Professional Documents
Culture Documents
Defibrillator/Monitor D500
EU representative
OBELIS S.A
Bd. Général Wahis, 53, 1030 Brussels, Belgium
Local Distributor
Manufacturer
Mediana Co., Ltd.
132, Donghwagongdan-ro, Munmak-eup,
Wonju-si, Gangwon-do, Korea
Tel: (82) 2 542 3375 (82) 33 742 5400
Fax: (82) 2 542 7447 (82) 33 742 5483
Warranty
Trademark
Product brand names shown in this manual are likely to be the trademark or registered
trademark of the company concerned.
ii
Figures
Figure 1. Front panel components .......................................................................................................... 8
Figure 2. Top panel components ........................................................................................................... 10
Figure 3. Rear panel components ......................................................................................................... 11
Figure 4. Left panel components ........................................................................................................... 12
Figure 5. Right panel components ........................................................................................................ 12
Figure 6. Manual Self-Test .................................................................................................................... 19
Figure 7. Parameter Default Setting ..................................................................................................... 25
Figure 8. Pressure Sensor Accuracy Test ............................................................................................. 32
Figure 9. Air Leakage Test .................................................................................................................... 33
Figure 10. Inflation Time Measurement ................................................................................................ 34
Figure 11. Deflation Rate Measurement ............................................................................................... 35
Figure 12. The access of Service menu via Setup menu ..................................................................... 47
Figure 13. Entering the pass code ........................................................................................................ 48
Figure 14. System Setting I ................................................................................................................... 49
Figure 15. System Setting II .................................................................................................................. 52
Figure 16. Printer Setting ...................................................................................................................... 55
Figure 17. Parameter Default Setting ................................................................................................... 56
Figure 18. ECG Setting ......................................................................................................................... 63
Figure 19. 12 Lead ECG Setting ........................................................................................................... 64
Figure 20. Shock Setting ....................................................................................................................... 66
Figure 21. AED Mode Setting ............................................................................................................... 68
Figure 22. Pacing Mode Setting ............................................................................................................ 70
Figure 23. 3G Network Setting .............................................................................................................. 71
Figure 24. Wi-Fi Network Setting .......................................................................................................... 72
Figure 25. Software update display....................................................................................................... 81
Figure 26. Disassembly Sequence Flow Chart ................................................................................... 100
Figure 27. Side option case disassembly (1) ...................................................................................... 101
Figure 28. Side option case disassembly (2) ...................................................................................... 102
Figure 29. Front case disassembly (1)................................................................................................ 104
Figure 30. Front case disassembly (2)................................................................................................ 106
Figure 31. Front case disassembly (3)................................................................................................ 108
Figure 32. Front case disassembly (4)................................................................................................ 109
Figure 33. Rear case disassembly (1) ................................................................................................ 110
Figure 34. Rear case disassembly (2) ................................................................................................ 111
Figure 35. Paddle mounting plate disassembly .................................................................................. 113
Figure 36. AC power supply (SMPS) disassembly ............................................................................. 115
Figure 37. D500 Exploded view .......................................................................................................... 116
Figure 38. Exploded View – Spare Parts ............................................................................................ 117
Figure 39. Biphasic Waveforms at 360 Joules and 360 Joules .......................................................... 138
Figure 40. Biphasic Waveforms at 200 Joules.................................................................................... 138
Figure 41. Biphasic Waveforms at 175 Joules.................................................................................... 139
Figure 42. Biphasic Waveforms at 150 Joules.................................................................................... 139
Figure 43. Biphasic Waveforms at 125 Joules.................................................................................... 140
Figure 44. Biphasic Waveforms at 100 Joules.................................................................................... 140
Figure 45. Biphasic Waveforms at 75 Joules...................................................................................... 141
Figure 46. Biphasic Waveforms at 50 Joules...................................................................................... 141
Figure 47. Biphasic Waveforms at 40 Joules...................................................................................... 142
Figure 48. Biphasic Waveforms at 30 Joules...................................................................................... 142
Figure 49. Biphasic Waveforms at 20 Joules...................................................................................... 143
Figure 50. Biphasic Waveforms at 10 Joules...................................................................................... 143
Figure 51. Biphasic Waveforms at 9 Joules........................................................................................ 144
Figure 52. Biphasic Waveforms at 8 Joules........................................................................................ 144
Figure 53. Biphasic Waveforms at 7 Joules........................................................................................ 145
Figure 54. Biphasic Waveforms at 6 Joules........................................................................................ 145
Figure 55. Biphasic Waveforms at 5 Joules........................................................................................ 146
iii
Figure 56. Biphasic Waveforms at 4 Joules........................................................................................ 146
Figure 57. Biphasic Waveforms at 3 Joules........................................................................................ 147
Figure 58. Biphasic Waveforms at 2 Joules........................................................................................ 147
Figure 59. Biphasic Waveforms at 1 Joules........................................................................................ 148
Figure 60. System Block Diagram ...................................................................................................... 153
Figure 61. Power Unit Block Diagram ................................................................................................. 154
Figure 62. Main Process Unit Block Diagram ..................................................................................... 154
Figure 63. Display Unit Block Diagram ............................................................................................... 155
Figure 64. User-Control Unit Block Diagram....................................................................................... 155
Figure 65. Audio Unit Block Diagram .................................................................................................. 155
Figure 66. Communication Unit Block Diagram .................................................................................. 156
Figure 67. GUI Unit Block Diagram ..................................................................................................... 156
Figure 68. Thermal Printer Unit Block Diagram .................................................................................. 156
Figure 69. NIBP Unit Block Diagram ................................................................................................... 157
Figure 70. EtCO2 Unit Block Diagram ................................................................................................. 157
Figure 71. ECG Unit Block Diagram ................................................................................................... 157
Figure 72. SpO2 Unit Block Diagram................................................................................................... 158
Figure 73. Temperature Unit Block Diagram ....................................................................................... 158
Figure 74. IBP Unit Block Diagram ..................................................................................................... 158
Figure 75. Oxyhemoglobin Dissociation Curve ................................................................................... 162
iv
Tables
Table 1. The defibrillator/monitor controls on front panel ........................................................................ 9
Table 2. The defibrillator/monitor controls on paddle ............................................................................ 10
Table 3. Panel and Label Symbols........................................................................................................ 13
Table 4. Required Equipment................................................................................................................ 17
Table 5. Earth Leakage Current Values ................................................................................................ 39
Table 6. Enclosure Leakage Current Values......................................................................................... 40
Table 7. Patient Leakage Current Values ............................................................................................. 41
Table 8. Patient Leakage Current Values - Mains Voltage on Applied Part .......................................... 42
Table 9. Test Lead Combinations .......................................................................................................... 42
Table 10. Allowable Leakage Current ................................................................................................... 43
Table 11. System Setting I..................................................................................................................... 49
Table 12. System Setting II ................................................................................................................... 52
Table 13. Printer Setting ........................................................................................................................ 55
Table 14. Parameter Default Setting ..................................................................................................... 56
Table 15. ECG Setting........................................................................................................................... 63
Table 16. 12 Lead ECG Setting ............................................................................................................ 64
Table 17. Shock Setting ........................................................................................................................ 66
Table 18. AED Mode Setting ................................................................................................................. 68
Table 19. Pacing Mode Setting ............................................................................................................. 70
Table 20. 3G Network Setting ............................................................................................................... 71
Table 21. Wi-Fi Network Setting ............................................................................................................ 72
Table 22. Parameter ranges and factory defaults ................................................................................. 74
Table 23. Required Equipments for software update ............................................................................ 81
Table 24. Problem Categories ............................................................................................................... 84
Table 25. Technical error codes ............................................................................................................ 96
Table 26. Part descriptions – Side option case disassembly (1) ........................................................ 101
Table 27. Part descriptions – Side option case disassembly (2) ........................................................ 102
Table 28. Part descriptions – Front case (1) ....................................................................................... 104
Table 29. Part descriptions – Front case (2) ....................................................................................... 106
Table 30. Part descriptions – Front case (3) ....................................................................................... 108
Table 31. Part descriptions – Front case (4) ....................................................................................... 109
Table 32. Part descriptions – Rear case (1) ........................................................................................ 110
Table 33. Part descriptions – Rear case (2) ........................................................................................ 111
Table 34. Part descriptions – Paddle mounting plate.......................................................................... 113
Table 35. Part descriptions – AC power supply (SMPS) ..................................................................... 115
Table 36. Part list................................................................................................................................. 118
Table 37. Delivered Energy at Every Defibrillator Settings into a Range of Loads ............................. 137
v
This page is intentionally left blank.
vi
Safety Information
Important! Before use, read carefully the manual, accessory directions for use,
all precautionary information and specifications.
Warnings
Do not connect grounding wire to gas pipes. This could cause fire.
Only doctors and officially certified personnel should use this D500. Do
not allow patients to touch this defibrillator/monitor. Allowing patients
to touch this defibrillator/monitor could cause accidents.
This defibrillator/monitor cannot be used when MRI is in progress. If
MRI is in use, keep patient attachments away from patients to prevent
accidents.
The defibrillator/monitor conforms to the requirements of the EMC
standard (IEC60601-1-2), and may therefore be used simultaneously
with pacemakers and other electrical simulators. It should, however, be
noted that the defibrillator/monitor may be affected by electrical
scalpels and microwave therapeutic apparatus. Please check operation
of the defibrillator/monitor during and after use of such equipment.
Do not take mobile phones or transceivers into a room where this
defibrillator/monitor is installed, as such devices may cause accidents.
In order to avoid accidents, do not use any unauthorized accessories or
options.
Thoroughly read the instruction manuals supplied with accessories and
options to ensure correct use. This instruction manual does not carry
the caution selections for such equipment.
Do not open cover or disassemble this defibrillator/monitor. Doing so
could cause electric shock or fire. It is prohibited by law to modify the
1
defibrillator/monitor without authorization.
Do not use power source other than the specified voltage, (100-
240V~50/60Hz) as this may cause fire or electric shock.
Pre-use inspection and preventive maintenance must be performed for
safe use.
The defibrillator/monitor may be used with electrical surgical
equipment.
Follow the instruction manuals for medical instruments – notably
electrosurgical and diathermy instruments – when used, as their high–
frequency energy units may cause burns to patients via attachments.
This defibrillator/monitor is protected against the discharge of a
defibrillator. However, do not touch the defibrillator/monitor when a
defibrillator is being discharged (electrified), as doing so may cause
electric shock.
The following cautions apply when connecting the defibrillator/monitor
with other equipment.
1. Ensure that the connected equipment is in accordance with the
IEC60601-1 or IEC safety standards, so that the system complies
with IEC60601-1.
2. Employ additional protective measures (e.g. additional protective
earthing) as necessary.
Do not connect devices that do not meet medical safety standards
(such as commercial PCs), as they may cause electric shock. This
defibrillator/monitor meets the restricted level of leakage current
required for medical devices. Therefore, this defibrillator/monitor
cannot be connected to a device that would give a combined total of
leakage current beyond the restricted level.
Avoid connecting the patient to several devices at once. Leakage
current limits may be exceeded. Do not use a second defibrillator on the
patient while pacing with the Mediana defibrillator/monitor.
Do not place anything on top of this defibrillator/monitor. If something
is spilled over the defibrillator/monitor or gets into it, such spillage may
cause fire or electric shock. If fluid spills on the defibrillator/monitor
accidentally, disconnect power cord, wipe dry immediately, and have
the defibrillator/monitor serviced to make sure that no hazard exists.
Do not place heavy objects on the power cord, as doing so may cause
fire or electric shock.
Before conducting maintenance work, turn the power Off and unplug
the power cord from the wall socket to prevent electric shock.
When the following occur, turn the power Off immediately and unplug
the power cord from the wall socket. Continued use in such situations
may cause fire or electric shock.
There is smoke or a strange odor leaking out of the device.
The device has been dropped or impacted by an object.
Liquid or foreign matter gets inside the device.
Device failure has occurred.
Also, when any of the above occurs, promptly do the following:
1. Check to see that the power cord has been unplugged from the
wall socket.
2. Place an “Out of Order” sign on the device and do not use it.
Do not connect more than one patient to the defibrillator/monitor. Do
2
not connect more than one defibrillator/monitor to a patient.
The defibrillator/monitor is a prescription device and is to be operated
by qualified personnel only.
As with any medical equipment, carefully route patient cabling to
reduce the possibility of patient entanglement or strangulation.
Never lift the defibrillator/monitor by the sensor cable, blood pressure
hose, power cord, or any other accessory. Such accessories could
detach, causing the defibrillator/monitor to fall on the patient.
Do not make any clinical judgments based on this
defibrillator/monitor’s measurement only.
Emergency defibrillation should be performed only by appropriately
trained, skilled and qualified personnel who are aware of the protocol
for handling a patient in medical emergency such as cardiac arrest and
have been certified in Advanced Cardiac Life Support (ACLS) or Basic
Life Support (BLS).
Synchronized electrical cardioversion should be performed only by
skilled personnel trained in Advanced Cardiac Life Support (ACLS) and
practiced in equipment operation. The precise cardiac arrhythmia must
be determined prior to performing defibrillation.
The defibrillator delivers up to 360 joules of electrical energy. Unless
properly used as described in this manual, this electrical energy may
cause serious injury or death. Do not attempt to operate this
defibrillator/monitor unless thoroughly familiar with this manual and the
function of all controls, indicators, connectors and accessories.
Do not discharge standard paddles on top of pads or ECG electrodes.
Do not allow standard paddles (or pads) to touch each other, ECG
electrodes, lead wires, dressings, transdermal patches, etc. Such
contact can cause electrical arcing and patient skin burns during
defibrillation and may divert defibrillating energy away from the heart
muscle.
Discharging the defibrillator with the standard paddle surfaces shorted
together can pit or damage the paddle electrode surface. Pitted or
damaged paddle surfaces may cause patient skin burns during
defibrillation. Discharge the defibrillator only as described in theses
operating instructions.
If a person is touching the patient, bed, or any conductive material in
contact with the patient during defibrillation, the delivered energy may
be partially discharged through that person. Clear every away from
contact with the patient, bed, and other conductive material before
discharging the defibrillator.
Do not discharge the defibrillator into the open air. To remove an
unwanted charge, change the energy selection, or change the mode, or
turn off the defibrillator.
Conductive gel on the paddle handles can allow the electrical energy to
discharge through the operator during defibrillation. Completely clean
the paddle electrode surfaces, handles, and storage area after
defibrillation.
A gel pathway on the skin between the standard paddles will cause
defibrillating energy to arc between paddles and delivery energy away
from the heart muscle. Do not allow conductive gel (wet or dry) to
become continuous between paddle sites.
During defibrillation checks, the discharged energy passes through the
3
cable connectors. Securely attach cable connectors to the simulator.
Do not touch the patient or any equipment connected to the patient
during defibrillation. Warn all persons around patient to STAY AWAY
prior to defibrillation.
The defibrillator/monitor should be out of contact with water (puddles
or water spray). It may cause electrical shocks and device failure.
Electrical safety of the device may not work properly when wet.
Defibrillation may cause implanted devices to malfunction. Place
standard paddles or pads away from implanted devices if possible.
Check implanted device function after defibrillation.
Implanted pacemakers may cause the heart rate meter to count the
pacemaker rate during incidents of cardiac arrest or other arrhythmias.
Pacemaker patients should be carefully observed. Check the patient’s
pulse; do not rely solely on heart rate meters. Dedicated pacemaker
detection circuitry may not detect all implanted pacemaker spikes.
Patient history and physical exam are important in determining the
presence of an implanted pacemaker.
Cautions
Caution statements identify conditions or practices that could result in damage
to the equipment or other property.
4
Manual Overview
This manual contains information for servicing the D500 defibrillator/monitor. The
D500 defibrillator/monitor will be referred to as the defibrillator/monitor throughout this
manual.
Read and understand all safety warnings and service notes printed in this service
manual and the operator’s manual.
Related Documents
To perform tests and troubleshooting procedures, and to understand the principles of
operation and circuit analysis sections of this manual, you must know how to operate
the defibrillator/monitor. Refer to the defibrillator/monitor operator’s manual.
To understand the various paddle and pads, SpO2 sensors, ECG leads, blood
pressure cuffs, IBP transducer, capnography accessories and temperature probes
that work with the defibrillator/monitor, refer to the individual directions for use that
accompany these accessories.
5
Indications for Use
Indications Contraindications
Manual Manual defibrillation is Defibrillation is
Defibrillation indicated for the termination contraindicated in the
of certain potentially fatal treatment of Pulseless
arrhythmias, such as Electrical Activity (PEA),
ventricular fibrillation and such as idioventricular or
ventricular tachycardia. ventricular escape
Delivery of this energy in the rhythms, and in the
synchronized mode is a treatment if asystole.
method for treating atrial Defibrillation is not
fibrillation, atrial flutter. intended for patient who is
conscious and has a
palpable pulse and is
contraindicated for use on
neonates.
Automated External AED mode is to be used only AED is not intended for
Defibrillation on patients of at least 8 years patient who is conscious
of age in cardiopulmonary and has a palpable pulse
arrest. The patient must be and is contraindicated for
unconscious, pulseless, and use on neonates.
not breathing normally before
using the defibrillator to
analyze the patient’s ECG
rhythm.
