You are on page 1of 5

Article

Applied Biosafety:
Journal of ABSA International
Validation and Verification of Steam 2017, Vol. 22(1) 33-37
ª ABSA International 2017

Sterilization Procedures for the Reprints and permission:


sagepub.com/journalsPermissions.nav
DOI: 10.1177/1535676017694147
Decontamination of Biological Waste journals.sagepub.com/home/apb

in a Biocontainment Laboratory

Jeremy J. Bearss1, Shelley P. Honnold1, Edwin S. Picado1, Neil M. Davis1,


and James R. Lackemeyer2

Abstract
One of the most critical aspects of safe operations in biosafety level 3 and 4 and animal biosafety level 3 and 4 biocontainment
laboratories is the effective inactivation of biological select agents and toxins in the waste generated in these environments. Steam
sterilization has been used for decades to effectively kill microbial contaminants in a variety of medical and commercial settings. In
biocontainment research settings, steam sterilization is often used to inactivate biological select agents and toxins in various types
of waste prior to removal from the containment laboratory. This method is especially important for types of waste for which
other decontamination methods are not as effective or practical, such as chemical decontamination (quaternary ammonium,
sodium hypochlorite, formalin fixation, paraformaldehyde), ultraviolet light, or incineration. Bulk trash, animal waste, and animal
carcasses in particular do not easily lend themselves to decontamination via chemical or radiation methods. Decontamination of
these types of waste remains most effectively achieved with steam sterilization. Here we review the method used in our institute
to validate the steam sterilizers (autoclaves) and associated procedures to ensure destruction of microbial contaminants from
these challenging types of waste.

Keywords
biocontainment, biosafety level 3, biosafety level 4, sterilization, biological select agents and toxins, autoclave

With biosafety level (BSL-3 and BSL-4) and animal biosafety time.2 Steam should be dry saturated steam (dryness fraction
level 3 and 4 (ABSL-3 and ABSL-4) containment laboratories, 97%).2 Chamber pressure and temperature are usually con-
it is imperative that all items leaving the containment labora- trolled in unison, with pressures above atmospheric pressure
tory are effectively decontaminated to eliminate the presence being used to achieve temperatures above the nominal tempera-
of biological select agents and toxins. Typically, at least one of ture of steam at atmospheric pressure. Achieving higher
the following forms of decontamination is used: chemical chamber temperatures results in more efficient (faster) killing
decontamination (quaternary ammonium, sodium hypochlorite, of microorganisms. Typical temperatures used in commercial
alkaline digestion, formaldehyde solutions, or formaldehyde steam sterilization units2 are 121 C and 132 C; however, other
vapor), radiation (ionizing radiation, ultraviolet light), or thermal temperatures can be used, provided that the time for which the
sterilization (steam, incineration). chamber is maintained at these temperatures is sufficient. In
Chemical decontamination and, to a lesser extent, radiation, fact, some modern computer-controlled autoclaves use the con-
are typically used for surface decontamination of impervious cept of an F0 (F-naught) value, in which the additive effect of
material and equipment. Some types of waste, however, do not time at various temperatures throughout the cycle are
lend themselves to decontamination via chemicals or radiation,
and some bacterial spores have significant dry heat resistance,
making incineration less desirable for this application.1 For 1
Pathology Division, US Army Medical Research Institute of Infectious
several decades, steam sterilization has been successfully used Diseases, Fort Detrick, MD, USA
2
at multiple government and academic research institutions for Safety Division, US Army Medical Research Institute of Infectious Diseases,
the decontamination of trash, carcasses, and other biological Fort Detrick, MD, USA
waste prior to their removal from BSL-3 and BSL-4 and
Corresponding Author:
ABSL-3 and ABSL-4 containment laboratories. Jeremy J. Bearss, US Army Medical Research Institute of Infectious Diseases,
Effective steam sterilization must address key principles: 1425 Porter Street, Fort Detrick, MD, 21702-5011, USA.
steam quality, chamber pressure, chamber temperature, and Email: jeremy.j.bearss.mil@mail.mil
34 Applied Biosafety: Journal of ABSA International 22(1)

