Professional Documents
Culture Documents
13
GENERAL INTRODUCTION
The present study incorporated in the thesis was taken up by the author with an
aim to develop more efficient and validated new high performance liquid
chromatographic methods with UV / mass spectrometric detection for estimation of
some important drugs namely felbamate, gemfibrozil, linezolid, pioglitazone and their
metabolites and lamivudine, zidovudine individually or in combination in human
plasma. The study design involves the development of new reverse phase HPLC & LC-
MS/MS methods for estimation of the selected drugs, validation of the methods thus
developed and testing their suitability for estimation of the drugs in plasma samples.
Out of a total of five methods proposed three were carried out by adopting reverse
phase HPLC technique and the remaining by LC-MS/MS technique. The methods were
validated as per FDA as well as ICH guidelines.
14
A literature survey on the analytical methods of felbamate, gemfibrozil,
linezolid, pioglitazone and their metabolites and lamivudine, zidovudine revealed that
some HPLC & LC-MS/MS methods are available for their estimation in plasma and
other biological fluids. Some of these methods have certain drawbacks like low
resolution, lesser sensitivity, long run time, incomplete recovery, large volumes of
plasma sample for the extraction technique and large volume of injection which results
in less number of injections on the column etc. Furthermore, some methods were
partially validated and not as per the desired guidelines. Hence, the author had
attempted to develop simple, faster, more reproducible methods for the determination
of these drugs, using low volumes of plasma samples for the extraction and injection
thereby ensuring longer column life. The methods proposed by the author are less
tedious and economical. The proposed methods can be used as alternative methods to
those reported by the earlier workers and provide good choice for the routine
determination of the chosen drugs in their plasma samples for their clinical,
pharmacokinetic, bioavailability and in bioequivalence studies.
The thesis has been presented in five chapters. Chapter-I describes the
introductory information about HPLC & LC-MS/MS and their techniques. This is
followed by the general guidelines and methodology to be followed for developing
methods for estimation of the drugs by HPLC & LC-MS/MS. Later, the procedures
adopted to determine various parameters for validation of the methods have been
reported.
Chapter- II, III and IV deals the HPLC method development for the assay of
three drugs namely felbamate, gemfibrozil, and linezolid in human plasma. The
method development is followed by the determination of various validation
parameters.
15
Chapter- IV and V describe’s the details of the authors experimentation and results
obtained in the LC-MS/MS method development for the assay of Pioglitazone and their
metabolites (Keto-pioglitazone and Hydroxy-pioglitazone), and simultaneously
determination of Lamivudine & Zidovudine in human plasma. The method
development is followed by the determination of various validation parameters.
1) Drug profile
2) Past work on the analytical aspects of the drug
3) Experimentation and results
A) Materials
i) Instrumentation
ii) Drug and internal standard
iii) Chemicals and solvents
iv) Dilutions for Calibration Curve standards and quality control samples
v) Calibration curve plasma standards and quality control plasma samples
C) Method validation
Auto sampler carry over test, screening of plasma lots and specificity, linearity,
Precision & accuracy, recovery, stability of drugs in stock solution, stability of
drugs in biological matrix, freeze-thaw stability, bench top stability, in-injector
stability, dry extract stability, long term stability, dilution integrity etc.
16
5) References
The results obtained in these experiments have been thoroughly discussed at the
end of each part. The references cited in the body of the thesis have been given at the
end of each part.
17
INTRODUCTION TO HPLC TECHINQUE
The development of any new or improved method for the analysis of an analyte
usually tailors the existing analytical approaches and instrumentation. Method
development usually requires selecting the method requirements and deciding on the
type of instrumentation1. In the development stage of an HPLC method, decision
regarding the choice of column, mobile phase, detector and method of quantitation
must be addressed.
During the optimization stage, the initial sets of conditions that have evolved
from the first stages of development are improved or optimised in terms of resolution,
peak shape, plate counts, peak asymmetry, capacity, elution time, detection limits, limit
18
of quantitation, and overall ability to quantify the specific analyte of interest. Results
obtained during optimization must be evaluated against the goals of the analysis set
forth by the analytical figures of merit. This evaluation may reveal that additional
improvement and optimization are needed to meet some of the initial method
requirements.
Optimization of the method should yield maximum sensitivity, good peak symmetry,
minimum detection and quantitation levels, a wide linearity range, and a high degree of
accuracy and precision. Other potential optimization goals include baseline resolution
of the analyte of interest from other sample components, unique peak identification,
online demonstration of purity and interfacing of computerized data for routine sample
analysis. Absolute quantitation should use simplified methods that require minimal
sample handling and analysis time.
