Umed Pharma Lab.Pvt. Ltd.

An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory


Plot No: SO/All, Sy. No. 334 to 337, Bachupalli (V), QuthubuUapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality*


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location Coating
Sampling Equipment Auto coater
INSTRUMENT CALIBIUTION DETAILS
Name of the Instruments Make Model No/S.N 0 Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755110129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS-
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 ::; -400 C -48.1
2 Suspended Particulate mg/m'' ISO 8573-1,2010 Less than 5 mg/m" 0.2
Matter Class-6
...,
.J Moisture Content ppm v/v EP NMT 67 ppm v/v 09
4 Oil Mist mg/rrr' EP Less than 0.10 mg/m" 0.01
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 11
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

ruFERENCE: After reviewing the above results it is foncluded thatthe quality of


-
cOlI!Pressed
-
air is found-
within specified limits .

Reviewed By

f:~~
.~~t~ucted By
/, ..,..
lie:: H '7.~

\VVrl"
~ C))
.09 ~

~'<~a Pharma Lab Pvt.Ltd For Rene Industries Ltd


Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: 50/AIl, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality=


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location Coating
Sampling Equipment Manual coating machine
INSTRUMENT CALIBRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- PlZ-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :s -40 ° C -50.3
2 Suspended Particulate mg/m'' ISO 8573-1,2010 Less than 5 mg/m" 0.2
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 13
4 Oil Mist mg/nr' EP Less than 0.10 mg/nr' 0.02
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 15
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017


-- -- - -- - ----

INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
within specified limits.

~~~uctedBy Reviewed By

·:10: y. .~
f~~
~Jt; ry; 0)/
y'af~d Pharma Lab Pvt.Ltd For Rene Industries Ltd
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: 501AlI, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality= ~


Customer Details Samgle Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala ,Uganda Date of test Report 21.08.2017
Sampling Location Blister
Sampling Equipment Blister packing machine 1
INSTRUMENT CALIDRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- ~-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 S -400 C -50.2
.)
2 Suspended Particulate mg/m ISO 8573-1,2010 Less than 5 mg/nr' 0.1
Matter Class-6
...,
.J Moisture Content ppm v/v EP NMT 67 ppm v/v 10
.)
4 Oil Mist mg/nr' EP Less than 0.10 mg/m 0.03
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 19
matter

For Moisture Content Gastec Tgbe QC No :61129, Vali_dup to 11/2017 - --- - - -

INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
within specified limits.

~
L=?--~~ucted By Reviewed By
(/-..'1 \~. --

~~~IP~
tl)*J
For Dmed Pharma Lab Pvt.Ltd For Rene Industries Ltd
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABL Accredited Laboratory
Plot No: 50/All, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality*


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
-
Sampling Location - Blister
Sampling Equipment Blister packing machine 2
INSTRUMENT CALIBRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755/1 0 129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTI- 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :s -40 ° C -51.1
2 Suspended Particulate mg/nr' ISO 8573-1,2010 Less than 5 mg/rrr' 0.2
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 08
4 Oil Mist mg/rrr' EP Less than 0.10 mg/m" 0.01
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 21
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

INFERENCE: After reviewing the above results it is concluded that.the.quality of compressed air is found __
within specified limits.

T",~l'lIil~dBy Reviewed By

fir~-O'l
'~ y, ~'l
~~fi ~,'iI
For 1.Jm:e."(f·PharmaLab Pvt.Ltd For Rene Industries Ltd
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: SO/All, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality"


Customer Details Sample Details
M/S Rene Industries Ltd Date of sample collection 15.08.2017
Kampala ,Uganda Date of test Report 21.08.2017
Sampling Location Blister
Sampling Equipment Blister packing machine 3
lNSrRTJ'MENT CALiBRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On -

Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017


Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTI- 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULis
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :s -40 ° C -52.2
2 Suspended Particulate mg/m" ISO 8573-1,2010 Less than 5 mg/rrr' 0.2
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 08
4 Oil Mist mg/nr' EP Less than 0.10 mg/m" 0.01
Viable Particulate
5 CFU/m3 - IS:5402 NMT 100 CFU/ cubic meter 14
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

[l\;J'F'EMMeE:-' 'Afterreviewing the above resultsitisconcluded that the 'quality of compressed-air is found -
within specified limits.

