Application Note
The precision sourcing and measurement 1. Connect a type K mini- 3. Once your connections are
capabilities of the rugged Fluke Documenting connector into the “TC” input made, starting from the DPC
Process Calibrator (DPC) make it an ideal choice of the DPC and connect to power up state (or measure
for the calibration of process instrumentation. the thermocouple input of screen), select the “mA” but-
The following examples show how to address trans- the transmitter (Figure 1). It ton to measure milliamps
mitter calibrations quickly and effectively using the is necessary that you use the (Figure 2).
Fluke 753 or 754 Documenting Process Calibrators. correct type of mini-connec-
tor for this connection to the
DPC.
2. Connect the center two ter-
minals of the DPC to the con-
Calibrating a 2 Wire Thermocouple nections on the transmitter
Transmitter with the Fluke 750 Series where loop power is applied
Documenting Process Calibrators (Figure 1).
The transmitter in this example is a 2 wire trans- Note: If you prefer to use house
mitter (needs loop power). The input stimulus for loop power rather than supplying
Figure 2.
this device is from a type K thermocouple with an loop power from the DPC, connect
input range of 100 °C to 300 °C at 0 % and 100 %. these two leads in series with the 4. Next depress the “MEAS/
The output drawn against loop power is 4-20 mA loop supply to the transmitter. SOURCE” button; now you are
at 0 % and 100 % with a tolerance of 0.25 %. The in the source screen.
steps below show how to: 1) perform a three point Since you want to source a
ascending (3↑) As Found (pre-adjustment) test at thermocouple, press “TC/RTD”
0-50-100 %, and Tag the results, 2) adjust the once. Press ENTER to select
zero and span (span = 100 %) of the transmitter, TC. Use the down arrow to
then 3) perform a three point ascending As Left move the cursor to “K” and
(post-adjustment) test at 0-50-100 %, Tag those then press “enter.” Select
results and review them in memory. 754 DOCUMENTIN
“Linear T” and press ENTER.
The DPC will respond with
G PROCESS
CALIBRATOR
TEST DC PWR
– ++ –
Fluke DPC
Figure 1.
5. Next press the “MEAS/ 0.25 %. As long as the error
SOURCE” button once and exceeds .25 % the error win-
you will be in the split screen dow in the measure half of
mode (Figure 4). the screen will be in reverse
video. Adjust the zero on the
transmitter until the error is
less than 0.25 %. Next, select
“Go to 100 %” and adjust the
Figure 7. transmitter span in the same
way. You may also “Go to
8. Select “Auto Test,” and watch 50 %” to check linearity at
the DPC take 0 %, 50 % and mid-span.
100 % measurements of cur-
12. After adjustment is complete,
Figure 4. rent (4, 12 and 20 mA nomi-
select “As Left,” “Done” and
Note: Skip to step 7 if you are nal) while sourcing 100 °C,
“Autotest.” Watch the 0-50-
measuring current in series using 200 °C and 300 °C.
100 % evaluations. If the
the house loop power source 9. Once the DPC has completed adjustment was successful,
mentioned in step (2). this evaluation, it will display all the errors in the post run
the post test summary screen summary should be in normal
6. Next, depress the “SETUP” with the SOURCE, MEASURE video (or less than .25 %)
button on the DPC. The cursor and ERROR columns. Failures (Figure 10).
should begin on loop power. in the error column will be
Depress “ENTER,” down in reverse video (Figure 8).
arrow until the DPC cursor Depress “Done.”
is on Enabled and depress
“ENTER.” Next depress the
“Done” softkey (Figure 5).
Figure 10.
754 DOCUMENT
ING PROC
ESS CALIB
RATOR
C-1
C-2
S-2
TEST DC PWR
– ++ –
S-4
S-1
S-3
Fluke DPC
Figure 11.
1. Make the RTD source 3. Once the connections are 4. At this point, enter the split
connections, S1, S2, S3 and made as per the diagram in screen mode of the DPC and
S4 (omit S3 and S4 for 2 wire Figure 11, enable the loop perform the As Found, Adjust
RTDs and S4 for 3 wire RTDs; 24 volt power from the setup and As Left tests as you did
S4 is only needed for 4 wire screen, source the pertinent in the previous thermocouple
RTD sourcing). RTD for the specific transmit- transmitter example.
2. Next, connect the loop power ter being evaluated, and set
and current measurement the measure mode to mA.
connections, C1 and C2.
Figure 14.
Figure 18.
Figure 20.
1. Connect the pressure module 5. Next depress the “MEAS/ 7. Next depress the “MEAS/
to the DPC. Connect the input SOURCE” button; now you are SOURCE” button once and
of the P to I and the output in the source screen (Figure you should be in the split
of a pressure hand pump 21). screen mode (Figure 23).
to the input of the pressure
module using a “T” connector
(Figure 20). If you are using a
differential pressure module,
ensure (for this example) it is
a 15 PSI module and connect
to input labeled “H.”
2. Connect from the center 2
banana type connections of Figure 21. Figure 23.
the DPC to the input of the P
to I where loop power 6. Since we wish to source
is supplied (Figure 20). a pressure (using a hand
3. Next, depress the “SETUP” but- pump) depress the pressure
ton on the DPC. The cursor button, (directly above the
should begin on loop power. HART or range button)
Depress “ENTER,” down (Figure 22).
arrow to Enabled and depress
“ENTER.” Next depress the
“Done” softkey.
754
Figure 22.
Fluke DPC/TRACK2™
On Time Support
Process/Track
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Y
MA
NAG E M E NT
SY
Fax +1 (425) 446-5116
IT
Web access: http://www.fluke.com
ST
AL
E
QU
‚
M
01
90
RT
IF
IE O
D T IS
O M EET
Modification of this document is not permitted
without written permission from Fluke Corporation.