Professional Documents
Culture Documents
Physica Medica
j o u r n a l h o m e p a g e : h t t p : / / w w w. p h y s i c a m e d i c a . c o m
Review Paper
A R T I C L E I N F O A B S T R A C T
Article history: For the maxillofacial region, there are various indications that cannot be interpreted from 2D images and
Received 3 July 2015 will benefit from multiplanar viewing. Dental cone beam CT (CBCT) utilises a cone- or pyramid-shaped
Received in revised form 15 September X-ray beam using mostly flat-panel detectors for 3D image reconstruction with high spatial resolution.
2015
The vast increase in availability and amount of these CBCT devices offers many clinical benefits, and their
Accepted 19 September 2015
Available online 23 October 2015
ongoing development has potential to bring various new clinical applications for medical imaging. Ad-
ditionally, there is also a need for high quality research and education. European guidelines promote the
use of a medical physics expert for advice on radiation protection, patient dose optimisation, and equip-
Keywords:
Dental ment testing. In this review article, we perform a comparison of technical equipment based on manufacturer
CBCT data, including scanner specific X-ray spectra, and describe issues concerning CBCT image reconstruc-
Characteristics tion and image quality, and also address radiation dose issues, dosimetry, and optimisation. We also discuss
Dosimetry clinical needs and what type of education users should have in order to operate CBCT systems safely.
Image quality We will also take a look into the future and discuss the issues that still need to be solved.
Guidelines © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
Development
http://dx.doi.org/10.1016/j.ejmp.2015.09.004
1120-1797/© 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
T. Kiljunen et al./Physica Medica 31 (2015) 844–860 845
Figure 1. (a) D. 16 is endodontically treated. In the intraoral radiograph the apical regions of the buccal roots are suspicious – distobuccal root (arrow), mesiobuccal root
(broken arrow). (b) A CBCT examination carried out for treatment planning shows a large lesion in the distobuccal root, indicating apical periodontitis (arrow) with perfo-
ration of the buccal cortex (broken arrow). Sagittal view (upper left), coronal view (upper right), 3D view (lower left) and axial view (lower right). There is mucosal swelling
in the floor of the right maxillary sinus in region 16 referring reactive changes. In the mesiobuccal root there was apical periodontitis palatoapically (not shown). Note also
the beam hardening artefacts caused by the metal fillings (Courtesy: Tapio Tammisalo).
adjustments related to detector dark-current, gain and pixel defects scanners. One of the low cost factors of CBCT technology is the x-ray
by applying offset and gain corrections. A possible latent image signal tube which can be implemented with similar technical specifica-
from the previous projection read-out also has to be erased by af- tions as in panoramic scanners. The basic structure of a dental CBCT
terglow correction, especially if higher frame-rates are applied. Other gantry can be designed similar to panoramic system, but in the case
processing methods can be utilised based on the physical proper- of CBCT, combined with more advanced digital detectors, soft-
ties of the acquisition system, such as X-ray beam spectral factors, ware and a reconstruction computer [4].
estimated scatter distribution, focus-detector distance, detector re- One of the main characteristics of CBCT is its ability to produce
sponse, among others [11]. scans with varying FOVs according to clinical indication. With user
From an economical perspective, typical dental CBCTs can be ob- adjustable FOV sizes and image quality settings, different image data
tained with significantly lower costs compared to general MSCT voxel sizes can be acquired. As the voxel size is connected to spatial
846 T. Kiljunen et al./Physica Medica 31 (2015) 844–860
Figure 2. Implant treatment planning with CBCT: Linear measurements in region 46. The patient wore a scanning appliance in the CBCT examination where the planned
location of the implant is marked. Axial image (upper left), cross sectional view (upper right), panoramic view (lower left) and 3D image (lower right).
resolution by sampling principles, it is clearly anticipated that smaller maxillofacial region where high spatial resolution of hard tissue is
voxel sizes can provide images with more accurate representa- typically the primary interest.
tions of fine structures than what could be achieved with larger voxel The 3D image reconstruction methods used in the currently avail-
sizes [4,12]. able CBCT scanners mainly still apply an approximate Radon
Furthermore, CBCT image data are typically isotropic which means inversion algorithm that was introduced originally by Feldkamp et al.
that the voxels have the same dimensional length in all three di- in 1984 [15]. However, due to the need of improving image quality
rections of the volumetric image data. As such, the isotropic 3D image further, iterative methods are being developed in many fields of CT
data can be reviewed from any projected direction using reformat- imaging [16].
ted slice orientation with similar image quality – a clear asset with Spatial resolution is essential for dental CBCT imaging where de-
regard to complex 3D anatomy (Figs. 1 and 2). Additionally, the 3D piction of fine details is crucial for reliable diagnosis, e.g. for
image data set can be post-processed into visualisations similar to periodontal applications where the periodontal ligament gap is pre-
traditional dental radiology representations, such as panoramic, sented in the range of 0.1 mm [17,18]. Current CBCT models achieve
cross-sectional, cephalometric, or bilateral multiplanar projec- a spatial resolution of 0.1 mm or even less [9,17]. Spatial resolu-
tions of the temporomandibular joint [4]. The volumetric CBCT image tion is closely related to the detector pixel size due to sampling
data do not have diverging geometry as in projection images, and principles, but there are also other important factors which affect
therefore the linear measurements can be done with high reliabil- the image quality. Focus size, contrast range, detector motion blur
ity and precision (Fig. 2) [13]. and fill factor, the number of raw-data projections (basis images),
CBCT has limited low contrast resolution due to various physi- noise level, and reconstruction technique significantly influence the
cal and technical factors which bring limitations for soft tissue observed resolution [6]. The main exposure parameters such as tube
evaluation [4]. Highly scattered radiation during image acquisi- voltage (effect on contrast, dose and image noise), tube current and
tion will adversely affect the contrast of the projection data and the exposure time (effect on dose, image noise and movement arte-
final reconstructed images. Also, the flat panel technology can cause fact probability through scan time) also have a fundamental effect
artefacts. There are various solutions to the beam hardening effects on the general perceived image quality. Therefore, the isotropic res-
which adversely affect the clinically observed image quality, espe- olution of CBCT which is achieved with a stationary phantom is
cially when imaging of patients with extensive restorations, multiple probably not achieved with living patients where e.g. hemody-
prostheses or previous implant treatments [14]. However, the prop- namic pulsation introduces movements also in the maxillofacial
erties of CBCT are well suited for dental examinations in the region. The resulting spatial resolution of CBCT is likely to be limited
T. Kiljunen et al./Physica Medica 31 (2015) 844–860 847
to around 500 μm in clinical practice [17]. This should be kept in standing position is the most commonly used for scanning, but it
mind when comparing CBCT image quality with MSCT [17,19,20]. is vulnerable to patient movements, especially when the head fix-
In the field of dental 3D imaging, CBCT acts not as a substitute but ation tools are insufficient. To minimise movement artefacts, standing
rather as a complement to MSCT in maxillofacial imaging [9]. devices can also be used with the patient in a sitting position. The
According to a recent review by Nemtoi et al. [9], a total of 47 scanners utilising a standing position typically also have a motorised
CBCT devices by 20 companies were available for dentomaxillofacial column that allows vertical translation of the x-ray tube and de-
imaging when the focus was particularly on the European market. tector system, thus providing access also to wheelchair patients. The
The vast increase in availability and amount of these CBCT devices seated position allows patient comfort and natural orientation of
and their increasing use in DMFR offer many benefits, but there is anatomy, and thereby reduces artefacts induced by patient motion.
