Professional Documents
Culture Documents
While the vast majority work safely, infusion pumps have been involved
in a number of safety incidents. In the UK alone, at least 1,000 incidents
have been investigated by the Medicines and Healthcare Products
Regulatory Agency (MHRA) between 2005 and 2010. The majority of
these relate to over infusion of drugs due primarily to user error
involving dosage or patient data. Some incidents have been caused by
product design and engineering or software malfunction.1,2 Therefore, it
is important that those with responsibility for medical device safety and
performance arrange for IV devices to be regularly tested and evaluated
to ensure that they are functioning to the manufacturer’s specification
and within clinical and environmental expectations.
Flow Measuring Principles
The primary aim of testing is to verify that the device delivers the
required flow rate, volume, and bolus accurately; that occlusion alarms
are activated when necessary; and that the device is safe for patient and
operator use. It’s important that testing conditions mirror real-life
settings and that they reflect what the manufacturer recommends to
ensure that the equipment is working within its specification.
Testing Methods
Optimal infusion is the ability of a device to reliably deliver the
prescribed dosage and volume to the patient, at a pressure that
overcomes all baseline and intermittent resistance, while causing no
harm to the patient.3 The reliability of infusion pumps is extremely
important because these devices are used for patients who could be in a
critical condition. Moreover, the safety incidents associated with
infusion devices demonstrate that there is a need to adequately validate
the accuracy and performance of these devices.4
There are several methods for testing reliability. Users are increasingly
moving away from basic measurements such as weighing scales and
burettes, where continuous user input is required to ensure accuracy,
toward automatic flow analyzers, which record real-time results
graphically and comply with both the IEC 60601-2- 24 standard and
manufacturer’s specifications, as outlined in service manuals.
Furthermore, it is important to ensure that the accuracy of an infusion
system is looked at in its entirety, taking into account all possible
inaccuracies. For example, the syringe used and other external
equipment, including the tubing set, could increase the inaccuracy to
10%. It is therefore generally accepted that the testing method and
equipment have to be more accurate than the equipment under test.2
Common methods for measuring volume or flow rate are as follows: