Professional Documents
Culture Documents
| Service Manual
SYSTEM VERSION
VERSION DESCRIPTION
1.0.1.43.4 HI application software
2.29D Firmware
COPYRIGHT
Copyright 2018. Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden,
Germany. All rights reserved.
No part of this documentation may be reproduced in any form, nor processed, copied or
distributed by means of electronic systems, without prior permission of HUMAN in wri-
ting. Since all precautionary measures were taken into account in producing these operating
instructions, the manufacturer accepts no responsibility for any errors or omissions. This
includes any liability for damage that could arise from possible incorrect operation based on this
information. Subject to changes without notice as result of technical development.
TABLE OF CONTENTS
1 SAFETY INSTRUCTIONS 13
1.1 INTRODUCTION 13
1.2 USER WARRANTY 13
1.3 INTENDED USE OF THE INSTRUMENT 13
1.4 GENERAL SAFETY WARNINGS 14
1.5 DISPOSAL MANAGEMENT CONCEPT 14
1.6 BIOHAZARD WARNING 15
1.7 INSTRUMENT DISINFECTION 15
2 TECHNICAL SPECIFICATION 17
2.1 OPERATION 17
2.2 SAMPLES 17
2.3 REAGENTS 17
2.4 PIPETTING 17
2.5 REACTION 18
2.6 SOFTWARE 18
2.7 ELECTROLYTES 18
2.8 ENVIRONMENT 18
3 INSTRUMENT OVERVIEW 19
3.1 GENERAL DESCRIPTION 19
3.2 MAIN CHARACTERISTICS 19
3.3 INSTALLATION AND REMOVAL FROM SERVICE 19
3.3.1 Shipping and delivery 19
3.3.2 Packaging 19
3.3.3 Responsibilities 20
3.3.4 Before setting-up 20
3.3.4.1 Site 20
3.3.4.2 Environment 21
3.3.4.3 Storage 21
3.3.4.4 Unpacking 21
3.3.5 Installation 23
3.3.5.1 Left side 23
3.3.5.2 Right side 23
3.3.5.3 PC connections 24
3.3.5.4 Instrument connections 24
3.3.5.5 Hydraulic connections 24
3.3.6 Moving 26
3.4 MAIN COMPONENTS 26
3.4.1 CPU board 28
3.4.2 Power board 0 28
3.4.3 Power board 1 29
3.4.4 Power board 2 29
3.4.5 Level and shock sensor board 30
3.4.6 Power supplies 30
3.4.7 Ground connections 31
3.4.8 Optical lamp 31
3.4.9 Hydraulics panel 32
3.4.10 Hydraulics panel - manifold 35
3.4.11 Wash station 35
3.4.12 External tanks 37
4 SERVICE MENU 39
4.1 ACCESS 39
4.2 TABS MENU 40
4.3 MOUSE USE 41
4.4 OPERATION 42
4.5 SERVICE MENU 43
4.6 HARDWARE TEST MENU (TE-OS-F1) 44
4.6.1 Hardware Test table 45
4.6.2 Logical positions 46
4.6.2.1 Inner plate 46
4.6.2.2 Outer plate 46
4.6.2.3 Filter selection 47
4.6.2.4 Wash station 47
4.6.2.5 Needle 1 rotation 48
4.6.2.6 Sampling needle 1 48
4.6.2.7 Diluter syringe 1 49
4.6.2.8 Needle 2 rotation 50
4.6.2.9 Sampling needle 2 50
4.6.2.10 Diluter syringe 1 51
4.6.3 Pumps and valves tests (TE-OS-F1-F4) 52
4.6.4 Continuous display of inputs (TE-OS-F1-F5) 53
4.7 MECHANICAL CALIBRATIONS MENU (TE-OS-F2) 54
4.7.1 Operations 54
4.7.2 Calibration rows 57
4.7.2.1 Reagent bottles (ARM 1 and ARM 2) 57
4.7.2.2 Diluent bottle (ARM 1 and ARM 2) 57
CONTENTS
5 ADJUSTMENTS 87
5.1 REACTION ROTOR TEMPERATURE 87
5.2 OPTICAL OFFSET AND GAIN 88
5.2.1 Necessary tools 88
5.2.2 Offset adjustment 88
5.2.3 Gain adjustment 89
5.2.4 Reading cell SELECTION SHORTCUT 90
5.2.5 INTERFERENCE FILTERS 90
5.2.6 READINGS OUT OF RANGE 90
5.3 FILTER WHEEL 90
5.3.1 NECESSARY TOOLS 90
5.3.2 OPTICAL HOME SWITCH ADJUSTMENT 91
5.3.3 FS BELT ADJUSTMENT 91
5.3.4 OFFSET ADJUSTMENT 92
5.4 REACTION ROTOR BELT TIGHTENING 92
5.4.1 NECESSARY TOOLS 92
5.4.2 OP BELT TIGHTENING 92
5.4.3 Access to the OP motor screws 93
5.5 SAMPLING NEEDLE VERTICAL ADJUSTMENT IN THE WASH WELL 94
5.6 WASH STATION DOWN ADJUSTMENT 95
5.6.1 ROUGH AUTOMATIC CALIBRATION 95
5.6.2 FINE CALIBRATION 95
6 SERVICING 97
6.1 LAMP REPLACEMENT 97
6.2 SAMPLE TRAY REPLACEMENT (40 / 60 POSITIONS) 99
6.3 SINGLE REACTION CELL REPLACEMENT 102
6.4 REACTION ROTOR REPLACEMENT 103
6.5 SAMPLING NEEDLE REPLACEMENT 104
6.6 FILTER REPLACEMENT AND EQUALIZATION 106
CONTENTS
9 TROUBLESHOOTING 133
9.1 THE ANALYZER DOES NOT CONNECT 133
9.1.1 Not connected 133
9.1.2 Notes on the USB-to-RS232 adapter 134
9.2 OPTICAL READINGS HAVE BAD REPEATABILITY 134
9.2.1 Noise on the optical signal 135
9.2.2 Imprecision in the reaction rotor positioning 135
9.2.3 Imprecision in the filter wheel positioning 135
9.2.4 Dispensation problems 135
9.2.5 Incorrect vertical positioning of the needle 136
9.3 AUTOZERO HAS BAD REPEATABILITY, UNSTABLE OPTICAL SIGNAL 136
9.4 REACTION ROTOR TEMPERATURE IS UNCONTROLLED 137
9.5 REACTION CUVETTES BECOME DIRTY AND EXCLUDED QUICKLY 138
9.5.1 First checks 138
9.5.2 Check the reaction cuvettes 138
9.5.3 Check the cause of high absorbances 139
9.5.4 Execute the pumps test 139
9.5.5 Execute cuvettes special wash 140
9.5.6 Instrument displays yellow rather than green cuvettes 140
9.6 A DROPLET FORMS ON THE NEEDLE TIP 141
9.6.1 At the end of the needle tip 141
9.6.2 On the side of the needle 142
9.7 ABERRANT OR NULL RESULTS AND REAGENT BOTTLES WRONGLY DE-
TECTED FULL (GREEN) 142
9.8 "NEEDLE SHOCK DETECTOR IS STUCK" ERROR MESSAGE 143
9.9 VACUUM PUMP ERRORS 143
9.9.1 Different causes 143
9.9.2 The TE-OS-F5-F5-1 Test Pumps procedure 144
9.10 NON-REPEATABLE TESTS TROUBLESHOOTING 144
9.10.1 Autozero test 145
9.10.2 Test pumps 145
9.10.3 OP+FS Home 145
9.11 OP CALIBRATION ERROR 145
9.12 INVALID AUTOZERO 146
9.13 SPIKES 146
9.14 WASH STATION DOWN ERROR 147
10 MAINTENANCE 149
10.1 UPDATE FIRMWARE AND SOFTWARE 149
10.2 COVER 149
10.3 OPTICAL LAMP 149
10.4 SAMPLING NEEDLE 150
10.5 WASH STATION NEEDLES 150
10.6 WIRING, CONNECTORS 150
10.7 HYDRAULIC TUBES, FLOATS 151
CONTENTS
1 SAFETY INSTRUCTIONS
1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read
the ”GENERAL SAFETY WARNINGS”, describing the appropriate operating pro-
cedures. Please contact your HUMAN authorised local Technical Service in the
event of instrument failure or other difficulties with the instrument.
The instrument must be used for its intended purpose. It must be ope- [IVD]
rated in perfect technical conditions, by qualified personnel, in such
working conditions and maintained as described in this manual, in the GENERAL
SAFETY WARNINGS. This manual contains instructions for qualified professional
operators.
14
Batteries, power packs and similar power sources must be removed from
electric/electronic parts and disposed of in accordance with applicable local
regulations.
FIGURE 1-1
Biological Hazard Symbol
2 TECHNICAL SPECIFICATION
2.1 Operation
2.2 Samples
2.3 Reagents
2.4 Pipetting
2 pipetting needles
Needle shock detector
Capacitive liquid level detector
Internal, external and method specific wash procedures
18
2.5 Reaction
2.6 Software
2.7 Electrolytes
2.8 Environment
3 INSTRUMENT OVERVIEW
Small desktop space, 950 x 780 x 520 mm (width x depth x height). 970 mm
height with open top cover.
Weight approximate 75 kg.
Up to 290 tests/h (without ISE).
Method, needle and cuvette management of incompatible tests.
Easy PAUSE request to add samples and reagents. Automatic recovery from em-
pty reagent bottles. Multiple reagent bottles management.
Extends range of kinetic tests through a dynamic processing of the readings.
Simple access to hydraulic pumps system and optical lamp for easy mainte-
nance.
Run-time sample predilution in reaction well. Automatic standards predilution.
The new analyzer and its accompanying accessories have been shipped in two
separate containers designed to provide adequate protection during transport
under normal conditions.
3.3.2 Packaging
3.3.3 Responsibilities
FIGURE 3-1
Indicator TIP-N-TELL (not tipped
or mishandled) Indicator
DROP-N-TELL (impact has not
exceeded 25G)
3.3.4.1 Site
The laboratory bench for the analyzer workstation should be a level surface of
solid construction to avoid vibration.
A space of 15 cm on both sides and behind the instrument is the minimum re-
quirement to allow adequate ventilation.
An additional work area on one or both sides of the instrument will contribute
greatly to the work efficiency of the technician during operation.
