Professional Documents
Culture Documents
9 Laminar Flowmeters FI
Design Temperature Up to 300°F (150°C) normally, but can be higher with special designs
Material of Construction Stainless steel, aluminum, or any alloy available in small bore tubing
Inaccuracy 0.5% to 1% of actual flow for commercial gas flow elements, if calibrated
Costs A 1/2-in. (13 mm) stainless-steel laminar flow element costs $700, a 2-in. (50 mm)
unit costs $1500, and a 16-in. (300 mm) all-stainless unit costs $11,000. The differential-
pressure readout devices are additional to the above element costs.
Laminar flowmeters fill a special need in flow measurement is theoretically linear with pressure drop, in practice, some
where the requirements might include low to extremely low nonlinearities are often encountered. In most cases, these are
flow rates, linear calibration and low noise, the ability to of little consequence.
measure high-viscosity liquids, or steady low-flow repeat- The theory for laminar flowmeters is based on the
ability and control accuracy. Laminar flowmeters are Hagen–Poiseuille Law for laminar flow and the Reynolds
intended for very low flow rates where other types of meters number as a means of defining the type of flow. Both are
are either marginal in performance or cannot be used at all. required to investigate and design a laminar flow element.
Laminar flowmeters can be constructed by various methods, More detailed explanations and discussions of theory can be
but the most common is with capillary tubes. Hence, the found in any standard textbook on fluid mechanics.
terms laminar flowmeter and capillary flowmeter are virtually
synonymous. Proprietary commercial units use other matrix
shapes and are intended for use with gases (Figure 2.9a). THEORY
Where gas is metered, it is preferable to use calibrated com-
mercial units instead of undertaking the design of a laminar Fluid flow in pipes and tubes is characterized by a nondi-
flowmeter. mensional number called the Reynolds number (Re). Up to
The flowmeter consists of the laminar flow element and approximately Re 2000, the flow is called laminar, viscous, or
a differential-pressure measuring instrument. While the flow streamline. Above 10,000, the flow is called fully developed
201
© 2003 by Béla Lipták
202 Flow Measurement
FIG. 2.9a
The laminar flowmeter and its matrix element with miniature triangular duct passage with under 0.1-mm effective diameters. (Courtesy of
Meriam Instrument Div. of Scott Fetzer Co.)
where where
Re = Reynolds number SG = specific gravity relative to air (Table 2.9b)
ρ=
3
density (lb/ft ) at flowing temperature P = flowing gas pressure in inches of mercury absolute
Q= flow rate (gal/min) ∆P = differential pressure in inches of water
D= internal tube diameter (in.) mµ = viscosity of the flowing gas in micropoise
µ= viscosity of flowing temperature (centipoise) (Table 2.9b)
W= flow rate (lb/h)
Hagen–Poiseuille Law
For gas flow,
Once the tube inside diameter required to give laminar flow
according to the Reynolds number calculation has been
6.32 ρ Q 6.32W defined, the length of the capillary has to be determined to
Re = or Re = 2.9(2)
Dµ Dµ design the laminar flowmeter system. These equations are as
follows.
where
For liquid flow,
ρ = density at standard conditions (lb/ft )
3
where Capillary
L = length of tube (in.) Temperature
∆P = differential pressure drop (in water) Control
D = tube internal diameter (in.)
µ = viscosity at flowing temperature (centipoise) Low
ρ
3
= density at flowing temperature (lbm/ft ) Pressure
Q = flow rate (gal/min) Steam
W = flow rate (lbm/h) Overflow Jet
to Drain Mixer
fix viscosity will also fix density. With some fluids, 2.9(7)
cooling may be required instead of heating, but the
overall principle is the same. ( L / D) Re = (15.7 / 0.1235)/128 = 0.993 2.9(8)
This is an easy length to work with in fabricating a meter When the fluid exits the capillary, the flow path enlarges.
element and a constant temperature bath, and it looks like a If the piping is similar to that described under inlet loss, the
reasonable design based on the criteria. loss can be calculated by
5.6 × 10 −7 W 2
∆Pe = 2.9(10)
ERROR SOURCES D4 ρ
Changes in viscosity and density can result in flow measure- Entrance loss occurs in addition to the normal capillary
ment errors. Viscosity changes in liquid as a result of tem- pressure drop in the initial fluid path distance or, to state it
perature can be substantial, while density changes are more in another way, for a short distance the pressure drop is higher
2,3
moderate. With gases, the reverse is usually true, with tem- than that predicted by the Poiseuille Equation. The addi-
perature having more influence on density and less on vis- tional loss is due to the work expended in the formation of
cosity. The need for careful control of the operating temper- the parabolic velocity distribution profile characteristic of
ature to minimize these effects must be emphasized. laminar flow. It can be expressed in terms of an equivalent
From Equations 2.9(4) and 2.9(5), it can be seen that length of capillary, Leq, added to that calculated by the Poiseuille
internal diameter of the tube is very important, because it is equation. Refer to Figure 2.9d for determining the Leq.
multiplied to the fourth power. While high-quality tubing will The following equation can be used for the pressure drop:
be very close to published specifications, manufacturing tol-
erances will result in variations from these dimensions, both
laterally and longitudinally. If the actual effective internal 1.96 × 10 −7 W 2
∆Pen = 2.9(11)
diameter of the capillary tube differs by 1% from the value D4 ρ
used in the calculation for a given ∆P, an error of about 4%
will result. Therefore, the laminar flowmeter should be cali- Table 2.9e can be used as a quick guide for judging the
brated on a known fluid before use, and appropriate design design factors that will minimize overall entrance effects. For
adjustments should be made as necessary. the conditions given in the table, the error involved will be
To measure the true capillary differential pressure drop less than 1%. In general, the effect of all of the above errors
according to the Poiseuille equation, it would be necessary will be minimized if the Reynolds number is low, the laminar
to put the pressure taps into the capillary at the calculated L flow element is long, and the pressure drop is high. The overall
dimension. This is impractical because of the small tubing.
