Professional Documents
Culture Documents
a r t i c l e i n f o a b s t r a c t
Article history: A manual method is described for the determination of nitrate and nitrite in milk and milk powders that
Received 5 April 2013 is intended to provide an alternative to conventional manual methods accomplished by cadmium
Received in revised form reduction. Reduction of nitrate is performed in solution utilising vanadium (III) and quantitation ach-
5 July 2013
ieved by concurrent reaction with Griess reagent. Performance data are acceptable in terms of precision
Accepted 22 August 2013
and accuracy, repeatability being about 6% and intermediate precision at 8% for both analytes, providing
the limit of detection is not approached. Limit of quantitation is 0.1 mg kg1 for both analytes.
Ó 2013 Published by Elsevier Ltd.
biological samples that can use small reagent volumes (Bories & 2. Materials and methods
Bories, 1995; Jobgen, Jobgen, Li, Meininger, & Wu, 2007).
The reduction of NO
3 to NO2 can be avoided using ion- 2.1. Equipment
chromatographic methods to separate NO
2 and NO3 which are
then measured individually by direct UV detection or conductivity Spectral scans and absorbance readings were taken with Varian
(Jobgen et al., 2007; Reece & Hird, 2000). Such methods have been Cary 50 spectrophotometer (Varian, Palo Alto, USA) with 1.0 cm
successfully deployed and validated internationally for foods and polymethyl methacrylate (PMMA) or polystyrene (PS) macro cu-
biological samples, although there remains significant opportunity vettes (Brand GMBH, Wertheim, Germany), and also equipped with
for interference because of the low wavelength necessary to detect a sipper flow-through cell and a fibre-optic coupler.
NOx (BS EN, 1997a, 2005b; Jedli ckova, Paluch, & Alusik, 2002). Reactions were performed in a Grant circulating water-bath
Another ion-exchange approach has been to perform NO 3 reduc- with GD100 controller (Chelmsford, UK) and suitable stainless
tion in post-column mode to maintain the sensitive and selective steel rack to take 20 mL (approx. volume) glass vials. Various
detection modes available with the NO 2 ion. For example, temperatures were used, the chosen setting being at 60 C 2 C.
Lookabaugh and Krull (1988) used photochemical reduction and Incubation at 25 C or 40 C within spectrophotometer cuvettes
electrochemical re-oxidation of NO 2 to improve detection, while were performed in a Contherm Series 5 (Lower Hutt, New Zealand)
Gapper, Fong, Otter, Indyk, and Woollard (2004) used cadmium convection oven.
reduction and post-column addition of Griess reagent to allow Centrifugation was performed at about 300 g (w1000 rpm)
colorimetric detection at 540 nm. Another HPLC technique has with a Gerber Instruments (Effretikon, Switzerland) centrifuge
been reported to combine pre-column enzyme reduction and re- fitted with a 28 mm radius rotor and buckets to carry 50 mL
action of NO2 with 2, 3-diaminonapthalene to a stable fluorescent polypropylene tubes (Tarson 500040, Kolkata, India). Reagent and
2,3-napthotriazole (Jobgen et al., 2007). A successful post-column sample dispensing were performed using 1 mL and 5 mL variable-
reduction strategy involved addition of trivalent vanadium to the volume pipettors, sourced locally through Labserve (from Fisher
column eluent, which is followed by Griess reagent (Casanova, Scientific, Vantaa, Finland). Nylon or cellulose acetate 0.45 mm high
Gross, McMullen, & Schenck, 2006). The principal advantage of performance liquid chromatography filters, with 3 mL syringes
vanadium (III) reduction is that it occurs in the acidic solution were obtained from various vendors for sample clarification. Some
compatible with the Griess reaction, the principle difficulty being filters required a prewash with 0.1 M HCl before use to remove NO x
the need for elevated temperatures and inert post-column equip- residues.
ment to prevent corrosion of stainless steel parts by hydrochloric
acid. 2.2. Reagents
Vanadium (III) has also been exploited as reductant in non-
chromatographic assay formats. Thus, Braman and Hendrix Incoming water was purified by commercial-scale deionisation
(1989) analysed NO x in water and various biological samples units with separate cation- and anion-exchange resins. This water
including foods by reduction to nitric oxide (NO) and detection by was then reduced to 18 MU resistivity by a Barnstead Epure system
highly sensitive chemiluminescence. Individual samples were with in-line charcoal filter to ensure the absence of NO x
placed into the reduction flask without prior clean-up, the liberated contamination.
