SERVICE INFO

To Reference
All service managers MvH/

From Info number


Service department SEL/021 2 pages

Date
13 April 1995

Subject
Monitor 6001-634 / Waste tube 6001-990

MONITOR 6001-634.

Date of implementation:
SELECTRA 1 instruments delivered in may 1995 and later.

The SELECTRA 1 monitor will be equipped with a new type of


powersupply. The line powercable connection of the new powersupply
is made by a solderconnection and not by a clamp.
To prevent problems when exchanging a monitor unit we included the
cables in the partnumber.

6001-634 Monitor assy (without covers)


including power-, videosignal- and brightnescontrolcable

Pag.-1-
SERVICE INFO

WASTE TUBING 6001-990

Date of implementing:
SELECTRA 1 instruments delivered in May 1995 and later.

Due to field information from country's like Brazil and France we


made a modification on the waste tubing, connecting V3 (the
collecting cask) with the rear panel.
In some instruments an obstruction was found by means of a deformed
tubing. For the future we offer you a tube assy 6001-990 consisting
of a metal pipe and a hard plastic 90' connector.

Pag.-1-
SERVICE INFO

The first 5 tube assy's will be send to you free of charge as soon
as they are available. In case more assy's are wanted they can be
ordered:

6001-990 tube assy V3/backpanel DFL. 15,-

With kind regards,

Marcel van Heertum


Service Manager