19
SYSTEMATIC APPROACH TO THE REVERSE PHASE CHROMATOGRAPHIC
SEPARATION OF PHARMACEUTICAL COMPOUNDS
Classifying the sample
20
different mobile phases as initial choice. These conditions provide reasonable plate
number (N=8000), a run time of < 15 min for a capacity factor k < 20 and a maximum
pressure drop < 440 kgf for any mobile phase made from mixtures of water, acetonitrile
or methanol.
Mobile phase
The preferred organic solvent (B) for the mobile phase mixture is acetonitrile (ACN)
because of its favorable UV transmittance and low viscosity. However, methanol
(MeOH) is a reasonable alternative. Amine modifiers like tetra hydro furan (THF) are
less desirable because they may require longer column equilibration times, which can
be a problem in method development and routine use of the method. They may
occasionally introduce additional problems like erratic base line and poor peak shape.
However, some samples may require the use of amine modifiers when poor peak
shapes or low plate number are encountered.
Separation temperature
Mostly the temperature controllers operate best above ambient (>300C). Higher
temperature operation also gives lower operating pressures and higher plate numbers,
because of decrease in mobile phase viscosity. A temperature of 30-400 C is usually a
21
good starting point. However, ambient temperature is required if the method will be
used in laboratories that lack column thermostating.
Sample size
Initially, a 10-50 µL injection (25-50 µg) can be used for maximum detection
sensitivity. Smaller injection volumes are required for column diameters of below 4.5
mm and /or particles smaller than 5 µm. The sample should be dissolved initially in
water (1mg/mL) or dilute solution of acetonitrile in water. For the final method
development stage, the best sample solvent is the mobile phase. The samples which
cannot be dissolved in water or the mobile phase should be dissolved initially in either
acetonitrile or methanol and then diluted with water or mobile phase before injection.
The analytical column is completely equilibrated with the mobile phase before
injecting the sample for analysis and retention data are collected for interpretation. This
is done for ensuring accurate retention data. Equilibration is required whenever the
column, mobiles phase or temperature is changed during method development; usually
by flow rate at least 10 column volumes of the new mobile phase before the first
injection. Some mobile phases may require a much longer column equilibration time
(e.g. mobile phases that contain THF amine modifiers such as tri ethylamine and tetra
butylamine and any ion pair reagent).
22
A mass spectrometer is an instrument that measures the masses of electrically
charged molecules, or ions. Mass spectrometer5 (MS) is an analytical technique that is
used for the identification of unknown compounds, the quantitation of known
compounds, and the elucidation of structural information and chemical properties of
molecules, works by using magnetic and electric fields to exert forces on charged
particles (ions) in vacuum. Therefore, a compound must be charged or ionized to be
analyzed by a mass spectrometer. Furthermore, the ions must be introduced in the gas
phase into the vacuum system of the mass spectrometer. This is easily done for gaseous
or heat-volatile samples. However, many (thermally labile) analytes decompose upon
heating. These kinds of samples require either desorption or desolvation methods if
they are to be analyzed by mass spectrometry. Although ionization and
desorption/desolvation are usually separate processes, the term "ionization method" is
commonly used to refer to both ionization and desorption (or desolvation) methods.
The choice of ionization method depends on the nature of the sample and the type of
information required from the analysis. So-called 'soft ionization' methods such as field
desorption and electrospray ionization tend to produce mass spectra with little or no
fragment-ion content. Mass spectrometers measure the mass-to-charge (m/z) ratios of
gas phase ions. Creating gas phase ions is the role of the ionization method. Ionization
methods available on the instruments within the MS Facility are described below. Use
this as a guide to determine which ionization method is best suited for your sample.
23
Fig-1: Schematic diagram of Mass Spectrometer
24
into the analyser of the mass spectrometer, which is held under high vacuum. The lens
voltages are optimised individually for each sample.
Source-Dependent Parameters
Nebulizer Gas (Neb): The Neb parameter controls the nebulizer gas. The nebulizer gas
helps generate small droplets of sample flow and affects spray stability and sensitivity.
(This parameter is called Gas 1 for Q TRAP™, 4000 Q TRAP™, API 2000™, API 3200
and API 4000™ systems).
GS1: The GS1 parameter controls the nebulizer gas. The nebulizer gas helps generate
small droplets of sample flow and affects spray stability and sensitivity. (This parameter
is called nebulizer gas for API 3000™ systems.)
25
GS2: The GS2 parameter controls the auxiliary or turbo gas. It is used to help evaporate
(This parameter is called auxiliary, or turbo, gas for API 3000 systems) the spray
droplets and prevent solvent from entering the instrument.