NO gas being removed by flow of helium to the detector. Similarly, Hydrochloric acid, 20%, was prepared by careful measurement,
Baskic, Jovanovic, Jakovljevi
c, Delibasi
c, and Aresenjevic (2005) in measuring cylinders, of 1080 mL of concentrated hydrochloric
used vanadium (III) to measure NO
x but captured the NO2 as a acid (37%, w/w). This was transferred to a 2 L volumetric flask and
diazo compound using Griess chemistry, avoiding the need for gas made to volume with water. All procedures were performed in a
handling associated with chemiluminescence but enhancing the fume cupboard.
need to control inadvertent loss of the NO 2 to NO. Miranda, Espey, To prepare 0.1%, w/v, vanadium trichloride (VCl3) solution,
and Wink (2001) successfully measured NO
3 and NO2 in clinical approximately 0.10 g of vanadium (III) chloride (Aldrich 20,827-2)
samples using trivalent vanadium reduction and Griess reaction in was dissolved in 100 mL 20% HCl. The green solution was stable for
microtitre wells and noted the better performance of HCl acidifi- many weeks under refrigeration as shown by its absorbance near
cation compared to H3PO4. Beda and Nedospasov (2005) modified 400 nm. Although some oxidation occurred over time, as revealed
the microtitre assay to overcome problems associated of low NO 3 by a lighter colour, the reagent remained functional as it was pre-
in the presence of high NO 2 , and described the preparation of pared in excess concentration. For enhanced stability, however, the
trivalent vanadium from V2O5 and magnesium and the importance solution was purged with nitrogen gas. Vanadium trichloride and
of leaving some tetravalent vanadium in the reaction mixture to its solutions should be handled with care due to their corrosive
improve analytical precision. The reaction was described by Equa- nature.
tion (1). Carrez solutions were prepared as follows from Merck reagents
(Darmstadt, Germany): 30.0 g potassium hexacyanoferrate (II) tri-
hydrate was dissolved in water and diluted to 200 mL (Carrez I).
2V3þ þ NO þ
3 þ 2H ¼ 2V
4þ
þ NO
2 þ H2 O (1)
This solution was stable several months in the absence of light,
Doane and Horwath (2003) reported a manual single-reagent preferably under refrigeration. Zinc acetate dihydrate (46.0 g) was
procedure using trivalent vanadium for NO 3 testing of water dissolved in water and diluted to 200 mL (Carrez II). This was stable
samples and extended to a number of other matrices. indefinitely, particularly under refrigeration.
The aim of the present study was to evaluate the potential of Individual Griess reagents were prepared separately from Merck
vanadium (III) chemistry to analyse NOx in dairy products using reagents (Darmstadt, Germany) as follows: 5 g of sulfanilamide
a manual, non-chromatographic technique. In this way, tradi- (Merck 1.08035) was dissolved in 10 mL concentrated hydrochloric
tional reduction of NO3 with cadmium, zinc or enzymes can be acid and made to 500 mL with water. This was stable for several
avoided, thereby simplifying the analysis. The reduction and months at 4 C. 0.5 g N-1-naphthyl-ethylenediamine dihydrochloride
reaction procedures have been optimised, performance param- (Merck 1.06237) was dissolved in water and diluted to 250 mL. This
eters estimated, and the method confirmed to be a useful routine solution was stored up to one month in a darkened bottle, avoiding
alternative for compliance testing of NO
x in dairy laboratories. exposure to light. The combined Griess reagent was prepared just
90 D.C. Woollard, H.E. Indyk / International Dairy Journal 35 (2014) 88e94
prior to use by mixing 100 mL sulfanilamide solution and 20 mL N-1- addition, since heating was not required, the reactions were
naphthyl-ethylenediamine dihydrochloride solutions. conveniently performed directly within disposable cuvettes,
Stock standard solutions (1 mg mL1) were prepared from pre- namely 1.0 mL of each sample filtrate, standard or blank, was mixed
dried chemicals (110 e120 C to constant weight) sourced from with 1.0 mL Griess reagent and 1.0 mL hydrochloric acid. After 10e
BDH Chemicals. Potassium nitrate (0.1629 g), containing 100 mg 15 min, the absorbance could be measured at 530 nm against water.
