Professional Documents
Culture Documents
User Manual
2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
0413
No part of this document may be reproduced, copied, adapted or transmitted in any form or by any means
without the written permission of Agfa HealthCare N.V.
2 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Table of contents
2544 Q EN 20111115
3
CR MAMMOGRAPHY SYSTEM
Q Disclaimer.
4 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
The following samples show how warnings, cautions, instructions and notes
appear in this document. The text explains their intended use.
Caution: Cautions are directions which, if they are not followed, can cause
damage to the equipment described in this manual or any other
equipment or goods and can cause environmental pollution.
Instruction: This sign typically is used in combination with the warning sign,
when providing a specific instruction, which if followed exactly, should avoid
the subject of the warning.
Note: Notes provide advice and highlight unusual points. A note is not
intended as an instruction.
2544 Q EN 20111115
5
CR MAMMOGRAPHY SYSTEM
Disclaimer
6 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Introduction to the CR
Mammography System
This section covers the following topics:
Q Configuration.
Q Intended Use.
Q System Documentation.
Q Training.
Q Compliance.
Q Connectivity.
Q Installation.
Q Cleaning.
Q Safety Directions.
2544 Q EN 20111115
7
CR MAMMOGRAPHY SYSTEM
Configuration
System components
Diagnostic
Workstation
Optional
components
CR Digitizer Mammography
Printer and Film
8 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
CR Digitizer
multi-plate DX-MTM
CR Mammography Detectors, Plates and Cassettes
available formats are 18x24cm and
CR HM5.0
24x30cm
CR Workstation
NX for Mammography
2544 Q EN 20111115
9
CR MAMMOGRAPHY SYSTEM
Intended Use
Diagnostic mammography
Diagnostic mammography is a radiographic examination. It is performed to
provide additional information about patients who have signs and/or
symptoms of breast disease or radiographic findings of concern. It can also be
performed in situations where direct supervision of the imaging is required by
the interpreting physician.
A diagnostic mammogram is performed under the direct supervision of a
qualified physician in mammography and may include mediolateral oblique
(MLO), craniocaudal (CC) and/or additional views.
Screening mammography
Screening mammography is a radiological examination to detect unsuspected
breast cancer in asymptomatic women. This examination may be performed
without a physician in attendance.
10 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
System Documentation
CR Mammography System
• CR Mammography System User Manual (this
CR Mammography System document).
User Documentation CD. • Agfa Healthcare System Safety User Manual
(3100).
CR Digitizer
DX-G / DX-M User Manual (2321).
Only DX-M and CR Mammography related content applies.
CR Mammography Plates and Cassettes
AGFA CR Detectors, Plates and Cassettes (CR HD5.x, CR MD4.xR, CR HM5.x,
CR MM3.xR) User Manual (2322).
Only CR HM5.x related content applies.
CR Workstation
NX User Manual (4420).
NX User Documentation CD. Only CR Mammography related content
applies.
NX Online Help.
2544 Q EN 20111115
11
CR MAMMOGRAPHY SYSTEM
Training
Agfa provides training and support for installation, calibration and use of the
CR Mammography System and components.
The Agfa training does not cover the interpretation of the diagnostic images.
At the outcome of the training, a Customer Acceptance Document will be
endorsed accordingly.
A learning curve is needed to allow the Radiologist to get entirely familiar
with the CR Mammography digital images, which have a different 'look and
feel' or image rendering in comparison to conventional film/screen.
It is the responsibility of the Radiologist to determine the conditions needed
to fulfill the learning curve process and to implement them.
12 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Compliance
Note: The CE-label is conform the Medical Device Directive (MDD) 93/42/
EEC for a class IIa system.
2544 Q EN 20111115
13
CR MAMMOGRAPHY SYSTEM
Sensitometric response
The CsBr phosphor of the CR HM5.0 detector is a linear detector. As the
amplifier in the digitizer responsible for the amplification of the PMT signal is
a square root detector, the relation between the radiation exposure level and
the PV is a square root relation. In a log-log graph the relation is linear and the
slope is ~ 0.5.