External Pacing External pacing is indicated External pacing is
for symptomatic bradycardia contraindicated for the
in patients with a pulse. treatment of ventricular
Increased heart rates in fibrillation and asystole.
response to external pacing External pacing in the
often suppress ventricular presence of severe
ectopic activity and may hypothermia may be
prevent tachycardia. contraindicated and is
contraindicated for use on
neonates.
Electrocardiography The 12-lead No known
electrocardiogram is used to contraindications
identify, diagnose and treat
patients with cardiac
disorders and is useful in the
early detection.
Noninvasive Blood Noninvasive blood pressure Noninvasive Blood
Pressure monitoring is intended for Pressure is not intended
detection of hypertension or for use with severe
hypotension and monitoring arrhythmia. Noninvasive
BP trends in patient Blood Pressure is not
conditions such as, but not intended for patients who
limited to, shock, acute are experiencing
dysrhythmia, or major fluid convulsion or tremors.
imbalance.
Pulse Oximetry Pulse oximetry monitoring is Pulse Oximetry is not
intended to be used to intended for use with
monitor functional arterial severe peripheral vascular
6
Indications Contraindications
oxygen saturation and pulse disease and severe
rate. anemia (decreased
Hemoglobin).
Invasive Blood Invasive pressure monitor is No known
Pressure indicated for use in contraindications
measuring arterial, venous,
intracranial and other
physiological pressures using
and invasive catheter system
with a compatible transducer.
End-Tidal CO2 EtCO2 monitoring is indicated No known
for detection of trends in the contraindications
level of expired CO2. It is
used for monitoring breathing
efficacy and treatment
effectiveness in acute
cardiopulmonary care.
Temperature Temperature monitoring is No known
indicated for use in patients contraindications
who require continuous
monitoring of body
temperature.
7
Defibrillator / Monitor Controls
12
13 - 16
17
18
2
3 19 - 21
4 22
5
6 23
24 - 25
7
8 26 - 30
9
10
11
31 32
8
Table 1. The defibrillator/monitor controls on front panel
Symbols Description
ANALYZE button
analyzes the patient’s ECG to determine whether or not to
deliver a shock.
CHARGE button
charges to the desired energy level automatically.
SHOCK button
delivers a shock.
RATE button
adjusts the rate of pacing waveform.
mA button
adjusts the current of pacing waveform.
NIBP button
toggles between starting and stopping NIBP measurements.
PRINT button
prints measured data.
Alarm button
pauses the audible alarm temporarily.
turns off the audible alarm by pressing over 2 seconds.
LEAD button
selects to display the desired ECG lead on the screen.
SIZE button
adjusts and selects the amplitude of an ECG waveform.
9
Top Panel Components 1 2 3 4 5
CHARGE button
charges to the desired energy level automatically.
REC button
Prints the measured data.
10
Rear Panel Components
3 4 5 6 7
11
Left Panel Components
12
Table 3. Panel and Label Symbols
Symbols Description Symbols Description
Environmental shipping/storage
IBP 2 connector
altitude limitations
Environmental shipping/storage
EtCO2 connector
humidity limitations
Environmental shipping/storage
SpO2 connector
temperature limitations
13
This page is intentionally left blank.
14
Routine Maintenance
Do not spray or pour any liquid on the defibrillator/monitor or its
accessories. Do not immerse the defibrillator/monitor or its accessories
in liquid or clean with caustic or abrasive cleaners.
Cleaning
The defibrillator/monitor may be surface-cleaned by using a soft cloth dampened with
either a commercial, nonabrasive cleaner or one of the solutions listed below. Lightly
wipe the top, bottom and front surfaces of the defibrillator/monitor lightly.
Quatemary Ammonium
Alcohol-70% Isopropyl
10% Chlorine bleach solution
PDI Sani-System
For cables, sensors, cuffs, and probes, follow cleaning instructions in the directions for
use shipped with those components.
15
Batteries
If the defibrillator/monitor has not been used for a long period of time, more than 6
months, the battery will need charging. To charge the battery, connect the
defibrillator/monitor to an AC outlet as described in the Battery Charge paragraph in
this service manual or the Battery Operation section of the operator’s manual.
Note: Storing the defibrillator/monitor for a long period without charging the battery
may degrade the battery capacity. The battery may require a full
charge/discharge cycle to restore normal capacity. Mediana recommends that
the defibrillator/monitor’s sealed, Li-ion batteries be replaced at 2 years intervals
(shelf-life of Li-ion battery is 2 years). Refer to the Disassembly Guide section.
Note: Due to the physical dimensions of the battery compartment, only batteries
supplied by Mediana should be used. Using other types of replacement
batteries may result in damage to the defibrillator/monitor and void the limited
warranty.
Environmental Protection
Follow local governing ordinances and recycling plans regarding disposal or recycling
batteries and other defibrillator/monitor components.
Note: The correct disposal of your old appliance will help prevent potential negative
consequences for the environment and human health.
Note: For more detailed information about disposal of your old appliance, please
contact your city office, waste disposal service or the shop where you
purchased the defibrillator/monitor.
16
Performance Verification
General
This section discusses the tests used to verify performance following repairs or during
routine maintenance. All tests can be performed without removing the
defibrillator/monitor covers. All tests except the battery charge and battery discharge
tests must be performed as the last operation before the defibrillator/monitor is
returned to the user.
If the defibrillator/monitor fails to perform as specified in any test, repairs must be
made to correct the problem before the defibrillator/monitor is returned to the user.
Required Equipment
Table 4 lists the equipment required for performance verifications.
Note: The sphygmomanometer must be calibrated periodically. The correct value can
not be found if the sphygmomanometer has not been calibrated.
Note: Contact Mediana Technical Service Department for pricing and ordering
information.
17
Performance Tests
The battery charge and battery discharge test should be performed before
defibrillator/monitor repairs whenever the battery is suspected as being a source of
problems. All other tests may be used following repairs or during routine maintenance
(if required by your local institution). Before performing the battery discharge test,
ensure that the battery is fully charged. This section is written using factory defaults
as power-up. Please refer to the Service Menu and Parameter default settings
section to set the factory defaults. If your institution has preconfigured custom defaults,
those values will be displayed.
Power
1. Connect the defibrillator/monitor to AC power source or DC power source using
proper power cord.
2. Verify that AC/DC in LED and Battery A charging status LED are lit.
3. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
4. Verify that the defibrillator/monitor is turned on.
5. After the defibrillator/monitor operates in normal mode, disconnect the power cord.
6. Verify that Battery status icon at top of the screen indicates the residual quantity
of battery capacity instead of AC and battery status icon. Verify that AC/DC in
LED and Battery A charging status LED are off.
7. Rotate the Mode select knob to OFF, and then verify that the defibrillator/monitor
is turned off.
Battery Charge
1. Connect the defibrillator/monitor to AC power source or DC power source using a
proper power cord.
2. Verify AC/DC in LED is lit with green color and Battery A charging status LED is
lit with orange color.
3. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
4. Verify that the Battery status icon appears at top of the screen. The bar in battery
status icon should be filled, indicating that the battery is charged.
5. To check for a full charge, perform the procedure in paragraph “Battery Discharge”.
Note: The battery may require a complete charge/discharge cycle to restore its normal
capacity, depending on its previous usage.
Battery Discharge
1. Disconnect the power cord from the defibrillator/monitor with fully charge battery.
2. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
3. Verify that the Battery status icon appears at top of the screen. The bar in battery
status icon should be filled and displays ‘100%’, indicates that the battery is fully
charged.
4. Connect ECG lead cable to appropriate terminals on the ECG simulator and make
no alarm condition.
5. The defibrillator/monitor must operate for 5 hours with one fully charged battery.
The monitor must operate for at least 15 minutes after the alarm message “Low
18
Battery” appears before the monitor powers down due to the low battery condition.
6. Verify that low priority audible alarm occurs and a alarm message “Low battery” is
displayed about 15 minutes before battery fully discharges.
7. Allow defibrillator/monitor to operate until it automatically powers down due to low
battery condition. Verify that high priority audible alarm occurs and a alarm
message “SYSTEM: Critically Low-Battery condition” is displayed about 5
minutes before defibrillator/monitor automatically shuts down.
If defibrillator/monitor passes this test, immediately recharge battery. (paragraph
“Battery Charge”)
Manual Self-Test
1. Connect the defibrillator/monitor to an AC power source.
2. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
3. Press Setup soft key.
4. Rotate Multi function knob and select Manual Self Test menu.
5. Manual Self-Test is started as Figure 6. After finish the test, the result is printed out
automatically.
19
Pads/paddle connection test
Test if the defibrillator/monitor detects the condition of pads/paddle connection.
System ROM test
Test if the ROM reads data in normal.
System RAM test
Test if the RAM reads and writes data in normal.
Real time clock test
Test if the RTC chip for real time data operates in normal.
Installed module alive test
Test if the modules which are installed on defibrillator/monitor operate in normal.
Battery test
Test if the defibrillator/monitor detects the condition of battery installation and
charging.
20
Alarm Volume Control
1. Connect the defibrillator/monitor to an AC power source.
2. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
3. Connect SpO2 simulator to sensor input cable and connect cable to
defibrillator/monitor.
4. Set SpO2 simulator as follows: SpO2 of 75% and pulse rate of 60bpm.
5. Verify SpO2 and pulse rate values are correctly displayed.
6. Verify the audible alarm sounds.
7. Press Setup soft key and the setup menu will be displayed.
8. Rotate the Multi function knob to highlight Volume and then press the knob to
select Volume menu.
9. Rotate the Multi function knob to select Alarm menu, and if the Alarm menu is
highlighted, press Multi function knob.
10. Speaker icon is activated.
11. Rotate the Multi function knob to adjust the volume level from 1 to 8.
12. Press the Multi function knob to save the desired volume.
13. Press Return soft key. Verify the alarm sound is increased.
14. Set the alarm volume level from 8 to 1.
15. Verify the alarm sound is decreased.
21
13. Press the Multi function knob to save the desired volume.
14. Press Return soft key. Verify the beep sound is increased.
15. Set the beep volume level from 7 to 0.
16. Verify the alarm sound is decreased.
22
Voice Prompt Volume Control
1. Connect the defibrillator/monitor to an AC power source.
2. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
3. Select the Setup menu by pressing the soft key and the setup menu will be
displayed.
4. Rotate the Multi function knob to highlight Volume and then press the knob to
select Volume menu.
5. Rotate the Multi function knob to select Voice prompt menu, and if the Voice
prompt menu is highlighted, press Multi function knob.
6. Speaker icon is activated.
7. Rotate the Multi function knob to adjust the volume level from 1 to 8.
8. Press the Multi function knob to save the desired volume.
9. Press Return soft key. Verify the voice prompt sound is increased.
10. Set the voice prompt volume level from 8 to 1.
11. Verify the voice prompt sound is decreased.
23
4. Rotate the Multi function knob to highlight Volume and then press the knob to
select Volume menu.
5. Rotate the Multi function knob to select Pacing alarm menu, and if the Button
menu is highlighted, press Multi function knob.
6. Speaker icon is activated.
7. Rotate the Multi function knob to adjust the volume level from 0 to 7.
8. Press the Multi function knob to save the desired volume.
9. Press Return soft key. Verify the Pacing alarm sound is increased.
10. Set the Pacing alarm volume level from 7 to 0.
11. Verify the Pacing alarm sound is decreased.
24
Restoring Power-On Default Settings
The following test procedures will verify that alarms are activated at the level of factory
default alarm limits and that any changed settings are saved and in effect when the
user changes alarm limit settings in Service menu.
Printer Test
The following test procedures will verify the printer performance.
25
6. Set Print on alarm, Print on charge, Print on shock, Print on mark, Print on BP,
Print on self-test, 12 lead auto printing and Analyze auto printing to On.
7. Press SAVE soft key then select Yes to save the setting.
8. Turn off the defibrillator/monitor.
9. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
10. Press the Setup soft key. (Setup is displayed on the initial screen of each mode.)
11. Rotate the Multi function knob to highlight Printer setting via Setup Menu, and
then press the Multi function knob to select Printer setting.
12. Set 12 lead auto printing and Analyze auto printing to On.
13. Test #1: Print on alarm
a. Set SpO2 simulator as follows: SpO2 of 75% and pulse rate of 60bpm.
b. Verify that the audible alarm sounds.
c. Verify that the defibrillator/monitor automatically prints out information
before and after 10 seconds when a physiological alarm condition is
activated.
14. Test #2: Print on charge
a. Press the Energy Level button and rotate the Multi function knob to
choose energy level as 100J.
b. Press the CHARGE button to charge the energy.
c. Verify that the defibrillator/monitor automatically prints out information
before and after 10 seconds when shock energy is charged.
Note: Do not press the SHOCK button during the test. The charged energy will be
automatically discharged after 60 seconds.
15. Test #3: Print on mark
a. Press the Setup soft key.
b. Rotate the Multi function knob to highlight Clinical action list and then
press the Multi function knob to select Clinical action list.
c. Select the first clinical action.
d. Verify that the defibrillator/monitor automatically prints out the first clinical
action.
16. Test #4: Print on BP
a. Ensure the defibrillator/monitor is set up with dummy can-large.
b. Press NIBP button to start measuring NIBP value.
c. Verify that the defibrillator/monitor automatically prints out the result of
NIBP measurement.
17. Test #5: Print on self-test
a. Press the Setup soft key.
b. Rotate the Mode select knob to highlight Manual self-test and then press
the Multi function knob to start Manual self-test.
c. Verify that the defibrillator/monitor automatically prints out the result of
manual self-test.
26
18. Test #6: 12 lead auto printing
a. Connect ECG 12 lead cable to the defibrillator/monitor.
b. Connect the ECG 12 lead cable to appropriate terminals on the ECG
simulator.
c. Press 12 Lead soft key and then press Acquire soft key.
d. Verify that the defibrillator/monitor automatically prints out the analysis
result of acquired 12 lead ECG waveform data.
19. Test #7: Analyze auto printing
a. Ensure the ECG lead cable is set up with simulator and the waveform is
displayed in normal.
b. Press ANALYZE button.
c. Verify that the defibrillator/monitor automatically prints out the analyzing
data for 10 seconds after a rhythm analysis is done.
27
Defibrillator Operation Tests
Pacing mode
1. Connect the defibrillator/monitor to an AC power source.
2. Turn on the analyzer and press pacer button.
3. Set the analyzer as follows:
Brand: Default algorithm
Load: 50Ω
AED mode
1. Connect the defibrillator/monitor to an AC power source.
2. Turn on the analyzer and press defibrillator button.
3. Set the analyzer as follows:
Brand: Default algorithm
Load: 50Ω
4. Press F1 button 3 times to enter to the waveform selection mode in the analyzer
and select VFib waveform.
5. Check the ECG waveform type is set as coarse in the analyzer then press F5
button to select Done.
6. Check the analyzer is set to ready.
7. Connect pads connector to defibrillator/monitor’s front panel.
8. Contact pads to the defibrillator analyzer’s adapter plates.
9. Rotate the Mode select knob to AED mode to turn on the defibrillator/monitor.
10. Defibrillator/monitor automatically starts the AED step with audio guidance and
STEP icon. When the audio guidance said “Press the SHOCK button now.”, press
SHOCK button on the defibrillator/monitor’s front panel. Delivered shock energy is
125J for the first shock.
11. Verify that the analyzer displays the delivered shock energy of 125J±15%.
Note: The accuracy of the defibrillator/monitor shock energy is ±15% when the load is
set to 50 Ω.
28
Manual mode
1. Connect the defibrillator/monitor to an AC power source.
2. Turn on the analyzer and press defibrillator button.
3. Set the analyzer as follows:
Brand: Default algorithm
Load: 50Ω
4. Press F1 button 3 times to enter to the waveform selection mode in the analyzer
and select VFib waveform.
5. Check the ECG waveform type is set as coarse in the analyzer then press F5
button to select Done.
6. Check the analyzer is set to ready.
7. Connect paddles connector to defibrillator/monitor’s front panel.
8. Contact paddles to the defibrillator analyzer’s adapter plates.
9. Rotate the Mode select knob to Manual mode to turn on the defibrillator/monitor.
10. Press the Energy Level button to select the energy setting. Rotate the Multi
function knob to set energy level as 125J.
11. Press SHOCK button on the defibrillator/monitor’s front panel.
12. Verify that the analyzer displays the delivered shock energy of 125J±15%.
Note: The accuracy of the defibrillator/monitor shock energy is ±15% when the load is
set to 50 Ω.
29
Measurement Parameter Operation Tests
ECG Operation
1. Connect the defibrillator/monitor to an AC power source.
2. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
3. Connect the ECG 3 lead cable to appropriate terminals on the ECG simulator.
4. Connect the ECG cable to ECG connector on the defibrillator/monitor’s front panel.
5. Set ECG simulator as follows:
Heart rate: 30 bpm
Amplitude: 1 mV
Lead select: II
Normal sinus rhythm
30
11. Set ECG Lead selection to Lead II.
12. Change ECG waveform size to all the selectable sizes and verify that an
appropriate size is displayed on the ECG waveform area.
