Professional Documents
Culture Documents
Table of Contents
PREFACE
IMPORTANT IMPORTANT-1
PRECAUTION ON CE CONFORMITY MARKING IMPORTANT-1
1. ELECTROMAGNETIC COMPATIBILITY REQUIREMENT IMPORTANT-1
2. SAFETY REQUIREMENT IMPORTANT-1
INFORMATION FOR USERS ON WEEE
(ONLY FOR EU COUNTRIES) IMPORTANT-2
INFORMATION ON DISPOSAL FOR USERS IMPORTANT-2
1. In the European Union IMPORTANT-2
2. In other Countries outside the EU IMPORTANT-2
WARRANTY ON PRODUCT IMPORTANT-3
INSTALLATION, RELOCATION AND
AFTER-SALE TECHNICAL SERVICE IMPORTANT-6
CUSTOMER TRAINING IMPORTANT-6
OTHER PRECAUTIONS IMPORTANT-6
1. INSTALLATION 1-1
1.1 Unpacking 1-1
1.2 Installation Conditions 1-2
1.2.1 Power Requirements 1-2
1.2.2 Installation Place 1-2
1.3 Installation Environment 1-3
1.4 Check of Contents 1-4
1.5 Assembly 1-4
1.6 Antiearthquake Measure 1-5
1.7 Check of Power Voltage and Fuse 1-6
-1 -
Table of Contents (Cont'd)
2. FUNCTION 2- 1
2.1 Outline 2- 1
2.2 Instrument Appearance 2- 1
2.2.1 Sample Compartment 2- 2
2.2.2 Right Side Panel of Spectrophotometer 2- 2
2.3 Principle of Operation 2- 3
2.3.1 Optical System 2- 3
2.3.2 Signal Processing & Control System 2- 4
2.3.3 Operating Panel 2- 5
2.3.4 Display 2- 6
2.3.5 Auto Calibration and Self-Diagnosis Functions 2- 6
2.4 Specifications 2- 7
3. OPERATION 3- 1
3.1 Booting the Equipment 3- 1
3.1.1 Initialization 3- 1
3.1.2 Main Menu 3- 2
3.1.3 Operation on the Main Menu 3- 2
3.2 Photometry 3- 3
3.2.1 Measurement of Transmittance (Absorbance) 3- 3
3.2.2 Concentration Measurement (Creation of Calibration Curve) 3- 9
3.2.3 Concentration Measurement (Entering Coefficient) 3-24
3.2.4 Measurement of Ratio 3-35
3.3 Wavelength Scan 3-42
3.4 Time Scan 3-61
3.5 Measurement in Enlarged Display Screen 3-76
3.6 Instrument Setup 3-78
3.6.1 Wavelength Calibration 3-79
3.6.2 0%T 3-80
3.6.3 <LampTime> Display and Reset the Lighting Time of Lamp 3-82
3.6.4 Validation Function of <GLP/GMP(Auto)> and
<GLP/GMP(Manual)> 3-84
3.6.5 Printer Setup 3-94
3.6.6 <TimeSetup> 3-95
3.7 Saving Measurement Parameters 3-97
3.8 Loading the Measurement Parameter File 3-98
3.9 List of Error Messages 3-99
Table of Contents (Cont'd)
4. MAINTENANCE 4- 1
4.1 Maintenance and Checkup 4- 1
4.1.1 Purpose of Maintenance 4- 1
4.1.2 Cleaning of Sample Compartment 4- 1
4.1.3 Cleaning of Spectrophotometer Cover 4- 2
4.1.4 Cleaning and Storage of Cell 4- 2
4.1.5 Lubrication 4- 2
4.1.6 Cleaning of Transmission Window Plate of Sample Compartment 4- 3
4.2 0%T Measurement 4- 4
4.3 Performance Check 4- 5
4.3.1 Wavelength Accuracy 4- 5
4.3.2 Wavelength Setting Repeatability 4- 6
4.3.3 Bandpass 4- 7
4.3.4 Noise Level 4- 8
4.3.5 Baseline Stability 4- 9
4.3.6 Baseline Flatness 4-10
4.4 Troubleshooting 4-12
5. REPLACEMENT OF PARTS 5- 1
5.1 General 5- 1
5.2 Consumable and Spare Parts 5- 1
5.3 Replacement of Light Source Lamp 5- 2
5.3.1 Light Source Lamp Replacement Procedure 5- 2
5.3.2 Lamp Turn-On Time Resetting 5- 5
5.3.3 Lamp Adjustment 5- 5
5.4 Storage 5- 9
5.4.1 After Measurement 5- 9
5.4.2 When Not Using the Instrument for a Long Time 5- 9
APPENDIX
APPENDIX 1. ABSORPTION SPECTROPHOTOMETRY APPENDIX- 1
APPENDIX 2. FOR PROPER USE OF SPECTROPHOTOMETER APPENDIX- 2
INDEX INDEX- 1
- in -
1.1
1. INSTALLATION
1.1 Unpacking
After unpacking the shipping crate, carefully take out the spectrophotometer and place it on a
table.
CAUTION
CAUTION on taking out a heavy article
U-1900 is as heavy as about 27 kg. In order to prevent the
spectrophotometer from hurting people when it is taken out, special
attention should be paid and two persons should hold it tight from
both sides when taking it out.
1 -1
1.2 Installation Conditions
Check that the following conditions are satisfied for the installation of the spectrophotometer.
Floor area : Width of 370 mm and depth of 550 mm or more. Space of at least
200 mm should be provided on either side of the main unit.
Work table strength : Flat surface capable of withstanding weight of 27 kg or more.
1 -2
1.3
(a) Atmosphere should be free of acidic or alkaline gas or other gases that may corrode
metal.
(b) Atmosphere should also be free of vapor from organic solvents (such as benzene and
thinner) that may dissolve paint.
(a) Avoid direct sunlight (otherwise optical performance might deteriorate or the housing
might be discolored). Install the instrument as far away from a window as possible.
(b) A strong vibration or shock that can be felt by the human body must not be transmitted
to the instrument (otherwise the fine adjustment mechanism might malfunction).
(c) Avoid the vicinity of a heat generating apparatus such as a gas burner, electric heater
or oven in order to prevent the mainframe cover being heated (beyond 60 °C).
(d) Avoid the vicinity of instruments that generate a strong electric field (such as an
electric welding machine, high frequency furnace or pole transformer).
(f) The instrument should be free from an abrupt variation of power supply voltage (which
constitutes a noise source).
(g) Do not turn on and off frequently electric instruments (such as stirrer or vibrator) that
are not provided with a noise-preventive device and are connected on the same power
line as the spectrophotometer mainframe.
NOTICE: The optical system is very delicate. And the control unit
incorporates high-density electronic circuit components
that function as a computer. The above cautions
should therefore be observed strictly.
1 -3
1.4 Check of Contents
After unpacking, check the contents of delivery against the packing list. If any part is missing or
damaged, or if you have any question, contact your nearest Hitachi High-Technologies
Corporation sales representative.
1.5 Assembly
WARNING
Contact with the power supply voltage may cause an electric shock,
resulting in fatal or serious injury. Do not connect the power cord to
the outlet until the assembly work is finished.
The light source switching lever in the light source compartment is held in place by sponge
material to prevent its moving during transport. Remove the sponge material in the following
procedure.
1 -4
1.6
Two threaded screw holes for a clamp are put both on the right and left sides of the body to
secure the spectrophotometer as an antiearthquake measure.
Use these threaded screw holes to secure the photometer as needed. (Each of the threaded
screw holes is for a 4 mm screw, and the spacing of two holes is 50 mm.) (The user needs to
prepare clamps and screws.)
Secure the personal computer on the laboratory table with retaining bands as needed.
' 1> I I f 1 r
A N V A M ^ X -J. ywj^- A-^.^,,^
¥ f
I I t I •\ -
? uMtfwJ . , fn-.w-rt™-.^rnnnrrrrr*rT^, , L , * Vv ^ *S • rft-jVj-Agj^.-j
1-5
1.7 Check of Power Voltage and Fuse
WARNING
Contact with the power supply voltage may cause an electric shock,
resulting in fatal or serious injury. Before checking the voltage and
fuse, make sure the power cord is disconnected.
First check if the line voltage to be used matches the allowable line voltage indicated on the label
beside the power connector.
.<*
Fuse holder
Now check if the fuse capacity is suitable for the line voltage to be used.
Table 1-1
1 -6
1.8
WARNING
Contact with the power supply voltage may cause an electric shock,
resulting in fatal or serious injury. Make sure the power switch is
OFF before the cable connection, and connect the power cord last.
To AC power source
Printer cable
Printer
Spectrophotometer
main unit
To AC power source
Computer PC
Printer
Spectrophotometer
main unit
1 -7
1.9 Connection of Power Cord and Grounding Wire
(1) Plug the power cord securely into the connector of the mainframe.
(2) When using a plug adapter or table tap, make sure its grounding wire is connected
securely to the grounding terminal.
To power receptacle
Grounding
terminal
Grounding Spectrophotometer
resistance 100 Q
or less
Grounding
wire (green)
Underground
fl-Ht Power cord
°*
300 mm
2 mm
X^
10 mmX
COMMON COMMON
W white
GROUND ..I. Si
/ \
COMMON
HOT HOT GROUND
S black green GROUND
1.10 Power-On
Insert the power cord adapter into a power receptacle and then turn ON the power switch of the
spectrophotometer.
1 -8
2.1
2. FUNCTION
2.1 Outline
The U-1900 is an instrument for photometric analysis in the ultraviolet and visible regions.
Described below are the functions of the instrument.
Printer (Option)
D020422
2-1
2.2 Instrument Appearance
The right side panel of the spectrophotometer contains the items shown below.