If sufficient space is available a surface 90 cm deep by 225 cm in working width
will allow a generous surface for the analyzer and the necessities of work (e.g.
tubes, reagents, samples, calibrators, controls, pipettes, user manuals but also
monitor, keyboard, mouse and printer – if necessary).
3.3.4.2 Environment
The location of the analyzer should be dust-free, away from drafts, heat sources
and direct sunlight.
Satisfactory operations may be conducted with temperature ranges from 16 °C
to 30 °C and with a variation during the testing process not to exceed ±2 °C.
Temperatures outside this range may cause erroneous operation.
Air conditioning may be required to ensure result quality if temperatures exceed
these limits.
Relative humidity should not fall below 10 % or rise above 80 % with no conden-
sation.
Ensure that the electromagnetic standards are met. Refer to European directive
on electromagnetic compatibility (see Document 89/336/CEE of 03/05/89).
3.3.4.3 Storage
If for any reason the instrument has been subjected to prolonged storage under
unfavorable conditions a revision by specialized technicians may be required be-
fore proceeding with installation.
3.3.4.4 Unpacking
The analyzer arrives in a wooden container, the top of which is secured with
screws.
The four walls are fixed to the lift-pallet with screws at the bottom line. Taking
off the bottom line screws allow lift up the complete box to get easy access to
the analyzer inside.
FIGURE 3-2
Wood container
The analyzer is fixed to the pallet by four mounting brackets. First unscrew the
brackets from the pallet and then loosen the counter nuts to remove the bra-
ckets from the analyzer feet.
FIGURE 3-3
Analyzer fixation
As you have already been told, and as HUMAN suggests, it is best to wait for the
arrival of a representative of the distributor to remove the analyzer from the
container (keep the wooden container in case you need to return the analyzer
for adjustment or repairs).
FIGURE 3-4
Shipment contents sheet
3.3.5 Installation
After selecting an appropriate site for the instrument and for the PC (inclusive
monitor, keyboard and mouse), you can proceed with the necessary connections.
Connections required are located in two points on the sides of each instrument.
FIGURE 3-5
Left side connections
From top:
- Floats connector (black)
- Liquid sensors connection plug snap-on connectors for:
- System solution tank (blue, not connected)
- Special wash solution tank (green
- Biologic waste tank (yellow)
- Normal waste tank (red)
FIGURE 3-6
Right side switches and connec-
tions
Cooling unit switch (yellow), main power switch (red) with power cord fitting
(incuding fuses compartment) and USB port type B for connection with the PC.
24
3.3.5.3 PC connections
FIGURE 3-7
"Schuko" connection
FIGURE 3-8
Vertical positions of the tanks
FIGURE 3-9
Routing of the tubes and the
sensor cables
Route the connecting tubes as straight as possible to avoid the creation of air
bubbles.
The waste tubes must always descend on its way to the waste tanks, in order to
avoid reflux or back pressure.
FIGURE 3-10
Connection of the floats and tubes
assembly to the manifold
Insert and rotate the floats bayonet connector (black) to secure the 4 liquid level
sensors.
Connect the 4 tanks to the corresponding hydraulic tube connections of the
analyzer.
26
3.3.6 Moving
The movement of an instrument after it has been put into operation should be
handled with extreme care. One person should never try to move an analyzer
alone. If the instrument needs to be moved over a large distance in or out of the
laboratory please follow the instructions in the sequence below:
- Execute the "Empty entire system" procedure (see User Manual > Mainte-
nance).
- Switch off the analyzer, the computer and the reagent refrigeration, and dis-
connect all lines and tubes connected to the instrument. Be sure to avoid
contact with potentially infectious liquid waste when disconnecting tubes.
- Fix the sampling arm in the uppermost position using the foam protection
tube provided with the shipment.
- At least two persons are now required for moving the analyzer. Lift the in-
strument slowly holding the metal base and in such a way that the cover
remains in the closed position.
- When the analyzer has been located in its new position, reconnect all lines
and tubes according to the installation procedure contained in this manual.
FIGURE 3-11
Optical preamplifier board
FIGURE 3-12
Internal USB converter board
FIGURE 3-13
2 level sensor boards
FIGURE 3-14
CPU board
FIGURE 3-15
3 power boards (numbered 0, 1, 2)
28
FIGURE 3-16
CPU board details
FIGURE 3-17
Power board 0 details
FIGURE 3-18
Power board 1 details
FIGURE 3-19
Power board 2 details
30
FIGURE 3-20
Level and shock sensor board
details
FIGURE 3-21
Level and shock sensor board
mounted and connected
FIGURE 3-22
Main power and cooling power
supplies
FIGURE 3-23
Ground terminal
FIGURE 3-24
Ground connection
FIGURE 3-25
Optical lamp
The HumaStar has an advanced lamp saving mode that helps to prolong the
lifetime of the halogen lamp.
It is controlled by the parameter (228) LMPSAV. The function is deactivated if
the parameter is set to zero.
If set (the parameter is not zero), after 10 minutes of inactivity and if the instru-
ment is in "Stop" state, the lamp voltage is reduced to 3V. Any user activity will
wake-up the lamp. The lamp returns to full voltage and needs 200 seconds to
be ready for measurement (200 is the recommended setting of the parameter).
FIGURE 3-26
Hydraulics panel
FIGURE 3-27
Counter-clockwise peristaltic
pumps operation
FIGURE 3-28
Hydraulics panel rear view (wiring
and tubing)
FIGURE 3-29
Hydraulics panel top view
(electrical connectors)
34
Electrical
Description Device destination
connector
EV1 Valve switching internal (on) / Pinch valve 1
external (off) needle wash 1
EV2 Valve switching internal (on) / Pinch valve 2
external (off) needle wash 2
MDS1 Diluter syringe stepper motor Diluter sampling arm 1
SDS1 Diluter syringe limit switch Limit switch diluter 1
MDS2 Diluter syringe stepper motor Diluter sampling arm 2
SDS2 Diluter syringe limit switch Limit switch diluter 2
PM10 Wash well 2 dispensing pump 10 Lateral nozzle well 2
PM9 Wash well 2 aspiration pump 9 Aspiration well 2
PM8 Wash well 1 dispensing pump 8 Lateral nozzle well 1
PM7 Wash well 1 aspiration pump 7 Aspiration well 1
PM6 Wash station aspiration pump 6 First long needle
PM5 Wash station dispensing pump 5 Short needle 5
PM4 Wash station dispensing pump 4 Short needle 4
PM3 Wash station dispensing pump 3 Short needle 3
PM2 Wash station dispensing pump 2 Short needle 2
PM1 Wash station dispensing pump 1 Short needle 1
FIGURE 3-30
Diluter tubing and pinch valve of
sampling arm 1 and 2
A Input from the needle wash dispensing pump 8/10 (arm 1/2)
B Output to the external needle wash nozzle
C Wash input to the diluter
D Wash output from the diluter to the sampling needle 1/2
E Output to the sampling needle
F Valve switching internal (on) / external (off) needle wash
FIGURE 3-31
Hydraulics panel manifold
FIGURE 3-32
Wash station top view with
tubing
FIGURE 3-33
Wash station needles assembly
FIGURE 3-34
Wash station top view without
tubing
36
FIGURE 3-35
Wash station bottom view with
tubing (manifold to hydraulics
panel)
FIGURE 3-36
Wash station assembly
FIGURE 3-39
External tanks
FIGURE 3-40
Floats and system solution
connectors - water tube
disconnected (left) / connected
(right)
The hydraulic tubes have snap-in connectors. The connector can be unplugged
pressing on the lateral metal button.
38
4 SERVICE MENU
The HI Service Terminal must be seen as a complex, unitary system enabling the
service engineer to perform all configuration, diagnostic, calibration and setup
tasks necessary to the correct functioning of the analyzer.
4.1 Access
To access to the Service Terminal, you must have logged to the HI software with
a "Service" or an "Installer" user name.
Press the "CLOUD" icon button at the center of the bottom part of the screen.
Select then the "TERMINAL" option.
FIGURE 4-1
Access to the Service Terminal
program
By the first window of the Service Terminal you are directly connected to the
analyzer's CPU. All the displayed screens are issued directly by the analyzer and
the PC is just a pass-through terminal.
40
FIGURE 4-2
Start window of the SERVICE
TERMINAL program
FIGURE 4-3
Logging enabled
- Right click inside the light gray area to select the "Screenshot" or "Print" func-
tions. On a touchscreen monitor these functions can be selected by touching
the enabled area for approximate two seconds.
- Selecting "Screenshot" opens a file selection dialog to define where to store
the screenshot as a ".JPG" file. The filename is automatically given and con-
tains a date and time stamp, e.g. all screenshots are stored in chronological
order.
FIGURE 4-5
Screenshot function selected
- Selecting "Print" allows to print the current screen or to show a preview first.
This function requires an installed printer.
FIGURE 4-6
Print menu selected
42
4.4 Operation
!
board.
Press the CAPS LOCK key to
set the upper case on the key-
To operate the Terminal program, click with the mouse inside the Terminal win-
dow to focus on it.
On a touchscreen monitor just touch inside the light gray area to focus on it.
This is mandatory for the use of
the Service Terminal. There are two function modes:
- "Command" mode, used by the HI software to control the analyzer.
- "Menu" mode, used by the service engineer to operate on the analyzer.
The service engineer can eventually use some of the commands of the "Com-
mand" mode to get some additional information on the instrument status, e. g.
PC<Enter>.
FIGURE 4-7
Result of the
"Menu" command PC
FIGURE 4-8
Start screen of the Service Menu
When the service operations are completed, you must exit the "Menu" mode ty-
ping <ESC> several times (depending on the menu level in which you are), until
the screen is cleared.
! The Service menu is control-
led by the function keys F1 to
F10 and by numerical keypad
numbers in case of submenus.
When you are in "Command" mode, typing <Enter> will generate the <NAK> re-
ply to be displayed on the screen. This can be used to verify that you are correctly
again in "Command" mode. ! To address a special function,
a short description will be
used, e.g: TE-OS-F4-F5-2 for the
The calibration, autocalibration and diagnostics system is uniquely powerful. Autozero Test function.
All the needed functions can be easily performed and the faulty components
detected. All the diagnostics can be run in remote control using software like
TeamViewer, allowing the service engineer to remotely acknowledge the pro-
blem and then go on site bringing with him the necessary spare parts.