still an obvious need for high quality research and education. Even Similar to some of the standing systems, some of the seated systems
though there is an abundance of research connected to CBCT in also have a column that serves access to wheelchair patients. There
DMFR, there is an obvious lack of evidence regarding the impact are also a few CBCT systems that utilise supine position in scan-
of CBCT examinations upon treatment decisions. In particular, more ning (MyRay SkyView, Cefla Dental Group, Imola, Italy; NewTom 3G
research is needed to show the possible benefits of the use of CBCT and NewTom 5G, both from Quantitative Radiology Systems, Verona,
to treatment outcomes in comparison to conventional dental imaging Italy). Although the supine position minimises the patient motion
methods. For a CBCT examination to be justified it needs to be po- and maximises the stability of the scanning process, and there-
tentially beneficial to the patient – it must have the ability to change fore may result in excellent image quality and patient satisfaction,
the treatment decision and patient outcome. For example, Larheim the needs for room dimensions are more demanding compared to
et al. [21] reported that current data concerning the use of CBCT both standing and seated systems. Device footprint is an impor-
in temporomandibular joint diagnostics are mainly limited to tech- tant practical aspect in dental practices where space is often limited
nical efficacy and diagnostics accuracy, i.e. the two first levels in the with office-like installations [9].
six-stage framework to assess the efficacy of imaging methods –
technical, diagnostic accuracy, diagnostic thinking, therapeutic,
Field of view (FOV)
patient outcome and societal efficacy – by Fryback and Thornbury
[22]. However, little attention has been paid to diagnostic think-
The FOVs available in dental CBCT systems vary from FOVs suit-
ing efficacy and therapeutic efficacy [21].
able for a single jaw (D × H usually 4 × 4 cm2 or 5 × 5 cm2) to full
European guidelines promote the use of a medical physics expert
craniofacial imaging (from approximately 15 × 15 cm 2 up to
for advice on radiation protection, patient dose optimisation, and
23 × 26 cm2). The dimensions of the FOV depend on the size and
equipment testing [23]. In this review article, we are performing a
shape of the detector, the beam projection geometry and the ability
comparison of technical equipment based on manufacturer data,
to collimate the beam [6]. The reduction of the FOV in most units
and describe issues concerning CBCT image reconstruction and image
is performed by using adjustable lead shields as a primary colli-
quality, and also radiation dose issues and optimisation. We also
mation at the radiation source. Moreover, most of the systems utilise
discuss clinical needs and what type of education users should have
either one or a few prefixed FOVs planned for different indica-
in order to operate CBCT systems safely. We will also take a look
tions, whereas in some devices one can freely adjust the FOV within
into the future and discuss the issues that still need to be solved.
certain limits in both the vertical and the horizontal cross-sectional
volumes. Additionally, in some CBCT devices it is possible to stitch
Technical characteristics of current dental CBCT scanners
together adjacent 3D volumes, resulting in larger FOVs. Impor-
tantly, users should remember, while scanning patients, that the FOV
General features
is the most important scanning parameter affecting the patient ra-
diation dose and image quality, and therefore it should be selected
Depending on the practical necessities, it is possible to choose
according to clinical questions (these are discussed more thor-
from CBCT systems with different imaging features and capabili-
oughly in later sections).
ties. The technical properties of the reviewed dental CBCT scanners
are summarised in Table 1 and discussed in the text. Data were col-
lected from manufacturers’ brochures available in Internet, and Data collection during the scan
furthermore, checked and completed by the representatives of each
vendor with a questionnaire (a few manufacturers did not answer CBCT data reconstruction with complete information requires the
the questionnaire). CBCT systems are typically divided into three acquisition of projection images from at least a scan arc of 180° [6].
different categories according to FOV size: (1) large FOV (>15 cm The rotation angle typically varies between 180° and 360° depend-
maximum scan volume height), (2) medium FOV (from 10 cm to ing on the CBCT device. In some scanners it is possible to select a
15 cm field height), and (3) small FOV (≤10 cm field height) [24]. variable angular range depending on scan protocol whereas most
Some of the devices available are multimodality imaging systems of the scanners function only with a single fixed rotation angle [9,25].
that also include digital 2D panoramic and cephalometric imaging The rotational movement and gantry of CBCT devices are mostly
features in the same unit, and thereby enable savings in office space based on existing dental panoramic gantry designs which use shorter
and investment cost. Additionally, in some devices there are also angular ranges. The number of projections collected during the ro-
possibilities to perform 3D photography (e.g. Planmeca ProFace for tation depends of the detector frame rate and varies from 128 to
all Promax 3D systems, Planmeca Oy, Helsinki, Finland) to assist pre- 1024 projections, giving different bases for the image reconstruc-
operative treatment planning and follow-up. tion [6,9]. The number of projections during the scan arc movement
is determined by the frame rate (number of projection images per
Gantry size and type second), the length of the rotational trajectory, and the speed of the
rotational movement [6]. Image quality is affected to some extent
The size and geometry of the available CBCT systems vary de- by the number of projection images, as an increased number of