Dynamic range
The useful dynamic range is > 40 dB. For 2mm Al 28 kV Mo / Rh exposure
conditions exposure levels ranging from ca 0.75 µGy to 8400 µGy are covered
by the imaging system.
Spatial resolutions
The pixel size of the DX-M reader is 50 µm, resulting in a Nyquist frequency of
10 lp/mm. The average MTF (averaged over both fundamental orientations)
at 1 lp/mm = 0.78; at 3 lp/mm = 0.39; at 5 lp/mm = 0.20.
Noise analysis
The DX-M/HM5.0 system is operating under quantum limited regime in the
relevant clinical detector exposure level range. Due to the high degree of
homogeneity of the physical deposited CsBr phosphor layer, the contribution
of screen structure noise to the total noise is very low. The electronic noise is
negligible. As a consequence the DQE for the system is nearly constant over
the relevant clinical detector exposure level range.
14 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Phantom testing
Sets of ACR phantom images were taken and evaluated by board certified
medical physicists with experience reading ACR phantom images for ACR
accreditation. Images were acquired at low- and standard-dose levels and
scored on hard- and softcopy. In addition 2 and 6 cm set-ups were evaluated.
The overall results for the 4.2 cm ACR phantom were 5.15 fibers, 3.80 specks
and 3.75 masses after deduction.
Contrast-detail analysis (CDMAM) was performed by UK physicists and
expert readers covering the clinical dose range. The threshold limits were
determined and compared to the predicate device.
For the Agfa system a threshold thickness of 1.34 µm for the 0.1 mm diameter
and 0.26 µm for the 0.25 mm diameter at the low-dose level (1.74 mGy) was
determined. Doses to reach the minimum acceptable threshold thickness
(0.1 mm diameter) are 1.24 mGy for the Agfa system.
2544 Q EN 20111115
15
CR MAMMOGRAPHY SYSTEM
Study Design
To meet the requirements of the guidance an image attribute review was done
with 7 patient studies which met the criteria as described in the FDA guidance
section 9 with both screening and diagnostic patient studies included.
Each patient study was processed with standard MUSICA Mammo Image
Processing and T2 MUSICA Mammo Image Processing for a total of 14 (7 x 2)
patient studies. The diagnostic studies, which contained magnification views,
were processed with a dedicated MUSICA processing algorithm for
Mammography Magnification and Spot exams.
The images were printed on Drystar Mammo blue media for hardcopy reading
and were available as DICOM files for softcopy reading.
The image attribute reviews were done by three US radiologists experienced
in Digital Mammography, that meet the qualifications required in the
guidance. Each radiologist did the readings in an independent session.
Patient Population
All mammography studies were made in Belgium or Germany. All studies
were from female patients between the age of 36 and 88.
Variables
The following variables per image are listed:
• The exposure data (kV, mAs, thickness and filtering) and the distance
between the compression paddle and the bucky are captured per image.
• The HVL data are taken from the literature derived from the captured kV
and filtering.
• BI-RADS breast composition:
1 <25% almost entirely fatty
2 25%-50% scattered fibroglandular densities
3 51%-75% for heterogeneously dense
4 >75% dense for the extremely dense category
• BI-RADS Assessment categories:
0 Assessment incomplete
16 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
1 Negative
2 Benign
3 Probably benign
4 Suspicious abnormality Perform biopsy
5 Highly suspicious of malignancy
6 Known biopsy-proven malignancy
• The calculated average glandular dose
2544 Q EN 20111115
17
CR MAMMOGRAPHY SYSTEM
Connectivity
18 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Installation
The System components and the modality AEC (Automatic Exposure Control)
have to be configured before mammography examinations will be performed.
