You are on page 1of 20

CHAPTER 1.

INTRODUCTION TO MASS SPECTROMETRY

Henry P. Longerich and Wilfredo Diegor


Department of Earth Sciences and Centre for Earth Resources Research
Memorial University of Newfoundland
St. John’s, NF A1B 3X5, Canada

INTRODUCTION
This is a general introduction to mass Most mass spectrometers can be easily
spectrometry, with special reference to mass divided into four separate parts, discussed in order
spectrometers (MS) used with an Inductively below. In some situations these parts become harder
Coupled Plasma (ICP) ion source, with further to separate, but the four divisions are still useful.
emphasis on the requirements for sample The four parts are the vacuum system, the ion
introduction to the ICP using Laser Ablation (LA). source, the analyser, and the detector (Fig. 1.1). The
Thus the topic is LA-ICP-MS, with all the hyphens vacuum system will be discussed first since in many
optional depending upon editorial preference. It is ways it is the most important and fundamental part,
however worth while to start in the beginning as all which will be the cause of most of the “head aches”
stories should. We will look first at the last part and of the operator. Secondly is the source from which
that is the MS part. Mass spectrometers come in ions are created from the sample, focussed, and
various kinds and sizes, but have also a lot of accelerated; which for our instrument importantly
similarities among them. It is worth while to contains the ICP where the ions are created. Thirdly
appreciate not only what makes instruments is the analyser which is where ions are separated not
different but those aspects which are the same. As by “mass” but by their mass to charge (m/e) ratio,
the field is changing with new instruments being using quadrupoles, magnetic fields, time, etc.
introduced, it is important to understand the Fourthly, and of great importance is the detector
fundamentals of the instrument.

Figure 1.1. Schematic diagram of an inductively coupled plasma mass spectrometer showing the 4 main parts, as
discussed in this chapter: the vacuum system; the ion source or ICP; the analyser; and the detector.

1
along with associated electronics, which converts of 25ºC and a molecular diameter of 370 pm.
the ion beam to a number which can be used by the At atmospheric pressure an ion does not go
computer system. very far (0.1 µm) before a collision occurs, clearly
not a satisfactory condition. At a pressure of 10-5
VACUUM SYSTEM torr, a minimum operational pressure of a quadru-
A vacuum is needed for the simple reason that ions pole ICP-MS instruments, the mean free path is a
will not get very far in the presence of one reasonable 5 m in length. This is adequate for many
atmosphere pressure of air. Without a mean path studies since there is a low probability of an ion
length of more than one metre, it is simply colliding with another ion, molecule, or atom on its
impossible to have much of a mass spectrometer. It path through the instument. On the other hand there
is not a trivial matter to obtain and maintain a are high precision applications where even lower
suitable low pressure and to monitor this pressure. pressures and their accompanying longer mean free
Experienced users will confirm that the vacuum path are needed because even a small amount of
system is the cause of the majority of the problems scattering can result in an ion being diverted. With
with instrument maintenance. This section will relative errors of a few ppm, especially when
discuss the instrumentation used to obtain the measuring large isotope ratios, lower pressures are
vacuum, and the equally important devices used to needed. Typical quadrupole and TOF instruments
monitor the system pressure. are operated at a slightly lower pressure, closer to
10-6 torr. Pressures in “high precision” MS are often
Mean Free Path 10-8 torr and even lower.
The fundamental parameter of importance
in vacuum systems is the mean free path which is Pressure Units of Measure
the mean or average distance which an ion (or Clearly in the “modern” system of SI units
neutral) can move before it hits another particle. the measurement of pressure should be simple, but
While clearly longer is better, very low pressures that is not true. As the life of instruments is
are not cheaply obtained. In ICP systems it is approximately 10 years, it takes several decades for
especially difficult to obtain very low pressures as any change to start to take place and even longer
the “hole” in the instrument is in the order of 1mm before we can hope for a more limited list of units
in diameter, through which a large volume of gas of pressure measurement. In the previous table one
flows. Table 1.1 gives mean free path distances in of the older units of pressure, the torr, was used
metres as a function of the system pressure without it having been rigorously defined. This was
expressed in units of torr. These values are suitable done to convince the reader, that a vacuum is
for N2 or O2, which together compose 99% of air. necessary, and now we will further define units of
The calculation is based on an ambient temperature pressure measurements. Several units are in

Table 1.1 Mean free path distances as a function of the system pressure

Pressure (torr) Mean Free Path

760 (1 atm) 0.0000001 m (0.1 µm)

1 0.00005 m (0.05 mm)

0.05 0.001 m (1 mm)

10-5 5m

10-6 50 m

10-7 500 m

10-8 5,000 m

2
Table 1.2. Units of pressure measurement.

Bar psi Torr Pascal

1Bar 1 14.5 750 100 000

1 psi 0.068 9 1 51.7 6 895

1 Torr 0.001 33 0.019 3 1 133

1 Pascal 0.000 01 0.000 145 0.007 5 1

from CRC, 2000.

popular and accepted use regardless if we like it or vacuum system measurement, is used for other
not (Table 1.2). pressure and vacuum measurements and simply will
Torr or mm of Hg is one of the oldest not go away in North America. There are 14.6959
units of pressure (vacuum) measure and just will not pounds/in2 (psi) in one atmosphere. This pressure
go away. It was originally defined as the height of a unit is still commonly used in the gas supplies in the
column of Hg in a barometer which exerts the same laboratory, and requires mention for that reason
pressure as does the gas in the vacuum system. alone. Vacuum pressure (for very unsophisticated
Noting that using units of distance for pressure did systems like aspirators and diaphragm pumps) can
not make sense it was renamed torr after a famous be expressed relative to ambient pressure rather
scientist. As MS system pressures are well less than than to a zero of no pressure, adding to the con-
one torr, the unit of milli-torr is often encountered. fusion. Note also that in high pressure gas systems
Torr still remains, if not the proper SI unit, as a (like your automobile tires) pressure is usually
common and well recognised unit. given in units of Pounds per Square Inch Gauge
The unit of atmosphere (atm), while (psig) which is the pressure relative to ambient
commonly used in other geological and chemical pressure and not relative to zero. The “gauge”
calculations, is not usually used in vacuum meaning the pressure difference across the gauge.
system measurements. An atmosphere is equal to
760 torr. Pressure measurement
The unit of a bar is of similar vintage as The measurement of the system pressure is
the atmosphere. High pressures in geological a non-trivial part of mass spectrometer operation
systems are commonly given in bars or more and requires some note. Pressure measurement is
practically in kilobars. Recent geochemistry text made difficult for several reasons, one being that the
books use the bar as a pressure unit in preference to devices used in the laboratory to measure pressure
the atmosphere. The bar is slightly less than one typically produce a signal which is not only a
atmosphere, the bar being equal to 750.062 torr. A function of the system pressure, but also of the
milli-bar is thus not very different from a torr (1 composition of “residual” gas causing the pressure.
mbar = 0.75 torr), and as a first approximation can A second major difficulty is the limited dynamic
be treated as being equivalent. The use of the bar as range of the available pressure measuring devices.
a pressure measurement in vacuum measurement To complicate this discussion some pressure
has returned, after an absence. measurement devices can also affect the pressure in
The proper SI unit of pressure measure is the system, both in good ways in that the device acts
the Pascal, which is slowly coming into use in as a vacuum pump but also in a degrading way in
vacuum pressure measurement alongside the torr that gas may be released by the measurement
and bar. There are exactly 100 kilopascals in one device. The latter is often caused when part of the
bar, explaining the return of the bar as a pressure measurement device is heated.
unit of measure. Pirani. This gauge is one of the oldest
The old English unit of Pounds per gauges, one which can be easily made from a pair of
Square Inch (psi), while not used for “high” incandescent light bulbs. It often forms the