pending mostly on the technique used for patient positioning. The samples basically result in higher image quality. The projections
weight of the systems themselves ranges from 128 kg to 600 kg. (basis images) form the total raw data which are then pre-processed
Scanning can be performed in a standing, sitting or supine posi- and used to reconstruct the final isotropic 3D image data. The
tion, depending on the scanner and the needs of a patient. The greyscale depth of the image data varies between 8 and 16 bits,
848
Table 1
CBCT device specifications of different systems.
Reconstruction
requirements,
Manufacturer
W × D × H (m)
Tube current
Weight (kg)
Trade name
positioning
Scan mode
FOV, D × H
Focal spot
Voxel size
Greyscale
Exposure
Scanning
potential
filtration
Rotation
Detector
time (s)
Patient
display
Space
(mm)
(mm)
angle
(mA)
Total
Tube
Dose
time
time
(cm)
(bit)
(kV)
(s)
(s)
(°)
Auge SOLIO Z CSi FPD 60–100 2–12 >2.5 mmAl 0.5 0.100–0.315 8.5–17 N/A Continuous N/A 180/360 12 1.90 × 1.37 N/A Standing 5.1 × 5.5, 9.7 × 10, 170
Cefla Sc* Carestream* Carestream* Carestream* Carestream* Carestream* Asahi Roentgen Asahi Roentgen Asahi Roentgen
CM × 2.40 16.1 × 10
JAP
Alioth CM CSi FPD 60–100 2–12 >2.8 mmAl 0.5 0.100–0.226 8.5–17 8.5–17 Continuous N/A 180/360 8–14 1.90 × 1.41 N/A Standing 5.1 × 5.5, 7.9 × 8, 314
× 2.29 11.6 × 8
JAP
20 × 17.9
CS8100 CMOS FPD 60–90 2–15 >2.5 mmAl 0.6 0.075–300 7–15 N/A Pulsed 10–55 N/A 14 1.0 × 1.2 × 2.2 DAP Standing/ 4 × 4, 5 × 5, 8 × 5, 92
wheelchair 8 × 8, 8 × 9
CS 9000 3D CMOS FPD 60–90 2–15 >2.5 mmAl 0.5 0.076–0.200 22 10.8 Pulsed 10–55 360 15 1.5 × 2.0 × 2.4 DAP Standing/ 5 × 3.75, 7.5 × 3.75 160
wheelchair (extended mode)
CS 9300 aSi FPD 60–90 2–15 >2.5 mmAl 0.7 0.090–0.500 12–28 6.2–20 Continuous/ 20–90 190/360 14 1.5 × 2.0 × 2.4 DAP Standing/ 5 × 5, 8 × 8, 10 × 5, 160
pulsed wheelchair 10 × 10, 17 × 6,
17 × 11, 17 × 13.5
CS 9300 Select aSi FPD 60–90 2–15 >2.5 mmAl 0.7 0.090–0.500 12–28 6.2–20 Continuous/ 20–90 190/360 14 1.5 × 2.0 × 2.4 DAP Standing/ 5 × 5, 8 × 8, 10 × 5, 160
pulsed wheelchair 10 × 10
9500 aSi FPD 60–90 2–15 >2.5 mmAl 0.7 0.200–0.300 24 10.8 Pulsed 140 360 14 1.73 × 1.64 DAP Standing/ 9 × 15, 18.4 × 20.6 176
× 2.4 wheelchair
MyRay HR image 90 1–10 11.4 mmAl 0.5–0.6 0.16–0.33 10–30 6.9 (std) Pulsed 120–180 190/360 12 1.54 × 2.51 DAP Supine 15, 11, 7 (spherical) 360
SkyView intensifier (std 15) × 1.72
– CCD
Reconstruction
requirements,
Manufacturer
W × D × H (m)
Tube current
Weight (kg)
Trade name
positioning
Scan mode
FOV, D × H
Focal spot
Voxel size
Greyscale
Exposure
Scanning
potential
filtration
Rotation
Detector
time (s)
Patient
display
Space
(mm)
(mm)
angle
(mA)
Total
Tube
Dose
time
time
(cm)
(bit)
(kV)
(s)
(s)
(°)
MyRay ASi FPD 60–90 1–10 N/A 0.5 0.075–0.15 18–26 3.6–9 Pulsed >15 360 14 1.31 × 1.52 × 2.4 DAP Standing/ 5 × 5, 8 × 5, 8 × 8, 170
J. Morita J. Morita Instrumentarium* Instrumentarium* Imtec Corp. Imaging Sciences Image Works Gendex Gendex Cefla Sc*
Hyperion X9 wheelchair 11 × 5 11 × 8,
11 × 13 (stitched)
GXCB-500 HD aSi FPD 90–120 3–8 N/A 0.5 0.125–0.400 8.9–23 N/A N/A 20–90 360 14 1.22 × 1.34 × 1.8 N/A Sitting 8 × 8, 14 × 8 231
GXDP-700 S aSi 57–90 −16 N/A 0.5 0.125–0.400 10–20 2.34– Pulsed 20–90 360 14 1.4 × 2.0 × 2.4 N/A Standing/ 4 × 6, 6 × 8 200
12.5 wheelchair
Panoura 18S CdTe 58–82 2–10 2.5 mmAl 0.5 0.08–0.1 N/A 11.5–23 N/A 90–180 190 (×2 16 1.89 × 1.14 N/A Standing/ 4 × 5.7, 7.7 × 5.4 205
(XP73) CMOS sliding × 2.32 wheelchair
sensor
system)
iCAT aSi 90–120 3–7 N/A 0.5 0.125–0.40 5–26.9 2–7.2 Pulsed 30 360 14 1.22 × 0.94 DAP Sitting 8 × 8, 4 × 16, 6 × 16, 231
International
Iluma Elite aSi 120 1–3.8 N/A 0.3 0.09–0.40 20–40 20–40 Continuous 120 190/360 14 2.14 × 1.83 × N/A Sitting Up to 10.8 × 9.6 350
(360), (360), 2.44 (small), up to
7.8 7.8 21.1 × 14.2 (large)
(190) (190)
OP-300 CMOS 57–90 3.2– 3.2 mmAl 0.5 0.085–0.33 10–20 1.2– Pulsed 10–60 200 16 2.01 × 1.41 DAP Standing/ 6.1 × 4.1, 6.1 × 7.8 250
16 12.6 × 2.41 wheelchair
OP-300 Maxio CMOS 57–90 3.2– 3.2 mmAl 0.5 0.085–0.42 10–40 1.2–9 Pulsed 10–60 200 16 2.01 × 1.41 DAP Standing/ 5 × 5, 6.1 × 7.8, 250
16 × 2.41 wheelchair 7.8 × 7.8, 7.8 × 15,
13 × 15
3D Accuitomo aSi (CsI) 60–90 1–10 3.1 mmAl 0.5 0.08–0.25 <18 <18 Continuous 20–120 360 14 1.62 × 1.20 CTDI Sitting 4 × 4, 6 × 6, 8 × 8, 400
170 × 2.08 10 × 10, 17 × 12
Veraviewepocs CMOS 60–80 1–10 N/A 0.5 0.125 9.4 9.4 Continuous 60–240 180 13 2.00 × 1.33 CTDI Standing/ 4 × 4, 4 × 8, 8 × 8 258
3D (CSi) × 2.36 wheelchair
849
850
Table 1 (continued)
Reconstruction
requirements,
Manufacturer
W × D × H (m)
Tube current
Weight (kg)
Trade name
positioning
Scan mode
FOV, D × H
Focal spot
Voxel size
Greyscale
Exposure
Scanning
potential
filtration
Rotation
Detector
time (s)
Patient
display
Space
(mm)
(mm)
angle
(mA)
Total
Tube
Dose
time
time
(cm)
(bit)
(kV)
(s)
(s)
(°)
Veraviewepocs CMOS 60–90 1–10 N/A 0.5 0.125 9.4 9.4 Continuous 30–60 180 14 2.00 × 2.30 CTDI Standing/ 4 × 4, 4 × 8, 8 × 5, 190/
J. Morita
× 1.83
GmbH
I-Max Touch aSi 60–86 6–10 2.5 mmAl 0.5 0.156 20 8 Pulsed <30 200/180 16 1.97 × 1.27 DAP Standing/ 8.3 × 9.3 140
Radiology*
Promax 3D aSi (CsI) 50–84 1–16 2.5 mmAl 0.5 0.075–0.4 18 4.8–15 Pulsed 30–150 200 15 2.15 × 1.63 CTDI/ Standing/ 4.2 × 4.2, 5 × 5, 119
Planmeca* Planmeca*
120
6.8 × 6.8, 8 × 8,
8.5 × 5, 10 × 6,
8.5 × 8.5, 10 × 10,
20 × 6, 20 × 10,
20 × 17
(continued on next page)
Table 1 (continued)
Reconstruction
requirements,
Manufacturer
W × D × H (m)
Tube current
Weight (kg)
Trade name
positioning
Scan mode
FOV, D × H
Focal spot
Voxel size
Greyscale
Exposure
Scanning
potential
filtration
Rotation
Detector
time (s)
Patient
display
Space
(mm)
(mm)
angle
(mA)
Total
Tube
Dose
time
time
(cm)
(bit)
(kV)
(s)
(s)
(°)
Promax 3D aSi (CsI) 60–90 1–12 2.5 mmAl 0.6 0.075–0.6 9–40 23.6–15 Pulsed 2–55 210 / 15 1.16 × 1.37 CTDI/ Standing/ 4.2 × 5, 5 × 5.5, 134
Max or 60– + 0.5 mmCu 360 × 2.39 DAP wheelchair 8.5 × 5, 10 × 5.5,
120 8.5 × 7,5, 10 × 9,
11 × 5, 13 × 5.5,
Planmeca*
11 × 7.5, 13 × 9
