Professional Documents
Culture Documents
Csoeregi1994 Design, Characterization, and One-Point in Vivo Calibration of A Subcutaneously Implanted Glucose Electrode PDF
Csoeregi1994 Design, Characterization, and One-Point in Vivo Calibration of A Subcutaneously Implanted Glucose Electrode PDF
1994,66, 3131-3138
A 0.29-mm-diameter flexible electrode designed for subcutane- include (a) safety, (b) clinical accuracy and reliability, (c)
ous in vivo amperometric monitoring of glucose is described. feasibility of in vivo recalibration, (d) stability for at least one
The electrode was designed to allow “one-point” in vivo hospital shift of 8 h, (e) small size, ( f ) ease of insertion and
calibration, Le., to have zero output current at zero glucose removal, and (g) a sufficiently fast response to allow timely
concentration, even in the presence of other electroreactive intervention.
species of serum or blood. A valid zero point, along with a Keys to safety are absence of leachable components,
measurement of the glucose concentration in a withdrawn biocompatibility, and limiting of the potentially hazardous
sample of blood at which the current is known, defined the foreign matter introduced into the body to an amount that is
sensitivity in the linear response range. The electrode was inconsequential in a worst case failure. The clinical accuracy
four-layered, with the layers serially deposited within a 0.125-
(1) Armour, J. C.; Lucisano, J. Y.; McKean, B. D.; Gough, D. A. Diabetes 1990,
mm recess upon the tip of a polyimide-insulated 0.25-mm gold 39, 1519-1526.
wire. The recessed structure reduced the sensitivity to (2)Shichiri, M.; Kawamori, R.; Yamasaki, Y. Merhods Enzymol. 1988, 137,
326-334.
movement and allowed, through control of the depth of the (3)Shichiri, M.; Yamasaki, Y.; Nao, K.; Sekiya, M.; Ueda, N. In Zmplanrable
recess, control of the transport of glucose and thus the range Glucose Sensors-The Srare of rhe Art; Pfeiffer, E. F., Reaven, G. M., Eds.;
Supplement Series/Hormone and Metabolic Research; (Reisensburg, 1987)
of linearity. The recess contained the four polymeric layers, Thieme Medical Publishers, Inc.: New York, 1988;pp 17-20.
with a total mass less than 5 pg and no leachable components. (4) Moatti-Sirat, D.;Capron, F.; Poitout, V.; Reach, G.; Bindra, D. S.; Zhang,
Y.; Wilson, G. S.; Thevenot, D. R. Diaberologia 1992,35,224230.
The bottom glucose concentration-to-current transducing layer (5) Karube, 1.; Yokoyama, K.; Tamiya, E. Biosens. Bioelecrron. 1993,8, 219-
consisted of the enzyme “wiring” redox polymer poly- 228.
(6) Brilckel, J.; Kerner, W.; Zier, H.;Steinbach, G.; Pfeiffer, R. F. Klin.
[(~inylimidazole)Os(bipyridine)2Cl]+~~+, complexed with re- Wochenschr. 1989,67,491495.
combinant glucose oxidase and cross-linked with poly(ethy1ene (7)Gernet, S.; Koudelka, M.; Rwij de, N. F. S e w . Acruarors 1989,17,537-540.
(8)Pickup, J. C.; Claremont, D. J.; In Zmplanrable Glucose Sensors-The Srare
glycol) diglycidyl ether, to form an electron-conducting ojthe Art;Pfeiffer, E. F., Reaven, G. M., Eds.;Supplement Series/Hormone
hydrogel. The layer was overcoated with an electrically and Metabolic Research; (Reisensburg, 1987)Thieme Medical Publishers,
Inc.: New York, 1988;pp 3636.
insulating layer of polyaziridine-cross-linked poly (allylamine), (9)Yokoyama, K.; Lee, S. M.; Tamiya, E.; Karube, I.; Nakajima, K.; Uchiyama,
on which an immobilized interference-eliminating horseradish S.; Suzuki, S.; Akiyama M.; Masuda, Y. Anal. Chim. Acra 1992,263,101-
110.
peroxidase based film was deposited. An outer biocompatible (10)Tamiya, E.; Karube, I. Sens. Acruarors 1988,15, 199-207.
layer was formed by photo-cross-linking derivatized poly- (11) Bindra, D.S.; Zhang, Y.; Wilson, G. S.; Sternberg, R.; Thevenot, D. R.;
Moatti, D.; Reach, G. Anal. Chem. 1991,63,1692-1696.