Auxiliary Gas (Aux): The Aux parameter controls the auxiliary or turbo gas. It is used
to help evaporate the spray droplets and prevent solvent from entering the instrument.
(This parameter is called Gas 2 for Q TRAP, 4000 Q TRAP, API 2000, and API 4000
systems.)
Temperature: The temeperrature parameter controls the temperature of the turbo gas in
the Turbo Ion Spray™ source or the temperature of the probe in the heated nebulizer
(or APCI) source. It is used to help evaporate the solvent to produce gas phase sample
ions.
Curtain Gas (CUR): The CUR parameter controls the Curtain Gas™, which flows
between the curtain plate and the orifice. Curtain Gas™ prevents solvent droplets from
entering and contaminating the ion optics. The Curtain Gas™ should be maintained as
high as possible without losing sensitivity.
Ion Spray Voltage (IS): The IS parameter controls the voltage applied to the needle that
ionizes the sample in the ion source. It depends on the polarity, and affects the stability
of the spray and the sensitivity. If you are using the PhotoSpray™ source on an API
2000, API 3200, API 4000, API 5000™, Q TRAP, or 4000 Q TRAP instrument, this
parameter is called Ion Transfer Voltage.
Interface Heater (ihe): The ihe parameter switches the interface heater on and off.
Heating the interface helps maximize the ion signal and prevents contamination of the
ion optics. For API 4000, API4000 Q TRAP and API3200 systems with the Turbo V
source, the interface plate is heated to 100 °C. For the API 2000 and Q TRAP systems,
the interface plate is heated to 100 °C. (This parameter does not apply to API 150EX and
API 3000 systems.) It also does not apply to QSTAR systems.
26
Interface Heater Temperature (IHT): The IHT parameter controls the temperature of
the NanoSpray interface heater and is only available if the NanoSpray source and
interface are installed. The temperature can be adjusted up to 250 °C. (This parameter
applies to the API 4000 and 4000 Q TRAP instruments.)
Nebulizer or Needle Current (NC): The NC parameter controls the current applied to
the corona discharge needle in the APCI (atmospheric pressure chemical ionization)
probe, used in the Turbo V™ source. The discharge ionizes solvent molecules, which in
turn ionize the sample molecules.
Compound-Dependent Parameters
The available compound-dependent parameters vary with instrument type. They
consist mostly of lens elements in the ion path. Optimal values for compound-
dependent parameters do not depend on LC flow conditions. Therefore, the parameters
can be optimized using any sample introduction technique. The parameters listed here
are generally the only ones that need to be optimized. For information on working with
other parameters, refer to the online Help.
Compound-Dependent Parameters for Both Quadrupole- and LIT-mode Scans The
following parameters are available for optimization if you are running a quadrupole-
mode scan or an LIT-mode scan.
27
Collisionally Activated Dissociation Gas (CAD): The CAD parameter controls the
pressure of collision gas in the collision cell during Q2 MS, MS/MS, and LIT scans. For
Q3 MS scans, the collision gas helps to focus the ions as they pass through the collision
cell. For MS/MS scans, the collision gas acts as a target to fragment the precursor ions.
When the parent ions collide with the collision gas, they can dissociate to fragment ions.
Although this parameter is on the Source/Gas tab, this parameter is compound-
dependent and not dependent on the sample flow.
Focusing Potential (FP): The FP parameter controls the voltage applied to the focusing
ring lens. The focusing potential helps focus the ions through the skimmer region of the
mass spectrometer interface. It can induce fragmentation in the interface area, similar to
the de clustering potential. (This parameter does not apply to the API 4000, or 4000 Q
TRAP, 5000 systems.)
Entrance Potential (EP): The EP parameter controls the potential difference between the
voltage on Q0 and ground. The entrance potential guides and focus the ions through
the high pressure Qo region, EP effects the value of all the ion path voltage. The
Entrance Potential uses the Denotes ID EP.
Collision Cell Entrance Potential (CEP): The CEP parameter controls the collision cell
entrance potential, which is the potential difference between Q0 and IQ2. It focus ions in
to Q2 (collision cell). The optimal CEP gives the greatest intensity for the ions of
interest. For MS type scans, the default value is appropriate. For MS/MS scans,
optimize CEP for the precursor ion. This is used to focus and accelerate the ions into the
collision cell (Q2). It is used in Q1 and MS/MS type scans and is mass dependent. (This
parameter applies only to API 2000, 3200 and Q TRAP systems.)