8.5 × 11, 10 × 13,
11 × 11, 13 × 13,
11 × 13.6, 13 × 16,
18.7 × 14.5, 22 × 17,
23 × 16, 23 × 26
Prexion 3D CSi FPD 90 4 N/A 0.2 0.101–0.147 8.6– 8.6– Continuous 30 217/ 13 1.57 × 1.17 N/A Sitting 8.1 × 7.5, 5.6 × 5.2 363
Prexion
11 × 13 (stitched)
NewTom VGi aSi 110 1–32 12.0 mmAl 0.3 0.1 15–18 1.8–4.3 Pulsed <60 360 16 1.20 × 1.13 CTDI Standing/ 5 × 5, 8 × 5, 8 × 8, 260
Systems*
12 × 8, 15 × 5,
15 × 12, 16 × 16,
24 × 19
NewTom 3G Image 110 1–15 1.4 mmAl @ 0.5–1.5 0.15–0.3 36 5.4 Pulsed >60 360 12 1.90 × 2.50 CTDI Supine 10, 15, 20 480
intensifier 70 kV × 1.98 (spherical)
+ CCD (inherent) +
QR Systems*
11.4 mmAl
@ 110 kVp
(12″ & 9″
FOV) OR
2.2 mmAl @
110 kVp (6″
FOV)***
NewTom 5G aSi 110 1–20 4.4 mmAl @ 0.3 0.075 18–26 3.6–6.7 Pulsed <60 360 14 1.75 × 3.60 CTDI Supine 6 × 6, 8 × 8, 12 × 8, 600
Systems*
70 kV × 1.78 15 × 5, 15 × 12,
QR
18 × 16
WhiteFox aSi 105 6–10 9.2 mmAl 0.5 0.10–0.50 18–27 6–9 Pulsed 30 360 16 1.89 × 1.58 × 2.38 N/A Standing/ 6 × 6, 8 × 8, 12 × 8, 275
Sirona Dental Sirona Dental Saletec
wheelchair 15 × 13, 20 × 17
GALILEOS Image 85 5–7 N/A 0.5 0.15–0.30 14 2–6 Pulsed 150–270 200 12 1.60 × 1.60 N/A Standing/ 15 (spherical) 120
Systems*
GALILEOS Image 98 3–6 N/A 0.5 0.125–0.25 14 in 2–5 Pulsed <240 200 12 1.60 × 1.60 N/A Standing/ 15.4 (spherical) 120
Systems*
851
852
Table 1 (continued)
Reconstruction
requirements,
Manufacturer
W × D × H (m)
Tube current
Weight (kg)
Trade name
positioning
Scan mode
FOV, D × H
Focal spot
Voxel size
Greyscale
Exposure
Scanning
potential
filtration
Rotation
Detector
time (s)
Patient
display
Space
(mm)
(mm)
angle
(mA)
Total
Tube
Dose
time
time
(cm)
(bit)
(kV)
(s)
(s)
(°)
Orthophos XG CMOS FPD 60–90 3–16 N/A 0.5 0.1–0.16 2–14 2–5 Pulsed 270 200 12 1.41 × 1.28 N/A Standing/ 5 × 5.5, 8 × 5.5, 110
Soredex* Sirona Dental
Systems*
3D × 2.25 wheelchair 8 × 8
Scanora 3D CMOS FPD 60–90 4–12.5 6.6 mmAl 0.5 0.133–0.35 10–26 2–6 Pulsed 60–240 360 12 1.60 × 1.40 DAP Sitting 6 × 6, 7.5 × 10, 310
× 1.97 7.5 × 14.5, 13 × 14.5
Cranex 3Dx CMOS FPD 57–90 4–16 3.2 mmAl 0.5 0.085–0.40 10–40 1–9 Pulsed 60–240 180/360 14 1.96 × 1.41 DAP Standing/ 5 × 5, 6.1 × 7.8, 200
× 2.41 wheelchair 7.8 × 7.8, 7.8 × 15,
13 × 15
Master 3Ds N/A 40–90 2–10 N/A 0.5 0.164–0.40 15–24 N/A N/A 9–51 360 14 1.02 × 1.24 N/A Standing/ 16 × 7, 16 × 10, N/A
× 2.29 sitting/ 20 × 15, 20 × 19
wheelchair
PaX-Reve 3D CMOS FPD 40–90 2–10 N/A 0.5 0.08–0.25 15–24 3–13.5 Pulsed <60 360 14 2.05 × 1.57 N/A Standing/ 5 × 5, 8 × 6, 12 × 8, 400
× 2.33 wheelchair 15 × 15, 15 × 19
PaX-Duo 3D CMOS FPD 60–90 2–10 N/A 0.5 0.08–0.20 15–24 <9 Pulsed 18–30 360 14 1.07 × 1.57 N/A Standing/ 5 × 5, 8.5 × 5, 400
× 2.35 wheelchair 8.5 × 8.5, 12 × 8.5
PaX-Zenith 3D CMOS FPD 50–120 4–10 N/A 0.5 0.08–0.30 15–24 <9 Pulsed <221 360 14 1.80 × 2.00 N/A Sitting/ 5 × 5, 8 × 8, 12 × 9, 493
Vatech
Figure 3. Mean photon energies for different dental CBCT scanners with 90 kVp tube voltage available were determined from the simulations using a 10° anode angle. The
highest mean energy, and thus the greatest total filtration, was observed to be in Planmeca’s CBCT systems that are using additional copper filtration in addition to alu-
minium filtration.
where the majority of modern CBCT models currently operate with with typical values between 60 and 90 kVp (see Table 1), resulting
14 bits [9]. in mean photon energies of 34–57 keV (based on SpekCalc [29] sim-
ulation, see also Fig. 3). Further kV optimisation could be possible
Voxel size in CBCT, where a medical physicist should be involved with possi-
ble voltage adjustments, if they become more available in the new
Currently, dental CBCT systems typically offer the possibility to CBCT models, and extending clinical indications [9]. The total fil-
choose the used voxel size according to the specific task. The small- tration of the systems, affecting the X-ray spectra (Fig. 4), varies
est voxel size available nowadays is 75 μm and the highest possible remarkably (from 2.5 mmAl at minimum up to 12 mmAl (NewTom
selection is 600 μm. The voxel size is related to spatial resolution, VGi evo, QR Systems srl, Verona, Italy)). Additionally, some scan-
and the smaller it is, the higher the spatial resolution can theoret- ners utilise additional copper filtration (Planmeca CBCT systems,
ically be. Thus, voxel size can influence diagnostic performance with Planmeca Oy, Helsinki, Finland) which hardens the photon beam
special tasks, such as the assessment of dental trauma, which re- remarkably.
quires a high level of detail [26,27]. However, smaller pixels capture
less X-ray photons, resulting in more image noise, and therefore the Scan time
radiation dose may have to be increased. In addition to voxel size,
the focal spot size also affects the spatial resolution. In most of the The scanning times in CBCT vary from approximately 5 seconds
CBCT systems the focal spot size is 0.5 mm, although the full range up to 40 seconds, depending among other things on the number
is from 0.2 mm to 1.5 mm. Although the theoretical spatial resolu- of acquired projection images. However, the exposure times are
tion of CBCT systems may be high due to small voxel and focal spot usually less than this due to pulsing the x-ray beam, ranging from
sizes, it is also affected by beam projection geometry, scatter and circa 1 second up to 40 seconds. Additionally, the reconstruction
patient movement, detector motion blur and fill factor, the number times differ between scanners from a few seconds up to several
of projections, and the reconstruction algorithm. As CBCT scan- minutes depending on the acquisition parameters (FOV, voxel size,
ning may take up to 40 s, the scanning is vulnerable to patient number of projections), hardware (processing speed, data transfer
motion. It has been discussed that the heartbeat alone induces a from acquisition to reconstruction computer) and software (recon-
slight but relevant movement of the patient’s head [17]. This move- struction algorithms) used [6].
ment has been shown to be approximately 80 μm per heartbeat [28].