3
backbone of the monitoring system, being cheap is not linear. The unit is calibrated at one
and very robust. With the exception of calibration atmosphere pressure and at a “low” pressure.
drift, these units typically last a very long time. The Resistance is measured in a Wheatstone bridge
device is based upon the same principle as that of arrangement in which another “identical” filament
your vacuum “Dewar” coffee carafe. As the contained in a sealed container is used for reference,
pressure in a system is lowered the thermal resulting in a measurement system of lower drift.
conductivity (the rate of transfer of thermal energy Thermocouple. This gauge has been
through the media) of the gas is reduced. Thus your around for some time and operates on the same
vacuum coffee carafe keeps your coffee warm fundamental principle as does the Pirani, that being
because of the low thermal conductivity of the the change in thermal conductivity of a system as a
vacuum. In the Pirani gauge a heated filament (not function of pressure. Having a wider dynamic range
too hot, just warm) is placed in the vacuum system than does the Pirani, it is being used more often in
(Fig. 1.2). Heat is transported from the heated systems, although sometimes it is not obvious which
filament through the residual gas in the system to kind of gauge head is actually present. The
the walls of the system. The robustness of the thermocouple gauge uses a heated filament similar
device is related to the fact that the filament is not to that of the Pirani, but instead of monitoring the
operated at incandescence, but at a much lower temperature of the heated filament, a thermocouple
temperature, so that convection but not radiation is is positioned near the heated filament. The
the primary means of heat transfer. The lower the temperature of the thermocouple is used to monitor
pressure the lower the heat transport and thus the temperature which is a function of pressure. The
higher the temperature of the filament. For metallic lower the pressure in the system, the lower the
filaments the resistance of the wire changes with thermal conductivity of the residual gas, and thus
temperature and it is thus the resistance which is the lower the temperature of the thermocouple. This
measured and calibrated to system pressure. gauge is also very robust and tolerant of system
Unfortunately this gauge is really only useful over a pressures reaching atmosphere. As mentioned, with
limited dynamic range from approximately 10-1 torr only slightly more complexity than the Pirani, this
to 10-3 torr, the range in which the thermal gauge has a wider (lower pressures) measurement
conductivity of the system changes. The calibration capability than does the Pirani and is thus often
replacing the Pirani in instruments.
Hot Cathode Ion Gauge. Due to the
inability of the above mentioned gauges, which
operate on the principle of changing thermal
conductivity with pressure, to monitor the pressures
typical of MS operation, systems require a gauge
which is useful at the ultimate system operational
pressure. This often means some kind of hot
cathode gauge. A heated filament is used to create
electrons, which are accelerated through a hundred
volts or so. The electron “beam” creates positive
ions from the residual gas molecules in the system.
The positive ions are then in turn collected on a
wire held at a negative potential. The ion current,
which is measured, is proportional to the pressure.
The “burn out” of the filament is the most common
mode of failure of this detector. However the life
time has been increased by the use of treated
filaments which produce high electron currents
while operating at lower temperatures. They are
thus less prone to “instant” failure if the system is
quickly (accidentally) brought to atmospheric
Figure 1.2. Schematic diagram of a Pirani gauge. pressure. During initial turn on, additional pressure

4
can be created and reported as the gauge outgases pressure which the pump can obtain, but the speed
from the filament heat. These gauges are useful to of removal of gas from the system. The pump speed
monitor pressures from 10-3 torr to the lowest is often related to pump size, with larger pumps
pressures obtained in MS. As the gauge should not having a higher speed. In modern MS, the user will
be used at high pressures, it is typically protectively usually find only “rotary mechanical oil pump(s)”
interlocked to a Pirani. used in conjunction with “turbo molecular pumps”.
Cold Cathode Ion Pump This device is Other pumps will be briefly mentioned.
mentioned here, even though as the name implies it Rotary Mechanical Oil Pumps are the
is firstly a pump; however the device also operates first line pumps, which are used to drop the pressure
as a gauge. Instead of having a hot cathode, this from atmospheric to less than one torr. They are
device simply has two electrodes, both at ambient ultimately capable of, or specified to have, ultimate
temperature. A “high” voltage of typically from one pressures of 10-4 torr, a pressure which can
to three kilo volts is applied. When a residual gas realistically only be reached with a system of almost
molecule gets close to the positive electrode an zero volume, as the pump speed of these pumps is
electron is removed to the electrode (field not high. They are based upon a simple mechanical
ionisation) and a positive gas phase ion results. This system in which an eccentric rotor isolates a volume
positive ion is accelerated to the negative electrode, of the system gas, and using oil to make gas tight
where it is “buried”. The opposite can occur in seals pushes this volume of gas out of the system to
which negative ions form, but positive ion the ambient atmosphere. They are mechanical and
formation is more probable. Magnetic fields are noisy; the oil becomes contaminated and for long
used to increase the path length so that pump life should be changed frequently (monthly).
multiplicative effects take place. These devices are Older pumps had belt drives, with frequently belt
very clean and are found in the lowest pressure failure, but modern units are direct drive. Care is
portion of MS systems. As a pressure measuring needed so that in the case of a system failure that oil
device, an error can be exhibited due the significant does not get “sucked” into the vacuum system,
ion currents which they create, however this current although modern pumps contain much more robust
is a good indicator of system pressure. They are protective devices than were available in older
extremely reliable, as they use only a moderate pumps. Systems often include automatic or manual
current, and a moderate high voltage supply. “Down To Air” valves which open to allow
atmosphere to enter the pump during power
Vacuum Pumps outages. Rotary pumps are fundamental to system
The “pumps” remove the air from systems operation as “high vacuum” pumps are invariably
creating the vacuum. Before listing different pumps not capable of operating between low pressures and
and their operation it is important to note that atmospheric, but do operate between low pressures
vacuum systems operate in equilibrium. The and that easily obtained using the rotary pump and
ultimate pressure is an equilibrium between the hence the common term “backing” pump in which
pumps removing gas from the system and gas the rotary “backs” the low pressure pump. In all
coming into the system from “leaks” and from “out ICP-MS units the rotary pump directly pumps the
gassing” from system walls (especially water from expansion region, in which a pressure of between
metallic surfaces). All systems leak and if the pump 0.5 and 2 torr is usual. This same rotary pump or an
is closed off from the system pressure will rise. If separate rotary pump maybe used to back the
the leak gets too high, then unsatisfactory operation several turbo (v. i.) pumps.
of the instrument can result, even though the Turbo Molecular pumps, usually simply
pressure gauges may be indicating a satisfactory called turbo pumps, are almost universal in modern
pressure. If the pressure rises linearly with time instrumentation. The technology has been around
when the pump is isolated from the system, this is for some time, but the widespread acceptance of the
an indication of a leak; otherwise out gassing is pump has awaited the “trivial” engineering problem
suggested. The pressure is only an indication that of creating bearings which will run 24 hours a day
the system leaks are “small enough” to allow good at speeds of up to 100,000 RPM, which is really
operation. Thus the concept of the “speed” of a “screaming”. The mode of failure is still the
pump is important, being not the lowest ultimate “simple” failure of the bearings resulting in the need