(ethylene oxide). The current output of a typical electrode at (12)Cronenberg, C.; Grocn van, B.; Beer de, D.; Heuvel van den, H. Anal. Chim.
10 mM glucose and at 37 OC was 35 nA, the apparent K, was Acta 1991, 242,275-278.
(13) Johnson, K. W.; Mastrototaro, J. J.; Howey, D. C.; Brunella, R. L.; Burden
20 mM, and the 10-90% response time was -1 min. The Brady, P. L.; Bryan, N. A.; Andrew, C. C.; Rowe, H. M.; Allen, D. J.; Noffke,
sensitivity varied only by f5% over a 7 2 4 test period. The B. W.; McMahan, W. C.; Morff, R. J.; Lipson, D.; Nevin, R. S . Biosens.
Bioelectron. 1992,7, 709-714.
electrode tracked the blood glucose concentration in a rat model (14)Koudelka, M.; Gernet, S.;Rwij de, N. F. Sens. Acruarors 1989,18,157-165.
before, during, and following intraperitoneal glucose infusion. (15)Shaw, G. W.; Claremont, D. J.; Pickup, J. C. Biosens. Bioelectron. 1991, 6,
401406.
Two failure modes were observed. The first, deactivation of (16)Pickup, J. C.; Shaw, G. W.; Claremont, D. J. Diaberologia 1989,32,213-217.
hydrogen peroxide-producing lactate oxidase in the interfer- (17)Gregg, B.; Heller, A. Anal. Chem. 1990,62,258-263.
ence-eliminating layer, resulted in inadequate preoxidation of (18) Gregg, 8.; Heller, A. J. Phys. Chem. 1991, 95,5976-5980.
(19)A h , T.; Lau, Y. Y.; Ewing, A. G. AMI. Chem. 1992,64,2160-2163.
interferants. The second was an abrupt drop in the sensitivity (20)Bartlett, P. N.; Caruana, D. J. Analyst 1992,117, 1287-1292.
(21)Chen, C.-Y.; Gotoh, M.; Makino, H.;Su, Y.-C.; Tamiya, E.; Karube, I. Anal.
of implanted electrodes, -7 h after implantation. The failed Chim. Acra 1992,265, 5-14.
electrodes promptly regained their sensitivity in buffer. (22)Kawagoe, J. L.; Niehaus, D. E.; Wightman, R. M. Anal. Chem. 1991, 63,
2961-2965.
(23)Murakami, T.; Nakamoto, S.; Kimura, J.; Kuriyama, T.; Karube, I. Anal.
In response to the need for frequent or continuous in vivo Lett. 1986,19,1973-1986.
(24)Pishko, M. V.;Michael, A. C.; Heller, A. Anal. Chem. 1991,63,2268-2272.
monitoring of glucose in diabetics, particularly in brittle (25)Tamiya, E.;Sugiura, Y.; Akiyama, A.; Karube, I. Ann. N.Y. Acad. Sci. 1990,
diabetics, a range of possible in vivo glucose electrodes have 613. 396-400.
been Thedesired characteristics of theseelectrodes (26)Turner, R. F. B.; Harrison, D. J.; Rajotte, R. V.; Bakes, H. P. Sens. Acruarors
B 1990,21, 561-564.
(27)Urban, G.; Jobst, G.; Keplinger, F.; Aschauer, E.; Tilado, 0.; Fasching, R.;
+ University of Lund. Kohl, F. Biosens. Bioelectron. 1992,7,733-739.
t University of Texas at Austin. (28)Velho, G.; Froguel, P.; ThCenot, D. R.; Reach, G. Diabetes Nurr. Merab.
I University of Uppsala. 1988,I , 227-234.
