ex-
posure levels, they most likely represent a fixed pickup
noise, perhaps from a switching power supply that is used in
the unit. The KESPR reader electronics operate with the
same digitization rate ~i.e., the same analog bandwidth! for
both pixel sizes and adjust the plate and laser scan speeds to
achieve the different size pixels. Therefore, peaks that appear
at the same relative position in both figures ~representing
different frequency scales! correspond to the same analog
frequency. Several of the peaks appear at similar positions,
FIG. 5. Normalized NPS in the fastscan and slowscan directions for the ~a!
Kodak GP-25 ~172 mm/pixel! plate, ~b! Kodak GP-25 ~100 mm/pixel! plate,
and ~c! Kodak HR ~100 mm/pixel! plate.
FIG. 11. Comparison of average presampling MTF curves for the Kodak-
400 and Fuji-7000 readers and the ~a! type V ST ~200 mm/pixel! Fuji and FIG. 12. Comparison of average NPS profiles at 0.03, 0.3, 3, and 30 mR for
GP-25 ~172 mm/pixel! Kodak plates, ~b! type IIIN ST ~100 mm/pixel! Fuji the Kodak-400 and Fuji-7000 readers and the ~a! type V ST ~200 mm/pixel!
and GP-25 ~100 mm/pixel! Kodak plates, and ~c! type V HR ~100 mm/pixel! Fuji and GP-25 ~172 mm/pixel! Kodak plates, ~b! type IIIN ST ~100 mm/
Fuji and HR ~100 mm/pixel! Kodak plates. pixel! Fuji and GP-25 ~100 mm/pixel! Kodak plates, and ~c! type V HR ~100
mm/pixel! Fuji and HR ~100 mm/pixel! Kodak plates. ~NPS data at 0.03 mR
for the Fuji 7000 reader and type V HR plate were not available.!
average of the NPS profiles for the two directions. Generally, trend. The leveling off of Kodak values at higher frequencies
the Kodak NPS curves are higher than the corresponding is also not as prevalent for the larger plates.
Fuji curves for each plate type and exposure level studied. The DQE curves for both systems are compared in Fig.
For all plates studied, differences in noise power between the 13. Each curve represents an average of the two directions.
two systems tend to increase with increasing exposure level As expected from the NPS calculations, differences in DQE
and frequency. For 100 mm pixel sampling, differences be- between the two systems tend to increase with increasing
tween the Kodak and Fuji NPS curves ~both standard and exposure. The exception occurs at 0.3 mR where the DQE
high resolution plates! remain relatively constant up to 2.5–3 curves for each system are a bit closer than the 0.03 mR
cycles/mm @see Figs. 12~b! and 12~c!#. In this frequency curves. The Kodak DQE curves are ;30%–40% lower at
range the Kodak NPS curves are about 20%–40% higher at 0.03 mR, 20%–35% lower at 0.3 mR, 30%–50% lower at 3
0.3 mR, 40%–70% higher at 3 mR and 100%–170% higher mR and 50%–70% lower at 30 mR.
at 30 mR. At higher frequencies the Kodak curves begin to
level out while the corresponding Fuji curves continue to
B. General discussion
drop. For the larger sampling distance ~172 and 200 mm
pixels! the Kodak NPS curves are again higher and a similar A very wide range of exposures was used for these ex-
trend of greater difference with increasing exposure is seen periments. This was done to test the system at very high and
@see Fig. 12~a!#. Only the 0.3 mR curves deviate from this very low exposure levels. In routine clinical practice, expo-
compared in this report. Lastly, slightly different methodolo- cies. At lower frequencies the 100 mm/pixel GP-25 plate is
gies were used to measure NPS and MTF. In spite of these much better. At all exposure levels and frequencies, the 100
differences, there are some areas where the results can be mm/pixel GP-25 plate outperformed the 172 mm/pixel GP-25
compared. The MTFs measured in both studies for standard plate.
resolution plates ~both Fuji and Kodak! were very similar. These measurements have shown the Kodak-400 and
We were unable to compare the large pixel size in the laser Fuji-7000 systems to be comparable in resolution perfor-
scan direction and high resolution plates ~both Fuji and mance. Noise power studies have shown differences among
Kodak! since that data was not available in the previous re- the two systems. At low exposure levels, the Kodak images
port. Also, NPS values could not be compared since data for contain ;20% more noise power and 40%–70% more at
similar measurements was not available. We found that our higher exposure levels. The DQE for the Kodak plates is
measurements of DQE for the Kodak system were ;20% from 20%–50% lower than the corresponding Fuji plates in a
lower at all frequencies than those of Samei and Flynn7 for clinically relevant exposure range ~0.3–3.0 mR!.