5
for a replacement or rebuild unit. The pump is very discussed in three parts; the ion source, the analyser,
clean as it consists of 10 to 20 “fan” blades similar and the detector. The ion source is the place and
to design of jet engine turbine blades. As residual means in which ions are created, accelerated in the
gas molecules get close to a blade they are “hit” by desired direction, and focussed to increase the ion
the blade and forced towards the next blade, etc. As beam current. For the purpose of this essay, of
mentioned the pump is “backed” by a mechanical course the Inductively Coupled Plasma (ICP) is the
pump, although in very simple situations the pump ion source of interest. Other ion sources will be
is capable of operating without backing, but with discussed briefly so that the user can appreciate the
higher ultimate pressures obtained. Progress has differences, advantages, and disadvantages of
been made in the development of bearings, with various MS systems. In the present time when
modern units often being magnetically suspended, finances are limited, it is important that efficient use
and used without lubrication. The bearings are of facilities be made. It is important that an
sometimes cooled by the instrument water system. optimum choice of instrumentation and method be
Other pumps Other pumps are not used for each application.
commonly encountered in new instruments, but will The ICP has long been recognised as a
be briefly mentioned for completeness and because good source of ions, with some of the choice optical
they may be found on older facilities. Ion pumps emission wavelengths being from the photon
have been mentioned and will still be found in very emission from ions. The problem which awaited the
low pressure systems; they are simple and reliable. mid 80’s was how to get ions from an atmospheric
Diffusion pumps were for decades the pump of pressure plasma into a suitable low pressure vacuum
choice. They are cheap to make and can be “home system containing a mass spectrometer. The ICP is
made” either in metal or glass. The operation a good source of ions because it has a temperature
principle is that oil or Hg is heated and the vapour which is in the vicinity of 8000 K, or close to the
passed though “jets” at super sonic speeds, after apparent temperature of the surface of the sun, the
which some kinetic energy is transferred from the plasma of most familiarity. Most of the elements in
pumping fluid to the residual gas and forced to the the periodic table which are classified by
other end of the pump to which is connected a geochemists as being in the lithophile or siderophile
rotary pump. The use of oil required periodic geochemical groups exist in the ICP at close to
change, however the most serious disadvantage of 100% ionised. On the other hand the chalcophile
“diff” pumps was in the case of system failure when elements, those being in the upper right hand
hot oil vapour was “sucked” into the MS, which in portion of the periodic table, can also be found as
some organic instruments made the system ions, but with lower degrees of ionisation. All the
unusable. First generation VG ICP-MS, of which a elements with the exception of He, Ne, and F can
few are still in use, were supplied with “diff” clearly be seen in sample spectra; which is not to
pumps. Cryo pumps which can be as simple as a say that all elements can be determined with
cooled trap are not commonly used in ICPMS, but satisfactory detection limits for the applications
still have a place in the specialised tool box of which the user may wish to explore. The capability
vacuum systems. The first generation ICPMS units and limitations of the detectability of all the
from Sciex used a cryo pump which is simply a elements in the periodic table will be explored later.
refrigeration system which used He as a carrier gas The interface between the atmospheric
and operated at 15 K, making a simple fast pump, pressure ICP and the 10-5 torr vacuum is solved by
however one which took hours to reach operational not trying to make the lowering of pressure of 8
pressures and required a weekly “defrosting” of the orders of magnitude in one step. The first step is to
system. Various cooled absorbers can be allow or rather “suck” the plasma though a hole
significantly enhanced by the use of molecular (orifice) of approximately 1 mm in diameter, in a
sieves, charcoal, etc. to enhance removal of gas metal disc (Fig. 1.3). This is called the sampler and
phase molecules from the system. is usually fabricated from Ni, although Al, Cu, and
Pt are also sometimes used. The requirement is that
ION SOURCES the metal has good thermal conductivity, so heat
At a pressure of less than 10-5 torr it is from the plasma can be dissipated to a water cooled
possible to have an ion path, which will be plate. Stainless steel, for example, is not suitable

6
a similar sized orifice. This cone is called the
skimmer, which “skims” the plasma, allowing
through approximately 1% of the sample. The
plasma at this point has expanded significantly, and
while the true temperature has dropped (due to the
expansion) tremendously, the ions in the plasma
seem to have been “frozen” in the more rarified
atmosphere. The pressure behind the skimmer is
typically around 10-4 torr. As for the sampler, this
cone is usually Ni, but Al, Cu, and Pt can be used.
The pressure in this region is, in all modern
instruments, maintained by a turbo pump backed by
a rotary oil pump. Usually, but not always, a
separate rotary oil pump is used to back the turbos
and one is used to pump the expansion region.
Some distance beyond the skimmer
another opening is placed. This opening can be
much larger than the sampler and skimmer orifices,
as it separates the region operating at 10-4 torr from
an slightly lower pressure region operating at 10-6
torr or lower pressure. This is the analyser and
Figure 1.3. Schematic cross-section of the interface detector region and is usually pumped by a second
region of the ICP-MS, with the plasma being turbo pump. An additional vacuum region is
“sampled” by the sampler cone, then “skimmed” possible in larger sector instruments, especially
further by the next cone, before being directed important where very high precision isotope ratios
toward the analyser to the left. are to be determined.
Various lenses to focus the ion beam into a
due to its poor thermal conductivity. The pressure smaller solid angle can be found following the
drop across the sampler is from atmospheric to sampler. Various configurations are used in
between 0.5 and 2 torr (a lowering of pressure by different instruments, but they are most commonly
three orders of magnitude), the vacuum being cylinders, “donut”-like plates with holes through
maintained by a rotary pump. The pressure is which the ions flow, and “discs” or stops around
affected by the size (speed) of the pump and the which the ions must flow. As well as focussing the
diameter of the orifice. Thus large volumes of gas ion beam, these “lenses” also serve the function of
are carried to the pump, which has both good and removing some of the high energy photons which
bad effects on the oil contained in the pump. The can also be detected by some detection systems.
high volume of gases tends to “clean” out the oil, This is accomplished by forcing the ions to go
thus the time needed between oil changes might be through various bending and contortions through
longer than expected, however in our experience a which photons, which only go in straight lines, are
monthly change contributes to the long life of the absorbed. Different systems are possible, but a
pump. The pressure in the region following the detailed understanding of their operation is not
sampler (expansion region) can affect the required for the operator. With experience on a
instrumental sensitivity. Gas flow through the particular instrument the operator will soon
sampler has been estimated at from 1 to 2 litres per understand the ranges over which each lens
minute (approximately 100% of the sample carrier normally operates, and which lens is most
gas passes this orifice), and shortly after the sampler important. To complicate operation, invariably there
orifice, the gas reaches supersonic velocities, giving is an interrelationship between the various lenses,
the ions a forward velocity towards the analyser which has seldom if ever well documented.
portion of the instrument. Especially on quadrupole instruments, there is
Following the sampler orifice, by usually a “trade off” of sensitivity with various lens
approximately one centimetre is a second cone with setting; that is, low mass or high mass sensitivity

7
can be “traded off” one with the other depending on each instrument has been designed for a particular
the application. For quadrupole instruments the application and for this application will be found to
natural circular ion beam is optimal, however for be better (more accurate, more precise, and
sector instruments a beam shape with a rectangular cheaper).
shape is needed requiring other focussing lenses. In organic MS instruments Electron
Collision/reaction cells v. i. act as ion lenses as well Ionisation (EI) is very commonly used. EI is the
as forming a place in which gas phase reactions take source in the first MS instruments and remains a
place. simple and useful technique to create ions. In EI, an
Dynamic Reaction Cells or Collision electron beam with an energy of a few hundred volts
cells are an recent addition to ICP instruments, is used to ionise a vapour phase sample. This is the
although their potential was suggested many years technique used in geological “stable isotope” mass
ago. Similar reaction cells have been in use in spectrometers in which H, C, N, O, and S isotope
organic MS for many years, often in the “triple ratios are measured on vapour phase samples (H2,
quad” configuration in which ion-molecule CO2, N2, SO2, or SF6). ICP-MS, for reasons to do
reactions took place in the centre quad. Reactions with high backgrounds and large interferences, is
have been shown to reduce the intensity of several essentially useless for these kind of applications,
poly-atomic ions, including the famous ArO+ which even though ion signals are observed from all of
interferes with the major isotope of Fe. It has also these elements.
been suggested that a “thermalisation” of the ion Thermal Ionisation (TI) must be
beam in reaction cells produces improved isotope mentioned here as it is one of the very important ion
ratio measurements. These cells are currently placed sources for the study of inorganic radiogenic
after the ion lenses and before the analyser. isotope systems in the geological sciences. TI-MS,
Quadrupoles, hexapoles, and octapoles are used by which has been used for a very long time for several
manufacturers to contain the reactants, and when the isotopic systems (esp. Rb/Sr, Nd/Sm, and Pb/U),
cell is placed off axis, improved backgrounds can has recently been put in competition with new ICP
result. instruments. In a TI source the sample is placed on
Other ion sources are found in a wide a flat metallic filament having dimensions of
variety of commercial MS instruments and will be approximately 1 by 10 mm. The analyte is always
mentioned here, so the reader will have a better separated and concentrated before it is deposited on
appreciation of the differences between various the filament, using various solution digestion and
instruments. All of these have the function of separation methodologies. The filament is then
converting a gas, liquid, or solid sample into ions. simply heated and if the chemical “magic” is done
Important differences in the hardware depend on the right, ions will be emitted from the filament into the
phase (gas, liquid, solid, or plasma) of the sample vapour phase from which they can be accelerated
which the user wishes to analyse. The process of into the instrument. We will focus here on the most
volatilization, breaking the sample into molecules, important characteristics of this source in order to
and atoms, and finally into ions is the function of all better understand the differences between an ICP
sources. A wide variety of sources are commercially and a TI source for geological applications.
available both in the area of organic MS and in Filament temperatures are less than 2000ºC,
inorganic MS. Some instruments support several compared to ICP sources of 8000ºC, thus TI has a
sources. While organic MS would seem to be limited capability of producing ions from elements
outside the interest of this essay, we have with high ionisation potentials. Thus the number of
participated in work in which Zn isotope ratios were elements in the periodic table is more limited to TI
determined using FAB (Fast Atom Bombardment) sources than is available to ICP sources. For
ion source MS and compared to ICP-MS analyses. example the Hf/Lu radiogenic system can be
Chemical Ionisation (CI), in which energetic efficiently measured using ICP sources, but not TI
molecules are used to ionise samples are widely sources.
used in organic MS, but are seldom used in Mass Fractionation is a term widely used,
inorganic MS. All MS instruments are potentially often with different meaning by different writers. In
capable of quantitative elemental determinations this essay the definition of mass fractionation will
and isotope ratio determinations, however usually be limited to the phenomena which is observed in