2-2
2.3.1
The optical system of the U-1900 is shown in Fig. 2-4. The white light emitted from the light
source is led to the monochromator of Seya-Namioka mount that utilizes a concave diffraction
grating (with grating constant of 1/600 mm, blaze wavelength of 250 nm, and diffraction area of
20 mm x 25 mm), where it is transformed into a monochromatic beam. The beam from the
monochromator then passes through a filter, toroidal mirror (M2), and is separated into sample
beam and reference beam by half mirror (HM).
The sample beam that has passed through the sample in the sample compartment enters a
detector where it is converted into electric signal. The reference beam also passes through the
sample and enters a detector in the same way, and this signal is used as a reference in
photometry. This ratio beam system in which the reference beam is also detected provides very
stable measured values that are unobtainable with a single beam spectrophotometer.
( 1
^ ' Mirror (IV
Wl lamp
Exit slit * " " \
Filter (?f\
Mirror (M4) \
,, ,, . \ \ Detector 2
Half mirror \ \
L }
(HM) 7X \
Sample Lens
/ \ \ Sample A _
/ \ / \ \ side
M »r
Mirror (M3)\> \ \ U LDetector 1
V
Di
Mirror (M2)
Monochromator Sample compartment
2-3
2.3 Principle of Operation
Fig. 2-5 shows the signal processing & control system. This system can be controlled either by
keyboard or by an external personal computer. The optical signal converted to an electric signal
by the detector is amplified and undergoes A/D conversion, after which it is log-converted by the
software and becomes absorbance data. The result of measurement appears on the LCD and
is printed out on the printer.
Monochromator section
Half mirror
Photometric \
section
I/O A/D
interface converter
Parallel
interface
c
RS-232C
serial
interface
Keyboard Display
Printer
Personal
computer
2-4
2.3.3
MAIN FOR
MENU WARD
00 TO
WL
AUTO
ZERO
CLEAR STOP Operating
ENTER Panel Face
RETURN START
Fig. 2-6
2-5
2.3 Principle of Operation
2.3.4 Display
Current
Current photometric
wavelength
value
-me
WL Scan ^ B a n j n m •BBSIABS
2 Print Setup
• Data Mode
Start WL(nm)
ABS
490.0
3 System Setup Stop WL(nm) 480.0
4 Proc Setup Upper Scale 1.000
5 Save Param Lower Scale -1.000
Speed(nm/min) 100
Measuring \
setting items Parameters
Select Function :
Enter the desired item number of measuring setting in the setting box located at the bottom of
screen.
On the setting screen, enter numerical value, or instead, choose from among pre-set
alternatives.
The U-1900 is equipped with the following auto calibration and self-diagnosis functions.
2-6
2.4
2.4 Specifications
2-7
3.1.1
3. OPERATION
This chapter describes how to handle the operation. As shown in Fig. 3-1-1, this equipment as
the following our operations depending on measurement mode. Read the section suitable to your
purpose.
Measurement of Transmittance Measurement of concentration
(Absorbance): Clause 3.2.1 (creation of calibration curve):
Measurement of Ratio: Clause Clause 3.2.2
3.2.4
Preparatory
operation
Section 3.1
3.1.1 Initialization
When you boot the equipment, it will automatically check the following items.
RAM RAM check
ROM ROM check
Wavelength Driver Wavelength drive-system
D2 Lamp D2 lamp lighting check
Wl Lamp Wl lamp lighting check
656.1 nm Wavelength calibration at 656.1 nm
OK will appear if every item has no problem, and NG will appear if some problems exist. If NG
appears, fix the problem with reference to Section 4.3.
At the time of 656.1 nm, if the lid of the sample chamber is opened (to allow entry of external
light) or if the power switch is turned on (for initialization) with the sample left in the sample
chamber, wavelength calibration cannot be accomplished normally to cause "NG" or wavelength
deviation, resulting in erroneous analyses. When the power switch is turned on (during
initialization), do not open the lid of the sample chamber nor leave a sample in the sample
chamber.
If NG appears in RAM, ROM, Wavelength Driver, D2 Lamp, or 656.1 nm, the system will stop.
If NG appears in Wl Lamp, you can proceed by pressing the CLEAR key.
If initialization ends without any problem, the main menu screen will appear.
3-1
3.1 Booting the Equipment
After the initialization, the parameter setup screen used in the previous measurement appears. If
you select a different measurement mode, press the |MAIN MENU|. The screen as shown in Fig.
3-1-2 will appear. You can select from the five modes on the main menu screen. Each mode will
be explained in and after the following section.
Main Menu
1 Photometry
2 WL Scan
3 Time Scan
4 Load Method
5 Instrument Setup
6 Data Display
Select Function :
Fig. 3-1-2
The following operations are required to select the function displayed on the main menu. Press
the key of the number put on the item using keypad. For example, if you select Photometry
(quantitative operation), select Photometry and press the |ENTER| key. Or enter (1) and press
the | ENTER] key. The screen proceeds to the Photometry screen. Press the | RETURN] key or
press the |MAIN MENU| to return to the original main menu screen.
3-2
3.2.1
3.2 Photometry
If you select Photometry on the main menu screen, the screen as shown in Fig. 3-2-1 will appear.
Select Function :_
Fig. 3-2-1
The following functions can be selected on this mode. Proceed to the explanation of the function
to use.
1) Trace transmittance or absorbance of the designated wavelength —"Clause 3.2.1
2) Calculate concentration from the calibration curve of the standard
liquid --Clause 3.2.2
3) Calculate concentration from the calibration curve by entering
coefficient —•Clause 3.2.3
4) Calculate ratio and difference between 2 or 3 wavelengths —"Clause 3.2.4
[Preparatory Operation!
Print Setup
^UTOZEROl
[Measurement!
Prinj
3-3
3.2 Photometry
Select Function :_
DON @OFF
Fig. 3-2-2
The following items are available on the parameter setup screen. If Data Mode is set at %T
or ABS, the setup of Curve and Standard will become null.
After you finish the setup, press the |RETURN| key, and press the |_AJ key three times,
and select "Param Setup." The screen will return to Fig. 3-2-1.
NOTE: The header information will be printed on all reports even if all print items are set OFF.
3-4
3.2.1
Select Function
0%T @ABS SCONC S RATIO
Fig. 3-2-3
The following items are available on the parameter setup screen. Set the Data Mode at %T
or ABS on the photometry mode.
WL1 to WL6 190.0 to 1100.0 nm To set the wavelength. The figure entered in
the Number of WL is displayed.
After you finish the setup, press the [ RETURN] key. The screen will return to Fig. 3-2-1.
NOTE: If the Data Mode is set at %T or ABS, neither "Curve Setup" nor "Curve Data" is
required for setup.
3-5
3.2 Photometry
If you select "System Setup" in Fig. 3-2-1 and press the [ENTER] key, the screen as
shown in Fig. 3-2-4 will appear.
Photometry
Select Function :_
Q Fast @ Medium Slow
Fig. 3-2-4
After you finish the setup, press the I RETURN] key. The screen will return to Fig. 3-2-1.
3-6
3.2.1
(4) Measurement
After the auto zero is executed, the sample will be measured. Change the sample, and
press the | START| key. After the range is shifted to the one set before and the data at
each wavelength are obtained, the screen as shown in Fig. 3-2-5 will appear.
Select Function :
Q Measure jtj Print Setup number of pages
Fig. 3-2-5
The number of pages will be displayed on the right of the screen. In the case of multiple
pages, you can change the screen to display using the arrow key in the keypad.
NOTE 1: You can execute the auto zero between measurements. Set the blank and press the
[AUTO ZERO! key.
2: Press the | RETURN] key and return to the parameter setup screen to change
parameters.
3-7
3.2 Photometry
Model U-1900!Spectrophotometer
Serial No. 3345997 1
Program No. 3J05300-00 1 Header
Name Information
Report Date 2006/10/23 13:46 J
Operator
Photometry
Data Mode %T
Number of WL 3
WL1(nm) 600.0
WL2(nm) 500.0
WL3(nm) 400.0
Response Medium
Band pass(nm) 4.0
Lamp Change Mode Auto
Lamp Change WL(nm) 340.0
D2 Lamp ON
Wl Lamp ON
Path Length(mm)
Sample
ID WL1(nm) WL2(nm) WL3(nm)
(600.0) ( 500.0 > (400.0)
1 95.3 98.6 95.0
2 65.3 28.6 45.0
NOTE: Fill in the blanks by hand in the items of Name (name of the sample), Operator (name of
the operator), and Path Length (length of the cell).
3-8
3.2.2
[Preparatory Operation
|Print Setupl
[Measurement of Sample
3-9
3.2 Photometry
Photometry
1 Param Setup ON
2 Curve Setup Curve ON
3 Curve Data Standard ON
4 Print Setup Sample ON
5 System Setup
6 Save Param
Select Function :
DON @OFF
Fig. 3-2-7
The following items are available on the parameter setup screen. Set an item desired to be
printed. (You can change the item to be printed after the measurement.)
After you finish the setup, press the | RETURN] key, and press the [_•] key three times,
and select "Param Setup." The screen will return to Fig. 3-2-1.
NOTE: The header information will be printed on all reports even if all print items are set OFF.
3-10
3.2.2
Select Function
fl%T |ABS § CONIC 0 RATIO
Fig. 3-2-8
The following items are available on the measurement parameter setup screen. Set CONC
at Data Mode in the Concentration measurement (creation of calibration curve).
After you finish the setup, press the |RETURN| key. The screen will return to Fig. 3-2-1
3-11
3.2 Photometry
If you select "System Setup" in Fig. 3-2-1 and press the | ENTER! key, the screen as
shown in Fig. 3-2-9 will appear.
Photometry Hnm BS
Select Function :_
QFast | Medium @Slow
Fig. 3-2-9
After you finish the setup, press the | RETURN] key. The screen will return to Fig. 3-2-1
3-12
3.2.2
press the |ENTER| key, the screen as shown in Fig. 3-2-10 will appear.
Select Function :_
Q 1st order @K-Factor
Fig. 3-2-10
3-13
3.2 Photometry
<Curve Type>
Select the type of calibration curve. Here, since a calibration curve is created by using the
standard liquid, select the "1st order." Press Q] and the ]ENTER| key on Fig. 3-2-11.