2544 Q EN 20111115
19
CR MAMMOGRAPHY SYSTEM
CR Digitizer
After proper installation of the CR Mammography System, the mammography
scan mode on the Digitizer is activated automatically when a CR Mammo
image Cassette, correctly initialized and identified, is introduced into the
Digitizer.
CR Workstation
The CR Mammography Workstation software needs to be configured to
ensure that the mandatory:
20 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Print Layouts
For NX CR Workstation standard layouts are used.
These layouts optimize the hanging order of the films of the left and right
breast on the light box, by minimizing the borders at the thorax side of both
images.
CR Mammography Printer
Hardcopies with a maximum optical density of at least 3.6 are recommended
to display the image details properly.
Diagnostic Workstation
The softcopy reading station shall be installed and configured in presence of
the CR Mammography specialist. In case of deviation from these
recommended settings, Agfa does not guarantee optimal performance of the
system.
Softcopy reading stations of diagnostic quality require a dual head five mega-
pixel mammo display system.
2544 Q EN 20111115
21
CR MAMMOGRAPHY SYSTEM
These settings are based on the recommendations of the EUREF guidelines for
digital mammography.
Alternatively, a PVI log-based dose setting or a less restrained EUREF dose
setting is possible with the DX-M system. With these settings the system is
operated within a wider range of dose and hence image quality, still providing
sufficient image quality and acceptable dose or close-to-EUREF adjustment
respectively.
22 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
The use of W/Rh is restricted to cases in which low dose is applied (Target
AGD for DX-M with HM5.0 Dose Optimized). For a thickness less than 3 cm,
W/Rh is not recommended as this exposure technique might lead to
inadequate results. In order to avoid long exposure times, the use of
considerably higher kV is necessary (28 kV or more for medium thickness,
32 kV or more for 6 cm, 34 kV or more for a thickness more than 6 cm). W/Rh
is not recommended for magnification exposures.
2544 Q EN 20111115
23
CR MAMMOGRAPHY SYSTEM
Caution: Do not use the film density adjustment of the X-ray modality.
Note: For needle biopsy specimen, Agfa recommends the lowest available kV
setting (typical 22kV) and 15mAs.
24 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Q If you can consult prior studies, an overview of both new and prior studies.
Q A one-on-one comparison view of images (e.g. right and left CC, left CC and left
MLO,...).
Other settings should be discussed by the customer with the Agfa
mammography application specialist.
2544 Q EN 20111115
25
CR MAMMOGRAPHY SYSTEM
26 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Cleaning
2544 Q EN 20111115
27
CR MAMMOGRAPHY SYSTEM
Safety Directions
For general safety directions, refer to the Agfa HealthCare System Safety User
Manual, document 3100.
For product safety directions, refer to the User Manuals listed in ‘System Docu-
mentation’ on page 11.
Caution: It is important that the same image plate is always used with the
same cassette.
28 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Getting started
This section consists of the following topics:
Q Basic workflow.
Q Limitations.
Basic workflow
The following procedure describes the workflow that will be followed when
using the CR Mammography System.
2544 Q EN 20111115
29
CR MAMMOGRAPHY SYSTEM
At the CR workstation:
At the CR digitizer:
3 The digitizer converts the image on the exposed cassette to a digital image, stores it as
a file (or data set) and transfers it to the CR workstation via the network.
Note: When using connectivity to the X-ray modality to retrieve the exposure
data, each cassette must be identified before the next exposure is taken.
Otherwise the exposure data can be lost or linked to the wrong image.
30 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
2544 Q EN 20111115
31
CR MAMMOGRAPHY SYSTEM
Limitations
32 2544 Q EN 20111115
CR MAMMOGRAPHY SYSTEM
Technical data
The digitizers scan the CR Mammography plate under the special conditions:
• 50µm pixel size for high resolution scanning,
• asymmetric scanning towards the thorax side.
For more information, contact your sales representative.
2544 Q EN 20111115
33
0413
Printed in Belgium.
Published by Agfa HealthCare N.V., B-2640 Mortsel, Belgium
2544 Q EN 20111115