Collision Cell Exit Potential (CXP): The CXP parameter controls the collision cell exit
potential, which is used to focus and accelerate the ions out of the collision cell (Q2). It
is used in Q3 and MS/MS type scans. In API 3000, API3200, API 4000, and 4000 Q
TRAP systems, CXP is the potential difference between RO2 and ST3 (the stubby lens
28
between Q2 and Q3), and is not mass-dependent. In API 2000 and Q TRAP systems,
CXP is the potential difference between RO2 and IQ3 (interquad lens 3) and is mass
dependent. The Collision Cell Exit Potential parameter uses the Access ID CXP. It is also
displayed by its Parameter ID ST3.
Rod Offset 2 (RO2): The RO2 parameter controls the potential applied to the collision
cell, Q2. In Q1 and Q3 scans, RO2 is used to focus and transmit the ions. In MS/MS type
scans, Q2 is accessed as CE.
Collision Energy (CE): The CE parameter controls the collision energy, which is the
potential difference between Q0 and Q2 for MS/MS-type scans. This is the amount of
energy that the precursor ions receive as they are accelerated into the Q2 collision cell,
where they collide with gas molecules and fragment.
Collision Cell Rod Offset: The RO2 parameter is also referred to as the Collision
Energy (CE). This parameter controls the potential applied to the collision cell (Q2). In
Q1 and Q3 scans, RO2 is used to focus and transmit the ions. In MS/MS scans CE is the
potential difference between Q0 and Q2. This is the amount of energy that the precursor
ions receive as they are accelerated into the Q2 collision cell, where they collide with gas
molecules and fragment.
Ion Energy 1(IE1): The IE1 parameter controls the potential difference between Q0 and
RO1. Although this parameter does affect the sensitivity, it has a greater impact on the
resolution of the peaks, that is, peak shape, and is considered a resolution parameter.
IE1 is used in Q1and MS/MS-type scans. In Q3 scans, the potential applied to Q1 is
called RO1 (Q1 Rod Offset) and helps to transmit ions.
Ion Energy 3 (IE3): The IE3 parameter controls the potential difference between RO2
and RO3. Although this parameter does affect sensitivity, it has a greater impact on the
resolution of the peaks, that is, peak shape, and is considered a resolution parameter.
29
IE3 is used in Q3 and MS/MS-type scans. In Q1 scans, the potential applied to Q3 is
called RO3 (Q3 Rod Offset) and helps to transmit ions.
Collision Energy Spread (CES): The CES parameter controls the spread of collision
energies used when filling the LIT and applies when you are using AutoFrag. For
example, if you have a CE of 30 and a CES of 5, collision energies of 25, 30, and 35 will
be used.
Fixed LIT Fill Time: The Fixed LIT Fill Time parameter controls amount of time that
you are filling the trap with ions. In general, the default time is appropriate. You want
to fill the linear ion trap so that you have the greatest peak intensity without saturation.
Dynamic Fill Time (DFT): The DFT parameter controls whether the LIT fill time is
dynamic. If DFT is turned on, the software will dynamically calculate the length of time
that ions are collected in the LIT. If you are using DFT, Q0 trapping is turned off by
default.
Q0 Trapping: The Q0 trapping parameter controls the storage of ions in the Q0 region,
which can increase sensitivity. This parameter has only two values: on or off. When Q0
trapping is on, ions are stored in the Q0 region, while ions are being scanned out of the
trap. If the sample is diluted, you will want to turn on Q0 trapping to increase the duty
30
cycle and get better sensitivity. If the sample is concentrated, you will want to turn off
Q0 trapping to prevent saturating the peaks, which results in poor peak resolution. If
you are using DFT, Q0 trapping is turned off by default.
Time Delayed Fragmentation Collision Energy (TDF CE): Available only for TDF
scans, the TDF CE parameter controls the collision energy that is used to fragment the
precursor ions.
Q3 Entry Barrier: The Q3 Entry Barrier parameter controls the potential difference
between RO2 and RO3. It is used to transfer the ions from Q2 into the LIT. If the
compound is fragile, you may want to decrease the value from the default to prevent
fragmentation.
Q3 Empty Time: Available only for EMC scans, the Q3 Empty Time parameter controls
the amount of time that you are removing the singly charged ions from the trap: the
greater the time, the more ions that leave the trap. If the time is too long, the multiply
charged ions will also exit the trap without detection, resulting in decreased sensitivity.
Q3 Cool Time: Available only in TDF scans, the Q3 Cool Time parameter controls the
amount of time that the precursor ions are allowed to cool prior to collecting all of their
fragment ions. As you increase this time, the number and intensity of fragment ions
decreases. In general, the default Q3 Cool Time value is sufficient.