0003-2700/94/0366-3131$04.50/0 Analytical Chemistry, Vol. 66, No. 19, October 1, 1994 3131
0 1994 American Chemical Society
must be such that even when the readings are least accurate, sensing site. Good correlation was observed between intra-
the clinical decisions based on these be still correct. Feasibility vascular and subcutaneous glucose concentration^.^^,^^ They
of prompt confirmation of proper functioning of the sensors also demonstrated the need for in vivo sensor ~ a l i b r a t i o n . ~ ~
and of periodic in vivo recalibration is of the essence if a Another approach to in vivo glucose monitoring was based on
physician is to allow the life of a patient to depend on the coupling subcutaneous microdialysis with electrochemical
sensor. This one-point calibration, relying on the signal at detection.46 To control and adjust the linear response range,
zero glucose concentration being zero and measuring the blood electrodes have been made glucose-diffusion limited, usually
glucose concentration at one point in time, along with the through glucose transport limiting membranes.
signal, is essential, but has heretofore been elusive. The Diffusional mediators, through which the 02 partial
sensitivity must be sufficiently stable for the frequency of the pressure dependence of the signals is reduced, are leached
required in vivo recalibration to not be excessive. The sensor from sensors. Such leaching introduces an unwanted chemical
must be small enough to be introduced and removed with into the body and also leads to loss in sensitivity, particularly
minimal discomfort to the patient and for minimal tissue in small sensors. In microsensors, in which outward diffusion
damage. It is preferred that the sensor be subcutaneous and of the mediator is radia1,47.48the decline in sensitivity is rapid.
that it be inserted and removed by the patient or by staff in This problem has been overcome in “wired” enzyme electrodes,
a physician’s office. Finally, its response time must be fast i.e., electrodes made by connecting enzymes to electrodes
enough so that the corrective measures, when needed, will be through cross-linked electron-conducting redox hydrogels
timely. Glucose oxidase has been wired with
In response to some of these needs, needle polyelectrolytes having electron relaying [Os(bpy)2C1]+/2+
typel.2A6.8.11-13,16,19,22,23,25-27,29-32and other subcutaneous am-
redox centers in their backbones. Hydrogels were formed
perometric sensors were considered. The majority of these upon cross-linking the enzyme and its wire on electrodes. These
utilized platinum, platinum-iridium, or platinum black to electrodes had high current densities and operated at a potential
electrooxidize H202 generated by the glucose oxidase (GOx) of 0.3 V vs SCE.17*50The electrooxidizable interferants were
catalyzed reaction of glucose and oxygen.2,5q6J1-13~16~19~26,27*30 eliminated through peroxidase-catalyzed preoxidation in a
In these sensors, the GOx was usually in large excess and second, nonwired, hydrogen peroxide generating layer on the
immobilized, often by cross-linking with albumin and wired enzyme ele~trode.~5.36
g l ~ t a r a l d e h y d e . ~ ~ To
- ~ ~exclude
? ~ ~ , ~electrooxidizable
~ inter- In this paper, we report on a 0.29-mm recessed gold wire
ferants, membranes of cellulose acetate and sulfonated electrode for subcutaneous in vivo glucose monitoring, ap-
polymers including Nafion were ~sed.~~,~6,33,34 Particular proaching in its performance all of the above listed require-
attention was paid to the exclusion of the most common ments, including in vivo one-point calibration. The electrode
electrooxidizable interferants: ascorbate, urate, and acetami- was constructed by depositing four layers into a recess. The
nophen. Also, to cope with the interferants, two-electrode recess was formed by etching away gold from an insulated
differential measurements were used, one electrode being gold wire. The polymer layers were protected within the recess
sensitive to glucose and electrooxidizable interferants and the against mechanical damage. The recess and its polymer layers
other only to i n t e r f e r a n t ~ .A~ ~strategy for overcoming the also reduced the transport of glucose to the gold contacting
problem of interferants, applied also in this work, involved sensing layer. By limiting the glucose flux, the desired linear
their p r e ~ x i d a t i o n . ~To
~ . ~make
~ the electrodes more bio- response range, spanning the clinically relevant glucose
compatible, hydrophilic polyurethanes,2.6*11.37,38 poly(viny1 concentration range, was obtained. The sensor had no
a l ~ o h o l ) and
, ~ ~ o ~ ~ H Emembranes M A ~ ~have been used.