44
FIGURE 4-9
Hardware Test menu
Function keys:
FIGURE 4-10
Hardware Test table
FIGURE 4-11
Inner plate logical positions
FIGURE 4-12
Outer plate logical positions
FIGURE 4-13
Filter selection
FIGURE 4-14
Wash station logical positions
FIGURE 4-15
Needle rotation 1 logical
positions
FIGURE 4-16
Needle up/down 1 logical
positions
FIGURE 4-17
Diluter syringe 1 volumina
FIGURE 4-18
Needle up/down 2 logical
positions
FIGURE 4-19
Needle up/down 2 logical
positions
(364) SN2WSH 364 At the bottom of the wash well white tip.
(368) SN2PTU 368 At the bottom of the primary sample tubes.
(367) SN2CUP 367 At the bottom of the sample cups inserted on the
sample tray cups adapter.
(370) SN2CU2 370 At the bottom of the sample cups on the inner ring
of the 20+20 samples tray.
(373) SN2DLC 373 At the bottom of the diluent bottle.
(366) SN2REL 366 At the bottom of the large reagent bottles.
(374) SN2RES 374 At the bottom of the small reagent bottles.
371 At the bottom of a tube in the reagent bottles
(371) SN2RET adapter.
(372) SN2REC 372 At the bottom of a cup in the reagent bottles
adapter.
FIGURE 4-20
Diluter syringe 2 volumina
FIGURE 4-21
Pumps and valves tests
FIGURE 4-22
Continuous display of inputs
When you select the calibration menu all motors will be homed first for safety.
FIGURE 4-23
Motos homing indication
4.7.1 OPERATIONS
The 39 calibrations are sorted into two pages, named "ARM 1" and "ARM 2":
- Page "ARM 1" contains all 18 calibrations for this arm. It also contains the
barcode reader, reading position and wash needles down calibration.
FIGURE 4-24
Sampling arm 1 calibration menu
FIGURE 4-25
Sampling arm 2 calibration menu
UP, DOWN IP/OP MOVE The up "" and down "" keys move the IP or OP
motor step by step.
LEFT, RIGHT RN MOVE The left "" and right "" keys move the RN
motor step by step.
+ - SN/WS MOVE The "+" and "-" keys of the numerical keypad move
the SN or WS motor step by step.
M MOTORS OFF Toggles the motors OFF and ON. When OFF, F5
MOTORS HOME switches the motors ON again.
The calibration point can be selected by direct address, typing the two digits of
the row. When one parameter of the calibration point is temporary modified, it
is shown in reverse (refer to Figure 4-26, row 01). With the F2 key, the new value
is recorded and the quote value returns to normal .
When a calibration point has been saved with F2, on the far right a v symbol
appears (refer to Figure 4-26, row 02).
Before access to a particular calibration row, the mnemonic names of the invol-
ved parameters are shown (refer to Figure 4-26, rows 03 and 04).
After access to a particular calibration row, the values of the involved parame-
ters are shown (refer to Figure 4-26, rows 01 and 02).
FIGURE 4-26
Marked saved calibrations;
mnemonic or value display
06 Sampling needle high, 4 mm over the diluent position, large bottle type.
F6 checks the liquid level in the large reagent bottle.
07 Sampling needle low at the bottom of the diluent position, large bottle
type.
Go down step by step, using the "+" key (numerical keypad) until the
vertical collision flag starts moving upwards. Go 7 steps up again, using
the "-" key (numerical keypad). Save the value.
58
18 Wash needles down calibration at the bottom of the cuvettes (one step
over the bottom contact).
In this calibration row, the OP position cannot be calibrated.
F6 is a calibration procedure that approximately finds out the (115) WS_
ASP quote for correct aspiration. After auto-search, the suggested value
is displayed on the far right of row 18. Use the lowest value, i.e. 364 (red
marked) in the example:
60
Some precision and stress tests are executed for one or more motors at the
same time.
The duration of the tests in seconds is predefined but changeable. The tests are
partially long lasting.
At the end of the test or at aborting the test pressing ESC, a report of the errors
or the home position variations is displayed.
FIGURE 4-27
Mechanical checks menu (bottom
part)
Function keys:
F1 RN1/RN2/IP Rotates the sampling arms and the sample tray
forward and backward.
F2 SN1/SN2/OP Moves the sampling arms up and down and the
reaction rotor forward and backward.
F3 WS/SN1/SN2/OP F2 in combination with the down and up move-
ment of the wash station.
F4 OP+FS HOME Executes the home search of the reaction rotor
and the filter wheel, starting from all their posi-
tions. After positioning of the reaction rotor and
before starting the home search, the wash station
moves down and up. The analyzer shows how
many uncertainties of the motor are detected. At
that point it is possible to decide if to improve the
tention of the motor belt or improve the motor
pulley position.
F5 ALL Executes the four F1, F2, F3 and F4 tests.
At the beginning it requests the duration of the F1,
F2 and F3 tests in seconds.
F6 OP READING MOV Tests the reaction rotor movement. The instru-
ment simulates the real movement that perform
during the reaction reading. There it is possible to
find any lost step or any critical rotation.
F7 DS1/DS2 Tests the function of the dilutors. Here it is possi-
ble to verify the steps of the diluters motors and
the diluters limit switches.
62
FIGURE 4-28
Example of the F5 ALL report
FIGURE 4-29
Example of the F6 OP reading
movement report
The instrument simulates the real movement that performs during the reaction
reading. There it is possible to find any lost step or any critical rotation.
FIGURE 4-30
Example of the F7 diluters check
report
Here it is possible to verify the steps of the diluters motors and the diluters limit
switches.
64
FIGURE 4-31
Example of the F8 FS inversion
check report
The motor moves forward and backward during the run and it is very important
that at the inversion moment no steps will be lost.
FIGURE 4-32
Example of the F9 OP belt check
report
The test suggests if the tension of the belt needs to be increased or decreased.
FIGURE 4-33
Example of the F10 OP inversion
check report
The motor moves forward and backward during the run and it is very important
that at the inversion moment no steps will be lost. This test reduces the delay of
the inversion step by step, starting from 40 cs to 1 cs.
66
FIGURE 4-34
Optical reading page
Function keys:
F1 SET O.D. REFER. Store the current A/D reading value in the refe-
rence value.
The O.D. value is the optical density calculated bet-
ween the current A/D value and the stored refe-
rence value.
F2 SELECT CUVETTE In the Reading Cuvette field, type the reaction cell
number (1 to 80) which should be used for the rea-
ding test. In position of cuvette 32, the two holes
for the adjustment of the gain and the offset of the
optical preamplifier are placed over the two trim-
mers to allow the access with a fitting screwdriver.
F3 CHANGE CUVETTE Positions the selected cuvette at the center of the
cuvette replacement hole - for cuvette inspection
or replacement.
F4 READ ALL FILT. Read and display all the autozero values of the fil-
ters.
F5 AUTOZERO Opens the autozero submenu.
F6 OP RDG. OFFSET Opens the reaction rotor reading offset submenu.
F7 FS WHEEL TEST Opens the filter wheel tests submenu.
Column 1 Number
Column 2 Wavelength in nm.
Column 3 Previous reading in Abs.
Column 4 Current reading in Abs.
Column 5 Difference in Abs. between previous and current reading.
Column 6 v = valid value.
Err = error, the value is out of range.
Dif = error, non repetitive.
Ovf = reading overflow.
Column 7 Current autozero reading or initial value of the autozero. It helps
to evaluate the filters downgrade.
The factory initial autozero values are stored in the CPU's memory
and can be retrieved by the command OZ9<Enter>.
Column 8 Virtual column on the right side, outside the table. Shows the "F4
Read All Filters function" values.
To get a reading of all filters for a selected empty cuvette in "column 8" of the
filters table.
If a cuvette 1 to 80 is selected, the reading is executed through the cuvette. So
the A/D value is lower because of the cuvette's absorbance. The reading through
cuvette 0 (gap between two cuvettes) is higher.
68
If filter and cuvette 0 are selected, the reading simulates an autozero reading. It
is faster then the command TE-OS-F4-F5-1 Autozero, because it reads only ones
and not 10 times.
FIGURE 4-35
Example of F4 Read all filters
FIGURE 4-36
Autozero
At the end of the autozero processing, a confirmation is requested. "1" stores the
current values in the old values,"0" restores the old values.
FIGURE 4-37
Autozero with Diff error
The analyzer could not obtain two autozero readings with the same value.
In case of Dif error, it is necessary to find out the origin of the unreliability:
1. Defective lamp, loose lamp screw
2. Lamp connections
3. Defective or broken or loose interference filter (TE-OS-F4-F5-2)
F5 AUTOZERO, 1:Test: Repeats ten times the autozero, calculates the average
and the CV% for every filter.
FIGURE 4-38
Autozero repeatability test
The Old row displays the last autozero values executed before the test. It is only
a reference and is not part of the test.
The CV% row displays for each filter the CV% variation coefficient of the ten
autozero repetitions.
Good values for the CV% are lower than 0.100 % for all the filters and lower than
0.150 % for filter 1.
70
If the CV% values are higher, the test must be evaluated in the following order:
1. FS OP
The "FS" column displays the variation in the home positioning of the filter
wheel. All values in the column must be zero. If not, the filter wheel must be
checked. The "OP" column displays the variation in the home positioning of the
reaction rotor. All values in the column must be zero. If not, the reaction rotor
must be checked.
2. Average O.D.
The "Aver." Column displays the average of all the autozero values. The "O.D."
column shows the O.D. of the average compared to the old average value. The
values of this column must not vary more than 0.0020 Abs. If the values have a
constant drift, the optical signal is not yet thermally stabilized and it must be re-
peated. If the values have a random variation, there is noise in the optical signal
(optical preamplifier, optical cable, CPU A/D converter) or the lamp contacts are
bad or oxidized (check the contacts, spray with contact cleaner).
3. DARK
The "Dark" column displays the value of the optical signal with the solid dark
filter. It must be in the range 10 to 100. If negative, it is suggested to increase
the dark (offset) to a positive value, acting on the Offset trimmer (Refer to "5.2.2.
Offset adjustment" and "5.2.3. Gain adjustment"). If the dark value in the ten
repetitions varies more than two units, the optical preamplifier of the CPU A/D
converter must be checked.