13. Disconnect 3 ECG leads and connect 5 ECG leads.
14. Verify that Waveform Setting menu displays I, II, III, aVR, aVL, aVF, V.
15. Repeat step 9 to 12.
16. Disconnect 5 ECG leads and connect 12 ECG leads.
17. Verify that Waveform Setting menu displays I, II, III, aVR, aVL, aVF, V, V1, V2,
V3, V4, V5, V6.
18. Repeat step 9 to 12.
19. Turn off the defibrillator/monitor.
NIBP Operation
These tests verify the functionality of the defibrillator/monitor pneumatic system. The
Bio-Tek simulator or any equivalent NIBP simulator is required to perform these tests.
Each of the tests must be performed to verify pneumatic system functionality.
Over-Pressure Test
1. Connect the cuff hose for adult to the NIBP connector on the defibrillator/monitor’s
front panel.
2. Connect the other end of the cuff hose to the NIBP simulator.
3. Set the NIBP simulator to Pressure Relief or Over-pressure Test mode.
4. Press the Start Test Button on the NIBP simulator. The simulator will pressurize the
system until the monitor’s over pressure relief system activates.
Verify that the peak point displayed on the NIBP simulator (point of protection
pressure) is within 330mmHg. Also, this point of protection pressure may be verified at
the moment of the monitor’s NIBP relief system activation.
These tests verify the functionality of the defibrillator/monitor’s pneumatic system. The
defibrillator/monitor must be placed in Service menu. For a detailed explanation to
access the service menu, refer to Service Menu and Parameter Default Settings
section.
1. Rotate the Multi function knob to select NIBP test mode in the service menu, and
then press the Multi function knob.
Note: Before accessing the NIBP test mode, ensure that current patient mode is
proper for the pneumatic system to be test. You can set patient mode:
Adult/Pediatric or Neonatal via the Setup menu.
31
Note: In the NIBP test mode, function buttons will have no effect except the Multi
function knob. All the tests will start to be performed by pressing or rotating the
Multi function knob. If you would like to stop the test during test progressing,
press the Multi function knob.
Note: Inflation time measurement and Deflation Rate Measurement are intended
for factory use only.
1. Connect the NIBP cuff hose to the NIBP connector on the defibrillator/monitor’s
front panel.
2. Connect the other end of the NIBP cuff hose to the NIBP simulator.
3. Rotate the Multi function knob to select NIBP test mode in the Service menu,
and then press the Multi function knob.
4. Rotate the Multi function knob to select Pressure Sensor Accuracy Test, and
then press the Multi function knob.
5. Press Select button on the simulator until simulator displays “Pressure Source
Set Test Pressure”. Adjust pressure on the simulator for 250, 150, 50 and 0
mmHg.
6. Press Start Pump button on simulator. The simulator will begin to pressurize.
Allow 15~20 seconds for pressure to stabilize.
7. Press Multi function knob on defibrillator/monitor to select Start.
The current pressure in mmHg will be displayed on both of the simulator and the
monitor displays. Ensure the monitor pressure sensor accuracy meets the
performance standard of ANSI/AAMI SP-10:2002+A1:2003 +A2:2006 (within the
specification by more than ±10 mmHg in Adult/Pediatric patient mode) to
successfully complete the test.
8. After finishing the test, press the Multi function knob to select Return. The
menu box will disappear. If Return is selected during the test progressing, the
test will stop and the menu box will disappear.
32
Air Leakage Test
The air leakage test verifies the integrity of the pneumatic system.
Note: The test will have been successfully completed if the pressure has dropped by
6 mmHg, or less, during the 1-minute period.
33
Inflation time measurement
34
Deflation rate measurement
35
criterion is 197 to 203 bpm.
c. The defibrillator/monitor will display:
- 90 % SpO2
- 200 bpm
- alarm: “High Heart Rate/Pulse Rate limits violated” message will
display and the HR/PR area will flash, indicating pulse rate is above
default high alarm limit (medium priority alarm).
d. Press PULSE RATE selection button on the SpO2 simulator. The
PULSE RATE 60 LED will light.
e. The defibrillator/monitor will decrease to 60 and stabilize at 60bpm.
The test pass criterion is 57 to 63 bpm.
f. The defibrillator/monitor will display:
- 90 % SpO2
- 60 bpm
- no alarm
- low level modulation
7. Test #3: Modulation Level
a. Press %MODULATION selection button on the SpO2 simulator.
The %MODULATION LED will light.
b. The defibrillator/monitor’s waveform area will spike and stabilizes at a
higher modulation level.
c. The defibrillator/monitor will display:
- 90 % SpO2
- 60 bpm
- no alarm
d. Disconnect all equipments and turn off the defibrillator/monitor.
Respiration Operation
1. Connect the defibrillator/monitor to an AC power source.
2. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
3. Connect ECG lead cable to an appropriate terminal on the respiration simulator.
4. Connect the ECG cable to the ECG connector on the defibrillator/monitor’ front
panel.
5. Set the respiration simulator lead selection to lead I.
6. Set the respiration simulator to 120 breaths per minute.
7. After the normal power-up sequence, verify the following reactions:
a. The defibrillator/monitor displays respiration rate of 120 ±1 breaths per
minute.
b. Audible alarm will sound, “High Respiration Rate limits violated”
message will display and the Respiration numerical area will flash,
indicating a respiration rate is above default high alarm limits. (medium
priority alarm)
8. Decrease the respiration rate setting on Respiration simulator to 20 breaths per
minute.
a. Verify that the defibrillator/monitor displays the respiration rate of 20 ±1
36
breaths per minute.
Note: The accuracy of Respiration rate measurements is ±1 breaths per minute. In the
procedure above, add the tolerance of the simulator to the acceptable range of
readings.
Note: The respiration measure function does not operate if pads or paddles are
connected to the defibrillator/monitor. Verify the pads or the paddles are not
connected to the defibrillator/monitor before operates respiration measurement.
Temperature Operation
1. Connect the defibrillator/monitor to an AC power source.
2. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
3. Connect the temperature probe (supplied with the temperature simulator) to an
appropriate terminal on the temperature simulator.
4. Connect the temperature probes to the temperature connector 1 and 2 on the
defibrillator/monitor’s front panel.
5. Set Temperature simulator as follows:
Temperature: 37°C (98.0°F)
Probe type: YSI-400 series Temperature Probes (Probe accuracy: ±0.1°C)
6. After the normal power-up sequence, verify that the temperature reads 37°C ±0.1°C
(98.6°F ±0.2°F if Fahrenheit is selected for the temperature units).
7. Turn off the defibrillator/monitor.
IBP Operation
1. Connect the defibrillator/monitor to an AC power source.
2. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
3. Connect the IBP test cable to the IBP simulator, then turn on the IBP simulator.
4. Press 6, zero button on the IBP simulator, then press Enter button on the IBP
simulator.
5. Connect the other end of IBP test cables to IBP channel 1 and 2 connectors on the
defibrillator/monitor’s front panel.
6. The defibrillator/monitor will display IBP waveform’s baseline on the screen.
7. Rotate the Multi function knob to select IBP numeric area, then press Zero
setting soft key and press Enter soft key to perform the zero calibration.
8. If zero calibration is successful, the defibrillator/monitor will display the value “0” on
the IBP numerical area.
9. Press the 5, wave button on the IBP simulator, and then press Enter button on the
IBP simulator.
10. The defibrillator/monitor will:
‐ display IBP waveform on the screen.
37
‐ display systolic, mean, diastolic measurement values on the IBP numerical area.
Note: Initial values of IBP simulator are systolic 120mmHg, diastolic 80mmHg at
channel 1, systolic 120mmHg, diastolic 0mmHg at channel 2.
Capnography Operation
Note: The recommended Capnostat5 CO2 sensor should be compared against
calibration gas every 6 months. Check the calibration status before the test.
Note: The airway adapter zero is required when changing to a different style of airway
adapter.
38
4. Test #2: Flow Rate Accuracy
a. Connect the Sidestream CO2 module into the CO2 connector on the
defibrillator/monitor’ front panel.
b. Attach the sidestream sampling lime to the inlet port located on the metal
container on the front of the CO2 module. You will hear a click when
properly inserted.
c. Connect the calibrated flow meter to the exhaust port of the module after
the defibrillator/monitor is warmed up and all messages have cleared.
d. Verify that the flow rate is 50 ml/min ± 10ml. If the measured flow rate is
outside the specified limits, remove the CO2 module form use and contact
a Mediana Technical Service Department.
5. Test #3: Occlusion
a. Block the exhaust port while the sidestream CO2 sensor module is running.
b. Verify that the “CO2 – Occlusion or leak” message displays.
Safety Tests
The defibrillator/monitor safety tests meet the standards of, and are performed in
accordance with, IEC 60601-1, Clause 19 (Second Edition, 1988; Amendment 1,
1991-11, Amendment 2, 1995-03), EN60601-1 for instruments classified as Class I for
AC mains and Class II for DC mains and Type BF and Type CF of applied part.
This test checks the integrity of the power cord ground wire from the AC plug to the
instrument chassis ground. The current used for this test is less than or equal to 4 Volt
RMS, 50 to 60 Hz, and 25 Amperes.
Electrical Leakage
39
SFC: Single Fault Condition / RM: Reverse Mains/Lines Voltage
Note: Earth leakage current is measured under various conditions of the AC mains
and protective earth conductor. For each condition, the measured leakage
current must not exceed that indicated in Table 5.
Note: The analyzer leakage current indication must not exceed the values listed in Table
6. Enclosure Leakage Current Values.
This test measures patient leakage current in accordance with IEC60601-1, clause 19,
for Class I, type BF and type CF equipment. Patient leakage current in this test is
measured from any individual patient connection to earth (power ground).
Note: Patient leakage current is measured under various conditions of the AC mains
and protective earth conductor. For each condition, the measured leakage
current must not exceed that indicated in Table 7.
40
Note: This test requires a test cable for each patient connector. For example, the
ECG test cable consists of the ECG cable connector, with all conductors
shorted together, connected to a test lead from the electrical safety analyzer.
Test cables for SpO2 and temperature can be configured in a similar manner,
by wrapping each sensor end individually with aluminum foil filled with
conductive gel (only enough gel to ensure conductivity). Attach a wire to the foil
that is connected to a test lead from the electrical safety analyzer.
This test measures patient leakage current in accordance with IEC60601-1, clause 19,
for Class I, type BF and type CF equipment. In this test, 110% of mains voltage is
applied between each patient connection and earth (power ground). Patient leakage
current is then measured from any individual patient connection to earth.
Note: Keep the patient test cable length as short as possible during the leakage test.
Note: This test requires the same test cables for each patient connector as described
in paragraph “Patient Leakage Current”.
Note: Patient leakage current is measured with normal and reverse mains polarity.
For each condition, the measured leakage current must not exceed that
indicated in Table 8.
41
Table 8. Patient Leakage Current Values - Mains Voltage on Applied Part
Allowable Leakage Current (microamps)
Test Condition
Type BF Type CF
Normal polarity (SFC) 5000 50
Reverse polarity (SFCRM) 5000 50
This test measures patient auxiliary current in accordance with IEC60601-1, clause 19,
for Class I, type BF and type CF equipment. The applied voltage for IEC60601-1 the
voltage is 264 volts, 50 to 60 Hz. Patient auxiliary current is measured between each
ECG test lead and between each sensor connection for all possible connections.
Note: Keep the patient test cable length as short as possible during the leakage test.
Note: This test requires the same test cables for each patient connector as described
in paragraph “Patient Leakage Current”.
42
Table 10. Allowable Leakage Current
Allowable Leakage Current (microamps)
Test Condition
Type BF Type CF
Normal Condition (NC) 100 10
SFC Open Supply (OS) 500 50
SFC Open Earth (SFC OE) 500 50
Normal Condition RM (NCRM) 100 10
SFC Open Supply RM (SFC OSRM) 500 50
SRC Open Earth RM (SFC OERM) 500 50
43
Verification Check Sheet
44
ITEMS RESULTS REMARKS
PERFORMANCE TEST
-Deflation rate measurement Pass / Fail
Pulse Oximetry operation
- SpO2 90 ± 2% Pass / Fail Value: %
- Pulse rate 200 ± 3bpm (High priority alarm condition) Pass / Fail Value: bpm
- Pulse rate 60 ± 3bpm Pass / Fail Value: bpm
- Modulation level Pass / Fail
Respiration operation Pass / Fail
Temperature operation Pass / Fail
IBP operation Pass / Fail
CO2 operation
-Display accuracy 76 ± 2 mmHg (for Mainstream)
Pass / Fail Value: mmHg
-Display accuracy 76 ± 2 mmHg (for Sidestream)
-Flow rate Pass / Fail
-Occlusion Pass / Fail
SAFETY TEST
TEST CONDITIONS LIMIT(uA) RESULTS REMARKS
Earth leakage current (NC) 500 Pass / Fail Value: uA
Earth leakage current (SFC OS) 1000 Pass / Fail Value: uA
Earth leakage current (NCRM) 500 Pass / Fail Value: uA
Earth leakage current (SFC OSRM) 1000 Pass / Fail Value: uA
Enclosure leakage current (NC) 100 Pass / Fail Value: uA
Enclosure leakage current (OS) 500 Pass / Fail Value: uA
Enclosure leakage current (SFC OE) 500 Pass / Fail Value: uA
Enclosure leakage current (NCRM) 100 Pass / Fail Value: uA
Enclosure leakage current (SFC OSRM) 500 Pass / Fail Value: uA
Enclosure leakage current (SFC OERM) 500 Pass / Fail Value: uA
Patient leakage current (NC) 10 Pass / Fail Value: uA
Patient leakage current (OS) 50 Pass / Fail Value: uA
Patient leakage current (SFC OE) 50 Pass / Fail Value: uA
Patient leakage current (NCRM) 10 Pass / Fail Value: uA
Patient leakage current (SFC OSRM) 50 Pass / Fail Value: uA
Patient leakage current (SFC OERM) 50 Pass / Fail Value: uA
Mains voltage on applied part (SFC) 50 Pass / Fail Value: uA
Mains voltage on applied part (SFCRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG LL (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG LL (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG LL (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG LL (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG LL (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG LL (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG RA (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG RA (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG RA (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG RA (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG RA (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA–ECG RA (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL–ECG RA (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL–ECG RA (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL–ECG RA (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL–ECG RA (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL–ECG RA (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL–ECG RA (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (OS) 50 Pass / Fail Value: uA
45
SAFETY TEST
TEST CONDITIONS LIMIT(uA) RESULTS REMARKS
Patient auxiliary current ECG LA-TEMP (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-TEMP (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-TEMP(SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-TEMP (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO2 (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO2 (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO2 (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO2 (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO2 (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LA-SpO2 (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO2 (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO2 (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO2 (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO2 (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO2 (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG LL-SpO2 (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO2 (NC) 10 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO2 (OS) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO2 (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO2 (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO2 (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current ECG RA-SpO2 (SFC OERM) 50 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO2 (NC) 10 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO2 (OS) 50 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO2 (SFC OE) 50 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO2 (NCRM) 10 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO2 (SFC OSRM) 50 Pass / Fail Value: uA
Patient auxiliary current TEMP-SpO2 (SFC OERM) 50 Pass / Fail Value: uA
Remarks
NC: Normal Condition
NCRM: Normal Condition Reverse
SFC: Single Fault Condition
OS: Single Fault Condition (Open Line/Neutral)
OSRM: Single Fault Condition (Open Line/Neutral) Reverse
OE: Single Fault Condition (Open Earth)
OERM: Single Fault Condition (Open Earth) Reverse
46
Service Menu and Parameter Default Settings
General
This section discusses use of the Service menu to configure system setting, printer
setting, parameter default setting, ECG setting, Shock setting, AED mode setting, 3G
network setting and Wi-Fi network setting. Also this section explains briefly the
Parameter Default Settings.
Service Menu
The purpose of the Service menu (Figure 14) is to allow the authorized user to create
a Power-on default for the settings in effect each time the defibrillator/monitor is
powered on. Once the Service menu is entered, physiological monitoring is
terminated. The screen layouts do not display any information associated with normal
monitoring operation. Use the following procedure to configure the Service menu for
the defibrillator/monitor (also see Using the defibrillator/monitor section, of the
operator’s manual):
1. Soft key menus are displayed on the lower of the normal screen.
2. Press Setup soft key to display Setup menu (see Figure 12).
3. Rotate the Multi function knob to highlight Service menu in Setup menu, and
then press the Multi function knob to access the Service menu.
4. Three digits are displayed in the Level 2 menu as shown in Figure 12.
47
Figure 13. Entering the pass code
5. Rotate the Multi function knob to highlight the top of the digits. Press the Multi
function knob to enter pass code.
6. Rotate the Multi function knob until “0” appears, then press the Multi function
knob.
7. Repeat step 5-6 to enter all the pass code “0” “0” “0”.
8. The Service menu will now be present (Figure 14). The available Service menu
items are explained in Figure 14 ~ Figure 24 and Table 11 ~ Table 21. Make
changes to these menu items as desired by rotating and pressing the Multi
function knob.