8
TI, that is - it is caused by the difference in vacuum, a problem which is awaiting an ingenious
volatility of isotopes of different masses, as a TI researcher. The spark/arc plasma, created using a
filament source is heated. Simply, a lighter isotope direct or alternating current source, was formerly
is (usually) more volatile than a heavier isotope. In available with MS detection, however its
a process akin to distillation, initially the vapour applications have been entirely replaced by ICP-MS
phase is enriched in the light isotope compared to and GD-MS (v. i.). The spark/arc is similar to the
the solid sample. With time, the solid sample DCP. Important fundamental elemental analysis was
becomes depleted in the light isotope and at some carried out in the past using spark source MS
time the vapour isotopic composition becomes instrumentation with isotope dilution procedures. In
equal to that of the original sample. This typically these earlier instruments the spark was used to
takes place after 2/3 of the sample has been create the ions for the MS. There have been reports
volatilised from the filament. If sample introduction in which the spark/arc source is used as a sampling
continues, the vapour phase composition becomes device from which material is carried to an ICP in a
depleted in the light isotope relative to the original manner similar to laser-ablation sampling.
sample composition. From a chemometric point of Other laser-based ion sources have been
view, the important fact is that the measured used in the past, but are no longer commercially
isotopic composition changes with time because available. The interest here is to contrast and
lighter isotopes are more volatile, the same compare them to Laser-Ablation (LA)-ICP
phenomena, which can drive stable isotope systems instruments. A laser beam focussed onto a sample
and makes them useful for geological applications. creates a plasma containing ions. LAMMA
Chemometric corrections must be applied to the instrumentation (Laser-Microprobe MAss spectro-
measured data for mass fractionation in order to metry) was commercially produced in the past
obtain the true isotopic composition of the original which relied on the laser-induced plasma to create
sample. Further discussions of the chemometrics the ions which were extracted into a MS for
will be made, when various data acquisition and analysis. Unfortunately the efficiency of the
data reduction paradigms are discussed. The ionisation process was very variable with different
important point here is to appreciate the sample matrices making quantification very
fundamentals of the phenomena of mass difficult. In LA sampling by the laser is separated
fractionation and that mass fractionation is a time from the ion source (ICP) making, this a very
dependent phenomena. Later we will compare this powerful combination. The separation is similar to
time dependent mass fraction phenomena with that discussed using spark/arc sampling in
instrumental variation of sensitivity with mass conjunction with an ICP ion source.
which we will call mass bias. Mass bias or mass Also deserving of mention is MALDI
discrimination is subject to change from the slow (Matrix-Assisted Laser-Desorption Ionisation)
“drift” or change of instrumental parameters, but source MS. This has become a very powerful
this “drift” is not regularly and simply related to a technique for organic analysis. In this method the
real change in the vapour isotopic composition. sample is mixed with a substance that absorbs laser
Other plasma sources are possible. Some radiation and assists in the ionisation of the sample;
of the fundamental studies which led to the ICP-MS only very low laser energies are used although there
were done using MIP (Microwave Induced are some obvious similarities with LA-ICP methods.
Plasma). No commercial MIP-MS has been This method has some similarities to reported laser-
produced, however the MIP source is used in ablation sampling of solutions in which compounds
elemental MIP emission spectrometer detectors for are added to the solution to enhance the absorbance
GC systems, and thus suitable sources are available of the laser energy by the liquid.
for experimental studies. The other commercially Ion Probes or Secondary Ionisation Mass
available plasma source is the DCP (Direct Spectrometers (SIMS) must be mentioned as a
Coupled Plasma). This plasma is very robust and powerful and older technique which is in
has a small but important place in elemental competition with LA-ICP-MS. One of the more
emission spectroscopy. Unfortunately the geometry powerful models has taken the name SHRIMP
of this plasma source is such that ions are not (Secondary ionisation High Resolution Ion
conveniently extracted from the plasma into a MicroProbe). In SIMS an ion beam is created, often

9
from O, Ar, or Cs and these ions are focussed using friendly conversation. Atomic Mass Unit (AMU or
a small MS onto the sample. The impact of the ion amu) is often used, without much formal
beam on the sample creates secondary ions from the recognition. “Mass” is clearly formally incorrect,
sample which are analysed with the main mass and when multiply charged ions are detected the
spectrometer. Capabilities are comparable to that of meaning is especially confusing. However, “mass”
LA-ICP-MS. LA systems are cheaper and much remains the colloquial word of choice, in the
faster than SIMS techniques, but the older SIMMS laboratory and will be used here, remembering that
instruments are still producing high quality data for m/e is always meant. Fortunately, in ICP-MS most
certain applications, and the comparison of the two ions have a single charge; only a few have doubly
instruments requires an extensive examination. plus charge (alkaline earths and light rare earths),
Glow Discharge (GD) instrumentation has and higher charged ions are not seen.
not had much impact in the geological sciences, but Later the concept, measurement, and use of
has been the subject of a lot of interest in the different “resolutions” will be made. Resolution is
analytical community, with a large number of one of the important figures of merit of the various
applications in the semiconductor and high purity kinds of analysers. While the qualitative definition
metals sciences. GD-MS instruments are similar to refers to the ability to separate one mass from
ICP-MS instruments, and some researchers have another, more quantitative measures are needed.
modified ICP sources to GD sources. The GD The parameter of choice depends also on the
source can be thought of, in a very crude sense, as a characteristics of the analyser, with it being
low pressure Ar ICP, in that instead of operating at convenient and conventional to chose a parameter
atmospheric pressures the GD operates at a few torr. which has a constant value over the mass range.
Solid samples can be analysed, although not at what More fundamental, is what resolution is needed for
would be considered micro dimensions, and depth the various applications which are anticipated by the
profiling can be done very well. GD-MS has in fact user, and what consequences result from the use of
replaced the spark-MS instrumentation in many different resolutions. This discussion will also
applications. explore the potentials of “high resolution” along
In conclusion it should be emphasised that with its limitations. But first we will consider the
a large part of the power of LA-ICP-MS is due to various analysers.
the separation of the sampling process (LA) and the
ion source (ICP). This allows the separate Analysers
optimisation of the two steps in a manner not Magnetic Sector Analyser (MSA) was historically
possible when both the sampling and ionisation are the first analyser and has now become a very
done in the same operation. The significant advance important player in the ICP-MS area. The
in the science which created LA-ICP-MS was the fundamental principle, which comes from electricity
combination of three reasonably well understood and magnetism, is simply that in the presence of a
parts into the final instrument. magnetic field, a moving charge (ion) will travel in
a curved path (Fig. 1.4). When all ions have the
ANALYSERS same kinetic energy, it can be shown that the paths
Now for the “guts” of the instrument or the of heavier ions (larger m/e) will be less deflected
analyser which forms the fundamental part of the than those of lighter ions. Qualitatively a heavy ion
instrument. This is where the ions are separated just carries on, while a light ion is easily pushed
according to their mass to charge ratio (m/e). A sideways by the magnetic field. In some
mass spectrometer does not measure mass in the configuration (Mattauch-Herzog), ions with a wide
sense that a balance measures mass. A mass range of masses are focussed on a plane; this
spectrometer does not work without ions, thus the configuration was popularly used in spark source
discussion of ion sources before that of the analyser. MS and there are research reports on its use with an
Due to the fact that m/e is measured and not mass ICP source, but no commercial instruments have
itself, there is a bit of confusion, but mainly plain been produced. Even though it has been demon-
disagreement of what to call the units of strated that the Mattauch-Herzog configuration can
measurement. Dalton is the more formal unit of be used, sensitivity was very poor in experimental
m/e, but remains one seldom used, at least in instruments. In a MSA the ions are accelerated with