Multi-Charge Separation (MCS) Barrier: The MCS Barrier parameter controls the
voltage used to remove singly charged ions from the LIT and is only available for EMC
scans. This parameter applies only to the 4000 QTRAP system
31
Detector Parameters:
In a quadrupole mode scan, the following parameters are available for optimization.
Channel Electron Multiplier (CEM): The detector degrades with time and the CEM
value should periodically be readjusted using the standard positive PPG calibrant. Do
not change the voltage unless the detector has been replaced or there has been a
reduction of sensitivity. A typical initial value is 2000V. A typical end-of-life value is
3000V.
Deflector (DF): The deflector shows no mass or energy dependence and optimizes over
at least a 100 V plateau. Each detector has its own optimum value. On some systems,
there may be a different optimization value for MS and MS/MS scans.
Note: Over time the detector begins to wear and require a greater voltage for the same
performance. Therefore, adjusting the detector voltage is an important part of ensuring
maximum sensitivity.
32
Also as the aqueous composition of the carrier solvent increases at high flow
rates (2 mL/ min), the more visible the spray becomes and the further away from the
orifice it should be directed. Refer to the fig-1. TurboIonSpray positioning across the
Orifice, where the areas indicated in the figure for the different flow rates are the
optimum target areas for the TurboIonSpray liquid spray. The circle immediately
around the orifice (for example the part of the orifice plate which is visible when
viewing the front of the interface) should remain clear of solvent or solvent drops at all
times. The best position is usually a few millimeters off axis to the left of the curtain
plate aperture. Multiply charged proteins and peptides introduced at a few micro liters
per minute usually require the sprayer to be less than 1 cm from the Curtain Plate.
33
flow until the signal starts to decrease). In order to prevent instrument contamination
the Curtain Gas flow should be optimized at the highest possible setting but never
below six that does not result in a significant reduction in signal intensity. Refer to the
System Reference Manual for further details of Vacuum Interface operation.
Turbo Temperature
The quantity and type of sample affects the optimal TurboIonSpray temperature.
At higher flow rates the optimal temperature increases. As the organic content of the
solvent increases the optimal probe temperature should decrease with solvents
consisting of 100 percent methanol or acetonitrile the probe performance may optimize
as low as 300°C. Aqueous solvents consisting of 100 per cent water at flows
approximately 1mL/min require a minimum probe temperature of 425°C. Normal
optimization is usually performed in increments of 25°C. The TurboIonSpray is
normally used with sample flow rates of 5 µL/min to 2000 µL/ min. The heat is used to
increase the rate of evaporation and this improves ionization efficiency resulting in
increased sensitivity.
34
De clustering Potential (DP) and Focusing Potential (FP) Voltages
Optimal de clustering potential and focusing potential operating conditions with
the TurboIonSpray source should be set high enough to reduce the chemical noise but
low enough to avoid fragmentation. Start with the de clustering Potential DP) at 300V
and the Focusing Potential (FP) at 30V.
Solvent Composition
Commonly used solvents and modifiers are acetonitrile, methanol, propanol,
water, acetic acid, formic acid, ammonium formate and ammonium acetate. The
modifiers such as TEA, sodium phosphate, TFA and dodecyl sodium sulfate are not
commonly used because they omplicate the spectrum with their ion mixtures and
cluster combinations. They may also uppress the strength of the target compound ion
signal. The standard concentration of ammonium formate or ammonium acetate is from
2 to 10 mmol per liter for positive ions and 2 to 50 mmol per liter for negative ions. The
concentration of the organic acids is 0.01% to 0.5% by volume.
35
Polarity selection in multiple reaction monitoring mode (LC-MS/MS)
In the case of method development for the known compound it is necessary to
consider the chemical structure of the molecule. Based on the functional groups
presented in the molecule the polarity of the compound is decided for the mass
spectrometer study. For example, if the compound is having basic functional groups
such as primary and secondary amines amides etc. it will accept a proton from the
solution and its molecular weight increases. Thus, for compounds with basic functional
groups positive polarity is selected in the multiple reaction monitoring (MRM).
36
METHOD DEVELOPMENT AND PLASMA EXTRACTION PROCEDURE
37
important if very low amounts of the analytes are present in the samples. Extracting
hydrophilic compounds from these aqueous matrices is an analytical challenge. Blood
contains many components, including a variety of proteins, fats, salts and suspended
cells. The red blood cells can be removed from the plasma by centrifugation after
addition of an anti-coagulant. The simplest form of sample preparation for this kind of
samples involves dilution, centrifugation, filtration and/or evaporation. Some
commonly used techniques for sample preparation, especially for biological fluid clean-
up are briefly described below.