calculated to achieve the sterilization equivalent to 121 C at off and remove any water in the chamber to assist in drying of
15 psi above atmospheric pressure for 15 minutes.3 This allows the load.
for time spent at relatively lower temperatures to count toward Animal carcasses, tissues, and waste are packaged prior to
the total sterilization effect of the cycle and can significantly steam sterilization to prevent personnel and environmental
reduce sterilization time requirements. Utilizing standard auto- contamination, contain the materials, and maintain autoclave
clave cycles typically results in a 10-3 to 10-6 reduction in cleanliness. Packaging typically consists of 1 to 3 layers of
viable organisms4; however, some organisms, such as prions, impermeable heat-resistant plastic autoclave bags and/or a
may require additional time and/or temperature to achieve the cardboard box that provides some structural rigidity to the
desired reduction in bioburden.5 The Centers for Disease Con- package. While necessary, this packaging has the potential to
trol and Prevention guidelines stipulate a sterility assurance further impede the penetration of steam and heat into the bio-
level of 10-6 for microbiological and biomedical laboratories.6 hazard waste. Additionally, any pockets of trapped air created
While the principles of steam sterilization used for biologi- by this packaging will act as an insulator and further impede the
cal waste decontamination are identical to those used for dis- conduction of heat. Therefore, care must be taken while packa-
infection and sterilization of instruments and supplies in the ging biohazard waste to ensure that none of the packaging is
health care industry, the nature of these types of waste intro- sealed air-tight and that some movement of air into and out of
duces additional challenges to the sterilization process and the the package is maintained.7
validation thereof. These challenges include variability in ther- The final challenge for sterilizing this type of material is
mal mass and density, high water content, presence of trapped the significant difference in efficiency that is inherent to the
air pockets, and differences in autoclave design, efficiency, and physical engineering of the multiple makes and models of
sterilization cycle capabilities and configurations. autoclaves that may be found in any given institute. The
Variability in thermal mass and density is an especially removal of air from the chamber and load prior to the
challenging aspect of biohazard waste sterilization, particu- exposure/kill phase of a cycle differs significantly among
larly when sterilizing large amounts of animal tissue or car- manufacturer, machine model, and cycle configuration.
casses. In standard autoclave cycles utilized in the health care Some autoclaves are only capable of gravity displacement
industry, steam sterilization is used to sterilize instruments with steam pressure to push air from the chamber out
and supplies that are constructed of impervious materials that through the drain. With this process, air evacuation is often
readily conduct heat, such as metal and plastic, or fabrics and incomplete and greatly impedes steam penetration and heat-
other porous materials that can be effectively penetrated by ing efficiency. In an autoclave that performs a prevacuum,
the steam within the autoclave chamber. However, this is not the contents are subjected to 1 vacuum pulses prior to the
the case when sterilizing animal waste, tissues, and carcasses. exposure/kill phase to remove entrapped air from the con-
These materials do not readily conduct heat; in the cases of tents. With this process, air evacuation is more efficient
carcasses, some tissues (eg, skin, fat, and muscle) act as effec- and complete, resulting in significantly faster steam penetra-
tive insulators and impede the conduction of heat to the tion and heating efficiency. Various modifications of these
deeper tissues in the carcass. Nor do these materials have processes—such as the steam flush-pressure pulsing and
sufficient porosity to allow for the effective penetration of super-atmospheric steam-vacuum pulsing—may further alter
steam. As a result, significant additional time must be pro- air evacuation efficacy.2
grammed into the autoclave cycle to ensure that sufficient
sterilization temperatures are reached at all levels of the bio-
hazard waste within the autoclave.
Methods
Water content, particularly in carcasses and animal waste, is An appropriate validation/verification protocol for steam ster-
another challenging aspect of sterilization. High water content ilization should include initial installation qualification (IQ),
further slows the conduction of heat through the tissues due to operational qualification (OQ), and performance qualification
the water’s large capacity to absorb heat energy. Additionally, (PQ) tests. 8 A robust system of documentation must be
the high water content of these materials dictates that a liquid established to accompany all 3 phases (IQ, OQ, PQ) of the
sterilization cycle must be used to avoid vaporization of auto- validation/verification process.
clave contents at the end of the cycle. Liquid sterilization IQ is completed once the autoclave has been installed; all
cycles differ from dry cycles most significantly in the way that aspects of the design requirements (established prior to acqui-
pressure is evacuated from the chamber at the end of the cycle. sition of the autoclave) must be satisfactorily addressed. For
In a liquid cycle, the rate of pressure release from the chamber autoclaves used for the sterilization of biological waste from
is controlled to minimize the flash boiling of water to steam. biocontainment suites, this validation should address unique
This allows any liquid contents that are in the chamber to cool containment laboratory characteristics, such as one-way door
as chamber pressure is gradually decreased to atmospheric interlock systems, chamber drain effluent routing, chamber
pressure. In other types of cycles, the pressure is evacuated effluent filtering, and vacuum pump discharge. In our institute,
from the chamber abruptly, often assisted by vacuum, which various vendors over the past 20 years have installed multiple
results in the flash boiling/vaporization of any liquid in the makes and models of autoclaves, and IQ was performed and
chamber. This abrupt reduction in pressure is designed to flash documented as part of the initial validation of the equipment
Bearss et al 35

