You are on page 1of 8

0099-2399/85/1106-0249/$02.

00/0
JOURNAL OF ENDODONTICS Printed in U.S.A.
COpyright 9 1985 by The American Association of Endodontists VOL. 11, No. 6, JUNE1985

An In Vitro Comparison of the Efficacy of the Step-


back Technique Versus a Step-back/Ultrasonic
Technique in Human Mandibular Molars
Comparacion In Vitro de la Eficacia de la Tecnica
Escalonada Versus la Tecnica Escalonada
Ultrasonica en Molares Inferiores de Humanos
Alan Goodman, DDS,MS, AI Reader, DDS,MS, Mike Beck, DDS,MA, Rudy Melfi, DDS,PhD,and
William Meyers, DMD,MEd

This study compared the effect of the step-back escalonada por ultrasonido. No se demostraron di-
preparation versus a step-back/ultrasound prepa- ferencias significativas en la capacidad de limpieza
ration on the tissue removal from the mesial root de los itstmos cuando se comparo el grupo de
canals of 60 extracted human mandibular molars. tecnica escalonada con el grupo de control, tanto
rhe step-back/ultrasound preparation significantly en el nivel de 1 mm como en el de 3 mm. En el nivel
cleaned isthmuses at both levels and canals at the de 1 mm, el operador 1 limpio los conductos en el
1-mm level more effectively than the step-back grupo de tecnica escalonada en forma significati-
preparation. Statistical analysis indicated no signif- vamente mas efectiva que el operador 2. Sin em-
icant difference, in canal cleanliness at the 3-mm bargo, en el grupo de tecnica escalonada por ultra-
level, between the step-back and the step-back/ sonido, no se demostraron diferencias de importan-
ultrasound groups. There were no significant differ- cia entre los operadores en su capacidad de limpiar
ences demonstrated in isthmus cleaning ability los conductos a este nivel. En ambos grupos exper-
when comparing the step-back group with the con- imentales, las comparaciones del operador no indi-
trol group, at both the 1- and 3-mm levels. At the 1- caron diferencias significativas en la limpieza de los
mm level, operator 1 significantly cleaned canals in istmos en un nivel de 1 mm.
the step-back group more effectively than operator
2. However, in the step-back/ultrasound group, no
significant differences were demonstrated between
operators in their ability to clean canals at this level. Complete tissue removal from the root canal system is
In both experimental groups, operator comparisons one of the major goals of successful endodontic therapy
indicated no significant differences in isthmus (1,2). Proper debridement and shaping are necessary
cleanliness at the 1-mm level. for microbial control and adequate obturation (3, 4).
However, studies (5-7) of root canal anatomy have
Este estudio comparo el efecto de la preparacion shown the complexity of this system, which makes
escalonada versus la preparacion escalonada por complete debridement by mechanical means difficult.
ultrasonido en la remoci6n de tejido del conducto The step-back (serialization, flaring) canal preparation
de la raiz mesial de 60 molares humanos inferiores technique has been demonstrated to be the most ef-
extraidos. La preparacion escalonada por ultrason- fective debridement technique currently in use (8-10).
ido limpi6 los istmos en ambos niveles significati- However, tissue and debris were found remaining
vamente y los conductos a un nivel de 1 mm mas within the canal system even when utilizing this tech-
efectivamente que la preparacion escalonada. El nique (8-10). There appears to be a definite need for
analisis estadistico no indico diferencias significa- new techniques and materials to be developed and
tivas en la limpieza del conducto a un nivel de 3 mm investigated in an attempt to debride the root canal
entre los grupos de t(~cnica escalonada y tecnica system more completely.
249
250 Goodman et al. Journal of Endodontics