Photometry K H n ni BEEEABS
Select Function :_
01st order @K-Factor
Fig. 3-2-11
<STDs>
Select the number of standard solutions. You can select up to 20 solutions. This screen is
displayed when you set "1st order" on the "Curve Type." If you wish to set the number of
standards at 3, press and the I ENTER] key on the screen Fig. 3-2-12.
Photometry nm
Input Value :_
Number of STD (2 to 20)
Fig. 3-2-12
3-14
3.2.2
you press the |ENTER| key, the screen as shown in Fig. 3-2-13 will appear.
Enter the concentration of the standard sample.
Select Function :_
Fig. 3-2-13
When you finish setting up parameters, press the |FOR WARD| key on Fig. 3-2-1 to move
to the measurement screen. Measurement parameters will be set up.
3-15
3.2 Photometry
Photometry finm
ID ABS CONC
STD1 0.000
Fig. 3-2-14
Set the first standard sample, here, standard sample of 0.00 concentration, and press the
START! key. Then, measure standard samples 2 and 3. When the measurement of the
standard samples completes, the screen as shown in Fig. 3-2-15 will appear.
Select Function :
Q Measure § Graph § Print Setup
Fig. 3-2-15
3-16
3.2.2
Photometry
m mPlililABS
ID ABS CONC
1
Fig. 3-2-16
Photometry m H BlililABS
ID ABS CONC
1 0.071 9.657
2
Fig. 3-2-17
Then, change the sample and press the | START[ key. Repeat this operation. Press the
| STOP| key to stop the measurement. The screen as shown in Fig. 3-2-18 will appear.
3-17
3.2 Photometry
Select Function :
Q Measure S Graph g Print Setup
Fig. 3-2-18
3-18
3.2.2
Select Function :_
Q Scale @K0,K1 SSTDData
Fig. 3-2-19
The following commands are available in the command selection column on this screen.
<Scale> : You can change the scale of the calibration curve.
<K0,K1> : You can see the information of the calibration curve.
<STD Data> : You can see the data of the calibration curve.
NOTE: While this screen appears, the calibration curve will not be printed even if you press the
| PRINT[ key- Current wavelength and data will be printed. Press the |RETURN| key
on Fig. 3-2-19 to return to the measurement result setup screen of the standard data,
and then press the |PRINT|T key. (If Curve is set "ON" on the Print Setup, the
calibration curve will be printed.)
3-19
3.2 Photometry
Change of Scale
If you select [T| on Fig. 3-2-19, the screen as shown in Fig. 3-2-20 will appear.
Scales of the X-axis and Y-axis of the calibration curve should be set.
Photometry Bnm BS
20.000
X-Min: 0.000
Y-Max: 0.154
Y-Min: 0.000
Input Value :_
X-Max (0.000 to 9999.000)
Fig. 3-2-20
After the completion of scale change, press the | RETURN! key. The screen will return to
the graph screen and the graph on the changed scale will appear.
3-20
3.2.2
Display of Coefficient
If you select-[2] on Fig. 3-2-19, the screen as shown in Fig. 3-2-21 will appear.
R=0.999
R2=0.999
K0= -0.023
K1=129.783
Select Function :_
Q Graph
Fig. 3-2-21
n
Yi VXi
n
Xi=Photometry of each standard
Yi=Entered concentration
n =Number of standards
R=CD =
(n-SXi2-(ZXi)2)(n-IYi2-(lYi)2)
R2=R 2 =CD 2 =
3-21
3.2 Photometry
Photometry
ID ABS CONC
STD1 0.000 0.000
STD2 0.077 10.000
STD3 0.154 20.000
Select Function :_
Q Graph
Fig. 3-2-22
3-22
3.2.2
Photometry
Data Mode CONC
WL(nm) 600.0
Curve Type 1st order
Number of STD 3
Response Medium
Bandpass(nm) 4.0
Lamp Change Mode Auto
Lamp Change WL(nm) 340.0
D2 Lamp ON
Wl Lamp ON
Path Length(mm)
Standard
ID ABS CONC DIFF RD
STD1 0.000 0.000 0.000 0.000
STD2 0.077 10.000 -0.030 -38.961
STD3 0.154 20.000 -0.036 -47.296
Sample
ID ABS CONC ID ABS CONC
1 0.071 9.657 2 0.075 9.984
3 0.150 19.786
Fig. 3-2-23
NOTE: Fill in the blanks by hand in the items of Name (name of the sample), Operator (name of
the operator), and Path Length (length of the cell).
3-23
3.2 Photometry
[Preparatory Operation|
V
Print Setupl
AUTOZEROl
[Measurement of Sample!
Prihl
3-24
3.2.3
Photometry •MMIMABS
Select Function :_
0 O N @OFF
Fig. 3-2-24
The following items are available on the parameter setup screen. Set an item desired to be
printed. (You can change the item to be printed after the measurement.) If Curve Type is
set at K-factor, the setup of Standard will become null.
After you finish the setup, press the | RETURN] key, and press the |_Aj key three times,
and select "Param Setup." The screen will return to Fig. 3-2-1.
NOTE: The header information will be printed on all reports even if all print items are set OFF.
3-25
3.2 Photometry
Photometry
Select Function :_
D%T @ABS @CONC 0 RATIO
Fig. 3-2-25
The following items are available on the measurement parameter setup screen. Set CONC
at Data Mode in the Concentration measurement (entering coefficient).
After you finish the setup, press the | RETURN! key. The screen will return to Fig. 3-2-1.
3-26
3.2.3
If you select "System Setup" in Fig. 3-2-1 and press the |ENTER| key, the screen as
shown in Fig. 3-2-26 will appear.
1 Param Setup
2 Curve Setup Lamp Change D2 Only
3 Curve Data D2 Lamp ON
4 Print Setup Wl Lamp ON
• 5 Svstem Setuo I
6 Save Param
Select Function :_
Q Fast § Medium § Slow
Fig. 3-2-26
After you finish the setup, press the | RETURN] key. The screen will return to Fig. 3-2-1.
3-27
3.2 Photometry
Photometry Knm
Select Function :_
D 1st order § K-Factor
Fig. 3-2-27
3-28
3.2.3
you press the |ENTER| key, the screen as shown in Fig. 3-2-28 will appear.
Enter the K factor.
Select Function :_
Fig. 3-2-28
If the K factor is used, the following formula is used to calculate the concentration.
When you finish setting up parameters, press the I FOR WARD | key on Fig. 3-2-1 to move
to the measurement screen. Measurement parameters will be set up.
3-29
3.2 Photometry
Photometry im
ID ABS CONC
1
Fig. 3-2-29
Photometry Bnm
ID ABS CONC
1 0.071 9.657
2
Fig. 3-2-30
Then, change the sample and press the I START| key. Repeat this operation. Press the
I STOP| key to stop the measurement. The screen as shown in Fig. 3-2-31 will appear.
3-30
3.2.3
Select Function :
Q Measure @ Graph § Print Setup
Fig. 3-2-31
3-31
3.2 Photometry
0.00'
0.000
Select Function :_
D Scale @K0,K1
Fig. 3-2-32
The following commands are available in the command selection column on this screen.
<Scale> : You can change the scale of the calibration curve.
<K0,K1> : You can see the information of the calibration curve.
NOTE: While this screen appears, the calibration curve will not be printed even if you press the
PRINT[ key. Current wavelength and data will be printed. Press the I RETURN] key
on Fig. 3-2-32 to return to the measurement result setup screen of the standard data,
and then press the | PRINT| key. (If Curve is set "ON" on the Print Setup, the
calibration curve will be printed.)
3-32
3.2.3
Change of Scale
If you select \T\ on Fig. 3-2-32, the screen as shown in Fig. 3-2-33 will appear.
Scales of the X-axis and Y-axis of the calibration curve should be set.
Input Value :_
X-Max (0.000 to 9999.000)
Fig. 3-2-33
After the scale was changed, press the | RETURN] key to return to Fig. 3-2-32. A
calibration curve will be redrawn in the changed scale.
Display of Coefficient
If you select [|] on Fig. 3-2-32, the screen as shown in Fig. 3-2-34 will appear.
Photometry nm \BS
K0= -0.023
K1=129.783
Select Function :_
Q Graph
Fig. 3-2-34
Press |JJ |ENTER| or the RETURN key to return to the previous screen (Fig. 3-2-32).
3-33
3.2 Photometry
1.000
ABS
^ - "
i
0.000 CONC 20.000
Photometry
Data Mode CONC
WL(nm) 600.0
Curve Type K-Factor
Response Medium
Bandpass(nm) 4.0
Lamp Change Mode Auto
Lamp Change WL(nm) 340.0
D2 Lamp ON
Wl Lamp ON
Path Length(mm)
Standard
K0=-0.023 K1=129."^83
Sample
ID ABS CONC ID ABS CONC
1 0.071 9.357 2 0.075 9.984
3 0.150 19. 786
Fig. 3-2-35
NOTE: Fill in the blanks by hand in the items of Name (name of the sample), Operator (name of
the operator), and Path Length (length of the cell).
3-34
3.2.4
[Preparatory Operation]
[Print Setupl
[AUTOZEROl
[Measurement!
iPrintj
3-35
3.2 Photometry
Select Function :_
DON @OFF
Fig. 3-2-36
The following items are available on the parameter setup screen. Set an item desired to be
printed. (You can change the item after the measurement.) If Data Mode is set at RATIO,
the setup of Curve and Standard will become null.
After you finish the setup, press the | RETURN! key, and press the |_AJ key three times,
and select "Param Setup." The screen will return to Fig. 3-2-1.
NOTE: The header information will be printed on all reports even if all print items are set OFF.
3-36
3.2.4
Photometry BIIIIIMABS
Select Function
D%T @ABS @CONC S RATIO
Fig. 3-2-37
The following items are available on the measurement parameter setup screen. Set the
Data Mode to RATIO for the ratio measurement.