Protein precipitation
When a drug strongly binds to the plasma proteins (in case of plasma samples) it
is often difficult to extract the drug from plasma by any means. Then protein
precipitation followed by extraction is only the process to extract the drug from plasma
samples. This separation technique removes proteins from the samples by denaturating
them directly. The protein precipitation is usually done by the addition of a water
miscible organic solvent (e.g. methanol, ethanol, acetonitrile or acetone) or a strong acid
such as trichloroacetic acid. The denatured proteins are then removed from the sample
by centrifugation. Efficient centrifugation will give clear and safe samples for injection.
38
solvent should match the analytes polarity while still being immiscible with water and
it should preferably be compatible with the following detection method. A larger
volume of the extraction solvent should be higher compared to the sample. However,
the sample extract can easily be evaporated if a volatile solvent is used to increase the
analyte concentration. Other factors, such as pH and ionic additives may affect the
extraction efficiency.
39
METHOD VALIDATION PARAMETERS AND THEIR ACCEPTANCE CRITERIA
Introduction
Method validation is the process of proving that an analytical method is
acceptable for its intended purpose. The methods are followed by the guidelines of
International Conference on Hormonization (ICH), and Food and Drug Administration
2- 4 (FDA). This guideline introduces the validation terms as defined by the ICH and the
purpose of these guidelines is to details the validation data necessary for High
Performance Liquid Chromatographic (HPLC), and Liquid Chromatography connected
with Mass spectrometer (LC-MS/MS).The validation parameters to be given below.
Selectivity
At least 6 different blank plasma lots were screened for the interference at the
retention times of analyte(s) and internal standard using specified extraction procedure
and chromatography. Spiked one LLOQ level from each plasma lot and extracted the
LLOQ sample as per the extraction procedure. The percent interference was calculated
for endogenous components present in plasma at the retention times of the analyte and
the IS.
Acceptance criteria
The interfering peaks at the retention time of the analyte must be < 20% of the
respective plasma blanks extracted mean LLOQ peak area. Response of interfering
peaks at the retention time of internal standard must be < 5% of the respective mean
response of internal standard in LLOQ sample. At least 80% of the blank screened
matrix lots should be meets the above acceptance criteria.
40
Precision and accuracy (within batch and global)
A minimum of 4 P&A batches have to be analyzed.
The Precision & Accuracy batches are organized in the following manner:
Reconstitution solution / Mobile Phase x 1
Standard blank (without Analyte, Internal Standard) x 1
Standard zero (with internal standard) x 1
6 – 8 non-zero CC standards (LLOQ and ULOQ)
LLOQ QC, LQC, MQC, HQC
The above mentioned set of QC samples (LLOQ QC, Low QC, Middle QC and
High QC) was injected for 6 times. The calibration curve and back calculated the
concentrations of quality control samples are generated. The precision and accuracy at
each concentration level of QC samples are then determined (both within batch and
global).
Acceptance criteria
1. Lower Limit of Quantification (LLOQ)
The lowest standard on the calibration curve should be accepted as the limit of
quantification if the following conditions are met:
The analyte response at the LLOQ should be at least 5 times the response compared to
the blank response. Analyte peak response should be identifiable, discrete, and
reproducible with a precision of 20% and an accuracy of 80-120%.
With respect to the calibration curve 75% or a minimum of 6 out of 8 non-zero
standards should be used to construct the calibration curve including the LLOQ and the
ULOQ. The back calculated concentration should fall within ±15% of the nominal value
for all the calibration standards except LLOQ where it can be ±20%.
Precision:
41
The within and between batch %CVs for low, medium and high concentrations
should be within 15% except LLOQ QC for which %CV should not exceed by more than
20%.
Accuracy:
The within and between batch mean value should not deviate by more than 15%
of the nominal value at low, medium and high QC concentrations except LLOQ QC
where it should not be more than 20%.
Recovery:
To evaluate the recovery of the analyte (s) and the internal standard (IS) the
respective peak areas are used. In case of combination of drugs, all the analytes must be
present in extracted samples as well as in unextracted samples. Six low, medium, and
high quality control samples from the freezer are retrieved and processed as per
extraction method and then injected. Unextracted quality control samples (post spiked,
spiked with internal standard) were prepared from the stock solutions having a
concentration equivalent to that of extracted samples and were injected.