9. Select SAVE soft key and select Yes using Multi function knob. All changes made
to the power-up defaults will be in effect the next time the defibrillator/monitor is
turned on.
10. Turn off the defibrillator/monitor, and then turn on the defibrillator/monitor again.
Note: The defibrillator/monitor must be powered off upon selecting “Save” and “Yes” to
save any changes into the defibrillator/monitor, and then the changes made to
the Power on defaults will be in effect next time the defibrillator/monitor is
powered up.
48
System Setting I
49
System Setting I
Level1 Level2 Level3
Audio pause time 90 sec
120 sec
Return
Audio off time 3 min
5 min
10 min
Return
Alarm reminder tone OFF
3 min
10 min
Return
Alarm active at power up Enable
Disable
Service menu data Save service menu data Start
Return
Load service menu data Start
Return
Auto block erase Enable
Disable
Return
Software information
This menu displays software versions of Main software, Display software, Analog
software, Charger, H/V, ECG, SpO2, TEMP, NIBP Main, NIBP Sub, CO2, and IBP.
Software update
Main software update
The main software can be updated in this menu.
Display software update
The display software (screen display and alarm messages) can be updated in this
menu.
Analog software update
The analog software (for ECG analyzes CPU) can be updated in this menu.
Voice prompt
The voice prompt can be updated from SD card via this menu.
Parameter color
The color of each parameter’s waveform and measured value can be set in this menu.
Brightness
The brightness can be adjusted in this menu. The brightness can be set at level 1 to 4.
Volume
The volume of alarm, beep, button, charging tone, voice prompt, CPR metronome and
pacing alarm sound can be set in this menu.
50
Audio pause time
Pressing the ALARM button temporarily off alarms for the period selected in this menu.
The factory default of alarm off period is 30 seconds.
51
System Setting II
52
System Setting II
Level1 Level2 Level3
Device ID Confirm
Return
12 lead institution name Confirm
Return
12 lead location code Confirm
Return
Communication phone numbers Confirm
Return
Manual mode protection Enable
Confirm
Password
Return
Manual mode locking passcode Pass number 1
Pass number 2
Pass number 3
Confirm
Return
Pacing mode protection Disable
Confirm
Password
Confirm
Return
Pacing mode locking passcode Pass number 1
Pass number 2
Pass number 3
Confirm
Return
Service menu passcode Pass number 1
Pass number 2
Pass number 3
Confirm
Return
Auto self-test
Auto self-test configuration can be set in this menu.
Self-test time
The time of self test operation can be set in this menu.
Self-test interval
The self test time interval can be set as Off, 24, 48 and 72 hours in this menu. If Self-
test interval is set to Off, the self-test is not performed.
53
Device installation ID
Device installation ID can be entered in this menu.
Device ID
Device ID can be entered in this menu.
54
Printer Setting
Print on alarm
If Print on alarm is set to On, the defibrillator/monitor will automatically print out
information before and after 10 seconds when a physiological alarm condition is
activated.
Print on charge
If Print on charge is set to On, the defibrillator/monitor will automatically print out
information before and after 10 seconds when energy is charged for shock.
Print on shock
If Print on shock is set to On, the defibrillator/monitor will automatically print out
55
information before and after 10 seconds when a shock is generated.
Print on mark
If Print on mark is set to On, the defibrillator/monitor will automatically print out when
Clinical action list is selected in Setup menu.
Print on BP
If Print on BP is set to On, the defibrillator/monitor will automatically print out when the
NIBP value is measured.
Print on self-test
If Print on self-test is set to On, the defibrillator/monitor will automatically print out the
result of Self-Test when Self-Test is performed.
56
Parameter Default Setting
Level1 Level2 Level3
Airway (high limit) : 1~150
Impedance (low limit) : 3~119
Airway (low limit) : 0~149
Return
RESP ON/OFF On
Off
Power on waveform select 1st Waveform I, II, III, Pads
2nd Waveform I, II, III, Pads, SpO2,
RESP, IBP1, IBP2,
EtCO2, Return
3rd Waveform I, II, III, Pads, SpO2,
RESP, IBP1, IBP2,
EtCO2, Return
4th Waveform I, II, III, Pads, SpO2,
RESP, IBP1, IBP2,
EtCO2, Return
Return
SpO2 alarm Limit (high limit) : 21~100
(low limit) : 20~99
Return
NIBP alarm Limit – SYS Adult/Pedi (high limit):
35 to 270 mmHg
(4.6 to 36.0 kPa)
Neonatal (high limit)
45 to 130 mmHg
(6.0 to 17.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
Adult/pediatric
30 to 265 mmHg
(4.0 to 35.3 kPa)
Neonatal
40 to 125 mmHg
(5.3 to 16.6 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
NIBP alarm Limit – MAP Adult/Pedi (high limit):
25 to 260 mmHg
(3.3 to 34.7 kPa)
Neonatal (high limit):
35 to 110 mmHg
(4.7 to 14.7 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
Adult/Pedi (low limit):
20 to 255 mmHg
(2.7 to 34.0 kPa)
Neonatal (low limit):
30 to 105 mmHg
(4.0 to 14.0 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
NIBP alarm Limit – DIA Adult/Pedi (high limit):
15 to 250 mmHg
(2.0 to 33.3 kPa)
Neonatal (high limit):
10 to 245 mmHg
57
Parameter Default Setting
Level1 Level2 Level3
(1.3 to 32.6 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
Adult/Pedi (low limit):
10 to 245 (mmHg)
1.3 to 32.6 (kPa)
Neonatal (low limit):
25 to 85 mmHg
(3.3 to 12.0 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
Return
NIBP inflate Pressure Adult/Pedi: 120 mmHg
Adult/Pedi: 140 mmHg
Adult/Pedi: 160 mmHg
Adult/Pedi: 180 mmHg
Adult/Pedi: 200 mmHg
Adult/Pedi: 220 mmHg
Adult/Pedi: 240 mmHg
Adult/Pedi: 260 mmHg
Adult/Pedi: 280 mmHg
Neonatal: 80 mmHg
Neonatal: 100 mmHg
Neonatal: 120 mmHg
Neonatal: 140 mmHg
Return
NIBP patient mode Adult/Pedi
Neonatal
NIBP unit mmHg
kPa
NIBP test mode Pressure sensor accuracy test Start
Return
Air leakage test Start
Return
Inflation time measurement Start
Return
Deflation rate measurement Start
Return
Return
Temp alarm limit Temp 1(high limit):
0.1 to 50.0 °C
(32.2 to 122.0 °F)
(0.1°C, 0.1° or 0.2°F steps)
Temp 1(low limit):
0.0 to 49.9 °C
(32.0 to 121.8 °F)
(0.1°C, 0.1° or 0.2°F steps)
Temp 2(high limit):
0.1 to 50.0 °C
(32.2 to 122.0 °F)
(0.1°C, 0.1° or 0.2°F steps)
Temp 2(low limit):
58
Parameter Default Setting
Level1 Level2 Level3
0.0 to 49.9 °C
(32.0 to 121.8 °F)
(0.1°C, 0.1° or 0.2°F steps)
Return
Temp unit °C
°F
IBP alarm limit P1 SYS(high limit):
-45 to 300 mmHg
(-6 to 40 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P1 SYS(low limit):
-50 to 295mmHg
(-6.7 to 39.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P1 MEAN(high limit):
-45 to 300 mmHg
(-6 to 40 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P1 MEAN(low limit):
-50 to 295 mmHg
(-6.7 to 39.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P1 DIA(high limit):
-45 to 300 mmHg
(-6 to 40 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P1 DIA(low limit):
-50 to 295 mmHg
(-6.7 to 39.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P2 SYS(high limit):
-45 to 300 mmHg
(-6 to 40 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P2 SYS(low limit):
-50 to 295mmHg
(-6.7 to 39.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P2 MEAN(high limit):
-45 to 300 mmHg
(-6 to 40 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P2 MEAN(low limit):
-50 to 295mmHg
(-6.7 to 39.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P2 DIA(high limit):
-45 to 300 mmHg
(-6 to 40 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
P2 DIA(low limit):
-50 to 295mmHg
59
Parameter Default Setting
Level1 Level2 Level3
(-6.7 to 39.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
Return
IBP unit mmHg
kPa
IBP1 label P1
ABP
Return
IBP2 label P2
CVP
PAP
LAP
Return
EtCO2 alarm limit EtCO2(high limit):
1 to150 mmHg
(0.13 to 20.0 kPa), (0 to 20.0 %)
(1 mmHg, 0.13 kPa, 0.13 % steps)
EtCO2(low limit):
0 to 149 mmHg
(0 to 19.9 kPa), (0 to 19.9 %)
(1 mmHg, 0.13 kPa, 0.13 % steps)
InCO2(high limit):
1 to 20 mmHg
(0.13 to 2.7 kPa), (0.13 to 2.6 %)
(1 mmHg, 0.13 kPa, 0.13 % steps)
InCO2(low limit):
0 to 19 mmHg
(0 to 2.5 kPa), (0 to 2.5 %)
(1 mmHg, 0.13 kPa, 0.13 % steps)
Return
EtCO2 unit mmHg
kPa
%
Return
RESP On/Off
The custom-default respiration function can be set in this menu.
60
SpO2 alarm limit
The custom-default SpO2 alarm limit can be set in this menu.
NIBP unit
The custom-default NIBP unit can be selected as mmHg or kPa.
Temp unit
The custom-default Temperature unit can be selected as ℃ or ℉.
IBP unit
The custom-default IBP unit can be selected as mmHg or kPa.
61
IBP1 label
The IBP1 label can be selected as P1 or ABP in this menu.
IBP2 label
The IBP2 label can be selected as P2, CVP, PAP or LAP in this menu.
EtCO2 unit
The custom-default EtCO2 unit can be selected as mmHg, kPa or %.
62
ECG Setting
AC line filter
The AC line filter can be set to 50Hz or 60Hz in this menu.
Pacer detection
If Pacer detection is set to Enable, defibrillator/monitor detects pacer pulses of ±2mV
to ±700mV with pulse widths of 0.1 to 2ms and rise times 10% of width not to exceed
100ms.
ECG size
The 1st waveform ECG gain is set to x0.5, x1, x1.5, x2 or x3.
63
ECG filter
The filter bandwidth frequency for ECG signal display can be set to 0.5Hz ~ 21Hz or
0.05Hz ~ 40Hz or 1Hz~21Hz.
Message display
If Message display is set to Enable, defibrillator/monitor sets ECG waveform message
to display.
64
12 Lead ECG Setting
Level1 Level2 Level3
12 lead print speed 25mm/s
50mm/s
Auto transmit
If the Auto transmit is set to On, the reading result of 12 lead ECG signal is sent by
3G or Wi-Fi.
65
Shock Setting
66
Elapsed time display
If Elapsed time display is set to Enable, the elapsed time of defibrillator/monitor
operation is displayed on the main screen each mode.
Message display
If Message display is set to On, the defibrillation guide line is displayed in the manual
mode.
Voice prompt
If Voice prompt is set to Enable, the voice prompt is played in the manual mode.
Auto change
If Auto change is set to On, the shock energy is auto changed when the shockable
rhythm is detected.
67
AED Mode Setting
68
Setup menu
If Setup menu is set to Enable, Setup soft key menu is enabled in AED mode.
12 Lead menu
If 12 Lead menu is set to Enable, 12 Lead soft key menu is enabled in AED mode.
Waveform display
If Waveform display is set to Enable, the waveforms from measurement parameters
are displayed in AED mode.
Icon display
If Icon display is set to Enable, the icons for AED guide line are displayed in AED
mode.
Value display
If Value Display is set to Enable, the measurement values are displayed in AED mode.
69
Pacing Mode Setting
Fixed pacing
If Fixed pacing is set to On, the defibrillator/monitor performs pacing in pacing mode.
70
3G Network Setting
3G
3G for 3G network connection can be set in this menu.
Destination IP address
Destination IP address for 3G network connection can be set in this menu.
Destination port
Destination port for 3G network connection can be set in this menu.
71
USIM1, USIM2
USIM for 3G network connection can be set in this menu.
APN
SSID for 3G network connection can be set in this menu.
Authentication protocol
Authentication protocol for 3G network connection can be set in this menu.
Authentication ID
Authentication ID for 3G network connection can be set in this menu.
Authentication password
Authentication password for 3G network connection can be set in this menu.
72
Configuration Menu – Wi-Fi Network Setting
Level1 Level2 Level3
Wireless security mode None
WEP
WPA
WPA2
EAP
Return
WEP key length 64bit (10 hex digits)
128 bit (26 hex digits)
Shared key Confirm
Return
Wi-Fi
Wi-Fi is set to On or Off on this menu.
IP configuration
IP address, Subnet mask and Gateway address for Wi-Fi network connection can be
set in this menu.
Destination IP address
IP address for Wi-Fi network connection can be set in this menu.
Destination port
Port number for Wi-Fi network connection can be set in this menu.
Wireless channel
Channel for Wi-Fi network connection can be set from channel 1 to channel13 in this
menu.
Shared key
WPA share key for Wi-Fi network connection can be set in this menu.
73
Factory Default Settings
Parameter ranges and Parameter Default Settings in Service menu are divided into 2
groups, adult/pediatric and neonatal as described in Table 22. Parameter ranges and
factory defaults.
The patient mode is preset to “Adult/Pediatric” mode. Alarm limits settings will be
automatically changed to the default settings for each patient mode as the mode is
changed to Adult/Pediatric or Neonatal mode.