10
spectrometers are not constructed using an ESA
alone, although the addition of an ESA to an
instrument can be very useful. The ESA separates
ions based upon their energy, so that the theoretical
MSA equation given above can be utilized. This
equation is based upon the assumption that all ions
have the same energy (equal to the accelerating
voltage). The ESA is simply constructed of two
curved conducting plates between which the ions
pass, and a potential difference is applied to the two
Figure 1.4. Schematic diagram of a Magnetic Sector
opposing curved surfaces (Fig. 1.5). In an operation
Analyser (MSA). Patterned after Turner et al.,
some what similar to the MSA, ions going too fast
Instrumentation for low- and high-resolution
hit the outside wall, while ions going too slow are
ICPMS, In: Montaser, A. (Ed.), Inductively
pulled into the inside wall, while ions with just the
Coupled Plasma Mass Spectrometry. Wiley-VCH,
NewYork, pp. 421-501. By permission of John
Wiley & Sons, Inc.

several kilovolts of energy, with higher energies


potentially yielding better sensitivity and resolution.
This is much higher energies than are used with
quadrupole instruments (v. i.), and slightly
complicates instrument construction as one end
(source) or the other (detector) must be at a high
potential. The mathematics is:
2 2
m H r Figure 1.5. Schematic diagram of an Electrostatic
 = Sector Analyser (ESA). ). Patterned after Turner et
e 2V
al., Instrumentation for low- and high-resolution
where: H is the magnetic field, r is the radius, V is ICPMS, In: Montaser, A. (Ed.), Inductively
the accelerating voltage, and (m/e) is the “mass”. In Coupled Plasma Mass Spectrometry. Wiley-VCH,
single detector instrumentation, as a property of the NewYork, pp. 421-501. By permission of John
hardware, r, is fixed, and the instrument can select Wiley & Sons, Inc.
the desired mass by changing either the field (H), or
the accelerating voltage (V). In multicollector
instrumentation, H and V are selected so that the
various ions impact on different detectors
positioned at various positions (different r). Both H
and V are used to select the detected mass, in some
very elegant programmed combinations in ICP-MS.
Note that for constant magnetic field (H) and
accelerating voltage (V), that mass (m/e) varies not
with r, but with r squared. The consequence of this
is that masses become closer to each other as mass
is increased. A practical definition of resolution will
Figure 1.6. Schematic diagram of a double-
be one which reflects this relationship. The
focussing analyser. ). Patterned after Turner et al.,
magnetic field could be supplied by a permanent
Instrumentation for low- and high-resolution
magnet, but with current instruments an
ICPMS, In: Montaser, A. (Ed.), Inductively
programmable electro magnet is always used.
Coupled Plasma Mass Spectrometry. Wiley-VCH,
Electrostatic Sector Analyser (ESA) now
NewYork, pp. 421-501. By permission of John
demand our attention, though the ESA is not
Wiley & Sons, Inc.
actually a mass analyser, it is an energy filter. Mass

11
“right” energy will exit between the two plates. RF frequency, but this is unstable in operation, so in
While the ICP source does produce ions of similar practice crystal controlled, fixed frequency, RF
energy, there is some variation, which degrades supplies are always used. Ions of the selected mass
resolution. Note that TI instruments are invariably are subject to several oscillations, but will exit from
single focussing MSA instruments, whereas ICP the end of the rods, while ions of either higher or
instruments are invariably double focussing lower mass, either pass between the rods or hit a
instruments, double meaning that an MSA is used in rod. Since the quad passes only one mass for a
tandem with a ESA (Fig. 1.6). If the ESA come first given set of applied RF and DC, this analyser
(after the source) the configuration is known as a operates only as an sequential analyser, with none
Nier-Johnson. If the two sectors are reversed, with of the simultaneous (multichannel) capability of
the MSA first, the configuration is called, as might sector instruments.
be expected, a reverse Nier-Johnson. A double Time Of Flight (TOF) is the new analyser
focussing sector instrument is capable of much high found in commercially produced ICP-MS
resolution (v. i.), than is possible with a single instruments, with two manufacturers now producing
focussing instrument. Since TI instrumentation does instruments. As the first author is most familiar with
not demand high resolution and the energy spread the Leco instruments, the specifications of this
of the ions is small, only single focussing manufacturer will be mentioned as an example of
instruments are used with a TI source. performance. The operation of a TOF is entirely
Quadrupole (quad) analysers will now be different from the sector and quad analysers
considered. Quad analysers have a history of more previously discussed. The principle of the TOF is
than half a century, and were the analysers which indeed the simplest of all, although the time frame
were used in the first ICP-MS instruments can be difficult to comprehend by geologists used to
produced, and continue to be the majority of all thinking in giga years but not nano seconds. In a
ICP-MS instruments used. There are various TOF instrument ions from the source are collected
reasons for this, but before entering a comparative in a contained volume for a period of time (5 µs).
discussion of the attributes of the various analysers, This packet of ions are then, after being given
we will first briefly review the operation of this kinetic energy by being accelerated by several
clever device. A thorough understanding requires an thousand volts, are “thrown” down a long empty
appreciation of second order differential equations tube (1 m long). Newtonian physics tells us simply
and an ability to visualise in three dimensions. The that:
student is referred to other references to this theory. mv2
Quad means four, and in this analyser there are four KE = = eV
rods, originally about the diameter of a broom stick, 2
but now getting smaller. The four rods, about one That is the Kinetic Energy (KE) is a function of the
foot long, are arranged in a perfect square mass (m) and the velocity (v). Since all ions have
arrangement. The rods are ideally of hyperbolic been given the same kinetic energy from the
shape rather than circular. Pairs of opposing rods accelerating potential (V), all ions have a Kinetic
are connected together electrically, with the two Energy (KE) equal to the product of the electron
connections brought out of the vacuum system to a charge (e) and the applied potential (V). The
combination RF (Radio Frequency), often of 1MHz velocity (v) is the path length (L) divided by time
frequency, and DC (Direct Current) electrical (t).
supply. Ions enter the region between the four rods L
with a low velocity, so that as the ions pass through v=
t
the rods they are effected by the RF oscillations.
Solving for m/e (“mass”) as a function of time (t):
The more oscillations to which an ion is subject, the
higher the possible resolution. The quad is a mass 2
filter, allowing to pass only ions of a selected mass.  m  2V 2V t  2V  2
Mass is selected by applying a calculated value of
 = 2 = 2 = 2 t
e v L L 
the DC plus RF, with the ratio between DC and RF
usually maintained constant at a fixed RF Thus mass is a simple function of time (t) for fixed
frequency. Theoretically it is possible to scan with accelerating voltage (V) and fixed flight length (L).