After you finish the setup, press the [RETURN] key. The screen will return to Fig. 3-2-1.
The following calculations will be performed using the set up wavelength.
3-37
3.2 Photometry
If you select "System Setup" in Fig. 3-2-1 and press the |ENTER| key, the screen as
shown in Fig. 3-2-38 will appear.
Photometry
Select Function :_
Q Fast § Medium § Slow
Fig. 3-2-38
After you finish the setup, press the | RETURN! key. The screen will return to Fig. 3-2-1.
3-38
3.2.4
When you finish setting up parameters, press the | FOR WARD [ key on Fig. 3-2-1 to move
to the measurement screen. Measurement parameters will be set up.
Photometry Knm
ID
1 WL1(600.0nm)
WL2(500.0nm)
WL3(400.0nm)
DIFF
RATIO
Fig. 3-2-39
Set the sample and press the [ START| key to move to the wavelength that was set,
collect the data, and display the result. The following ID number will appear.
Photometry
ID
3 WL1(280.Onm) 0.214
WL2(260. Onm) 0.064
WL3(320. Onm) 0.012
DIFF 0.150
RATIO 3.885
4
: Prev • : Next (1/1)
Fig. 3-2-40
3-39
3.2 Photometry
Then, change the sample and press the | START| key. Repeat this operation. Press the
| STOP| key to stop the measurement. The screen as shown in Fig. 3-2-41 will appear. The
screen will move to the set wavelength, and obtained data to display the result.
Photometry BlililABS
ID
3 WL1(280.0nm) 0.214
WL2(260.0nm) 0.064
WL3(320.0nm) 0.012
DIFF 0.150
RATIO 3.885
Select Function :
Q Measure § Print Setup
Fig. 3-2-41
3-40
3.2.4
Photometry
Data Mode RATIO
Number of WL 3
WL1(nm) 600.0
WL2(nm) 500.0
WL3(nm) 400.0
Response Medium
Band pass(nm) 4.0
Lamp Change Mode Auto
Lamp Change WL(nm) 340.0
D2 Lamp ON
Wl Lamp ON
Path Length(mm)
Sample
ID WL1(nm) WL2(nm) WL3(nm) DIFF RATIO
(600.0) (500.0) (400.0)
1 0.103 0.032 0.012 0.071 4.550
2 0.112 0.037 0.012 0.075 4.000
3 0.214 0.064 0.012 0.150 3.885
NOTE: Fill in the blanks by hand in the items of Name (name of the sample), Operator (name of
the operator), and Path Length (length of the cell).
3-41
3.3 Wavelength Scan
When you select \2\ on the main menu screen, the screen as shown in Fig. 3-3-1 will appear.
Select Function :_
Fig. 3-3-1
[Preparatory Operation]
Print Setup]
[Baseline Calibration]
i
[Measurement
Data Processing
3-42
3.3
Select Function :_
Fig. 3-3-2
The following items are available on the parameter setup screen. Set an item desired to be
printed. (You can change the item to be printed after the measurement.)
3-43
3.3 Wavelength Scan
After you finish the setup, press the | RETURN] key, and press the [_Aj key once, and
select "Param Setup." The screen will return to Fig. 3-3-1.
NOTE: The header information will be printed on all reports even if all print items are set OFF.
3-44
3.3
<Data Mode>
If you press the|ENTER| key on Fig. 3-3-1, the screen as shown in Fig. 3-3-3 will appear.
Select Function
fl%T @ABS 1E(S) flE(R)
Fig. 3-3-3
<Speed>
Set a value for the speed after you select the scan speed by using up- and down-arrow
keys in Fig. 3-3-3. The screen as shown in Fig. 3-3-4 will appear.
3-45
3.3 Wavelength Scan
Select Function :_
D10 @ 100 @200 0400
@800 §1200 02400 @3600
Fig. 3-3-4
The sampling interval varies with measurement range of the wavelength and scan speed.
Whichever the bigger one of the two intervals will be effective.
For example, the sampling interval will be 0.5 nm if the measurement range is between 600
nm and 400 nm is scanned at 400 nm/min.
3-46
3.3
If you select "System Setup" in Fig. 3-3-1 and press the | ENTER) key, the screen as
shown in Fig. 3-3-5 will appear.
WL Scan
Select Function :_
Q Fast | Medium @ Slow
Fig. 3-3-5
3-47
3.3 Wavelength Scan
After you finish the setup, press the | RETURN.! key. The screen will return to Fig. 3-3-1.
3-48
3.3
When you finish setting up parameters, press the [FOR WARD| key on Fig. 3-3-1 to move
to the measurement screen. Measurement parameters will be set up.
Baseline Correction
Fig. 3-3-6
Set the blank and press the | START I key. If the baseline has already been measured and
the measurement parameters are not changed, the screen as shown in Fig. 3-3-7 will
appear.
Fig. 3-3-7
Select the [START\ key to measure the baseline again, and select the |CLEAR[ key to
use the previous baseline.
3-49
3.3 Wavelength Scan
1.000
ABS
0.000
400.0 500.0 600.0 nm
Set a sample and press START key:
RETURN key: Baseline Correction
Fig. 3-3-8
Set the sample and press the I START| key. The measurement starts. While the
measurement is being performed, the screen as shown in Fig. 3-3-9 will appear.
Wavelength Scan im
1.000
ABS
0.000
400.0 500.0 600.0 nm
Scanning
Fig. 3-3-9
3-50
3.3
When the measurement ends, the screen as shown in Fig. 3-3-10 will appear and the
spectrum will appear.
1.000
ABS
0.000
400.0 500.0 600.0 nm
Fig. 3-3-10
3-51
3.3 Wavelength Scan
WL Scan/Scale
600.0
X-Min(nm): 400.0
Y-Max(ABS): 1.000
Y-Min(ABS): 0.000
Auto Scale
Input Value :_
X-Max(nm) (400.0 to 600.0)
Fig. 3-3-11
After you changed the scale, press the [RETURNl key. The scale will be changed, and
the spectrum will appear.
3-52
3.3
ABS
0.000
400.0 500.0 600.0 nm
Select Function:_
Q All Data @ Peak @ Processing
Fig. 3-3-12
3-53
3.3 Wavelength Scan
WL Scan/All Enm
1.0001
ABS
0.000
400.0 500.0 600.0 nm
Select Function :_
Q Data List § List Interval (40.0nm)
Fig. 3-3-13
If you select [T| on Fig. 3-3-13, the screen as shown in Fig. 3-3-14 will appear. The
wavelength and photometry value will appear.
Select Function :
flData List S List Interval (40.0nm)
Fig. 3-3-14
If data extends to multiple pages, change the page using the arrow keys.
When you print the list, press the | PRINT] key. If you select in Fig. 3-3-13 or Fig.
3-3-14, you can change the List interval.
NOTE: All data lists will be printed instead of the data list on the screen only.
3-54
3.3
ABS
0.000
A
400.0 500.0 600.0 nm
Select Function :_
Q Data List @ Detective
(T=0.001 S=1)
Fig. 3-3-15
If you select [T] on Fig. 3-3-15, the screen as shown in Fig. 3-3-16 will appear.
WL Scan/Peak \BS
1.000
ABS
0.000
400.0 500.0 600.0 nm
Select Function :_
QPeak gj Valley S Peak&Valley
Fig. 3-3-16
3-55
3.3 Wavelength Scan
Peak Detection
If you select [Tj on Fig. 3-3-16, the screen as shown in Fig. 3-3-17 will appear.
Select Function :
QPeak @ Valley @ Peak&Valley
Fig. 3-3-17
If data extends to multiple pages, change the page using the arrow keys
Press the |PRINT| key to print the list. If the peak cannot be detected, "No Peak" will
appear. If you select \2\ in Fig. 3-3-15, you can change the peak detection parameters.
3-56
3.3
Valley Detection
If you select [2] on Fig. 3-3-16, the screen as shown in Fig. 3-3-18 will appear.
The valley wavelength and photometry value will appear. If the valley cannot be detected,
"No Valley" will appear. If you select [2] in Fig. 3-3-15, you can change the peak detection
parameters.
Select Function :
QPeak S Valley g Peak&Valley
Fig. 3-3-18
3-57
3.3 Wavelength Scan
ABS
0.000
400.0 500.0 600.0 nm
Select Functions
0 Smooth @ Derivative § Reset
Fig. 3-3-19
(a) Smoothing
If you select Q] on Fig. 3-3-19, the screen as shown in Fig. 3-3-20 will appear. The
screen shows the smoothed spectrum.
WL Scan/Smooth
1.000
ABS
0.000
400.0 500.0 600.0 nm
Select Functions
Q Scale
Fig. 3-3-20
3-58
3.3
WL Scan/Deriv
1.000
ABS
0.000
400.0 500.0 600.0 nm
Select Function^
E| Scale
Fig. 3-3-21
3-59
3.3 Wavelength Scan
1.000
A
ABS
0.000
Wavelength Scan
Data Mode ABS
Start WL (nm) 600.0
Stop WL (nm) 400.0
Scan Speed(nm/min) 100
Sampling Interval (nm) 0.5
Response Medium
Bandpass(nm) 4.0
Lamp Change Mode Auto
Lamp Change WL(nm) 340.0
D2 Lamp ON
Wl Lamp ON
Path Length(mm)
Fig. 3-3-22
NOTE: Fill in the blanks by hand in the items of Name (name of the sample), Operator (name of
the operator), and Path Length (length of the cell).
3-60
3.4
When you select (Time Scan) on the main menu screen, the screen as shown in Fig. 3-4-1
will appear.
Select Function :_
Fig. 3-4-1
[Preparatory Operation!
iPrint Setup]
AUTOZERO
[Measurement)
|Data Processing
3-61
3.4 Time Scan
Select Function :_
Fig. 3-4-2
Press the [ENTER| key on this screen to set the print parameters. The following items are
available on the parameter setup screen. Set an item desired to be printed. (You can
change the item after the measurement.)