Several researchers tested GOx-based glucose sensors in
vivo and obtained acceptable results in rats,4*30,37*42 rabbit~,13?~3
-
leachable components, and its four cross-linked polymer layers
contained only 5 pg of immobilized material and only a few
nanograms of polymer-bound osmium. Preoxidation of the
d 0 g ~ , 2 , pigs,8-41142
~ + ~ ~ , ~sheep,6and h u m a n ~ . l ~These
, ~ ~ studies
>~ interferants in one of the four layers made possible one-point
validated the subcutaneous tissue as an acceptable glucose- in vivo calibration of the sensor.
(29) Wang, J.; Angnes, L. Anal. Chem. 1992,64, 456-459.
(30) Koudelka, M.; Rohner, J. F.; Terrettaz, J.; Bobbioni, H . E.; Rooij de, N . F.;
Jeanrenaud, B. Biosens. Bioelectron. 1991, 6, 31-36. EXPERIMENTAL SECTION
(31) Velho, G.; Froguel, P.; Sternberg, R.; Thevenot, D. R.; Reach, G . Diabetes Materials. The electrodes were made of a polyimide-
1989, 38, 164-171.
(32) Pishko, M. V.; Katakis, I.; Lindquist, S.-E.; Heller, A. Mol. Cryst. Liq. Crysr. insulated 250-pm-diameter gold wire, 290-pm outer diameter
1990, 190, 221-249. (0.d.) (California Fine Wire Co., Grover City, CA). Heat-
(33) Shiono, S.;Hanazato, Y.; Nakako, M. Anal. Sei. 1986, 2, 517-521.
(34) Gorton, L.; Karan, H. I.; Hale, P. D.; Inagaki, T.; Okamoto, Y.; Skotheim, shrinkable tubing (RNF 100 3/64 in. BK and ‘ / I 6 in. BK,
T.A. Anal. Chim. Acta 1990, 28, 23-30. Thermofit, Raychem, Menlo Park, CA) and a two-component
(35) Maidan, R.; Heller, A. Anal. Chem. 1992, 64, 2889-2896.
(36) Maidan, R.; Heller, A. J . Am. Chem. SOC.1991, 113, 9003-9004.
(37) Koudelka, M.; Rohner, J. F.; Terrettaz, J.; Bobbioni, H. E.; Rooij de, N . F.; (44) Shichiri, M.; Asakawa, N.; Yamasaki, Y.; Kawamori, R.; Abe, H. Diabetes
Jeanrenaud, B. Biomed. Biochim. Acta 1989.48.953-956. Care 1986, 9, 298-301.
(38) Velho, G.; Froguel, P.; Thevenot, D. R.; Reach, G. Biomed. Biochim. Acta (45) Kerner, W.; Zier. H.; Steinbach, G.; Briickel, J.; Pfeiffer, E. F.; Wiess, T.;
1989, 48, 951-964. Camman, K.; Plack, H. In Implantable Glucose Sensot-The State of the
(39) Rebrin, K.; Fischer, U.; Woedtke, T.; Abel, P.; Brunstein, E. Diabetologia Art; Pfeiffer, E. M., Raven, G. M., Eds.; Supplement Serics/Hormone and
1989, 32, 573-576. Metabolic Rcaearch; (Reiscnsburg, 1987) Thieme Medical Publishers Inc.:
(40) Fischer, U.; Ertle, R.; Abel, P.; Rebrin, K.; Brunstein, E.; Dorschevon, H. H.; New York, 1988; pp 8-13.
Freyse. E. J. Diabetologia 1987, 30, 940-945. (46) Moscone, D.; Pasini, M.; Mascini, M. Talanta 1992, 39, 1039-1044.
(41) A h , T.; Lau, Y. Y.; Ewing, A. G. J. Am. Chem. Soc. 1991,113,7421-7423. (47) Aoki, K. Electroanalysis 1993, 5, 627639.
(42) Claremont. D. J.; Pickup, J. C.; Sambrook, I. E. Life Support Sysr. 1986, (48) Fleischmann, M.; Pons, S.;Rolison, D. R.;Schmidt, P. P. Ultramicroelectrodes;
369-370. Publishers Press: Morganton, NC, 1987.
(43) Ito, K.; Ikcda, S.;Asai, K.; Naruse, H.; Ohkura, K.; Ichihashi, H.; Kamei, H.; (49) Heller, A. J . Phys. Chem. 1992, 96, 3579-3587.