Richman (11), in 1957, first reported on a technique with the exceptions of working length determination,
of root canal therapy using an ultrasonic instrument, were performed with the teeth stabilized in a modified,
which he felt had important potentials for the future. In bench-mounted cast metal vice. Customized rubber
a series of studies, Martin and his colleagues (12-19) pads were glued to the inner surface of the vice's
reported on the use of ultrasound as an intracanal holding jaws to aid in stabilization and to prevent tooth
cleaning technique. They found that the root canals of movement during the experimental procedure. The ex-
the teeth that were ultrasonically filed and irrigated, perimental teeth were randomly divided into two equal
when compared with conventional methods, were sig- groups, for preparation by two operators, in order to
nificantly cleaner and the smear layer appeared to be evaluate any differences in operator technique. Follow-
greatly reduced. Cameron (20) found that ultrasonics ing working length determination, the entire filing pro-
was effective in removing the smear layer in biome- cedure was performed utilizing a constant flow irriga-
chanically prepared teeth. Tauber et al. (21) found no tion/aspiration system. A 2.62% sodium hypochlorite
statistical difference between canals prepared with ul- irrigating solution was placed in an elevated 2000-ml
trasonics, hand filing, or ultrasonics with hand filing. pyrex flask, to which an angled 18-gauge, 11/2-inch
Cymerman et al. (22) reported no difference in debride- needle was connected via an intravenous extension set
ment between hand instrumentation and ultrasonic in- (Travenol Labs., Inc., Deerfield, IL). The irrigation needle
strumentation. Weller et al. (23) quantitatively com- was positioned next to a high-speed suction tip, which
pared the debridement efficacy of a hand-instrumented provided a constant flow irrigation/aspiration system.
serialization preparation, ultrasonic preparation, and a The gravity-propelled flow of irrigating solution was
hand-instrumented serialization preparation followed by calibrated, using the flo-trol clamp, to a rate of 40 ml/
ultrasonics. They found that the most effective debride- rain.
ment occurred when ultrasonication was used after A number 15 K-type file (Star Dental Mfg., Co., Valley
completion of hand instrumentation. These authors Forge, PA) was placed into the canal until the file tip
concluded that ultrasonication was not an alternative was just visible at the apical foramen. This length was
to conventional hand instrumentation, but it was a measured with an endodontic ruler and used as the
significant aid in increasing the efficiency of endodontic canal working length.
debridement. Step-back preparation (8-10), using an alternating
Therefore, the purpose of this study was to compare, technique, was performed in the following sequence:
histologically, the efficacy of the step-back preparation #15, 20, 25, and 30 (25 and 30 files were short of
versus a combined step-back/ultrasonic preparation in working length); # 2 and 3 Gates Glidden. Alternating
the mesial root canals of extracted human mandibular sequence: #15, 35 (short of working length), 20, 40
first and second molars. (short), 25, 45 (short), 30, 50 (short); #5 Gates Glidden.
Step-back sequence: #35, 40, 45, 50. Final file (30F).
MATERIALS AND METHODS Each mesial root canal was enlarged approximately
three file sizes larger than the first file that bound in the
Freshly extracted, human first and second mandibu- apical third of the canal. This corresponded to a #25
lar molars were collected and stored at 0~ until ready or 30 size file. Individual operator time of canal prepa-
for use. The age, sex, pulpal status, or reason for ration was recorded for evaluation. Following comple-
extraction was not recorded. Radiographs were taken tion of canal debridement and shaping, the irrigation/,
of all teeth from the buccal direction and the degree of aspiration system remained functioning for an additional
curvature of each mesial root was determined using period of 3 min/canal. After drying, each canal was
Schneider's method (24). In order to standardize canal irrigated with 10% formalin to assure optimum fixation
curvature, only those roots exhibiting a 15 to 35 degree of any remaining pulpal tissue.
of curvature were used in this study. Standard root The teeth in the step-back/ultrasonic group were
canal access openings were made using a # 4 round treated like those teeth in the step-back preparation
bur, The presence or absence of coronal pulp tissue group except that an ultrasonically energized file was
was noted and recorded. Those teeth exhibiting no used following completion of the step-back preparation.
pulp tissue were discarded from the study. The 60 The ultrasonic power source was Buffalo model S-40
remaining teeth were randomly divided into three Piezo Electric ultrasonic dental unit (Buffalo Dental
groups of 20 teeth,, with both mesial canals of each Mfg., Brooklyn, NY). This unit utilized a piezoelectric
root being treated alike according to the group. For ceramic rather than a magnetostrictive transducer (26-
evaluation and standardization purposes, only those 28). The unit was operated at an automatically adjusted
mesial roots demonstrating a type 111 (25) canal con- frequency of 40,000 cycles/s. Since almost no heat
figuration were used in the study. This was verified, was generated, the unit could be operated without
during working length determination, when both file tips using a coolant. A manually operated power control
were seen just protruding from two separate apical knob, corresponding to a tip amplitude displacement
foramens. All procedures involving canal preparation, range of 0.001 to 0.004 inches, was set at the one-half
Vol. 11, No. 6, June 1985 Step-beck Versus Step-back/Ultrasound 251

