Professional Documents
Culture Documents
Johnson 1992
Johnson 1992
In vivo evaluation of an
electroenzymatic glucose sensor
implanted in subcutaneous tissue
Medical Devices and Diagnostics Division, Lilly Laboratory for Clinical Research, Eli Lilly & Co, Lilly Corporate
Center, Indianapolis, IN 46285, USA
(Received 1 April 1992; revised version received 1 September 1992; accepted 17 September 1992)
710
Biosensors & Bioektronics Evaluation of an electroenzymatic glucose sensor
IiV WO (RABBIT) EV-UATION change in the plasma glucose values. This lag was
thought to be a physiologic phenomenon and not
New Zealand White rabbits, with venous and related to the response time of the sensor. This
arterial cannulas surgically implanted, were used study was conducted 24 h after implantation of
for these studies. The sensors were implanted in the sensor and initial polarization. The
the subcutaneous tissue between the scapulas or calibration parameters from the in vitro calibration
above the lumbar muscle area near the most appear still to be valid, suggesting no appreciable
posterior rib and sutured to the skin. A custom drift or loss of sensitivity due to encapsulation.
connector was used to make electrical contact This type of study also confirms the subcutaneous
with the three leads of the sensor. Prior to tissue as a viable and accurate location in which
implantation, a two-point in vitro calibration was to monitor steady-state glucose values as an
performed with the sensor in the manner indicator of actual blood glucose concentrations.
mentioned previously by exposing the sensor to Experiments conducted in our laboratories,
solutions with glucose concentrations of 100 and incorporating a microdialysis probe implanted in
200 mg dl-’ following a settling period of the subcutaneous tissue, have also verified this
approximately 1 h. The calibration equation correlation (Phebus ef al., 1991).
derived from this study was used to convert the
in vivo current values from the sensor to glucose
concentration values. 11v IWO (HUMAN) EVALUATION
Arterial blood samples were collected inter-
mittently throughout the duration of the glucose This study was designed to assess the feasibility of
infusion study. The plasma was separated from safely and accurately monitoring blood glucose
the sample and analyzed by a commercial clinical levels continuously for 72 h via the glucose sensor
chemistry analyzer. The current output from the implanted in the subcutaneous tissue of the
sensor was recorded at regular intervals. A sterile abdomen. These tests were conducted on six male
glucose solution was infused into the venous volunteers without diabetes, with an individual
cannula to elevate the rabbit’s blood glucose sensor implanted in each of them. A parallel
levels. comparison of data from the glucose sensor, data
Figure 2 illustrates the change in plasma from a Biostator@’ instrument which analyzed
glucose values during the continuous intravenous venous whole blood continuously, and capillary
infusion of a glucose solution (20 mg glucose blood glucose measurements analyzed by a YSI
min-’ kg-‘). The output from the sensor follows 2300G STAP was performed.
this trend very well, with a slight lag of approxi- The sensors implanted in the volunteers were
mately 20 min visible during periods of rapid sterilized by electron beam irradiation. The
portion of the sensor containing the electrodes
was aseptically inserted into the subcutaneous
STOP INNSION
tissue of the abdomen following the creation of a
tunnel in the subcutaneous tissue with an 18
gauge hypodermic needle. The site of implantation
was perfused with xylocaine, a local anesthetic. A
battery powered potentiostat, fabricated in our
laboratory, was connected to the sensor and a
potential of +0*6 V was applied. A portable data
logger was connected to the analog output from
the potentiostat to collect and store data at 10 s
l PLASMA GLUCOSE (mg/dL)
START INNSION
- SENSOR GLUCOSE intervals. This electronic system allowed the
4 volunteer to be ambulatory during the test period,
-50 0 50 100 150 200 250 300 350
except during Biostator use.
TIME (MIN) After an average settling time of approximately
Fig. 2. The change in glucose concentration of a rabbit 2.5 h, a venous cannula was placed in the
during the constant infusion of a glucose solution as volunteer and conncected to the Biostator. This
measured by the sensor in the subcutaneous tissue and by device continuously removed blood and measured
analysis of arterial plasma samples. the glucose concentration. The results were stored
711
K. W. Johnson et al. Biosensors & Bioelectronics
712
Biosensors & Bioelectronics Evaluation of an electroenzymatic glucose sensor
713
K. WVJohnson et al. Biosensors h Bioelectronics
R J. (1989). ExtendedAbstracts - 175th Mtg Electro- D. &Andrew, C. C. (1992). Sensors and Actuators B,
them. Sot., 89(l), pp. 921-2. 5(1-4), 139-44.
Koudelka, M., Rohner-Jeanrenaud, F., Terrettaz, J., Phebus, L. A., Morff, R. J., Johnson, K. W. & Bryan, N.
Bobbioni-Harsch, E, DeRooij, N. F. & Jeanrenaud, (1991). Current Separations, 10(4), 133-6.
B. (1991). Biosensors and Bioelectronics, 6, 3 l-6. Shichiri, M., Kawamori, R., Hakui, N., Yamasaki, Y. &
Mastrototaro, J. J., Johnson, K W., Morff, R. J., Lipson, Abe, H. (1984). Diabetes, 33, 1200-2.
714