Image reconstruction
Exposure settings
Image reconstruction in CBCT (and for CT in general) is a math-
Some of the dental CBCT devices still use a continuous radia- ematical process where the measured projection raw-data are related
tion exposure whereas most of the scanners are now utilising a to the image data which describe the patient object attenuation (i.e.
pulsed x-ray beam exposure. The generators operating in the pulsed reduction of beam intensity caused by photon interactions) with
mode typically result in lower radiation doses to patients as the ra- the used X-ray spectrum and beam geometry. The attenuation dis-
diation is turned off intermittently during the image acquisition tribution of the patient is presented in the 3D CBCT image data by
process. the greyscale value of each voxel. The image reconstruction is an
In some CBCT scanners the tube potential and tube current are inverse problem where traditional reconstruction methods do not
fixed for 3D imaging, whereas in others they can be changed within offer a direct solution. The reconstruction technique mostly used
certain limits, allowing better possibilities for scan optimisation in in CBCT utilises 3D filtered back projection (FBP) via the Feldkamp–
terms of radiation dose and image quality. Available tube currents Davis–Kress (FDK) algorithm [11] due to its simplicity and fast
vary between 1 and 32 mA. Tube voltages range from 40 to 120 kVp reconstruction times. Applied filtering in FBP can be adjusted to
854 T. Kiljunen et al./Physica Medica 31 (2015) 844–860
Figure 4. X-Ray spectra for different dental CBCT scanners with 90 kVp tube voltage available were determined from the simulations using a 10° anode angle. As total fil-
tration was increased, the beam was hardened, which can be seen as a higher mean energy of the beam.
control the level of sharpness of the resulting image data. However, and causes artefacts related to highly attenuating materials in the
the sharper image also accentuates noise. Balance between the level patient, such as metal implants. Beam hardening also causes a
of edge detection and noise must be set by the selection of an ap- cupping effect when the X-rays which pass through a thicker centre
propriate filter for each specific protocol or diagnostic task. Cone part of a cylindrical object are filtered to higher photon energies
beam CT with a limited amount of projections, shorter angular range compared to the X-rays which pass through the edges of an object,
in data acquisition, and truncated data could benefit considerably and thus are filtered less by the object thickness. The spectrum of
from more advanced iterative reconstruction techniques which are the X-ray beam can be narrowed by increasing the added filtra-
discussed in the later chapter (future aspects). tion of the beam to decrease the basis for beam hardening, but this
Typical clinical presentation of the dental CBCT image data in- will increase the power requirements of the X-ray source and/or
cludes axial, coronal and sagittal planes, in addition to a volume cause longer exposure times in order to acquire a sufficient amount
rendered 3D visualisation (Fig. 1). Image data can be navigated of quanta for the image signal collection. However, if added filtra-
through mouse-driven cursor lines with rotating and tilting func- tion can be applied, it is also beneficial due to the decreased radiation
tions as needed. Various specially designed visualisations may be exposure of the patient, as the softer X-rays are more optimally
used for dentomaxillofacial applications, including e.g. virtual pan- removed from the primary beam.
oramic display computed as a curved multi-planar reformat of the Scatter and beam hardening give rise to characteristic image ar-
3D image data (Fig. 2). tefacts in CBCT where dark shading and bright streaks are typically
shown especially around highly attenuating metallic objects within
Stitching volumes the image volume (Fig. 1) [11,31]. Scatter removal may include phys-
ical grids with a related increase in dose level, computational scatter
In some cases, the single FOV does not have enough coverage correction methods, including simple (polynomial) background signal
for the clinical volume of interest. Thus, separate scans producing estimations, or more elaborate Monte Carlo simulation based
FOVs adjacent to and partly overlapping each other may be stitched methods applied to each raw-data projection. Metal artefact cor-
together to provide a larger effective field of view. This stitching may rections are more application-specific and may take advantage of
be used to combine horizontal or vertical image volumes to compile an iterative reconstruction approach and prior information of the
a larger image volume. The stitching process itself may utilise exact metallic material in the scan volume [11]. Artefacts in the maxil-
information on patient positions between the separate original lofacial region are commonly produced by metal objects – metal
scans or automatic matching of the images using image registra- fillings and crowns, fixed orthodontic appliances, etc. – causing prob-
tion [6,11]. lems, for example, in caries diagnostics (Fig. 1).
An X-ray tube produces a beam with a continuous spectrum of The intrinsic greyscale values in MSCT image data are cali-
energies (Fig. 4). When the polyenergetic beam is attenuated and brated to the Hounsfield unit (HU) scale which describes the relative
scattered in the patient, the mean energy of the remaining primary X-ray attenuation in a voxel in relation to the attenuation of air
photons becomes higher. Therefore, the beam is hardened. Basic re- (−1000 in HU range) and water (zero in HU range). The applicabil-
construction algorithms such as FBP assume a monoenergetic beam. ity of the HU scale in CBCT image data is limited due to several
For the actual polyenergetic beam, the attenuation values of struc- physical factors including, most of all, X-ray scatter due to the cone
tures that cause beam hardening are thus falsely considered lower beam geometry, and projection data truncation, which bring un-
than their true attenuation values [16,30]. Beam hardening changes certainty and inconsistency to the claimed HU values and ultimately
the calculated attenuation values in the reconstructed image data limit low-contrast (soft-tissue) resolution. Therefore, the exact
T. Kiljunen et al./Physica Medica 31 (2015) 844–860 855
Table 2
The range and mean effective doses (μSv) for dental CBCT with different FOVs [24,41].
* Ludlow: L = large (>15 cm), M = medium (10–15 cm), S = small (<10 cm), Bornstein:
L = large (>100 cm2), M = medium (40–100 cm2), S = small (<40 cm2).
Table 3
The mean equivalent doses (μSv) for critical organs with different FOVs and standard imaging settings [24].
FOV* Bone marrow Bone surface Skin Oesophagus Brain Thyroid Salivary glands Remainder
Adult
L 359 1457 189 122 2182 1130 3484 475
M 233 844 163 75 1211 762 3675 442
S 94 299 62 43 211 413 2259 316
Child
L&M 130 448 142 79 962 1621 2918 404
S 60 213 65 34 171 843 2388 336
or are not expected to answer the diagnostic question and it is ex- detector noise, beam hardening, and truncation could be resolved
pected that CBCT will add new relevant information. by more realistic system and physics modelling in the course of the
reconstruction process.
Quality assurance Image reconstruction and post-processing techniques are im-
portant prerequisites for the overall image quality improvements
Quality assurance (QA) is an important part of the clinical use in dental CBCT imaging, and for evolving computed tomography tech-
of any radiological modality, including CBCT, meant to assure that nology in general. Challenges in image reconstruction are typically
the technical specifications and performance level of the scanner related to low photon counts and incomplete projection data with
are maintained during its life cycle. The manufacturer typically per- respect to acquisition angular range, the number of projections, or
forms quality assurance procedures during the original installation. truncation. Truncation effects are generated in traditional recon-
However, the user organisation should also pay attention to the per- struction techniques when the acquired projection data do not cover
formance of regular QA tests, related to further constancy and the entire cross-section of the attenuating patient object. Thus, the
maintenance testing, and ensure that QA phantoms and guide- attenuation measured in certain projections includes signal from
lines are in place and implemented during the clinical use of the outside the field-of-view, which causes higher calculated voxel values
device. This approach was also an essential part of the SEDENTEXCT near the lateral aspects of the reconstructed image volume. Itera-
project, including the formulation of a quality assurance pro- tive methods may provide a sound solution to the truncation problem
gramme. From the practical point of view, it should be emphasised in a form of interior tomography where the limitations of FBP data
that the staff performing dental imaging must have competence in consistency requirements can be surpassed [16,74–76].