power setting and maintained at this position through- tions from each level were chosen blindly by an expe-
out the experiment. To minimize cutting ability, a #15 rienced histologist. One section was stained with he-
tapered, finger plugger was used as the ultrasonic tip. matoxylin and eosin and the other with Masson stain.
The T-4 ultrasonic scaling tip (Buffalo Dental Mfg.) was A total of 120 sections were used in this study for
modified by sectioning the triple-bended tip at its middle evaluation. All mounted sections were projected on
bend, which produced a shortened unibend tip. A white tracing paper using the Kramer Model XM Xenon
groove was placed in the outer surface of the modified Microprojector (Kramer Scientific Corp., NY, NY) at a
tip to a depth just short of and running parallel to the magnification of • The slides were coded and
internal water channel. The plastic handle of the finger randomly projected so the evaluator was unaware of
plugger was removed, leaving an additional 5-mm the corresponding level or group. A black tracing pen
shank for soldering. The plugger was then positioned was used to outline each canal and isthmus wall. Pulpal
within the groove and soldered to the ultrasonic tip tissue debris was outlined with a red tracing pen. When
assembly using silver solder and the Torit Model 21-A isthmuses were encountered, a blue tracing pen was
electric soldering machine (Torit Mfg. Co., St. Paul, used to delineate the canal-isthmus interface. A com-
MN). After completion of the step-back preparation, the pensating polar planimeter, model 620005 (Keuffel &
ultrasonic file was used in an up and down, circumfer- Esser Co., Morristown, NJ), was used to determine the
ential, brush-stroke motion. Prior to activation, the ul- area of each canal, isthmus, and remaining pulpal tis-
trasonic tip was placed into the canal until binding sue. The values were then calculated into percentage
occurred or working length was reached. This depth of tissue removed. The collected raw data were ana-
was recorded on the instrument shaft with the use of lyzed nonparametrically utilizing the Kruskal-Wallis one-
a silicone stop. In canals in which binding took place, way analysis of variance. A posteriori testing was done
the stop was adjusted 1 mm short of this length which using a nonparametric analogue to the Scheffe test.
indicated the maximum depth of instrument insertion Differences between operators were analyzed nonpar-
without binding in the canal. During a pilot study, the ametrically using the Mann-Whitney U test. Nonpara-
silicone stop remained fixed in place during activation metric tests were used because of large differences in
of the ultrasonic instrument. However, ultrasonic acti- the variances among most of the groups analyzed.
vation frequently prevented insertion of the instrument
to the previously established depth of the inactivated RESULTS
file. Therefore, after ultrasonic preparation, the rubber
stop was repositioned to the actual depth following The mean values of percentage clean (canal and
usage, remeasured, and recorded. The ultrasonic file isthmus) for each instrumentation technique, at each
was used for a period of 3 min in each canal and at no level, are presented in Table 1. The comparisons be-
time was the instrument used at a depth beyond the tween the canal groups are presented in Table 2. At
established working length. Irrigation was identical to the 1-mm level, the step-back/ultrasound group (mean
that used in the previous step-back group. After drying, value 97.3%) showed significantly greater canal clean-
each canal was irrigated with 10% formalin. liness than either the step-back group (mean value
The control teeth were opened and the presence of 91.3%) or control group (mean value 18.7%, p < 0.05).
pulp tissue was confirmed. These teeth were not in- There were also significant differences demonstrated
strumented and served as a control for the processing between the step-back group (mean value 91.3%) and
and sectioning procedures. the control group (mean value 18.7%, p < 0.05). At the
At the conclusion of preparation, a single vertical 3-mm level, the step-back group (mean value 98.9%)
groove was placed in the mesial buccal root surface to and the step-back/ultrasound group (mean value
a depth of 0.5 mm. This groove was used as an aid in 99.9%) showed significantly greater canal cleanliness
determining canal identity during the processing and than the control group (mean value 27.6%, p < 0.05).
evaluation procedures. Each tooth was placed into At this level, there were no significant differences be-
COvered glass jars containing 10% formalin for a mini-
mum period of 48 h. TAaLE 1. Summary of results for the three groups* at the 1- and
3-ram levels
All teeth were decalcified and prepared for histologi-
cal sectioning. Seven-micrometer sections were ob- Level No. of
Average % No. of Average %
tained using the Reichert-Jung 1140/Autocut micro- Group (mm) Canals Clean/
Canal Isthmuses Clean/
Isthmus
tome (R-Jung, Heidelberg, West Germany) equipped
With a carborundium knife. All initial sections were 1 1 40 91.33 9 21.67
1 3 40 98.98 10 37.26
COllected and microscopically examined until verification 2 1 40 97.29 8 72.03
of the apical foramen occurred. Subsequent sections 2 3 40 99.97 10 92.03
Were numbered, with sections from the 1-mm (143rd 3 1 30 18.69 4 19.85
Section: _+4) and 3-mm (429th section: _+4) levels col- 3 3 31 27.61 7 3.41
lected for evaluation. Two technically error-free sec- * Group1, step-back;group2, step-back/ultrasound;group3, contr..
252 Goodman et al, Journal of EndodonticB