12
Qualitatively, if all the ions have the same energy, upon the application and analyser, indeed different
the light ones are moving fast and arrive at the end definitions may be important and useful.
of the path in a short time, while the heavy ions Peak width is the simplest definition of
move slowly and get there later. Note the syntax, in resolution. The peak width can be measured at any
that detection in this analyser is sequential in time, relative height, and can, depending upon the desires
the sampling is “simultaneous”. With a path length of a writer, be used to make the analyser look better.
on the order of one metre, the entire mass spectra is In quad instruments the peak width at 10% peak
acquired in 50 µs, with a data value being acquired maximum is commonly used for alignment. It is
every 2 ns. The analyser has the characteristics of a normal practice to set up the instrument so that
simultaneous device and is expected to have all the resolution, defined this way, is constant. Typical
advantages of sampling each ion at the same time. values are between 0.7 and 1.0 mass units. As is
We might call the analyser pseudo-simultaneous. As characteristic of almost all instrumentation, higher
with the MSA analyser the peaks get closer together resolution results in a loss of sensitivity.
as the mass increased, because mass is proportional Valley definition is useful as it gives a
to the square of time (t). A complete spectra more qualitative indication of what is normally
contains 25,000 values. Each second 20,000 thought of as resolution. When a valley of 10% of
complete spectra are acquired, consisting of peak maximum between two peaks of equal height
500,000,000 values. Clearly some very fast occurs, resolution is specified as ∆m, the difference
electronics and data preprocessing is required to between the position of the two peaks. Since the
make this instrument a reality, and this is one of the possibilities of finding two peaks of equal height is
reasons that TOF have only been recently low, it is useful to note that if two hypothetical
introduced to the commercial market, although the peaks give a valley of 10%, then the peak width at
theoretical concepts have been well known and used 5% peak maximum is equal to ∆m.
for a long time. m/∆m definition is useful for sector and
Ion Trap (IT) is another analyser, which TOF instruments, where “m over ∆m” is constant.
has been demonstrated in research instruments, but Implied is the previously mentioned fact that
has not at the present time been introduced in resolution, defined as peak width, is better at low
commercial ICP instrument. The IT is related to the mass and worse at high mass. Low resolution is
quadrupole, but in operation ions are collected and about 300, being equivalent to a peak width
held in the trap. Later ions are then released as a resolution of one at mass 300. Sector instruments
function of mass. The IT has found a place in can be operated at “high” resolution of around
organic MS, and likely will be used in ICP 10,000. This resolution is easily calculated by
instruments in the future, if only as a component of taking the mass (peak centre) divided by the peak
a much more complicated system. width at 5% peak maximum, although
Ion Cyclotron Resonance (ICR) is a very manufacturers will be found to prefer 10% peak
sophisticated device in research development. Very width which gives a better appearing value.
high magnetic fields are used, of such strength that Abundance Sensitivity is not usually
they are only found in very specialised research thought of as a resolution definition, however this
facilities. Ions are trapped and held in a circular parameter is of fundamental importance, especially
motion in the magnetic field. The motion of the ions when isotope ratios are determined which have
are perturbed by applied electrostatic fields, which values far from unity. Abundance sensitivity is the
results in the production of a Fourier transform of ratio of the intensity of a mass peak relative to the
the mass spectra. Very high mass resolution is intensity of the “tail” of that peak at mass plus one
possible with this analyser and hence the research and at mass minus one. It is thus a measure of the
interest. “interference” of a mass peak on its two neighbours.
Notably for quad instruments there is a larger tailing
Resolution definition on the low mass side compared to the high mass
Before discussing the comparative side. Similarity TOF analysers exhibit a tailing,
differences in capability of the various analysers it except on the high mass side instead of the low
is important to quantitatively define and discuss mass side.
various measures of mass resolution. Depending

13
High Resolution which is equal to the total count of the number of
What is the interest in having an instrument neutrons and protons. Further, there is not a
which is capable of operating in a high resolution constant binding energy due to each neutron and
mode? That is, with resolutions better than unit proton as shown in Fig. 1.7. The figure is a plot of
resolution (peak width at 10% peak height) or better the “exact mass” divided by the integer “mass” vs.
than an equivalent m/∆m of 300. To understand the the integer “mass”. Clearly seen is the extra stability
capabilities and limitations of high resolution, a at mass 4He. Not as clear in this figure, is the extra
consideration of the “mass” of the proton and stability at mass 12C, 16O, and 20Ne. Note the
neutron need to be made, and the loss of mass as minimum which is found at mass 56Fe, the heaviest
these fundamental particles combine to form nuclei “major” element, which reflects that light elements
where some of the mass is converted to binding “fuse” to form heavier nuclei while heavy elements
energy. While the manufacturers of high resolution undergo “fission” forming lighter nuclei, with Fe the
instrumentation would like buyers to believe that all “goal”.
possible mono and poly atomic ions can be The “mass defect” varies with mass. In
separated, that is simply not true. There are however theory, it is possible, knowing the exact mass of an
some interesting applications of high resolution ion, to determine the element. For example 14C
instrumentation. (important for archeological dating) has a mass of
The mass of each neutron and proton 13.996 758 Dalton, which is different from 14N (the
which makes up a nuclei each have a mass of major interference in the determination of 14C by
approximately 1 amu or 1 Dalton each. However, mass spectrometric means) which has a larger mass
when neutrons and protons combine to form a of 13.996 926 Dalton. The mass difference is 0.000
nuclei, there is a loss of mass (mass defect) as some 168 Dalton, so a resolution of 83,000 would be the
of the mass is converted to binding energy of the minimum required to separate these two ions.
nuclei. The loss of mass is in accordance with the Noting that standard double focussing sector
Einsteinian law that “E” equals “m c squared”. Thus instruments have a high resolution of around
the exact mass of an ion is not an integer number 10,000, the determination of 14C is not practical on

Figure 1.7. Plot of the “exact mass” divided by the integer “mass” vs. the integer “mass”.

14
this kind of mass spectrometer. An important 1,000 cps. However at this signal intensity, the
conclusion is that all interferences cannot be digital mode is superior, a superiority which the
resolved using standard double focussing sector digital mode maintains until near the upper range of
instruments, and this includes importantly elemental the digital mode. Cross calibration is required if
ion overlaps. The existence of the minimum at 56Fe both modes are to be used, which converts the
has the unfortunate consequence that two nuclei to analogue value into an equivalent digital count rate.
the left of the minimum can have a combined mass This is done by making measurements where both
very close to one nucleus to the right. For example modes are operational (100,000 to 1,000,000 cps).
51 40
V Ar has an exact mass of 90.906 345 6 and There is a mass dependence of the cross calibration,
could only be separated from 91Zr (90.905 644 2) so the cross calibration is usually done with several
with an unobtainable resolution of 130,000. In elements across the mass range and a curve fitting
conclusion high resolution mass spectrometers have applied to the results. The analogue mode is
a powerful ability to resolve some overlaps of two operational up to somewhere close to 1,000,000,000
ions having the same nominal mass, but not all over equivalent cps, limited by the largest potential
laps can be resolved. Note also that the calculation (voltage) input which the analogue digital converter
of the minimum required resolution here assumes can allow.
the unlikely occurrence that the overlapping ions are The most common detector (transducer) in
of the same intensity. In the more likely situation in first generation instruments was the Channel
which one peak is much larger than the other, even Electron Multiplier (CEM). This device is a one-
higher resolution would be required. piece-cone-like device manufactured in glass, with a
high resistance coating. A moderately high potential
DETECTORS of a few kilo-volts is applied. When an ion impacts
Now for the final and “tail end” of the on the first surface several electrons are released.
instrument, where the ion beam is converted to a The following operation is similar to that of a photo
computer-usable number, either in pulse counting multiplier, in that each subsequent impact of
(PC) or digital mode in which ions are counted, or electrons increases the net electron current flow,
in an analogue mode in which the ion beam current until at the end of the “cone” a current pulse of
(amperage) is converted to a potential (voltage) multiple electrons is collected. This current pulse is
which is in turn converted to a number using some then counted in the digital mode or the combined
kind of analogue-to-digital conversion hardware. current is measured in the analogue mode. Often to
Ideally the user would like to have a detector which increase the lifetime of the detector, the applied
can operate in either mode because the two different “high voltage” is decreased when in the analogue
modes are optimum in different intensity ranges. mode. These devices are moderate cost ($K), and
For low intensity signals, the digital mode is moderate lifetime (year), but are subject to
optimum, from the point of view of the best relative deterioration with each ion impact.
standard deviation of the signals and minimum The Micro Channel Plate (MCP) device
detection limits. At high ion count rates (greater is based upon the same principle, but contains a
than approximately 1,000,000 counts per second or large number of small tubes. Spatial resolution is
cps) pulse pile up or dead time corrections degrade possible, but the more common application has been
quality. Hardware or software corrections may be large area detectors in TOF instruments. However
applied, but the correction theory has problems as the more interesting future application will be as a
intensities get too high, and eventually fails entirely. potential detector for multiple channel applications.
As a rule of thumb, below about 100,000 cps dead Currently MCP are plagued by low dynamic range
time is not significant, and the user should be along with low amplification.
careful at higher count rates that a correction has More recently a device based upon the
been applied. In any case at sufficiently high count same principles but with discrete dynodes have
rates (several million cps), the sequence of pulses been introduced. A resistor chain is used to apply
turn into one giant pulse and the digital mode is no separate and different potential to each dynode. The
longer operative. reputed advantage is the lack of need for the high
For most hardware, the analogue mode resistance coating, making what should be a more
starts at a ion beam intensity equivalent to about reliable device. With some minor mechanical

