Professional Documents
Culture Documents
2010 - Validation of A New Device To Measure Postsurgical Scar Adherence
2010 - Validation of A New Device To Measure Postsurgical Scar Adherence
Scar Adherence
Giorgio Ferriero, Stefano Vercelli, Ludovit Salgovic,
Valeria Stissi and Francesco Sartorio
PHYS THER. 2010; 90:776-783.
Originally published online March 11, 2010
doi: 10.2522/ptj.20090048
The online version of this article, along with updated information and services, can be
found online at: http://ptjournal.apta.org/content/90/5/776
S
carring after surgery can lead to
a wide range of disorders such
as pain, movement limitation,
functional impairment, and aesthetic
or psychological disturbance.1,2 The
assessment of pathological postsur-
gical scars is crucial for planning
their treatment.3– 6 It usually includes
evaluation of physical characteristics
(eg, height, pliability, relief, adhesion),
cosmetic appearance (color, cosmetic
defects), and the patient’s symptoms
(pain, itching). In particular, scar ad-
herence (defined as the failure of the
tissues to successfully establish inde-
pendent layering)7 may produce sev-
eral clinical problems, limiting range
of motion and muscle strength (force-
Figure 1.
generating capacity) and altering the
The Adheremeter. The diameter of the largest concentric ring is 28 mm, and the
local proprioceptive input.6,7 external edge of the device is 17.5 mm from the center.
Figure 2.
Scar adherence (marked with a black fine-point pen) in original position (O) at rest (left) and at maximal caudal excursion (C) (right)
when stretched with maximal force within a comfort range for the patient. Red arrow shows stretching direction. In this example,
maximal caudal excursion of the adherence (from O to C) is 3 mm.
the contralateral anatomic position landmark returned to the Adhereme- rostro-caudal) and 1 represents com-
of the adherence and marked them ter’s center and, if not, repeated the pletely normal scar mobility. In both
with a fine-line pen. Both surfaces measurement. Markers on the skin indexes, an increase of values means
were cleaned. In linear scars, the were cleaned at the end of each mea- a better scar condition (ie, a higher
rater reported on the patient record surement. The whole procedure surface mobility for the SMA and a
the position of the worst adherent generally took a few minutes per scar surface mobility approaching
point by measuring its distance from landmark. that of normal skin for the AS).
the 2 extremities of the scar. The
Adheremeter was positioned so that Data Analysis
the rings were centered on the land- The 4 measurements (ie, caudal, ros-
mark. Skin was relaxed, and nearby tral, and the 2 side maximal land-
joints were in a loose-packed posi- mark excursions from the rest posi-
tion. The rater held the device in the tion), taken both for the scar and for
hand, supporting the hand on the the normal contralateral skin, were
patient’s body in such a way that used to obtain a couple of indexes of
there was no contact between the surface mobility: the adherence’s
Adheremeter and the patient’s skin. surface mobility index for the scar
The other thumb was positioned (SMA) and the surface mobility index
close to the external edge of the de- for the normal contralateral skin
vice (17.5 mm from the center) (SMN). The score of each index of
(Fig. 2). Before stretching the skin surface mobility was obtained by cal-
with the thumb with maximal force culating the area of the quadrilateral
within a comfortable range for the whose diagonals, which are orthog- Figure 3.
patient, the rater said to the patient, onal to each other, are the side-to- Graphic representation of the surface mo-
“Now, I’m beginning to stretch the side and rostro-caudal landmark max- bility index. O is the original position of
skin; if you feel any discomfort, tell imal excursions (Fig. 3). Then, the the evaluation point, S1 and S2 represent
the 2 maximal lateral excursions, and C
me immediately.” Traction was ap- SMA was compared with the SMN, and R represent the maximal caudal and
plied centrifugally in 4 directions: thus giving an index of adherence rostral excursions. Because the diagonals,
caudal, rostral, right side, and left severity (AS). The AS estimates the S1S2 (side-to-side, red) and RC (rostro-
side. For every traction, the rater ratio between the SMA and the SMN: caudal, blue), intersect at right angles, the
read on the Adheremeter the posi- AS!SMA/SMN (Fig. 4). Its values thus area of the quadrilateral (yellow) is com-
puted as: (S1S2 # RC)/2. In this example,
tion of the landmark at the maximal calculated range from 0 to 1, where S1S2!7$5!12 mm, RC!8$2!10 mm,
excursion. Once the tension was re- 0 represents scar immobility in at and, consequently, the surface mobility
leased, the rater verified that the least one diagonal (side-to-side or index is scored as 12 # 10/2!60 mm2.
