Table 2.2.2.-6. - Reference solutions R the other is the reference electrode (e.g. a silver-silver chloride
electrode). They are often combined as 1 compact electrode,
Volumes in millilitres together with a temperature probe.
Reference Standard solution R Hydrochloric acid
solution (10 g/L HCl) Apparatus. The measuring apparatus is usually a voltmeter
R1 100.0 0.0 with an input resistance at least 100 times that of the electrodes
R2 75.0 25.0 used. It is normally graduated in pH units and has a sensitivity
such that discrimination of at least 0.05 pH unit or at least
R3 50.0 50.0 0.003 V may be achieved.
R4 37.5 62.5
Recent pH meters are microprocessor-controlled and are
R5 25.0 75.0
operated using the manufacturer’s firmware or software,
R6 12.5 87.5 according to given instructions.
R7 5.0 95.0
Management of electrodes. The electrodes are stored
Storage appropriately and according to the manufacturer’s
recommendations (e.g. in an electrolyte solution or a suitable
For Method I, the reference solutions may be stored in sealed storage solution). Before measurement, the electrodes are
tubes of colourless, transparent, neutral glass of 12 mm visually checked. Refillable electrodes are checked for the
external diameter, protected from light. absence of air bubbles in the glass bulb and to ensure that
the inner electrolyte solution level is satisfactory. The filling
For Method II, prepare the reference solutions immediately orifice has to remain open during the measurement. It is also
before use from the standard solutions. recommended that the diaphragm of the reference electrode is
checked. Before first use, or if the electrode has been stored
out of electrolyte solution, it is usually necessary to condition
it, according to the recommendations of the manufacturer.
If pH stabilisation is too slow (i.e. a long response time), or
a zero point shift, reduced slope or any other difficulties in
calibration are observed, the electrode will probably need to
07/2016:20203 be cleaned or replaced. The cleaning is performed depending
on the type of sample and as prescribed in the manufacturer’s
manual. Regular cleaning is recommended.
Potassium dihydrogen phosphate 0.0087 M + Disodium Strongly alkaline > 10 Phenolphthalein paper R
hydrogen phosphate 0.0303 M. Dissolve 1.18 g of KH2PO4
and 4.30 g of Na2HPO4, both previously dried for 2 h at Thymol blue solution R
120 ± 2 °C, in carbon dioxide-free water R and dilute to
Neutral 6–8 Methyl red solution R
1000.0 mL with the same solvent.
Disodium tetraborate 0.01 M. Dissolve 3.80 g of Phenol red solution R
Na2B4O7,10H2O in carbon dioxide-free water R and dilute Acid <6 Methyl red solution R
to 1000.0 mL with the same solvent. Store protected from
atmospheric carbon dioxide. Bromothymol blue solution R1
Sodium carbonate 0.025 M + Sodium hydrogen carbonate Slightly acid 4–6 Methyl red solution R
0.025 M. Dissolve 2.64 g of Na2CO3 and 2.09 g of NaHCO3
in carbon dioxide-free water R and dilute to 1000.0 mL with Bromocresol green solution R
the same solvent. Store protected from atmospheric carbon
dioxide. Strongly acid <4 Congo red paper R