Table 1. Recurring OQ Tests for Sterilizer Validation. carcasses were used for the liquid cycle verification because,
in our experience, they are the most challenging type of waste
OQ Test Performance
Description Interval Standards
to sterilize; these carcasses were sourced from Institutional
Animal Care and Use Committee–approved animal protocols
Vacuum leak test Quarterly <0.78 psi leak over 15 min on a that were occurring in the laboratory serviced by the autoclave
sealed chamber at maximum being verified, thus avoiding the need for using additional ani-
vacuum mals specifically for the verification process. The dry/wrapped-
Air evacuation/ Quarterly Getinge Assure AccuFast
goods cycle was verified via a mixed load of 3 or 4 bags
steam Biological Test Pack (Steam)
penetration test rotated among 4 locations (enough to fill the chamber while maintaining a minimum
within chamber must be 2-inch clearance between the load and all interior surfaces of
negative the autoclave chamber) of dry trash and/or laundry, also
Equipment Quarterly Check door seal integrity, sourced from the trash/laundry being generated as part of
preventive mechanical operation, address normal laboratory operations. These load size limits were
maintenance software updates determined by the volume capacity and efficiency of the
Steam quality test Semiannual Dry steam at 100 psi
autoclaves within our facility. The temperature limits were
Temperature Semiannual Calibrated with NIST traceable
sensor instrument determined by the ability of the bags, packaging materials, and
calibration fabrics to withstand high temperatures. All test runs were
Pressure sensor Semiannual Calibrated with NIST traceable conducted through our facility’s standardized packaging
calibration instrument procedures for the appropriate types of waste.
Integrator strips Daily (every Getinge Assure Steam Sterilization Biological indicators (BIs) were used to evaluate the
load) Integrator, 1-2 strips per load efficacy of each sterilizer run. Our institute used MagnaAmp
Abbreviations: NIST, National Institute of Standards and Technology; Geobacillus stearothermophilus spore BIs for steam steriliza-
OQ, operational qualification. tion, supplied by Mesa Labs. Each BI contains 2.6  106 spores
of G stearothermophilus 7953 (traceable to a recognized cul-
ture collection identified in USP and ISO 11138) suspended in
jointly by both the vendor and the institute per the acceptance a pH-sensitive color-changing growth medium sealed inside a
criteria established contractually. 1-mL glass ampule. The medium is able to support germination
OQ includes verification that the installed equipment is of low numbers of spores and reliably indicate pH via color
operating properly within predetermined limits when used by change following extended heat exposures. For liquid cycles,
trained personnel in accordance with standard operating pro- each carcass contained 2 BIs; 1 BI was placed in the abdominal
cedures.8 Unlike, IQ, OQ is not a onetime event but rather cavity among the abdominal viscera, and 1 BI was placed
should be accomplished on a recurring basis to ensure contin- within a deep incision in the thigh muscles along the shaft of
ued verifiable operation of the autoclave over time by institute the femur. For dry/wrapped-goods cycles, each trash or laundry
staff. The formal tests to perform should include, at a mini- bag contained 2 or 3 BIs placed at different depths throughout
mum, a vacuum leak test, air evacuation/steam penetration test the contents of the bag. Each run contained a total of at least
(Bowie-Dick or equivalent test), steam quality testing, tem- 10 BIs, as well as 1 negative control ampule (containing growth
perature and pressure sensor calibration, equipment preventive media but no spores). For further cycle verification, each ster-
maintenance, and integrator strip testing.8 An example of an ilization run also contained a National Institute of Standards
OQ program used by our institute is described in Table 1. and Technology traceable, self-contained, wireless temperature
The PQ phase of autoclave verification must be designed to data logger to monitor internal carcass/laundry/trash tempera-
verify that the equipment is able to deliver the desired lethality ture during the sterilization cycle. Data loggers were supplied
when used by qualified personnel in accordance with standard by MadgeTech (model HITEMP140) and were set to record 1
operating procedures and therefore has, as a prerequisite, a temperature reading per minute. For liquid cycles, the data
robust and successful IQ/OQ program.8 PQ testing should logger was placed with the BI in the abdominal cavity of one
include both physical and biological measurements to deter- of the nonhuman primate carcasses. For dry/wrapped-goods
mine repeatable, uniform lethality in a typical autoclave load. cycles, the data logger was placed in the center of one of the
Due to the large number of autoclaves in use at our institute and bags being used for verification.
to account for differences in performance, autoclaves were Following the sterilization cycle, the contents of the auto-
subgrouped according to control system, design specification, clave were brought back into the containment laboratory where
and operational characteristics; PQ verification was performed they originated, and the BIs and data logger were retrieved,
on representative autoclaves from each subgroup. decontaminated, and removed from the laboratory for incuba-
For this validation, each autoclave subgroup was verified for tion. For easier retrieval and positive identification following
both liquid (for carcasses) and dry/wrapped goods (for trash the sterilization cycle, both the BIs and the data logger were
and laundry) cycles. The liquid cycle was verified with 6 non- identified by a plastic tag with a unique identifier, which was
human primate carcasses, each weighing no more than 10 kg, secured to the BI or data logger with Kevlar dental floss.
for a total maximum load of 60 kg. Nonhuman primate Following confirmation of successful autoclave verification,
36 Applied Biosafety: Journal of ABSA International 22(1)