mined on the basis of the pilot study 3. The standard error of measure-
expecting to obtain ICC values of ment (SEM) at T1, reflecting the
about .90, with a 95% confidence extent of expected errors in dif-
interval (CI) close to .2.16 The in- ferent raters’ scores, computed
trarater reliability of the SMN (T1 ver- as follows: SEM!SD # &1 " R,
sus T2) was calculated for both rat- where SD is the standard devia-
ers. Interrater reliability was analyzed tion of test scores and R is the
by comparing the SMA, SMN, and AS test-retest reliability coefficient,
at both T1 and T2 for both raters. which in this study was the ICC18;
and
Given the sample size of 25 pa-
tients15 and the link between scar 4. The minimum detectable change
adherence, pliability, and general in single subjects (MDC), com-
scar healing,5 to provide evidence of puted from the SEM, to indicate
Figure 4. concurrent validity, we tested our the amount of change required to
Graphic representation of the surface mo-
a priori hypothesis, which was to be adequately confident that the
bility index (patient 7) showing differ-
ences between the adherence’s surface find at least a moderate correlation change that has occurred is not
mobility index for the scar (SMA) (patho- (r %.50) between the SMA and the AS attributable to measurement error
logical condition, red quadrilateral) and and both VSS and PL-VSS. Correlation or chance variation. The MDC
the surface mobility index for the normal coefficients (rs) were calculated us- was estimated using a previously
contralateral skin (SMN) (normal skin con-
ing the Spearman rank method, cor- described method (1.96 # SEM
dition, black quadrilateral) at the initial
examination (T1) and the SMA at the end rected for ties. Data were analyzed # &2, where 1.96 is the 2-sided
of treatment (T2) (outcome, blue quadri- using SPSS statistical software.† tabled z value for a 95% CI).18
lateral). The figure clearly shows, in this
patient, an improvement in scar mobility The sensitivity to change (ie, the abil- Results
after the treatment (the blue quadrilateral
ity to detect change in general, re- The mean duration of the rehabilita-
is larger than the red quadrilateral), but
also that maximal rostral excursion did gardless of whether the change was tion intervention was 17 days (inter-
not change. In this example: at T1, clinically relevant) of the SMA and quartile range!12–30 days). No pa-
SMA!10 mm2, SMN!60 mm2, and, con- the AS was determined by: tient reported discomfort during
sequently, AS is scored as 10/60!0.17; measurement with the Adheremeter.
at T2, SMA!12 mm2, SMN!60 mm2,
1. Wilcoxon signed rank tests; Table 2 shows the mean values for
and, consequently, AS is scored as 12/
60!0.20. the SMA and the AS at T1 and T2.
2. The effect size, defined as mean Both scores increased significantly
change score (T2"T1) divided by during the testing period (for all,
the standard deviation of the T1 P'.001). Figure 5 shows the corre-
Intrarater and interrater reliability
(admission) scores (values around lation between the AS values at T1
were calculated by computing the
0.2, 0.5, and 0.8 are considered, and changes that occurred after the
intraclass correlation coefficient
respectively, small, moderate, treatment period, calculated for each
(ICC [2,1]) at T1 and T2. Intraclass
and good)17; patient with the following formula:
correlation coefficient values higher
than .75 were considered good, and [(AS score at T2)"(AS score at T1)].
those above .90 were considered ex- Table 3 shows the mean values for
cellent.15 The sample size of 25 pa- the VSS and the PL-VSS at T1 and T2.
tients assessed by 2 raters was deter- †
SPSS Inc, 233 S Wacker Dr, Chicago, IL 60606.
Table 2.
Mean (SD) Values for the Adherence’s Surface Mobility Index for the Scar (SMA) and the Index of Adherence Severity (AS) at the
Initial Examination (T1) and at the End of Treatment (T2)
Physical Therapist Physical Therapist Student
Index T1 T2 T1 T2
To verify the validity of the Adher- AS .88 (.75, .94) .87 (.72, .94)
The Adheremeter might not be reli- This article was received February 18, 2009, 11 Cleary C, Sanders AK, Nick TG. Reliability
and was accepted December 20, 2010. of the skin compliance device in the as-
able for measuring scars situated in sessment of scar pliability. J Hand Ther.
highly concave or convex anatomi- DOI: 10.2522/ptj.20090048 2007;20:232–237.
cal zones. In the absence of a con- 12 Draaijers LJ, Botman YA, Tempelman FR,
et al. Skin elasticity meter or subjective
tralateral landmark (eg, amputation
References evaluation in scars: a reliability assess-
or scar on the midsagittal axis), we 1 van de Kar AL, Corion LU, Smeulders MJ,
ment. Burns. 2004;30:109 –114.
suggest comparing the adherence et al. Reliable and feasible evaluation of 13 Truong PT, Lee JC, Soer B, et al. Reliability
linear scars by the Patient and Observer and validity testing of the Patient and Ob-
with the nearest healthy skin. Scar Assessment Scale. Plast Reconstr server Scar Assessment Scale in evaluating
Surg. 2005;116:514 –522. linear scars after breast cancer surgery.