TABLE 2. Comparisons between the canal groups* at the 1- and cantly greater canal cleanliness than operator two (me-
3-mm levels dian value 93.2%, p < 0.05). However, in the step-
Level Kruskal-Wallis df p Group back/ultrasound group, at this level, there were no
(mm) Statistic Comparison P significant differences in canal cleanliness demonstra-
1 79.59 2 0.000 1 versus 2 <0.05 ted between operator one (median value 100.0%) and
1 versus 3 <0.05 operator two (median value 100.0%).
2 versus 3 <0.05
Isthmus cleanliness, in the step-back group, at the
3 84.70 2 0.000 1 versus 2 NS
1 versus 3 <0.05
1-mm level, showed no significant differences between
2 versus 3 <0.05 operator 1 (median value 19.8%) and operator 2 (me-
9Group1, step-back;group2, step-back/ultrasound;group3, control.
dian value 16.3%). Isthmus cleaniness, in the step-
back/ultrasound group, at this level, showed no signif-
TABLE 3. Comparisons between isthmus groups* at the 1- and
icant differences between operator 1 (median value
3-mm levels 86.1%) and operator 2 (median value 67.7%).
Level Kruskal-Wallis Group
(mm) Statistic df p Comparison P DISCUSSION
1 10.36 2 0,000 1 versus 2 <0.05 The results of this study showed that canals prepared
1 versus 3 NS
with the step-back/ultrasonic technique were signifi-
2 versus 3 <0.05
3 19.88 2 0.000 1 versus 2 <0.05 cantly cleaner than those in the step-back group. This
1 versus 3 NS finding can be attributed to the action of the ultrasoni-
2 versus 3 <0.05 cally activated irrigated solution within the canal sys-
* Group1, step-back;group2, step-back/ultrasound;group3, control. tem. When a chemical containing fluid medium, such
as sodium hypochlorite, is subjected to ultrasonic vibra-
TABLE 4. Operator mean and range of canal preparation time
tions, which produces cavitation, chemical activity is
utilizing only hand instruments* accelerated as a result of both mechanical and thermal
activity in the fluid (27). Most power ultrasonic proc-
Operator 1 Operator 2
Group (rain) (min) esses, involving liquids, utilize the phenomenon of cav-
itation as the action producing mechanism. This refers
1
Mean 11.3 13.6
to the growth and collapse of small bubbles in a liquid
Range 10-13 10-15 under the cyclic phases of rarefaction--compression
that occur during ultrasonic vibration. In any ordinary
Mean 9.9 13.2 liquid, bubble nuclei exist in the form of very small gas
Range 7-14 11-16 bubbles, dust particles, or other impurities. For given
* Group1, step-back;group2, step-back/ultrasound. conditions of frequency and alternating pressure level,
bubble nuclei of a critical radius will grow during the
tween the step-back group (mean value 98.9%) and rarefaction phase. During the compression phase, the
the step-back/ultrasound group (mean value 99.9%). bubble undergoes catastrophic collapse in a small frac-
The comparisons between the isthmus groups are tion of a vibration cycle. Extremely high pressures are
presented in Table 3. At the 1-mm level, the step-back/ thus instanteously developed and result in outward
ultrasound group (mean value 72.0%) showed signifi- propagating shock waves (28). Ewen (27) states that
cantly greater isthmus cleanliness than either the step- cavitation has a complex combination of characteris-
back group (mean value 21.7%) or control group (mean tics-mechanical, thermal, and chemical. Cavitation is
value 19.9%, p < 0.05). However, there were no sig- accompanied by sharp, but localized, increases in tem-
nificant differences between the step-back group (mean perature around the vapor bubbles. However, the over-
value 21.7%) and the control group (mean value all temperature of the fluid median is usually not sub-
19.9%). At the 3-mm ' level, the step-back/ultrasound stantially raised by this effect because of the relatively
group (mean value 92.0%) showed significantly greater
TABLE 5. Interoperator comparisons of canal and isthmuses
isthmus cleanliness than either the step-back group cleanliness at the l-ram level*
(mean value 37.3%) or control group (mean value 3.4%,
p < 0.05). However, there were no significant differ- Median % Clean:
Group Operator 1 U
ences between the step-back group (mean value versus Statistic P
37.3%) and the control group (mean value 3.4%). Operator 2
The mean and range of time comparisons between 1 (canal) 99.05 versus 93.20 124.5 <0.05
operators, in each experimental group, are presented 2 (canal) 100.00 versus 100.00 170.0 NS
in Table 4. Operator comparisons, at the 1-mm level, 1 (isthmus) 19.75 versus 16.30 7.0 NS
are presented in Table 5. In the step-back group, 2 (isthmus) 86.10 versus 67.70 6.5 NS
operator 'one (median value 99.0%) showed signifi- * Group1, step-back;group2, step-back/ultrasound.
Vol. 11, No. 6, June 1985 Step-back Versus Step-back/Ultrasound 253