the remaining contents of the autoclave verification run were 180


resterilized prior to leaving the containment laboratory. 160
BIs were incubated per manufacturer instructions at 57 C 140
120
for 48 hours and observed for color change and/or turbidity,

Minutes
100
which indicates spore germination; 1 ampoule containing 80
spores that was not sterilized (positive control) and 1 ampoule 60
40
not containing spores that was not sterilized (nonsterilized neg- 20
ative control) was incubated concurrently with the BIs from 0
105 110 115 120 125 130 135 140
each verification run. Data from the data logger were down- Temperature °C
loaded for analysis.
Figure 1. Time versus temperature curve for predicting 100%
lethality for Geobacillus stearothermophilus spores. Data derived from G
Results stearothermophilus biological indicators placed directly in an empty
During the validation of the numerous autoclaves at our insti- autoclave.
tute, we encountered considerable differences in unit effi-
ciency. Not surprising, the more modern autoclaves that chamber. As mentioned in the discussion of the F0 sterilization
employed prevacuum air evacuation at the beginning of the concept, even relatively lower temperatures can have a lethal
cycle were far more efficient. They were able to achieve effect on microbes, provided they are exposed for a sufficient
sterilization of test loads in liquid cycles in <4 hours and in amount of time. Peak core carcass temperatures of 110 C to
dry/wrapped-goods cycles in <2 hours. Other older autoclaves 115 C registered by the data logger were uniformly lethal for
relying on gravity displacement required as much as 7 hours to the spores in the adjacent BIs, provided the time held at these
achieve sterilization in liquid cycles and as much as 3 hours in temperatures was sufficient. To predict lethality more accu-
dry/wrapped-goods cycles. To account for these differences in rately during our verification process, we developed a time/
efficiency, we used the verification results of the least efficient temperature lethality curve (Figure 1) for the BIs. This was
autoclave, with an additional safety margin, to develop a single accomplished in an empty autoclave with BIs placed directly
standard for all autoclaves within the facility. Our facility’s in the autoclave chamber. The results of this lethality curve
liquid cycle standard for carcass decontamination is an correlated accurately with the combined results of the BIs and
8-hour exposure time at 121 C, and the dry/wrapped-goods data logger time/temperatures encountered in our liquid cycle
cycle standard for trash and laundry is 4 hours at 121 C. These and were predictive of successful validation tests in carcasses.
exposure times and temperatures are unique to our institute’s For the purposes of uniformity and reducing potential for
existing equipment and standard operating procedures; operat- user error, load limits in our institute are standardized for all
ing parameters for any new autoclaves will have to be validated autoclaves, based on the capacity of the smallest units. How-
to confirm that the standard facility procedures for autoclaving ever, during the verification of some of the larger autoclaves,
carcasses and dry waste are adequate for effective decontami- the size of the loads were increased to determine the maximum
nation of these types of waste. No attempt should be made to potential load size for these larger units. In our experience,
extrapolate these values to equipment in other facilities without physical volume of the chamber was the only significant lim-
proper validation. iting factor on the maximum load volume. Data from our ver-
The need for these extended cycle times is a direct result of ification tests indicate that the more substantial steam supply
the list of challenges for autoclaving waste reviewed earlier. In and greater thermal mass of chamber and steam jacket walls
our experience, the most challenging type of biological waste to were more than sufficient to offset the increased autoclave
sterilize is animal carcasses. Various barriers—including multi- load. However, one factor that had a significant effect on ster-
ple layers of sealed plastic autoclave bags, cardboard boxes with ilization efficacy was the temperature of the contents at the
tight-fitting lids, high tissue density and water content, and the beginning of the cycle. Verification tests carried out on frozen
insulating properties of other animal tissues (eg, fat and hair)— carcasses required significantly more time (hours) to achieve
significantly hamper air extraction and steam penetration into sterilization. As a result, standard operating procedures at our
the center of the carcass. One cannot overstress the importance institute dictate that the contents of the autoclave must be >0 C
of removing as many of these barriers as is feasible. In our at the beginning of any cycle.
experience, reducing the number of bags, maintaining a loose
closure on bags, and predrilling holes in the cardboard boxes
used for rigidity is critical to ensure proper sterilization. How-
Discussion
ever, even with these controls in place, extended exposure times Reliable lethality of sterilization cycles used for all types of
were required to achieve sterilization at the center of the carcass. waste (BSL-3, BSL-4, ABSL-3, and ABSL-4 biocontainment
The results of the data logger used in our verification pro- laboratories) is a critical component to safe biocontainment
cess demonstrated that successful sterilization of a carcass does laboratory operation and is crucial to protect public, occupa-
not necessarily require attaining the same temperature at the tional, and environmental health. A number of challenges must
center of the carcass tissue mass as is attained in the autoclave be overcome when using steam sterilization to process these
Bearss et al 37