3-62
3.4
After you finish the setup, press the [RETURN] key, and press the _A_ key once, and
select "Param Setup." The screen will return to Fig. 3-4-1.
NOTE: The header information will be printed on all reports even if all print items are set OFF.
<Data Mode>
You can choose a data mode from the following four.
1-%T Transmittance
2-ABS Absorbance
3- E(S) Energy (Sample energy)
4- E(R) Energy (Reference energy)
<WL (nm)>
Sets the measurement wavelength. The entry range is between 190.0-1100.0 nm.
3-63
3.4 Time Scan
<Scan Time>
Sets the scan time. (Minimum entry time is 60 seconds.)
The sampling interval will be set automatically depending on the scan time.
3-64
3.4
If you select "System Setup" in Fig. 3-4-1 and press the | ENTER] key, the screen as
shown in Fig. 3-4-3 will appear.
1 Param Setup
2 Print Setup Lamp Change D2 Only
• 3 Svstem Setup D2 Lamp ON
4 Proc Setup Wl Lamp ON
5 Save Param
Select Function :_
Q Fast § Medium § Slow
Fig. 3-4-3
After you finish the setup, press the | RETURN] key. The screen will return to Fig. 3-4-1.
3-65
3.4 Time Scan
Time Scan
1.0001
ABS
0.000
0.0 30.0 60.0 s
Select Function or press START key:_
RETURN key: Parameter setup
Fig. 3-4-4
(4) Measurement
After performing the auto zero, replace the sample and press the | START| key to start the
measurement. The screen as shown in Fig. 3-4-5 will appear during measurement.
1.000T
ABS
0.000.!
30.0 60.0 s
Scanning
Fig. 3-4-5
3-66
3.4
When the measurement ends, the screen as shown in Fig. 3-4-6 will appear.
i.ooq
ABS
0.000
30.0 60.0 s
Select Function or press START key:_
Q DataProc @ Scale | Print Setup
Fig. 3-4-6
60.0
X-Min(s): 0.0
Y-Max(ABS): 1.000
Y-Min(ABS): 0.000
Auto Scale
Input Value :_
X-Max(s) (0.0 to 60.0)
Fig. 3-4-7
After you changed the scale, press the [RETURNl key. The scale will be changed, and
the spectrum will appear.
3-67
3.4 Time Scan
1.000
ABS
0.000
30.0 60.0 s
Select Function :_
Q All Data | Peak @ Processing
Fig. 3-4-8
3-68
3.4
All Data
If you select Q] on Fig. 3-4-8, the screen as shown in Fig. 3-4-9 will appear.
ABS
0.000
0.0 30.0 60.0 s
Select Function :_
Q Data List @ List Interval (1.0 s)
Fig. 3-4-9
If you select Q] on Fig. 3-4-9, the screen as shown in Fig. 3-4-10 will appear. The
wavelength and photometry value will appear.
Data List
ID TIME(s) ABS
1 0.0 0.104
2 1.0 0.104
3 2.0 0.104
4 3.0 0.104
5 4.0 1.104
Select Function :
0Data List I List Interval (1.0 s)
•s
\
Current
list interval
Fig. 3-4-10
If data extends to multiple pages, change the page using the arrow keys.
When you print the list, press the | PRINT) key. If you select \2\ in Fig. 3-4-9 or Fig.
3-4-10, you can change the list interval.
NOTE: All data lists will be printed instead of the data list on the screen only.
3-69
3.4 Time Scan
ABS
0.000
0.0 30.0 60.0:
Select Function :_
Q Data List § Detective
(T=0.001 S=1)
Fig. 3-4-11
If you select [T] on Fig. 3-4-11, the screen as shown in Fig. 3-4-12 will appear.
Time Scan/Peak
1.000
ABS
0 000 /
0.0 30.0 60.0 s
Select Function :
QPeak g Valley § Peak&Valley
Fig. 3-4-12
3-70
3.4
Peak Detection
If you select Q] on Fig. 3-4-12, the screen as shown in Fig. 3-4-13 will appear.
The peak time and photometry value will appear.
Peak
ID TIME(s) ABS
1 5.2 0.104
2 50.4 0.097
Select Function :_
flPeak § Valley @Peak&Valley
Fig. 3-4-13
If data extends to multiple pages, change the page using the arrow keys .
Press the | PRINT] key to print the list. If the peak cannot be detected, "No Peak" will
appear. If you select [2] in Fig. 3-4-11, you can change the peak detection parameters.
3-71
3.4 Time Scan
Valley Detection
If you select \2\ on Fig. 3-4-12, the screen as shown in Fig. 3-4-14 will appear.
The valley time and photometry value will appear.
Time Scan/Peak
Valley
ID TIME(s) ABS
1 28.1 0.028
2 60.0 0.034
Select Function :_
QPeak @ Valley g Peak&Valley
Fig. 3-4-14
If data extends to multiple pages, change the page using the arrow keys
Press the | PRINTl key to print the list. If the valley cannot be detected, "No Valley" will
appear. If you select \2\ in Fig. 3-4-11, you can change the peak detection parameters.
3-72
3.4
1.000
ABS
0.000
0.0 30.0 60.0 s
Select Function:_
Q Smooth @ Derivative § Reset
Fig. 3-4-15
(a) Smoothing
If you select |T| on Fig. 3-4-15, the screen as shown in Fig. 3-4-16 will appear.
The smoothed spectrum will appear.
1.000
ABS
0.000
0.0 30.0 60.0 s
Select Function :_
Q Scale
Fig. 3-4-16
3-73
3.4 Time Scan
(b) Derivation
If you select \2\ on Fig. 3-4-15, the screen as shown in Fig. 3-4-17 will appear. The
spectrum after the first derivation will appear.
i.ooq
ABS
0.000
30.0 60.0 s
Select Function :_
Q Scale
Fig. 3-4-17
3-74
3.4
1.000
ABS
0.000
0.0 15.0 30.0 45.0 60.0 s
Time Scan
Data Mode ABS
WL(nm) 600.0
Scan Time (s) 60.0
Sampling Interval (s) 0.1
Response Medium
Bandpass(nm) 4.0
Lamp Change Mode Auto
Lamp Change WL(nm) 340.0
D2 Lamp ON
Wl Lamp ON
Path Length(mm)
Fig. 3-4-18
NOTE: Fill in the blanks by hand in the items of Name (name of the sample), Operator (name of
the operator), and Path Length (length of the cell).
3-75
3.5 Measurement in Enlarged Display Screen
If you select [6] Data Display on the main menu screen, the screen as shown in Fig. 3-5-1
will appear. This mode enlarges the displayed characters for easy reading. It is the
effective mode to confirm the photometry by wavelength.
Wavelength:
500.0 nm
Data:
0.001 ABS
Select Data Mode:
0%T @ABS @E(S) 0E(R)
Fig. 3-5-1
To change the wavelength, press the [GO TO WL|, enter the wavelength, and press the
I ENTER] key.
The following is the example to measure absorbance or transmittance at wavelength of 500
nm.
1) Press the | GO TO WU and enter 500, and press the |ENTER| key.
The screen will move to that of 500 nm wavelength.
2) Enter the number of the corresponding data mode and press the [ENTER! key to
change the data mode.
3) Set the blank.
4) Press the |AUTOZERO|. (The absorbance of the blank is set at zero. In the case of
transmittance, it is set at 100.)
5) Set the sample.
6) Scan the data displayed on the LCD screen. If the printer is connected, press the
|PRINT| key to print them.
3-76
3.5
Print sample
500.Onm 100.0%T
500.0nm 100.0%T
NOTE: In the case of the print by a printer, only the measurement result is printed without titles
such as the ID Number. If you wish to print the number, use the mode of item 3.2.1.
3-77
3.6 Instrument Setup
If you select [5] Instrument Setup on the main menu screen, the screen as shown in Fig.
3-6-1 will appear.
1 WL Calibration
2 0%T
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Select Function :
Fig. 3-6-1
Item Details
Detects D2 lamp's emission line of 656.1 nm and
WL Calibration
performs the wavelength automatically.
0%T Performs 0%T.
Lamp Time Sets lighting of the lamp, displays and resets the used
time.
3-78
3.6.1
Press the |ENTER| key in Fig. 3-6-1. The screen as shown in Fig. 3-6-2 will appear.
1 WL Calibration
2 0%T
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Fig. 3-6-2
After confirming that nothing is set in the cell holder, close the lid of the sample room. Press the
START key to start the wavelength calibration. Detects D2 lamp's emission lines of 656.1 nm
and perform the wavelength calibration. The STOP key does not work while calibration is being
performed.
Instrument Setup
1 WL Calibration
2 0%T
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Executing WL Calibration...
Fig. 3-6-3
NOTE: If an error appears, the screen will return to the system menu. Perform this function
again. If it ends without problem, the screen will return to Fig. 3-6-2. Press the RETURN
key if you select another item.
3-79
3.6 Instrument Setup
3.6.2 0%T
Perform this function when more than two hours have passed since power was turned on. Select
the "0%T" on the equipment setup screen on Fig. 3-6-1. The screen as shown in Fig. 3-6-4 will
appear.
Instrument Setup Bnm
U/VLCalibration
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Select Function :
Fig. 3-6-4
Press the )ENTER| key. The confirmation screen as shown below will appear.
Instrument Setup
IWLCalibration
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Fig. 3-6-5
3-80
3.6.2
Press the | START] key to start the 0%T. The screen as shown in Fig. 3-6-6 will appear while it
is being performed.
1 WL Calibration
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Executing...
Fig. 3-6-6
If the calibration ends, the screen will return to Fig. 3-6-5. Press the | RETURN] key to select
another item.
3-81
3.6 Instrument Setup
3.6.3 <Lamp Time> Display and Reset the Lighting Time of Lamp
If you select the "Lamp Time" on the equipment setup screen in Fig. 3-6-1, the screen as shown
in Fig. 3-6-7 will appear.