- -

4~
II~.b.LI':-'" ~~~te
n.
dB y Reviewed By

-»For Umed Pharma Lab Pvt.Ltd


~-

For Rene Industries Ltd


Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABL Accredited Laboratory
Plot No: 501AJI, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@urnedlabs.com

TEST CERTIFICATE

"
.... *Determination of Compressed Air .Quality * .c_
Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location Blister
Sampling Equipment Blister packing machine 4
INSTRUMENT CALIBRATION DETAILS
Name of the Instruments Make Model N o/S.N 0 Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755110129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :s -40 ° C -49.9
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/m" 0.1
Matter Class-6
...,
.) Moisture Content ppm v/v EP NMT 67 ppm v/v 10
4 Oil Mist mg/nr' EP Less than 0.10 mg/m" 0.01
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 12
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017


-_ -- - --- -
INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
within specified limits.

;ftt6\f-bA~ ted By Reviewed By

r~ cy .§YI
~~
-

~n*
For"tJiiied Pharma Lab Pvt.Ltd For Rene Industries Ltd
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: 50/A/!, Sy. No. 334 to 337, Bacbupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE
-
*Determination of COll)pressed Air Quality*' r':

Customer Details Sample Details


MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
- -
Sampling Location Blister
Sampling Equipment Blister packing machine 5
INSTRUMENT CALm RATION DETATh8
Narne of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05Tf.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTI- 06/668 09.08.2017 08.08.2018
India
OBT AINE'D TEST :RES'11LTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :S -40 ° C -50.2
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/rrr' 0.2
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 07
4 Oil Mist mg/m ' EP Less than 0.10 mg/m" 0.04
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFUI cubic meter 10
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

rnEEREN.(J.R~fter reviewing the above results it is concluded.that.the.quality of compressed air.is found__


within specified limits.

~i\.!~ -
£~~~ctedBy Reviewed By
{f?: L

~J.1~
o ---/q;;
,;yt-t; ;;
n ..-~ >
For Umed Pharma Lab Pvt.Ltd For Rene Industries Ltd
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: 501AIl, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

~ <s-: ~ *Determination of Compressed Air .Quality=


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date oftest Report 21.08.2019
Sampling Location Blister
Sampling Equipment Blister packing machine 6
INSTRUMENT CALmRA TION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755/1 0129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 061668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 S -40 ° C -52.3
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/nr' 0.2
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 04
4 Oil Mist mg/rrr' EP Less than 0.10 mg/m" 0.01
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFUI cubic meter 13
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017


_
- --- --- --
INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
within specified limits.

roo
~~uctedBy Reviewed By

'/:5: ~
''';';'

~a..
o '))
;-iJ
r-:
'~.:~t . "'y

FO~~"-Pharma Lab Pvt.Ltd For Rene Industries Ltd


Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABL Accredited Laboratory
Plot No: 50IA/1, Sy. No. 334 to 337, Bachupalli (Y), QuthubuUapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality *


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala ,Uganda Date of test Report 21.08.2017
Sampling Location ORS
Sampling Equipment Pouch packing machine
INSTRUMENT CALIDRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 S -40 ° C -53.2
2 Suspended Particulate mg/nr' ISO 8573-1,2010 Less than 5 mg/m' 0.2
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 09
4 Oil Mist mg/rrr' EP Less than 0.10 mg/rrr' 0.01
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 24
matter

For Moistu!~ Content Gastec Tube QC No :§JE~Valid up to 11/2.Q~_7_


INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
within specified limits.

//~J8f~~cted By Reviewed By

For TImed Pharma Lab Pvt.Ltd For Rene Industries Ltd


Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: SOINI, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-S00090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality=


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala ,!lganda - -
Date _Qf test Rep011 21.08.2017 ~
Sampling Location Mono Block
Sampling Equipment Bottle washing machine
. INSTRUME1'qT CALffiRATION DET.A.tL£
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755110129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :::;-40 ° C -50.2
2 Suspended Particulate rng/nr' ISO 8573-1,2010 Less than 5 mg/rrr' 0.2
Matter Class-6
,.,
j Moisture Content ppm v/v EP NMT 67 ppm v/v 05
4 Oil Mist mg/nr' EP Less than 0.10 mg/rrr' 0.03
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 19
matter -

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

~E~NGE: After reviewing the above results it-is concluded that the quality of compressed-air is found-
within specified limits.