The recovery was calculated from the mean peak response of extracted samples
and the unextracted samples. The percent recovery at each concentration of LQC, MQC
and HQC levels and the overall mean recovery were computed. Similarly, the percent
recovery of the IS was estimated at MQC level only. The percent recovery of analyte(s)
was determined by using the formula
42
Acceptance criteria:
The percent recovery of the analyte and the internal standard should not be more than
115%. The CV for the % recovery of analyte across LQC, MQC and HQC levels should
be ≤ 15%.
Stabilities
The stock solutions of the analyte and the IS for stability evaluation were prepared in an
appropriate solvent.
The following stability experiments were conducted.
1. Stock solution stability (short term and long term)
2. Auto injector stability or in-injector stability
3. Dry extract stability
4. Bench top stability
5. Freeze thaw stability
6. Long-term stability of Analyte (s) in matrix
Short-term stability
From the fresh stock solution approximately one mL aliquot is taken into a pre
labeled tube and kept on bench for short term stock stability. After a minimum of 6
hours dilutions from this solution are prepared and injected in 6 replicates. Dilutions
from the original stock solution, which is kept in the refrigerator are made and injected
(n=6). The stability is assessed by comparing the mean response of the stability samples
against the mean response of the comparison samples.
43
Acceptance criteria:
The mean peak response of the freshly prepared stock solution of analyte or internal
standard versus the comparable stored solution should be within the range of 90-110%.
Mean peak response of stability stock
% Stability of stock solution = ---------------------------------------------- X 100
Mean peak response of fresh stock
The short-term stability time is computed by subtracting the time when the stability
stock was kept on the work bench from the time when the fresh (comparison) stock is
retrieved from refrigerator.
Acceptance criteria:
The mean peak response of the freshly prepared stock solution of analyte or internal
standard versus the comparable stored solution should be within the range of 90-110%.
Mean peak response of stability stock
% Stability of stock solution = ---------------------------------------------- X 100
Mean peak response of fresh stock
The Long-term stock solution stability time is computed by subtracting the time when
the stability stock was kept in the refrigerator from the time when the stock is retrieved
from refrigerator.
44
Six replicates of low and high QC samples are processed and loaded into the
auto injector for a minimum period of 24 hours. After the auto injector storage period,
the stability samples are analyzed against freshly spiked calibration curve standards
and 6 replicates of LQC, HQC samples (comparison samples).
The time interval specified is only an example. The time intervals are selected on
the basis of anticipated analytical batch run time. The Mean, SD, %CV and %nominal
are calculated for the stability samples as well as the comparison samples, %CV and
%nominal should be within 15%. The percent in-injector stability is assessed by using
the formula
Acceptance criteria
The percent stability of stability samples should be within 85-115%. The in-injector
stability time is calculated by subtracting the time when the stability QC samples are
loaded in to the auto sampler from the time when the fresh CC and QC samples are
loaded in to the auto sampler for the assay.
45
Mean concentration of stability samples
% Stability = ---------------------------------------------------------- x100
Mean concentration of comparison samples
Acceptance criteria
The percent stability of stability samples should be within 85-115%. The dry
extract stability time is calculated by subtracting the time when the stability samples are
loaded into the refrigerator from the time when they are retrieved from the refrigerator
for the assay.
Acceptance criteria
46
The percent stability of stability samples should be within 85-115%. The bench top
stability time is calculated by subtracting the time when the stability samples were kept
on bench for thawing from the time when the processing of these samples was started.
Freeze-thaw stability
In Freeze- thaw stability, minimum three cycles are to be conducted. All the six
replicates of LQC and HQC samples after three freeze-thaw cycles are processed along
with freshly prepared calibration curve standards and comparison LQC and HQC
samples. If the third freeze thaw cycle is not stable, then the first and the second freeze
thaw samples are processed. If an analyte is unstable at the storage temperature of –
200C, then stability sample should be frozen at -700C during the three freeze and thaw
cycles.
Three sets of samples (containing six each at LQC and HQC levels) are chosen for
the freeze-thaw stability studies. The samples are initially stored in a deep freezer at -
700C. After a minimum storage of 24 hrs these samples are taken out and allowed to
thaw at room temperature. After the thawing is complete (around 45 min) the samples
are kept back in the freezer.
After a minimum of 12 hours freezing, retrieve two sets of the samples from the
deep freezer and allowed to thaw again (second freeze thaw cycle) and the samples are
again stored in the deep freezer. After a minimum of 12 hours freezing, retrieve one set
of the samples and allowed to thaw. This completes three freeze thaw cycles. Now the
samples are processed along with freshly spiked CC standards and freshly spiked LQC
and HQC samples (comparison samples). The freshly prepared calibration standards,
comparison samples and stability samples are analyzed. The calibration curve is
generated and the concentrations are back calculated.