74
Parameter Ranges / Selections (Adjust step) Factory Defaults
(5 mmHg, 0.6 or 0.7kPa steps)
NIBP SYS Low Alarm Limits adult/pediatric adult/pediatric
30 to 265 mmHg 90 mmHg
(4.0 to 35.3 kPa) (12.0 kPa)
neonatal neonatal
40 to 125 mmHg 40 mmHg
(5.3 to 16.7 kPa) (5.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
NIBP DIA High Alarm Limits adult/pediatric adult/pediatric
15 to 250 mmHg 90 mmHg
(2.0 to 33.3 kPa) (12.0 kPa)
neonatal neonatal
25 to 90 mmHg 60 mmHg
(3.3 to 12.0 kPa) (8.0 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
NIBP DIA Low Alarm Limits adult/pediatric adult/pediatric
10 to 245 mmHg 50 mmHg
(1.3 to 32.7 kPa) (6.7 kPa)
neonatal neonatal
20 to 85 mmHg 20 mmHg
(2.7 to 11.3 kPa) (2.6 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
NIBP MAP High Alarm Limits adult/pediatric adult/pediatric
25 to 260 mmHg 110 mmHg
(3.3 to 34.7 kPa) (14.7 kPa)
neonatal neonatal
35 to 110 mmHg 70 mmHg
(4.7 to 14.7 kPa) (9.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
NIBP MAP Low Alarm Limits adult/pediatric adult/pediatric
20 to 255 mmHg 60 mmHg
(2.7 to 34.0 kPa) (8.0 kPa)
neonatal neonatal
30 to 105 mmHg 30 mmHg
(4.0 to 14.0 kPa) (4.0 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP1 MENU
IBP1: Scale Auto, 0~50mmHg, 0~100mmHg, 0~200mmHg, Auto
0~300mmHg
IBP1: Label P1, ABP P1
IBP 1 SYS High Alarm Limits adult/pediatric adult/pediatric
-45 to 300 mmHg 120 mmHg
(-6 to 40 kPa) (16.0 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 1 SYS Low Alarm Limits adult/pediatric adult/pediatric
-50 to 295mmHg 70 mmHg
(-6.7 to 39.3 kPa) (9.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 1 DIA High Alarm Limits adult/pediatric adult/pediatric
-45 to 300 mmHg 70 mmHg
(-6 to 40 kPa) (9.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 1 DIA Low Alarm Limits adult/pediatric adult/pediatric
-50 to 295 mmHg 40 mmHg
75
Parameter Ranges / Selections (Adjust step) Factory Defaults
(-6.7 to 39.3 kPa) (5.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 1 MEAN High Alarm Limits adult/pediatric adult/pediatric
-45 to 300 mmHg 90 mmHg
(-6 to 40 kPa) (12.0 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 1 MEAN Low Alarm Limits adult/pediatric adult/pediatric
-50 to 295 mmHg 50 mmHg
(-6.7 to 39.3 kPa) (6.7 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP2 MENU
IBP2: Scale Auto, 0~50mmHg, 0~100mmHg, 0~200mmHg, Auto
0~300mmHg
IBP2: Label P2, CVP, PAP,LAP P2
IBP 2 SYS High Alarm Limits adult/pediatric adult/pediatric
-45 to 300 mmHg 120 mmHg
(-6 to 40 kPa) (16.0 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 2 SYS Low Alarm Limits adult/pediatric adult/pediatric
-50 to 295mmHg 70 mmHg
(-6.7 to 39.3 kPa) (9.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 2 DIA High Alarm Limits adult/pediatric adult/pediatric
-45 to 300 mmHg 70 mmHg
(-6 to 40 kPa) (9.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 2 DIA Low Alarm Limits adult/pediatricl adult/pediatric
-50 to 295mmHg 40 mmHg
(-6.7 to 39.3 kPa) (5.3 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 2 MEAN High Alarm Limits adult/pediatric adult/pediatric
-45 to 300 mmHg 90 mmHg
(-6 to 40 kPa) (12.0 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP 2 MEAN Low Alarm Limits adult/pediatric adult/pediatric
-50 to 295mmHg 50 mmHg
(-6.7 to 39.3 kPa) (6.7 kPa)
(5 mmHg, 0.6 or 0.7kPa steps)
IBP unit* mmHg, kPa mmHg
EtCO2
EtCO2 On, Off ON
N2O Gas On, Off OFF
O2 Gas On, Off OFF
Scale 0~40 mmHg, 0~60 mmHg, 0~80 mmHg, Auto Auto
EtCO2 High Alarm Limits adult/pediatric adult/pediatric
1 to 80 mmHg 80 mmHg
(0.13 to 10.7 kPa), (0.13 to 10.5 %) (10.7 kPa), (10.5 %)
(1 mmHg, 0.13 kPa, 0.13 % steps)
EtCO2 Low Alarm Limits adult/pediatric adult/pediatric
0 to 79 mmHg 0 mmHg
(0 to 10.5 kPa), (0 to 10.4 %) (0 kPa), (0 %)
(1 mmHg, 0.13 kPa, 0.13 % steps)
InCO2 High Alarm Limits adult/pediatric adult/pediatric
1 to 20 mmHg 20 mmHg
76
Parameter Ranges / Selections (Adjust step) Factory Defaults
(0.13 to 2.7 kPa), (0.13 to 2.6 %) (2.7 kPa), (2.6 %)
(1 mmHg, 0.13 kPa, 0.13 % steps)
InCO2 Low Alarm Limits adult/pediatric adult/pediatric
0 to 19 mmHg 0 mmHg
(0 to 2.5 kPa), (0 to 2.5 %) (0 kPa), (0 %)
(1 mmHg, 0.13 kPa, 0.13 % steps)
EtCO2 unit* mmHg, kPa, % mmHg
Temperature MENU
TEMP1 High Alarm Limits 0.1 to 50.0 °C 39.0 °C
(32.2 to 122.0 °F) (102.2 °F)
(0.1°C, 0.1° or 0.2°F steps)
TEMP1 Low Alarm Limits 0.0 to 49.9 °C 36.0 °C
(32.0 to 121.8 °F) (96.8 °F)
(0.1°C, 0.1° or 0.2°F steps)
TEMP2 High Alarm Limits 0.1 to 50.0 °C 39.0 °C
(32.2 to 122.0 °F) (102.2 °F)
(0.1°C, 0.1° or 0.2°F steps)
TEMP2 Low Alarm Limits 0.0 to 49.9 °C 36.0 °C
(32.0 to 121.8 °F) (96.8 °F)
(0.1°C, 0.1° or 0.2°F steps)
TEMP Unit* °C, °F °C
Respiration MENU
Respiration ON ON
OFF
RR Source Auto Auto
Impedance
Airway
Size x0.5, x1, x1.5, x2 x1
Apnea time setting Off, 10 seconds, 20 seconds, 30 seconds 10 seconds
40 seconds, 50 seconds, 60 seconds
RR High Limits 4 to 120 BPM (1 BPM step) 30 BPM
RR Low Limits 3 to 119 BPM (1 BPM step) 8 BPM
Manual, AED, Pacing, Monitor Mode MENU
Alarm Volume 1, 2, 3, 4, 5, 6, 7, 8 5
Beep Volume OFF, 1, 2, 3, 4, 5, 6, 7 4
Button Volume OFF, 1, 2, 3, 4, 5, 6, 7 4
Charging alarm Volume 1, 2, 3, 4, 5, 6, 7, 8 4
Voice prompt Volume 1, 2, 3, 4, 5, 6, 7, 8 4
CPR metronome Volume OFF, 1, 2, 3, 4, 5, 6, 7 4
Pacing alarm Volume OFF, 1, 2, 3, 4, 5, 6, 7 2
2nd Waveform ECG I, ECG II, ECG III, ECG aVL, ECG aVF, ECG SpO2 Waveform
aVR, ECG V1, ECG V2, ECG V3, ECG V4, ECG V5,
ECG V6, Pads, SpO2, RESP, IBP1, IBP2, EtCO2
3rd Waveform ECG I, ECG II, ECG III, ECG aVL, ECG aVF, ECG IBP1 Waveform – P1
aVR, ECG V1, ECG V2, ECG V3, ECG V4, ECG V5,
ECG V6, Pads, SpO2, RESP, IBP1, IBP2, EtCO2
4th Waveform ECG I, ECG II, ECG III, ECG aVL, ECG aVF, ECG EtCO2 Waveform
aVR, ECG V1, ECG V2, ECG V3, ECG V4, ECG V5,
ECG V6, Pads, SpO2, RESP, IBP1, IBP2, EtCO2
Print on alarm** On, Off Off
Print on charge** On, Off Off
Print on shock** On, Off Off
Print on mark** On, Off Off
77
Parameter Ranges / Selections (Adjust step) Factory Defaults
Print on BP** On, Off Off
Print on self-test** On, Off On
12lead auto printing** On, Off On
Analyze auto printing** On, Off Off
Elapsed time display* Enable, Disable Enable
CPR guide display* Enable, Disable Enable
Message display* Enable, Disable Enable
Voice prompt* Enable, Disable Enable
Basic auto energy escalation* Enable, Disable Enable
Above 200J setting* Enable, Disable Disable
Manual shock energy – 1st 1~200J (Adult) 125J (Adult)
energy* 1~100J (Pediatric) 50J (Pediatric)
Manual shock energy – 2nd 1~200J (Adult) 150J (Adult)
energy* 1~100J (Pediatric) 50J (Pediatric)
Manual shock energy – 3rd 1~200J (Adult) 200J (Adult)
energy* 1~100J (Pediatric) 50J (Pediatric)
Analyze continuous mode* On, Off Enable
Auto charge* On, Off On
Remain in sync mode after On, Off On
shock*
Charging hold time* 15 sec, 60 sec 60 sec
ECG display with paddle in On, Off On
openair*
Setup Menu* Enable, Disable Disable
Alarm Limit Menu* Enable, Disable Disable
12 Lead Menu* Enable, Disable Disable
AED display setting – Enable, Disable Enable
Waveform display*
AED display setting – Enable, Disable Enable
Icon display*
AED display setting – Enable, Disable Enable
Value display*
AED display setting – Enable, Disable Disable
AED step 1 display*
AED display setting – Enable, Disable Disable
AED step 2 display*
AED display setting – Enable, Disable Disable
AED step 3 display*
AED shock energy – Enable, Disable Disable
Above 200J setting*
AED shock energy – 1~200J (Adult) 125J (Adult)
1st energy* 1~100J (Pediatric) 50J (Pediatric)
AED shock energy – 1~200J (Adult) 150J (Adult)
2nd energy* 1~100J (Pediatric) 50J (Pediatric)
AED shock energy – 1~200J (Adult) 200J (Adult)
3rd energy* 1~100J (Pediatric) 50J (Pediatric)
Fixed Pacing* On, Off Off
Pace Pulse Rate* 30 ~ 180 bpm 70 bpm
Default Pacing Energy* 0, 30, 70, 100, 140 mA 0 mA
Others
Parameters Color - ECG* 00FA00(Green), FAFAFA(White), 64FAFA(Cyan), 00FA00(Green)
6464FA(Light blue), FA96FA(Pink), FA0000(Red),
FAFA00(Yellow), 9696C8(Light purple), FA9600,
78
Parameter Ranges / Selections (Adjust step) Factory Defaults
0096FA, FA00FA, 96FA32, C8C8C8, 0096C8,
C89600, C8C832
Parameters Color - NIBP* 00FA00(Green), FAFAFA(White), 64FAFA(Cyan), FAFAFA(White)
6464FA(Light blue), FA96FA(Pink), FA0000(Red),
FAFA00(Yellow), 9696C8(Light purple), FA9600,
0096FA, FA00FA, 96FA32, C8C8C8, 0096C8,
C89600, C8C832
Parameters Color - SpO2* 00FA00(Green), FAFAFA(White), 64FAFA(Cyan), 64FAFA(Cyan)
6464FA(Light blue), FA96FA(Pink), FA0000(Red),
FAFA00(Yellow), 9696C8(Light purple), FA9600,
0096FA, FA00FA, 96FA32, C8C8C8, 0096C8,
C89600, C8C832
Parameters Color - RESP* 00FA00(Green), FAFAFA(White), 64FAFA(Cyan), 9696C8(Light blue)
6464FA(Light blue), FA96FA(Pink), FA0000(Red),
FAFA00(Yellow), 9696C8(Light purple), FA9600,
0096FA, FA00FA, 96FA32, C8C8C8, 0096C8,
C89600, C8C832
Parameters Color - TEMP* 00FA00(Green), FAFAFA(White), 64FAFA(Cyan), FA96FA(Pink)
6464FA(Light blue), FA96FA(Pink), FA0000(Red),
FAFA00(Yellow), 9696C8(Light purple), FA9600,
0096FA, FA00FA, 96FA32, C8C8C8, 0096C8,
C89600, C8C832
Parameters Color - IBP 1* 00FA00(Green), FAFAFA(White), 64FAFA(Cyan), FA0000(Red)
6464FA(Light blue), FA96FA(Pink), FA0000(Red),
FAFA00(Yellow), 9696C8(Light purple), FA9600,
0096FA, FA00FA, 96FA32, C8C8C8, 0096C8,
C89600, C8C832
Parameters Color - IBP 2* 00FA00(Green), FAFAFA(White), 64FAFA(Cyan), FAFA00(Yellow)
6464FA(Light blue), FA96FA(Pink), FA0000(Red),
FAFA00(Yellow), 9696C8(Light purple), FA9600,
0096FA, FA00FA, 96FA32, C8C8C8, 0096C8,
C89600, C8C832
Parameters Color - EtCO2* 00FA00(Green), FAFAFA(White), 64FAFA(Cyan), 9696C8(Light purple)
6464FA(Light blue), FA96FA(Pink), FA0000(Red),
FAFA00(Yellow), 9696C8(Light purple), FA9600,
0096FA, FA00FA, 96FA32, C8C8C8, 0096C8,
C89600, C8C832
Brightness 1, 2, 3, 4 4
Audio pause time* 30 sec, 60 sec, 90 sec, 120 sec 60 sec
Audio off time* 3 min, 5 min, 10 min, Indefinite Indefinite
Alarm reminder tone* 3 min, 10 min, Off Off
Alarm activate at power up* Enable, Disable Enable
Auto self-test : 0 ~ 23 o‘clock 0 o‘clock
Self-test time*
Auto self-test : 24 hours, 48 hours, 72 hours, Off Off
Self-test time interval*
Auto self-test : 1~10J 10 J
Internal shock energy*
Clinical action list* None, CPR, EPI, Atrop, DOPA, PHEN, Bicarb, Aspirin, None
Oxygen, IV, Morphine, Valium, β-block, LIDO, Mag
Sulf, Thrombo, Sedation, Heparin, Procain, Cordar,
Thiamine, Dilantin, Intubate, Narcan, Atrovent,
Adenosin, Fentanyl, Digoxin, Vasopr, Dextrose,
Paralytic, Nitro, Albuterol, Amrinon, Benadryl,
79
Parameter Ranges / Selections (Adjust step) Factory Defaults
Demerol, Oral Glu, Lasix, Calcium
Manual mode protection* Disable, Confirm, Passcode Disable
Manual mode locking 000~999 (Passcode)
passcode*
Pacing mode protection* Disable, Confirm, Passcode Disable
Pacing mode locking 000~999 (Passcode)
passcode*
Service menu passcode* 000~999 (Passcode)
Print on charge** On, Off Off
Print on mark** On, Off Off
Print on BP** On, Off Off
Print on self-test** On, Off Off
Power on waveform select – ECG I, ECG II, ECG III, Pads ECG I
1st waveform*
Power on waveform select – ECG I, ECG II, ECG III, Pads, SpO2, RESP, IBP1, SpO2
2nd waveform* IBP2, EtCO2
Power on waveform select – ECG I, ECG II, ECG III, Pads, SpO2, RESP, IBP1, IBP1
3rd waveform* IBP2, EtCO2
Power on waveform select – ECG I, ECG II, ECG III, Pads, SpO2, RESP, IBP1, EtCO2
4th waveform* IBP2, EtCO2
Wi-Fi* On, Off On
Wireless Channel* Channel 1 ~ 13, Auto Channel 1
Wireless security mode* None, WEP, WPA, WPA2, EAP None
WEP key length* 64 bit (10 hex digits), 128 bit (24 hex digits) 64 bit (10 hex digits)
Outer Authentication* EAP-FAST, EAP-TLS, EAP-TTLS, EAP-PEAP EAP-FAST
Inner Authentication* EAP-MSCHAP, EAP-GTC EAP-MSCHAP
3G* On, Off On
USIM1* Enable, Disable Enable
Authentication protocol* None, PAP, CHAP PAP
USIM2* Enable, Disable Enable
Note: An asterisk (*) by a parameter in the above table indicates that the parameter can only be changed by authorized
personnel as described in the service manual
Note: Asterisks (**) by a parameter in the above table indicate the settings only when an optional printer is installed in the
defibrillator/monitor.
80
Software Update
General
This section is for the purpose for reloading software into the defibrillator/monitor
when the possibility of corrupted software exists, or updating software with a new
system revision (system/device version). Call Mediana Technical Service Department
for the latest version of software utility required.
Equipment Needed
Table 23 lists the equipment required for software update.
How to Update
1. Connect the defibrillator/monitor to an AC power source.
2. Connect a SD memory card containing the software to the SD card port on the right
panel of the defibrillator/monitor.
3. Rotate the Mode select knob to monitor mode to turn on the defibrillator/monitor.
4. Press Setup soft key then select Service menu using Multi function knob.
5. Enter the pass code.
6. The defibrillator/monitor will display the software update screen as shown below
Figure 25.
81
7. Choose Software update menu and then select Main software menu or Analog
software menu or Display software menu or Voice prompt menu. In Main
software menu, the main CPU software for various parameters can be updated. In
Display software menu, the software for screen display, alarm messages. In
Analog software menu, the software for ECG analysis can be updated. In Voice
prompt menu, the voice prompt can be updated.
82
Troubleshooting
General
This section provides information that can be helpful in troubleshooting the
defibrillator/monitor.
83
Troubleshooting Guide
Problems with the defibrillator/monitor are separated into categories for further
troubleshooting instructions.
Note: Taking the recommended actions discussed in this section will correct the
majority of problems you will encounter. However, problems not covered here
can be resolved by calling Mediana Technical Service Department.
84
Categories Symptoms
11.3: Poor ECG signal quality (noisy trace, wandering baseline,
etc.) from signal acquired through paddles/pads.
11.4: Solid flat line - no waveform, no Lead Off in operation.
11.5: ECG value or waveform does not display.
11.6: Lead select button does not respond.
11.7: One or more controls do not respond (e.g., select lead or soft
keys).
12.1: No display on the numeric area or waveform though sensor
12. IBP
is in normal operation.
13.1: Pacing does not start.
13. Defibrillation 13.2: Pacing (energy output) failure.
and Pacing 13.3: HV capacitor charging does not start.
13.4: Defibrillation failure.
14.1: Paper won't move.
14.2: Paper moves then stops.
14. Printer
14.3: Paper moves but printing is faint or absent.
14.4: Loud buzzing or grinding noise.
85
1. General
Symptom 1.1: The report data and customer settings are not saved.
Checkpoint Action Remark
CPU board Replace CPU board.
Main board Replace main board.
Note: If the report data were deleted in setup menu, the deleted data cannot be
recovered.
Symptom 1.2: The date and time printed on the recording paper is incorrect.
Checkpoint Action Remark
Date and time setting Reset the date and time in the
Date and Time menu.
Coin cell battery 1. If the Coin cell battery is
disconnected, connect Coin cell
battery properly.
2. If the Coin cell battery is
rusted, replace the Coin cell.
Main board or CPU board Replace main board or CPU
board.
Symptom 1.3: The screen does not turns on or the screen stop working with
buzzer sound though the defibrillator/monitor turns on.
Checkpoint Action Remark
CPU board Replace CPU board.
Main board Replace main board.
2. Power
Power problems are related to AC/DC and/or Battery as follows. If the action requires
replacement of the components, refer to the Disassembly Guide section.
Symptom 2.1: Defibrillator/monitor does not turn on though rotate the Mode
select knob.
Checkpoint Action Remark
AC/DC line power cord Check AC/DC line power cord is
connected properly.
Battery If battery is depleted or defective,
replace battery.
AC power supply (SMPS) or Replace AC power supply
charger board (SMPS) or charger board.
Main board Replace main board.
86
Symptom 2.6: The defibrillator/monitor is suddenly turned off even thought the
AC power is connected.
Checkpoint Action Remark
AC/DC line power cord Turn off the device then turn on
again to check the AC power is
supplied properly.
AC power supply (SMPS) or Replace AC power supply
charger board (SMPS) or charger board.