1 WL Calibration
2 0%T
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Select Function :
Fig. 3-6-7
Next, press the [ENTER] key. The following screen will appear. The date when the lamp was
replaced and the used time of the lamp will appear.
Lamp exchange:
D2 Lamp Usage(h) 33
Exchange day 10/09/2003
Wl Lamp Usage(h) 23
Exchange day 10/09/2003
Select Function :_
Q D2 Lamp | Wl Lamp
Fig. 3-6-8
3-82
3.6.3
If you select Q] or {2}, used hours of the lamp is reset. Reset the time when the lamp is
replaced. The following explanation is the operation case that the D2 lamp is replaced.
Q] on Fig. 3-6-8 and press the [ENTER] key. The confirmation message as shown below will
appear.
Lamp Time
Lamp exchange:
D2 Lamp Usage(h) 33
Exchange day 10/09/2003
Wl Lamp Usage(h) 23
Exchange day 10/09/2003
Fig. 3-6-9
3-83
3.6 Instrument Setup
This function confirms the performance of the equipment. Parameters for measurement and
judgment are already programmed. The following two measurement methods are available.
GLP/GMP <AUTO> : Function to confirm all items without interruption.
GLP/GMP <MANUAL>: Function to select and confirm each item.
The items set here are those to be measured with nothing being placed in the cell holder.
Confirm there is nothing in the sample room to perform the measurement.
3-84
3.6.4
3.6.4.1 <GLP/GMP(Auto)>
If you select the "GLP/GMP(Auto)" on the instrument setup screen in Fig. 3-6-1, the screen as
shown in Fig. 3-6-10 will appear.
1 WL Calibration
2 0%T
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Select Function :
Fig. 3-6-10
Next, press the |ENTER| key. The following screen will appear. Press the | START] key to
perform the automatic measurement.
GLP/GMP(Auto)
WLAccuracy(656.1)
WL Accuracy(486.0)
WL Repeatability
Bandpass
Noise Level
Baseline Flatness
Baseline Stability
Hardware
Fig. 3-6-11
3-85
3.6 Instrument Setup
If you press the [ START| key, the screen as shown below will appear.
GLP/GMP(Auto) im BS
WLAccuracy(656.1)
WL Accuracy(486.0)
WL Repeatability
Bandpass
Noise Level
Baseline Flatness
Baseline Stability
Hardware
Fig. 3-6-12
pi] If you select Yes, the automatic measurement will be performed. Result will be printed when
each item is measured.
[2] If you select No, the automatic measurement will be performed but automatic print is not
performed. Result will be displayed without automatic print.
When you press the [ RETURN] key, the screen will return to Fig. 3-6-10.
GLP/GMP(Auto) im \BS
WL Accuracy(656 1)
WL Accuracy(486.0)
WL Repeatability
Bandpass Item under current execution is
Noise Level
highlighted
Baseline Flatness
Baseline Stability
Hardware
Executing...
Fig. 3-6-13
3-86
3.6.4
Press the |STOP[ key to stop the automatic measurement. Result of the active item is not
displayed. If you select the GLP/GMP (Manual), you can select the item and measure it. (Result
of GLP/GMP (Automatic) will be kept.)
The following screen will appear when the automatic measurement ends.
WLAccuracy(656.1) O.Onm OK
WL Accuracy(486.0) O.Onm OK
WL Repeatability +/-0.05nm OK
Bandpass 3.8nm OK
Noise Level +/-0.00000ABS OK
Baseline Flatness +/-0.00020ABS OK
Baseline Stability 0.0001 OABS OK
Hardware OK
Fig. 3-6-14
You can print the summary report of the automatic measurement by "Summary Report" on the
GLP/GMP (Manual).
3-87
3.6 Instrument Setup
3.6.4.2 <GLP/GMP(Manual)>
If you select the GLP/GMP (Manual) on the equipment setup screen in Fig. 3-6-1, the screen as
shown in Fig. 3-6-15 will appear.
1 WL Calibration
2 0%T
3 Lamp Time
4 GLP/GMP(Auto)
ElBEBffilJllill
6 Printer Setup
7 Time Setup
Select Function :
Fig. 3-6-15
Next, press the | ENTER) key. The following screen will appear. Select the item to measure
using the up- and down-arrow keys, and press the | START] key to start the measurements.
GLP/GMP(Manual) Knm
WL Accuracy(656.1)
WL Accuracy(486.0)
WL Repeatability
Bandpass
Noise Level
Baseline Flatness
Baseline Stability
Hardware
Summary Report
Fig. 3-6-16
3-88
3.6.4
<WL Reproducibility>
First a WL range from 660.0 to 650.0 nm is measured, then wavelength is shifted to 190 nm, and
range from 660.0 to 650.0 nm is measured again. At the third time wavelength is shifted to
1100 nm and the range from 660.0 to 650.0 nm is measured again.
The difference between maximum and minimum values for the three measurements is then
determined. If one half of the difference is within 0.3 nm, OK will be indicated.
<Bandpass>
WL range from 660.0 to 650.0 nm is measured the same as for WL accuracy.
The wavelength width (half-value width) at one half of the height versus maximum peak height
(A) of the obtained spectrum is then determined.
Wavelength
Half-value width
(Bandpass)
Fig. 3-6-17
3-89
3.6 Instrument Setup
<Noise Level>
Perform the auto zero processing, and then measure the time-dependence of the absorbance
fluctuation for 300 seconds at 500nm of wavelength.
Calculate the half value of the difference {AI2) between the maximum and the minimum in the
measured absorbance.
WhenZl/2 falls within ±0.00015Abs (i.e., -0.00015^Zl/2^0.00015), "OK" is indicated.
<Baseline Flatness>
Perform the wavelength scanning measurement from 950.Onm to 200.Onm of wavelengths.
Obtain the baseline, and then conduct the measurement.
Divide 750 items of data, excluding the beginning data, into 150 blocks each having wavelength
spread of 5nm (i.e., 1 block = 5nm wide). Then, expressing a given block as "block i", find the
maximum a(i) and the minimum b(i) within this given block i; and then calculate a value given by
(a(i) - b(i))/2 + b(i).Let this calculation be c(i), which defines the flatness c(i) of the block i for the
wavelength spread of 5nm.
Calculate the flatness c(i) for all the data (150 groups). Find the maximum c(i): A, and the
minimum c(i): B from among c(i) calculated as above. The baseline flatness {AI2) is then defined
as the value given by (A-B)/2.
When the baseline flatness (AI2) falls within ±0.002Abs (i.e., -0.002^^1/2^0.002), ±Al2 (see
Note) and "OK" are indicated.
<Baseline Stability>
Measure the time-dependence of the absorbance fluctuation for 300 seconds at 500nm of
wavelength, and then apply one time the smoothing processing.
Calculate the difference between the maximum and the minimum of the absorbance after the
smoothing.
When this value falls within 0.0004Abs, "OK" is indicated.
3-90
3.6.4
When measurement of all the items ends, results and OK or NG will appear as shown
in Fig. 3-6-18.
GLP/GMP(Manual) finm
WLAccuracy(656.1) O.Onm OK
WL Accuracy(486.0)
WL Repeatability
Bandpass
Noise Level
Baseline Flatness
Baseline Stability
Hardware
Summary Report
Select Function :_
Q Print RETURN key: Cancel
Fig. 3-6-18
When measurement of each item ends, the selection column to ask whether to print
the result will appear in the command setup column.
If you select Q], the result will be printed. Printed sample is as shown in Fig. 3-6-20.
If you press the | RETURN] key, you will finish the item without printing it.
The screen as shown in Fig. 3-8-12 will appear only for the items of "Hardware".
Enter [T) to print the result.
Hardware
RAM OK
ROM OK
Wavelength Driver OK
D2 Lamp OK
Wl Lamp OK
D2 Lamp Usage(h) 23
Wl Lamp Usage (h) 33
Select Function :_
Q Print RETURN key: Cancel
Fig. 3-6-19
Lastly, select the "Summary Report" and [T] (Print)to print the list of the result. Spectrum will not
be printed in this print. Fig. 3-6-21 shows the print sample.
3-91
3.6 Instrument Setup
400
E(R)
/ \
/
/
GLP/GMP
WL Accuracy (656.1nm) +/- 0.5 nm OK
Wavelength Scan
Data Mode E(R)
Start WL (nm) 660.0
Stop WL (nm) 650.0
Scan Speed(nm/min) 100
Sampling Interval (nm) 0.1
Response Fast
Bandpass(nm) 4.0
Lamp Change Mode D2 Lamp Only
Lamp Change WL(nm) 340.0
D2 Lamp ON
Wl Lamp ON
Path Length(mm)
3-92
3.6.4
GLP/GMP
WL Accuracy (656.1nm) +/- 0.5 nm -0.2 nm OK
WL Accuracy (486.Onm) +/- 0.5 nm 0.0 nm OK
WL Repeatability +/- 0.3 nm +/- 0.1 nm OK
Bandpass 3.5-4.5 nm 3.8 nm OK
Noise Level +/-0.00015ABS +/- 0.00000ABS OK
Baseline Flatness +/- 0.002ABS +/-0.00010ABS OK
Baseline Stability 0.0004ABS 0.00020ABS OK
Hardware
RAM OK
ROM OK
Wavelength Driver OK
D2 Lamp OK
Wl Lamp OK
D2 Lamp Usage (h) 24
Wl Lamp Usage (h) 33
3-93
3.6 Instrument Setup
If you select the "Printer Setup" on the instrument setup screen in Fig. 3-6-1, the screen as
shown in Fig. 3-6-22 will appear.
1 WL Calibration
2 0%T
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Select Function :
Fig. 3-6-22
Next, press the |ENTER| key. The following screen will appear. Set the printer type.
Printer Setup
flnm
Printer Type ESC/P (Cut)
Select Function :_
0DPU @PCL3
S ESC/P(Cut) S ESC/P(Continuous)
Fig. 3-6-23
3-94
3.6.6
If you select the "Time Setup" on the instrument setup screen in Fig. 3-6-1, the screen as shown
in Fig. 3-6-24 will appear.