A pressure tap must be perfectly flush with the inside of the
tube and must be clean with no burrs or other projections Leq /D
into the tube. Otherwise, considerable differential-pressure
measurement error will result. Using practical methods of Reference 3
0.40
constructing a capillary flowmeter, there are three additional
sources of pressure drop in addition to the capillary loss.
These are all additive and will give a greater indicated pres- 0.35
sure drop than the capillary flow alone. These three sources
of error are inlet loss, exit loss, and capillary entrance loss. 0.30
These losses also contribute to nonlinearity. Recommended for Recommended for
Re < 800 Re > 1000
There is very little loss from the entrance fitting into the 0.25
capillary tube if laminar flow conditions exist. But if the
piping cavity ahead of the capillary is extremely large relative
0.20
to the capillary (approximating a reservoir) and the fluid
velocity is thus extremely low (approaching zero), there can
1
be an inlet effect and pressure loss. This is a result of the 0.15
100 Re 1000 10,000
sudden contraction from the large reservoir to the small tube
bore, forming a bell-mouth shape approach flow. This loss FIG. 2.9d
can be expressed as Equivalent length of capillary (Leq).
2.8 × 10 −7 W 2
∆Pi = 2.9(9)
D4 ρ TABLE 2.9e
L/D Ratio to Minimize Entrance Effect
Re 10 50 100 500 1000 2000
This equation is derived from Bernoulli’s equation for flow
out of a reservoir. L/D> 15 75 150 750 1500 3000
8 0.10
CONCLUSION
0.08
7
0.06 Laminar flowmeters are highly useful in measuring low flow
D
Dc rates of liquids and gases. Design of the elements is based
0.04
6 on the use of the Reynolds number and Poiseuille’s law.
C Design for most units is relatively simple, but fabrication of
5 0.02 a complete unit and system can be complex. Simple capillary
units can be fabricated by the user, but most require manu-
4 facturers’ skills and design knowledge.
It is highly recommended that the final system be cali-
3 brated with the same type of fluid as the fluid upon which
the sensor will operate, such as air or nitrogen for gas services
2 and water for liquid services. After such calibration, the con-
version is easily made to the actual fluid. The critical con-
1 sideration is to calibrate the unit under conditions that will
100 1000 10,000
NRe approximate the actual in-service Reynolds number of the
application. Some sources of calibration services, other than
FIG. 2.9i the manufacturers, are the National Institute of Standards and
Correction factors for coiled capillary flowmeters (data adapted
Technology (NIST), Edison ESI, and the Colorado Engineer-
from reference 5).
ing Experiment Station, Inc. (CEESI).
where
L = capillary length (ft) References
2
∆P = pressure drop (lbf/ft )
2
gc = gravity constant 32.17 (ft/sec ) 1. Kreith, F. and Eisenstadt, R., Trans. ASME, 1070–1078, 1967.
D = capillary internal diameter (ft) 2. Willoughby, D. A. and Kittle, P. A., Industrial and Eng. Fundamentals,
6(2), 304–306, 1967.
ρ = fluid density (lbm/ft) 3. Rivas, M. A. Jr. and Shapiro, A. H., Trans. ASME, 489–497, April 1956.
V = fluid velocity (ft/sec) 4. Greef, C. E., and Hafckman, J. R., ISA J., 75–78, August 1965.
5. Powell, H. N. and Browne, W. G., Rev. Sci. Instr., 28(2), 138–141, 1957.
Fain, D. E., Calibration of a laminar flowmeter, in Flow, Its Measurement Roth, R. E., Four weeks to fill a bucket, Instrumentation, 7(6).
and Control in Science and Industry, Volume 2, Instrument Society of Spitzer, D. W., Flow Measurement, 2nd ed., ISA Press, Research Triangle
America, Research Triangle Park, NC, 1981. Park, NC, 2001.
Gann, R. G., J. Chemical Ed., 51(11), 761–762, 1974. Thomas, D. L., Laminar flow elements, Control, March 1991.
Hughes, R. A., New laminar flowmeter, Instrum. Control Syst., April 1962. Todd, David, A., A universal calibration curve for laminar flowmeters, in
Instrum. Control Syst., 75–76, 1976. Flow, Its Measurement and Control in Science and Industry, Volume
Lipták, B. G., Flow measurement trends, Control, June 2000. 2, Instrument Society of America, Research Triangle Park, NC, 1981.
Mahood, R. F. and Littlefield, R., private communications, March 1952. Weigand, J. and Lombardo, L., The use of laminar flow element in comput-
Polentz, L. M., Capillary flowmetering, Instrum. Control Syst., April erized flow measurement, 1989 ISA Conference, Paper #89–0002.
1961. Welch, J. V., Trends in low gas flow metering, InTech, February 1991.