NO3 , was weighed quantitatively into a 100 mL volumetric flask
and diluted with water. Similarly, 0.1850 g potassium nitrite, con- 2.6. Calculations
taining 100 mg NO 2 , was weighed quantitatively into a 100 mL
volumetric flask and diluted with water. These were stored in a From the two standard graphs, best linear fits were calculated
refrigerator and replaced monthly. For longer storage, the stock and used to determine the NOx concentrations in each unknown
solutions were subdivided and frozen (20 C). Alternatively, sample in mg kg1 using Equation (2):
standards can be prepared from sodium salts using 0.1371 g and
0.1500 g of sodium nitrate and sodium nitrite, respectively. ½Absorbance of Sample Sample Blank
mg=kg ¼
Intermediate NO 1
3 and NO2 standards (10 mg mL ) were pre- Slope of Calibration Graph
pared daily by dilution of 1.00 mL of each stock solution into DF
separate 100 mL flasks and made to volume. These were further (2)
Weight Sample ðgÞ
diluted in 10 mL volumetric flasks to generate six-point calibration
curves, including blanks, in the range 0e1 mg mL1. Dilution factor (DF) during sample dissolution and clarification
was usually 19 (ie 15 mL water plus 2 mL of each Carrez solution).
2.3. Sample preparation Dilutions during colorimetry were the same for standards and
samples.
Milk powder samples, approximately 1 g, were weighed to four The concentrations of NO 3 were determined by difference (with
decimal places into 50 mL polypropylene centrifuge tubes. Fifteen molecular weight correction) of total NO
x and NO2 (Equation (3))
millilitres NO3-free deionised water was added, the tubes capped
mg mg mg
and vortex mixed to dissolve. If necessary, the extracts were NO3 ¼ NOX 1:35* NO2 (3)
warmed to about 40 C to achieve full dissolution. In the case of kg kg kg
liquid milk, 15.0 mL was added to the centrifuge tube. A quality
control milk powder sample with known NO x concentrations was
2.7. Optimisation
included to confirm run-to-run precision.
Two mL Carrez I was added to the sample followed by 2.0 mL The reactions with Griess reagents were well established from
Carrez II to precipitate protein. The well-mixed extract was left 15e literature reports and previous experience. The reduction step by
30 min to facilitate full precipitation, and then centrifuged at about vanadium was optimised for reagent strength, time and temperature.
300 g to produce a clear supernatant. If a clear supernatant was
not achieved, then a further 2 mL of each Carrez solution was added 2.8. Validation
and centrifugation repeated.
The clear upper phase was filtered through a 0.45 mm membrane The optimised procedure was validated for intra-laboratory
into a suitable glass vial or test-tube ensuring sufficient volume was repeatability by replicate analysis of single samples. In addition,
collected, sometimes requiring a second filter. A duplicate blank was duplicate data of different samples over many months of testing
prepared similarly, by omitting the sample but including all reagents. yielded further robust intermediate precision data. Intermediate
precision was also determined from control charts of three nitrite-
2.4. Nitrate determination (NO
3 þ NO2 ) positive samples obtained from six months of testing with multiple
analysts. Reproducibility trials with other laboratories were not
Two mL of each sample extract, standard and blank were conducted.
pipetted into separate glass vials. To each was added 2.0 mL of Accuracy was estimated by spiked recoveries and by comparison
Griess reagent, followed by 2.0 mL of vanadium chloride solution. with conventional cadmium-reduction methods performed within
The vials were capped tightly to prevent evaporation and incubated a New Zealand inter-laboratory proficiency program.
in a 60 C water bath for 40e45 min. Progress of the reaction could Limit-of-detection estimates were performed by dilution of
be seen by a developing red colouration. standards until the absorbances were no longer distinguishable
The vials were then mixed by inversion, cooled to room tem- from spectrophotometer noise (3s). In addition, samples with un-
perature in cold water and portions transferred to disposable 1 cm detectable NO x were repetitively analysed to determine data
disposable cuvettes. The absorbances were read at 530 nm against scatter around zero concentration.