1 WL Calibration
2 0%T
3 Lamp Time
4 GLP/GMP(Auto)
5 GLP/GMP(Manual)
6 Printer Setup
7 Time Setup
Select Function :
Fig. 3-6-24
3-95
3.6 Instrument Setup
Next, press the |ENTER| key. The following screen will appear. Set the printer type. You can
set time, the time of printing the report, and the format of time display by selecting the item using
the up- or down-arrow key.
Input Value :_
Minute (0 to 59)
Fig. 3-6-25
NOTE: This clock allows adjustment only to minute, and does not allow adjustment to second.
Accuracy is within about ±1 min. a month.
3-96
3.7
The parameters that you set remain even if the equipment is turned off, and the remaining set
parameters will be set as measurement parameters when the equipment is turned on again. If
you save these measurement parameters as a file, the following procedure needs to be
performed. Press [6] Save Param [ENTER| in Fig. 3-2-1, or press [E\ Save Param Enter in
Fig. 3-3-1 or Fig. 3-4-1. The screen as shown in Fig. 3-7-1 will appear.
\
Parameter Mode Date and time of last save
Input Value :_
File No. (1 to 5)
Fig. 3-7-1
The parameters that you saved in the Save Method of the quantity measurement parameter,
wavelength scan measurement parameter, and time scan measurement parameter setup
screens can be displayed up to five. The measurement mode of saved parameter will be
displayed next to the file number.
You can save the measurement parameters either by selecting the file by using the up- or
down-arrow key and press the |ENTER| key, or pressing the | ENTER! key after entering the
file number. If the file you designate the number is already saved, the following message will
appear.
No. 1 already exists. Overwrite?_
j|Yes @No
|T|| ENTER]: Overwrites and saves.
[2]! ENTER): DO not save (The same is true when the | RETURN] key is pressed.)
3-97
3.8 Loading the Measurement Parameter File
On this screen, the 1 FORWARD] key is not acceptable. Return to the parameter setup screen
by pressing the | RETURN| key in Fig.3-7-1, and press the | FORWARD] key. The screen will
move to the measurement screen.
Select [4] Load Method on the main menu screen. The screen as shown in Fig. 3-8-1 will
appear.
Load Method im
1 PHT 2005/11/0515:30
2 WLS 2005/11/0513:44
3 TMS 2005/11/05 9:21
4 WLS 2005/11/0519:32
5
\
Parameter Mode Date and time of last save
Input Value :_
File No. (1 to 5)
Fig. 3-8-1
The parameters that you saved in the Save Method of the quantity measurement parameter,
wavelength scan measurement parameter, and time scan measurement parameter setup
screens can be displayed up to five. The measurement mode of saved parameter will be
displayed next to the file number.
You can save the measurement parameters either by selecting the file by using the up- or
down-arrow key and press the | ENTER! key, or pressing the |ENTER| key after entering the
file number.
After loading the measurement parameters, set the measurement parameters and move to the
corresponding measurement screen.
3-98
3.9
The following messages will appear. Please refer the measure to fix the trouble.
3-99
4.1.1
4. MAINTENANCE
This spectrophotometer has been constructed precisely with extreme care to provide superlative
accuracy in analytical operation. For ensuring the high accuracy of this spectrophotometer, it is
required to carry out its maintenance properly.
If a sample spills inside the sample compartment, remove and clean its base immediately (see
Fig. 4-1).
Guide pin
4-1
4.1 Maintenance and Checkup
(2) Remove the setscrews on the sample compartment base. Then, take out the sample
compartment base.
g| ' rf
Fig. 4-2
(3) Clean the sample compartment base thus removed, and the inside of the sample
compartment.
(4) After cleaning, mount the sample compartment base along the guide pins.
Avoid placing such items as samples on the top of the spectrophotometer to prevent possible
deterioration of its optical performance.
If a sample spills on the instrument, be sure to clean it using a soft cloth moistened with water or
non-stimulating detergent as soon as possible. After cleaning, let it dry up. Take particular
care not to spill a sample over the keypad.
After measurement, wash each cell with distilled water adequately. Swish water off the inside of
the cell, and then wipe the cell with a soft cloth or gauze thoroughly.
For storing the cell, use a desiccator. Alternatively, immerse the cell into a beaker of distilled
water, put a lid on the beaker, and store it in a clean place.
If the transmission face of the cell is contaminated, clean it using a soft cloth moistened with a
1:1 mixture of alcohol and ether.
4.1.5 Lubrication
This spectrophotometer is provided with movable internal parts that have been lubricated at the
factory before shipment. If an abnormality is suspected in the internal mechanisms of the
instrument, notify your local service office authorized by Hitachi High-Technologies Corporation.
4-2
4.1.6
If the transmission window plate of the sample compartment is contaminated, remove and clean
it in the following manner.
(2) Detach the black window collar with tweezers or the like.
(4) Clean the window plate using a soft cloth moistened with a 1:1 mixture of alcohol and
ether.
Optical axis
Fig. 4-3 How to Disassemble the Transmission Window Plate of Sample Compartment
4-3
4.2 0%T Measurement
1 Power ON
1
Initializing
2
I
0%T Measurement*
3 Parameter Setting
* 0%T Measurement
This calibration should be done once the unit becomes stable after the power has been switched
on (typically about 2 hours after the power is switched on), or each time prior to sample
measurement (as per the measurement flow above).
The procedure for calibration is as follows.
® MAIL
1) Press I •"» I key.
ENTER
6) Press key; calibration will be complete when the message "Executing ..."
disappears from the screen.
4-4
4.3.1
After turning power on, leave the spectrophotometer for at least 30 minutes until it warms up.
Note that a baseline stability check should be made at least two hours after power is turned on.
The following presents the procedures for respective performance check items.
[ Outline ]
Measure bright lines from the D2 lamp, determine a wavelength indicating a maximum intensity,
and check a deviation of a wavelength indication value from a specific wavelength.
[ Procedure ]
(1) Set up the following conditions.
a) Main Menu
• Measurement mode WL scan
b) Param Setup
• Data mode E(R)
• Start WL(nm) 660
• StopWL(nm) 650
• Upper Scale 100
• Lower Scale 0
• Speed (nm/min) 100
(2) After setting up the above conditions, press the | FORWARD! key (Parameter Setup).
The measurement screen will then appear on the display monitor.
(3) Make sure that no sample is placed in the sample compartment. Then, press the
I START I key. (Measurement is started.)
(4) On completion of the measurement, press | PRINT| key. If necessary, print out the
spectrum after adjusting the scale.
(6) Change the start and stop wavelengths to 490 nm and 480 nm, respectively.
Then, conduct measurement in the same manner.
4-5
4.3 Performance Check
(8) Check that the peak wavelengths thus attained are within the allowable range.
t
0
c
til
Wavelength
Peak wavelength
[ Outline ]
Measure the same bright line from the D2 lamp repeatedly, and determine variations in
wavelength readings.
[ Procedure ]
(1) Set up the following conditions.
a) Main Menu
• Measurement mode WL scan
b) Param Setup
• Data mode E(R)
• Start WL(nm) 660
• StopWL(nm) 650
• Upper Scale 100
• Lower Scale 0
• Speed (nm/min) 100
(2) After setting up the above conditions, press the | FORWARD! key (Parameter Setup).
The measurement screen will then appear on the display monitor.
(3) Make sure that no sample is placed in the sample compartment. Then, press the
I START| key. (Measurement is started.)
(4) On completion of the measurement, press [PRINT] key. If necessary, print out the
spectrum after adjusting the scale.
(6) In the second measurement, shift the wavelength to 1100 nm using the [Go TO WL| key.
Then, repeat steps (3) to (5).
4-6
4.3.3
(7) In the third measurement, shift the wavelength to 190 nm using the |GoTO WLi key.
Then, repeat steps (3) to (5).
(8) Determine the difference between maximum and minimum values in the results of the three
measurements. Check that 1/2 of the difference thus determined is within the allowable
range.
4.3.3 Bandpass
[ Outline ]
The spectral width of monochromatic light emitted from the exit slit of the monochromator is
determined by the wavelength bandpass.
[ Procedure ]
(1) Set up the following conditions.
a) Main Menu
• Measurement mode WL scan
b) Param Setup
• Data mode E(R)
• Start WL(nm) 660
• StopWL(nm) 650
• Upper Scale 100
• Lower Scale 0
• Speed (nm/min) 100
(2) After setting up the above conditions, press the [FORWARD] key (Parameter Setup).
The measurement screen will then appear on the display monitor.
(3) Make sure that no sample is placed in the sample compartment. Then, press the [STARTJ
key. (Measurement is started.)
(4) On completion of the measurement, press |PRINT| key. If necessary, print out the
spectrum after adjusting the scale.
4-7
4.3 Performance Check
(5) Determine a bandpass width at 1/2 of the maximum peak height of the spectrum attained.
Wavelength
Wavelength width (Bandpass)
[ Outline ]
With no sample in the sample compartment, determine a variation range of absorbance values
measured in a short time under constant measurement conditions.
[ Procedure ]
(1) Set up the following conditions.
a) Main Menu
• Measurement mode Time scan
b) Param Setup
• Data mode ABS
• WL(nm) 500
• Upper Scale 0.001
• Lower Scale -0.001
• Scan Time (s) 300
(2) After setting up the above conditions, press the | FORWARD] key (Parameter Setup).
The measurement screen will then appear on the display monitor.
(3) Make sure that no sample is placed in the sample compartment. Then, press the
^UTOZEROl key.
(5) On completion of the measurement, press |PRINT| key. If necessary, print out the
spectrum after adjusting the scale.
4-8
4.3.5
(6) Determine the difference between maximum and minimum photometric values
(peak-to-peak value).
Peak to Peak
300 s
[ Outline ]
With no sample in the sample compartment, determine a variation range of absorbance values
measured in the vicinity of absorbance zero under constant measurement conditions.
