Professional Documents
Culture Documents
B Inggris
B Inggris
I. preliminary
Hematology is a branch of health science concerning blood, blood-forming organs, and blood-
related diseases. In general there are two main components are blood, namely blood cells and blood
plasma. Blood cells consist of erythrocytes (red blood cells), leukocytes (white blood cells), and platelets.
Blood plasma consists of about 90% of the remaining water in the form of a solution of protein, glucose,
factor congulation, hormones, enzymes, electrolytes, other substrates, and carbon dioxide.
The goal is to find out the number of blood cells. Like red blood cells or Red Blood Cells (RBC), white
blood cells or White Blood Cells (WBC), platelete and other blood parameters. As well as the average
volume of blood cells can be known.
The principle of measurement and absorption of light due to interactions between light that has a
certain wavelength with the solution or sample in its path. This tool works based on the flow cytometer
principle.
Hematology examination is a routine examination carried out on almost all patients who come to
the hospital. Hematological examination is performed on venous blood samples that have been mixed
with EDTA antichogeneous (ethylene diamine tetraacetic acid) to prevent blood clots.
Hematology analyzer tool requires a patient's blood sample for the purpose of counting the number
of blood cells. This tool has several methods and systems for calculating blood cells, namely:
The Electrial Impedance method has the ability to count 3 types of blood cells only, namely:
The method for detecting a sample is based on physical or chemical differences in the sample. So,
the components in the blood will enter with the help of certain electrodes so that they will enter the
device through small gaps and will be read by a beam of light, then will be read by a detector.
Blood counts using this tool also have limitations, namely when there are abnormal cells, such as
many immature cells in leukemia, sepsis, the cell count is so high that this tool is unable to count them.
Therefore it is also advisable to use manual checks.
The results issued by this tool are usually already through quality control carried out by internal
laboratories, both hospital and clinical laboratory installations.
II. METHOD
Material :
1. EDTA blood
2. Reagents (Lyse, Probe Cleanser, Diluent, EZ Cleanser)
B. Prosedur
1.
A. Observation Results
B. Discussion
In this practicum discussing hematology analyzer, this tool is used to examine complete blood by
automatically counting other blood cells based on the impedance of electric current or beams of light to
the cells being passed.
Hematology analyzer is a device used to measure components in the blood. This tool is the main
instrument used in clinical laboratories. This tool is used in pratama laboratories (at the level of
puskesmas) up to the main laboratory or referral laboratory.
Erythrocytes are the most cells in the blood with normal values of 4-5 million per microliter of women's
blood and 5-6 million per microliter of men's blood. Normal erythrocytes have a volume of about 9 fL (9
femtoliters) and one third of the volume is filled with hemoglobin. There are about 270 million
hemoglobin molecules in each cell, where each molecular cell carries four heme groups. This heme is
what functions to carry oxygen to the body's tissues through the blood.
Leukocytes are cells in the blood that function for the immune system, with a diameter of 7-15 µm. In
normal conditions the amount is 4,000-10,000 per microliter of blood, but in conditions of infection the
number can increase. There are several types of leukocytes in peripheral blood with different forms of
function, namely basophils, eosinophils, stem ethrophils, neutrophil segments, lymphocytes, and
monocytes. Under abnormal conditions (eg leukemia), young leukocytes (such as lymphoblasts,
myeloblasts, promyelocytes, myelocytes, and metamyelocytes) in the peripheral blood, whereas these
cells should only be found in the spinal cord.
Platelets are nucleated cells with irregular shapes and measuring 2-3 µm. The normal value of platelets
in the blood is 15,000-450,000 per microliter. Platelets play a role in the process of hemostasis (blood
clotting) at the cellular level.
Time efficiency, faster in the examination only takes 2-3 minutes compared to the manual
method. Because the manual method must be calculated first such as platelets, leukocytes, and
so forth.
Samples, hematology analyzer examination uses a small sample while using manually required
more blood samples.
Accuracy of results, the results of this tool have been through quality control carried out by the
internal laboratory, both at the hospital institution or the Pratama clinical laboratory.
This tool cannot count abnormal cells, so this examination does not always go well. As in cell count
counts, it could be low because there are countless cells because the cells have an abnormal shape, for
example in platelet or leukocyte count results.
2. Incorrect specimen storage and examination time is delayed for too long so that changes in
morphology of blood cells occur.
4. Run out of lyse reagent so that all cells are not destroyed when measuring certain cells.
4. Make sure the tool has been warmed up and a background has been made.
5. check the condition of the volume and packaging of the reagents diluents, lyse and rinse.
3. Buttons (menus, print, page up, page down, count, enter, etc.), for operating tools.
6. Sample probe, for entering the sample into the hematology analyzer.
Calibration of this tool every 6 months using Mindray BC2300 series calibrators. As for the
treatment, namely:
1. The temperature must be 18-20ºC, try to use the air conditioner in the room.
IV. CONCLUSION
Hematology analyzer is a tool for blood tests or aids for diagnosing diseases that can be checked
through sarah cells.
V. REFERENCES
Ginting Br, Raheliya. 2016. The System for Detecting Problems of Hematology Analyzer Damage Using
the Forward Channing Method. Medan: Independent Polytechnic Magazine Bina Putera.
Husni Faruk, Zulfikar. 2018. Analysis of Blood Lysis of the Thrombocyte Value Using the Electrical
Impedance Method. Semarang: Muhammadiyah University Semarang.
Okttiyani, Neni. 2017. Accuracy Calculate the Number of Erythrocytes in the Manual Method and the
Automatic Method. Banjarmasin: Banjarmasin Health Polytechnic.