CBL-11076-01-A
pag. 1 / 9
Customer Bulletin
Chemistry Systems: Miura, Single and Double Sampler ARM (all versions)
1 INTRODUCTION
This technical bulletin applies to MIURA systems, both Single and Double ARM.
As announced with our previous presentation “Miura Single Sampler ARM and Double Sampler ARM
– New Software Version 2.2.0.10” issued at the launch of the last and newest Miura software version
by our Marketing office, it gives you evidence that next Miura systems will be released in the new
configuration that includes the new software.
2 INTRODUCTION
These new configurations will be on the systems from the following serial numbers:
Double sampler ARM: from s/n A1294.
Single sampler ARM: from s/n A1278.
The monitor of the PC is now on-board, fixed on an appropriate jointed arm integrated with
the instrument cabinet.
Both single and double sampler arm models can be supplied in two main sample tray
configurations:
The new sample tray configurations together with the new software features give the system the
main following advantages:
Small “dilution cups” no longer in use,
Dynamic sample positioning (as in Miura One system),
Sample dilution process improved,
Automatic sample pre-dilution by method,
Automatic sample repetition with post-dilution,
More sample positions,
Fully compatible with MS Windows 7 ™,
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-11076-01-A
pag. 2 / 9
Customer Bulletin
Possibility of upgrading older 50-sample position systems on the field and complete software
compatibility.
New MIURA systems, single and double sampler Arm, adopt a new cuvette Washing Station.
The purpose of this modification is to allow the system to use of a new Cleaner Solution cheaper
and easier to ship. This modification gives the operator the following main advantages:
Larger aspiration channels.
Higher aspiration power per channel.
All needles have self-recovering position (given by springs).
First aspiration channel has separate waste path.
Possibility to use the New Cleaner Solution which is more effective.
Foam formation limiting.
New User Manual and new Quick Start Guide have been issued and the following link is for
downloading the PDF versions; they are valid for any instrument versions:
http://www.logotech-ise.com/files/MProject/Manuals/MNU-10563-01-D_ENG,Miura-PDF.zip
Note: all new features and performances have been detailed into the new User Manual. Refer to
that publication for detailed descriptions on how to use.
In the next paragraph a brief introduction has been given to announce and explain the main
changes.
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-11076-01-A
pag. 3 / 9
Customer Bulletin
4 MONITOR OF PC IS ON BOARD
The monitor of the PC has been integrated on-board to save space on the
working bench.
It has been fixed onto an appropriate support integrated with the instrument
cabinet, so it doesn’t need to be located on the lab bench anymore.
The cabinet is provided with a hub where to fix the monitor support (a jointed
arm).
The monitor must be fixed on the support through the standard VESA fixings.
The joints of the support can be adjusted as preferred by the user.
This support gives full adjustment margin.
Note: the new system packing list will include this new jointed arm – for mounting instructions refer
to the last Miura User Manual.
The system Miura 79-sample positions, single and double ARM, can only use the following supports
indifferently for the samples / standard and calibrators / QCs:
Sample tubes diameter from 12mm to 13mm and height from 75mm to 100mm,
3mL sample cups without any special adapter.
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-11076-01-A
pag. 4 / 9
Customer Bulletin
12mm-13mm OD tubes
Note: the new system packing list will NOT include old dilution cups.
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-11076-01-A
pag. 5 / 9
Customer Bulletin
Part of the process related to the generation of a dilution starting from a concentrated has been
changed:
The sample aspiration timing, when aspirating from concentrated sample, has been
optimized. By avoid quicker depressions a higher precision in bigger volume aspiration of
thicker fluids has been reached: substantially it has been given longer time for sample
aspiration.
Higher diluted volume percentage mixed twice during preparation of the dilution in
cuvette.
Possibility of probe’s outer surface cleaning (in washing sink – w/systemic solution) after aspiration of
the diluted sample from the cuvette for analysis (this feature requires new washing sink tips).
10 MS WINDOWS 7™ COMPATIBILITY
The new software requires a new installer. The new installer has been designed to be fully
compatible with MS Windows 7™ and it has also been optimized in terms of memory used and
operating timing.
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-11076-01-A
pag. 6 / 9
Customer Bulletin
pump is the same controlling the aspiration needle pump and it has been derived through
a wiring jumper.
This choice allows the possibility to treat foamy reagents that, containing detergents, could lead to
the formation of foam that limits the performance.
Note: the new system packing list will include the new Waste Tank wiring.
Tip
Aspiration
2-heads Pump
Needles
Aspiration
Pump
Needle #1 Aspiration
Needle w/springs
Waste connectors
Waste fittings
Fittings support - Top view Fittings support - External view Washing station - Rear view
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-11076-01-A
pag. 7 / 9
Customer Bulletin
11.1 The New Washing Station in Miura single and double sampler arm
The new Washing Station in Miura systems has the following configuration:
5 couples of needles, each composed by one shorter and smaller needle for dispensing
and by one longer and larger one for aspiration (ASPx/DISx);
1 aspirating needle, large size (ASP6);
1 drying tip (white Teflon).
The architecture is shown in the following pictures:
TIP DISP.1/ASP.1
POSITION #8 POSITION #1
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-11076-01-A
pag. 8 / 9
Customer Bulletin
Each of the 5 dispensing peristaltic pumps is directly connected to only one dispensing needle
(from position 1 to position 5 - short ones) of the washing station.
One additional peristaltic pump is on the aspiration needle in position 1; it aspirates reaction fluids
limiting the foam generation (and then the impedance) along the waste path to enhance the
flow.
The double-heads (single motor) diaphragm pump is connected in parallel to 5 of the 6 aspirating
needles (long and smaller ones): one head aspirates in parallel from needle 2 and 3, the other
aspirates from needles 4, 5 and 6. The single-head diaphragm pump is dedicated to the drying tip
only.
During the washing cycle all cuvettes progressively run under each of the needles from position 1
up to position 8, moving one position on at any machine cycle. During the washing cycle,
composed from a descent phase, a waiting phase and a rising phase, the needles, placed in the
different positions, behave as follow:
Positions 1 (couple of needles): the cuvette is fully discharged of the reaction fluid
and then filled with Systemic Solution; solution level
left is about 450ul;
Positions 2 (couple of needles): the cuvette is fully discharged and then filled with
Cleaner Solution; solution level left is about 400ul;
Positions 3÷4 (couple of needles): the cuvette is fully discharged and then filled with
Systemic Solution; solution level left is about 450ul;
Position 5 (couple of needles): the cuvette is fully discharged and then filled with
Systemic Solution; solution level left is about 450ul. It
is optically measured and checked (transparent or
zero-value updating);
Position 6 (single needle): the cuvette is fully discharged;
Position 7 (no needle): no actions;
Position 8 (tip): the cuvette is dried from residual liquid.
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-11076-01-A
pag. 9 / 9
Customer Bulletin
OUTLET
INLET
Note: The new system packing list will include this new Cleaner Solution.
The previous Cleaner solution can be anyway used without any performance degradation in
systems with the new Washing Station.
NOTE: the new Cleaner Solution can be used also into older systems only after special upgrade of
the instrument. This special retrofit upgrading the system will be described in a forthcoming
dedicated technical bulletin that will be sent soon after this one.
Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.