5. Summary
To recapitulate, if the CV% values are high, it is recommended to check the fol-
lowing points:
- The TE-OS-F3-F4 OP+FS Home check gives information on the precision of
the home positioning.
- TE-OS-F4-F7-2 FS offset: If the filters are not centered on the light beam, the
light variation and the CV% increase.
- TE-OS-F4-F6-1 OP offset: If the cuvettes are not centered, the light variation
and the CV% increase.
- Defective lamp, loose lamp screw.
- Defective or oxidized lamp connections.
- Defective or broken interference filter, cracked gray filter, loose locking ring.
- Optical preamplifier, optical cable, CPU A/D converter.
- TE-OS-F3-F4 OP+FS Home: Filter wheel belt check.
- TE-OS-F3-F4 OP+FS Home: Reaction rotor belt check.
To adjust the best reading points through the cuvettes and through the gaps
between the cuvettes.
FIGURE 4-39
Best reading points
FIGURE 4-40
Fine and rough offset
The procedure verifies if the current reaction rotor reading offset is correct or
must be modified. It verifies only the fine tuning and it is supposed that the
offset is already set within the range of the cuvette 1 (next procedure ROUGH
OFFSET). The test shows the current value of the reading offset (022) OPOFRD
and the suggested value. If the values are the same, the value of (022) OPOFRD
is not to be changed.
The +½ step is just to show the exact offset. It must of course be ignored when
updating the (022) OPOFRD parameter.
72
If the values are different, the service engineer can type Y to store the new value
in (022) OPOFRD or N to keep the old value.
It is used to find the rough reading offset of cuvette 1. The purpose it to adjust
the reaction cuvette 1 position in the center of the cuvette access opening in the
reaction rotor cover to ensure that the cuvette 1 will be used by the fine offset
adjustment.
FIGURE 4-41
Rough offset adjustment menu
FIGURE 4-42
Filter wheel test submenu
The home positioning is repeated 100 times and a correlation check is perfor-
med.
FIGURE 4-43
FS home repeatability test
The first two numbers are only for reference and must not be considered.
All the numbers in the rows with eight columns must be higher than 0.9990. In
this case the result is marked as -OK-
If the result is lower than 0.9990, the result is failing. In this case the result is
marked as FAIL. The final result is then "FAIL".
FIGURE 4-44
FS offset adjustment
FIGURE 4-45
Best reading point
The test scans automatically all the filters searching for the light peak.
For each filter is displayed if a correction has been applied to the offset parame-
ters (031) FSOFS1, (032) FSOFS2, …, (038) FSOFS8.
! Filter 9 (additional filter) is
not checked in this test.
The value between the two asterisks is the currently set peak value, before the
correction performed by the test. If the peak is correctly detected, no correction
is performed.
74
If some of the readings are dispayed with a red background, the values are in
overflow (3,276) or too close to the overflow (higher than 1,940). The analyzer
will automatically get the filter reading in the best position out of the center to
avoid the overflow.
In this case it is suggested to reduce the optical gain (refer to "5.2.3. Gain adjust-
ment") and then repeat this test until there are no more red areas.
FIGURE 4-46
Cuvettes status
The upper table is the current reading (in 0.1 mAbs) of the 80 cuvettes at the
filter 1 wavelength of 340 nm.
The lower table is the difference of the current reading to the startup reference
reading (in 0.1 mAbs) of the 80 cuvettes at the filter 1 wavelength of 340 nm.
1 The wash position is the position of the reaction rotor that is assigned to
operate under the first (rightmost) needle of the wash station. It ranges
from 1 to 80.
2 On the upper right corner: Continuous monitoring of temperature, floats
status and cover safety switch.
!
3 The reaction cells wash status bar. It displays the status of the 80 reaction If an error appears, the error
cells: message is displayed white-
0: Clean cell. on-blue in the line above the sta-
1: Dirty cell. tus bar 3.
2 to 9: Cell passed wash station 1 to 8 wash cycles.
F: Wash failed 1 or 2 times (the cell didn't pass the reading test after fill in
wash cycle 6).
W: Cell in warning to be excluded (wash failed 3 or 4 times).
X: Cell excluded (wash failed for 5 times).
The white-on-blue-background group of eight cells represents the eight
positions of the wash station. The first on the left is position 8 of the
wash station, the last on the right is position 1 of the wash station.
4 The pumps and lamp working hours.
The pumps dispensation in μl (last, average, flow rate, timers, pump sta-
tus).
Function keys:
F3 FILL W.CUVETTE The cuvette selected by the wash position will be filled
with water through the sampling needle.
F4 STARTUP Executes the startup procedure.
F5 PUMPS TESTS Opens the pumps tests submenu page.
F6 WASH NEEDLE Executes a needle wash cycle in the wash well.
F7 SING.WASH CYCLE Executes a single cycle of the wash station and moves
the wash position to the next cell.
F8 CLEAR ERRORS Clears the display of an error message.
F9 SPECIAL WASH Executes the Rinse Cuvettes procedure.
To select the current wash position, enter its number and confirm with <Enter>.
FIGURE 4-48
Select wash position
The wash station indicator moves with its rightmost position (wash needle 1) to
the selected position.
FIGURE 4-49
Execute washes
Fill the cuvette selected by the wash position with water through the sampling
needle.
FIGURE 4-50
Special pumps tests menu
FIGURE 4-51
Test pumps
78
For the pumps test needs, it is requested a primary tube filled with 5 ml of water
in sample position 1. If it is missing, the procedure is halted.
The pumps test performs a precise measure of the flow of the six dispensing
pumps of the wash station and the needle wash dispensing pumps. It also
checks the wash station vacuum pump.
The results of the test can be verified in the monitoring section of the mainte-
nance.
5 According to the new calibration of the level sensor in the reaction cells –
(278) OPL150 and (279) OPL450 – the analyzer asks – if necessary – if you
want to adjust the dispensing quote (053) SN1DS2 and (369) SN2DS2.
Type Y to adjust, N to keep the old value.
This procedure is used to fill all the hydraulic tubes, after tank replacement or
for the initial filling.
In the TE-OS-F5-F5 Pumps Tests menu only the fill all option is possible.
To avoid dangerous water overflow caused by aspiration pumps malfunctioning
or unpredictable results caused by wash dispensing failure, during the fill hy-
draulics procedure a self-test of the pumps is performed.
The self-test is performed mainly to check if the aspiration and vacuum pumps
are performing. It also checks the flow rate of the needle wash dispensation,
both internally and externally.
The self-test is based on the use of the level sensor. So the first step of the self-
test is to check the level sensor in the diluent bottle. If the diluent bottle is em-
pty, the procedure is halted.
FIGURE 4-52
Fill tubes + pumps selftest
Performs the same procedure as fill hydraulic tubes, except that its duration is
fixed and it doesn't check when the tubes are filled.
Extract the floats from the tanks before executing the empty tubes procedure.
Do not unplug the external tubes from the analyzer until the procedure is com-
pleted.
This procedure can be performed to verify the liquid residue left by the wash
station inside the reaction cells.
This test measures precisely the volume dispensed by the five wash station peri-
staltic pumps and allows eventually to adjust the pump timers values to obtain
the specified dispensation volume.
Here it is possible to find the main tests that are in the previous pages. It is use-
ful to perform these diagnostics tests in order to improve the optical path or in
order to calibrate and verify all pumps, all dispensations, all aspirations.
Furthermore here it is possible to perform all the sequence tests by F9 and save
automatically all screenshots on the desktop. If you want to save the screens-
hots automatically, you must set the parameter (205) SVSCRN to 1.
82
FIGURE 4-55
Parameters menu
Exit the Edit Parameters pages by pressing <ESC>. Following menu is shown:
FIGURE 4-56
Parameters exit menu
The name of the parameter pages is displayed in the upper right corner.
Trying to input a number outside the parameter's values range causes a beep,
the new value is rejected and the old value is kept.
A blue field indicates a parameter which you are allowed to change.
84
A blue reverse field indicates a value which has not yet been saved in EEPROM.
A red field indicates a parameter which you must not change.
FIGURE 4-57
Select a parameter by its number
Type the number of the parameter to be found (3 digits with leading zeros). The
cursor will move to the selected parameter.
The parameter values shown in blue cannot be edited if the status in the upper
left corner is Read only. With the status changed to Unlocked the parameter
values can be edited.
FIGURE 4-58
Select a parameter by its name
Type the name of the parameter to be found. The cursor will move to the se-
lected parameter.
FIGURE 4-59
Backup all parameters
Press F7 to restore the parameters values recorded during the last backup.
A confirmation is requested after typing three times 0 (zero):
FIGURE 4-60
Restore all parameters
5 ADJUSTMENTS
FIGURE 5-1
Thermometer probe placement
The sensor is in a feed-back loop for temperature control, so the system will al-
ways slowly drive the temperature to the same (apparent) reference.
If the measured temperature is higher than the set-point, decrease the value of
the parameter (285) OPTCAL so that the displayed temperature goes close to the
measured temperature and wait for stabilization.
If the measured temperature is lower than the set-point, increase the value of
the parameter (285) OPTCAL so that the displayed temperature goes close to
the measured temperature and wait for stabilization. The procedure must be
repeated sometimes until the displayed value is identical to the measured value.
88
Adjust the parameter (285) OPTCAL until the displayed OP temperature and an
external thermometer temperature, placed in the reaction cell, are the same.
The optical preamplifier board is located in the inner side of the reaction rotor.
The two trimmers for gain and offset are accessible in TE-OS-F4 Reading Test
menu selecting the cuvette 32 with the F2 Select Cuvette key or with the rea-
ding cells selection shortcut keys G and Z.
FIGURE 5-2
Access holes for Gain and Offset
adjustment
Offset and gain adjustment must be made after the thermal stabilization of the
optical group, the reaction rotor and the optical preamplifier. Wait at least 15
minutes.
Flat screwdriver 1.5 mm, with minimum 22 mm length, for the trimmers
(16890/171).
During the reading test, the operator can select all the filters with the keys 0 to
9 and get the corresponding reading.
Check on the auto zero readings list which is the filter with the highest rea-
ding and select this filter pressing the numeric key corresponding to the filter
number.