Kondo, T. ACS Symp. Ser. 1986, 309, 373-382. (50) Ohara, T.; Rajagopalan, R.; Heller, A. Anal. Chem. 1993, 65, 24-29.
i 150 -
-
__r
140
80 130 -
120 -
70 110 -
- 60 / h
s
100 -
c
v
F 50 - P
P
v
F
u
90
80
-
-
70 -
0
60 -
50 -
40 -
‘ 1 ‘ 1 1 1 1 1 1 1 1 1 1 1 1
0 5 1 0 15 2 0 2 5 3 0 3 5 40 45 5 0 5 5 6 0 6 5 70 75 80 8 5 0 5 1 0 15 2 0 2 5 3 0 3 5 40 45 50 5 5 60 6 5 70 75 80 8 5
Glucose Conc. (mM) Glucose Conc. (mM)
Figure 2. Dependence of the sensitivity on the Os loading of the wire: Figure 3. Dependenceof the sensitivity on the recess depth: (0)125-
(V)PVI,-Os and (V)PV15-Os. and (0)250-pmdeep electrodes modified with the same amount of
sensing layer; E = +0.3 V vs SCE; 20 mM phosphate buffer containing
0.15 M NaCi at pH 7.15 and 25 OC.
Because the glucose electrooxidation current was limited by 140 I I I I
-
and then dropped. For the preferred 125-pm recess, 10
coatings, producing a 13-pm-thick wired-rGOx sensing
layer, yielded sensors that had the desired characteristics for
1 mM, reacted with 0 2 to form HzOz and pyruvate. H202,
in the presence of HRP, oxidizes ascorbate, urate, and
acetaminophenol, being reduced to ~ a t e r . The
~ ~ preferred
, ~ ~
in vivo use. coimmobilization process involved two separate steps: pe-
The Insulating Layer. This layer electrically insulated riodate oxidation of oligosaccharide functions of HRP to
the redox enzymes of the interference-eliminating layer (HRP aldehydes, followed by mixing with LOX and formation of
and LOX) from the wired rGOx layer. PAL cross-linking multiple Schiff bases between HRP aldehydes and LOXamines
with PAZ, forming a polycationic network at pH 7.0, was (e.g., lysines) and between H R P aldehydes and amines. The
i(nA) j(pA/cm*)
KmpPP(mM)
EH LB tr(s)
current
variance(%)
33.9 69.1 18.5 33.4 30-90 5.0
-
from an Eadie-Hofstee plot, was 20 mM (Table 1). The
10-90% response time was 1 min. As expected, and as can
be seen in Figure 5, with thinner films the glucose mass
transport increased, Le., the current was higher, while for
thicker films the stability improved. Because of the high
sensitivity of thin sensing film (- 1 pm) electrodes (>10-*A
cm-* M-l), a 1 order of magnitude decrease in sensitivity
could be traded for stability, while the currents remained high
enough to be easily measured. As seen in Figure 5, the
sensitivity of the stabilized sensors did not change by more
than f5% for 72 h of operation at 37 OC. After a small initial
decrease in sensitivity, it increased to a maximum after 40 h,
and the final 72-h sensitivity was almost identical with the
initial. There was no measurable difference in the apparent
stability of the sensitivity of continuously operated and resting
Flgure 6. Interferenceelimination by the glucose sensor: A indicates
the injections of ascorbate, producing a 0.1 mM concentration: L electrodes measured at 6-h intervals.
indicates inlections of lactic acki; 0, indicates glucose injections; 20 The characteristics of the electrodes are summarized in
mM phosphate buffer containing 0.15 M NaCi at pH 7.15, 37 OC; E Table 1. Each entry represents an average value for five
= +0.3VvsSCE. Toaccommodateallthedata,thescale waschanged
between the first and second glucose Injection. electrodes. The baseline currents were typically <0.5 nA and
the noise <10 PA. The currents observed throughout the
thickness of the interference eliminating layer was -85 pm. physiological glucose concentration range (2-20 mM) ex-
The layer was made by applying six successivecoatings. Figure ceeded the noise equivalent current by at least a factor of 100.