Excluding peaks due to one-time noises, determine the difference between maximum and
minimum absorbance values.
[ Procedure ]
(1) Before proceeding to this measurement, wait for at least two hours after power-on.
a) Main Menu
• Measurement mode Time scan
b) Param Setup
• Data mode ABS
• WL(nm) 500
• Upper Scale 0.001
• Lower Scale -0.001
• Scan Time (s) 3600
(3) After setting up the above conditions, press the | FORWARD! key (Parameter Setup).
The measurement screen will then appear on the display monitor.
(4) Make sure that no sample is placed in the sample compartment. Then, press the
lAUTOZERO] key.
4-9
4.3 Performance Check
(6) On completion of the measurement, press | PRINT] key. If necessary, print out the
spectrum after adjusting the scale.
(7) Draw a line across the center of the noise region as shown below, and determine the
difference "D" between maximum and minimum values.
3600 s
[ Outline ]
With no sample in the sample compartment, determine a systematic variation range* of
absorbance values measured in a specified wavelength region.
* The systematic variation range represents a range of variations having reproducibiiity, not
including noises and the like.
[ Procedure ]
(1) Set up the following conditions.
a) Main Menu
• Measurement mode WLsc
b) Param Setup
• Data mode ABS
• Start WL(nm) 950
• Stop WL (nm) 200
• Upper Scale 0.002
• Lower Scale -0.002
• Speed (nm/min) 200
(2) On the condition setup screen, select | FORWARD! key (User 1) for [User Baseline].
(User-baseline determination is started.)
(3) On completion of the baseline determination, press [START] key (Parameter Setup).
The measurement screen will then appear on the display monitor.
4-10
4.3.6
(5) On completion of the measurement, if necessary, print out the spectrum after adjusting the
scale. Draw a line along the center levels of noises as shown below, and determine the
difference between maximum and minimum values.
Then, determine a distance from the midpoint level.
Flatness
200 to 950 nm
4-11
4.4 Troubleshooting
4.4 Troubleshooting
If the instrument does not work normally even after a countermeasure is taken according to the
above troubleshooting table, contact your local sales representative or service office of Hitachi
High-Technologies Corporation.
WARNING
If you touch the internal electric parts of the instrument, you may
receive an electric shock. To prevent this, refer inspection of the
internal parts to qualified service personnel.
4-12
5.1
5. REPLACEMENT OF PARTS
5.1 General
Shown below are the consumable parts to be prepared as spares. For purchase of these parts,
contact your local sales representative or service office of Hitachi High-Technologies
Corporation.
Guaranteed Hours of
Guaranteed Period Part No.
Operation
Wl lamp 1000 hr 6 months after purchased by the user 885-1200
D2 lamp 350 hr 6 months after purchased by the user 2J1-1500
5-1
5.3 Replacement of Light Source Lamp
WARNING
CAUTIONS in replacement of light source lamp
• If you come into contact with the D2 lamp power supply (500 V),
you may incur fatal or serious injury due to an electric shock.
Before proceeding to lamp replacement, make sure that the
power switch of the photometer is turned off and then unplug the
power cord.
• If you touch hot parts of the D2 lamp or Wl lamp, you may get
burnt. After turning the power switch off, wait for at least ten
minutes until the lamp becomes sufficiently cool.
• When mounting a new lamp, wear clean gloves so that your
fingerprints will not be left on the surface of its bulb.
When handing a new D2 lamp, use particular care not to leave
your fingerprints on its protruded part.
• When you check the lighting of the lamp after its replacement, do
not look into the lamp with the unprotected eye at a near position.
After mounting the light source cover, check the lighting of the
lamp from a distant position.
5-2
5.3.1
(1) Turn the power switch off, and unplug the power cord. Then, wait for about 30 minutes
until the lamp part becomes sufficiently cool.
Loosen the setscrew P on the side face of the light source cover, and remove the light
source cover.
0* Setscrew P
When the cover is removed, the lamps appear as shown in Fig. 5-2.
D2 lamp
Wl lamp
Fig. 5-2
5-3
5.3 Replacement of Light Source Lamp
Loosen the retaining spring, and remove the Wl lamp. When mounting a new Wl lamp,
push it securely so that its root part will be seated properly.
Wl lamp
D2 lamp
Connector
5-4
5.3.2
(b) Holding the metallic part of the D2 lamp, turn it counterclockwise. The D2 lamp can
thus be removed.
After lamp replacement, clear the lamp turn-on time indication to zero. For details, refer to Item
3.6.2 in Chapter 3.
WARNING
CAUTION in replacement of light source lamp
No adjustment of the light source position is required in common
practice. If it becomes necessary to adjust the light source position,
carry out the adjustment while observing the cautionary instructions
given below.
1. The Wl lamp and D2 lamp become extremely hot when they are
turned on. Be careful not to touch them.
2. The D2 lamp emits intense ultraviolet rays.
Wear ultraviolet-blocking glasses at the time of lamp check.
5-5
5.3 Replacement of Light Source Lamp
Photometry Bnm
Select Function:_
QON @OFF
Fig. 5-6
Choose |T| D2 lamp and choose \2] OFF. Press | FORWARD!; then D2 lamp light
off out.
5-6
5.3.3
(b) Press the 1 GO TO will key, and set up a wavelength of more than 400 nm.
The light source mirror will be oriented to the Wl lamp, causing its beam to reflect to
the slit. Open the light source cover, and check that the image of the lamp is formed
as shown in Fig. 5-7.
Sidewise Lengthwise
deviation deviation
< Good > < No good >
Fig. 5-7
If the lamp image deviates sidewise, adjust the nut A and screw A in the light source
compartment.
If the lamp image appears on a vertically deviate position with respect to the Wl lamp
position, contact your local service office of Hitachi High-Technologies Corporation.
5-7
5.3 Replacement of Light Source Lamp
(b) Press the |GO TO WL| key, and set up a wavelength of less than 300 nm.
The light source mirror will be oriented to the D2 lamp, causing its beam to reflect to
the slit. Open the light source cover, and check that the image of the lamp is formed
as shown in Fig. 5-8.
If the lamp image appears on a horizontally deviate position, adjust it with the nut B
and screw B. After it is adjusted propelly, turn the nut B tightly.
5-8
5.4.1
5.4 Storage
(1) Turn the power switch off, and unplug the power cord from power outlet.
(1) Store the instrument so that it will not be exposed to a high temperature (60 °C or higher), a
low temperature (-20 °C or lower), a high humidity (80% or higher), and intense vibration.
(3) Take care so that a detrimental gas such as a strong acid/alkali vapor will not intrude into
the instrument.
(4) Before transporting the instrument to a storage place, be sure to fix the light source mirror.
If the instrument is moved with the light source mirror not fixed, a failure may be caused by
impact or vibration. For fixing the light source mirror, use the sponge part indicated in
Item 1.1.
5-9
APPENDIX 1.
The U-1900A spectrophotometer is applicable to absorption analyses of liquid, solid and gaseous
samples in ultraviolet and visible spectral regions.
Assuming that a monochromatic beam with intensity "/0" travels through a liquid having
concentration "c" and optical path length "I" and the intensity of the monochromatic beam
decreases to "/t" as shown in Fig. A1-1, the relationship expressed by Equation (1) can be set up.
I,/I0=10-~t=t (1)
where, "s" is an absorption coefficient that is determined depending on the sample under
measurement.
Equation (1) represents the Bouguer-Beer law, and "t" indicates a transmittance value.
^Transparent cell
Liquid
Monochromatic beam
Concentration
APPENDIX-1
APPENDIX 2. FOR PROPER USE OF SPECTROPHOTOMETER
When selecting a solvent for sample preparation, make sure that the following
requirements are satisfied:
Table A2-1 shows the applicable wavelength ranges of organic solvents to be used in
common applications.
(For details, refer to "Lectures on Experimental Chemistry" Vol. 15, Part I, edited by the
Chemical Society of Japan.)
Wavelength
200 nm 300 nm 400 nm or more
Solvent
J J J
Cyclohexane 200 nm
APPENDIX - 2
APPENDIX 2.
Note that Equations (1) and (2) indicated in APPENDIX 1 are not applicable to analysis of
the following samples:
• Fluorescing samples
• Samples having high turbidity
In measurement of a solid sample such as a glass plate piece, a loss (r) in optical energy
occurs due to reflection on the surface of the solid sample. In this case, the following
equation is given.
where, "r" varies depending on the reflectance of the solid sample under measurement.
APPENDIX-3
INDEX
0%T 3-80,4-4
A
Absorption coefficient APPENDIX-1
Absorption spectrophotometry APPENDIX-1
B
Bandpass 3-84, 4-7
Bouguer-Beer law APPENDIX-1
C
Check of contents 1-4
Cleaning of sample compartment 4-1
Communication connector 2-2
Concentration measurement 3-9, 3-24
Connection of grounding wire 1-8
Consumables 5-1
Control system 2-4
D
Diffraction grating 2-3
E
Error messages 3-99
F
Fuse 1-6
Installation 1-1
Instrument appearance 2-1
K
K factor 3-13, 3-28
L
Light source changeover 2-7
INDEX-1
M
Maintenance 4-1
N
Noise level 2-7, 3-84,4-8
o
Operation panel 2-5
Optical system.. 2-3
P
Photometric accuracy 2-7
Photometric range 2-7
Photometric repeatability 2-7
Photometry 3-3
Power supply voltage 1-2, 1-3
Printer output connector 2-2
R
Ratio mode 3-11
Replacement of light source lamp 5-2
Replacement parts 5-1
S
Scan speed 3-42
Self-diagnosis 2-6
Signal processing 2-4
Spares 5-1
Specifications 2-7
Stability 2-7
Stray light 2-7
T
Troubleshooting table 4-12
V
Validation function 3-84
W
Wavelength accuracy 2-7, 3-84, 4-5
Wavelength calibration 3-79
Wavelength setting repeatability 2-7, 4-6
Wavelength scan 2-7, 3-42
INDEX-2