Adjust the gain with left trimmer that the maximum reading of the highest
(most transparent) filter is close to 56,000 and the lowest one (least transpa-
rent) is higher than 40,000. Note that the maximum reading of the A/D con-
verter is 65,000. The adjustment of the gain can also be done by parameter
(341) OPGAIN. It is possible to increase or decrease the value in order to improve
the filters values. The default value is 75. This process is very useful in remote
control (e.g. by TeamViewer).
In TE-OS-F4 Reading Test menu press S to modify the parameter (341) OPGAIN.
Since cuvette 32 is selected, the current reading is obtained with the light beam
through it. Depending on the absorbance of this cuvette, the A/D reading during
the adjustment of the trimmers is 10 % lower (or even less) than the real auto-
zero value. To check the true A/D reading, take the screwdriver off the holes and
select cuvette 0 with the F2 Select Cuvette key. Press the F4 Read All Filters key
to get a quick reading of all wavelengths.
In the A/D reading of the bare beam (cuvette 0 selected), it is preferable not to
go beyond 60,000 because it is needed some margin of tolerance to avoid rea-
ding overflow when changing the lamp. Lamps can have an emission tolerance,
mainly in the low wavelengths range (340 to 400 nm).
When all A/D reading values are satisfactory, press the F5 Autozero key and then
the 1 Single Autozero key.
Verify that the difference between all filters must be lower than 300 units to
avoid problem when you replace the lamp. In the TE-OS-F4 Reading Test menu
it is possible to press "A" to perform the automatic gain adjustment. The instru-
ment will set the highest filter to 56,800 units and will run the autozero. Wait
for the end of the procedure.
90
The interference filters are factory equalized, so that the maximum difference in
the A/D readings is typically lower than 20 %.
Anyway, to reduce the difference between the final transmittances, it is possible
that a light quartz grey filter has been mounted on some interference filters for
finer matching. Grey filters are available for 38 mAbs and 100 mAbs.
The difference between all filters must be lower than 700 units.
Interference filters have a very high stability and are mechanically protected. It
is very uncommon that they must be replaced. But it is eventually possible that
a quartz grey filter has cracked if it has been locked too tight by the black locking
ring.
If a filter has an autozero reading out of range, all the methods employing this
filter will be marked with AUTOZERO ERROR and these methods cannot be exe-
cuted.
The filter 1 (340 nm) is also used for the check of the cuvettes washing. An au-
tozero error on this filter will give an AUTOZERO ERROR during the washings.
To check if the filter wheel is operating properly, there are two basic tests to be
executed in this sequence: ! The optical limit switch must
never be moved. Only if it is
broken. All the adjustments will
1. TE-OS-F4-F7-1 FS Home Repeatability. be done without moving the opti-
2. TE-OS-F3-F4 FS Home Test. cal switch.
3. TE-OS-F4-F7-1 FS Home Repeatability.
The first test checks if the mechanics are working properly and if the stepper
motor is well phased on the optical limit switch. If all the results are not OK
(excluding the first row with two numbers), once you have excluded a big pro-
blem in the mechanics, it is possible that the home positioning is not repeatable
because the optical limit switch position is phased in the middle of two mecha-
nical steps.
In this case, the cogwheel of the FS motor must be shifted a bit. This is a blind
operation, because you can check the result only later, repeating the FS Home
Repeatability test.
1. Open the filter wheel cover (Allen screw M5 placed over the FS motor).
2. Release with a #2 Allen Key the two M3 Allen screws on the cogwheel of the
FS motor.
3. Rotate just a bit the cogwheel on the motor axis.
4. Tighten the two Allen screws of the motor's cogwheel.
5. Repeat the TE-OS-F4-F7-1FS Home Repeatability test. If it fails, go back to
step 2.
6. At the end, close the filter wheel cover.
The second test, FS Home Test, will check the Home search from all the filter
positions. When the filter wheel is operating perfectly, it must give all 0 (zero)
values.
If it gives non-zero values higher than 2, the filter wheel belt must be tightened.
To tighten the belt, release the four M4 Allen screws of the motor, pull with one
hand the motor, sliding outward, to tighten the belt and then tighten the four
motor screws with the other hand.
92
Once the FS positioning is accurate and repeatable, the offsets of the single fil-
ters must be detected. This operation is done automatically by the instrument
running TE-OS-F4-F7-2 Adjust FS Offsets.
1. Take off the sample tray and the instrument upper cover.
2. Move the reaction rotor with the hand that you can identify the OP motor.
3. Using the Allen Key, loosen the 4 screws that hold the OP motor.
FIGURE 5-3
Unscrew the OP motor screws
4 - While keeping the belt well tighten pulling the OP motor, tighten the four
motor's screws.
FIGURE 5-4
Fasten the OP motor belt
The 4 screws are placed on a square of approximately 5 cm. Two screws are
placed on the internal side and two screws are placed on the external side.
To find the internal screws it is necessary to move the IP and OP plate together.
FIGURE 5-5
First internal screw
FIGURE 5-6
Second internal screw
The 2 outer screws are accessible rotating the reaction rotor in such a way that
the hole placed on the bottom of the rotor will be, in sequence, exactly over the
2 screw heads.
To find the external screws it is necessary to move just the OP plate.
94
FIGURE 5-7
Third external screw
FIGURE 5-8
Fourth external screw
The 2 inner screws are accessible rotating first the sample rotor in such a way
that the hole on the IP plate will be overlying on the vertical of a second hole of
the reaction rotor (the IP hole and the OP hole must be then on the same radius).
Move then the IP and OP rotors over the heads of the 2 inner screws of the OP
motor.
Press F6 to enter the special check function and then press the space bar to step
through the check:
1. The water springs out from the needle and washes the needle internally and
externally. Check if the water flows well ((196) INWELV = pump speed).
2. The flow stops and the needle moves up to the drip position. The needle must
stay about one millimeter inside the water drop (the bottom of the needle must
be at the same level of the top of the white well nipple (adjust (045) SN1WSH).
3. The water flows from the lateral nozzle and washes the needle externally.
The water jet must hit the needle gently with a downward parabola and not hit
horizontally. In this way the drops will not remain attached on the needle side
(adjust (195) EXWL1V and (194) EXWL2V).
When the F6 special check function is completed, with the +/- keys you can ad-
just the vertical position of the needle inside the well that, when it goes up to
the drip position, the needle remains 1 mm inside the water drop.
3. At this point press the "+" key to go down until the shock sensor is engaged.
TTo be used only by operators not able to execute the FINE CALIBRATION proce-
dure.
Select calibration step 18 in the TE-OS-F2 Mechanical Calibration menu.
6 SERVICING
The lamp life has a wide distribution. It can range from 500 to 2,000 hours, with
typical life 1,000 hours. ! Do not touch the glass part
of the lamp with your fin-
gers. Fat, dust and humidity shor-
1. Switch the instrument off. ten the life of the lamp and limit
2. Open the lamp window in the back side of the instrument. the lamp emission.
3. Unplug the lamp connector (A).
FIGURE 6-1
Lamp and connector
FIGURE 6-2
Lamp knob
FIGURE 6-3
Lamp holder
FIGURE 6-4
Optical group window
13. Click the "Replace lamp" button to reset the lamp life timer and execute the
auto-calibration of the lamp. This procedure will adjust the mechanical off-
sets for the best optical performance and stores the autozero values in me-
mory.
range trimming the optical preamplifier gain. See "5.2.2 Offset adjustment"
and "5.2.3 Gain adjustment".
16. In TE-OS-F4-F7, verify the filter offsets with the function 2 "Adjust FS offsets"
and eventually adjust the filter offsets. See "4.9.5.2. Adjust FS offsets".
17. In TE-OS-F4-F6, verify the reaction rotor offset with the function 1 "Search
fine reading offset" and eventually adjust the reaction rotor reading offset.
See "4.9.4.1. Fine offset adjustment".
18. In TE you can run the command OZ6<Enter> to perform the automatic cali-
bration of the optical path and for the gain. The instrument will perform the
"Adjust FS offsets", the "Search fine reading offset" and will perform the gain
adjustment to set the higher filter(s) to 56,000.
19. Execute the start-up of the analyzer.
FIGURE 6-5
Standard sample tray
FIGURE 6-5
Alternative sample tray
100
Then in TE-OS-F2 Mechanical Calibrations the high and low position of the first
tube of the inner and the outer ring have to be calibrated for sampling arm 1
and sampling arm 2:
FIGURE 6-7
Reaction cuvettes window
FIGURE 6-8
Selected cuvette
You can verify that the engraved number corresponds to the selected one.
With the "Cuvette extraction pliers" 16890/54 (from the accessory box), extract
the cuvette placed at the center of the opening.
FIGURE 6-9
Reaction cuvettes replacement
Insert the new cuvette. The cuvette has a rectangular foot. It must be inserted
oriented so that the marks (one or two) on the bottom face are situated on the
external side of the rotor ring.
Confirm on the screen that the cuvette has been changed.
FIGURE 6-10
Reaction cuvettes replacement
confirmation
This will reset all the cuvette references and autozeros and mark the cuvette as
dirty, to enable its washing before it can be used in a reaction.
Disassemble the needles group of the wash station (two screws A and B) and
move it aside, but don't damage the TEFLON tubes.
FIGURE 6-11
Wash station needle group
replacement
Disassemble the black reaction rotor cover (three screws D, E and F) and remove
it.
FIGURE 6-12
Reaction rotor cover replacement
104
Unscrew the reaction carousel (six screws A, B, C, D, E and F) and remove it.
FIGURE 6-13
Reaction carousel replacement
Insert the new carousel taking care to keep the orientation (one pin marked with
a red circle in above figure).
Tighten the six screws progressively, taking care the rotor is the correct flat po-
sition.
Reassemble the reaction rotor cover and the wash station needle group.
FIGURE 6-14
Sampling arm cover
FIGURE 6-15
Needle holder cap replacement
FIGURE 6-16
Needle holder cap replacement
FIGURE 6-17
TEFLON tube replacement
Pull up the old needle and put down the new one.
FIGURE 6-18
Needle replacement
After reassembling, check that the needle slides correctly in the needle guide
without friction, so that the spring holds the needle in the lower position and
the shock sensor is correctly activated in the upper position.
If the needle doesn't slide freely, try to loosen the white needle holder cap and
verify the spring position.
