Diagnostic Software
Note: LUNAR Diagnostic service software options are not available to the customer. This
software is available to LUNAR authorized representatives only.
The Service software gives the Technician access to all areas of the enCORE
Software, DO NOT adjust settings.
39
Printed Documentation
Stability Run
Used for the detection of arcing and testing the detectors performance
Signal Monitor
View count rates for low and high-energy channels for the detector.
Primary Calibration
Not used on DPX Bravo/ DPX Duo
Secondary Verification
Used in the manufacturing process
Pileup
Used in the manufacturing process
Detector Tests
Not Used
Training
Used to enter applications training data
Install Scanner
An built-in function in the software, which has to completed when the scanner is installed.
Error Log
Ability to view and print the contents of the diagnostic error log (see section 3.4).
Statistics Log
This provides a statistics on the system for the number and type of scans done, the X-ray on
time, the last PM done.
Send Configuration
Can be used to send the error log, patient database and QA database to email or to a disk.
Address Book
The address book is linked to the Microsoft Outlook address book.
List Manager
Used to maintain a list of patients and physicians in the software
User Options
Set up user configured options for the scanner.
40
System_Software
Service Options
Service Mode scanner configurations
Connectivity options
Used to set up the options for Fax, email, Dicom, HL7, Report Delivery and Dexter
Service History
Information on the service performed on the system. A detail of each service visit has to be
entered in the service history option.
41
Printed Documentation
Calibration Pileup
Not used for service.
Spillover
Perform and Acquire a Spillover Measurement (test of source spectrum and/ or detector
resolution).
Count Rate
Acquire and print Reference Counts (count rate of detector must be performed through lucite).
Scanner Motion
Control manually and perform scanner motion diagnostics and motion configuration.
• Motion Tests Tab
o Limit Switch Monitor - used to verify limit switch status i.e. Engaged (Switch
closed) or Disengaged (Switch open)
o Limit to Limit Test - used to verify limit switch positions
o Swing Arm Switch Monitor (DPX Bravo/ DPX Duo Only)- used to test the 2 limit
switches for the swing arm
o Scan tests- to test motion for various scan types
o Configure check box- this box must be checked when limit to limit is run to
establish the scanners range of motion (scan window).
o Home scanner - moves mechanics to Home position
Note: When the Home position is changed from the head end of the table to the foot or
vice versa, the Limit-to-Limit test (see Motion Tests tab above) must be run with the
Configure box selected.
o Move Steps - moves the scan arm to a specific location on the table
o Move Absolute moves an absolute distance from home
o Move relative - move relative to last position
o Joystick and Laser - Enable and Control Scanner Motion manually, change motor
speeds, toggle laser on and off
Lower portion of the screen displays the status of limit switches; scan arm position, and
positioning switch (joystick) status.
• Find Block Tab
Secondary Calibration - can be used to find / verify that the scanner is able to find the
different chambers on the secondary calibration (daily QA) block. This tool can be useful
for troubleshooting “cannot find Block errors” and for Transverse binding issues.
Scanner X-Ray
• X-ray Tests Tab
o Manual Test - Control Manually the Shutter, X-ray on Indicator and patient
Positioning laser
42
System_Software
o Automatic test – this will automatically toggle the Shutter, Laser and X-Ray Relay
o X-Ray Source - Turn on source (ramp and make x-rays), warm up a tube head at
install or replacement.
o X-ray cycle test - cyclically ramp the source - (can be used to test ramping
failures or for arcing)
o Warm Up - Button to start tube head warm up.
o X-rays Tab
Ramp the kV and / or the mA of the x-ray generation system and view feedback real-time.
Faults are displayed and polled real-time in the window on the right.
• Ports Tab
View status of and send commands to cSBC ports.
• Download Tab
Verify the firmware version in the cSBC. Can also be used to download the firmware
version on the cSBC.
Scanner Detector
• Detector Tests Tab
Gain Control - adjusts the gain on the AGS amplifier used in the detector peak setting -
peak adjusts only the AGS amplifier.
Bias Control - adjusts the bias of the detector - bias peak (For DPX Bravo/ DPX Duo
Only) adjusts both the detector bias and the AGS amplifier gain.
Note (For DPX Bravo/ DPX Duo Only): The bias and peak setting of the detector work
together to maximize the number of counts - the bias peak adjusts both the bias on the
detector and the AGS amplifier peak, peak adjusts only the AGS amplifier peak. When
the AGS peak falls out of range the enCORE software will automatically perform a bias
peak. The operator will not see any difference in operation, however, the Daily QA may
take about 40 seconds longer.
DCA window settings - view DCA window settings - these settings should not be edited.
Detector status - allows the service engineer to view the detector status real time -
information displayed includes Peak setting Bias setting, and DCA windows.
• Detector AGS Tab
AGS DAC control (verify AGS DAC settings), check on detector rollover, and check on
channel count rates.
DO NOT adjust the detector Bias, DCA windows or AGS DAC settings -
changes in the Detector bias, DCA windows or AGS DAC’s will change
the scanner’s results.
AGS Mode - the operate / calibrate signal (disables the AGS systems so that the detector
can be peaked correctly) can be toggled.
Lin/Rep
Not Used for Service
Scanner Disconnect
Disconnect the scanner from the DPX Bravo/ DPX Duo software (reset the cSBC).
43
Printed Documentation
Performed under F2 Measure - only visible when the scanner is in service mode.
44
System_Software
Error Log
The enCORE software will maintain an error log file whenever the scan table is in operation. All
errors, fatal or nonfatal will be logged here. The Error log gives a brief description of the error and
if possible the origin of the fault. The error log is one of the best troubleshooting tools for scanner
malfunction.
Note: To aid in a quick and accurate response when requesting assistance from GE-LUNAR,
please have an error log available for review by GE-LUNAR Customer Support.
The error log file is named errorlog.bin, this file is the current error log session. Older sessions
are named errorlog.1, errorlog.2 and so on up to errorlog.4. When requesting help from GE-
LUNAR, E-mailing the current error log file and a description of the symptoms will aid in a rapid
diagnosis.