low ratio of vapor bubbles to total fluid. Graft (28)


described other physical effects associated with ultra-
sound, in liquids, which are generally of lesser impor-
tance than cavitation but still play important roles in
ultrasonic technology. One such phenomenon is acous-
tic streaming, in which intense, local circulation and
vortex flow fields are observed near the vibrating trans-
ducer face. Such flow fields result from nonlinear fluid
behavior and can enhance the rates of heat and mass
transfer in the ultrasonic process.
In a pilot study, using ultrasonically energized K-type
files and diamond-coated files with a noncutting tip
(Abrasive Technology-Medical Inc., Columbus, OH), un-
controllable cutting and instrument fracturing were en-
countered. Cameron (29) reported using Hedstrom files
FiG 2. Photomicrograph of cross-section of the mesial root at the l-
in a Cavitron ultrasonic insert. He found instrument
mm level, group 2 (step-back/ultrasonic). Mesiobuccal canal (/eft),
breakage was a problem, especially if the file was bent mesiolingual canal (right), and connecting isthmus demonstrating
to gain access to a posterior tooth. Therefore, a #15 100% cleanliness. Masson stain; original magnification x13.
ultrasonically energized finger plugger was used follow-
ing completion of the serial preparation with conven-
tional hand and rotary instruments, not for preparing or cleaned as the step-back/ultrasonic group (Tables 1
shaping the canal, but for its ultrasonic activation of the and 2 and Figs. 1 and 2). This finding agrees with the
irrigating solution. The step-back preparation in the results reported by Weller et al. (23) in which ultrasonic
step-back/ultrasonic group was utilized prior to ultra- cleaning, following serial preparation, was significantly
sonic cleaning to enhance debridement (23), to permit more effective in removing radioactive-laden gelatin
a greater volume of fresh irrigating solution in contact than the serial technique alone.
with the canal system (30, 31), and to make the canal At the 1-mm level, the canals prepared with the step-
large enough for adequate obturation with gutta-percha back technique demonstrated an average of 91.33%
(4). cleanliness per canal (Table 1). This result appears to
Baker et al. (30) reported that the removal of debris be greater than those reported in previous studies (8,
and microorganisms from the root canal system 9), possibly due to the differences in the irrigating
seemed to be a function of the quantity of irrigating solution and volume used. However, due to differences
solution rather than the type of solution used. The in evaluation and statistical techniques, a true compar-
flushing action of the solutions, and not their tissue ison is not possible.
dissolving qualities, appeared to be the significant fac- At the 3-mm level, there were no significant differ-
tor. In the present study, even though a maximum ences in canal cleanliness found between the step-back
volume of fresh irrigating solution was utilized, the step- and the step-back/ultrasonic groups (Table 2 and Figs.
back group, at the 1-mm level, was not as effectively 3 and 4). Previous studies (8, 9) comparing the step-
back technique, reported canal cleanliness was en-
hanced at each level coronal to the root apex. This can
be attributed to a straighter canal found at the more
coronal levels of the root, resulting in more planed walls
(10). This would also yield a larger preparation with an
increase in volume of fresh irrigating solution contacting
the canal walls during instrumentation (30, 31). Cun-
ningham et al. (15) found that teeth prepared entirely
by ultrasonic instrumentation, at the 3-mm level, were
significantly cleaner than those prepared by a conven-
tional hand-filing technique. However, they used a con-
stant flow, high volume, irrigation/aspiration system in
conjunction with those teeth prepared with ultrasound,
whereas teeth prepared by conventional hand filing
were irrigated intermittently during canal preparation.
To eliminate the variable in canal cleanliness due to
FIG 1. Photomicrograph of cross-section of mesiobuccal canal and
isthmus at the 1-mm level, group 1 (step-back). Mesiobuccal canal
unequal volumes of irrigating solution used, all experi-
demonstrating 68.9% cleanliness, with the isthmus 2.0% cleaned. mental teeth were prepared utilizing a high volume,
Masson stain; original magnification x31. constant flow irrigation/aspiration system in this study.
254 Goodman et al. Journal of Endodontics