FIGURE 6-19
Needle holder and shock flag
components
FIGURE 6-20
Filter wheel
The filters are equalized (matched) so that the reading of each filter is within
a ±20 % difference in relation to the 340 nm reference filter. Typical final ad-
justment of the Autozero of the filters is a maximum of 56,000 units and a mi-
nimum of 36,000 units but the difference between them must be lower than
10,000 units.
The filters supplied belong to a set of 8 or 9 factory matched filters. This me-
ans that, typically, the filters don't need a strong equalization with dark gray
filters or pinhole disks. It can happen that light gray quartz filters of 38 mAbs, or
100°mAbs (in any applicable combination) are added to the filters with a higher
transmittance to reduce their emission and to have a better equalization.
10. If the A/D reading of the new filter is lower than the others, you must check if
you can increase the gain of the optical preamplifier (if no readings are high-
er than 60,000). If not, you can leave it at its low level if its reading is higher
than 36,000. Else you must add gray filter(s) to the filters with the highest
readings and then increase the optical preamplifier gain again.
- Select the DARK position on the filter wheel with the '0' key and cuvette 32
with the F2 key. Adjust the optical preamplifier offset in the range 500 to
1,000.
- Select cuvette 0 with TE-OS-F4-F2 Select Cuvette. One by one, adjust all the
mounted filters.
- Adjust the readings with gray filters assembled together with the interfe-
rence filters.
- Adjust the optical preamplifier gain selecting cuvette 32, then go back to
cuvette 0.
The order of the filter assembly has to be gray filter, O-ring, interference filter,
locking ring.
The reading of all the filters must be in the range 40,000 to 60,000. Try to reach
50,000 ± 5,000 units.
Execute TE-OS-F4-F5-1 Autozero to verify the final readings (they can be slightly
different from those obtained from the immediate reading) and eventually re-
peat the matching.
The two vacuum pumps are electrically connected in parallel on the same out-
put signal.
FIGURE 6-21
Vacuum pumps
108
The vacuum pump on the left is aspirating from the wash station aspiration
needles.
The vacuum pump on the right is aspirating from the wash station cleaning noz-
zle.
The outputs of the vacuum pumps are connected to a Y nipple (E) via the hydrau-
lics panel manifold (C) to the output for the waste tank (D).
FIGURE 6-22
Vacuum pumps connections
FIGURE 6-23
Vacuum pumps head unscrewed
FIGURE 6-24
Vacuum pumps head opened
The two rubber membrane valves are inserted in two pins and surrounded by
rubber O-rings.
Take off the two membrane valves.
Center the two new membrane valves on the two pins and carefully position
them.
Ensure the right position of the two rubber O-rings.
FIGURE 6-25
Vacuum pump membrane valves
Close the pump head and fix it tight by the four screws.
Open the side cover of the hydraulic unit placed on the left side of the instru-
ment.
Find which pump you want to replace.
Extract the two tubes of the pump head from the barbed fittings.
Extract the pump head clenching the two clips placed on the two sides of the
pump head.
FIGURE 6-26
Peristaltic pump head replace-
ment
Push the new pump head down until you hear a click and reconnect the two
tubes.
Don't forget to prime the tubing to avoid any trapped air bubbles.
Disconnect the motor connector (MDS1 or MDS2) and the limit switch connec-
tor (SDS1 or SDS2) and take them off the hydraulic panel frame.
110
FIGURE 6-27
Electrical connections of the
diluter
FIGURE 6-28
Hydraulic and mechanic connec-
tions of the diluter
Unscrew on the hydraulic panel the old diluter with the four screws (green mar-
ked) at the base of the diluter head.
Place the new diluter and fix it with the four screws.
Connect the two hydraulic fittings.
Insert the new MDS1 or MDS2 and SDS1 or SDS2 connectors on the hydraulic
panel frame and connect them to the wiring.
The parameter (068) DSTYPE is normally set to 3 (default setting). If, due to
hydraulic narrowings, the diluter is not working properly, set parameter (068)
DSTYPE to 4 and the parameter (133) DSVELD to 1,520 or 1,400 to have higher
torque margins.
These values are automatically updated on exit of the TE-OS-F7 Edit Parameters
procedure, after having changed the (068) DSTYPE parameter.
On exit, press the F2 key to accept the modifications.
The sealing gasket must be replaced before it doesn't seal anymore and there is
a water leakage from the back of the head or from the optical switch opening.
The replacement is also part of the 12 months preventive maintenance.
Remove the system solution connector (blue) to avoid to completely emptying
the tubes.
Unscrew the diluter head with an Allen key and extract the head.
Remove the old sealing gasket.
Place the new sealing gasket, oriented with the orange part on the inner side.
Insert gently the head on the piston, avoiding to damage the new gasket.
Tighten gently the two screws.
Reconnect the system solution connector.
Execute a filling of the water tubes.
FIGURE 6-29
Diluter sealing gasket replace-
ment
112
- Disconnect the wiring connectors and the blue needle tube fitting.
- Unscrew the two screws on the two lower pillars.
- Unscrew the two screws on the two upper pillars. One of the two screws will
remain trapped under the reaction rotor incubator, but the sampling arm
can be anyway disengaged. This screw is the last to be unscrewed during the
disassembling and the first to be screwed during the reassembling.
- Replace the sampling arm, tighten first the two screws of the upper pillars,
then the two screws of the lower pillars, connect the electrical connectors
and the blue needle tube fitting.
- Go to the TE-OS-F2 Mechanical Calibrations menu and adjust all the quotes
related to the sampling needle positions.
Refer to the sampling arm replacement procedure for the disassembly and re-
assembly of the arm.
Insert the following wirings from bottom after removing the old ones:
FIGURE 6-31
Sampling arm SN optical limit
switch wiring (16890/113-1)
FIGURE 6-32
Sampling arm level sensor board
wiring (16890/112-1)
FIGURE 6-33
Small motor wiring (16890/201-
1)
Insert the TEFLON tube from bottom after removing the old one:
FIGURE 6-34
TEFLON tube with two fittings
(16890/117-1)
114
FIGURE 6-35
Lengths of wirings and TEFLON
tube
FIGURE 6-36
Insert the connectors in the ap-
propriate sockets
FIGURE 6-37
Add the previously cut cable
clamp
FIGURE 6-38
Fix the wirings with the clamp
FIGURE 6-39
Sampling arm positioning
Since the new wirings are internally twisted, it is necessary to turn the sampling arm
(top view) counter-clockwise until mechanical blocking. This is to avoid further twi-
sting of the wirings and must be done before proceeding with the central part as-
sembly.
FIGURE 6-40
Pass all wirings through the hole
FIGURE 6-41
Assemble the small motor wiring
FIGURE 6-42
Routing and connecting
! Swapping/short-circuiting
the connectors can create se-
rious damages.
Legend:
A Small motor wiring
B Small motor
C Limit switch wiring
D Limit switch
FIGURE 6-43
Central part wiring
Fix the limit switch with the upper tie-wrap on the metal support.
Also fix the wirings and the TEFLON tube with the lower tie-wrap on the metal
support. Place the tie-wrap on the rubber coating part overlapping the wirings
or at the blue coating.
FIGURE 6-44
Correct lengths
The wiring is composed of six ultra-flexible wires. They are supplied – on the
level sensor board side – already crimped but not yet inserted in the MOLEX 8
pins connector case.
The wires must be passed through the rectangular duct without the connector
case.
They must be routed carefully and inserted into the MOLEX connector. Swap-
ping the wires can damage the boards. Short-circuiting the wires will also da-
mage the boards.
FIGURE 6-45
MOLEX connector
The pins 1 to 3 are used for the SN level sensor and the pins 6 to 8 are used for
the vertical shock detection.
The pin numbers of the MINIFIT connector are printed on the back side as follow-
ing:
118
Recommended:
Go to the TE-OS-F5 Washings menu. The status of the floats is displayed in the
upper right corner of the screen.
Normal status:
Not recommended:
Go to the TE-OS-F1-F5 Display Inputs menu. The status of the floats is displayed
in a different way in the upper right corner of the screen.
Normal status:
Each float contains a toroidal magnet in one side of the white body. The magnet
has a specific position when it belongs to a water tank (normally open) and must
be mounted in the opposite way when it belongs to a waste tanks (normally
closed).
FIGURE 6-46
Lock washer reassembly
FIGURE 6-47
Float body inversion
Disassemble the needles group of the wash station (two screws A and B).
FIGURE 6-48
Wash station needle group
replacement
120
Disassemble the black reaction rotor cover (three screws D, E and F) and remove
it.
FIGURE 6-49
Reaction rotor cover replacement
FIGURE 6-50
Unscrew the reaction carousel (six screws A, B, C, D, E and F) and remove it.
Reaction carousel replacement
Rotate the reaction rotor support until the notch is positioned over the optical
preamplifier cover, then unscrew the optical preamplifier cover (2 screws) from
the incubator.
FIGURE 6-51
Positioning of the optical pream-
plifier cover
Push up the optical preamplifier cable (red wires with black sleeve) with the
hand in such a way that the optical preamplifier board is lifted up out of the in-
cubator.
FIGURE 6-52
Pushing up the optical preampli-
fier cable
FIGURE 6-53
Optical preamplifier board disas-
sembly
FIGURE 6-54
Optical preamplifier cable disas-
sembly
7 FIRMWARE UPDATING
7.1.1 DOWNLOAD
6. Go to 'File' and select 'Save' or 'Save as…' to open the Windows dialog menu
fo storing a file. You can either store the parameters as an ordinary text file
with the extension '.PAL' or as a tagged file with the extension '.XML'.
FIGURE 7-2
PAL type download file example
FIGURE 7-3
XML type download file example
7.1.2 UPLOAD
The upload of the saved system parameters works in the opposite way.
FIGURE 7-4
Upload overwrite warning
FIGURE 7-5
Special USB cable
1. Open the front cover of the HS300SR and remove the boards cover.
2. Connect the special USB cable to the CPU (connector on the left marked
"USB" on the picture).
3. Connect the special USB cable to an USB port of the computer.
7.4.3 UPGRADE
8 HI SOFTWARE INSTALLATION
8.1 Settings
8.2.1 LOG ON
To install the software you need to log on your computer as Windows admini-
strator.
If the new software is on an USB stick, just insert the stick and the installation
will normally start automatically. If not, execute the "HiSetup.exe" program.