Plast Reconstr Surg. 2007;119:487– 494.
Conclusions 2 Moran M, Khan A, Sochart DH, Andrew G.
Evaluation of patient concerns before total 14 Truong PT, Abnousi F, Yong CM, et al.
In our sample, this new method to knee and hip arthroplasty. J Arthroplasty. Standardized assessment of breast cancer
measure adherent scars showed an 2003;18:442– 445. surgical scars integrating the Vancouver
Scar Scale, Short Form McGill Pain Ques-
adequate level of reliability, validity, 3 Roques C. Massage applied to scars. tionnaire, and patients’ perspective. Plast
and sensitivity to change. The Adher- Wound Repair Regen. 2002;10:126 –128. Reconstr Surg. 2005;116:1291–1299.
emeter could be considered a useful 4 Roques C, Téot L. A critical analysis of 15 Portney LG, Watkins MP. Foundations of
measurements used to assess and manage Clinical Research: Applications to Prac-
device for clinicians working with scars. Lower Extremity Wounds. 2007;6: tice. Upper Saddle River, NJ: Prentice Hall
patients with scars. Caution should 249 –253. Health; 2000.
be applied in generalizing the results 5 Idriss N, Maibach HI. Scar assessment 16 Bonett DG. Sample size requirements for
scales: a dermatologic overview. Skin Res estimating intraclass correlations with de-
of this study because further studies Technol. 2009;15:1–5. sired precision. Stat Med. 2002;21:1331–
are needed to confirm our results in 6 Vercelli S, Ferriero G, Sartorio F, et al. 1335.
larger cohorts and for other types of How to assess postsurgical scars? A review 17 Kazis LE, Anderson JJ, Meenan RF. Effect
of outcome measures. Disabil Rehabil. sizes for interpreting changes in health
scars. 2009;31:2055–2063. status. Med Care. 1989;27(suppl 3):S178 –
7 Kobesova A, Morris CE, Lewit K, Safarova S189.
M. Twenty-year-old pathogenic “active” 18 Taylor R, Jayasinghe UW, Koelmeyer L,
Dr Ferriero and Dr Vercelli provided concept/ postsurgical scar: a case study of a patient et al. Reliability and validity of arm volume
idea/project design. All authors provided with persistent right lower quadrant pain. measurements for assessment of lymph-
J Manipulative Phys Ther. 2007;30:234 – edema. Phys Ther. 2006;86:205–214.
writing. Dr Vercelli, Dr Stissi, and Dr Sartorio 238.
provided data collection. Dr Ferriero, Dr Ver- 19 Zhang Y, Goldgof DB, Sarkar S, Tsap LV. A
8 Sutton GS, Bartel MR. Soft-tissue mobiliza- modeling approach for burn scar assess-
celli, and Dr Salgovic provided data analysis. tion techniques for the hand therapist.
Dr Ferriero provided project management ment using natural features and elastic
J Hand Ther. 1994;7:185–192. property. IEEE Trans Med Imaging. 2004;
and institutional liaisons. Dr Salgovic pro- 9 Nedelec B, Shankowsky HA, Tredget EE. 23:1325–1329.
vided consultation (including review of Rating the resolving hypertrophic scar: 20 Copay A, Subach B, Glassman S, et al. Un-
manuscript before submission). The authors comparison of the Vancouver Scar Scale derstanding the minimum clinically impor-
thank Dr Franco Franchignoni for his con- and scar volume. J Burn Care Rehabil. tant difference: a review of concepts and
2000;21:205–212. methods. Spine J. 2007;7:541–546.
tinuing guidance and advice.
10 Silverberg R, Johnson J, Moffat M. The ef- 21 Domholdt E. Rehabilitation Research:
The research reported in this article was un- fects of soft tissue mobilization on the im- Principles and Applications. 3rd ed. Phil-
dertaken in compliance with the Helsinki mature burn scar: results of a pilot study. adelphia, PA: Elsevier Saunders; 2005.
J Burn Care Rehabil. 1996;17:252–259.
Declaration and the international principles
governing research on animals.