dental imaging quality assurance issues [71]. The radiation exposure to patients can be decreased and spatial
resolution improved by pulsing the X-ray exposure during the gantry
Future aspects rotation in image acquisition. The evolving iterative image recon-
struction techniques might also require fewer projections as
Evolving detectors compressed sensing methods are coming into use. In those tech-
niques, the number of projections might be possible to reduce even
Current image detectors in X-ray modalities are energy- to only a few dozen, with a corresponding reduction in radiation
integrating. Thus, for each detector element, they accumulate all the exposure to the patient [16,77]. This type of X-ray exposure could
energy deposited by the x-ray quanta during the acquisition time be readily implemented with modern dental CBCTs which increas-
period to produce the detected signal charge. No information is ac- ingly utilise pulsed exposures. Overall, iterative reconstruction
quired on the single detected X-ray quantum energy. This may methods with more advanced modelling may bring significant im-
change in the future as photon counting detectors (PCD) are de- provement particularly in the field of CBCT imaging with sparse and
veloped further for general radiological use. In a photon counting truncated data, high local attenuation from metals and preferably
detector, information on the detected quantum energy deposi- low photon counts for reduced radiation exposure.
tions can be utilised to improve image quality, especially related
to soft-tissue low-contrast resolution and scatter correction, which Beam shaping filters
are prevailing challenges in dental CBCT imaging. PCDs are also likely
to allow reduced radiation exposures where (detector related) noise Unlike CBCT systems, the MSCT scanners include pre-patient
performance is otherwise a limiting factor in the low-dose range beam shaping filters to compensate for patient attenuation at the
[16]. detector signal level by spatially shaping the X-ray field intensity
within the scan field of view [78]. The function of a beam shaping
Iterative reconstruction (or bowtie) filter is to allow maximum X-ray intensity to the thick-
est part of a patient, which also attenuates the most x-rays, and to
As an alternative to traditional FBP methods, iterative recon- reduce X-ray intensity in peripheral areas with less attenuation,
struction methods approach the pursued image data solution in thereby reducing X-ray scatter, cupping effects, and the radiation
steps. An iteration step includes comparing the true measured raw- dose of surface tissues [79]. Such filters could also be beneficial in
data projections with forward-projected simulated projections based CBCT imaging as the larger field-of-views cover a more variable and
on the current image data estimate during the iteration process. The cylindrical attenuation geometry, and taking into account that the
aim of the iterative method is to reduce the difference between the prior reduction in the beam intensity also reduces scatter contri-
simulated raw-data and the real measured raw-data until a certain bution in image reconstruction and total radiation dose to the patient.
level of convergence has been achieved in the image estimate. Thus, Although the reduced scatter and cupping effects could bring ben-
the resulting image estimate corresponds to the final recon- efits, the use of beam shaping filters may also reduce detector
structed image. Actually, the first CT images back in the 70s were efficiency and low-contrast detectability as the effective mean photon
calculated based on the simple algebraic reconstruction tech- energy in the attenuated beam is somewhat higher [80]. At the
nique (ART) [30,72]. However, ART is much more computationally moment, beam shaping filters are not utilised in the commercial-
expensive than FBP and therefore the latter became the dominant ly available dental CBCT scanners.
method for decades. Recently, iterative methods have raised pop-
ularity since their strengths in improving image quality, especially Tube current modulation
related to noise and artefacts, have been proven in the field of MSCT
[73]. Modern iterative methods can model the acquisition chain more In MSCT, automatic tube current modulation is also used to
faithfully with respect to noise statistics and systems optics. Al- optimise the scan exposure, taking into account the patient atten-
though the computational burden remains and even increases (with uation in different parts of the longitudinal scan range and in different
more elaborate system and noise statistics modelling), current com- rotational angles. The purpose of mA-modulation is similar to beam
puter systems have processing power which enables the use of shaping filters: to equalise the detector signal by pre-patient mod-
iterative methods also in routine MSCT clinical settings with rea- ulation of the exposure. Therefore, a higher tube current is used in
sonable calculation times [16,73]. Iterative methods are likely to angulations where patient attenuation is higher (lateral directions
appear soon also in CBCT imaging, where the basic image quality during the rotation) and a lower tube current where the patient is
limitations due to a lower number of projections, X-ray scatter, thinner (ap/pa direction). A similar approach could also be used in
T. Kiljunen et al./Physica Medica 31 (2015) 844–860 859
CBCT to optimise the scan settings. However, this would require tech- [17] Brüllmann D, Schulze RKW. Spatial resolution in CBCT machines for dental/
maxillofacial applications – what do we know today? Dentomaxillofac Radiol
nical developments to add real-time mA-modulation features to the
2015;44:20140204.
X-ray generators in CBCT scanners [11]. [18] Ralph WJ, Jefferies JR. The minimal width of the periodontal space. J Oral Rehabil
1984;11:415–18.
[19] Dillenseger JP, Matern JF, Gros CI, Bornert F, Goetz C, Le Minor JM, et al. MSCT
General future aspects versus CBCT: evaluation of high-resolution acquisition modes for dento-
maxillary and skull-base imaging. Eur Radiol 2015;25(2):505–15.
Collaboration between medical physicists, radiologists and cli- [20] Kyriacou Y, Koditz D, Langner O, Krause J, Kalender W. Digitale
Volumentomographie (DVT) und Mehrschicht-Spiral-CT (MSCT): eine objective
nicians, radiographers, and engineers is essential for an efficient Untersuching von Dosis und Bildqualität. Fortschr Röntgenstr 2011;183:144–53.
optimisation process in modern radiology, also covering the use of [21] Larheim TA, Abrahamsson AK, Kristensen M, Arvidsson LZ. Temporomandibular
CBCT [81,82]. If and when the image quality can be improved by joint diagnostics using CBCT. Dentomaxillofac Radiol 2015;44(1):20140235.
[22] Fryback DG, Thornbury JR. The efficacy of diagnostic imaging. Med Decis Making
various developments in image acquisition technology, image post-
1991;11:88–94.
processing, and reconstruction methods, CBCT technology is likely [23] European Commission (EC). Radiation protection No. 172: evidence based
to gain ever increasing popularity in dental applications, but also guidelines on cone beam CT for dental and maxillofacial radiology. Luxembourg:
in other fields of medical imaging, e.g. in interventional and intra- Office for Official Publications of the European Communities; 2012 Available
from: <https://ec.europa.eu/energy/sites/ener/files/documents/172.pdf>.
operative imaging [82]. CT imaging has undergone huge [24] Ludlow JB, Timothy R, Walker C, Hunter R, Benavides E, Samuelson DB, et al.
developments from the first scanners in the 1970s to this day when Effective dose of dental CBCT – a meta analysis of published data and additional
CBCT is an essential and growing technological branch in the overall data for nine CBCT units. Dentomaxillofac Radiol 2015;44:20140197.
[25] Horner K, Jacobs R, Schultze R. Dental CBCT equipment and performance issues.
field of computed tomography. In the future, this rapid and ongoing Rad Prot Dosimetry 2013;153:212–18.
development is likely to bring various new clinical applications which [26] Melo SL, Bortoluzzi EA, Abreu M Jr, Corrêa LR, Corrêa M. Diagnostic ability of
may revolutionise the entire field of medical imaging once a cone-beam computed tomography scan to assess longitudinal root fractures
in prosthetically treated teeth. J Endod 2010;36:1879–82.
again. [27] Kamburoğlu K, Kursun S. A comparison of the diagnostic accuracy of CBCT
images of different voxel resolutions used to detect simulated small internal
Acknowledgements resorption cavities. Int Endod J 2010;43:798–807.
[28] De Kinkelder R, Kalkman J, Faber DJ, Schraa O, Kok PHB, Verbraak FD, et al.
Heartbeat-induced axial motion artifacts in optical coherence tomography
The authors would like to thank the Radiation and Nuclear Safety measurements of the retina. Invest Ophthalmol Vis Sci 2011;52:3908–13.
