Professional Documents
Culture Documents
CALIBRATION
zero oxygen and the electrical zero should be identified by passing nitrogen over the
and, since the reaction goes virtually to completion, this procedure is equivalent to
poisoning of, the membrane. For these reasons it is best avoided. If there is no
In this way, the true zero is established and, if preferred, this can be “backed-off” to
coincide with electrical zero. Once the dithionite solution is removed (or N2 flushing
is stopped), and the cathode is exposed to air, the signal generated by 21% oxygen
(the oxygen content of air) can be determined. In a closed system the amount of O2 is
governed by the partial pressure of O2 and the volume of the chamber. Thus 1 ml
(1000 μl) of air will contain 210 μl of oxygen. Similarly, if 1 ml of the space, in a 5 ml
greater. In fact, the difference between the two signals can be used to determine the
volume of a leaf-disc with some accuracy (4b and Fig. 4.2). In general, however,
although allowances can be made for differences in atmospheric pressure etc., the
instrument was designed for simplicity rather than absolute measurement of oxygen.
Although it has proved to be much more versatile than was originally imagined it was
originally intended for the measurement of the oxygen evolved when a relatively
large piece of leaf was illuminated in a relatively small chamber. For most purposes,
removal) of small volumes of air into the chamber (using a gas-tight syringe) can be
taken as the basis of calibration and all that remains to be done is some simple
arithmetic (see Fig. 4b). Experiment 1 incorporates several aspects of the calibration
procedure but this can be shortened (4e) or extended according to need and the degree
of accuracy sought.
Using a 1 ml gas-tight syringe, introduce successive 200 μl aliquots of air into the
leaf-disc chamber through one of the two gas-vents. (The other should be kept closed
during this procedure). As each volume of air is pushed into the chamber the reading
on the pen-recorder should rise quickly to a new level and stay there (Fig. 4.1). You
will see that the initial rise is fast but that it decreases as it approaches its final level.
This is because the rate of diffusion of oxygen to the detector will become limiting as
the difference in concentration between the atmosphere in the chamber and that at the
about one minute (i.e. if the response is sluggish) the electrode may require cleaning.
If the signal rises rapidly but then falls equally quickly, the chamber is leaking and
you should check to see that you have remembered to close the other gas-tap, that the
electrode disc is pressed securely against its O-ring, that the O-ring which seals the
top of the chamber to the bottom is in place and that the clips which hold the two parts
of the chamber together are properly closed. If the response seems too slow for the
experiments that you have in mind, it can be speeded by using a thinner membrane, by
cutting a hole in the spacer immediately above the cathode or by dispensing with the
spacer altogether. If no spacer is used, departure from linearity may sometimes result
(see below).
A 1 ml, gas-tight, syringe, with its plunger fully withdrawn (to the 1 ml position) was
attached to an open tap on the leaf chamber and the other tap closed. The figure
illustrates the responses which were observed as the plunger was then pressed
inwards in 200 μl stages and finally withdrawn again in a similar fashion. Note that
each successive addition produced a larger signal (for explanation see text