unique types of waste. The process described here has been Funding
used to successfully validate the numerous models and designs The author(s) received no financial support for the research,
of autoclaves operated in our institute. As with any validation authorship, and/or publication of this article.
process, robust documentation of the validation plan and all
results is essential. Some parts of the validation process, such References
as the IQ portion, will occur only at the time of installation and 1. Wood J, Lemieux P, Betancourt D, Kariher P, Gatchalian N. Dry
following any major upgrades to design or function. Other parts thermal resistance of Bacillus anthracis (Sterne) spores and spores
of the validation process, such as the OQ, must be performed on of other Bacillus species: implications for biological agent destruc-
a recurring basis to ensure that operation of the sterilization tion via waste incineration. J Appl Microbiol. 2010;109(1):99-106.
units is reliable and consistent. Similar to IQ, the PQ process 2. Centers for Disease Control and Prevention. Guideline for Disin-
must be conducted following initial installation and any time fection and Sterilization in Healthcare Facilities, 2008.
there are significant changes in equipment, design, or user Washington, DC: US Department of Health and Human Services;
standard operating procedures, particularly when sterilization 2008. https://www.cdc.gov/hicpac/pdf/guidelines/Disinfection_
of waste is such a critical component of safe biocontainment Nov_2008.pdf.
laboratory operation. In performing PQ and developing stan- 3. Peleg M, Normand M, Corradini M. Generating microbial survival
dard operating procedures, physical characteristics unique to curves during thermal processing in real time. J Appl Microbiol.
the biohazard waste generation practices of the facility that 2005;98(2):406-417.
might affect sterilization efficiency should be carefully consid- 4. Woedke T, Kramer A. The limits of sterility assurance. GMS
ered. While time-consuming, this steam sterilization validation Krankenhhyg Interdiszip. 2008;3(3):Doc19.
process is essential to maintain the public trust in our ability to 5. Akikazu S, Yasuhisa A, Takashi O, et al. Fundamentals of prions
safely work with biological select agents and toxins. and their inactivation (review). Int J Mol Med. 2011;27(4):
483-489.
Authors’ Note 6. Centers for Disease Control and Prevention. Biosafety in Micro-
Opinions, interpretations, conclusions, and recommendations biological and Biomedical Laboratories. 5th ed. Washington, DC:
stated within the article are those of the authors and are not US Department of Health and Human Services; 2009. http://
necessarily endorsed by the US Army or the Department of www.cdc.gov/biosafety/publications/bmbl5.
Defense. Special thanks to Philip Haines and Jeremiah DeWolf 7. Stinson M, Galanek M, Ducatman A, Masse F, Kuritzkes R. Model
for their extensive technical expertise in steam sterilizer theory, for inactivation and disposal of infectious human immunodefi-
operation, and validation. ciency virus and radioactive waste in a BL3 facility. Appl Environ
Microbiol. 1990;56(1):264-268.
Declaration of Conflicting Interests 8. Hideharu S. Validation study and routine control monitoring of
The author(s) declared no potential conflicts of interest with respect to moist heat sterilization procedures. Biocontrol Sci. 2012;17(2):
the research, authorship, and/or publication of this article. 57-67.

You might also like