Authority of Finland (STUK) and Alexey Sofiev (HUS Medical Imaging [29] Poludniowski G1, Landry G, DeBlois F, Evans PM, Verhaegen F. SpekCalc: a
program to calculate photon spectra from tungsten anode x-ray tubes. Phys
Center) for the simulation facilities and assistance in the MC cal- Med Biol 2009;54(19):N433-8. doi:10.1088/0031-9155/54/19/N01. [Epub 2009
culations. The authors also would like to thank Alexander Meaney Sep 1].
(HUS Medical Imaging Center) for the proofreading of the [30] Pauwels R, Jacobs R, Singer SR, Mupparapu M. CBCT-based bone quality
assessment: are Hounsfield units applicable? Dentomaxillofac Radiol
manuscript. 2015;44(1):20140238.
[31] Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U,
References et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol 2011;40(5):265–73.
[32] Pauwels R, Cockmartin L, Ivanauskaité D, Urboniené A, Gavala S, Donta C, et al.
Estimating cancer risk from dental cone-beam CT exposure based on skin
[1] Boeddinghaus R, Whyte A. Current concepts in maxillofacial imaging. Eur J dosimetry. Phys Med Biol 2014;59:3877–91.
Radiol 2008;66:396–418. [33] Theodorakou C, Walker A, Horner K, Pauwels R, Bogaerts R, Jacobs R, et al. The
[2] Suomalainen A, Pakbaznejad Esmaeili E, Robinson S. Dentomaxillofacial SEDENTEXT Project consortium 2102 Estimation of paediatric organ and
imaging with panoramic views and cone beam CT. Insights Imaging 2015;6:1– effective doses from dental cone beam ct using anthropomorphic phantoms.
16. Br J Radiol 2012;85:153–60.
[3] Durack C, Patel S. Cone beam computed tomography in endodontics. Braz Dent [34] Al Najjar A, Colosi D, Dauer LT, Prins R, Patchell G, Branets I, et al. Comparison
J 2012;23(3):179–91. of adult and child equivalent doses from 2 dental cone-beam computed
[4] Adibi S, Zhang W, Servos T, O’Neill PN. Cone beam computed tomography in tomography units. Am J Orthod Dentofacial Orthop 2013;143:784–92.
dentistry: what dental educators and learners should know. J Dent Educ [35] UNSCEAR 2013. Sources, effects and risks of ionizing radiation. Report of the
2012;76(11):1437–42. United Nations Scientific Committee on the Effects of Atomic Radiation to the
[5] Mozzo P, Procacci C, Tacconi A, Martini PT, Andreis IA. A new volumetric CT General Assembly. UN, New York, 2013.
machine for dental imaging based on the cone-beam technique: preliminary [36] International Commission on Radiation Protection (ICRP). The 2007
results. Eur Radiol 1998;8:1558–64. recommendations of the international commission on radiation protection. ICRP
[6] Scarfe WC, Li Z, Aboelmaaty W, Scott SA, Farman AG. Maxillofacial cone beam publication 103. Ann ICRP 2007;37:2–4.
computed tomography: essence, elements and steps to interpretation. Aust Dent [37] UNSCEAR 2010. Summary of low-dose radiation effects on health. Report of
J 2012;57(Suppl. 1):46–60. the United Nations Scientific Committee on the Effects of Atomic Radiation to
[7] Scarfe WC, Farman AG. Cone beam computed tomography: a paradigm shift the General Assembly. UN, New York, 2010.
for clinical dentistry. Aust Dent Pract 2007;Jul–Aug:102–10. [38] Tubiana M, Feinendegen LE, Yang C, Kaminski JM. The linear no-threshold
[8] Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of a compact relationship is inconsistent with radiation biologic and experimental data.
computed tomographic apparatus for dental use. Dentomaxillofac Radiol Radiology 2009;251:13–22.
1999;28:245–8. [39] Suomalainen A, Kiljunen T, Käser Y, Peltola J, Kortesniemi M. Dosimetry
[9] Nemtoi A, Czink C, Haba D, Gahleitner A. Cone beam CT: a current overview and image quality of four dental cone beam computed tomography scanners
of devices. Dentomaxillofac Radiol 2013;42:20120443. compared with multislice computed tomography scanners. Dentomaxillofac
[10] Baba R, Ueda K, Okabe M. Using a flat-panel detector in high resolution cone Radiol 2009;38:367–78. Erratum in: Dentomaxillofac Radiol. 2009;
beam CT for dental imaging. Dentomaxillofac Radiol 2004;33:285–90. 38:554.
[11] Pauwels R, Araki K, Siewerdsen JH, Thongvigitmanee SS. Technical aspects of [40] Koivisto J, Kiljunen T, Tapiovaara M, Wolff J, Kortesniemi M. Assessment of
dental CBCT: state of the art. Dentomaxillofac Radiol 2015;44(1):20140224. radiation exposure in dental cone-beam computerized tomography with the
[12] Razavi T, Palmer RM, Davies J, Wilson R, Palmer PJ. Accuracy of measuring the use of metal-oxide semiconductor field-effect transistor (MOSFET) dosimeters
cortical bone thickness adjacent to dental implants using cone beam computed and Monte Carlo simulations. Oral Surg Oral Med Oral Pathol Oral Radiol
tomography. Clin Oral Implants Res 2010;21(7):718–25. 2012;114:393–400.
[13] Suomalainen A, Vehmas T, Kortesniemi M, Robinson S, Peltola J. Accuracy of [41] Bornstein MM, Scarfe WC, Vaughn VM, Jacobs R. Cone beam computed
linear measurements using dental cone beam and conventional multislice tomography in implant dentistry: a systematic review focusing on guidelines,
computed tomography. Dentomaxillofac Radiol 2008;37(1):10–17. doi:10.1259/ indications, and radiation dose risks. Int J Oral Maxillofac Implants
dmfr/14140281. 2014;29(Suppl.):55–77.
[14] Esmaeili F, Johari M, Haddadi P, Vatankhah M. Beam hardening artifacts: [42] Morant JJ, Salvadó M, Hernández-Girón I, Casanovas R, Ortega R, Calzado Y.
comparison between two cone beam computed tomography scanners. J Dent Dosimetry of a cone beam CT device for oral and maxillofacial radiology using
Res Dent Clin Dent Prospects 2012;6(2):49–53. Monte Carlo techniques and ICRP adult reference computational phantoms.
[15] Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J Opt Soc Am Dentomaxillofac Radiol 2013;42:92555893.
1984;1:612–19. [43] Prins R, Dauer LT, Colosi DC, Quinn B, Kleiman NJ, Bohle GC, et al. Significant
[16] McCollough CH, Chen GH, Kalender W, Leng S, Samei E, Taguchi K, et al. reduction in dental cone beam computed tomography (CBCT) eye dose through
Achieving routine submillisievert CT scanning: report from the summit on the use of leaded glasses. Oral Surg Oral Med Oral Pathol Oral Radiol Endod
management of radiation dose in CT. Radiology 2012;264(2):567–80. 2011;112:502–7.
860 T. Kiljunen et al./Physica Medica 31 (2015) 844–860
[44] Hall P, Granath F, Lundell M, Olsson K, Holm LE. Lenticular opacities in [63] Deak P, van Straten M, Shrimpton PC, Zankl M, Kalender WA. Validation of a
individuals exposed to ionizing radiation in infancy. Radiat Res 1999;152:190– Monte Carlo tool for patient-specific dose simulations in multi-slice computed
5. tomography. Eur Radiol 2008;18(4):759–72.