water. For large sample numbers, a sipper cell was used with low-
volume cell or a fibre-optic probe for the Cary 50 spectrophotometer. 3. Results and discussion
An alternative procedure was also evaluated whereby reduced
volumes were pipetted directly into spectrophotometer cuvettes, 3.1. Colorimetric reaction
with lower temperatures necessary during incubation to prevent
loss of volume by evaporation. Thus, 1.0 mL sample extract, stan- To optimise the reduction of NO
3 to NO2 a range of temperature
dard or blank, 1.0 mL mixed Griess reagent and 1.0 mL vanadium conditions were evaluated. At room temperature (20 C) the
chloride solution were mixed in a capped cuvette and incubated at reduction was slow, whereas near-boiling temperatures facilitated
either 40 C for 3 h, or overnight at 25 C. a rapid reaction but also significant degradation of the coloured
Griess azo product. A temperature of 60 C was selected and pro-
2.5. Nitrite determination (NO
2) vided an acceptable compromise between colour stability and re-
action time. Fig. 1 illustrates the change in absorbance with time at
Nitrite levels were determined in the same way except the va- 20 C and 60 C for a milk powder extract, and Table 1 gives the
nadium chloride solution was replaced with hydrochloric acid. In time to optimum signal at other temperatures. The maximum
D.C. Woollard, H.E. Indyk / International Dairy Journal 35 (2014) 88e94 91
always near zero (intercept/slope < 0.001) but small positive in-
tercepts (typically 0.02e0.03) were common for the NO 3 blanks.
Potential NO 3 sources in the calibrations were water, HCl and VCl3
but these reagents were also used at the same quantity in all
samples, so intercepts were unimportant. Sample blanks compen-
sated for contamination from all sources including Carrez solutions,
disposable equipment and the 0.45 mm nylon filters, and were
typically 0.003e0.005 absorbance units for NO 2 and 0.05e0.07 for
NO 3 . Filters presented a situation of variable blank because a new
one was used for each sample. Filters with low endogenous NO 3
were selected to avoid the need for each to be washed with HCl and
air-dried prior to use. To obtain best precision, sample blanks were
performed in replicate.
Repeatability relative standard deviation (RSDr) estimates were
determined by same-day analyses (n ¼ 6) of various milk powders
Table 2
Intermediate precision estimates (RSDiR) for four typical milk powders.
Table 3 Acknowledgements
Recovery estimates of nitrate and nitrite added to a UHT liquid milk and ready-to-
drink (RTD) infant formulation.
The authors wish to acknowledge the technical assistance of
Drink Nitrate (mg L1) Nitrite (mg L1) Mayson Kay of NZ Laboratory Services, Auckland, NZ.
type
Added Found Recovery Added Found Recovery
NO
3 NO3 (%) NO
2 NO2 (%) References
UHT 0 4.91 e 0 0.62 e
10.00 15.13 102.2 1.00 1.69 107.0 Baski c, D., Jovanovi c, P., Jakovljevic, V., Delibasi
c, E., & Aresenjevi
c, N. (2005).
20.00 24.66 98.8 3.00 3.45 94.3 Spectrophotometric method for simultaneous detection of nitrate and nitrite.
40.00 44.01 97.8 5.00 5.62 100.0 Medicus, 6, 49e52.
Beda, N., & Nedospasov, A. (2005). A spectrophotometric assay for nitrate in an
60.00 65.43 100.9 10.00 11.12 105.0
excess of nitrite. Nitric Oxide, 13, 93e97.
RTD 0 6.62 e 0 0.22 e
Bories, P. N., & Bories, C. (1995). Nitrate determination in biological fluids by an
10.00 16.12 95.0 1.00 1.20 98.0 enzymatic one-step assay with nitrate reductase. Clinical Chemistry, 41, 904e
20.00 25.98 96.8 3.00 3.34 104.0 997.
40.00 44.99 95.9 5.00 5.04 96.4 Braman, R. S., & Hendrix, S. A. (1989). Nanogram nitrite and nitrate determination in
60.00 66.57 99.9 10.00 10.12 99.0 environmental and biological materials by vanadium (III) reduction with
Average ¼ 98.4 Average ¼ 100.5 chemiluminescence detection. Analytical Chemistry, 61, 2715e2718.