the results reported by previous authors (8, 9), in which


isthmuses were determined to be inadequately cleaned,
even when prepared with the step-back technique.
Even though the step-back/ultrasonic preparation
was found to clean isthmuses and canals (1-mm level)
significantly better than the step-back preparation, tis-
sue debris was still found to some degree within these
teeth (Fig. 4). Therefore, none of the canal preparation
techniques used in this study were found to clean
consistently the canal system completely. A factor
which we believed might have accounted for this incom-
plete tissue removal was the inability to insert the
activated ultrasonic tip to the full working length in all
teeth. Of the 40 experimental canals in the step-back/
ultrasonic group, the average ultrasonic tip depth of
FIG 3. Photomicrograph of cross-section of the mesiolingual canal
and isthmus at the 3-ram level, group 1 (step-back). Mesiolingual penetration was 2.7 mm short of the working length,
canal demonstrating 99.3% cleanliness, with the isthmus 13.5% with a range of 0.0 to 5.0 mm. When initially placing
cleaned. Masson stain; original magnification x20. the inactivated ultrasonic tip in the canal, a deeper
penetration was achieved, compared with a "backing-
out" or withdrawing motion encountered when the tip
was activated. Both operators in this study reported
any canal irregularities, or curvatures, that became
ledged (minor) with the ultrasonic instrument prevented
further apical penetration of the activated tip. This
inability of complete canal penetration by the activated
ultrasonic tip can be attributed to the inherent flexibility
and physical properties of the finger plugger ultrasonic
tip design, and the ultrasonic power unit amplitude
setting. However, we found that in 31 of 40 canals in
which the ultrasonic tip did not negotiate to the full
working length, complete tissue removal was shown at
the 1-mm level. As a result, complete penetration of
the ultrasonic tip to the established working length
might not be necessary to obtain ultrasonic cleaning of
FIG 4. Photomicrograph of cross-section of the mesiolingual canal the apical levels.
and isthmus at the 3-mm level, group 2 (step-back/ultrasonic). Me-
siolingual canal demonstrating 100% cleanliness, with the isthmus
82.5% cleaned. Hematoxylin and eosin; original magnification x20.

The step-back/ultrasonic preparation significantly


cleaned isthmuses better than the step-back prepara-
tion, at both the 1- and 3-mm levels (Table 3 and Figs.
1 to 4). Apparently, t~e ultrasonically activated irrigating
solution penetrated areas of the canal system and
removed the pulpal tissue which hand instrumentation
left untouched.
Even when using at least forty times the volume of
irrigating solution as reported in previous studies (8, 9),
we found no significant differences in isthmus cleanli-
ness, at either level, when comparing the step-back
group with the noninstrumented control group (Table
3). As a result, areas of this variable and complex FIG 5. Photomicrograph of cross-section of the mesial root at the 3-
system (5-7) which are not directly contacted by hand mm level, group 2 (step-back/ultrasonic). Mesiobuccal canal (right),
mesiolingual canal (/eft), and connecting isthmus demonstrating
instrumentation, regardless of the volume of irrigating 100% cleanliness. Note transposition of the mesiobuccal canal, re-
solution used, ap.pear to be unaltered by this method sulting in a near perforation of the root surface. Hematoxylin and
of preparation (Figs. 1 and 3). This finding corroborates eosin; original magnification x8.
Vol. 11, No. 6, J u n e 1985 Step-back Versus Step-back/Ultrasound 255