6 shows that electrooxidizable interferants were eliminated in The apparent Km was 20 mM, and the 10-90% response time
the presence of lactate at physiological levels. LOX slowly was, for aged electrodes, -90 s at the lowest physiologically
lost its activity in the cross-linked HRP-LOX layer. This led relevant glucose concentration (2 mM) and 30 s a t the highest
to degradation of the ability of the layer to eliminate (20 mM). The baseline of nil at 0 mM glucose was stable for
interferants. After 36 h of operation at 37 O C , a measurable 36 h in the presence of 0.1 mM ascorbate. The stability
current increment was noted when enough ascorbate was added observed and the existence of a valid zero point in the presence
to produce a 0.1 mM concentration. of interferants suggested that the sensor could be used in vivo
The Biocompatible Layer. The biocompatible layer for 72 h and be tested/recalibrated in vivo through a single-
consisted of photo-cross-linked tetraacrylated 18 500-Da poly- point calibration, Le., by withdrawing only a single sample of
(ethylene oxide).51 The thickness of this layer, made by
sequential photo-cross-linking of two coatings, was 20 pm.
The 1-min UV exposure required for the photo-cross-linking
- blood for independent analysis.
In Vivo Experimentation. The objective of this experiment
was to establish the validity of a one-point in vivo calibration.
process reduced the sensitivity by 16 f 2%. Two sensors were simultaneously implanted subcutaneously
25 I I300
50
5 I I I I 4 I 1 Io
50 100 150 200 250 300 350 400 450 500 50
Time (min.)
Flgure 7. Current output of two subcutaneously Implanted sensors, 0 50 100 150 200 250 300 350 400
tracking the blood glucose levels before, during, and after an
intraperitonealInfusion of glucose In a rat. The glucose concentrations Reference Glucose(mg/dl)
in blood samples from the tail vein of the rat were measured with a Flgure 8. Clark-type clinical error grkl analysis for the two subcutane-
YSI analyzer. The circles (0)represent the blood glucose levels ously implantedelectrodes calibrated In VIVOusingan arbitrarilychosen
measured on the withdrawn samples, line A shows the output of the point (0)electrode implantedsubcutaneously in the lntrascapulararea
electrode implantedsubcutaneouslyin the chest, and line B represents and (0)electrode Implanted subcutaneously in the chest. Readings In
the output of the electrodeImplantedsubcutaneouslyinthe intrascapular zone A are clinically accurate, those in zone B are considered clinically
area of the rat. correct.
in a rat: oneon the thorax and the second between the scapulas.
To make the difference between the blood sampled and the 350
' A
subcutaneous fluid probed with the sensors as extreme as E c .
possible, i.e., to probe whether the one-point calibration holds
even if the organs sampled are different and the sampling
sites are remote, blood was withdrawn from the tailvein. Blood
glucose levels were periodically measured in withdrawn
samples, while the absolute uncorrected sensor current output
was continuously monitored. The results are shown in Figure
7. As seen in Figure 7, at 410 min the current of this electrode
dropped precipitously. Such a drop was observed in other
subcutaneously implanted electrodes between 400 and 600
min, but was never observed in electrodes operated in buffer
at 37 OC. When the failed electrodes were withdrawn and
retested in buffer, most of their original sensitivity was found L
:' A 0%
to be intact. The cause for this sudden and reversible 0%
6 I 1 I
deactivation is under study. By use of an arbitrarily chosen
point to calculate a calibration curve for each electrode, Le.,
one blood glucose level determination and one current
measurement to establish the scales, all the data from Figure
7 were plotted in a Clarke-types2 clinical grid (Figure 8),
without further correction. In this analysis, points falling in
region A of the grid are considered clinically accurate, while deactivation of the lactate oxidase, resulting in loss of
those in region B are considered clinically correct. Points interference elimination, one-point calibration should be
falling in region C are not correct, but would not lead to practical. After such calibration, the readings of the sub-
improper treatment. Points in regions D or E are incorrect, cutaneous sensors will provide, without any correction,
and if treatment would rely on these, it would be improper. clinically useful estimates of blood glucose levels.
All of the points, from both electrodes, are in regions A and Figure 9 shows the distribution of all possible correlations
B, with 43 of the 48 points being in region A. The three points obtained when each of the 24 glucose analyses was used for
in region B near 100 mg/dL glucose, for the electrode single-point calibration of either implanted electrode. There
implanted between the scapulas, were the last three points of are 2 X 24 X 24 = 1152 points in the distribution. Of these,
theexperiment, at -410 min. Notwithstanding theunknown 78% are in region A, 15% are in region B, 1% are in region
failure mode at 400-600 min and failure after 36 h by C, 6% are in region D, and no points are in region E. In
(52) Clarke, W. L.; Cox, D.; Gonder-Frederick, L. A,; Carter, W.; Pohl, S. L.