BS EN. (1997a). Foodstuffs: Determination of nitrate and/or nitrite content. HPLC/IC
method for the determination of nitrate content of vegetables and vegetable
products. BS EN 12014-2. Brussels, Belgium: European Committee of
Concentrations of 0.03 mg kg1 in milk powders were therefore Normalisation.
deemed at the limit of detection, yielding a limit of quantitation of BS EN. (1997b). Foodstuffs: Determination of nitrate and/or nitrite content. Enzy-
matic determination of nitrate content of vegetable-containing food for babies
0.1 mg kg1. and infants. BS EN 12014-5. Brussels, Belgium: European Committee of
Maximum contamination levels (MCLs) vary between country Normalisation.
and matrix but for the New Zealand dairy industry the self-imposed BS EN. (1998). Foodstuffs. Determination of nitrate and/or nitrite content. Continuous
flow method for the determination of nitrate content of vegetables and vegetable
limits for milk powders are 50 mg kg1 NO 3 and 1 mg kg
1
NO2. products after cadmium reduction. BS EN 12014-7. Brussels, Belgium: European
The current method achieves these sensitivities and so is adequate Committee of Normalisation.
for routine quality control purposes and is likely to meet the future BS EN. (2005a). Foodstuffs: Determination of nitrate and/or nitrite content e Part 3:
Spectrophotometric determination of nitrate and nitrite content of meat products
requirements of the milk and milk powder industries. As new-born after enzymatic reduction of nitrate to nitrite. BS EN 12014-3. Brussels, Belgium:
infants are especially susceptible to the detrimental effects of these European Committee of Normalisation.
contaminants, upper limits applied to the production of milk-based BS EN. (2005b). Foodstuffs: Determination of nitrate and/or nitrite content. Ion-
infant formulae are generally lower at 25e40 mg kg1 for NO
exchange chromatographic (IC) method for the determination of nitrate and ni-
3 and trite content of meat products. BS EN 12014-4. Brussels, Belgium: European
0.5 mg kg1 for NO 2 . These requirements are also met by the cur- Committee of Normalisation.
rent method. BS EN. (2008). Milk and milk products: Determination of nitrate content e Method by
enzymatic reduction and molecular-absorption spectrometry after Griess reaction.
As a consequence of analyte instability, no food-based reference
BS EN ISO 20541. Brussels, Belgium: European Committee of Normalisation.
materials are available that contain certified nitrate or nitrite levels. Casanova, J. A., Gross, L. K., McMullen, S. E., & Schenck, F. J. (2006). Use of Griess
The accuracy of the proposed method has therefore been evaluated reagent containing vanadium (III) for post-column derivatization and simulta-
by comparison with conventional methods and spiked recovery. neous determination of nitrite and nitrate in baby food. Journal of AOAC Inter-
national, 89, 447e451.
Thus, over the course of two years, the proposed method was tested Doane, T. A., & Horwath. (2003). Spectrophotometric determination of nitrate with
within a New Zealand national proficiency scheme (Fig. 4). NO 2 a single reagent. Analytical Letters, 36, 2713e2722.
data was centrally located among all participants because the Ellis, G., Adatia, I., Yazdanpanah, M., & Makela, S. K. (1998). Nitrite and nitrate an-
alyses: a clinical biochemistry perspective. Clinical Biochemistry, 31, 195e220.
chemistry of the proposed manual NO 2 method was equivalent to Fox, J. B. (1979). Kinetics and mechanisms of the Griess reaction. Analytical Chem-
the automated conventional methods used by other laboratories. istry, 51, 1493e1502.
However, NO 3 results derived from the proposed V
3þ
reduction Gapper, L. W., Fong, B. Y., Otter, D. E., Indyk, H. E., & Woollard, D. C. (2004).
Determination of nitrite and nitrate in dairy products by ion exchange LC with
method were typically higher than median data from conventional spectrophotometric detection. International Dairy Journal, 14, 881e887.
reduction methodology, although the Z statistic was generally Griess, J. P. (1879). Bemerkungen zu der Abhandlung der H.H. Weselsky und Ben-
acceptable (<2). A UHT liquid milk sample and a ready-to-feed edikt, ueber einige Azoverbindungen. Berichte der Deutschen Chemistry Gesell-
schaft, 12, 426e428.
infant formula, both in 250 mL tetrapak cartons, were spiked by Hord, N. G., Tang, Y., & Bryan, N. S. (2009). Food sources of nitrates and nitrites: the
pipetting standard solutions to achieve NO
2 and NO3 concentra- physiological context for potential health benefits. American Journal Clinical
tions in representative ranges. Both analytes returned acceptable Nutrition, 90, 1e10.