Even when using a smooth, noncutting, blunted tip SUMMARY


finger plugger as the ultrasonic instrument, cutting of
the canal wall was observed in 22 of the 40 canals. At This study compared the effect of the step-back
both the 1- and 3-mm levels, ultrasonic cutting resulted preparation versus a step-back/ultrasonic preparation
in transposition (21 of 40 canals) (Fig. 5) and perforation on tissue removal from the mesial root canals of 60
(1 of 40 canals) of the root surface, whereas 10 of 40 extracted, human mandibular molars.
canals demonstrated transposition in the step-back The teeth were randomly chosen and equally divided
group, with no perforations observed. It may be that into three groups. Twenty teeth were prepared with the
this instrument system is too powerful for safe clinical step-back technique utilizing K-type files and Gates
application. However, it did demonstrate remarkable Glidden drills (group 1). Group 2 contained 20 teeth
cleaning ability. Perhaps, further research with the com- prepared with the step-back technique followed by the
mercially available ultrasonic instruments can provide use of an ultrasonically energized, intracanal instru-
an answer to the question of complete tissue removal ment. Twenty teeth in group 3 were uninstrumented
versus safe clinical use. and served as controls for the processing and section-
Operator one (99.05%) significantly cleaned the ca- ing procedures. Both experimental groups were pre-
nals in the step-back group, at the 1-mm level, more pared utilizing an equal volume, constant flow (40 ml/
effectively than operator two (93.20%; Table 5). Even min), irrigation/aspiration system of 2.62% sodium hy-
though operator two consistently, in both groups, av- pochlorite solution. Following histological preparation,
eraged more time for canal preparation with hand in- sections were obtained from the apical 1- and 3-mm
struments (13.4 min), resulting in an increased irrigating levels and projected on a screen. Tracings were then
volume and possibly a greater degree of planed walls, made outlining each canal, isthmus, and any remaining
the canals at this level were not as clean as those of pulpal tissue. The percentage of tissue removed within
operator 1 (average time 10.6 min) (Table 4). This is in each canal and isthmus was measured and recorded
agreement with Littman (32) who found that the differ- using a compensating polar planimeter.
ences in canal debridement where more dependent on The step-back/ultrasonic preparation significantly
the operators than on the technique used. As shown in cleaned isthmuses at both levels and canals at the 1-
Table 5, there were no significant differences found mm level more effectively than the step-back prepara-
between the operator's ability to clean the canals in the tion. Statistical analysis indicated no significant differ-
step-back/ultrasonic group and the isthmuses in either ence, in canal cleanliness at the 3-mm level, between
the step-back/ultrasonic preparation or step-back tech- the step-back and the step-back/ultrasonic groups.
nique, at the 1-mm level. The ability of both operators There were no significant differences demonstrated in
to clean canals and isthmuses equally in the step-back/ isthmus cleaning ability when comparing the step-back
ultrasonic group could be due to the lack of individual group with the control group, at both the 1- and 3-ram
technique required to operate the ultrasonic instrument levels. At the 1-mm level, operator 1 significantly
effectively. Operation of the ultrasonic instrument is cleaned canals in the step-back group more effectively
more of a mechanical procedure, with the instrument than operator 2. However, in the step-back/ultrasonic
doing the actual work. However, a careful brushstroke group, no significant differences were demonstrated
technique is still required. Forcing the instrument or between operators in their ability to clean canals at this
inappropriate planing may result in perforation in serially level. In both experimental groups, operator compari-
prepared canals. Whereas in a hand-instrumented prep- sons indicated no significant differences in isthmus
aration, the operator's technique appears to be an cleanliness at the 1-mm level.
important factor in cleaning efficiency (32). The use of
an ultrasonic instrument, following serial preparation
with hand instrumentation, might compensate for the This study was supported by research funding from the Ohio Association
of Endodontists.
variable in operator technique, resulting in comparable
preparations. This article was adapted from a thesis submitted by Dr. Goodman in partial
It may be that complete tissue removal from the root fulfillment of the requirements for the MS degree at Ohio State University,
Columbus, OH. A portion of this article was presented at the 41st Annual
canal system will always be an enigma to the endodon- American Association of Endodontists meeting, Toronto, Canada, and was
tist, or further refinements in ultrasonic technology may honored by a graduate student research award.
some day lead to 100% cleaning. We do want to