- - -

Tes~w'f~By Reviewed By

N~
If;!
~~-~:!J
~-YfA A~j
For tline~tharina
--6

Lab Pvt.Ltd
.- - -

--

For Rene Industries


- - -

Ltd
-
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABL Accredited Laboratory
Plot No: SOIAlI, Sy. No. 334 to 337, BacbupaUi (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality*


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location ~-
Mono- Block
Sampling Equipment Bottle filling machine
INSTRUMENT CALIBRATION UETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755110129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTI- 06/668 09.08.2017 08.08.2018
India
aRT Am{ED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :s -40 ° C -48.9
2 Suspended Particulate mg/nr' ISO 8573-1,2010 Less than 5 mg/rrr' 0.2
Matter Class-6
..,
.) Moisture Content ppm v/v EP NMT 67 ppm v/v 06
4 Oil Mist mg/rrr' EP Less than 0.10 mg/m" 0.01
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 10
matter
-

For Moisture Content Gastec Tube QC No :61129, Valid up to 1112017

INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
within-specified limits> ---- --~---- - - - -------- -

- - -

-
T8~~il.~d By Reviewed By

ft~,
T~
\'0
\~~'!!!r;rl)~~(\1!--
* >~...
For UlirEfd;Pharma Lab Pvt.Ltd --
-

For Rene Industries


-

Ltd
Dmed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABL Accredited Laboratory
Plot No: 501Afl, Sy. No. 334 to 337, BachupaUi (v), Quthubullapur eM), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determinatiop. of Compressed Air Quality=


Customer Details Sam ple Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location Linear
Sampling Equipment Filling Machine
INSTRUMENT CALIBRATION DETAILS
Name of the Instruments Make Model N o/S.N 0 Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 ~ -40 ° C -50.8
2 Suspended Particulate rng/rrr' ISO 8573-1,2010 Less than 5 mg/nr' 0.2
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 09
4 Oil Mist mg/rrr' EP Less than 0.10 mg/nr' 0.04
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 14
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

INFERENCE: After reviewing the above results-it is concluded that tlie quality of compressed air is found -
within specified limits.

-=--
~~~
tI'?"/Test 0\~ cted By Reviewed By

\~\:Y~~~
~""'"
q;f}"* .'17'///
".-.
.. .n'
. -::~.~.".

For Umed Pharma Lab Pvt.Ltd For Rene Industries Ltd


Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: SOIAJi, Sy, No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-S00090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

_ _ ~ *Determination of Compressed Air Quality=


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location Linear
Sampling Equipment Bottle filling machine
INSTRUMENT CALIBRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- PLZ-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 ~ -40 ° C -52.3
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/rrr' 0.2
Matter Class-6
.,.)
Moisture Content ppm v/v EP NMT 67 ppm v/v 12
4 Oil Mist mg/nr' EP Less than 0.10 mg/rrr' 0.03
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 21
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017


- - --- -- - --
INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
within specified limits.

1.~<D~edBy Reviewed By

~u~
~o~ 'c>""'j}!J
VT/rv\:;

For~arma
.
Lab Pvt. Ltd For Rene Industries Ltd
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: 50/AlI, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality=


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala ,Uganda Date of test Report 21.08.2017
Sampling Location Beta-lactarn
Sampling Equipment Bottle cleaning machine
INSTRUMENT CALIBRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- ~-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :s -40 ° C -5l.2
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/rrr' 0.2
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 06
4 Oil Mist mg/nr' EP Less than 0.10 mg/m' 0.01
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 12
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017


-- --- ---- -- - -- -- -

INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
within specified limits.

~uctedBy Reviewed By

f;'i'"
~~
0.. ~ . !l-J
\o . ;...
¥tt-~ .()Y
F~harma Lab Pvt.Ltd For Rene Industries Ltd
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: SO/All, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

=Determination of Compressed Air Quality=


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location Beta-lactam
Sampling Equipment. Auto capsule filling machine
INSTRUMENT CALIBRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.1l.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- ~-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point DC ISO 8573-1,2010 ~ -40 D C -50.1
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/rrr' 0.1
Matter Class-6
.J
'")
Moisture Content ppm v/v EP NMT 67 ppm v/v 05
4 Oil Mist mg/nr' EP Less than 0.10 mg/rrr' 0.01
Viable Particulate
5 CFU/m3 IS:5402 ]\TMT100 CFU/ cubic meter 20
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017


--- - ---- - - - --- --1-

'INFERENCE:After reviewing the above results it is concluded that the quality of compressed air is found
within specified limits.

- Reviewed By
,..ff:$~~~cted By

~
«~
;.~ L~
v-;
I
/fin *.?j
For Olifed Pharma Lab Pvt.Ltd For Rene Industries Ltd
Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABL Accredited Laboratory
Plot No: SO/AJI, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-S00090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality= •


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala ,Uganda Date of test Report, 21.08.2017
Sampling Location Beta-lactam
Sampling Equipment Blister packing machine
INSTRUMENT CALIDRA nON DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755110129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 S -400 C -48.4
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/nr' 0.1
Matter Class-6
3 Moisture Content ppm v/v EP NMT 67 ppm v/v 06
4 Oil Mist mg/rrr' EP Less than 0.10 mg/nr' 0.01
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 24
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

INFERENCE: After reviewing-the above results it is conclude-d 'that tlie quality of compressed air is fOUncr -
within specified limits.

/~ctedBy Reviewed By

Fo~etrPharma Lab Pvt. Ltd For Rene Industries Ltd


Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: 50/A/I, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality=


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location External Department
Sampling Equipment Filling Machine
INSTRUMENT CALIBRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755/10129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :s -40 ° C -50.3
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/rrr' 0.2
Matter Class-6
...,
.) Moisture Content ppm v/v EP NMT 67 ppm v/v 04
4 Oil Mist mg/m" EP Less than 0.10 mg/rrr' 0.03
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 22
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

INFERENCE: After reviewing the above results it is concluded that the quality of compressed air is found
--- -- -
within specified liinits."

---
~~~tedBy Reviewed By