1. Only if the old installation has not been correctly uninstalled, a new window
will ask you to remove the old installation. In this case, confirm that the old
installation must be removed, then press the "CLOSE" button and restart the
installation.
2. For the "FTDI CDM - USB-RS232" drivers installation, press "INSTALL". A DOS
command window will open during this installation.
3. MICROSOFT LICENSE AGREEMENT window. (If the Microsoft Framework
4.6.2 tools are already present in your computer, this step can be skipped by
the installer). Select the "I agree" option 3 times, first time for Microsoft Fra-
mework, second time for Microsoft Report Viewer. In this case, the system
will ask you to reboot the computer. Accept and wait the computer restarts
and enters again in the installer.
4. HI (HUMAN INTERFACE) LICENSE AGREEMENT window. Select the "I agree"
option and press the "NEXT" button.
5. Select the "COMPLETE INSTALLATION" option.
6. CONFIRM INSTALLATION window. Press "NEXT" button. The software is in-
stalled in a few seconds.
7. INSTALLATION COMPLETE window. Press the "CLOSE" button.
8. The installer creates a data folder named "Hi" with only void files. If you want
to keep your old data, you have to copy the old "Hi" data folder. Follow the
steps described in the next chapter. Else refer to your distributor for your de-
fault data folder. In the installation USB stick a reference default data folder
is supplied.
9. An "hi" icon is now on the desktop to start the program.
FIGURE 8-1
hi software logo and icon
10. The first time you start the program, you must select the COM port that will
be used for the communication with the instrument.
11. Since the internal method structure could have been changed, log on as Ad-
ministrator or Installer and execute the "METHODS UPLOAD" to synchronize
the instrument with the PC.
To maintain the application data from an old installation, copy the old "Hi" fol-
! The ProgramData folder and
all subfolders are hidden fol-
ders by default. To view them,
der and paste it in the following folder for Windows 10 or Windows 7: "C:\Pro- you must go to "Windows Control
gramData". Panel > Folder Options > View"
Once copied, rename your new data folder to "Hi" if the old folder was renamed. and set the "View Hidden Files"
option.
To avoid the risk of data loss, it is anyway safe to backup periodically the "HI"
data folder.
DEFAULT METHODS: In the installation USB stick there is a set of files that will
permit the automatic generation of default data sets at the first run of the soft-
ware.
When the Microsoft Framework 4.6.2 is installed, there is full compatibility with
Windows 7, 8, 8.1, and 10.
Note that this version of the framework is not compatible with Windows XP or
previous versions of Windows.
Check that the following drivers have been loaded (normally installed by the HI
software installation):
- Microsoft.net Framework 4.6.2
- Microsoft Report Viewer 2010 redistributable
- FTDI CDM Driver package 17/02/ 2009 24.16 (this driver reinstalls at every
installation).
132
9 TROUBLESHOOTING
FIGURE 9-1
"Not connected" display and
"Connect" icon
Press the "Connect" icon. If it doesn't connect:
1. Check if power is ON (red lamp on the main switch and orange lamp on the
reagents cooling switch).
2. Switch the instrument OFF, wait five seconds, and switch ON again. At po-
wer-on, the instrument must shake the sample plate three times to show
it is working. If it doesn't shake or shake only twice, contact the service. No
shake means that the CPU or the power board 0 are not working. Try again
to connect. If it doesn't connect, continue with step 3.
3. Verify if the USB cable between the computer and the instrument is correctly
inserted. Try again to connect. If it doesn't connect, continue with step 4.
4. Verify the USB port. Exit from the software. Enter the Windows Control Panel
> System > Device Manager. Unplug and insert again the USB cable from the
instrument. Verify that the computer correctly detects the COM port (Ports
COM & LPT > USB serial port (COMxx)). If not, check the USB cable and the
USB port on the computer. Replace the USB-to-RS232 converter to check if
the problem is in the internal converter board. If the USB port is detected
correctly by the PC, note which is the detected COM port and continue with
step 5.
5. Check the internal serial port. Start the EdifTerminal_2_3_5.exe program. Se-
lect in the upper left corner the COM port detected in the previous step. Click at
the center of the window to focus on it. Shut the instrument OFF and then ON.
Wait for the three shakings of the sample plate and for waste pumps movement.
134
The instrument motors are OFF and is displayed on the screen as follow:
Press ENTER two times. The instrument motors go ON (hold) and a must ap-
pear on the screen, after the characters already appeared, as follow:
There can be several reasons that can cause bad precision in readings:
1. Spray a contact cleaner on all the lamp and optical contacts, in the front and
in the back.
2. The lamp is possibly defective or close to end-of-life.
3. The optical signal cable from the optical preamplifier to the power board 0.
4. The optical preamplifier.
See next chapter "9.3. Autozero has bad repeatability, unstable optical signal"
for more detailed troubleshooting.
The imprecision in a single filter can be caused by a loose filter locking ring.
Check if the silicone tube is well inserted in the pinch valve and that it is not
worn.
Check that all the pipetting TEFLON fittings are well tight.
136
Incorrect vertical positioning during the dispensation and the mixing in the re-
action rotor (053) SN1DS2 and (369) SN2DS2 quote.
Execute the TE-OS-F5-F5-1 Test Pumps function.
At the end of the test the program will eventually suggest to update the (053)
SN1DS2 and/or (369) SN2DS2 mixing quote.
Go to TE-OS-F4 Optical Reading Test menu, wait that the instrument is well
warmed-up and do the following tests:
1. Select filter 0 (dARK) and disconnect the optical cable from the Power Board
0.
The A/D signal on the screen must be close to 20 and stable.
If not, there is a problem on the Power Board 0 or on the optical preamplifier/
multiplexer.
2. Re-connect the optical cable on the Power Board 0 and disconnect the opti-
cal preamplifier.
The A/D signal must be close to 20 and stable.
If not, there is a problem on the optical cable.
4. In the TE-OS-F4 Optical Reading Test menu, select filter 1 (340 nm).
The A/D signal must not oscillate more than 3.0 mV (30 A/D units).
If not, it's possibly a problem of the lamp either the lamp contacts or optical
preamplifier contacts.
Clean all contacts with a contact cleaner.
5. Select filter 2 and then again filter 1. Then filter 3 and again filter 1, etc.
Wait 10 seconds. The A/D signal on filter 1 must come back to the original
value within 4 mV (40 A/D units).
If not, there is a problem in the filter wheel positioning.
Check in the TE-OS-F4-F5-2 Autozero Test function the filter wheel offset.
If the filter wheel belt is well tightened and it is not damaged (TE-OS-F3-F4
OP+FS Home function), it can happen that the belt doesn't look damaged
but it doesn't work well anyway. If you are in doubt, replace the filter wheel
belt anyway.
Involved devices:
2. The displayed temperature is correctly 38 ±0.2°C, but the reaction rotor tem-
perature is wrong.
The temperature measurement is not adjusted. Increase or decrease the pa-
rameter (285) OPTCAL.(10 units are 1 °Celsius).
3. The displayed temperature and the reaction rotor are at room temperature
and PWM R = 70.
The heating is not working. Check the voltage on the resistance connector:
138
4. The displayed temperature is close to zero and the reaction rotor tempera-
ture is high:
The Power Board 0 is not working or the temperature sensor is short-circui-
ted.
It is important to check and find out in which way the cuvettes are dirty, because
it is very important to understand the cause.
The absorbance threshold for the cuvette to be considered clean is 120 mAbs at
340 nm. Typically the absorbance for a new cuvette at 340 nm is lower than 60
mAbs, and for the other wavelengths is lower than 35 mAbs.
If the spectrum is very high at 340 nm and low at the other wavelengths, the
cuvette looks clean to the eye, but it is anyway excluded.
High absorbances on all the wavelengths are typical for scratches or uncolored
(white/grey) deposits.
- If there are internal deposits, the washing is poor. If the deposit is coloured,
you can try to understand which reagents cannot be washed.
- If there are internal vertical scratches, the wash needle(s) touches/touch the
inner side of the reaction cuvette.
- If there are external horizontal scratches, the reaction cuvette touches so-
mewhere and is scratched during the rotation.
- If there are external "clouds", the vacuum pump is not working well or the
operator made some mistake and water overflowed outside. The "cloud" is
the solid deposit that remained on the outer side of the cuvettes after the
water is evaporated. In this case the two external sides of the cuvettes can
be cleaned with a soft cloth.
- If the area is highly humid, the two external optical faces of the cuvettes can
become dirty after some months. Take off the cuvettes rotor (Refer to "6.4.
Reaction rotor replacement ") and pass a soft optical cloth on the two optical
faces of the 80 cuvettes mounted on the rotor.
If the Aspiration test failed, the membrane valves of the vacuum pump or the
entire pump must be replaced.
If the Wash Station Dispensation is low, the peristaltic pumps timers must be
increased. This can be done manually, increasing the (160) DIWSH1 to (164) DI-
WSH5 parameters in the following way:
140
Example:
- Pump dispensation = 300
- Old value of DIWSH1 = 50
- New value of DIWSH1 = 50 x 400 / 300 = 67
The DIWSHx parameters range is 45 to 80. If the parameter must be set higher
than 80, the pumps head must be replaced.
The correction of the pump timers can also be made automatically executing
the TE-OS-F5-F5-5 Adjust Pumps procedure.
To eliminate protein and lipid deposits in the reaction cuvettes, put a reagent
bottle with 26 ml of NaOH solution at 4% in reagent position 1 and start the
Special Wash Procedure (HI – Maintenance – Special cuvette wash).
At the end, a Startup procedure is executed.
To eliminate salt deposits in the reaction cuvettes, put a reagent bottle with 26
ml of HCl solution at 5% in reagent position 1 and start the Special Wash Proce-
dure (HI – Maintenance – Special cuvette wash).
At the end, a Startup procedure is executed.
2. Check if there has been a liquid overflow from the wash station. The water
goes inside the incubator, evaporates and condenses on the walls of the cu-
vettes, increasing suddenly the O.D. of all the cuvettes.
When the condensation vanishes, the O.D. returns back to the right levels.
3. If the problem persists, verify if the cuvettes O.D. is really bad. Go to the TE-
OS-F4 Optical Reading Test menu in and follow the instructions below:
Please note:
The difference value between the filter 340 nm in real time and the filter 340 nm
on the Autozero must be lower than 700 units. If this difference is higher, there
is some discrepancy on the optical path (lamp, fine offset, Adjust FS offset).