87
Symptom 2.7: The defibrillator/monitor does not turn on though external DC
power is connected.
Checkpoint Action Remark
External DC module Replace External DC module
1. If DC connector cable is
disconnected, connect the cable
DC connector cable properly.
2. If DC connector cable is cut,
replace the cable.
Charger board Replace charger board.
3. Display
88
4. Sound
5. Controls
Symptom 5.1: Soft keys, Multi function knob and menu keys control failure.
Checkpoint Action Remark
Soft key and Multi function knob 1. If the cable is disconnected,
connection cable connect the cable properly.
2. If the cable is cut, change the
cable.
Front board and main board 1. If the cable is disconnected,
connection cable connect the cable properly.
2. If the cable is cut, change the
cables.
Front board Replace front board.
CPU board Replace CPU board.
Main board Replace main board.
6. NIBP
7. SpO2
90
8. Temperature
9. Respiration
10. Capnography
Symptom 10.1: Flow rates fall and function is not working properly.
Checkpoint Action Remark
Pump If pump is not working properly,
check CO2 module.
EtCO2 isolation board Replace EtCO2 isolation board.
Main board Replace main board.
11. ECG
Symptom 11.1: QRS beeper inaudible or beeps do not occur with each QRS
complex.
Checkpoint Action Remark
Beep volume setting If the QRS volume is configured
to OFF or the setting is too low,
configure the beep volume.
Lead If the amplitude of the QRS
complex is too small to detect,
select a different lead.
91
Symptom 11.2: Poor ECG signal quality (noisy trace, wandering baseline, etc.)
from signal acquired through monitoring electrodes.
Checkpoint Action Remark
Electrodes connection with patient 1. Check that the monitoring
electrodes are properly applied.
If necessary, prepare the
patient's skin and apply new
electrodes.
2. Check the date code on the
electrodes. Do not open the
electrode package until
immediately prior to use.
Power noise Check the connecting condition
between ground and product.
ECG cable Run the Operational check with
the ECG cable. If the test fails,
run it without the ECG cable. If
the test passes, replace the
cable. If not, remove the device
from use and call for service.
Symptom 11.3: Poor ECG signal quality (noisy trace, wandering baseline, etc.)
from signal acquired through paddles/pads.
Checkpoint Action Remark
Paddles/pads connection with Ensure proper skin preparation
patient. and correction application. If
necessary, apply new pads.
Pads condition If the pads are dried-out, check
the date code on the pads. Do
not open the pads package until
immediately prior to use.
Power noise Check the connecting condition
between ground and product.
Paddles/pads cable 1. If the cable is disconnected,
connect the cable properly.
2. If the cable is cut, change the
cable.
92
Symptom 11.7: One or more controls do not respond (e.g., select lead or soft
keys).
Checkpoint Action Remark
Front membrane cable 1. If the cable is disconnected,
connect the cable properly.
2. If the cable is cut, change the
cable.
Front board Replace front board.
Front board and main board 1. If the cable is disconnected,
connection cable connect the cable properly.
2. If the cable is cut, change the
cable.
CPU board Replace CPU board.
Main board Replace main board.
12. IBP
Symptom 12.1: No display on the numeric area or waveform though sensor is in
normal operation.
Checkpoint Action Remark
IBP connector cable 1. If cable is disconnected,
connect the cable properly.
2. If cable is cut, change the
cable.
IBP module Replace IBP module.
CPU board Replace CPU board.
Main board Replace main board.
93
13. Defibrillation and Pacing
94
14. Printer
95
Checkpoint Action Remark
printing.
2. If dirt on printerhead, clean
the printerhead.
Printer mecha or control board Replace printer mecha or control
board.
Charger board Replace charger board.
If such an error occurs during operation, the defibrillator/monitor will sound a low-
priority alarm. Audible alarm can be terminated by pressing ALARM button, but it
depends on error codes and conditions.
Table 25. Technical error codes provides a complete list of error codes and problem
identification.
96
Note: An asterisk (*) by an error code explanation in the above table indicates that an audible alarm
occurred by the error code cannot be silenced by pressing ALARM button.
98
Disassembly Guide
Performance Verification. Do not place the defibrillator/monitor into
operation after repair or maintenance has been performed, until all
Performance Tests and Safety Tests listed in Performance Verification
section of this service manual have been performed. Failure to perform
all tests could result in erroneous defibrillator/monitor readings.
Before attempting to open or disassemble the Ornage1
defibrillator/monitor, disconnect the power cord from the
defibrillator/monitor.
Observe ESD (electrostatic discharge) precautions when working within
the unit.
General
This section describes disassembly procedures with detailed disassembly instructions
and illustrations. Disassembly Sequence Flow Chart that is used to access
replaceable parts of the defibrillator/monitor is illustrated in Figure 26. The boxes on
the flow chart represent the various components or sub-assemblies. A complete listing
of the available spare parts and part numbers is in Spare Parts section. Follow the
reverse sequence of the disassembly procedures for reassembly.
99
D 500
Prior to Disassembly
1. Turn off the defibrillator/monitor by rotating Mode select knob.
2. Disconnect the defibrillator/monitor from the AC or DC power source.
Battery Replacement
1. Turn off the defibrillator/monitor and remove AC or DC power code/
2. Press the SMPS/battery push button on the rear panel of defibrillator/monitor.
3. Separate battery from defibrillator/monitor.
4. Replace new battery.
100
101
Figure 28. Side option case disassembly (2)
102
103
Front Case Disassembly (B)
This section describes the items that may be removed on the front case disassembly.
104
105
Figure 30. Front case disassembly (2)
106
107
Figure 31. Front case disassembly (3)
108
109
Rear Case Disassembly (C)
This section describes the items that may be removed on the rear case disassembly.
110
111
C2. HV capacitor disassembly
1. Remove 2 machine screws (3 × 6, S8007) fastening capacitor bracket (T4395) and
inner case (T0600).
2. Disconnect HV connector cable (W0375) from HV board (P1236).
3. Separate HV capacitor (M2027) from inner case (T0600).
112
113
Part Codes Descriptions
T4397 Paddle ground
W0514 Paddle mounting cable
C8. Paddle mounting plate disassembly
1. Remove 4 flat head screws (3 × 6, S8075) on the rear case (T0599).
2. Separate paddle mounting plate (T0607) from rear case (T0599).
3. Remove 2 screws (3 × 6, S8007) from paddle mounting holder (T1069).
4. Separate paddle mounting holder (T1069) from paddle mounting plate (T0607).
5. Remove paddle mounting cable (W0514) from paddle mounting holder (T1069) and
paddle mounting resistor (M4078).
6. Remove 1 screw (3 × 6, S8007) from paddle mounting locker (T1068). Repeat 8
times to remove each mounting locker (T1068).
114
115
Figure 37. D500 Exploded view
116
Spare Parts
Follow local government ordinances and recycle instructions regarding
disposal or recycling of device components, including batteries.
General
Spare parts, along with part numbers, are shown in Table 36. Part list.
117
Table 36. Part list
Item Part Code Description
1 T0645 Rubber stopper
NIBP cuff connector (EU/AP/JP) /
2 E4114/E4232
NIBP cuff connector (US)
3 T0618 Side option connector
4 W0153 IBP connector cable 1 (4pin)
5 W0154 IBP connector cable 2 (5pin)
6 W0450 EtCO2 cable (8pin)
7 T0620 Side option cover
8 P1239 Parameter connector PCB
9 P1241 EtCO2 PCB
10 P1072 IBP PCB
11 P1240 Temp PCB
12 P1242 Temp connector PCB
13 T0630 Side option case
14 T4397 Paddle ground
15 T0607 Paddle mounting plate
16 T0599 Rear case
17 E9030 Speaker
18 T0622 Battery lock button L
19 T4410 Spring lock Φ3.8
20 T0617 Battery lock case L
21 T0600 Inner case
22 P1236 HV PCB
23 T4395 Capacitor bracket
24 M2027 HV capacitor
25 T0623 Battery lock hook
26 T4409 Spring lock Φ4.2
27 T0616 Battery lock case R
28 T0621 Battery lock button R
29 P1237 Charger PCB
30 P1244 Battery connector PCB
31 P1248 Wireless connector PCB
32 T0632 Wireless inner case
33 T0644 USB cover
34 T0631 Wireless case
35 M0057/M0039 3G module / Wi-Fi module
36 P1246/P1247 3G interface PCB / Wi-Fi interface PCB
37 M4074/M4075 3G antenna / Wi-Fi antenna
38 T0625 Wireless cover
39 W0447 External DC connector
40 W0398 AC power wire
41 T4126 Ground pin
42 T0605 AC power supply (SMPS) case top
43 M0055 AC power supply (SMPS)
44 T0606 AC power supply (SMPS) case bottom
45 M0058 Li-ion battery
46 M0015 NIBP module
47 T4313 NIBP bracket (CPU cover)
48 T4394 LCD bracket
49 M4076 8.4" LCD
50 T0536 LCD window
51 M4065 Thermal printer module
52 T0624 Print door lock
118
Item Part Code Description
53 T0601 Printer case
54 M4066 Thermal printer mecha
55 T0602 Printer door
56 T0627 Mode knob
57 P1243 Mode knob PCB
58 P1234 Front PCB
59 T0642 Shock button
60 T0634 LED membrane
61 T0597 Front case deco
62 T0638 Front membrane (5key)
63 T0639 Front membrane (7key)
64 T0635 Front membrane (3key)
65 T0633 Multi function knob
66 T0598 Front case
67 W0354 ECG connector cable (13pin)
68 W0375 HV connector cable
69 W0486 SpO2 PI cable
70 P1233 Main PCB
71 M0007 SpO2 module
72 P1176 CPU PCB
73 P1245 ECG PCB
- A0343 Printer paper (80mm)
- A7225 D500 service manual
- A7226 D500 operator’s manual
Note: Please refer to the operator's manual for accessories.
119
This page is intentionally left blank.
120
Packing for Shipment
General Instructions
To ship the defibrillator/monitor for any reason, follow the instructions in this section.
Pack the defibrillator/monitor carefully. Failure to follow the instructions in this section
may result in loss or damage not covered by the Mediana warranty.
Prior to shipping the defibrillator/monitor, contact your supplier or the Mediana office
(Technical Service Department) for a returned goods authorization (RGA) number.
Mark the shipping carton and any shipping documents with the returned goods
authorization number.
Pack to shipping the defibrillator/monitor, contact your supplier or the Mediana office
(Technical Service Department) for a returned goods authorization number. Mark the
shipping carton and any shipping documents with the returned goods authorization
(RGA) number. Return the defibrillator/monitor by any method that provides proof of
delivery.
121
Repacking In a Different Carton
If the original carton is not available, use the following procedure to pack the
defibrillator/monitor:
122
Specification
Display
Screen Size 8.4” measured diagonally across the TFT-LCD screen
Screen Type Liquid Crystal Display (LCD) Color
Size 170 x 128 mm
Number of Traces 4 waveforms
Controls
Multi function knob; Mode select knob (Off, AED, Manual,
Pacing and Monitor); 11 buttons (Shock, Energy Level,
Standard
Charge, Analyze, NIBP, LEAD, Alarm, Size, Print, Rate,
mA); 5 soft key
Alarms
Categories Patient Status and System Status
Priorities Low, Medium and High Priorities
Notification Audible and Visual
Setting Default and Individual
Alarm Volume Level 45 to 85 dB
Distributed Alarm Less than 3 sec.
System Delay
123
Electrical
Instrument
Power Requirements AC Mains
100 to 240 V, 50/60 Hz, 120 - 130VA
DC Mains
18Vdc, 7.0A with DC/DC adapter, Model:MDD150-1218
(MDD150-1218: Input: 12-16Vdc, 160 - 160VA,
Output: 18Vdc, 7.0A)
Note: For 120 Volt applications, use only UL Listed detachable power cord with
NEMA configuration 5-15P type (parallel blades) plug cap. For 240 Volt
applications use only UL Listed Detachable power supply cord with NEMA
configuration 6-15P type (tandem blades) plug cap.
Battery (option)
Type Li-ion battery
Dimension 150 × 90 × 20 (mm) (W×H×D)
Voltage/Capacity 4S2P 14.4V/5200mAh
Discharge A minimum of 200 shocks at 200 Joules (per battery)
Operating time 5 hours (per battery), 10 hours (with two batteries)
At the following condition:
no printing
no external communication
no audible alarm sound
room temperature : 25°C
Recharge Over 8 hours with defibrillator/monitor turned on/off
(It would take about 8 hours to fully charge the battery from
the moment that low battery alarm is activated.)
6 months, new battery fully-charged
(- The battery life cycle may vary with the number of
recharging, operating temperature, and storage condition.
Typically, about 80% capacity of battery may remain after
300 cycles of recharge. If one cycle of recharging would be
Life Cycle about 2 or 3 days, the battery may have 80% capacity after
about 24 months.
- The battery is designed not to be charged for safety
reason when it reaches 40°C.)
After 2 months storage the defibrillator/monitor would run
for 50% of stated battery life.
124
Environmental Conditions
Operation
Temperature 0 to 50°C (32 to 122°F)
Humidity 15 to 95% RH, non-condensing
Altitude -170 to 4,877m (-557 to 16,000 ft)
Note: The system may not meet its performance specifications if stored or used
outside the specified temperature and humidity range.
Note: The battery will not be charged for safety if the operating temperature
exceeds 40°C.
Shipping and Storage (in shipping container)
Temperature −20°C to 70°C (−4°F to 158°F)
Humidity 15 to 95% RH, non-condensing
Altitude -304 to 6,096m (-1,000 to 20,000ft)
Note: The system may not meet its performance specifications if stored or used
outside the specified temperature and humidity range.
125
Tone Definition
High Priority Alarm Tone
Volume level Adjustable (level 1~8)
Pitch (± 48.8Hz) 976 Hz
Pulse width (± 10ms) 210 ms
Number of pulses 10 pulses per 4 sec, 10 sec inter burst
Repetitions Continually
Medium Priority Alarm Tone
Volume level Adjustable (level 1~8)
Pitch (± 34.85Hz) 697 Hz
Pulse width (± 10ms) 210 msec
Number of pulses 3 pulses per 1 sec, 15 sec inter burst
Repetitions Continually
Low Priority Alarm Tone
Volume level Adjustable (level 1~8)
Pitch (± 9.4Hz) 488 Hz
Pulse width (± 10ms) 210 msec
Number of pulses 1 pulse per 0.25 sec, 30 sec inter burst
Alarm Reminder Tone
Volume level Not changeable
Pitch (± 40Hz) 800 Hz
Pulse width (± 10ms) 200 msec
Number of pulses 1 pulse per 1 sec, 3 min, 10 min inter burst
Repetitions Continually
HR/PR Tone
Volume level Not changeable
650 Hz (ECG)
Pitch (± 32.5Hz)
(162 + 5*SpO2) Hz
Pulse width (± 5ms) 100 msec
Number of pulses N/A
Repetitions No repeat
Key Beep
Volume level Adjustable (Off, level 1~7)
Pitch 440 (± 22) Hz (valid)
168 (± 8.4)Hz (invalid)
Pulse width (± 5ms) 100 msec
Number of pulses N/A
Repetitions No repeat
126
Measurement Parameters
Pacing Mode
Pacing Mode
Pacing rate Variable from 30 bpm(ppm) to 180 bpm(ppm) ± 1.5%
(increments or decrements by a value of 2
bpm(ppm))
Resolution 2 bpm(ppm)
Accuracy ± 1.5 %
Output current 0 mA to 140 mA
Resolution 2 mA
Accuracy ± 5% or 5 mA, whichever is greater.