15
changes channel electron multipliers can be For a one second integration this gives a RMS noise
replaced with the new discrete dynode multipliers. of 29 µV, which, using Ohms law is equivalent to an
In analogue mode the analogue signal can be taken ion current of 1790 ions/sec. Notably the noise is
off at an intermediate dynode, with no change in constant with increasing ion currents. Thus there is
applied potentials or removal of the applied a rapid improvement in the relative noise (RSD)
potential to the later dynodes. with increasing signal compared to a digital signal,
The electron multiplying detectors have where the noise is the square root of the total
low noise gain and are found in essentially all single counts, thus the relative standard deviation (RSD)
detector instruments. They can be operated in either improves with the square of the intensity. The
digital or analogue mode, but if the applied important conclusion of this calculation is that the
potential is not reduced when high ion beams are base line noise is approximately 1790 cps, or a three
encountered (analogue mode) a reduced life time is sigma detection limit, expressed as a signal, is 5,400
expected. TOF instruments due to their very high ions/sec, approximately three orders of magnitude
rate of data acquisition do not have sufficient time higher than for pulse counting using a quadrupole
to switch modes and are operated simultaneously in analyser. The upper limit of the Faraday detector is
both analogue and digital modes. There is when the potential across the resistor reaches the
insufficient user information to know how much this maximum allowed by the following analogue to
will shorten the life time of the detector, although it digital converter. With an upper limit across the
should be noted that the detectors used in TOF resistor of 10 volts, which is equivalent to 10-10
instruments are larger in size, which is expected to amp, then the maximum possible intensity is
result in comparable lifetime. Note that the 620,000,000 ions/sec or a range of more than five
operation of all instruments, including TOF orders of magnitude.
instruments, can and should be operated to “skip” A disadvantage of the Faraday detector is
over very intense ion signals. that along with the resistor comes some capacitance,
For multiple collector instruments (MC), which makes switching of signal to the detector very
the detector of choice is the Faraday detector, slow. Why is such a very high resistance used, as
named after the famous scientist. The operation is clearly from the equation, lower resistance would
very simple in that for each singly charged ion lead to lower noise? Note however that lower
which hits the “bucket” an electron flows through a resistance reduces the signal linearly with
resistor, creating a potential difference across the decreasing resistance, while the noise would
resistor. The operation is very linear and very decrease more slowly with decreasing resistance,
stable, allowing the use of multiple collectors being proportional to the square root of the
simultaneously. The detector is 100% efficient resistance. Thus the best RSD results when the
meaning that all ions, which hit and are retained in highest value possible of the resistor is used, and for
the bucket, are detected. The gain is stable as gain is practical reasons of availability that usually means
dependent only on the value of the resistor, 100 GΩ.
potentially a very stable quantity. Unfortunately the Daly detector systems have been used on
resistor has a very high resistance (1011 Ω or 100 G TI instruments for some time, with a similar system
Ω). Noise (standard deviation or root mean square) now available on ICP instruments. The Daly system
in a 100 G Ω resistor, which is due to random is based upon positive ions impacting an aluminium
movements of electrons is given by: knob held at a negative potential. The impact
produces electrons, which in turn impact a
1 scintillator at which even more photons are
Noise = 4kTR = 29 µV produced. The photons are in turn detected by a
2t
traditional sealed photomultiplier. This device can
≡ 1.6 x10 −19 amp = 1790 ions / sec be used either in digital (fast phosphor) or analogue
(persistent phosphor) and has recently been used
Where k is Plank’s constant, T is the absolute
with ICP sources. The system is notable for its long
temperature (298°K), R is the resistance 1011 Ω, and
life with the initial ion impact being on a relative
1/(2t) is the frequency band pass where t is the
inert aluminium “knob”.
integration time (1 second equivalent to 0.5 Hz).

16
Photographic detection is obsolete, “masses” at which intensity data acquisition is
requiring processing to “develop” the image, required, with from 50 sweeps per second (n=2) to
followed by subsequent digitisation. Silver based 3 sweeps per second (n=33) possible. The analyser
films for MS require an emulsion made with silver is low cost due its light weight, requiring only
halide only, without the gelatin used in all other precision machining and precision component
films, making for a difficult medium. As well, the alignment. Resolution is normally unity (peak width
dynamic range of halide films are limited, requiring definition), although research is being done to
several exposures with different integration times to exploit other operating modes and has been
acquire the required dynamic range. Photographic demonstrated to allow higher resolution. Sensitivity
detection was always used with the Mattuch-Herzog is high, and until recently the highest. Backgrounds
analyser configuration in which the entire spectrum in the continuum region are typically 10 cps but are
was simultaneously obtained. This detector is now less than 1 cps with newer instrumentation.
mentioned as the future may include a simultaneous Sector instruments (MSA, ESA), were the
MS instrument using the Mattuch-Herzog second kind of analysers to be introduced to the
configuration and some kind of array detector, commercial ICP market. If magnetic field changes
possibly a scintillator followed by a suitable array are required these instruments are much slower,
photo-detector. Current problems are effects of the although speed is continuing to be improved as
magnetic field on the electronic photo-detectors and better laminated magnets and magnet power
very low sensitivity of the instrument, both supplies are being produced. Scanning over limited
problems which it is hoped will be solved by mass ranges (20% of starting mass) is possible using
advanced engineering. This could be the “dream” a fixed magnetic field (MSA) and simultaneously
instrument of the future. Again it is noted that this scanning the ESA with the accelerating potential.
the Mattuch-Herzog configuration was used for The Finnigan instrument allows an elegant
years with spark source MS and contributed combined ESA/MSA scan in which the MSA is
significantly to geoanalysis. scanned linearly while a saw tooth wave form is
applied to the ESA. While this mode allows
INSTRUMENTS efficient data acquisition, it forces the operator to
Having defined basic terminology and an obtain data at masses which are neither wanted or
understanding of various kinds of analysers, it is useful. Settling speed is thus very difficult to
appropriate to qualitatively discuss similarities, quantitate, as the time is different for every possible
differences, and figures of merit of various mass jump and depends on the mode used. As many
instruments. All of these instruments have a place in problems require the signal be acquired for low
the analyst’s tool box, and all have demonstrated mass major elements and high mass trace elements,
some special applications. the speed of this analyser will often be found to be
Quad instruments, being the oldest and inferior to quad instruments. The analyser is of high
most common ICP analyser are considered first. A cost due, in part to its ton(s) of magnet weight. High
quad is sequential, but is fast sequential. Settling resolution is possible with these instruments.
time between mass jumps is between 2 and 0.1 ms, However a severe loss of sensitivity when changing
and getting faster and more intelligent. Using “dwell from low to high mass, roughly proportional to the
times” of 8 to 20 ms allows efficient use of the resolution is incurred. Changing from low (300) to
analyser operating at well above 90%. The settling high resolution (10,000) can result in more than a
time is not a function of the mass change, but jumps 100 fold loss of sensitivity. As previously discussed,
over major intensities (e.g. Ar) can require not all mass overlaps cannot be resolved at the high
additional time. In first generation instruments, we mass resolution setting, which makes high
used an additional mass at the beginning of a sweep, resolution not a panacea. When changing from low
following the “fly back” to the lowest mass from the to high resolution, also there is a “sharpening” of
highest mass in the acquisition. Measurement of 100 the peak shape, as the peak changes from “flat
intensities per second (10 ms per mass total dwell topped” to approaching a triangular shape, which
plus settling) is easily obtained with good requires a very high “mass calibration” stability
efficiency. The number of complete sweeps per which is not easily obtained. This can force a “scan”
second, n, obviously depend upon the number of over each peak to ensure that the peak is accurately