Dr. Goodman is in private practice limited to endodontics, Atlanta, GA. Dr.
emphasize that the commercially available ultrasonic Reader is associate professor, Department of Endodontics, Ohio State Univer-
instruments, and the studies supporting their use, are sity, Columbus, OH. Dr. Beck is associate professor, Department of Oral
Diagnosis, Oral Medicine, Ohio State University. Dr. Melfi is professor and
not directly comparable to our results. Further studies assistant dean, Department of Oral Biology, Ohio State University. Dr. Meyers
of amplitude settings versus time (how long the instru- is professor and chairman, Department of Endodontics, Ohio State University.
Address requests for reprints to Dr. Reader, Department of Endodontics,
ment should be used) need to be done in order to find College of Dentistry, Ohio State University, 305 W. 12th Ave., Columbus, OH
the optimum cleaning ability of these instruments. 43210.
256 Goodman et al. Joumal of Endodontics
References Oral Surg 1982;53:527-31.
17. Martin H, Cunningham WT. The effect of endosonic and hand manipu,
1. Heuer MA. The biomechanics of endodontic therapy. Dent Clin North lation on the amount of root canal material extruded. Oral Surg 1982;53:611-
Am 1963;7:341-59. 3.
2. Cohen S, Burns RC. Pathways of the pulp. St. Louis: CV Mosby Co., 18. Martin H, Cunningham WT. An evaluation of postoperative pain inci-
1980:111-32. dence following endosonic and conventional root canal therapy. Oral Surg
3. Stewart GC. The importance of chemomechanical preparation of the 1982;54:74-6.
root canal. Oral Surg 1955;8:993-7. 19. Cunningham WT, Martin H, Pelleu GB, Stoops DE. A comparison of
4. Allison DA, Weber CR, Walton RE. The influence of the method of antimicrobial effectiveness of endosonic and hand root canal therapy. Oral Surg
canal preparation on the quality of apical and coronal obturation. J Endodon 1982;54:238-41.
1979;5:298-304. 20. Cameron JA. The use of ultrasonics in the removal of the smear layer:
5. Hess W. The anatomy of the root-canals of the teeth of the permanent a scanning electron microscope study. J Endodon 1983;9:289-92.
dentition--part 1. London: John Bale, Sons and Danielsson, 1925:3-49. 21. Tauber R, Morse DR, Sinai IA, Furst ML. A magnifying lens comparative
6. Skillen WG. Morphology of root canals. J Am Dent Assoc 1932;19:719- evaluation of conventional and ultrasonically energized filing. J Endodon
35. 1983;9:269-74.
7. Skidmore AE, Bjorndal AM. Root canal morphology of the human 22. Cymerman JJ, Jerome I.A, Moodnik RM. A scanning electron micro-
mandibular first molar. Oral Surg 1971 ;32:778-84. scope study comparing the efficacy of hand instrumentation with ultrasonic
8. Coffae KP, Brilliant JD. The effect of serial preparation versus nonserial instrumentation of the root canal. J Endodon 1983;9:327-31.
preparation on tissue removal in the root canals of extracted mandibular human 23. Weller RN, Brady JM, Bernier WE, Efficacy of ultrasonic cleaning, j
molars. J Endodon 1975;1:211-4. Endodon 1980;6:740-3.
9. Klayman SM, Brilliant JD. A comparison of the efficacy of serial prepa- 24. Schneider SW. A comparison of canal preparations in straight and
ration versus Giromatic preparation. J Endodon 1975;1:334-7. curved root canals. Oral Surg 1971 ;32:271-5.
10. Walton RE. Histologic evaluation of different methods of enlarging the 25. Weine FS. Endodontic therapy. St. Louis: CV Mosby Co., 1976:152-6.
pulp canal space. J Endodon 1976;2:304-11. 26. Peacocke AR, Pritchard NJ. Some biophysical aspects of ultrasound.
11. Richman MJ. The use of ultrasonics in root canal therapy and root Prog Biophys Mol Bio11968;18:187-208.
resection. J Dent Med 1957;12:12-8. 27. Ewen SJ. Ultrasonic therapy in periodontics. Springfield, IL: Charles C
12. Martin H. Ultrasonic disinfection of the root canal. Oral Surg Thomas, 1968:12-32.
1976;42:92-9. 28. Graft KF. Process applications of power ultrasonics--a review. IEEE
13. Martin H, Cunningham WT, Norris JP, Cotton WR. Ultrasonic versus Symposium on Sonics and ~trasonics, Milwaukee, Wl, 1974, 1-14.
hand filing of dentin: a quantitative study. Oral Surg 1980;49:79-81. 29. Cameron JA. The use of ultrasound in the cleaning of root canals: a
14. Martin H, Cunningham WT, Nords JP. A quantitative comparison of the clinical report. J Endedon 1982;8:472-4.
ability of diamond and K-type files to remove dentin. Oral Surg 1980;50:566- 30. Baker NA, Eleazer PD, Averbach RE, Seltzer S. Scanning electron
8. microscopic study of the efficacy of various irrigating solutions. J Endedon
15. Cunningham WT, Martin H, Forrest WR. Evaluation of root canal 1975;1:127-35.
debridement by the endosonic ultrasonic synergistic system. Oral Surg 31. Salzgeber RM, Brilliant JD. An in vivo evaluation of the penetration of
1982;53:401-4. an irrigating solution in root canals. J Endodon 1977;3:394-8.
16. Cunningham WT, Martin H. A scanning electron microscope evaluation 32. Littman SH. Evaluation of root canal debridement by use of a radio-
of root canal debridement with the endosonic ultrasonic synergistic system. paque medium. J Endodon 1977;3:135-8.

You might also like