Pulse Type 40 msec constant current pulse
Pulse Amplitude Variable 0 mA to 140 mA ± 5% or 5 mA, whichever is
greater. Digitally displayed on the defibrillator/monitor
(increments or decrements by a value of 2 mA)
Defibrillator
AED mode
Lead Lead ll
Patient Impedance 25 to 175 Ω
Heart Rate 20 to 300 BPM
Accuracy 1 per min
Detection V/F ≤ 200 µV
V/T ≤ 160 per min
Shock Analysis Time < 10 seconds typical
Charging Time to 200J* Within 6 seconds with rated main voltage/DC main
voltage
Within 7 seconds with fully charged battery
*Charging time for other cases, refer to defibrillator (technical specification)
Manual Mode
Shock Energy Level External paddles:
1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 40, 50, 75, 100,
125, 150, 175, 200, 300, 360 J
Automatic Discharge Time 60 seconds
Charging Time to 200J* Within 6 seconds with rated main voltage/DC main
voltage
Within 7 seconds with fully charged battery
Charging time to 360J Within 8 seconds with rated main voltage/DC main
voltage
Within 9 seconds with fully charged battery
*Charging time for other cases, refer to defibrillator (technical specification)
127
ECG
Heart Rate
Measurement Range 0, 20 to 300 BPM
Resolution 1 BPM
Accuracy ± 10% or ±5 bpm, whichever is greater
Average Response Time 5 seconds (from 80 to120 BPM)
9 seconds (from 80 to 40 BPM)
Tall T-wave Rejection maximum T-wave amplitude 1.8 mV
ECG (Electrocardiograph)
Leads 3 / 5 / 12 Lead
Lead I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5,
V6, Paddle, Pads
Lead Off Detection Detected and displayed
Pacer Detection Detect pacer pulses of ±2mV to ±700mV with pulse
widths of 0.1 to 2msec and rise times 10% of width
not to exceed 100msec
Input
Input Impedance 5 Mohm or more
Input Dynamic Range ±5 mV AC, ±300 mV DC
Voltage Range ±0.5 mV ~ ±5 mV
Signal Width 40 to 120 ms (Q to S)
Output
Filter mode 0.5 to 21 Hz
0.05 to 40 Hz
1 to 21 Hz
ECG Size Auto, 5.0, 10.0, 15.0, 20.0, 30.0 mm/mV
Display Sweep Speeds 25.0 mm/sec
Wave Size x0.5, x1, x1.5, x2, x3
Display Sensitivity 10 mm/mV (×1)
Pacing Pulse Detection On, Off
Electrode Disconnect Alarm Display and/or sound
CMRR 90 dB or more
Defibrillator Discharge <5 sec per IEC60601-2-27
Recovery
Defibrillator Protection Protected
ECG (Arrhythmia Supplemental Information as required by AAMI EC13)
Respiration, leads-off amp 0.03 µA
sensing, and active noise common 0.25 µA
suppression
Tall T-wave rejection Maximum T-Wave amplitude 1.8 mV
capability
Accuracy of input signal Displacement: 0.1 mV
reproduction Slope: 0.1 mV/s
Heart rate meter accuracy Provides correct heart rates, as follows:
and response to irregular Ventricular bigeminy: 79 to 83 bpm
rhythm Slow alternating ventricular bigeminy: 51 to 63 bpm
Rapid alternating ventricular bigeminy: 119 to
121bpm
Bidirectional systoles: 82 to 94 bpm
Response time of heart rate HR change from 80 to 120 bpm: 3.16 sec
meter to change in heart HR change from 80 to 40 bpm: 3.05 sec
rate
128
Time to alarm for Vent Tachycardia 1 mVpp, 206 bpm:
tachycardia Amplitude 0.5mV, Alarm activated time 6.79 sec
Amplitude 1mV, Alarm activated time 9.80 sec
Amplitude 2mV, Alarm activated time 8.19
Respiration
IM Respiration
Technique Impedance Pneumography
Range 0, 3 to 120 breaths/min
Resolution ±3 breaths/min
Leads RA to LA
Base impedance 500 to 2000 Ω
Delta impedance 0.5 to 3 Ω
Lead Off Condition Detected and displayed
Defibrillator Protection Protected
Display Sweep Speeds 25 mm/s
Wave Size x0.5, x1, x1.5, x2
AW Respiration
Technique Nondispresive Infrared Spectroscopy
Range 0 to 150 breaths/min
Accuracy ±1 breaths/min
129
NIBP
Pulse Rate
Pulse Rate Range Adult/Pediatric 40 to 200 BPM
Neonatal 40 to 240 BPM
Resolution 5 BPM
Pulse Rate Accuracy ±2 BPM or ±2%, whichever is greater
NIBP (Non-Invasive Blood Pressure)
Technique Oscillometric Measurement
Measurement Modes Off, Cont, 1, 2.5, 3, 5, 10, 15, 30, 60, 90 minutes
Measurement Range Adult/Pediatric
SYS 60 to 250 mmHg
MAP 45 to 235 mmHg
DIA 40 to 200 mmHg
Neonatal
SYS 40 to 120 mmHg
MAP 30 to 100 mmHg
DIA 20 to 90 mmHg
NIBP Accuracy Mean error and standard deviation per ANSI/AAMI
SP10:2002+A1:2003+A2:2006, IEC 60601-2-30:2009,
EN 1060-4:2004
Pressure Display Range Adult/Pediatric 0 to 300 mmHg
Neonatal 0 to 150 mmHg
Pressure Display Within ±3 mmHg or 2%
Accuracy
Initial Cuff Inflate Adult/Pediatric
Pressure 120, 140, 160, 180, 200, 220, 240, 260, 280 mmHg
(16.0, 18.7, 21.3, 24.0, 26.7, 29.3, 32.0, 34.7, 37.3 kPa)
Neonatal
80, 90, 100, 110, 120, 130, 140 mmHg
(10.7, 12.0, 13.3, 14.7, 16.0, 17.3, 18.7 kPa)
Automatic Cuff Deflation Measurement time exceeding 180s in adult/pediatric
(90s in neonatal) or maximum pressure value
exceeding 300 mmHg in adult (150 mmHg in neonatal).
Overpressure Protector Adult/Pediatric: 300 mmHg
Neonatal: 150 mmHg
Defibrillator Protection Protected
Measurement Speed About 20 seconds
At the following condition:
Adult
Cuff size 12 cm
SYS 120 mmHg
MAP 90 mmHg
DIA 80 mmHg/ PR 80 BPM
Manual Measurements (180 mmHg)
130
IBP
Pulse Rate
Pulse Rate Range 20 to 250 BPM
Accuracy ± 1 % or ± 1 bpm
Pulse Rate Resolution 1 BPM
IBP (Invasive Blood Pressure)
Parameter Displayed P1, ABP
P2, CVP, PAP, LAP
Measurement Range -50 mmHg to 300 mmHg
20 bpm to 250 bpm
Resolution 1 mmHg
Accuracy ± 3 mmHg
Input sensitivity 5µV/V/mmHg
Transducer Volume 0.1mm3/100mmHg
Displacement
Zero calibration range ± 100mmHg
Frequency response DC - 10 Hz ± 1 Hz, -3db
Wave Size 0 to 50, 0 to 100, 0 to 200, 0 to 300, Auto Size
Display Sweep Speeds 25.0 mm/s
Defibrillator Protection Protected
SpO2
Measurement Ranges
SpO2 saturation range 1% to 100%
Pulse rate range 20 to 300 beats per minute (bpm)
Perfusion range 0.03% to 20%
Display sweep speed 25.0 mm/s
Measurement Accuracy
Pulse rate accuracy 20 to 250 beats per minute (bpm) ±3 digits
SpO2 saturation accuracy 70% to 100% ±2 digits, neonates: ±3 digits
Note: SpO2 saturation accuracy - Monitoring system measurements are statistically
distributed; about two-thirds of monitoring system measurements can be
expected to fall in this accuracy (ARMS) range. Reference the Clinical Studies
section for test results. For a complete listing of SpO2 accuracy across the full
line of available Nellcor™ sensors, contact Covidien, a local Covidien
representative, or locate it online at www.covidien.com.
Operating Range and Dissipation
Red Light Wavelength Approximately 660 nm
Infrared Light Wavelength Approximately 900 nm
Optical Output Power Less than 15 mW
Power Dissipation 52.5 mW
131
Capnography
Capnography
Parameter Displayed EtCO2, InCO2
Range 0 mmHg ~ 150 mmHg (0kPa ~ 20kPa, 0% ~ 20%)
Accuracy 0-40mmHg ±2mmHg of reading
41-70mmHg ±5% of reading
71-100mmHg ±8% of reading
101-150mmHg ±10% of reading
(No degradation due to respiration rate or
Inspiration/Expiration ratio (I/E ratio))
Display Accuracy ±2mmHg
Response Time Mainstream: Less than 60msec
Sidestream: Less than 3sec
Baromatric Pressure -152.4 to 4572 meters
Correction (-500 to 15,000 feet),
775 to 429 mmHg, Automatic
Gas Compensation User selective at O2 >60% and N2O >50%
Stability and drift Short term drift: Less than 0.8 mmHg over 4 hours.
Long term drift: Accuracy specification will be
maintained over a 120 hour period.
Accuracy Change for 0-40 mmHg ±1 mmHg additional error
interfering Gases and 41-70 mmHg ±2.5% additional error
Vapors Anesthetic and 71-100 mmHg ±4% additional error
interfering agents 101-150 mmHg ±5% additional error
132
Temperature
Thermistor Temp
Probe Type YSI 400 series and YSI 700 series Probes
Measurement Method Thermistor
Operating Mode Direct
Range 0 to 50°C (32.0 to 122°F)
Display Accuracy ±0.1
Probe Accuracy YSI 400 series and YSI 700 series Probes:
±0.1°C from 25 to 45°C,
±0.2°C from 0 to 25°C and from 45 to 50°C
Transient Response Time About 17 sec
Defibrillator Protection Protected
Event
Type 12 lead, Event
Data storage Internal memory, SD card
Memory 12 lead
saves total 100 data
saves ECG waveform
saves ECG analysis result data
saves ECG analysis date and time
saves HR/PR, NIBP, SpO2, Respiration,
Temperature, IBP 1, IBP 2, EtCO2 numeric data
saves alarm condition
Event
saves total 250 data
saves defibrillation shock information (number of
shock, energy level, actual passed energy,
impedance)
saves pacing information (pace rate, pace current,
fixed mode)
saves clinical action list
saves 1 channel ECG waveform
saves Event date and time
saves HR/PR, NIBP, SpO2, Respiration,
Temperature, IBP 1, IBP 2, EtCO2 numeric data
saves alarm condition
133
Defibrillator (Technical Specification)
AED Mode
AED Mode
Charging Time - 200J Charging condition Time (sec)
With Rated Mains Voltage 5.8
With DC Mains Voltage 5.9
With fully charged battery 6.4
Charging Time - 200J Charging condition Time (sec)
(including time from the With Rated Mains Voltage 18.1
initiation of rhythm With DC Mains Voltage 18.4
analysis with a clear ECG With fully charged battery 19.3
signal to readiness for With Mains Voltage of 90% of the Rated
discharge.) 18.1
value
With DC Mains Voltage of 90% of the
18.8
Rated value
After 15 maximum energy discharges
19.6
taken from a new fully charged battery
With Mains Voltage of 90% of the Rated
value, but measured from initially
26.4
switching power on to ready for
discharge at maximum energy
With DC Mains Voltage of 90% of the
Rated value, but measured from initially
26.3
switching power on to ready for
discharge at maximum energy
After 15 maximum energy discharges
taken from a new fully charged battery,
but measured from initially switching 26.7
power on to ready for discharge at
maximum energy
Note: A used battery normally takes more time than the
charging time specified in above.
Waveform parameters The defibrillator/monitor delivers shocks to load
impedances from 25 to 175 ohms. The duration of each
pulse of the waveform is dynamically adjusted based on
delivered charge, in order to compensate for patient
impedance variations, as shown below:
(When 200J is selected)
Load resistance (Ω) Delivered energy(J)
25 195
50 200
75 196
100 199
125 199
150 200
175 200
134
Manual Mode
Manual Mode
Charging Time – 200J Charging condition Time (sec)
With Rated Mains Voltage 5.8
With DC Mains Voltage 5.9
With fully charged battery 6.4
With Mains Voltage of 90% of the
5.8
Rated value
With DC Mains Voltage of 90% of the
6.4
Rated value
After 15 maximum energy discharges
6.6
taken from a new fully charged battery
With Mains Voltage of 90% of the
Rated value, but measured from
11.7
initially switching power on to ready for
discharge at maximum energy
With DC Mains Voltage of 90% of the
Rated value, but measured from
11.9
initially switching power on to ready for
discharge at maximum energy
After 15 maximum energy discharges
taken from a new fully charged battery,
but measured from initially switching 12.5
power on to ready for discharge at
maximum energy
Note: A used battery normally takes more time than the
charging time specified in above.
Charging Time – 360J Charging condition Time (sec)
With Rated Mains Voltage 7.4
With DC Mains Voltage 7.5
With fully charged battery 8.0
With Mains Voltage of 90% of the
7.4
Rated value
With DC Mains Voltage of 90% of the
8.0
Rated value
After 15 maximum energy discharges
8.5
taken from a new fully charged battery
With Mains Voltage of 90% of the
Rated value, but measured from
13.0
initially switching power on to ready for
discharge at maximum energy
With DC Mains Voltage of 90% of the
Rated value, but measured from
14.0
initially switching power on to ready for
discharge at maximum energy
After 15 maximum energy discharges
taken from a new fully charged battery,
but measured from initially switching 14.3
power on to ready for discharge at
maximum energy
Note: A used battery normally takes more time than the
charging time specified in above.
Number of discharges The number of maximum energy discharges which are
available from a new and fully charged battery at 20°C
135
ambient temperature is above 200 times for 360J and
250 times for 200J
136
Biphasic Waveform Characteristics
The efficiency of Mediana’s Biphasic waveform has been clinically verified during a
ventricular fibrillation (VF) and ventricular tachycardia (VT) defibrillation study. This
study (which was conducted using D500 defibrillator/monitors) and the findings are
described below.
Table 37. Delivered Energy at Every Defibrillator Settings into a Range of Loads
Selected Load
Accuracy
Energy 25 50 75 100 125 150 175
1J 1.0 1.0 1.0 1.0 1.0 1.0 1.0 15%
2J 2.0 2.0 2.0 2.0 2.0 2.0 2.0 15%
3J 3.0 3.0 3.0 3.0 3.0 3.0 3.0 15%
4J 4.0 4.0 4.0 4.0 4.0 4.0 4.0 15%
5J 5.0 5.0 5.0 5.0 5.0 5.0 5.0 15%
6J 6.0 6.0 6.0 6.0 6.0 6.0 6.0 15%
7J 7.0 7.0 7.0 7.0 7.0 7.0 7.0 15%
8J 8.0 8.0 8.0 8.0 8.0 8.0 8.0 15%
9J 9.0 9.0 9.0 9.0 9.0 9.0 9.0 15%
10J 11.0 10.0 10.0 10.0 10.0 10.0 10.0 15%
15J 16.0 15.0 15.0 15.0 15.0 15.0 15.0 15%
20J 21.0 20.0 20.0 20.0 20.0 20.0 20.0 15%
30J 31.0 30.0 30.0 30.0 30.0 30.0 30.0 15%
40J 41.0 40.0 40.0 40.0 39.0 39.0 39.0 15%
50J 51.0 50.0 50.0 50.0 49.0 49.0 49.0 15%
75J 77.0 76.0 75.0 75.0 74.0 74.0 74.0 15%
100J 102 101 101 100 99.0 99.0 99.0 15%
125J 127 126 126 124 124 122 123 15%
150J 152 151 150 150 148 147 148 15%
175J 177 174 173 173 173 172 172 15%
200J 203 198 200 199 198 197 197 15%
300J 304 302 300 298 296 297 296 15%
360J 370 371 365 363 363 356 356 15%
137
Figure through show the biphasic waveforms that are produced when the
defibrillator/monitor is discharged into loads of 25, 50, 75, 100, 125, 150 and 175
ohms at each energy setting (360, 300, 200, 175, 150, 125, 100, 75, 50, 40, 30, 20,
15, 10, 9, 8, 7, 6, 5, 4, 3,2 and 1 joule[s]).
The vertical axis shows the voltage in volts (V); the horizontal axis shows the duration
in milliseconds (ms).
138
Figure 41. Biphasic Waveforms at 175 Joules
139
Figure 43. Biphasic Waveforms at 125 Joules
140
Figure 45. Biphasic Waveforms at 75 Joules
141
Figure 47. Biphasic Waveforms at 40 Joules
142
Figure 49. Biphasic Waveforms at 20 Joules
143
Figure 51. Biphasic Waveforms at 9 Joules
144
Figure 53. Biphasic Waveforms at 7 Joules
145
Figure 55. Biphasic Waveforms at 5 Joules
146
Figure 57. Biphasic Waveforms at 3 Joules
147
Figure 59. Biphasic Waveforms at 1 Joules
148
Compliance
Item Standard Description
Classification IEC60601-1:1988 Class I (on AC power)
+A1:1991+A2:1995, Internally powered (on battery power)
EN60601-1:1990+A1:1993
/AC:1994
Type of protection IEC60601-1:1988 Type BF and Type CF – Applied part
+A1:1991+A2:1995,
EN60601-1:1990+A1:1993
/AC:1994
Mode of operation IEC60601-1:1988 Continuous
+A1:1991+A2:1995,
EN60601-1:1990+A1:1993
/AC:1994
Degree of IEC60529:1989+A1:1999+A2 IP34 (provided by enclosures)
protection :2013,
EN60529:1991+A1:2000
+A2:2013
General 93/42/EEC as amended by Directives for medical devices
2007/47/EC
21CFR820 Code of federal regulations
2012/19/EU Waste electrical and electronic equipment
directive (WEEE)
2011/65/EU Restriction of the use of Hazardous
Substances in electrical and electronic
equipment II
93/86/EEC Battery disposal directive
2006/66/EC as amended by Battery directive
2008/103/EC
ISO13485:2003/Cor1:2009 Quality systems - Medical Devices -
EN ISO13485:2012 Requirements for regulating purposes
/AC:2012
ISO14971:2007, Risk analysis managements – medical
EN ISO14971:2012 devices
IEC60601-1:1988 General requirements for safety of medical
+A1:1991+A2:1995, electrical equipment
EN60601-1:1990+A1:1993
/AC:1994
IEC60529:1989+A1:1999+A2 Degree of protection provided by
:2013, enclosures (IP34)
EN60529:1991+A1:2000
+A2:2013
ISO14155:2011/Cor1:2011E Clinical investigation of medical devices for
N ISO14155:2011 human subjects – part 1: General
requirements
IEC60601-1-1:2000, Collateral standard for medical electrical
EN60601-1-1:2001 systems
IEC60601-1-6:2010, Collateral standard for usability
EN60601-1-6:2010
149
Item Standard Description
ISO 10993-1:2009/ Biological evaluation of medical devices –
Cor1:2010, Part 1: Evaluation and testing
EN ISO109931:2009
/AC:2010
ISO10993-5:2009, Biological evaluation of medical devices –
EN ISO10993-5:2009 Part 5: Tests for in vitro cytotoxicity
ISO10993-10:2010, Biological evaluation of medical devices –
EN ISO10993-10:2010 Part 10: Tests for irritation and delayed-
type hypersensitivity
IEC60601-2-49:2001, Particular requirements for multifunction
EN60601-2-49:2001 patient monitoring equipment
IEC62366:2007 Medical devices – Application of usability
EN62366:2008 engineering to medical devices
IEC62304:2006 Medical device software – Software life-
EN62304:2006/AC:2008 cycle processes
Ambulance EN1789:2007+A1:2010 Medical vehicles and their equipment –
road ambulance
EN13718-1:2008 Medical vehicles and their equipment – Air
ambulances – Part1:Requirements for
medical devices used in air ambulances.