[45] Minamoto A, Taniguchi H, Yoshitani N, Mukai S, Yokoyama T, Kumagami T, et al. [64] Carter L, Farman AG, Geist J, Scarfe WC, Angelopoulos C, Nair MK, et al. American
Cataract in atomic bomb survivors. Int J Radiat Biol 2004;80:339–45. academy of oral and maxillofacial radiology executive opinion statement on
[46] International Commission on Radiation Protection (ICRP). ICRP statement on performing and interpreting diagnostic cone beam computed tomography. Oral
tissue reactions / early and late effects of radiation in 1209 normal tissues and Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106:561–2.
organs – threshold doses for tissue reactions in a radiation protection 1210 [65] Horner K, Islam M, Flygare L, Tsiklakis K, Whaites E. Basic principles for use
context. ICRP Publication 118. Ann ICRP 2012;41(1/2). of dental cone beam computed tomography: consensus guidelines of the
[47] Hidalgo A, Davies J, Horner K, Theodorakou C. Effectiveness of thyroid gland European Academy of Dental and Maxillofacial Radiology. Dentomaxillofac
shielding in dental CBCT using a paediatric anthropomorphic phantom. Radiol 2009;38:187–95.
Dentomaxillofac Radiol 2015;44:20140285. [66] Spin-Neto R, Gotfredsen E, Wenzel A. Impact of voxel size variation on CBCT-
[48] Qu X, Li G, Zhang Z, Ma X. Thyroid shields for radiation dose reduction during based diagnostic outcome in dentistry: a systematic review. J Digit Imaging
cone beam computed tomography scanning for different oral and maxillofacial 2013;26(4):813–20.
regions. Eur J Radiol 2012;81:e376–80. [67] Horner K, O’Malley L, Taylor K, Glenny AM. Guidelines for clinical use of CBCT:
[49] Tsiklakis K, Donta C, Gavala S, Karayianni K, Kamenopoulou V, Hourdakis CJ. a review. Dentomaxillofac Radiol 2015;44:20140225.
Dose reduction in maxillofacial imaging using low dose Cone Beam CT. Eur J [68] Brouwers MC, Kho ME, Browman GP, Burgers JS, Cluzeau F, Feder G, et al. AGREE
Radiol 2005;56:413–17. II: advancing guideline development, reporting and evaluation in health care.
[50] Lofthag-Hansen S, Thilander-Klang A, Ekestubbe A, Helmrot E, Gröndahl K. CMAJ 2010;182:E839–42.
Calculating effective dose on a cone beam computed tomography device: 3D [69] Petersen LB, Olsen KR, Matzen LH, Vaeth M, Wenzel A. Economic and health
Accuitomo and 3D Accuitomo FPD. Dentomaxillofac Radiol 2008;37:72– implications of routine CBCT examination before surgical removal of the
9. mandibular third molar in the Danish population. Dentomaxillofac Radiol
[51] Vassileva J, Stoyanov D. Quality control and patient dosimetry in dental cone 2015;44(6):20140406. doi:10.1259/dmfr.20140406. [Epub 2015 Mar 18].
beam CT. Radiat Prot Dosimetry 2010;139:310–12. [70] Brown J, Jacobs R, Levring Jäghagen E, Lindh C, Baksi G, Schulze D, et al. Basic
[52] Thilander-Klang A, Helmrot E. Methods of determining the effective dose in training requirements for the use of dental CBCT by dentists: a position paper
dental radiology. Radiat Prot Dosimetry 2010;139:306–9. prepared by the European Academy of DentoMaxilloFacial Radiology.
[53] Holroyd JR, Walker A. Recommendations for the design of X-ray facilities and Dentomaxillofac Radiol 2014;43:20130291.
the quality assurance of dental cone beam CT (computed tomography) systems [71] Metsälä E, Ekholm M. Quality assurance in digital dental imaging: a systematic
– a report of the HPA Working Party on dental cone beam CT. HPA-RPD-065. review. Acta Odontol Scand 2014;72:362–71.
Didcot, UK: Health Protection Agency; 2010. [72] Gordon R. A tutorial on ART (algebraic reconstruction techniques). IEEE Trans
[54] Endo A, Katoh T, Vasudeva SB, Kobayashi I, Okano T. A preliminary study to Nucl Sci 1970;21:471–81.
determine the diagnostic reference level using dose-area product for limited- [73] Beister M, Kolditz D, Kalender WA. Iterative reconstruction methods in X-ray
area cone beam CT. Dentomaxillofac Radiol 2013;42:20120097. CT. Phys Med 2012;28(2):94–108.
[55] Ludlow JB. Dose and risk in dental diagnostic imaging: with emphasis on [74] Yu L, Zou Y, Sidky EY, Pelizzari CA, Munro P, Pan X. Region of interest
dosimetry of CBCT. Korean J Oral Maxillofac Radiol 2009;39:175–84. reconstruction from truncated data in circular cone-beam CT. IEEE Trans Med
[56] Pauwels R, Theodorakou C, Walker A, Bosmans H, Jacobs R, Horner K, et al. Dose Imaging 2006;25(7):869–81.
distribution for dental cone beam CT and its implication for defining a dose [75] Yu H, Wang G. Compressed sensing based interior tomography. Phys Med Biol
index. Dentomaxillofac Radiol 2012;41:583–93. 2009;54(9):2791–805.
[57] International Commission on Radiation Units and Measurements (ICRU). Patient [76] Kolditz D, Kyriakou Y, Kalender WA. Volume-of-interest (VOI) imaging in C-arm
dosimetry for x-rays used in medical imaging. ICRU Report 74 J. ICRU 2005;5:1– flat-detector CT for high image quality at reduced dose. Med Phys
113. 2010;37(6):2719–30.
[58] Mori S, Endo M, Nishizawa K, Tsunoo T, Aoyama T, Fujiwara H, et al. Enlarged [77] Sidky EY, Kao CM, Pan X. Accurate image reconstruction from few-views and
longitudinal dose profiles in cone-beam CT and the need for modified dosimetry. limited angle data in divergent-beam CT. J Xray Sci Technol 2006;14(2):119–39.
Med Phys 2005;32:1061–9. [78] Toth T, Ge Z, Daly MP. The influence of patient centering on CT dose and image
[59] Dance DR, Young KC, van Engen RE. Further factors for the estimation of mean noise. Med Phys 2007;34:3093–101.
glandular dose using the United Kingdom, European and IAEA breast dosimetry [79] Toth TL. Dose reduction opportunities for CT scanners. Pediatr Radiol
protocols. Phys Med Biol 2009;54:4361–72. 2002;32:261–7.
[60] International Atomic Energy Agency (IAEA). Dosimetry in diagnostic radiology: [80] Miracle AC, Mukherji SK. Conebeam CT of the head and neck, part 1: physical
An international code of practice. Technical report series No. 457. IAEA, Vienna. principles. AJNR Am J Neuroradiol 2009;30(6):1088–95.
2007. [81] European Union, COUNCIL DIRECTIVE 2013/59/EURATOM of 5 December 2013,
[61] European Commission (EC). European guidelines for quality assurance in breast laying down basic safety standards for protection against the dangers arising
cancer screening and diagnosis. 4th ed. Luxembourg: European Commission; from exposure to ionising radiation, Official Journal of the European Union,
2006. Volume 57, 2014.
[62] Morant JJ, Salvadó M, Casanovas R, Hernández-Girón I, Velasco E, Calzado A. [82] Anderson PJ, Yong R, Surman TL, Rajion ZA, Ranjitkar S. Application of three-
Validation of a Monte Carlo simulation for dose assessment in dental cone beam dimensional computed tomography in craniofacial clinical practice and research.
CT examinations. Phys Med 2012;28:200–9. Aust Dent J 2014;59(Suppl. 1):174–85.