The Mean, SD, %CV and %nominal are calculated for the stability samples as
well as the comparison samples, %CV and %Nominal should be within 15%. The
percent freeze thaw stability is assessed by using the formula
47
Mean concentration of stability samples
% Stability = ---------------------------------------------------------- x 100
Mean concentration of comparison samples
Acceptance criteria
The percent stability of stability samples should be within 85-115%.
48
when the stability samples were logged into the freezer/cold room from the date and
time when they were retrieved from the freezer/cold room for processing.
Dilution integrity
The dilution integrity is assessed by assaying six diluted quality control (DQC)
samples (spiked in screened blank matrix having less than twice the concentration of
ULOQ) diluted by a factor of 1/2 and six DQC samples diluted by a factor of 1/4 with
screened blank matrix prior to extraction, against calibration curve standards.
Acceptance criteria
The mean concentration obtained for dilution integrity QCs should be within
±15% of their nominal concentration and the %CV should not exceed 15%.
Acceptance criteria
The percent response ratio at LQC and HQC level should be within 85-115%.
49
THE VALIDATION PARAMETERS AND THEIR ACCEPTANCE CRITERIA
Parameters Acceptance Criteria
Response of interfering peaks at the retention time of analyte (s) must
be ≤ 20% of the mean response of LLOQ standard. Response of
Screening of plasma
interfering peaks at the retention time of IS must be ≤ 5% of the mean
lots and specificity
response of Internal standard. At least 80% matrix lots should meet the
above criteria
A minimum of 6 out of 8 standards, including LLOQ and ULOQ shall
Calibration
fall within ± 15% except LLOQ for which it shall be within ± 20% when
Curve
back calculated. Regression factor should be > 0.98 (r).
Percent recovery for analyte or IS should be ≤ 115%. The %CV of the %
Recovery
Recovery for the analyte at L, M and H QC levels shall be ≤ 15%.
A minimum of 4 P & A batches shall be evaluated. The precision shall
Precision (%CV) not exceed 15% for all the QC samples except for the LLOQ QC where
it shall not exceed 20%.
A minimum of 4 P & A batches shall be evaluated. The mean value
Accuracy
shall be within ± 15% for all QC samples except for the LLOQ QC,
(% Nominal Conc.)
where it shall not deviate by more than ± 20%.
Short term stock
Short term stability against comparison samples for analyte and
solution stability
internal standard shall be within the range of 90-110%.
The mean concentration obtained for LQC & HQC samples should be
In injector stability within ± 15% of nominal concentration and the %CV shall not exceed
15%. The % Stability when compared with comparison samples shall be
within ± 15%.
The mean concentration obtained for LQC & HQC samples should be
Dry extract
within ± 15% of nominal concentration and the %CV shall not exceed
stability
15%. The % Stability when compared with comparison samples shall be
within ± 15%.
The mean concentration obtained for LQC & HQC samples should be
Bench top stability within ± 15% of nominal concentration and the %CV shall not exceed
15%. The % Stability when compared with comparison samples shall be
within ± 15%.
The mean concentration obtained for LQC & HQC samples should be
Freeze-thaw within ± 15% of nominal concentration and the %CV shall not exceed
stability 15%. The % Stability when compared with comparison samples shall be
within ± 15%.
50
THE VALIDATION PARAMETERS AND THEIR ACCEPTANCE CRITERIA
(contd.)
The mean concentration obtained for LQC & HQC samples should be
Long term stability within ± 15% of nominal concentration and the %CV shall not exceed
in plasma 15%. The % Stability when compared with comparison samples shall
be within ± 15%.
The mean concentration obtained for LQC & HQC samples should be
within ± 15% of nominal concentration and the %CV shall not exceed
Dilution Integrity
15%. The % Stability when compared with comparison samples shall
be within ± 15%.
Accuracy should be within ±15% for all QC samples except for the
LLOQ QC, where it should not deviate by more than ± 20%.
Matrix effect
Precision should not exceed 15% for all the QC samples except for the
LLOQ QC where it shall not exceed 20%.
References
1. Lloyd R. Snyder, Joseph J. Kirkland and Joseph L. Glajah Practical HPLC method
development 2nd Edition, New York, 1997.
2. http://bebac.at/Guidelines.htm.
3. http://www.fda.gov/Drugs/GuidanceComplianceRegulatoryInformation/Gui
dances/default.htm.
4. http://www.emea.europa.eu/docs/en_GB/document_library/Scientific_guidel
ine/2009/12/WC500018062.pdf.
51