C Fat t
Thickness = = 0.725t [mm]
2
Error in C Fat = ± 3.5%
t (Example thickness error 10.0 ± 0.35 [mm])
Sources of Thickness Error
Bicep 3.8 mm
Fat
Layer
49 mm
Muscle
Layer
BodyMetrix
Fat 3.6 mm,
Muscle 48.0 mm
Comparison to X-ray CT
7.8 mm
Thigh Fat Layer
58.0 mm
Muscle Layer
BodyMetrix
Fat 7.6 mm,
Muscle 53.0 mm
Comparison of X-ray CT
23.4 mm Waist
Fat Layer
40 35
35 30
30 25
25 y = 1.9306x
20 y = 2.3038x
R2 = 0.4326
20 R2 = 0.9027
15
15
10
10
5
5
0 0
0 5 10 15 20 25 0 2 4 6 8 10 12 14 16
Thigh BX2000 (mm)
Chest BX2000 (mm)
60
50 Summary
Abdomen Caliper (mm)
40
%BF < 10% Difference ± 0.5%
30 y = 1.4522x
R2 = 0.4269
20
13% < %BF < 20% Difference ± 0.8%
10 25%< %BF < 30% Difference < 3 %
0
0 5 10 15 20 25 30 35 40
Abdomen BX2000 (mm)
Measurements with BodyMetrix
(Normal 1-dimensional mode)
• When used in it’s
normal mode the
BodyMetrix Device
is similar to
traditional A-mode
ultrasound. This
simply means that
the device gets
tissue structure
along one line.
FAT
MUSCLE
BONE
Muscle -
Bone
BodyView 2D Image
8 mm
6.5 mm
Fat
~ 4 cm
Beyond %BF, Health Risk
• Waist circumference,
and Waist to Hip Ratio
show good correlation
• Measuring abdominal
subcutaneous to detect
Superficial AT and
Deep AT could provide
better diagnostic.
Validation Studies