RTCA/ DO-160 section 20, Environmental Conditions and Test
21 Procedures for Airborne Equipment
Alarms IEC60601-1-8:2006, Alarm systems requirements, tests and
EN60601-1-8:2007/AC:2010 guidance in medical electrical equipments
systems
Electrocardiograph IEC60601-2-27:2005, Particular requirements for the safety of
EN60601-2-27:2006 Electrocardiographic monitoring equipment
/AC:2006
IEC60601-2-25:1993 Particular requirements for the basic safety
+A1:1999 and essential performance of
EN60601-2-25:1995 electrocardiographs
+A1:1999
AAMI EC13:2002/(R)2007 Cardiac monitors, heart rate meters and
alarms
AAMI EC53:1995/®2008 ECG cable and leads
AAMI EC57:2012 Testing and reporting performance results
of cardiac rhythm and ST-segment
measurement algorithms
Non-invasive blood AAMI SP10:2002 Electronic or automated
pressure +A1:2003+A2:2006 sphygmomanometers
EN1060-3:1997 Supplementary requirements for electrical-
+A2:2009 mechanical blood pressure measuring
systems
EN1060-4:2004 Non-invasive sphygmomanometers - Test
procedures to determine the overall system
accuracy of automated non-invasive
150
Item Standard Description
sphygmomanometers
IEC60601-2-30:1999, Particular requirements for the safety,
EN60601-2-30:2000 including essential performance, of
automatic cycling indirect blood pressure
monitoring equipment
Oxygen saturation ISO9919:2005, Basic safety & essential performance of
EN ISO 9919:2009 pulse oximeter for medical use
Temperature EN12470-4:2000 +A1:2009 Performance of electrical thermometers for
monitoring continuous Measurement
Invasive blood IEC60601-2-34:2000 Particular requirements for the safety,
pressure EN 60601-2-34:2000 including essential performance, of invasive
blood pressure monitoring equipment
Capnography ISO21647:2004, Particular requirements for the basic safety
EN ISO 21647:2009 and essential performance of respiratory
gas monitors
Defibrillator IEC 60601-2-4:2002 Safety of cardiac defibrillators
EN 60601-2-4:2003
AAMI DF80:2003 Safety of cardiac defibrillators (including
automated external defibrillators)
Electromagnetic IEC60601-1, sub clause 36, Electromagnetic compatibility-requirements
compatibility and & test
IEC60601-1-2:2001
EN60601-1-2:2007
/AC:2010
IEC61000-3-2:2014 Harmonic emission Ed 3.2
EN61000-3-2:2014
IEC61000-3-3:2013, Voltage fluctuations/Flicker emission Ed 2.0
EN61000-3-3:2013
IEC61000-4-2:2008, Electrostatic discharge Ed 2.0
EN61000-4-2:2009
IEC61000-4- Radiated RF electromagnetic field Ed 3.1
3:2006+A1:2007+A2:2010
EN61000-4-3:2006
+A1:2008+A2:2010
IEC61000-4-4:2012 Electrical fast transient/burst Ed 2.1
EN61000-4-4:2012
IEC61000-4-5:2014, Surge current Ed 2.0
EN61000-4-5:2014
IEC61000-4-6:2013, Conducted disturbances, induced by RF
EN61000-4-6:2014 field Ed 3.0
IEC61000-4-8:2009, Power frequency (50/60Hz) magnetic field
EN61000-4-8:2010 Ed 2.0
IEC61000-4-11:2004, Voltage dips, short interruption and voltage
EN61000-4-11:2004 variation on power supply input lines Ed 2.0
CISPR11:2009 +A1:2010 Limits and methods of measurement of
EN55011: 2009 +A1:2010 radio disturbance characteristics of
industrial scientific and medical (ISM) radio-
151
Item Standard Description
frequency
equipment RF Emissions Group 1, Class B
CISPR22:2008 Information technology equipment - Radio
EN55022:2010/AC:2011 disturbance characteristics - Limits and
methods of measurement
Package ISTA (Procedure 1A, 2001) Pre-Shipment test procedures (Package)
IEC60068-1:2013 Environmental testing, Part1: General
EN 60068-1:2014 guidelines
IEC60068-2-1:2007 Environmental testing - Part 2-1: Tests -
EN60068-2-1:2007 Test A: Cold
IEC60068-2-2:2007 Environmental testing - Part 2-2: Tests -
EN60068-2-2:2007 Test B: Dry heat
IEC60068-2-30:2005 Environmental testing - Part 2-30: Tests -
EN60068-2-30:2005 Test Db: Damp heat, cyclic (12 h + 12 h
cycle)
Reliability IEC60068-2-27:2008, Environmental testing – Shock
EN60068-2-27:2009
IEC60068-2-6:2007, Environmental testing – Vibration
EN60068-2-6:2008
IEC60068-2-31:2008 Environmental testing: Rough handling
EN60068-2-31:2008 shocks, primarily for equipment-type
specimens
IEC60068-2-64:2008, Environmental testing: vibration, broad-
EN60068-2-64:2008 band random (digital control) and guidance
Labeling EN1041:2008 Information supplied by the manufacturer
with medical devices
Marking IEC /TR60878:2003 Graphical symbols for electrical equipment
in medical practice
ISO15223-1:2012 Symbols to be used with medical device
ENISO 15223-1:2012 labels, labelling and information to be
supplied -- Part 1: General requirements
ISO15223-2:2010 Symbols to be used with medical device
labels, labelling, and information to be
supplied -- Part 2: Symbol development,
selection and validation
ISO7000:2014 Graphical symbols for use on equipment-
index and synopsis
EN980:2008 Graphical symbols for use in the labeling of
medical devices
EN50419:2006 Marking of electrical and electronic
equipment in accordance with article II (2)
of directive 2002/96/EC (WEEE)
152
System Processing Description
System Overview
The D500 is a defibrillator and multi-function monitor. Manual and Automated external
defibrillation, External pacing and the other monitoring functions (Electrocardiography
(ECG), functional arterial oxygen saturation(SpO2), respiration(RESP),
temperature(TEMP), invasive blood pressure(IBP) and end tidal CO2(EtCO2)) are
intended for use on adult and pediatric patients. Non-invasive blood pressure is
intended for use on adult, pediatric and neonatal patients.
In addition to monitoring and displaying the status of these physiological parameters,
the instrument performs various microprocessor-programmed analytical functions;
Creating both visual and audible alarm signals when settable limits are violated;
Creating and displaying warning/error messages when conditions are detected that
would degrade or prevent valid measurements;
Creating and displaying graphical or tabular trend data;
Providing input to an optional recorder for printout of current or tabular trend data.
153
Unit Description
Power unit: consists of power entry module, power supply, battery charger, battery,
external DC input and DC/DC unit.
Main process unit: consists of Samsung S3C2440A ARM9 CPU, SDRAM and
Flash.
154
Display unit: consists of Samsung S3C2440 CPU, SDRAM and Flash.
155
Communication unit : 8-channel UART
GUI (graphic user interface) unit: consists of TFT LCD, inverter for backlit and
internal video controller.
156
NIBP unit: measures non-invasive blood pressure data.
157
SpO2 unit: measures oxygen saturation data
External Interface
SD card : SD type for firmware and voice prompt update and real time data
download
Wireless Communication : Wi-Fi, 3G Module
158
ECG Processing
The measurement of the skin surfaces electrocardiogram is based on the electrical
signals on the skin surface, produced as the heart muscle contracts and relaxes. The
signals are detected by electrodes placed on the patient body. The information on
heart activity carried by these signals varies with the placing of the electrodes.
The technique used in ECG senses the varying potential difference between two
points at the skin surface which respond to the chemical actions of the muscular
activity of the heart.
Three electrodes are attached to the patient’s right arm (RA), left arm (LA) and left leg
(LL). The varying potentials at these locations are cable-connected to the ECG circuit
inputs where they are conditioned, and the difference of potential between two
selected leads is digitized before transmitting through opto-isolators to the processor.
The processor-installed algorithms operate on the signals to develop drivers for the
graphic display and to compute the heart rate in beats per minute (bpm).
In addition to the acquisition of the QRS waveform complex, the ECG input and
subsequent signal processing computing circuitry perform a number of other
functions:
NIBP Processing
Overview
The oscillometric technique does not use Korotkoff sounds to determine blood
pressure. The oscillometric technique monitors the changes in cuff pressure caused
by the flow of blood through the artery. The defibrillator/monitor inflates the cuff to a
pressure that occludes the artery. Even when the artery is occluded, the pumping of
the heart against the artery can cause small pressure pulses in the cuff baseline
pressure. The defibrillator/monitor lowers cuff pressure at a controlled rate. As the cuff
pressure goes down, blood starts to flow through the artery. The increasing blood flow
causes the amplitude of the pressure pulses in the cuff to increase. These pressure
pulses continue to increase in amplitude with decreasing cuff pressure until they reach
a maximum amplitude at which point they begin to decrease with decreasing cuff
pressure. The cuff pressure at which the pulse amplitude is the greatest is known as
Mean Arterial Pressure (MAP). The manner in which the pulse amplitudes vary is
often referred to as the pulse envelope. The envelope is an imaginary line that
connects the peak of each pressure pulse and forms an outline. The shape of the
envelope is observed by the defibrillator/monitor using a variety of techniques to
determine the diastolic and systolic blood pressure.
159
MAP
● Oscillometric Response
Distole (Pressure Pulses)
Systole
TIME
Systole
Distole
Systole MAP
MAP ● Cuff Pressure
Distole
TIME
The popular standard for the overall system accuracy is AAMI, SP-10 2002+A1;2003
(Electronic or automated sphygmomanometers).
The main test conditions are as follow:
A. Data comparing the Intra-arterial or the auscultatory by the clinical experts with
the automated sphygmomanometer.
B. For data collection and the data analysis, Bland-Altman Plot is used.
C. On the systolic, diastolic, and MAP, the Deltas of all measurements shall be met
under +/- 5mmHg of mean difference (MD), and +/- 8mmHg of standard deviation
(SD).
(Delta = Intra-arterial or Auscultatory – Automated sphygmomanometer)
160
SpO2 Processing
Pulse oximetry works by applying a sensor to a pulsating arteriolar vascular bed. The
sensor contains a dual light source and photodetector. Bone, tissue, pigmentation,
and venous vessels normally absorb a constant amount of light over time. The
arteriolar bed normally pulsates and absorbs variable amounts of light during systole
and diastole, as blood volume increases and decreases. The ratio of light absorbed at
systole and diastole is translated into an oxygen saturation measurement (SpO2).
Because a measurement of SpO2 is dependent upon light from the sensor, excessive
ambient light can interfere with this measurement.
161
Automatic Calibration
Because light absorption by hemoglobin is wavelength dependent and because the
mean wavelength of LEDs varies, an oximeter must know the mean wavelength of the
SpO2 sensor’s red LED to accurately measure SpO2. During monitoring, the
instrument’s software selects coefficients that are appropriate for the wavelength of
that individual sensor’s red LED; these coefficients are then used to determine SpO2.
Additionally, to compensate for differences in tissue thickness, the light intensity of the
sensor’s LEDs is adjusted automatically.
pH
100
Temperature
PCO2
2,3-DPG
Fetal Hb
Saturation (%)
pH
50 Temperature
PCO2
2,3-DPG
0
50 100
PO2 (mmHg)
SpO2 Accuracy
The saturation (SpO2) accuracy specification and/or pulse rate (PR) performance
were analyzed by comparative oximetry performance (COPS) tests between
defibrillator/monitor SpO2 module and MP100 module with the same version of the
oximetry algorithm from Nellcor. This was to demonstrate that the performance of
defibrillator/monitor SpO2 module was equivalent to that of MP100, which had been
validated during both standard motion, combined motion and cold-induced peripheral
vasoconstriction (low perfusion) conditions by direct comparison to measurements of
arterial oxygen saturation (SpO2) obtained from arterial blood samples analyzed with
Instrumentation Laboratory (IL) CO-Oximeters under Nellcor’s clinical protocol,
Invasive Controlled Hypoxia Studies.
162
Respiration Processing
The respiration monitoring is designed to use the variation of this thoracic impedance.
The chest contains various materials, ranging from bone to air. Each of these
materials has different electrical properties and is located in a different portion of the
chest. The materials of the chest vary in electrical resistivity (the amount of electrical
resistance between opposite faces of a cube of that material), which is an important
determinant of electrical impedance in the body.
Two of the major components of the chest, blood and air, are at opposite ends of the
scale. Furthermore, the volume of each of these materials varies with time over the
cardiac and breathing cycles. The variation of the thoracic impedance is caused by
the difference between air and blood in the thoracic impedance. Blood has relatively
low resistivity, which varies over the cardiac cycle owing to changing blood volumes in
the heart and in the vascular compartment. Air, on the other hand, has high electrical
resistivity and hence impedance, and it undergoes wide volume changes in the lungs
during normal breathing. i.e. the impedance of blood is 150 Ω/cm and the one of air is
5000 Ω/cm.
The patient’s respiration is detected by using two of the three leads of the ECG
electrodes (RA and LA, or RA and LL) and cable. The electrical impedance between a
pair of electrodes is determined by dividing the voltage difference between the two
electrodes by the current that passes between them. When the electrodes are placed
on the actual structure, respective structures change.
A low-level excitation signal is applied to these leads, and the variation of the thoracic
impedance caused by the breathing is sensed and processed for display and
measurement. This variation is processed to the voltage value for the measurement.
In order to transfer the thoracic impedance by a transformer, it is used a minimum
constant current of the sine wave carrier signal. The transferred thoracic impedance is
changed to the voltage signal by using bridge circuit and differential amplifier. Then,
ECG signal is removed by filter, and carrier frequency is removed by full wave rectifier
and filter in order to extract only thoracic impedance in amplifying at the definite level
of signal. This extracted thoracic impedance signal is used to measure the respiration
by digital signal processing.
Temperature Processing
Measurement of patient temperature is accomplished by processing the signal from a
probe containing a resistor whose resistance is temperature dependent. The class of
such components is called thermistor.
163
Invasive Blood Pressure Processing
The pressure transducer is connected to a pressure line which, by means of a
catheter is invasively connected to the patient blood stream. The force of movement of
the blood in the patient vessels in transported by the fluid column in the pressure line
to the transducer. This movement cause an electrical signal to be generated which is
then amplified to display the pressure wave and the numeric for the systolic, diastolic
and mean pressure values.
Capnography Processing
The defibrillator/monitor CO2 sensor measures CO2 by using the infrared technique.
The principle is based on the fact that CO2 molecules absorb infrared (IR) light energy
of specific wavelengths, with the amount of energy absorbed being directly related to
the CO2 concentration. When an IR beam is passed through a gas sample containing
CO2, the electronic signal from the photodetector (which measures the remaining light
energy) is measured. This signal is then compared to the energy of the IR source and
adjusted to accurately reflect CO2 concentration in the sample. The
defibrillator/monitor CO2 sensor’s response to a known concentration of CO2 is stored
at the factory in the sensor’s memory. A reference channel accounts for optical
changes in the sensor, allowing the system to remain in calibration without user
intervention.
Defibrillation Processing
The defibrillator is energy-based, meaning that the defibrillator charges a capacitor to
an appropriate level of voltage (either by an operator or automatically) and upon
initiation of the shock (automatically or upon the press of a button) current is delivered
directly to the heart to interrupt the arrhythmia and restore normal conduction. The
defibrillator delivers energy in a biphasic waveform by paddle or pads. With a biphasic
waveform, current is delivered to the heart in two vectors. Because of the two-vector
approach, the peak current required to convert the arrhythmia is reduced, and the
efficacy of the shock is greatly enhanced. It is generally accepted that biphasic
defibrillation results in less myocardial damage from the shock itself.
164