If the droplet is formed at the end of the needle tip, there can be several causes.
Most common causes are:
If the drop hangs on a higher position on the side of the needle, it has probably
been left there by the jet of the external wash nozzle.
Clean the needle with alcohol to ease the drop fall.
You can reduce the water jet speed with the parameter (188)°EXWELD, so that
the jet hits the needle in a descending parabola, allowing a better leaching of
the droplet. Try to reduce the (188)°EXWELD parameter by 5 or 10 units.
If the external wash nozzle splashes the needle at the end of the internal needle
wash, the delay for the valve commutation (186)°ASWELD is too low (increase it
by 10-30 units) or the Y tubing connected to the well dispensation pump, placed
on the back side of the hydraulic panel, is too resilient and balloons. Eventually
replace the affected silicon tube.
It happens more or less frequently that some reagent bottles are detected as
completely full (green color) even if they are not.
At the same time it happens that the needle sometimes fails to pipet the sample
or reagent 2 (the results are null) or fails to pipet the reagent 1 (the results are
aberrant).
Most probably the sensor level wiring of the sampling arm is broken.
Open the sampling arm cover and check the level sensor LED during the execu-
tion.
If the red LED blinks when the arm is still in a higher position, before having
detected the liquid level, the level sensor board wiring is broken and must be
replaced.
The "Needle shock detector is stuck" message is issued when the needle is in an
upper position and begins to move downwards.
If a shock signal is detected before starting the movement, the shock sensor is
not working properly and the error message is issued.
- The needle is bent or the white cap op top of the needle is too tight. In these
cases the needle doesn't slide well and it can remain in the upper "shock"
position.
- The optical sensor of the needle shock is broken. Replace the level sensor
board.
- If the level sensor wiring is broken and the white wire of the shock signal is
open, the system detects this as a shock situation and the alarm is issued.
Replace the level sensor board wiring. (Refer to "6.13. Level sensor wiring re-
placement".)
The vacuum pump error can be generated by several different causes. The first
two causes are hydraulic problems. The last two are simply problems of bad me-
chanical calibration of the sampling needle in the reaction cuvettes.
1. The vacuum pump 1 (connected to the wash station needles) is not working
properly. The most common cause is that the pump has aspirated some dirt
which remains on the input valve disk disturbing the aspiration. Alterna-
tively the valve disks could be worn and must be replaced (Refer to "6.8. Va-
cuum pump membrane valves replacement").
2. One (or more) wash station needles are clogged and must be cleaned and
unclogged with a metal wire.
3. The sampling needles rub on the side of the reaction cuvette when it checks
the level (in the OP dispensation position).
144
At the end of the pumps test, if the (053) SN1DS2 and (369) SN2DS2 quotes are
not to the target this pumps test will allow you to calibrate the dispensation
vertical position (053) SN1DS2 and (369) SN2DS2 (to avoid problem 4). Duri-
ng the needle dispensation, you can check through the dispensation hole if the
needles are rubbing on the side of the reaction cuvette (to avoid problem 3).
It allows also to understand which kind of problem can have the vacuum pump.
If one needle remains with a very high level of liquid, it means that this needle
is clogged (problem 2).
If all the needles are high, the vacuum pump is not working (problem 1).
When there are several inconsistent tests we could have three different pro-
blems, mechanical, hydraulics and electronics. It is very important to immedia-
tely understand how to find the solution.
FIGURE 9-2
How to troubleshoot
Perform the Autozero test (TE-OS-F4-F5-2). Here it is possible to check the lamp,
the optical preamplifier, OP, FS and the complete reading unit.
Perform the OP+FS home test (TE-OS-F3-F4). Here it is possible to check the OP
and FS movements. OP and FS are the most important units to achieve good
readings.
Instrument reads through the cuvettes more clear than through the gaps bet-
ween the cuvettes.
2. Try to reduce the OP speeds (125) OPVEL and (126) OPVELZ) or try to adjust
the belt tension or modify manually the (022) OPOFRD parameter 1 unit.
The difference values between the 340 nm filter in real time and the 340 nm fil-
ter on Autozero must be lower than 700 units. If this difference is higher means
that there is some discrepancy on the optical path (lamp, fine OP offset, adjust-
ment of the FS offset).
One or more filter values are lower than 5,000 mV or in overflow. You can ex-
tend the range, in emergency case, by decreasing the parameter (352) MINAUZ.
1. Verify in TE-OS-F4 Optical Reading Test menu if the filter values are good in
real time or if the values in real time are also lower.
3. If in TE-OS-F4 Optical Reading Test menu the autozero value is really lower
than the range, verify the optical preamplifier pins and the lamp pins. If the
problem persists, replace the lamp.
The difference values between the 340 nm filter in real time and the 340 nm fil-
ter on Autozero must be lower than 700 units. If this difference is higher means
that there is some discrepancy on the optical path (lamp, fine OP offset, adjust-
ment of the FS offset).
9.13 Spikes
There are one or more reading points that are very close to zero or in positive or
negative overflow (unusable result).
Check all optical wiring pins and cables, perform the FS tests or try to adjust
the FS belt tension, try to reduce the FS speed.
Wash station needles or pipe cleaner could hit on the cuvettes or the wash sta-
tion down limit switch could be broken.
1. If the down limit switch is okay, check the wash station calibration in rows 17
and 18 (TE-OS-F2 Mechanical Calibration).
2. If the calibration is okay, reduce the OP speed or adjust the OP belt tension.
148
10 MAINTENANCE
The instrument requires daily, weekly, monthly and special maintenance as de-
scribed in the User's Manual.
If the firmware and the software of the instrument are significantly old, proceed
to update the firmware and the HI software. Check the release on the "Info" but-
ton on the lower left corner of the software screen.
FIGURE 10-1
Instrument information
10.2 Cover
Check the lamp life. Lamp life ranges between 1,000 and 2,000 hours. If the
lamp has more than 1,000 hours of work and a new maintenance check-up is
not scheduled, you can consider changing the lamp (remember to reset the
lamp timer).
150
Check if the wash station needles are clogged, especially the first long needle
on the right.
Eventually unclog them with a probe or a copper wire. The deposit on the need-
les depends on the type of reagents employed. Clean them with the special
wash solution (basic) to remove protein or lipids deposits. Use an acid solution
to remove salt deposits.
Check if all the needles are not bent and are well centered in the cuvettes (TE-
OS-F2-17).
Check if the dryer needle tip is not worn or broken or rotated.
Replace all the damaged needles.
Check the lamp connectors (in the back and on the power board) and spray the
contacts with a contact cleaner.
In TE-OS-F4, filter 1 selected, move the lamp wires and connectors. The reading
value must not change for more than a few A/D units. Else, check the lamp wires.
Check the arm wiring: wires can be stressed or, inside the isolation, the copper
could be broken. The continuity of the signal is then precarious. If there is
any sign of wear or breakage, replace the wiring (level sensor board wiring
(16890/112-1), optical limit switch wiring (16890/113-1) and the small motor
wiring (16890/122-1)).
10.8 Diluter
On the diluter (16890/94), replace the white & orange gasket (16890/95) inside
the diluter head – see "6.11 Diluter sealing gasket replacement").
10.9 Pumps
If the flow rate of any pump is lower than 500 μl/s or if the pump head work
time is higher than 30 to 50 hours, replace the pump's head, reset the pump's
work timer and then execute the pumps adjustment procedure (TE-OS-F5-F5-5).
On the pumps test report (TE-OS-F5-F5-1), check the vacuum pump flow rate.
If lower than 2,000 μl/s, do as follow:
- First check if any aspiration needle (the longer ones) is clogged and eventu-
ally unclog them.
- Else replace the pumps membrane valves (see "6.8 Vacuum pump memb-
rane valves replacement") or the vacuum pumps (see "6.7 Vacuum pump
replacement").
- Else again, check if the hydraulic circuit waste is clogged.
If necessary, clean it with a sterilizing solution and then rinse abundantly, wipe
and dry.
If it is a 20 + 20 positions sample plate (16890/11), check if the external tubes
clips are well inserted and not damaged.
Check on the reaction rotor the cuvettes absorbances. If absorbances are typi-
cally higher than 800, take the reaction rotor off and clean gently the two exter-
nal faces of the cuvettes wiping them with a lens cleaning cloth. Pay attention
to avoid dust deposits on the two faces. After the cleaning, execute a wash of all
the cuvettes (TE-OS-F5-F2, N. of washes: 80).
Replace cuvettes which show an O.D. higher than 1,000 units (= 100 mAbs). See
"6.3 Single reaction cell replacement".
After these two operations it is necessary to execute the startup. There is a fast
version of the startup in the Washings menu.
The following menus must be called from the TERMINAL SERVICE menu to exe-
cute the test procedures:
Check all the mechanical calibrations, especially the sampling needle dispensa-
tion quote (053) SN1DS2 and (369) SN2DS2, the wash station down quote (115)
WS_ASP, the sampling needle wash in the well drop (045) SN1WSH and (364)
SN2WSH, the sampling needle in the reagent bottles and in the sample tubes
and cups.
Verify the belts and the mechanics executing the mechanical check tests, espe-
cially TE-OS-F3-F4.
If FS or OP fails, tighten eventually the belt(s).
TE-OS-F5-F5-2: Fill the hydraulics if the tubes are empty, else the following tests
will be faulty.
TE-OS-F5-F5-1: Pumps test, adjustment of the reaction cuvettes level check and
! Instead of TE-OS-F5-F5-2 you
can also run the hidden com-
mand TE-OS-F5-F5-6 which re-
of the needles dispensation quote. ports more data.
TE-OS-F5-F5-5: Adjust pumps.
Run the 2 μl pipetting test procedure with E124 (O.D. at 505 nm: 30 to 50 abs, 1
ml in sample tube #1).
154
OK1,1
XM6
To report:
WS1
XM0
The CV% must be lower than 2.5 to 3 %. Factory value is lower than 2 %
12 ERROR CODES
13.1 IP motor
13.2 OP motor
13.3 FS motor
13.4 SN motor
13.5 DS motor
13.6 RN motor
13.7 WS motor
13.9 Pumps
13.11 Barcode
13.13 Temperatures
13.15 Readings
13.18 Miscellaneous