both the standard and high resolution plates. This difference Measurements have shown that a beveled slit can be used
may be partially explained by a 3.5% difference in the value to measure the MTF of a system. A simple correction factor
used for the square of the ideal signal-to-noise ratio (2.61 can be applied to account for transmission through the bev-
3105 vs 2.73105 ). It is not clear to what the remaining eled jaws. The primary advantage of the beveled slit is that
difference can be attributed. the relief edges allow alignment of the slit to be a signifi-
The MTF values for two Fuji systems using standard cantly less critical issue. For the 4° relief angle of the jaws
resolution plates, measured by both Samei and Flynn,7 and the focal spot must only be positioned within 64°. At an
Dobbins et al.,4 were quite comparable. For the DQE results SSD of 1.8 meters that is 612 cm, a tolerance that was easily
there is only one curve where the two papers have reasonably accommodated by eye without the need for any alignment
comparable data. All the other curves are for different expo- mechanism. Since we also had a slit fabricated with straight
sures, kVps, or plates and readers. Comparing the PCR 7000 ~nonbeveled! jaws to compare our results, we were able to
reader from Dobbins et al. ~with ST-V plates, 200 micron find an accurate correction factor.
sampling and 0.3 mR exposure, and with both scan direc-
tions averaged! with the FCR 9501-HQ reader from Samei
and Flynn ~with ST-Va plates, 100 micron sampling, 0.3 mR
exposure, and with results for the plate scan direction only!
ACKNOWLEDGMENTS
shows approximately a 5% difference at low frequency up to
almost 20% difference at high frequencies. Since the differ- This work was supported in part by Grant No. 1 RO1
ences in the two results have a frequency dependence, meth- CA58521 from the National Cancer Institute and Grant No.
odological differences may have some effect. Since the DAMD17-94-J-4055 from the Department of the Army. The
present report used the same methodology as Dobbins et al., contents of this publication are solely the responsibility of
the methodology may account for some of the differences the authors and do not necessarily represent the official
between the present results and those of Samei and Flynn. views of the National Cancer Institute or the Department of
To correct for transmission through the beveled edges, we the Army. We would like to thank Eastman Kodak Co. for
chose to apply a sinc function correction factor to each MTF their help and the use of a 18 cm324 cm HR plate. We
curve. As shown in Fig. 10 this correction worked very well. would also like to thank the University of Wisconsin, Ac-
A more accurate description of the transmission profile, credited Dosimetry Calibration Laboratory for the use of an
through the slit, would presumably provide a more accurate ion chamber and electrometer.
correction. However, considering the small magnitude of the
MTF and DQE values involved at the higher frequencies, an
error of less than 1.75% should be accurate enough for most
situations. For the larger plate ~172 mm pixel! the correction 1
J. C. Wandtke, ‘‘Bedside chest radiography,’’ Radiology 190, 1–10
factor results in an error less than 0.8% up to the Nyquist ~1994!.
limit ~;3 cycles/mm!. The correction factor determined for 2
K. Aoki, ‘‘Computed Radiography in Pediatrics,’’ J. Digital Imaging 8,
these experiments is only applicable for this energy spec- 92–96 ~1995!.
3
trum. A different factor would be needed for different spectra F. E. Lindhardt, ‘‘Clinical experiences with computed radiography,’’ Eur.
J. Radiol. 22, 175–85 ~1996!.
since higher energy beams would penetrate the beveled jaws 4
J. T. Dobbins III, D. L. Ergun, L. Rutz, D. A. Hinshaw, H. Blume, and D.
of the slit to a greater extent and lower energy beams to a C. Clark, ‘‘DQE( f ) of four generations of computed radiography acqui-
lesser extent. sition devices,’’ Med. Phys. 22, 1581–1593 ~1995!.
5
W. Hillen, U. Schiebel, and T. Zaengel, ‘‘Imaging performance of a digi-
tal storage phosphor system,’’ Med. Phys. 14, 744–751 ~1987!.
6
VI. CONCLUSION H. Fujita, D.-Y. Tsai, T. Itoh, K. Doi, J. Morishita, K. Ueda, and A.
Ohtsuka, ‘‘A simple method for determining the modulation transfer
Among the three plates studied for the Kodak system, the function in digital radiography,’’ IEEE Trans. Med. Imaging 11, 34–39
~1992!.
100 mm/pixel HR plate outperformed both GP-25 plates at 7
E. Samei and M. J. Flynn, ‘‘Physical Measures of Image Quality in Pho-
the higher frequencies and exposure levels. At low expo- tostimulable Phosphor Radiographic Systems,’’ SPIE Proc. 3032, 328–
sures, all three plates are comparable at the higher frequen- 338 ~1997!.
8
T. Bogucki, D. P. Trauernicht, and T. E. Kocher, Characteristics of a tative radiographic imaging using a photostimulable phosphor system,’’
storage phosphor system for medical imaging, Technical and Scientific Med. Phys. 17, 454–459 ~1990!.
Monograph No. 6, Publication No. N-331 ~Health Sciences Division, 10
J. T. Dobbins III, ‘‘Effects of undersampling on the proper interpretation
Eastman Kodak C., Rochester, NY, 1995!. of modulation transfer function, noise power spectra, and noise equivalent
9
C. E. Floyd, H. G. Chotas, J. T. Dobbins III, and C. E. Ravin, ‘‘Quanti- quanta of digital imaging systems,’’ Med. Phys. 22, 171–181 ~1995!.