t
'lf
1\.10.
f~
1.0
y)
H

'1-'\
n •. <>fj
.
;-

For tJmed-Pharma Lab Pvt.Ltd


~~--

For Rene Industries Ltd


Umed Pharma Lab Pvt. Ltd.
An ISO 9001 : 2015 Certified Company & NABLAccredited Laboratory
Plot No: 50/AIl, Sy. No. 334 to 337, Bachupalli (V), Quthubullapur (M), Hyderabad,M.M.Dist. Pin-500090. Ph: 040-29804558, Email: sales@umedlabs.com

TEST CERTIFICATE

*Determination of Compressed Air Quality=


Customer Details Sample Details
MIS Rene Industries Ltd Date of sample collection 15.08.2017
Kampala, Uganda Date of test Report 21.08.2017
Sampling Location External Department
Sampling Equipment Bottle filling machine
INSTRUMENT CALmRATION DETAILS
Name of the Instruments Make Model No/S.No Calibrated Due date
On
Digital Electronic Balance Mettler Toledo 1125252637 05.11.2016 04.11.2017
Glass Tube Rota Meter INDUS 236-02 24.04.2017 23.04.2018
Stop Watch Racer 1507201601 25.07.2017 24.07.2018
CO2 Temperature Meter ---- AZ-7755/l0129817 15.09.2016 14.09.2017
Bio Technics
Incubator BTl - 06/668 09.08.2017 08.08.2018
India
OBTAINED TEST RESULTS
S.No Test Parameters Units Guide line as Acceptance criteria Results
per
1 Dew Point °C ISO 8573-1,2010 :S -40 ° C -52.3
2 Suspended Particulate mg/rrr' ISO 8573-1,2010 Less than 5 mg/nr' 0.2
Matter Class-6
..,.)
Moisture Content ppm v/v EP NMT 67 ppm v/v 09
4 Oil Mist mg/rrr' EP Less than 0.10 mg/nr' 0.03
Viable Particulate
5 CFU/m3 IS:5402 NMT 100 CFU/ cubic meter 07
matter

For Moisture Content Gastec Tube QC No :61129, Valid up to 11/2017

- lNFERENCE: After reviewing the above results it is concluded t11Mtlie 'quality of compressed air is fOUlla-
within specified limits.

~~~tedBy Reviewed By

ti:C
f.~
\d1.
~o
H ..
r-:
t - -

~~6o/:·~9
For {Ji'i'J.ec(Pharma Lab Pvt.Ltd For Rene Industries Ltd