Figure 10, we test for the improvement of the single-point
Diabetes Care 1987, 5, 622427. calibration through use of redundant electrodes. First, the
r,(
amount of polymers and enzymes was -5 pg. The glucose
0
D
response through the physiologically relevant 2-20 mM
0%
1% concentration range was close to linear. The electrodes did
not respond to ascorbate, urate, or acetaminophen01 for 36 h.
50
Their 10-90% response time was 90 s at 2 mM glucose and
C 30 s at 20 mM glucose. Their sensitivity, after 30-min
A 0%
equilibration, was stable for 72 h at 37 OC in 10 mM glucose,
, . I
0%
0 .'
0 50 100 150 200 250 300 350 400 the current deviating from the average by less than f5%. Two
Reference Glucoae(mg/dl) electrodes implanted subcutaneously in a rat tracked the blood
Flguro 10. Distribution of 242 glucose concentrations, based on 11 glucose levels, and their absolute, uncorrected current output
single-point calibrations for each sensor, for redundant implanted was proportional to the blood glucose concentration. Analysis
sensors. I f the difference in the readlngsof the two sensors (electrode
A readlng minus corrected electrode B reading) was greater than the
of the correlation between the blood glucose levels in the tail
standard deviation for all of the differences, the data were rejected vein and the current output of the sensors in the subcutaneous
both as possible one-point calibrations and as data points. regions of the thorax and between the scapulas of the same
rat showed that even when the probed sites and organs differed
readings of electrode A were normalized with respect to those
in the extreme, one-point in vivo calibration was valid. The
of electrode B by multiplying each reading by the average
analysis also showed thevalue of implanting redundant sensors.
output of electrode B divided by the average output of electrode
Had clinical decisions been made based on individual sensor
A. Next, the standard deviation was calculated for the
readings, calibrated at one point, 94% would have been
differences between the 24 sets of readings of implanted
clinically correct. By using redundant sensors and accepting
electrode B and corrected readings of implanted electrode A.
only those pairs of readings that were within one standard
Then, all those sets of readings that differed by more than the
deviation, the percentage of the clinically correct decisions
standard deviation were rejected. The number of sets was
was increased to 99%.
reduced thereby from 24 to 11; 82% of the points were in
region A, 17% in region B, 1% in region D, and no points in ACKNOWLEDGMENT
regions C and E. The distribution demonstrates that the The National Institutes of Health (DK42015) is acknowl-
sensors can be calibrated through a single independent edged for supporting this work. The Royal Swedish Academy
measurement of the glucose concentration in a withdrawn of Sciences/Per-Eric Lindahl Foundation is acknowledged
blood sample. They also demonstrate the improvement in by E.C. and the Swedish National Board for Industrial and
clinical accuracy resulting from the use of redundant sub- Technical Development is acknowledged by S.-E. L. for
cutaneous sensors. The selection of those data points that financial help. The Chiron Corp. is acknowledged for its gift
differed by less than the standard deviation for the entire set of the rGOx. The authors thank Dr. Tim Ohara and Mr.
led to a 6-fold reduction in the probability of clinically erring Ravi Rajagopalan for synthesizing the PVI-Os polymer, and
in a decision based on readings of the implanted sensors. Mr. Mark Vreeke for valuable discussions. The authors thank
Finally, the observation that a particularly large number of Mr. Andrew Young for the use of his bipotentiostats for the
readings differed by more than the standard deviation at low in vivo experiments. We also acknowledge Dr. Wolfgang
glucose concentrations shows that the sensors are less accurate Kerner for his assistance in constructing the recessed elec-
at the low glucose levels. trodes.
CONCLUSIONS Received for review January 26, 1994. Accepted June 1, 1994."
Insulated gold wire based 290-pm-0.d. subcutaneous
glucose sensors, allowing one-point calibration in vivo, can be Abstract published in Aduonce ACS Abstracts. August 1, 1994.