17
located. Note further that a change in resolution is TOF instruments are sequential in
slow because it requires the mechanical operation of sampling and extremely fast, with a complete mass
changing entrance and exit slits, making data scan acquired in 50 µs equivalent to 20,000
acquisition using different scan modes slow. complete scans per second. Due to hardware
Sensitivity is currently the highest of any limitations, the Leco instrument currently saves
instruments, but note that for some time it was “only” 64 integrated intensities, more than sufficient
claimed that the sensitivity was high simply because for most anticipated applications. The unique
it was a sector instrument. Recent sensitivity capability of this instrument is currently limited by
improvements came with the use of the shield torch low sensitivity, so low that the low sensitivity is not
which decreases ion energy variations yielding traded off for by the multiplex advantage of its
better transmission. This improvement is especially simultaneous sampling. Current state-of-the-art TOF
important in sector instruments, but of much less instruments are around 1000 times less sensitive
importance in quadrupole instruments which are than state of the art sector instrument, therefore the
much more tolerant of energy variations in the multiplex advantage will not improve signal
incoming ion beam. Backgrounds are very low (0.2 integration as 1000 “masses” are clearly ridiculous.
cps) in the continuum region of the spectra in sector Improvement in future instruments is hoped for,
instruments due to the long curved ion path, which however there may be a limitation to the ability to
usually is a half circle. Photons are very efficiently hold ions during the ion collection period. If the
eliminated, along with other ions which may in fact collection period could be increased from 5% to
be important in the formation of the continuum 100%, a significant (10 fold) improvement in
background. Lower backgrounds lead to lower sensitivity is possible, but this has not yet been
detection limits, but note that the improvement is engineered. Backgrounds are at first view, similar to
reduced at low total integration times of one or so quadrupole instruments at 10 cps, however due to
seconds often used in laser-ablation sampling. the simultaneous acquisition of the TOF it is
Multi-Collector (MC) instruments are advantageous to integrate over a number of
sector instruments, but are discussed separately. windows (default of 19 in the Leco instrument) this
This instrument avoids the need for magnet gives a background of around 200 cps (10*19).
changing by incorporating a number of detectors, Resolution is comparable to low resolution modes
which allows simultaneous detection, however only of sector instruments and are typically between 300
over a limited mass range. As mentioned, v. s., a and 500. Anticipated “killer applications” of the
Mattauch-Herzog configuration would not have this TOF will be in very fast transient signal acquisition,
limitation, but this instrument has not been but these are yet to be reported due to the very low
commercially produced with an ICP source, and the sensitivity of the current generation of TOF
literature report on research instruments had instrumentation, which means that although the
extremely low sensitivity. These MC instruments scanning hardware is very fast the detector is
are uniformly operated in low resolution. MC essentially blind.
instruments are truly simultaneous, giving them all The use of the analogue and digital ranges
the advantages thereof. However, for multiple in TOF instruments is distinctly different than
collection, most instruments use only Faraday described for the other single collector and multiple
detectors, due to the need to very precisely “match” collector instruments. In the Leco TOF instrument
the separate detector/amplifiers used. One or more the maximum digital count rate is approximately
digital detectors are often available, but the high 20,000 cps, much lower than the approximately
precision analysis required for geochronological 3,000,000 cps of other systems. Dead time
and stable isotope applications are almost always considerations are also more complicated as a only
carried out using multiple Faraday detectors. Thus zero or one ion can be collected in each 2 ns
there is an requirement for high ion currents, interval, and following the detection of a single ion,
preferably close to 100,000,000 ions per second. the detector is “dead” for several subsequent bins.
The user must carefully consider the total amount of
sample available in order to obtain such a high CONCLUSION
signal for a reasonable period of time (minutes). Inductively coupled plasma mass
spectrometry, while still undergoing continued

18
evolutionary development is now a well established analytical methods. This course has developed since
instrument in geoanalytical laboratories. All three the mid 1980s, and was always deemed to be his
commercial analysers, quad, sector, and time of better lectures in the course. The number of mass
flight have been demonstrated to be useful spectrometers in the department is and continues to
instruments when used with laser-ablation sample be large. The department has ICP-MS, gas source
introduction. As a very small portion of the ions in stable isotope MS, TI-MS, and organic MS. At one
the ICP actually reach the detector, there is, time there were nine operating MS instruments in
theoretically, the potential for further large increases the department. Other geoscience departments have
in sensitivity. This sensitivity can be used to or had spark/arc MS, glow discharge MS, rare gas
measure lower concentrations, or more usefully for MS, and secondary ion MS. Thus mass
the analysis of smaller volumes of sample. Other spectrometry has and continues to have a very
MS configurations which are not currently important impact on the geosciences.
commercially available, including notably the Most of the writing in this chapter came
Mattauch Herzog configuration with true entirely from memory, with only an occasional look
simultaneous detection is an interesting potential at lecture notes to make sure no important item was
instrument for the future. missed in the discussion. It is very difficult to find
Collision/reaction cells will be discussed in references to knowledge, some of which HPL has
a later chapter in this volume. These cells seem been familiar with since graduate student days when
poised to have a very significant impact on LA he did his first research using vacuum systems. Text
analysis in the future, considering their important books from various analytical instrumental
capability of reducing background and background chemistry sources will contain additional reading on
ions, without an accompanying loss of sensitivity. mass spectrometry that will be of use to the reader.
The use of these cells will create many new A very useful book on geoanalysis is that by Phil
problems and require many new choices of Potts, which covers mass spectrometry; however the
operating conditions, which will require some time ICP-MS chapter is seriously outdated and it was
for development. written in the very early development days of ICP-
Detector technology is undergoing MS. This book has been reissued in paper back, and
continuing development as well. Accompanying the is a highly recommend text covering most of the
higher sensitivity of new MS instruments, the need important areas of geoanalysis, including other
for increased dynamic range continues in types of mass spectrometry. In the area of ICP-MS
importance. The speed of switching modes from the volume edited by Akbar Montaser (1998) is a
analogue to digital is undergoing development, with must desktop reference book. Attendance at the
a new system being introduced which reduces annual Winter Conference on Plasma
sensitivity by changing applied detector potential Spectrochemistry held in alternative years in the
instead of switching to an analogue mode. Another southern United States and in Europe is highly
important need is for multicollector ion counting recommended to keep up to date in the field.
systems. At present, the drift between separate
detectors is large enough to make their operation REFERENCES
unsatisfactory for high precision isotope ratio CRC (2000): CRC Handbook of Chemistry and
measurements. This results in multicolletor Physics, 81st edition, Lide, DR, editor-in-
experiments being carried out with multiple Faraday chief, CRC Press, Boca Raton.
detection, with possibly one ion counter detector.
As explained, v. s., while Faraday detection is very MONTASER A. (1998): Inductively Coupled Plasma
efficient and stable, they have a background noise Mass Spectrometry, Wiley-VCH.
level near 1,000 cps, a severe limitation for many
applications.

ACKNOWLEDGEMENTS
This essay is almost entirely based on a
lecture given by HPL as part of an Earth Science
graduate course at Memorial University on

19
20

You might also like