IDF. (2004a). Milk and milk products e Determination of nitrate and nitrite contents e
recoveries as shown in Table 3, with no evidence of positive bias
Part 1: Method using cadmium reduction and spectrometry. ISO 14673-1:2004,
for NO3. IDF 189-1. Brussels, Belgium: International Dairy Federation.
IDF. (2004b). Milk and milk products e Determination of nitrate and nitrite contents e
Part 2: Method using segmented flow analysis (routine method). ISO 14673-2:
4. Conclusions 2004, IDF 189-1. Brussels, Belgium: International Dairy Federation.
IDF. (2004c). Milk and milk products e Determination of nitrate and nitrite contents e
Part 3: Method using cadmium reduction and flow injection analysis with in-line
The manual method for determination of nitrate and nitrite in dialysis (routine method). ISO 14673-3:2004, IDF 189-1. Brussels, Belgium: In-
milk and milk powders has been developed with sufficient benefit ternational Dairy Federation.
to replace traditional methods that require the perfusion of ex- Indyk, H. E., & Woollard, D. C. (2011). Nitrates and nitrites as contaminants. In (2nd
ed.)., Encyclopedia of dairy sciences (Vol. 1; pp. 906e991) San Diego, CA, USA:
tracts through cadmium columns. Nitrate is detected colorimet- Academic Press.
rically with Griess reagent following reduction to nitrite with Jedlickova, V., Paluch, Z., & Alusik, S. (2002). Determination of nitrate and nitrite by
trivalent vanadium in solution. The method avoids lengthy high-performance liquid chromatography in human plasma. Journal of Chro-
matography B, 851, 51e70.
manipulative procedures involving excess glassware and demon-
Jobgen, W. S., Jobgen, S. C., Li, H., Meininger, C. J., & Wu, G. (2007). Analysis of nitrite
strates figures of merit that are fit-for-purpose in dairy products. and nitrate in biological samples using high-performance liquid chromatog-
The recommended reduction conditions are 60 C for 30 min to raphy. Journal of Chromatography B, 851, 71e82.
meet high turnaround speeds. The method relies on reaching a Lookabaugh, M., & Krull, I. S. (1988). Determination of nitrite and nitrate by
reversed-phase high-performance liquid chromatography using on-line post-
constant absorbance end-point but suits later automation using column photolysis with ultraviolet absorbance and electrochemical detection.
kinetic measurements. Journal of Chromatography, 452, 295e308.
94 D.C. Woollard, H.E. Indyk / International Dairy Journal 35 (2014) 88e94
Merino, L., Edberg, U., Fuchs, G., & Aman, P. (2000). Liquid chromatographic Sun, J., Zhang, X., Broderick, M., & Fein, H. (2003). Measurement of nitric oxide
determination of residual nitrite/nitrate in foods: NMKL collaborative study. production in biological systems by using Griess reaction assay. Sensors, 3, 276e
Journal of AOAC International, 83, 365e375. 284.
Miranda, K. M., Espey, M. G., & Wink, D. A. (2001). A rapid, simple spectrophotometric Tsikas, D. (2007). Analysis of nitrite and nitrate in biological fluids by assays
method for simultaneous detection of nitrate and nitrite. Nitric Oxide, 5, 62e71. based on the Griess reaction: appraisal of the Griess reaction in the l-
Moorcroft, M. J., Davis, J., & Compton, R. G. (2001). Detection and determination of arginine/nitric oxide area of research. Journal of Chromatography B, 851,
nitrate and nitrite: a review. Talanta, 54, 785e803. 51e70.
Reece, P., & Hird, H. (2000). Modification of the ion exchange HPLC procedure for Woollard, D. C., & Forrest, L. J. (1980). Levels of nitrate and nitrite in casein products
the detection of nitrate and nitrite in dairy products. Food Additives and Con- using an automated procedure. New Zealand Journal of Dairy Science and Tech-
taminants, 17, 219e222. nology, 15, 83e90.